New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection

Similar documents
New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

Efficient Separations Obtained by Using Smaller Particle Size Analytical Columns with a High-Pressure Ion Chromatography System

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Utility of the Charge Detector in Ion Chromatography Applications

Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Dionex IonPac AS28-Fast-4µm column

Thermo Scientific Dionex IonPac AS23 Anion-Exchange Column

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water

Thermo Scientific. Anion-Exchange Column

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection

Thermo Scientific Dionex IonPac AS25 Anion-Exchange Column. Minutes

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

Keywords Haloacetic acids, Water analysis, 2-D ion chromatography, ICS-3000

columns IonPac AS17-C Anion Exchange Column

Expanded capabilities in ammonium and amine detection

Thermo Scientific Dionex IonPac CS17 Cation-Exchange Column

Thermo Scientific Dionex IonPac AS11-HC-4µm Anion-Exchange Column

Determination of Trace Cations in Power Plant Waters Containing Morpholine

Determination of Organic Acids in Beer Samples Using a High-Pressure Ion Chromatography System

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory

Thermo Scientific Dionex IonPac AS24A Anion-Exchange Column

columns IonPac AS22 Anion-Exchange Column

Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples

Determinations of Heat Stable Amine Salts and Amines in Amine Scrubbing Solutions Using Suppressed Conductivity and Charge Detection

Inline sample preparation for the determination of anions in sodium hydroxide

columns IonPac AS18 Anion-Exchange Columns

Determination of Methylamine in Drug Products

Determination of trace anions in concentrated hydrofluoric acid

anion-exchange column

Determination of Bromate in Bottled Mineral Water Using the CRD 300 Carbonate Removal Device

columns IonPac SCS 1 Silica Cation Separator

Determination of urea in ultrapure water by IC-MS/MS

Thermo Scientific Dionex IonPac AS26 Anion-Exchange Column

Thermo Scientific Dionex eluent suppressors for ion chromatography

Superior Chromatographic Performance Recommended anion-exchange column for separation of haloacetic acids prior to MS or MS/MS detection

Determination of Adsorbable Organic Halogen in Wastewater

Thermo Scientific Dionex IonPac AS11 and AS11-HC Anion-Exchange Column

Direct Determination of Small Organic Acids in Sea Water by IC-MS

Monitoring Protein PEGylation with Ion Exchange Chromatography

Concentrator and Trap Columns. Column: IonPac AG15, AS15, 2 mm. Eluent Source: EG40 Concentrator. Temperature: 30 C Column: IonPac AC15 (2 50 mm)

Determination of silicate in high-purity water using ion chromatography and online sample preparation

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Overcoming Challenging Matrices in Ion Chromatography

Accurate and Precise Automated Dilution and In-line Conductivity Measurement Using the AS-AP Autosampler Prior to Analysis by Ion Chromatography

Determination of Silicate in High Purity Water Using Ion Chromatography and AutoPrep

columns I o n P a c AS11 and AS11-H C Anion-E x c h a n g e C o l u m n s

The power to increase productivity. Thermo Scientific Dionex Reagent-Free Ion Chromatography (RFIC) System Capabilities

Advances in Sample Preparation for Accelerated Solvent Extraction

Determination of Inorganic Anions in Drinking Water by Ion Chromatography

for IonPac AG11-HC IonPac AS11-HC

suppressors SRS -ULTRA Self-Regenerating Suppressor

Determination of Sulfate and Chloride in Ethanol by Ion Chromatography

Application of Eluent Generation for Trace Anion Analysis of Borated Waters

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

The determination of trace anions in concentrated phosphoric acid

suppressors Eluent Suppressors for Ion Chromatography

Trace Level Determination of Bromate in Ozonated Drinking Water Using Ion Chromatography

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM

Simultaneous Determination of Pharmaceutical Peptides and Acetate Counterions by HPLC Using a Mixed-Mode Weak Anion-Exchange Column

For new orders of the following parts discussed in this manual, please use the updated part numbers listed below.

Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution

Enhancement of Linearity and Response in Charged Aerosol Detection

Determination of Verapamil Hydrochloride Purity Using the Acclaim PA Column

Determination of trace anions in high-purity waters by high volume/direct injection ion chromatography

Determination of an anionic fluorochemical surfactant in a semiconductor etch bath

Accurate and Reproducible Determination of Organic Halogens Using Combustion Ion Chromatography

The Determination of Trace Anions in Concentrated Phosphoric Acid

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

Sample Preparation for Ion Chromatography

Easy Method Transfer from HPLC to RSLC with the Dionex Method Speed-Up Calculator

High-Performance Chromatography Superior isocratic separation of

Novel Electrodeionization Devices: Applications in Inorganic Analysis

Determination of trace anions in basic solutions by single pass AutoNeutralization and ion chromatography

Application Note 187. Brian DeBorba and Jeff Rohrer Thermo Scientific, Sunnyvale, CA, USA

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Quantitation of Surfactants in Samples by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Determination of trace anions in high-nitrate matrices by ion chromatography

Determination of the Composition of Natural Products by HPLC with Charged Aerosol Detection. Introduction. Black Cohosh. Corona Detector Parameters

Determination of Trace Anions in Concentrated Bases Using AutoNeutralization Pretreatment and Ion Chromatography

Determination of Monochloroacetic Acid in Carbocisteine

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

New Stationary Phases for Solid-Phase Extraction. Pranathi R. Perati, Rosanne Slingsby, and Carl Fisher Thermo Fisher Scientific, Sunnyvale, CA, USA

Determination of Inorganic Cations and Ammonium in Environmental Waters by Ion Chromatography Using the IonPac CS16 Column

Hydrophilic Interaction Liquid Chromatography: Method Development Approaches

Determination of trace anions in organic solvents using matrix elimination and preconcentration

IonPac Trace Cation Concentrator

Development of the Thermo Scientific Dionex IonPac AS26 Column for Haloacetic Acid Analysis

Determination of Trace Anions in Organic Solvents

Determination of Organic Acids and Inorganic Anions in Lithium-Containing Boric Acid-Treated Nuclear Power Plant Waters

For new orders of the following parts discussed in this manual, please use the updated part numbers listed below.

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

ION CHROMATOGRAPHY SYSTEM S 150

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Product Manual for Dionex IonPac CS12A and CG12A Columns

User Manual. IonPac NS2 Columns Revision 02 October For Research Use Only. Not for use in diagnostic procedures.

A Platform Method for Pharmaceutical Counterion Analysis by HPLC

Transcription:

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection Barbara Shao, Terri Christison, Fei Pang, Cathy Tanner, and Frank Hoefler, Thermo Fisher Scientific, Sunnyvale, CA, USA

Overview Purpose: New solutions highlighting the applications and benefits of high-pressure capillary ion chromatography with 4 µm particle size columns in a compact, integrated IC system are discussed. Furthermore, the benefits of a Reagent-Free IC system with eluent generation capabilities to achieve consistent results, day-to-day, instrument-toinstrument and operator-to-operator are presented. Advantages of suppressed conductivity with charge detection (QD) are demonstrated. Methods: All applications were developed on an integrated, high-pressure Thermo Scientific Dionex ICS-4000 Capillary HPIC system using capillary Thermo Scientific Dionex IonPac 4 µm particle-size columns. Suppressed conductivity and the new charge detectors were used in series. Eluents were electrolytically generated, using the Reagent-Free IC system capabilities of the Dionex ICS-4000 HPIC system. Results: Fast separations of biogenic amines using capillary high-pressure IC are achieved in less than 9 minutes. High-resolution separations of inorganic anions and organic acids are presented using 4 µm particle columns. More peaks can be detected in orange juice with the new charge detector. Capillary Ion Chromatography Ion chromatography (IC) has become one of the preferred technologies for determining ionic analytes in any type of matrix, from ultrahigh purity water, to drinking and wastewater, to fruit juices and bodily fluids. Recent developments in ion chromatography, including capillary IC, which was introduced with the Dionex ICS-000 system, have generated great interest. In 2012, the first dedicated capillary highpressure IC system was introduced, targeting routine analysis. In capillary IC, packed columns with internal diameters of 0.4 mm are typically used at flow rates of 10 µl/min, thus enabling continuous operation with less than 1 ml of mobile phase per day or.2 L per year: Format Standard bore Capillary Column i.d. 4 mm 0.4 mm Typical Flow Rate 1.0 ml/min 10 µl/min Typical Injection Volume 2 µl 0.4 µl Eluent Consumption/ Waste Generation 43.2 L/month 0.432 L/month ECG Lifetime (at mm) 28 days 18 months Mass Detection Limits 000 fg 0 fg Further advantages include: Less sample volume is required due to increased mass sensitivity. This is important for sample limited applications, e.g., the analysis of ions in biological fluids. The low flow rates are beneficial for interfacing with mass spectrometry, providing increased sensitivity for trace level analysis. Two-dimensional separations using a 4 mm inner diameter column in the first dimension and a capillary column in the second dimension allow ng/l (ppt) level detection with only 1 ml of sample and suppressed conductivity detection. Capillary ion chromatography was built with technologies currently available for standard and microbore IC, including Reagent-Free IC with eluent generation, electrolytic suppression with conductivity detection, and electrochemical detection. Capillary IC also utilizes new technologies such as charge detection, high-pressure eluent generation, and the use of high efficiency 4 μm particle columns. 2 New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection

Methods Sample Preparation Typical sample preparation includes dilution and filtration using 0.4 μm nylon membrane filters. Ion Chromatography Equipment and Data Analysis Thermo Scientific Dionex ICS-4000 Capillary HPIC System consisting of a capillary Pump, an Eluent Generator module, an IC Cube housing the 0.4 µl injection valve and all capillary consumables, a conductivity (CD), electrochemical (ED), or charge detector (QD), and a Dionex AS-AP autosampler. Thermo Scientific Dionex Chromeleon Data System software, ver. 6.8 was used for instrument control and data analysis and management. Conditions The conditions are included in the figures. High-Pressure Ion Chromatography Capillary HPIC systems allow continuous operation with eluent generation at pressures up to 34 MPa (000 psi). To extend the pressure range from 3000 to 000 psi requires the use of HPIC components, including the pump heads, eluent genaration cartridge, and the eluent generation degasser. This enables the use of 4 μm particle size columns, which produce higher backpressures under standard conditions. The advantages of these new columns include: Smaller particle diameter = higher chromatographic efficiency Higher chromatographic efficiency = faster separations at higher flow rates without sacrificing resolution higher resolution at standard flow rates or Figure1 shows an example for a fast separation of inorganic cations and biogenic amines in less than 9 minutes. All peaks are baseline resolved due to the high efficiency of this column. To achieve these fast run times the flow rate was increased from the standard of 0.010 ml/min to 0.020 ml/min and a steep gradient was applied. Examples for highresolution separations are shown in Figure 4. FIGURE 1. Fast separation of nine inorganic cations and biogenic amines. µs 0 3 4 2 1 6 8 9 11 10 12 13 Column: Dionex IonPac CG19-4µm/ CS19-4µm (0.4 x 20 mm) Eluent: 9 mm MSA, gradient to 0 mm MSA at min, isocratic to. min, back to 9 mm MSA at.8 minutes Eluent Source: Dionex EGC-MSA (Capillary) Flow Rate: 0.020 ml/min Inj. volume: 0.4 µl Temperature: 30 C Detection: Suppressed Conductivity, Thermo Scientific Dionex CCES 300, AutoSuppression, recycle mode Peaks: 1. Lithium 0.0 mg/l 2. Sodium 0.2 3. Ammonium 0.2 4. Potassium 0.. Magnesium 0.2 6. Calcium 0.. Impurity 8. Putrescine. 9. Cadaverine 4. 10. Histamine 6. 11. Agmatine.0 12. Spermine 3.0 13. Spermidine 1. 0 2 4 6 8 10 Minutes NOTE: Concentrations are only approximate Thermo Scientific Poster Note PN0144_e 06/12S 3

Principle of Charge Detection The charge detector uses two membranes, an anion and a cation exchange, to separate the eluent channel from two detector channels. The anode and cathode are located in the two detector channels. A constant voltage is applied to the electrodes, causing ions to migrate from the eluent channel through the membrane into the detector channels (Figure 2). When ions reach the electrodes a current proportional to the charges is generated. The resulting response is similar for the same number (molarity) of similarly charged ions independent from the degree of dissociation, thus providing universal response. Therefore, the charge detector signal is twice as strong for doubly and three times as high for triply charged ions, e.g., phosphate (Figure ). FIGURE 2. Schematic of a charge detector cell. HY A + Y - AOH H 2 O (+) Anion Exchange Cation Exchange (-) Membranes The charge detector responds to all ionic compounds and therefore is best used with hydroxide-based eluents for anion analysis or with methanesulfonic acid eluents for cation analysis, as a second detector installed in series with a suppressed conductivity detector. The charge detector removes all ions from the eluent stream and is therefore always installed last in series with other detectors. A typical set up is shown in Figures 3 and 4. FIGURE 3. Schematic set up of a Capillary RFIC system with suppressed conductivity detection (CD) and charge detection (QD). H 2 0 Data Management High-Pressure Non-Metallic Pump Eluent Generator (OH or H + ) Separation Column Waste CR-TC Sample Inject (Autosampler) Electrolytic Eluent Suppressor Conductivity Detector Charge Detector Recycle Mode 4 New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection

FIGURE 4. Capillary components in a Dionex ICS-4000 Capillary HPIC System. Suppressor (CES) Column Cartridge Eluent Degasser Carbonate Removal Device (CRD) Injection Valve Conductivity Detector (CD) Charge Detector (QD) FIGURE. Overlaid chromatograms of a Seven Anion Standard, obtained with suppressed conductivity detection and charge detection. 8.0 µs 1 2 3 4 6 Column: Dionex IonPac AS1-9µm, 0.4 x 20 mm Eluent: 38 mm Potassium hydroxide Eluent Source: Dionex EGC-KOH (Capillary) Flow Rate: 0.012 ml/min Inj.Volume: 0.4 µl Column Temp.: 30 C Detection: Suppressed conductivity, Thermo Scientific Dionex ACES 300, AutoSuppression, Recycle mode -1.0 0.0 A: Charge Detection B:Conductivity Detection A Peaks: 1. Fluoride 1.0 mg/l 2. Chloride 2. 3. Nitrite.0 B 4. Sulfate.0. Bromide 10.0 6. Nitrate 10.0 10 20 30. Phosphate 1.0 Minutes Analysis of Organic Acids and Inorganic Ions in Orange Juice The separation of organic acids and inorganic ions in an orange juice sample demonstrates the high efficiency of the Dionex IonPac AS11-HC-4µm capillary column (Figure 6). Only 0.4 µl of a 1:0 diluted, filtered sample was injected onto the column. The black trace shows the chromatogram obtained with suppressed conductivity detection and the blue trace represents the one with charge detection. The two chromatograms are scaled to the chloride peak. This example demonstrates that weakly dissociated ions such as organic acids, acetate and maleate, as well as multi-valent ions phosphate and citrate show significantly higher response with charge detection. Thermo Scientific Poster Note PN0144_e 06/12S

FIGURE 6. Determination of organic acids and inorganic anions in orange juice using a Dionex IonPac AS11-HC-4µm, 0.4 x 20 mm anion-exchange column with suppressed conductivity and charge detections..6 2 16 Current [µa] 1 3 4 6 8 9 11 12 14 10 13 1 1 18 19 20 A B 3.8 0.0 Column: Instrument: Eluent Source: Eluent: 10 20 30 42 Minutes Dionex IonPac AS11-HC-4µm Capillary Chromatograms: A: Charge Detection (0.4 x 20 mm) B: Conductivity Detection Dionex ICS-000 HPIC system Peaks: EGC-KOH (Capillary) 1 mm KOH from 0 to 8 min, 1 mm to 30 mm KOH from 8 to 28 min, 30 mm to 60 mm KOH from 28 to 38 min, 60 mm from 38 to 42 min Flow Rate: 0.01 ml/min Inj. Volume: 0.4 µl Column Temp.: 30 C Detection: Suppressed conductivity, Dionex ACES 300 AutoSuppression, Recycle mode Samples: Orange juice, 1:40 dilution with DI water 1. Quinate 11. Carbonate 2. Fluoride 12. Maleate 3. Lactate 13. Sulfate 4. Acetate 14. Oxalate. Formate 1. Unknown 6. Pyruvate 16. Phosphate. Chloride 1. Citrate 8. Unknown 18. cis-aconitate 9. Bromide 19. trans-aconitate 10. Nitrate 20. Unknown Conclusion Newly developed solutions for capillary ion chromatography include high-pressure IC with 4 µm particle columns that enable high chromatographic resolution or fast analysis. Highpressure capillary ion chromatography is ideal for operation at higher flow rates due to the low eluent consumption. In addition, charge detection provides more information, universal detection and similar response for weakly and strongly dissociated ions. Capillary ion chromatography provides the benefits of low eluent consumption and low waste generation, and is ideal for analyses where sample volumes are limited. High pressure capillary ion chromatography using 4 µm particle-size columns provides high efficiency separations of complex samples or fast analysis. Reagent-Free IC systems are ideal for gradient separations, providing consistent results and eliminating the handling of corrosive acids or bases. The new charge detector provides increased information by detecting all ionic analytes, thus simplifying calibration and quantification. References For more information regarding the Dionex ICS-4000 HPIC system, charge detection, and Dionex IonPac CS19-4µm and AS11-HC-4µm columns, refer to following Thermo Fisher Scientific application notes and website. 1. Thermo Fisher Scientific website. Dionex ICS-4000 with Charge Detection, http://www.dionex.com/en-us/products/ion-chromatography/ic-rfic-systems/ics- 4000/lp-11162.html 2. Thermo Fisher Scientific. Dionex Application Note 143, Determination of Organic Acids in Fruit Juices. 3. Thermo Scientific Dionex IonPac CS19-4µm Cation-Exchange Column, http://www.dionex.com/en-us/products/columns/ic-rfic/cation-packed/ionpac-cs19-4um/lp-11100.html 4. Thermo Scientific Dionex IonPac AS11-HC Hydroxide-Selective Anion-Exchange Column, http://www.dionex.com/en-us/products/columns/ic-rfic/hydroxide-selectivepacked/ionpac-as11hc/lp-32.html 6 New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection

www.thermoscientific.com/dionex Thermo Scientific Dionex products are designed, developed, and manufactured under an ISO 9001 Quality System. 2012 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. U.S./Canada (84) 29 00 Brazil () 11 331 140 Austria (43) 1 616 1 2 Benelux (31) 20 683 968 (32) 3 33 42 94 Denmark (4) 36 36 90 90 France (33) 1 39 30 01 10 Germany (49) 6126 991 0 Ireland (33) 1 644 0064 Italy (39) 02 1 62 126 Sweden (46) 8 43 3380 Switzerland (41) 62 20 9966 United Kingdom (44) 126 69122 Australia (61) 2 9420 233 China (82) 2428 3282 India (91) 22 264 23 Japan (81) 6 688 1213 Korea (82) 2 3420 8600 Singapore (6) 6289 1190 Taiwan (886) 2 81 66 PN0144_E 06/12S