Jasco V-670 absorption spectrometer

Similar documents
Instytut Fizyki Doświadczalnej Wydział Matematyki, Fizyki i Informatyki UNIWERSYTET GDAŃSKI

Protocol for Use of BCMP Jasco J-815 Circular Dichroism Spectropolarimeter

Experiment 13. Dilutions and Data Handling in a Spreadsheet rev 1/2013

Spectrometer User s Guide

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III)

UV/VIS SPECTROMETRIC DETERMINATION OF AN EQUILIBRIUM CONSTANT

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III)

Measuring ph with Smart Cuvettes

Ocean Optics Red Tide UV-VIS Spectrometer (Order Code: SPRT-UV-VIS)

Connect the Vernier spectrometer to your lap top computer and power the spectrometer if necessary. Start LoggerPro on your computer.

Space Objects. Section. When you finish this section, you should understand the following:

Lab 5: Calculating an equilibrium constant

Miniature Fiber Optic Spectrometry StellarNet Tutorial

MassHunter Software Overview

2. To measure the emission lines in the hydrogen, helium and possibly other elemental spectra, and compare these to know values.

ON SITE SYSTEMS Chemical Safety Assistant

BIOLIGHT STUDIO IN ROUTINE UV/VIS SPECTROSCOPY

DOAS measurements of Atmospheric Species

Astro 3 Lab Exercise

Spectroscopy II Introduction: Spectrophotometry and Fluorometry

Size Determination of Gold Nanoparticles using Mie Theory and Extinction Spectra

Watershed Modeling Orange County Hydrology Using GIS Data

Determining the Concentration of a Solution: Beer s Law

Photoluminescence Spectrometer (FLS980)

RATE LAW DETERMINATION OF CRYSTAL VIOLET HYDROXYLATION

UV-Vis optical fiber assisted spectroscopy in thin films and solutions

Raman Spectroscopy of Liquids

Background The power radiated by a black body of temperature T, is given by the Stefan-Boltzmann Law

Hydration of Propionaldehyde by Temperature-Jump Relaxation 1

Login -the operator screen should be in view when you first sit down at the spectrometer console:

Chemical Kinetics: Integrated Rate Laws. ** updated Procedure for Spec 200 use **

Cerno Application Note Extending the Limits of Mass Spectrometry

Kinetics of Crystal Violet Bleaching

You w i ll f ol l ow these st eps : Before opening files, the S c e n e panel is active.

Effect of mass attached to the spring: 1. Replace the small stopper with the large stopper. Repeat steps 3-9 for each spring set.

Understanding Your Spectra Module. Agilent OpenLAB CDS ChemStation Edition

Optical density measurements automatically corrected to a 1-cm pathlength with PathCheck Technology

Lab VI Light Emitting Diodes ECE 476

Ocean Optics Educational Spectrometers. Getting Started Manual

Using SkyTools to log Texas 45 list objects

ISIS/Draw "Quick Start"

Experiment 4 Radiation in the Visible Spectrum

Fourier Transform Infrared Spectroscopy (Perkin Elmer - Spectrum One)

UV-Visible Absorption Analysis of a Two Component System

Experiment: Oscillations of a Mass on a Spring

Determining the Concentration of a Solution: Beer s Law

Lab 6: Spectroscopy Due Monday, April 10

Shown below is a sample titration curve for a diprotic acid. Note the two equivalence points.

C101-E112. BioSpec-nano. Shimadzu Spectrophotometer for Life Science

1. Prepare the MALDI sample plate by spotting an angiotensin standard and the test sample(s).

Physical Chemistry II Laboratory

Absorption Spectra of Conjugated Molecules

EXPERIMENT 4 ONE DIMENSIONAL MOTION

How to Make Photometric & Colorimetric Measurements of Light Sources using an Ocean Optics Spectrometer and SpectraSuite Software

Beer's Law and Data Analysis *

Electric Fields and Equipotentials

v Prerequisite Tutorials GSSHA WMS Basics Watershed Delineation using DEMs and 2D Grid Generation Time minutes

Experiment 6a: Excited-state properties of 2-naphthol (the acidity constants) (Dated: October 29, 2009)

Quantum States and Spectra of Gases

North Carolina State University Department of Chemistry Varian NMR Training Manual

Software BioScout-Calibrator June 2013

REVIEW OF LAB TECHNIQUES

NMR Predictor. Introduction

OCEAN/ESS 410 Lab 4. Earthquake location

User Manual RGB Lasersysteme GmbH Qstick Spectrometer Version: Date: November 04, 2013

Fluorescence Spectrophotometry

Stoichiometric Reactor Simulation Robert P. Hesketh and Concetta LaMarca Chemical Engineering, Rowan University (Revised 4/8/09)

X-ray spectroscopy: Experimental studies of Moseley s law (K-line x-ray fluorescence) and x-ray material s composition determination

Spectrophotometer for Life Science. BioSpec-nano C101-E112D

EXPERIMENT 6 INTRODUCTION TO SPECTROSCOPY

The data for this lab comes from McDonald Forest. We will be working with spatial data representing the forest boundary, streams, roads, and stands.

CHEMISTRY SEMESTER ONE

Measuring Planck s Constant By Martin Hackworth

Simulation of Second Order Spectra Using SpinWorks. CHEM/BCMB 8190 Biomolecular NMR UGA, Spring, 2005

CH 112 Special Assignment #4 Chemistry to Dye for: Part C

Vibrations of Carbon Dioxide and Carbon Disulfide

Zetasizer Nano-ZS User Instructions

REVIEW OF LAB TECHNIQUES

Investigating Transition Metal Complexes

Molecular Modeling and Conformational Analysis with PC Spartan

Automatic Watershed Delineation using ArcSWAT/Arc GIS

Quality Measures Green Light Report Online Management Tool. Self Guided Tutorial

Refine & Validate. In the *.res file, be sure to add the following four commands after the UNIT instruction and before any atoms: ACTA CONF WPDB -2

Introduction to Astronomy Laboratory Exercise #1. Intro to the Sky

Work with Gravity Data in GM-SYS Profile

Force vs time. IMPULSE AND MOMENTUM Pre Lab Exercise: Turn in with your lab report

Applying MapCalc Map Analysis Software

Work and Energy. This sum can be determined graphically as the area under the plot of force vs. distance. 1

CLEA/VIREO PHOTOMETRY OF THE PLEIADES

Laboratory Measurements and Procedures

SuperCELL Data Programmer and ACTiSys IR Programmer User s Guide

SPECTROPHOTOMETERS. Visible and UV-Visible

Figure 2.1 The Inclined Plane

NIST DTSA-II ( Son of DTSA ): Step-by-Step

EMISSION SPECTROSCOPY

Newton's 2 nd Law. . Your end results should only be interms of m

Physics 476LW Advanced Physics Laboratory Atomic Spectroscopy

v1.3, 11/9/2017 CASSIS as a spectroscopy tool for VESPA S. Erard, J.-M. Glorian, P. Le Sidaner, C. Chauvin

Experiment 9. Emission Spectra. measure the emission spectrum of a source of light using the digital spectrometer.

17-Nov-2015 PHYS MAXWELL WHEEL. To test the conservation of energy in a system with gravitational, translational and rotational energies.

Transcription:

Laser Spectroscopy Labs Jasco V-670 absorption spectrometer Operation instructions

1. Turn ON the power switch on the right side of the spectrophotometer. It takes about 5 minutes for the light source to stabilize. Wait until the light source has stabilized before starting measurement. 2. Double click on Spectra Manager icon on a desktop: Measurement programms a. Spectra measurement This program obtains UV/VIS absorption spectra of a sample. The spectrum acquired using this program is automatically transferred to the [Spectra Analysis] program for analysis. b. Quantitative measurement

This program creates a calibration curve by measuring a standard sample of known concentration according to the common quantitative analysis method and measures an unknown sample to determine its concentration. c. Time Course Measurement This program is used to measure changes to a sample over time at a fixed wavelength. The spectrum acquired is transferred by this program to the [Spectra Analysis] program for analysis. d. Fixed Wavelength Measurement This program measures the absorbance or transmittance of a sample at a fixed wavelength. Up to eight wavelengths can be set and measured. e. Abs/%T Meter The Abs/%T program is used to monitor photometric values. It is useful when a simple photometric value readout is required. f. Validation The validation program is used to check the basic performance of the spectrophotometer. It conducts tests based on various official methods. Analysis program a. Spectra Analysis This program saves, prints, and processes (difference spectra, peak picking, smoothing, derivative, vertical axis conversion, etc.) spectrum data or time course data. b. JASCO Canvas Use this program to layout and print spectra, measurement parameters, comments, etc. It can also be used to create drawings and enter characters. 3. Double click on Spectra Measurement. 4. Setting Measurement Parameters a. Select [Parameters] from the menu [Measure] (or click the button):

b. The [Parameters] dialog will be displayed. The [General] tab is open by default in the Parameters dialog. c. The measurement conditions are set in the [General] tab. The setting conditions for [Basic Mode] are used as an example here. To toggle between [Basic Mode] and [Advanced Mode], click the <Basic Mode > or <Advanced Mode > button. d. It is suggested to have NIR bandwidth 4 times greater than UV-Vis bandwidth for better data appearances. e. In [Control] tab the lamp and/or grating point in spectrum can be change. The margins of the changes will be seen when one put the mouse cursor on it. One can change the point of the change if the absorption lines coincide with the currently set point. f. [Information] tab. Enter the [Sample Name], [Operator], [Division], and [Comment] as desired. The information entered here is saved as comment information together with the spectral data when the spectrum is saved. To display the [Information] dialog prior to taking a measurement, check the [Display the Comment] dialog box before taking measurement] box. g. [Data] tab. Set the [Auto Save], [Send to Analysis] and [Print] options for the spectrum obtained.

5. Baseline measurement. a. The baseline defines the absorbance 0 (or transmittance 100%) and is subtracted from the measurement result to give a correct sample spectrum (or is divided in the case of transmittance). The baseline characteristics are inherent to an instrument, but in an actual measurement they differ slightly depending on the measurement parameters, such as response, scanning speed and other settings. It is recommended to measure the baseline under the same conditions as the actual spectrum measurement conditions in order to obtain an accurate spectrum. b. To perform baseline measurement, click [Measure] - [Baseline Measurement] (or click the button): c. Starts a baseline measurement. Click the [Measure] button to start a measurement. 6. Sample measurement. a. Select the [Measure] - [Sample] menu (or click the button) b. Sample measurement starts and the progress of the measurement will be displayed. c. After the measurement has been finished the data is sent to Spectra Analysis if this option has been checked in [Data] tab. Otherwise go to File-Send to Spectra Analysis option menu.

7. Saving the data. a. The data is saved and analyzied in Spectra Analysis. b. To save the data, go to File-Save As. Always save the data in internal program format. Also, save the data in ASCII code (.txt,.csv or.dat) format through File- Export.

8. Shutting down procedure. a. Close Spectra Analysis. b. Close Spectra Measurement or another measurement program that has been used. c. Close Spectra Manager. The following window will appear and drivers will be terminated. d. Only after the drivers have been terminated switch off the instrument.