Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

Similar documents
Quantitation and Characterization of Copper and Nickel Plating Bath Additives by Liquid Chromatography

Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution

Quantitation of Surfactants in Samples by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Determination of the Composition of Natural Products by HPLC with Charged Aerosol Detection. Introduction. Black Cohosh. Corona Detector Parameters

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Enhancement of Linearity and Response in Charged Aerosol Detection

Direct Analysis of Surfactants using HPLC with Charged Aerosol Detection

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Charged Aerosol Detection and Evaporative Light Scattering Detection Fundamental Differences Affecting Analytical Performance

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

Introduction to Fourier Transform Infrared Spectroscopy

Fitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

of mass spectrometry

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

Analytical Methods to Qualify and Quantify PEG and PEGylated Biopharmaceuticals

Comprehensive Neurochemical Profiling of Brain Tissue Samples

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

Monitoring Protein PEGylation with Ion Exchange Chromatography

Method Development for a Simple and Reliable Determination of PCBs in Mineral Insulating Oil by SPME-GC-ECD

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Fast methods for the determination of ibuprofen in drug products

Discover what you re missing

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency

Determination of Verapamil Hydrochloride Purity Using the Acclaim PA Column

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

A Platform Method for Pharmaceutical Counterion Analysis by HPLC

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

HILIC Method Development in a Few Simple Steps

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications

Advances in Sample Preparation for Accelerated Solvent Extraction

Determination of trace anions in concentrated hydrofluoric acid

Expanded capabilities in ammonium and amine detection

Introduction to Fourier Transform Infrared Spectroscopy

for XPS surface analysis

Semi-Targeted Screening of Pharmaceutically- Related Contaminants in the Thames Tideway using LC-HRMS

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Use of Charged Aerosol Detection as an Orthogonal Quantification Technique for Drug Metabolites in Safety Testing (MIST)

New Multi-Collector Mass Spectrometry Data for Noble Gases Analysis

New Stationary Phases for Solid-Phase Extraction. Pranathi R. Perati, Rosanne Slingsby, and Carl Fisher Thermo Fisher Scientific, Sunnyvale, CA, USA

Hydrophilic Interaction Liquid Chromatography: Method Development Approaches

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System

Thermo Fisher Scientific, San Jose, CA; 2 Kelleher Lab, Northwestern University, Evanston, IL; 3

( )( ) Selectivity Choices in Reversed-Phase Fast LC. Introduction. R s = 1 a 1 k 4 a 1 + k

Detection of Mycotoxins in Corn Meal Extract Using Automated Online Sample Preparation with Liquid Chromatography-Tandem Mass Spectrometry

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Determination of Methylamine in Drug Products

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD)

Metoprolol and Select Impurities Analysis Using a Hydrophilic Interaction Chromatography Method with Combined UV and Charged Aerosol Detection

Determination of Carbonyl Compounds in Workplace Air

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

[ Care and Use Manual ]

Thermo Scientific Accucore XL HPLC Columns. Technical Manual

Simultaneous Determination of Pharmaceutical Peptides and Acetate Counterions by HPLC Using a Mixed-Mode Weak Anion-Exchange Column

Mixed-Mode, Weak Anion-Exchange, Solid-Phase Extraction Method for the Extraction of Niflumic Acid from Human Plasma

A Total Solution for Explosives Analysis by Reversed-Phase HPLC with a Parallel HPLC System

Determination of Trace Cations in Power Plant Waters Containing Morpholine

XPS Surface Characterization of Disposable Laboratory Gloves and the Transfer of Glove Components to Other Surfaces

Dionex IonPac AS28-Fast-4µm column

The key to your laboratory s success

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection

Keywords Haloacetic acids, Water analysis, 2-D ion chromatography, ICS-3000

Paul A. Ullucci, Bruce Bailey, Ian N. Acworth, Christopher Crafts, Marc Plante Thermo Fisher Scientific, Chelmsford, MA, USA

Elemental Analysis: CHNS/O characterization of polymers and plastics

Automated and accurate component detection using reference mass spectra

Thermo Scientific FLASH HT Plus Elemental Analyzer for IRMS

Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC

Clinical Toxicology. Biomass Component Extraction: The uneaten cooked plant specimen was prepared for

Transcription:

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection Marc Plante, Bruce Bailey, Ian N. Acworth Thermo Fisher Scientific, Chelmsford, MA, USA

Overview Purpose: Analytical methods to determine quantities of copper plating bath additives are described. These methods must be stable, sufficiently sensitive, and retain flexibility for use with the many formulations of copper plating baths that exist. Methods: Two HPLC methods are run simultaneou sly to quantitatively measure the three additives that are typically used in copper plating baths. Both methods use reversed-phase high pressure liquid chromatography (RP-HPLC), which can be run simultaneously on the Thermo Scientific Dionex UltiMate 3000 HPLC. The accelerator and suppressor are measured using the Thermo Scientific Dionex Corona charged aerosol detector; the accelerator and leveller are measured using the Thermo Scientific Dionex Coulochem III electrochemical detector (ECD). Results: The methods are precise and sensitive for the determination of all additives. A quantitative measure of suppressor and suppressor degradation is presented. Calibration curves and sample analysis results are reported for all additives. Both analyses can be run using the same sample preparation. Introduction Copper plating baths are used in the manufacture of a multitude of products, from the relatively humble cooking pot to the most advanced integrated circuits and satellites. In order to provide the highest quality and most consistent products with copper plated components, the plating process must be well characterized and tightly controlled. One of the most common approaches to the copper plating is the acid bath, using copper sulfate, sulfuric acid, and a number of additives, namely the accelerator (typically a bis(sulfoalkyl) disulfude), suppressor (a polyalkylglycol), and leveller (either a large molecular weight polymer or small molecule containing nitrogen or sulfur). Each modifier serves a particular function controlling the speed of plating, surface wetting, and gap-filling in order to provide a smooth surface. The most commonly used technique, cyclic voltammetric stripping (CVS), measures these additives separately or combined, and has been cited as being slow (hours) and not very accurate. 1 HPLC has also been investigated, but with few published results. The accelerator and leveller, are present in minute concentrations, and a lack of a chromophore for most modifiers limits the choice of detectors that can be used for quantitation. HPLC can provide selective quantitation of these additives, without the use of sulfuric acid mobile phases, 1,2 which cause rapid column deterioration. A faster, quantitative measure of additives can be achieved using the UltiMate 3000 x2 Dual LC system with two detectors: the Corona ultra RS and the Coulochem III detectors. The Corona charged aerosol detector is a non-selective detector, capable of measuring any nonvolatile analyte at nanogram sensitivities regardless of whether the analyte possesses a chromophore or not. The Coulochem III detector is both extremely sensitive and selective and is ideal for measuring low levels of electrochemically active analytes in complex matrices. Both methods can be run simultaneously, as shown in the system schematic in Figure 1. With this configuration, the autosampler can be exchanged between the two systems without interrupting flow to either system. FIGURE 1. Schematic of the parallel setup HPLC solution for the simultaneous operation of both analytical methods: system 1 for the accelerator and leveller by ECD, and system 2 for the accelerator and suppressor by charged aerosol detector. 2 Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

The charged aerosol detector is a sensitive, mass-based detector, especially wellsuited for the determination of any nonvolatile analyte independent of chemical characteristics. As shown in Figure 2, the detector uses nebulization to create aerosol droplets. The mobile phase evaporates in the drying tube, leaving analyte particles, which become charged in the mixing chamber. The charge is then measured by a highly sensitive electrometer, providing reproducible, nanogram-level sensitivity. This technology has greater sensitivity and precision than evaporative light scattering (ELS) and refractive index (RI), and it is simpler and less expensive to operate than a mass spectrometer (MS). Typical characteristics of charged aerosol detection include: lownanogram on column (o.c.) amounts detected, over four orders of magnitude of dynamic range, and high precision results, typically less than two percent of peak area RSD. Analyte response is also largely independent of chemical structure, providing clear relationships among different analytes in a sample. FIGURE 2. Schematic and functioning of charged aerosol detection. 1 3 10 2 4 8 6 9 5 7 1. Liquid eluent enters from HPLC system 2. Pneumatic nebulization occurs 3. Small droplets enter drying tube 4. Large droplets exit to drain 5. Dried particles enter mixing chamber 6. Gas stream passes over corona needle 7. Charge gas collides with particles and charge is transferred 8. High mobility species are removed 9. Charge measured by electrometer 10. Signal transferred to chromatographic software The Coulochem III ECD uses unique coulometric working electrodes that offer extreme sensitivity and selectivity, well beyond those achieved by traditional amperometric electrodes. The selectivity of serially placed coulometric electrodes is presented in Figure 3. Typically the first electrode is held at a low potential, the second at a higher potential. As the analytes pass through from one electrode to the other, labile compounds will respond (oxidize) at the first electrode, leaving the second (downstream) electrode free to measure more stable compounds. Electrodes are 100% efficient, which provides the selectivity. In the example below, analyte A oxidizes to P on electrode 1 (E1) held at 100 mv applied potential effectively removing it from further reaction. Analyte B remains unchanged until it encounters electrode two (E2) at 500 mv applied potential. At E2, analyte B oxidizes to Q. This provides a selective means of determining amounts of different analytes. FIGURE 3. Functioning of electrochemical detection using coulometric cells. A 100 mv P + e - E1 A P A P E1 B 500 mv Q + e - E2 Flow B Q B Q E2 Thermo Scientific Poster Note PN70008_e 05/13S 3

Methods Sample Preparation All copper plating bath standard solutions and samples must be properly neutralized prior to injection onto the HPLC system. Liquid Chromatography HPLC System: UltiMate 3000 x2 Dual LC, parallel setup solution Mobile Phase 1A: Water, 2.5 g/l Sodium perchlorate, 2.5 ml/l 10% perchloric acid Mobile Phase 1B/2B: Methanol Mobile Phase 1C: Acetonitrile/methanol (900:100) with perchlorates as above Mobile Phase 2A: (70% Ethyl amine)/acetic acid in water (6 ml/l:4 ml/l), ph 5-6 Injection Volumes: 100 µl HPLC Column 1: Thermo Scientific Accucore C18, 2.6 µm, 4.6 150 mm HPLC Column 2: Thermo Scientific Acclaim 120 C18, 5 µm, 4.6 150 mm Column Temperature: 40 C Detector 1: Coulochem III ECD with Thermo Scientific Dionex 5010A Standard Analytical Cell Electrode 1: +650 mv Electrode 2: +900 mv, relative to Pd Detector 2: Corona ultra RS Data Rate: 10 Hz Filter: 4 Power Function: 2.00 (5.8 8.5 minutes) Sample Temperature: 20 C Analysis Time: 16 minutes Gradients: Time (min) -5 0 1 2 3 5 6 6.1 7 10 10 11 Flow Rate 1 (ml/min) 1.0 0.7 0.7 0.9 0.9 1 1 1.1 1 1.0 %1B 2 2 2 0 0 0 0 2 2 2 %1C 0 0 0 35 40 100 100 0 0 0 Flow Rate 2 (ml/min) 0.7 0.7 0.7 0.75 0.65 0.6 0.55 0.8 0.7 $2B 2 2 2 2 25 55 100 100 2 Data Analysis The HPLC system, data collection, and processing were all operated by and performed on the Thermo Scientific Dionex Chromeleon 7.1 software. Results The samples of plating bath and of plating bath additive stock solutions ( standards ) were from two, unrelated sources. As a result, the components of the plating bath samples may not be the same as those of the standards. Consequently, samples were analyzed according to nominal concentrations (NC) used by the source of the standards, where analytes matched. Accelerator and Leveller Analysis by Electrochemical Detection Standards of stock additive solutions were prepared in concentrations of 300% NC, and serially diluted to low concentrations. Analytical runs were made in triplicate to determine calibration curves and instrument precision, shown in Figures 4 and 5, respectively. At 100% NC, five samples were individually prepared and analyzed once each. The linear correlation coefficients were high with R 2 = 0.9987 and 0.9945 for accelerator and leveller, respectively. Peak area percent RSD values ranged from 0.6 to 2.3 for accelerator and 4.9 to 18.6 for leveller, with the higher values at the low concentrations. FIGURE 4. Calibration curve for accelerator by LC-ECD at +900 mv from 10 300% NC. FIGURE 5. Calibration curve for leveller by LC-ECD at +650 mv from 10 200% NC. Accelerator External ECD_2 18.0 µa*min 17.0 16.0 15.0 14.0 13.0 12.0 11.0 10.0 9.0 8.0 7.0 6.0 5.0 4.0 3.0 2.0 1.0 %-Nominal 0.0 0 25 50 75 100 125 150 175 200 225 250 275 300 325 350 Leveller External ECD_1 2.00 µa*min 1.90 1.80 1.70 1.60 1.50 1.40 1.30 1.20 1.10 1.00 0.90 0.80 0.70 0.60 0.50 0.40 0.30 0.20 0.10 %-Nominal 0.00 0 20 40 60 80 100 120 140 160 180 200 220 4 Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

FIGURE 6. HPLC-ECD chromatogram, two potentials overlaid, of used copper plating bath containing accelerator and leveller(s). Peak at 4.895 minutes is likely a free amine-polymer, based on the peak shape and potential of oxidation. 370 µa 300 200 100 2-4.895 7.0 µa 6.0 1 - Leveller - 7.038 5.0 4.0 1 - Accelerator - 3.032 3.0 2.0 1.0 0.0-1.0-2.0 min -3.0 6.00 6.20 6.40 6.60 6.80 7.00 7.20 7.40 7.60 7.80 8.00 +900 mv 0 1 - Leveller - 7.037 +650 mv -100 min -130 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.1 Spike and recovery values were determined for the accelerator and leveller: a bath sample was diluted to 50% to find initial concentrations, and then a second sample was diluted with 100% NC standard, yielding a 50% spike. The overlaid chromatograms are shown in Figure 6. The recovery values were found to be 99% for the accelerator, and 70% for the leveller. This leveller recovery value was confirmed using a second bath sample, and by the method of standard addition, indicating that the recovery value is stable. The signal-to-noise (S/N) value of the leveller (blue) in this spiked sample was 66, indicating that sufficient sensitivity was available for these determinations. Signal-to-noise values of 10 and 3.3 were used to calculate the limits of quantitation (LOQ) and detection (LOD), respectively. The LOQ and LOD values were 1 and 0.3% NC for the accelerator and 20% and 7% NC for the leveller, respectively. Accelerator and Suppressor Analysis by Charged Aerosol Detection Standards of stock additive solutions were prepared in concentrations of 300% NC, and diluted to low concentrations. Analytical runs were made in triplicate to determine calibration curves and instrument precision, shown in Figures 7 and 8, respectively. At 100% NC, five samples were individually prepared and analyzed once each. Calibration curves for Corona detectors are non-linear. However, with the use of the Power Function, a linear calibration fit was obtained. This power function adjusts the outgoing signal to provide for linear calibration data, which are important when peak shape changes occur between standard and sample analyte peaks. The correlation coefficients were high with R 2 = 0.9987 and 0.9957 for accelerator and suppressor, respectively. Precision RSD values, based on peak areas varied between 1.0 and 5.3% for the accelerator, and were less than 1% for the suppressor. As with the ECD experiments, the RSD precision values of the sample at 100% NC showed similar values using the charged aerosol detector, indicating that sample preparation is reproducible. In addition to method accuracy and precision data, the LC-charged aerosol detector method was evaluated for spike recovery in a similar manner as indicated for the ECD evaluation. Recovery values for the accelerator was 103%, and for the suppressor, 95-100%. Analyte resolution was also evaluated. Four peaks were resolved for this accelerator (not shown). The sensitivity for the accelerator was found to be 3% NC for LOQ, based on a S/N ratio of 10. In the sample chromatograms shown in Figure 9, two plating bath samples are overlaid consisting of a new and a used (diluted 50%) plating bath. The suppressor is seen as the largest peak in the chromatograms, along with many smaller peaks with earlier retention times. These smaller peaks represent lower molecular weight fractions of the suppressor. Compared to the new bath, the suppressor concentrations differed by nearly seven-fold in the used bath, along with a four-fold increase in the relative amounts of the smaller molecular weight fractions. These changes are related to suppressor degradation as the bath is operated. 3 As the plating bath is used, additives are consumed by the process and eventually replenished, as indicated by analytical determinations. For example, the ECD results indicate that the amount of accelerator in the used bath increased by 2.5-fold, relative to the new bath. Thermo Scientific Poster Note PN70008_e 05/13S 5

FIGURE 7. Inverted calibration curve for accelerator by LC-charged aerosol detector from 0.6-20 ml/l (200% NC) Accelerator 4.00 pa*min External CAD_1 3.80 3.60 3.40 3.20 3.00 2.80 2.60 2.40 2.20 2.00 1.80 1.60 1.40 1.20 1.00 0.80 0.60 0.40 0.20 ml/l 0.00 0.0 2.0 4.0 6.0 8.0 10.0 12.0 14.0 16.0 18.0 20.0 22.0 25.0 FIGURE 8. Inverted calibration curve for suppressor by LC-charged aerosol detector from 0.6-20 ml/l (200% NC) 30.00 28.75 27.50 26.25 25.00 23.75 22.50 21.25 20.00 18.75 17.50 16.25 15.00 13.75 12.50 11.25 10.00 8.75 7.50 6.25 5.00 3.75 2.50 1.25 0.00 Suppressor External CAD_1 pa*min ml/l 0.0 2.0 4.0 6.0 8.0 10.0 12.0 14.0 16.0 18.0 20.0 22.0 25.0 FIGURE 9. Overlay chromatograms of new (in black) and used (in blue, diluted 50%) copper plating baths, using HPLC with charged aerosol detection. 400 pa 350 2 - Suppressor 300 250 200 150 100 50 1 - Accelerator 0-50 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 min 11.0 Conclusions Using charged aerosol detection and electrochemical detection running in parallel on the UltiMate 3000 2x Dual LC provided excellent quantitative data on the three copper plating bath additives Analyses have sufficient sensitivity and range to be applicable to many plating bath compositions. Analysis was rapid, providing results for neutralized samples in as little as 22 minutes. The method is extremely versatile, enabling resolution for different additives of the same class, and can be adjusted to accommodate different bath compositions. The assay also resolved a number of potentially important degradants. References 1. Hong, K. J. Korean Phy. Soc., 43(2), 2003, pp. 286-289. 2. Application Note 139: Determination of Additives and Byproducts in an Acid Copper Plating Bath by Liquid Chromatography, LPN1251-01, 2002. Dionex, Part of Thermo Fisher Scientific. http://www.dionex.com/en-us/webdocs/4755- AN139_V16.pdf (accessed 24 Jan 2011). 3. Pavlov, M.; Shalyt, E; Bratin, P. A New Generation of CVS Monitors Cu Damascene Plating Baths. Electro IQ.[Online] http://www.electroiq.com/articles/ sst/print/volume-46/issue-3/features/chemical-handling/a-new-generation-of-cvsmonitors-cu-damascene-plating-baths.html (accessed 24 Jan 2012). 6 Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

www.thermofisher.com/dionex 2016 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Africa-Other +27 11 570 1840 Australia +61 3 9757 4300 Austria +43 1 333 50 34 0 Belgium +32 53 73 42 41 Canada +1 800 530 8447 China +86 10 8419 3588 Denmark +45 70 23 62 60 PN10326_E 11/16S Europe-Other +43 1 333 50 34 0 Finland/Norway/Sweden +46 8 556 468 00 France +33 1 60 92 48 00 Germany +49 6103 408 1014 India +91 22 6742 9494 Italy +39 02 950 591 Japan +81 45 453 9100 Latin America +1 561 688 8700 Middle East +43 1 333 50 34 0 Netherlands +31 76 579 55 55 New Zealand +64 9 980 6700 Russia/CIS +43 1 333 50 34 0 South Africa +27 11 570 1840 Spain +34 914 845 965 Switzerland +41 61 716 77 00 UK +44 1442 233555 USA +1 800 532 4752