Elemental Analysis: CHNS/O characterization of polymers and plastics

Similar documents
Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

Elemental Analysis: NCS characterization of paper

Accurate and Precise Sulfur Analysis using the Thermo Scientific FlashSmart Elemental Analyzer

Nitrogen/Protein determination of infant food by the Thermo Scientific FlashSmart Elemental Analyzer using helium or argon as carrier gases

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

Defining quality standards for the analysis of solid samples

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

Thermo Scientific FLASH HT Plus Elemental Analyzer for IRMS

Overcoming Interferences with the Thermo Scientific icap 7000 Plus Series ICP-OES

ECS8020 ORGANIC ELEMENTAL ANALYZER CHNS-O Analyzer

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD)

Expanded capabilities in ammonium and amine detection

Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

costech instruments ECS 4010 Elemental Combustion System CHNS-O

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Finnigan TC/EA. High Temperature Conversion Elemental Analyzer

Introduction to Fourier Transform Infrared Spectroscopy

Mass Spectral Studies of Polypropylene Chromatographic Well Plates

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Accurate measurement of elemental impurities in metals and metal alloys using the Thermo Scientific icap TQ ICP-MS

Analysis of trace elements in polymer samples to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES

Speed and flexibility

Automated and accurate component detection using reference mass spectra

Structural characterization begins with a purity check!

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

of mass spectrometry

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Determination of trace anions in concentrated hydrofluoric acid

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

XPS Surface Characterization of Disposable Laboratory Gloves and the Transfer of Glove Components to Other Surfaces

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter

Determination of Carbon d13c isotopic composition from DIC samples

Supported Liquid Extraction (SLE) Guide and FAQ s. User Guide

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Simplified workflows for the Thermo Scientific icap TQ ICP-MS

AN INTEGRATED SYSTEM USING TEMPERATURE BASED SAMPLING FOR POLYMER CHARACTERIZATION

Ideal and powerful solution

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Introduction to Fourier Transform Infrared Spectroscopy

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Complete Materials Deformulation Using TGA-IR

Introduction to Gas Chromatography

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

The key to your laboratory s success

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

PRINCIPLE. III. EQUIPMENT a. FlashEA 1112 b. Micro balance c. Computer IV. MATERIALS a. Two forceps (curved tip)

2401 Gas (liquid) Chromatography

δ 13 C of DIC and carbonate samples: comparison of traditional mass spectrometry methods with an infrared spectrometry method

Pyrolysis process for the treatment of food waste. Barbora Grycová VSB - Technical University of Ostrava Czech Republic

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

AOAC Official Method 2016.xx. Determination of Total Sulfur in Fertilizers by High Temperature Combustion

Agilent G2350A Atomic Emission Detector (AED)

Fitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS

Combustion Ion Chromatography. Fast and reliable determination of halogens and sulfur

Gas chromatography. Advantages of GC. Disadvantages of GC

Dionex IonPac AS28-Fast-4µm column

CHNS Analyser CHNS+O COMBO ANALYSER EPC / PRODUCTS / APPLICATION / SOFTWARE / ACCESSORIES / CONSUMABLES / SERVICES. Analytical Technologies Limited

NON-METHANE ORGANIC CARBON ANALYZER (NMOC Method 25)

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications

Quantification of total nitrogen according to the Dumas method (Musts and Wines) (Resolution Oeno 13/2002)

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

Determination of urea in ultrapure water by IC-MS/MS

The art of elemental analysis

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Determination of an anionic fluorochemical surfactant in a semiconductor etch bath

Fast methods for the determination of ibuprofen in drug products

Determination of chlorine, bromine, and sulfur in polyethylene materials using combustion IC with a carbonate/bicarbonate eluent

Standard Test Method for Determination of Trace Thiophene in Refined Benzene by Gas Chromatography 1

all suspects Identify Thermo Scientific Extractables & Leachables Workflow Unknown impurity identification and quantification workflows for:

Analysis of Polymers and Plastics. Innovation with Integrity. Quality Control & Failure Analysis FTIR

High-Speed Gas and Headspace Analysis for the Process-Line and Laboratory: SIFT- MS IFPAC 2017

( )( ) Selectivity Choices in Reversed-Phase Fast LC. Introduction. R s = 1 a 1 k 4 a 1 + k

Method Development for a Simple and Reliable Determination of PCBs in Mineral Insulating Oil by SPME-GC-ECD

Analysis of Polymers and Plastics. Innovation with Integrity. Quality Control & Failure Analysis FT-IR

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

Halogen Free Flame Retardant Based on Organophosphorus Compounds for Expandable Polystyrene Produced by Supension Polymerization

Thermo Scientific HyperSep Dispersive SPE Products. Efficient sample preparation and clean-up using the QuEChERS Method

Thermo Scientific GasBench II

Oxygen Permeation Through FEP Teflon and Kapton Polyimide

Transcription:

APPLICATION NOTE 0 Elemental Analysis: CHNS/O characterization of polymers and plastics Authors Introduction Dr. Liliana Krotz and Dr. Guido Giazzi Thermo Fisher Scientific, Milan, Italy The chemical composition of polymers and plastics is connected with their chemical, physical, mechanical properties. The development and production of polymers and plastics requires quality control of raw materials, additives, stabilizers, intermediate and finished products. The analysis of the behavior of polymers and plastics during molding and the evaluation of their lifetime contribute to define their quality. For the material characterization and the quality control testings of polymers and plastics, nitrogen, carbon, hydrogen, sulfur and oxygen are determined. Nitrogen determination is crucial, and the importance of sulfur determination has increased also. Nitrogen containing compounds are used in the production process of polymers and plastics to trigger a polymerization reaction. They can also be used as additives for the addition of specific properties to polymers and plastics. Keywords CHNS/O, Dumas method, Plastics, Polymers Goal This application note shows CHNS/O determination for polymers and plastics with the FlashSmart EA, needed for material quality control. As additives, nitrogen containing compounds provide the final product with specific properties, and they act as: stabilizing emulsion polymers, chain transfer agents and other polymerization modifiers to control molecular weight, plasticizers to increase flexibility, stabilizers to prevent polymer degradation, crosslinkers used to modify polymer networks. As the demand for material characterization testing has grown in recent years and elements are present at trace levels, the classical analytical methods showed to be no longer suitable, for their time-consuming sample preparation and for their use of hazardous reagents. For this reason an automated technique providing accurate data at trace levels is the requirement for modern laboratories dealing with routine analysis.

The Thermo Scientific FlashSmart Elemental Analyzer (Figure ), meets laboratory requirements such as accuracy, day to day reproducibility and high sample throughput. It improves the productivity of the laboratory over traditional methods, as it is automated and modular. CHNS determinations can be performed in a single analysis run and oxygen determination by pyrolysis in a second run. The FlashSmart EA can be configured to perform a single nitrogen determination or trace sulfur analysis when coupled to the Flame Photometric Detector (FPD). This note presents data on CHNS/O determination in polymers and plastics to show the performance of the FlashSmart Elemental Analyzer. Methods For CHNS or CHN determination the FlashSmart Elemental Analyzer operates with dynamic flash combustion of the sample. Samples are weighed in tin containers and introduced into the combustion reactor via the Thermo Scientific MAS Plus with oxygen. After combustion, the produced gases are carried by a helium flow to a layer containing copper, then swept through a GC column, which provides the separation of the combustion and finally, detected by a Thermal Conductivity Detector (TCD). CHNS Configuration Oxygen Configuration Figure. Thermo Scientific FlashSmart Elemental Analyzer. Total run time less than 0 minutes (Figure ). For trace sulfur determination, the gases produced by combustion are carried by a helium flow to a layer containing copper, then swept through a water trap, a short GC column and finally the sulfur is measured by the Flame Photometric Detector (FPD). Total run time is 5 minutes (Figure ). For oxygen determination, the system operates in pyrolysis mode. Samples are weighed in silver containers and introduced into the pyrolysis chamber via the MAS Plus. The reactor contains nickel coated carbon at a temperature of 060 C. The oxygen in the sample, combined with the carbon, forms carbon monoxide which is chromatographically separated from other products and detected by the TCD Detector (Figure ). Oxygen Pyrolysis Figure. CHNS/O configuration. HO Trap TCD For nitrogen determination, the Elemental Analyzer operates by dynamic flash combustion of the sample. Samples are weighed in tin containers and introduced into the combustion reactor via the MAS Plus with oxygen. After combustion, the produced gases are carried by a helium flow to a second reactor containing copper, then swept through CO and H O traps, and a GC column, and finally, detected by a Thermal Conductivity Detector (TCD). Total run time less than five minutes (Figure ). SO N Oxygen Oxygen CO Trap H O Trap H O Trap FPD TCD Figure. Sulfur configuration by FPD Detector. Figure. Nitrogen configuration.

A complete report is automatically generated by the Thermo Scientific Data Handling Software. Results Different types of polymers and plastics, raw materials and additives, with different content and aspect (pellets, powders and films), were analyzed. Table shows the repeatability of CHNS/O determination in polyimide sample and Table shows the repeatability of NCS determination in polymers powders. For CHNS and NCS analysis, the system was calibrated with mg of BBOT* standard using K factor as calibration method. The sample were weighed at mg. For oxygen determination, mg of benzoic acid was analyzed as standard using K factor. The sample were weighed at mg. Table shows the CHN/O data of polyacrilonitriles. The calibration was performed with mg of acetanilide using K factor as calibration method, and sample was weighed at mg. Acrilonitrile (monomer) is the main component of many types of compounds (polyacrylonitriles) such as plastics. Table. CHNS/O data of polyimide. Element N% C% H% S% O%.8 6.08.6.950 0.0.9 6.09.69.90 0..78 6.68.60.9 0.6 Average.85 6..60.9 0.5 RSD% 0.5 0.066 0. 0.5 0. Table. NCS data of polymers powders. Sample N% RSD% C% RSD% S% RSD% 5.89 5.98 0. 5.770 0.0 5.950 0.58 0. 0.58 5.756 5.89 0..775 9.0.66.76 0.05 8.889 0.9.8 0.70.777 8.97.67 * BBOT:,5-Bis (5-tert-butyl-benzoxazol--yl) thiophene Table. CHN/O data of polyacrilonitriles. Sample N% RSD% C% RSD% H% RSD% O% RSD% A.7.5.7 0.5 6.6 6.5 6. 0. 5.7 5.76 5.68 0.7.79.59.6.5 B.77.5. 0.75 6.65 6. 6. 0. 5.7 5.66 5.6 0.98 5. 5.7 5.0 0.7 C.97.88.88 0. 6.88 65.0 65. 0.0 5.67 5.66 5.70 0.7.99.0.98 0.5

Table shows the nitrogen data of polymers in pellets. The system was calibrated with 0 mg atropine standard using K factor as calibration method. Samples and were weighed at 00 50 mg. For samples to 5 the calibration was performed with 5 6 mg atropine and samples were weighed at 0 0 mg. Table 5 shows the nitrogen data of polycarbonate polymers in pellets analyzed in duplicate. The system was calibrated with 5 0 mg atropine using K factor as calibration method. Sample was weighed at 50 70 mg. Table 6 shows nitrogen data of polyethylene films. As standard, mg atropine was analyzed using K factor as calibration method. Samples were weighed at 0 5 mg. Table 7 shows the nitrogen data of plastics Table. Nitrogen data of polymers (pellets). Sample N (ppm) RSD% 5 985 998 98 00 09 999 005 00 008 0 8 6 6 5 7 9 967 97 969 9 7 7 Table 5. Nitrogen data of polycarbonate polymers (pellets). 9 89 90.589.6 0.06.00. Sample 5 N% 0.675 0.678 0.656 0.65 0.67 0.6 0. 0. 0.0 0. RSD% 0. 0.5 0.57.6 0.60 additives. The calibration was performed with 5 0 mg of atropine using K factor as calibration method. Samples were weighed at 5 0 mg. Table 6. Nitrogen data of polyethylene films. Sample N (ppm) RSD% 5 6 7 8 0 0 97 569 590 55 07 00 06 78 779 788 9 500 9 66 5 8 0 8 0 07 Table 7. Nitrogen data of plastics additives. Raw Material Raw Material Raw Material Raw Material 0...57 0.60 0.8 5.556.66 5.08 Sample N% RSD% Raw Material (Dimeric Benzotriazole) Polycarbonate + Dimeric Benzotriazole Raw Material (Hindered Amine Light Stabilizer) White Polystyrene Granules (Contains Hindered Amine Light Stabilizer) 8. 8.6 8. 9.69 9.6 9.606..07.7...9.799.808.8 0.6 0.65 0.69 5.767 5.75 5.70 0.506 0.50 0.509 0.0 0.087 0.09 0.569 0.057 0.8 0. 0.808

For trace sulfur determination by FPD Detector, the system was calibrated with Thermo Scientific Pasta and Soil Reference Materials with 0.5 S% and 0.0 S%, respectively using quadratic fit as calibration method. Table 8 shows the sulfur data of a polyethylene film sample. Table 9 shows the sulfur data of plastic powders while Table 0 shows the sulfur data of polymers. The samples were cut into small pieces and the samples were weighed at 0.5 mg. Table 0. Sulfur data of polymers. Sample S (ppm) Av. S (ppm) RSD% 95 66 A 67 66.87 56 7 876 B 90 80 869.0 879 Table 8. Sulfur data of polyethylene film sample. S (ppm) RSD% C 69 5 9.0 57 57 50 5 558 Table 9. Sulfur data of plastic powders..79 Sample S (ppm) RSD% 56 58 566 58 5 60 9 8 66 5 70.857 5.9 0.657 0.8 D E F G H 866 856 8 9 969 77 06 8 50 8 7 0 6 78 76 8 59 59 5 6 56 867.6 89.0 9.80 8.0 57 6. Conclusion The Thermo Scientific FlashSmart Elemental Analyzer enables to perform accurate and reproducible CHNS/O determination in polymers and plastics samples. I 0 6 8 9 6. No matrix effect was observed when changing the sample and element content indicating the complete combustion of the samples. J 5 9 7.79 Nitrogen only analyses, simultaneous CHNS, NCS, CHN and oxygen can be performed on the Thermo Scientific FlashSmart Elemental Analyzer with a simple upgrade. Trace sulfur determination can also be performed when coupled to the FPD Detector The FlashSmart Elemental Analyzer meets laboratory requirements in terms of automation, high sample throughput. Find out more at thermofisher.com/oea 07 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representatives for details. AN0-EN 07S