HPLC Praktikum Skript

Similar documents
High performance liquid chromatography

Determination of Caffeine by HPLC

Chemistry 3200 High Performance Liquid Chromatography: Quantitative Determination of Headache Tablets

Chemistry Instrumental Analysis Lecture 31. Chem 4631

High Performance Liquid Chromatography

High Pressure/Performance Liquid Chromatography (HPLC)

Chemistry Gas Chromatography: Separation of Volatile Organics

HPLC. GRATE Chromatography Lab Course. Dr. Johannes Ranke. September 2003

Preparative Separation of Active Components from Natural Products using Low-pressure Gradient Preparative HPLC

Liquid storage: Holds the solvent which is going to act as the mobile phase. Pump: Pushes the solvent through to the column at high pressure.

Instrumental Chemical Analysis

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Draft Method proposal: determination of glucoheptonic acid (HGA) in fertilizers.

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

RESOLUTION OENO 33/2004 DETERMINATION OF SHIKIMIC ACID IN WINE BY HPLC AND UV-DETECTION

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Determination of Trace Cations in Power Plant Waters Containing Morpholine

LC III: HPLC. Originally referred to as High-Pressure Liquid Chromatography. Now more commonly called High Performance Liquid Chromatography

Biochemistry. Biochemical Techniques HPLC

Determination of Polymer Modifier in Asphalt

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Chromatography. Gas Chromatography

Open Column Chromatography, GC, TLC, and HPLC

Assay Transfer from HPLC to UPLC for Higher Analysis Throughput

The Theory of HPLC. Quantitative and Qualitative HPLC

Prep 150 LC System: Considerations for Analytical to Preparative Scaling

Figure 1. Structures for Vitamin B2 and Vitamin B1.

High Performance Liquid Chromatography (HPLC) of Three Analgesics

Fall 2012 Due In Class Friday, Oct. 19. Complete the following on separate paper. Show your work and clearly identify your answers.

High Performance Liquid Chromatography

High Performance Liquid Chromatography

Quantitative Analysis of Caffeine in Energy Drinks by High Performance Liquid Chromatography

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

637. Thiamethoxam. HPLC method

Determination of Volatile Substances Proof of Food Adulteration

Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system

COMPREHENSIVE LC X LC ANALYSIS OF FLAVORING INGREDIENTS IN DISTILLED SPIRITS

Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System


Routine MS Detection for USP Chromatographic Methods

Qualification Specifications for Agilent 1100/1200 HPLC Systems

Dilution(*) Chromatography

LEARNING OBJECTIVES CHEM 212: SEPARATION SCIENCE CHROMATOGRAPHY UNIT. Thomas Wenzel, Bates College. In-class Problem Set Extraction.

Varian Galaxie Chromatography Data System for Preparative HPLC

Experiment UPHPLC: Separation and Quantification of Components in Diet Soft Drinks

Transferring HPLC methods from the HP 1090 Series to the Agilent 1100 Series HPLC system

HPLC. High Performance Liquid Chromatography (HPLC) Harris Chapter 25

Introduction of Hitachi Chromaster 5610 MS Detector for High Performance Liquid Chromatograph

Chromatography- Separation of mixtures CHEM 212. What is solvent extraction and what is it commonly used for?

A RP-HPLC METHOD DEVELOPMENT AND VALIDATION OF PARA- PHENYLENEDIAMINE IN PURE FORM AND IN MARKETED PRODUCTS

Analytical Chemistry

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

H 3 CO H 3 CO S CH 3

TECHNICAL TEMEPHOS. 1. Specification. Full specification WHO/SIT/19.R4 Revised 10 December Description

High Performance Liquid Chromatography

This method describes the identification of the following prohibited colorants in cosmetic products:

Using the Agilent 1290 Infinity II Multicolumn Thermostat with Extreme Temperatures and Flow Rates

Maximizing Performance Through GPC Column Selection

Impact of Instrument Dispersion on Performance of HPLC Capillary Columns

Luminescence transitions. Fluorescence spectroscopy

Basic principles of HPLC

Packed Column for Ultra-Fast Reversed-Phase Liquid Chromatography, TSKgel Super-ODS. Table of Contents

Chromatography and other Separation Methods

LECTURE 2. Advanced Separation Science Techniques Present and Future Separation Tools

LC Technical Information

Abstract: An minimalist overview of chromatography for the person who would conduct chromatographic experiments, but not design experiments.

GPC/SEC Practical Tips and Tricks. Thomas Dent Applications Scientist Agilent Technologies. October, 2011 Gulf Coast Conference

Dionex IonPac AS28-Fast-4µm column

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

CIPAC. CIPAC Free relevant impurities methods:

Chapter 14. Molar Mass Distribution.

British American Tobacco Group Research & Development. Method - Determination of ammonia in mainstream smoke

2501 High Performance Liquid Chromatography

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

for Acclaim Mixed-Mode HILIC-1 Column

Volume 6, Issue 2, January February 2011; Article-015

High Speed 2D-HPLC Through the Use of Ultra-Fast High Temperature HPLC as the Second Dimension

High Performance Liquid Chromatography. Table 1: Allowed HPLC Adjustment of USP <621> and EP <2.2.46>

NEVIRAPINE ORAL SUSPENSION Final text for addition to The International Pharmacopoeia (February 2009)

Fast methods for the determination of ibuprofen in drug products

LC800. Smart HPLC. Until now UHPLC From now Smart HPLC

Introduction to Chromatographic Separations

Introduction to Chromatographic Separations (Chapter 1) Many determinations involve separation followed by analysis chromatography electrophoresis

CHEM Experiment Five Gas Chromatographic Analysis Of A Phenolic Extract. Objectives:

Selective Formation of Benzo[c]cinnoline by Photocatalytic Reduction of 2,2 Dinitrobiphenyl with TiO 2 and UV light irradiation

3590/ Laboratory

Chromatography & instrumentation in Organic Chemistry

Chromatographic Analysis

IDENTIFICATION AND DETERMINATION OF HYDROQUINONE IN COSMETIC PRODUCTS 2 14/11/17 ACM 003 BY TLC AND HPLC

HPLC Columns. HILICpak VT-50 2D MANUAL. Shodex HPLC Columns Europe, Middle East, Africa, Russia

Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

How do you explain HPLC? 2 AZURA Educational System for tomorrow s HPLC professionals

TEPZZ A_T EP A1 (19) (11) EP A1 (12) EUROPEAN PATENT APPLICATION. (51) Int Cl.:

PYRIPROXYFEN TECHNICAL

DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

Packings for HPLC. Packings for HPLC

Transcription:

HPLC Praktikum Skript Assistants: Gianluca Bartolomeo HCI D330, 3 46 68, bartolomeo@org.chem.ethz.ch Sahar Ghiasikhou HCI E330, 2 29 29, ghiasikhou@org.chem.ethz.ch 1. Introduction In chromatographic techniques, the separation of different components is based on different physical or chemical properties of substances, i.e. distribution ratios between two phases, mobility within a phase, etc. A general introduction ("Einführung Chromatographie") can be found at www.analytik.ethz.ch. Please read carefully to get the theoretical bases of the subject, since in order to be able to work efficiently on the test days, a serious preparation is necessary. 2. HPLC device High-Performance Liquid Chromatography (HPLC) is a method of liquid chromatography in which the separation is carried out under high pressure (up to 300 bar at a flow of typically 1 ml / min). The most common stationary phase used is silica gel (normal phase HPLC) or silica gel modified with long alkyl chains (reversed phase HPLC). The particle size is 1.7-5 μm. This allows theoretical separation steps of up to N = 100,000 (According to Van Deemter s equation). A schematic representation of a typical setup is shown in Figure 1: Figure 1: HPLC components scheme.(from Chromedia.nl)

1 - Solvent reservoir: The mobile phase can differ according to the analyte of interest. The most common are water, acetonitrile, methanol. Before being pressurized, solvents must be degassed to avoid pressure instability in the column. 2 Pump: An ideal pump should provide solvent compatibility, constant flow delivery, low dead volume for minimum problems on solvent changeover and convenience of replacement. Three commonly used pump types: syringe type, constant pressure and the one used in the proposed experiment, the reciprocating piston pump. The reciprocating piston pumps deliver solvent(s) through reciprocating motion of a piston in a hydraulic chamber. The solvent deliver of reciprocating pumps is smooth because they have complementary actions; while one pump is in the filling cycle, the other one is in the delivery cycle. High pressure output is possible at constant flowrates and gradient operation is also possible. 3 - Injector: The injection system is positioned after the pump head, and is therefore located on the high-pressure side of the solvent delivery system. The injection of a sample at atmospheric pressure into the system, at high pressure, represents a critical step in the chromatographic process. Sample injection valves, or switching valves, are used to introduce reproducible amounts of sample into the HPLC eluent stream without causing changes in pressure or flow. Injecting directly into the high-pressure flow of mobile phase eliminates the requirement to stop the mobile phase flow, effectively minimizing any disturbance in the chromatographic baseline by allowing an essentially pulse-free change in the direction of the mobile phase flow through the autosampler valve. The sample is introduced using a syringe into the sample loop via a port on the injection valve (Fig. 2). Figure 2. Injection system in load position (left) and inject position (right).

Once the sample is loaded the valve is turned to the inject position. The rotor seal channels move and connect different ports on the stator. The pump is now able to push mobile phase through the sample loop, transferring the sample onto the column. Once the injection has been made the valve is returned to the load position. This enables loop filling for the next injection. Some workers prefer to leave in the inject position until the next injection is required to completely flush the loop and reduce carryover. 4 Column: Columns contain the stationary phase, which is selected according to the physical and chemical properties of the analyte of interest. The most common packing material is silica gel. Different modifications on the silica beads surface (Fig. 3) are required for separation of different analytes. For instance OH terminal groups on the silica surface allow interaction with highly polar compounds, whereas alkyl-modified beads are suitable for non-polar compounds (so-called reversed phase chromatography). In addition, surface area of the stationary phase plays a crucial role on the separation performance. Normally, the particle diameter varies from 1.5 to 5 µm. Although using smaller particles requires higher operational pressure, it provides better separation resolution due to higher surface area. Figure 3. Schematic representation of the different stationary phases: unfunctionalized silica (on top) Vs. C18-modified silica (bottom). 5 Detector: Evaluation of the separation performance is possible through detection of the analytes. The most common detectors are: ultraviolet/visible (UV-Vis), photo-diode array (PDA), and mass spectrometry (MS). 6 Data System: The otucoming data are acquired and saved in digital format using a computer interface. Altough it could be theoretically possible to work with a stand-alone machine, digital data treatment allows to achieve more reliable results, especially in quantification.

7 Waste: Once analyzed, the eluent mixture, possibly containg analytes, is collected and disposed of. 2.1 Experimental apparatus The proposed experiments will be performed on a modular device, manufactured by Shimadzu Corporation (Japan), one of the leading firms in the field. For the sake of convenience, operating conditions are summarized in the following Table 1: Table 1. Operating conditions for experiments Eluent mixture Operating pressure (Max) Injection volume Column type Diameter of silica beads Detector (photo-diode array) range Spectral resolution Water (H2O) and acetonitrile (ACN) in variable proportions 200 bar 20 μl C18-functionalized silica gel 5 μm 190-800nm 2 nm 3. Tasks 3.1 Influence of the solvent composition: Qualitative interpretation of the chromatogram of a homologous phenone series: Procedure: A dilute solution of acetophenone, propiophenone, n-butyrophenone and n-valerophenone (Figure 4) in acetonitrile: water (1: 1) is available (concentration ~ 100 mg / L). Figure 4. From left to right, acetophenone, propiophenone, butyrophenone and valerophenone.

The influence on the separation behavior of the following parameters is investigated: eluent composition (ACN: H2O, 50:50, 80:20, 95: 5) and flow rate (0.5 ml / min, 0.75 ml / min, 1 ml / min).the following combinations should be measured at 250 nm: Table 2. Composition of the separation parameters to be examined. 5 μm column (manual sampler) Solvent ratio Flow rate [ml/min] [ACN : H2O] 0.5 50 : 50 0.75 1 0.5 80 : 20 0.75 1 95 : 5 1 Evaluation: The chromatograms are evaluated by hand and submitted together with the report. On the chromatograms, the reversal tangent (analogous substances) are drawn in and the basic peak width (in cm) is measured and converted into a temporal peak width. Further, the following chromatographic parameters are calculated: retention time, net retention time, retention factor, relative retention, plate number, resolution. The following assumption is made: V0 = 1 ml at a flow rate of 1 ml / min. Discussion: Using the measured chromatographic parameters, discuss the properties of the separation, e.g. influence of eluent mixture, analysis time, separation performance. Determine and justify the optimal parameters for the separation problem. 3.2 Quantitative determination of acetophenone using an internal standard Procedure: An acetophenone and a propiophenone stock solution (both ~ 100 mg/l in acetonitrile: water (1: 1)) are available. The assistant prepares an acetophenone solution with an unknown concentration (= analyte). In order to determine the concentration of acetophenone, three calibration solutions with different acetophenone but constant propiophenone concentrations are prepared by the students in 10 ml volumetric flasks. The three calibration solutions and the analyte are quantified by HPLC. The separation is carried out using the optimum separation parameters determined in Exercise 3.1. All measurements are determined in triplicate.

Table 3. Acetophenone and propiophenone concentrations used. Measurement solution cacetophenon [mg/l] cpropiophenon [mg/l] 1 2 5 2 4 5 3 6 5 Analyte unknown 5 Evaluation: The area integrals of the acetophenone and propiophenone peaks are related to the factor f: A Acetophenon A Propiophenon = f c Acetophenon c Propiophenon The factor f is determined by the evaluation of the chromatograms of solutions 1-3. The figure of f against the concentration of Acetophenone including error bars should be present and the average f should be calculated. By applying this factor to the chromatograms of the analyte, the unknown acetophenone concentration can be determined. Discussion: After determination of the acetophenone concentration, the actual concentration can be announced by the assistant. The report must state the experimental concentration and the actual concentration. Possible sources of error are identified and discussed. 3.3 Quantitative analysis of caffeine in energy drinks or similar drinks Procedure: A 5 point calibration in the range of 0.01-0.1 mg / ml caffeine is made. The three samples taken (coffee without milk and sugar, energy drinks, coke, etc.) are analyzed separately by each student. As a method, a gradient from 10% ACN to 70% ACN is used (see Figure 5). The analysis sequence is worked out together with the assistant. Each sample (including calibration) is performed for three times and detected at a wavelength of 272 nm.

Acetonitrile [%] 100 80 60 40 20 0 0 2 4 6 8 10 12 Time [min] Figure 5. Solvent gradient of caffeine quantification method. The beverages containing caffeine are shaken for degassing, then filtered through a 0.45 μm cellulose filter and diluted by the amount of caffeine (Figure 6) declared by the manufacturer. Care must be taken to ensure that the diluted concentration corresponds to the selected calibration range. Figure 6. Structure formula of caffeine. Evaluation: The calibration curve is calculated by HPLC. The caffeine concentrations of the samples are determined by comparing the area integration with the calibration curve. The dilution factor used can be used to calculate the actual concentration. The measurement error is determined by Student t distribution. Discussion: The measured value should be set in relation to the manufacturer's specification. Potential sources of error should be determined and discussed. Also possible improvements in the analysis protocol are described.