Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

Similar documents
Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Method 8326 (0.006 to mg/l Al 3+ ) Powder Pillows

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Quaternary Ammonium Compounds

Mercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right.

Chloramine (Mono) and Nitrogen, Free Ammonia

PAR Method Method to 2.0 mg/l Pb TNTplus 850

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

Chromium, Total. Alkaline Hypobromite Oxidation Method 1 Method to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters)

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

Phenolphthalein and Total Alkalinity Method to 4000 mg/l as CaCO 3 Digital Titrator

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

DR/4000 PROCEDURE NITRATE. 2. The display will show:

DR/4000 PROCEDURE. IRON, Total

Method to 0.50 mg/l NH 3 N Powder Pillows

Nitrogen, Ammonia. Test Preparation. Powder Pillows Method Method 8155 Salicylate Method 1 Powder Pillows

Salicylate Method Method HR (2 to 47 mg/l NH 3 N) TNTplus 832

Method to 0.50 mg/l NH 3 N Powder Pillows

Sulfide DOC Methylene Blue Method 1 (0 0.70, and mg/l) Method Test procedure

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials

PAR Method Method to 2.0 mg/l Pb TNTplus 850

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows

Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters)

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials

Powder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters)

Oxygen Demand, Chemical

Oxygen Demand, Chemical

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV

Chlorphosphonazo Colorimetric Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow. Instrument Adapter Sample cell

Cadion Method Method to 0.30 mg/l Cd TNTplus 852

Persulfate Digestion Method Method to 40 mg/l N (HR) TNTplus 827

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials

Chlorine, Free and Total, High Range

Nitrogen, Total Inorganic

Chromazurol S Method Method to 0.50 mg/l Al TNTplus 848

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

DR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water.

Chlorine, Free and Total, High Range

Chlorine, Total. USEPA DPD Method 1 Method to mg/l as Cl 2 Chemkey Reagents. Test preparation. Before starting.

Trihalomethane Formation Potential (THMFP)

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ )

Dimethylphenol Method Method to mg/l NO 3 N or 1.00 to mg/l NO

Peracetic Acid (PAA) and Hydrogen Peroxide (H 2 O 2 )

CHLORINE, TOTAL (0 to 4.00 mg/l)

Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls)

Steam-Assisted Gravity Drainage Water Analysis Handook

DR/4000 PROCEDURE. HARDNESS, Total

Chlorine, Free and Total, Low Range

Scope and application: For wastewater, seawater, drinking water, surface water and process water.

DR/4000 PROCEDURE SELENIUM. 4. Measure 100 ml of sample into a second 500-mL erlenmeyer flask (label the flask sample ).

Mercaptoacetic Acid Method*

TPH (Total Petroleum Hydrocarbons)

NITRITE, Low Range (0 to mg/l NO 2- -N)

GENERAL INSTRUCTIONS GENERAL PREPARATION

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

Steam-Assisted Gravity Drainage Water Analysis Handbook

Method (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows

Ammonia Gas Sensing Combination Electrode

Sodium Sealed Electrode Sensor Bundle Product Number: ENSOD051

User instructions. Gel filled ORP/Redox electrode: Model MTC10101, MTC10103, MTC10105, MTC10110, MTC10115 or MTC10130

Ammonia Gas Sensing Combination Electrode

DR/4000 PROCEDURE CYANIDE. 1. Press the soft key under HACH PROGRAM. Select the stored program for cyanide by pressing 1750 with the numeric keys.

DR/4000 PROCEDURE LEAD. 2. The display will show: 3. Fill a 250-mL graduated cylinder to the 250-mL mark with sample.

GENERAL INSTRUCTIONS GENERAL PREPARATION

Fluoride Sealed Electrode Sensor Bundle Product Number: ENFLU049

Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter

Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste

Use a pipet bulb when pipeting cyanide solutions, as these solutions are highly toxic.

Standard Operating Procedure for GLNPO Specific Conductance: Conductivity Bridge

Ammonium Sealed Electrode Sensor Bundle Product Number: ENAMN020A

Operating Instructions for Ammonium ISE Specifications

DPD Test N Tube Method *

EXPERIMENT 6. Properties of Buffers INTRODUCTION

DOC /2013, Edition 1. User Manual

DR/4000 PROCEDURE MERCURY. Phase 1: Sample Digestion must be done in a fume hood! Toxic gases may be produced.

CHLORINE, FREE, Ultra-high Range ( mg/l Cl 2 ) Method 10069

NITROGEN, TOTAL, HR, Test N Tube

symphony Multisensor Probes

COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

symphony Reference Electrodes

Potentiometric measurement of ph

The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium ions in aqueous solutions.

C = concentration as ppm sulfide V t = volume of titrant at endpoint V s = volume of standard used (10 ml)

S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) Fax: +44 (0)

COLE-PARMER LABORATORY NITRATE ION ELECTRODE INSTRUCTION MANUAL

1) A ph/mv meter or an ion meter, either line operated or portable.

NITROGEN, AMMONIA, High Range, Test N Tube

PRETREATMENT TECHNICAL DATA SHEET CHROMIUM-FREE ORGANIC PASSIVATING RINSE PRODUCT DESCRIPTION

METHOD 9040B. ph ELECTROMETRIC MEASUREMENT

Tex-620-J, Determining Chloride and Sulfate Contents in Soil

PRETREATMENT TECHNICAL DATA SHEET A CHROME-FREE FINAL RINSE PRODUCT DESCRIPTION

PHS-25CW/3BW Benchtop ph Meter Instruction Manual

Transcription:

, 10255 DOC316.53.01322 Direct Measurement ISE Method Method 10255 3.55 g/l to 35 g/l Cl Powder Pillow ISA Scope and Application: For the determination of high concentrations (1 M) of chloride in brine solutions, produced waters and hydraulic fracturing waters. Test preparation Procedure notes: The instructions in this procedure are specific to the HQd meters. The SensION+ meters can also be used but the navigation and menus will be different. Refer to the meter user manual for meter operation. Refer to the probe user manual for probe maintenance. Set the date and time in the meter before the probe is attached. The probe must have the correct service-life time stamp. If complete traceability is necessary, enter a sample ID and operator ID. Refer to the HQd meter manual for more information. Calibrate the probe regularly for the best measurement accuracy. Refer to Calibrate the probe. Stir the standards at a slow and steady rate to prevent the formation of a vortex. Air bubbles under the probe tip can cause slow response or measurement errors. Gently shake the probe to remove bubbles. Keep the temperature of the calibration standards within ±2 ºC for best results. Between measurements, rinse the probe with deionized water. Blot dry with a lint-free cloth. Items to collect: Description Quantity Beaker, 100-mL, polypropylene 4 Bottle, wash, 500-mL 1 Chloride ion-selective electrode 1 Chloride Ionic Strength Adjustor (ISA) Buffer Powder Pillows 4 Deionized water varies Meter, ion-selective electrode 1 Sodium chloride 11.55 g Stir bars 4 Stirrer, electromagnetic 1 Volumetric flask, 200-mL 3 See Consumables and replacement items for reorder information. Prepare the probe 1. Remove the sensor protection cap from the probe. Page 1 of 6

Set up the meter 2. Rinse the probe with deionized water. Blot dry with a lint-free cloth. 3. Put the probe in a 3.55 g/l chloride standard solution for a minimum of 30 minutes. Refer to Prepare the calibration standards. Change the default settings in the meter for high-range measurements. Save the new settings as a new method. 1. Connect the probe to the meter. 2. Push. Calibrate the probe 3. Go to ISECl181 Settings>Modify Current Settings>Measurement Options. 4. Set the measurement options for high-range measurements: Units g/l Measurement limits set the lower limit to 3 g/l. 5. Go to Calibration Options. 6. Set the calibration options for high-range measurements: Std set custom Calibration units g/l Std set values 3.55, 12.5 and 35 g/l 7. Enter a name for the new settings, for example HR Cl. 8. Push EXIT until the display shows the measurement mode. Prepare the calibration standards Prepare a 3.55 g/l, a 12.5 g/l and a 35 g/l chloride standard solution for calibration. Items to collect: Sodium chloride, NaCl Volumetric flasks (3x), 200 ml, Class A Laboratory balance Deionized water 1. Prepare a 35 g/l chloride standard solution: a. Weigh 11.55 g of sodium chloride. b. Quantitatively transfer the NaCl into a volumetric flask. c. Dilute to the mark with deionized water. Mix fully. 2. Prepare a 12.5 g/l chloride standard solution: a. Transfer 71.43 ml (or g) of the 35 g/l standard solution into a volumetric flask. b. Dilute to the mark with deionized water. Mix fully. 3. Prepare a 3.55 g/l chloride standard solution: a. Transfer 56.8 ml (or g) of the 12.5 g/l standard solution into a volumetric flask. Page 2 of 6

b. Dilute to the mark with deionized water. Mix fully. HR chloride calibration procedure 1. Add 25 ml of the 3.55 g/l, 12.5 g/l and 35 g/l chloride standard solution to three beakers. 2. Add a stir bar and stir at a moderate rate. 3. Add the contents of one Chloride Ionic Strength Adjustment (ISA) Powder Pillow to each standard solution. 4. Rinse the probe with deionized water. Blot dry with a lint-free cloth. 4-7 5. Put the probe into the 3.55 g/l chloride standard solution. Tap the probe to remove any air bubbles. 6. Push Calibrate. The display shows the standard solution value. 7. Push Read. The display shows Stabilizing. When the reading is stable, the display shows the next standard solution value. 8. Repeat steps 4 to 7 for the 12.5 g/l and the 35 g/l chloride standard solutions. 9. Push Done to view the calibration summary. 10. Push Store to accept the calibration. Rinse the probe with deionized water and blot dry. Page 3 of 6

Measure samples Dilute samples that are greater than 35 g/l If the chloride concentration is greater than 35 g/l (1 M), dilute the sample to a lower concentration. Complete the steps that follow to make a 1:10 (10-fold) dilution. 1. Measure 2.5 ml of the sample in a 25-mL graduated cylinder. 2. Add deionized water to the 25-mL line. 3. Pour the diluted sample into a beaker. 4. Measure the concentration with the HR chloride measurement procedure. 5. Multiply the result by 10 to get the concentration of the sample before dilution. HR chloride measurement procedure 1. Add 25 ml of sample to a 50-mL beaker. 2. Add a stir bar and stir at a moderate rate. 3. Add the contents of one Chloride Ionic Strength Adjustment (ISA) Powder Pillow per 25 ml of sample. 4. Rinse the probe with deionized water. Blot dry with a lint-free cloth. 1-6 5. Put the probe into the sample. Tap the probe to remove any air bubbles. 6. Push Read. The display shows Stabilizing and then the chloride concentration of the sample. 7. Repeat steps 1 through 6 for each sample. 8. When done, rinse the probe. Page 4 of 6

Sample collection, preservation and storage Collect samples in clean plastic or glass containers. Samples can be stored for a minimum of 28 days at room temperature. Interferences The sensing element can respond to other ions in addition to chloride and cause a positive error. If Chloride ISA is added to the standards and samples, the effect of interfering ions is minimized. Samples can contain oxidizing agents such as Copper (Cu 2+ ), Iron (ferrous) (Fe 2+ ) and Permanganate (MnO 4 - ). Refer to Table 1. Table 1 Interfering substances Substance Mercury Ions that form insoluble salts of silver Strong reducing solutions Interference Must be absent from samples. Can deposit a layer of salt on the sensing surface and cause probe errors. Can form a surface layer of silver. Consumables and replacement items Required reagents and apparatus Description Quantity/Test Unit Item no. Chloride Ionic Strength Adjustor (ISA) Buffer Powder Pillows 1 100/pkg 2318069 Beaker, 50-mL, polypropylene 1 1 108041 IntelliCAL Chloride Ion Selective Electrode (ISE), 1 m cable 1 1 ISECL18101 Recommended standards Description Unit Item no. Sodium chloride, ACS grade 454 g 18201H Optional reagents and apparatus Description Unit Item no. Bottle, wash, 500 ml 1 62011 Cylinder, graduated, 25-mL, poly 1 108140 Deionized water 4 L 27256 IntelliCAL Chloride Ion Selective Electrode (ISE), 3 m cable 1 ISECL18103 sension+ 9652C Chloride Combination Ion Selective Electrode (ISE), 1 m cable 1 1 LZW9652C.97.002 Volumetric flask, 200 ml, Class A, glass 1 1457445 1 Use with sension+ meters or other meters that have a BNC connection. Page 5 of 6

FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: In the U.S.A. Call toll-free 800-227-4224 Outside the U.S.A. Contact the HACH office or distributor serving you. On the Worldwide Web www.hach.com; E-mail techhelp@hach.com HACH COMPANY WORLD HEADQUARTERS Telephone: (970) 669-3050 FAX: (970) 669-2932 Hach Company, 2012. All rights reserved. Printed in U.S.A. 05/2012, Edition 1