Determination of Benzene, Toluene, Ethylbenzene and Xylene in River Water by Solid-Phase Extraction and Gas Chromatography

Similar documents
Analysis of BTEX in Natural Water with SPME

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

AppNote 2/2000. Stir Bar Sorptive Extraction (SBSE) applied to Environmental Aqueous Samples

Analyst, February 1998, Vol. 123 ( ) 195

METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY

CHEM Experiment Five Gas Chromatographic Analysis Of A Phenolic Extract. Objectives:

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Sulfotepp impurities in Chlorpyrifos EC formulations

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

CHEM340 Tutorial 4: Chromatography

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Application. Gas Chromatography February Introduction

STANDARD OPERATING PROCEDURES

Volatile organic compounds (VOCs):

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

Chromatography. Gas Chromatography

METHOD 3600B CLEANUP

METHOD 8100 POLYNUCLEAR AROMATIC HYDROCARBONS

637. Thiamethoxam. HPLC method

Anethole. Gas chromatograhpy determination of trans-anethole in Spirit srinks of viti-vinicultural origin

OPTIMISATION OF SOLID PHASE MICROEXTRACTION (SPME) CONDITIONS FOR HEADSPACE ANALYSIS OF ORGANOPHOSPHATE PESTICIDES IN WHOLE BLOOD

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

1. A gas chromatograph equipped with flame ionization detection and a computing integrator.

TECHNICAL BRODIFACOUM

Chemistry Gas Chromatography: Separation of Volatile Organics

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Chromatographic Methods of Analysis Section: 5 Gas Chromatography (GC) Prof. Tarek A. Fayed

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

METHOD 3600C CLEANUP

Application Note. Abstract. Authors. Introduction

Course goals: Course goals: Lecture 1 A brief introduction to chromatography. AM Quality parameters and optimization in Chromatography

METHOD 8000A GAS CHROMATOGRAPHY

Determination of trace anions in concentrated hydrofluoric acid

CH 2252 Instrumental Methods of Analysis Unit V Gas Chromatography. M. Subramanian

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

GC Instruments. GC Instruments - Sample Introduction

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

ENVIRONMENTAL analysis

Determination of Pesticides in Aqueous Samples by On-Line Coupling Solid-Phase Extraction to Gas Chromatography with At-Column Concentrating Interface

CHAPTER 6 GAS CHROMATOGRAPHY

Introduction to Gas Chromatography

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

GAS CHROMATOGRAPHY (GC)

Yokogawa Analytical Systems Inc., Nakacho, Musashino-shi, Tokyo , Japan.

Chemical Analysis Problem

Available online Journal of Chemical and Pharmaceutical Research, 2012, 4(6): Research Article

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

Residual solvents analysis using an Agilent Intuvo 9000 GC system

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Abstract: An minimalist overview of chromatography for the person who would conduct chromatographic experiments, but not design experiments.

METHOD 8111 HALOETHERS BY GAS CHROMATOGRAPHY

Method - Determination of aromatic amines in mainstream cigarette smoke

Chapter 27: Gas Chromatography

Chapter 27: Gas Chromatography. Principles Instrumentation Detectors Columns and Stationary Phases Applications

Magnitudes of Back Diffusion During Long-Term Diffusive Sampling of Volatile Organic Compounds Using Carbotrap and Chromosorb 106

Atlantic RBCA. Guidelines for Laboratories. Tier I and Tier II Petroleum Hydrocarbon Methods

Date 10/22/97 Page 1 Revision 4.1 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY METHODS 8000/8100 (MODIFIED) DIESEL RANGE ORGANICS (DRO)

Fractionation of Acidic, Basic, and Neutral Drugs from Plasma with an SPE Mixed Mode Strong Cation Exchange Polymeric Resin (Agilent SampliQ SCX)

Gas Chromatography (GC)

Diquat 1,1 -ethylene-2,2 -bipyridium dibromide salt Paraquat 1,1 -dimethyl-4,4 -bipyridium dichloride salt Initial Preparation

LAMBTON SCIENTIFIC (A Division of Technical Chemical Services Inc.)

Determination of Pharmaceuticals in Environmental Samples

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Real Time On-Site Odor and VOC Emission Measurements Using a znose

SPE AND GC MS INVESTIGATION OF ORGANIC CONTAMINANTS IN ATMOSPHERIC PRECIPITATION

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Supporting information Online HPLC-ESI-HRMS Method for the Analysis and Comparison of Different Dissolved Organic Matter Samples

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.

METHOD 5000 SAMPLE PREPARATION FOR VOLATILE ORGANIC COMPOUNDS

AYOUB ET AL.: JOURNAL OF AOAC INTERNATIONAL VOL. 85, NO. 6,

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Ch24. Gas Chromatography (GC)

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Gas Chromatography. Introduction

Understanding Gas Chromatography

Journal of Chemical and Pharmaceutical Research

Volatile Hydrocarbons (VH) in Waters by GC/FID

AUTOMATED ONLINE IDENTIFICATION AND MONITORING OF IMPURITIES IN GASES

Characterization of Carbon SPE for the Extraction of Polar Analytes

Headspace Technology for GC and GC/MS: Features, benefits & applications

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Determination of Trace Cations in Power Plant Waters Containing Morpholine

Study of Residual Solvents in Various Matrices by Static Headspace

APPLICATION NOTE. A Capillary Approach to ASTM D3606: Test Method for Determination of Benzene and Toluene in Finished Motor and Aviation Gasoline

Date 02/24/96 Page 1 Revison 4.0 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY METHODS 8020/8015 (MODIFIED) GASOLINE RANGE ORGANICS (GRO)

Transcription:

2003 The Japan Society for Analytical Chemistry 1365 Determination of Benzene, Toluene, Ethylbenzene and Xylene in River Water by Solid-Phase Extraction and Gas Chromatography Mohammad A. MOTTALEB,* Mohammad Z. ABEDIN,** and Mohammad S. ISLAM** *Department of Chemistry, University of Rajshahi, Rajshahi 6205, Bangladesh **Department of Chemical Technology and Polymer Sciences, Shahjalal University of Science and Technology, Sylhet 3114, Bangladesh A rapid and reproducible method is described that employs solid-phase extraction (SPE) using dichloromethane, followed by gas chromatography (GC) with flame ionization detection for the determination of benzene, toluene, ethylbenzene, xylene and cumene (BTEXC) from Buriganga River water of Bangladesh. The method was applied to detect BTEXC in a sample collected from the surface, or 5 cm depth of water. Two-hundred milliliters of n-hexane-pretreated and filtered water samples were applied directly to a C 18 SPE column. BTEXC were extracted with dichloromethane and the BTEX concentrations were obtained to be 0.1 to 0.37 µg ml 1. The highest concentration of benzene was found as 0.37 µg ml 1 with a relative standard deviation (RSD) of 6.2; cumene was not detected. The factors influencing SPE e.g., adsorbent types, sample load volume, eluting solvent, headspace and temperatures, were investigated. A cartridge containing a C 18 adsorbent and using dichloromethane gave a better performance for the extraction of BTEXC from water. Average recoveries exceeding 90 could be achieved for cumene at 4 C with a 2.7 RSD. (Received October 1, 2002; Accepted July 17, 2003) Introduction Volatile organic compounds (VOC), e.g., benzene, toluene, ethylbenzene, xylene and cumene (BTEXC), are important environmental contaminants because of their high toxicity and widespread occurrence. They are present in aviation fuel (gasoline) and are widely used as industrial solvents and raw materials for the production of different commodities. 1 Benzene, toluene and ethylbenzene are among compounds designated as priority pollutants by the US EPA, and the action and risk levels of benzene, toluene, ethylbenzene, and xylene are described in the Dutch Government Quality Standards for the Assessment of Soil and Water Contamination. 2 The determination of BTEXC in environmental matrices is difficult because of their trace-level presence and losses incurred during sample handling, extraction etc. Recently, there has been considerable interest in the development of SPE columns for the clean-up, extraction, and pre-concentration of liquid samples. 3 SPE applications are found in different environmental areas, such as soils and sediments, 4,5 industrial effluents 6 and water samples. 7 9 SPE has been used for the extraction of BTEXC 10 and pesticides 11 from soils and sediments. A few SPE applications have also To whom correspondence should be addressed. E-mail: Mottaleb.Mohammad@epamail.epa.gov M. A. M. present address: EPA/NRC Postdoctoral Research Associate, Environmental Chemistry Branch, Environmental Sciences Division, National Exposure Research Laboratory, U.S. Environmental Protection Agency, P.O. Box 93478, Las Vegas, NV 89193-3478, USA. appeared for VOC components in the BTEXC analysis of water samples; a cartridge containing C 18 adsorbent was used to extract BTEX from industrial effluents 6 and benzene and toluene from seawater. 7 Although solid-phase micro-extraction (SPME) was also applied to isolate BTEX from water samples either by direct adsorption from the liquid 8 or via headspace sampling, 9 the limitation of SPME for the quantitation of sulfurbased VOC has been reported. 12 The VOCs analyzed from air and water samples by open-tubular, wall-coated columns, 13 solvent trapping 14 and semi-voc from air by atmosphericpressure chemical ionization mass spectrometry 15 were also described. Buriganga river water is important because it flows through the capital of Bangladesh and has an enormous impact on the socio-economic development of the country, especially industrial and shipping-marine activities. Tanker-washing sewage, shipping scrap particles and oil spillage are common features on the river at different ghats (marine terminals). As a result, the water is continually polluted by various organic compounds, especially hydrocarbons. In addition, aromatic solvents are increasingly used in industry, 1 and the wastes are sometimes disposed of in the aquatic environment, which becomes increasingly contaminated. To monitor contaminants, we recently determined the concentrations of various normal saturated hydrocarbons in Buriganga river water 16 and pesticides in soil 17 by GC, using liquid-liquid extraction and SPE methods, respectively. We also reported on the development of a chromatography method for the determination and characterization of anionic detergents in river water. 18 The present paper describes the concentration levels of benzene, toluene, ethylbenzene, xylene and cumene in water samples collected at two depths from the Sadarghat, one of the biggest

1366 ANALYTICAL SCIENCES OCTOBER 2003, VOL. 19 Table 1 Concentration of a BTEXC standard solution used for GC optimization Solution Concentration of BTEXC standard solution/µg ml 1 Benzene Toluene Ethylbenzene Xylene Cumene Standard 1 8.77 8.67 8.67 8.61 8.61 Standard 2 17.14 17.34 17.35 17.22 17.22 Standard 3 26.31 26.01 26.01 25.83 25.83 Standard 4 35.08 34.68 34.68 34.44 34.44 marine terminals, on the Buriganga river. Also discussed are the recoveries of BTEXC using different adsorbents for SPE columns and factors influencing SPE adsorption, such as the sample load volume, eluting solvent, headspace and temperatures. Experimental Apparatus and reagents A Varian gas chromatograph (Model 3300) equipped with a flame ionization detector (FID) was used in this study. A DB-1 fused-silica mega-bore analytical column (15 m 0.53 mm i.d., 1.5 µm thickness) and a phenyl-methyl deactivated guard column were used. An integrator (Varian Model 4290) gave the peak area and retention time of the peaks separated by GC-FID. Methanol (HPLC grade), hexane, dichloromethane and chloroform were obtained from Merck Ltd. (Germany). VOCs and internal standards were purchased from Sigma Chemicals Ltd., USA. The C 18, C 8 and phenyl (PH) (500 mg, 3 ml) SPE cartridges were obtained from the Supelco Ltd. Stock and working standard solutions Stock solution. An aliquot (10 µl) of each of the BTEXC constituents and internal standard dichlorobenzene (DCB) were dissolved in 100 ml of dichloromethane. The concentrations of the BTEXC components corresponded to 87.7, 86.7, 86.7, 86.1, 86.1 µg ml 1, respectively and the concentration of DCB corresponded to 130.5 µg ml 1. Working standard solutions. Four standard solutions were made for measuring the linearity of the GC response. The concentrations of the standard solutions are given in Table 1. Collection of river water samples Contaminated water samples were collected in 1-l dark glass bottles on September 21, 1999 from the Sadarghat. A map of the Buriganga river and the location of the Sadarghat area have been reported. 16 Cleaned bottles were rinsed with sample water prior to sample collection. Ten-liter samples were collected at the surface and ten samples were also collected at a depth of 5 cm. The distance of the sample collection point from the river bank was about 200 m. Method development work was carried out using double-distilled water containing BTEXC at the same concentrations as those found in the river water. Extraction of VOC from synthetic and river water samples A 10 ml volume of a synthetic sample was extracted with 10 ml of hexane in a 30 ml vial; the layers were allowed to separate. Prior to SPE work, 2 to 4 ml of the organic layer was removed and stored in a sealed glass vial at 4 C. The extraction of a river-water sample was carried out essentially according to a reported method. 16 Each (500 ml) water sample was shaken Fig. 1 GC-FID detection chromatogram for a standard solution of BTEXC, and internal standard (dichlorobenzene). Conditions: gas flow rates: N 2 (carrier gas), 4 ml min 1 ; H 2, 33 ml min 1 ; air, 330 ml min 1 ; detector temperature, 180 C; injector temperature, 180 C; mega-bore, DB-1 fused-silica analytical column; oven temperature program: 35 C for 5 min, to 70 C at 5 C min 1, to 180 C at 15 C min 1 and hold 10 min at 180 C; and sample injected volume, 1 µl. vigorously for 30 min with 50 ml of hexane at 4 C. The aqueous layer was separated and extracted again with 25 ml of hexane. The combined extracts were then stored at 4 C for SPE. Prior to direct use of the water sample in SPE, the sample (200 ml) was filtered by a 0.45 µm nylon membrane. Solid-phase extraction The column was activated with 3 ml 50 methanol and preequilibrated with 3 ml 1 methanol. The river-water (100 ml) sample was loaded on the column at 3 ml min 1. Elution was carried out with 2 portions of 2 ml aqueous 1 methanol. Finally, solutes were eluted with two aliquots of 2 ml of dichloromethane. Similar elution profiles were obtained for recovery experiments. DCB (200 µl or 5) was added as an internal standard prior to a GC analysis. Passing samples through a dryer containing sodium sulfate only eliminated traces of water. Calculation of response factor and concentration of components The relative response factor of a component (R F) to the internal standard of DCB is given by C DCB A DCB R F = Cc, where C DCB and C c represent the concentrations of DCB and the component analyte, respectively, in terms of µg/ml. The terms A DCB and A c indicate the peak-area counts from the integrator for DCB and the component analyte, respectively. The response factors for all components were calculated as mentioned above, and the concentration of each component (C c) was calculated as follows: A c A DCB A c C DCB C c =. RF

1367 Table 2 SPE of river-water samples at 4 C (C 18 column, sample volume 200 ml) Concentration of BTEXC component in river water/µg ml 1 Analyte Average conc. 0 cm depth SD a RSD, b Average conc. 5 cm depth SD a RSD, b Benzene 0.350 ± 0.034 9.8 0.372 ± 0.0230 6.2 Toluene 0.234 ± 0.033 14.3 0.167 ± 0.0289 17.4 Ethylbenzene 0.145 ± 0.0230 15.8 0.104 ± 0.0209 20.2 Xylene 0.124 ± 0.0206 16.1 0.144 ± 0.0135 9.36 Cumene a. SD represents standard deviation, which was calculated from each of three measurements. b. RSD means relative standard deviation. Fig. 2 Calibration curve for optimization of the GC-FID system for the analysis of BTEXC in river-water samples. Concentration of the injected BTEXC solution (please see Experimental, preparation of stock and working standard solutions). The operating conditions are the same as in Fig. 1. Fig. 3 GC-FID chromatograms for the river-water samples. For (A), water was collected from the surface, and for (B), water was collected from 5 cm depth. The operating conditions are the same as in Fig. 1. Results and Discussion GC optimization The GC-FID system used was optimized before the VOC measurement. Separations were achieved with different temperature programs. A good separation of individual BTEXC constituents, including DCB, was obtained under the following conditions: injector temperature, 180 C; column oven temperature, 35 C for 5 min, to 70 C at 5 C min 1, to 180 C at 15 C min 1 and hold at 180 C for 10 min. The detector temperature was 180 C and the flow rate of the nitrogen carrier gas was 4 ml min 1. Figure 1 is a GC-FID chromatogram of a standard solution of the BTEXC components, showing that the components and DCB were well-resolved. The linearity of the detector response was also demonstrated by injecting the working standard solutions into the GC-FID instrument. Figure 2 depicts the calibration graphs of the peak area versus the concentration. The FID gave good linearity of the response for the detection of each of the BTEXC constituents. Hence, it was decided that the above conditions could be used for the determination of BTEXC from the river-water samples. Presence of VOC in water samples Table 2 summarizes the concentration of BTEXC in the riverwater samples analyzed by SPE-GC-FID. The presence of ethylbenzene, xylene and cumene was not detected using the experimental conditions described previously, although trace levels of benzene and toluene were found. To detect the other constituents of the BTEXC family, an increased volume of 200 ml of water sample was directly applied to the SPE at 4 C. An appreciable amount of benzene and toluene, including trace levels of ethylbenzene and xylene, were obtained in the riverwater sample. However, cumene was never found. Figure 3 shows a representative GC-FID chromatogram of the riverwater samples. The chromatograms in Figs. 3(A) and 3(B) correspond to the surface and 5 cm depth of water, collected from the Buriganga river. These were obtained when 2 µl SPE eluted samples were injected into the GC and showed a similar chromatographic elution pattern with different magnitudes of the BTEX components peak. Blank experiments were performed prior to sample injection. Selection of adsorbents and eluting solvent To select the suitability of adsorbents and eluting solvents, the percentage recovery of BTEXC constituents was investigated using C 18, C 8 and PH cartridges with CH 2Cl 2 and CHCl 3 solvents. The recoveries obtained when 2 ml portions of standard VOC solutions were passed through different SPE columns and eluted with two portions of 2 ml of dichloromethane or chloroform at 4 C. The recovery results are presented in Table 3. The extractions were performed simultaneously for each solvent. Regardless of the solvents used, higher recoveries were obtained for xylene and cumene. This may have been due to a less evaporative loss of the two components because of their higher boiling point. Moreover, in comparison between CH 2Cl 2 and CHCl 3 solvents, it was observed that slightly better recoveries were obtained when CH 2Cl 2 was used as the eluting solvent (Table 3). This is probably due to a more non-polar interaction between a bonded

1368 ANALYTICAL SCIENCES OCTOBER 2003, VOL. 19 Table 3 Recovery of BTEXC from different SPE columns eluted with CH 2Cl 2 and CHCl 3 at a temperature of 4 C Solvent SPE column Recoveries () ± standard deviation (SD) a Benzene Toluene Ethylbenzene Xylene Cumene CH 2Cl 2 C 18 76.0 ± 2.5 80.5 ± 2.5 83.7 ± 3.5 87.1 ± 4.0 90.3 ± 3.3 C 8 70.0 ± 3.5 68.3 ± 2.0 74.6 ± 3.5 83.7 ± 1.5 79.6 ± 1.4 PH 73.0 ± 1.5 77.8 ± 3.5 70.4 ± 4.0 76.3 ± 1.5 82.5 ± 2.3 CHCl 3 C 18 74.0 ± 3.0 78.0 ± 2.5 80.0 ± 2.0 84.0 ± 3.0 87.0 ± 4.0 C 8 69.0 ± 3.5 67.0 ± 2.5 72.0 ± 3.0 80.0 ± 2.5 77.0 ± 3.0 PH 71.2 ± 3.0 76.0 ± 3.0 69.0 ± 4.0 75.0 ± 3.0 78.0 ± 2.5 a. SD values were calculated from each of five measurements. Table 4 Effect of the temperature on the recovery of BTEXC with CH 2Cl 2 using a C 18 cartridge Analytes At a temperature of 20 C Recovery, SD a RSD b, At a temperature of 4 C Recovery, SD a RSD b, Benzene 76 ± 3.0 3.9 79 ± 2.8 3.5 Toluene 82 ± 2.9 3.5 86 ± 2.5 2.9 Ethylbenzene 84 ± 3.3 3.9 91 ± 3.2 3.5 Xylene 87 ± 3.5 4.0 88 ± 2.7 3.1 Cumene 89 ± 2.9 3.3 91 ± 2.5 2.7 a. SD represents standard deviation, which was calculated from each of five measurements. b. RSD means relative standard deviation. Fig. 4 Effect of the sample load volume on the recoveries of the BTEXC constituents with a C 18 SPE column using dichloromethane solvent. phase and the CH 2Cl 2 system. Effect of the sample load volume The effect of the sample volume on the SPE recovery is one of the most important factors, because the SPE performance is affected by the amount of sample volume loaded on a particular column. 6,10 The break-through volume of a C 18 column (500 mg, 3 ml capacity) was determined by passing a number of VOC standard solutions to a volume of up to 500 ml. Known masses of the analytes were introduced. There were no appreciable changes in the recovery rates up to a sample volume of 400 ml. The percentage recovery of benzene and toluene decreases more rapidly than ethylbenzene, xylene and cumene. This observation confirms the fact that ethylbenzene, xylene and cumene possess higher breakthrough volumes than benzene and toluene (Fig. 4). Similar break-through volume curves were obtained for C 8 and phenyl substituted SPE columns. These are not shown. Factors affecting the SPE performance investigations Temperature. To investigate the effect of the temperature on SPE performance, a cartridge containing C 18 material was employed with CH 2Cl 2 solvent at temperatures of 20 C and 4 C. An effect of temperature on the SPE recovery of BTEXC constituents was found to occur (Table 4). It has been observed that both liquid-liquid and solid-phase extractions provided slightly better recoveries when experiments were carried out at Table 5 Effect of the headspace on the recovery of BTEXC with CH 2Cl 2 using a C 18 cartridge at 4 C Analyte Headspace No headspace Recovery, SD a Recovery, SD a Recovery difference, Benzene 67.0 ±2.0 76.0 ±3.0 11.8 Toluene 72.0 ±3.5 81.0 ±2.5 11.1 Ethlbenzene 74.0 ±3.5 84.0 ±2.5 11.9 Xylene 79.0 ±2.7 87.0 ±4.0 9.2 Cumene 82.0 ±2.8 90.0 ±3.5 8.9 a. SD represents the standard deviation, which was calculated from each of five measurements. a temperature of 4 C. The relative standard deviations (RSD) were calculated for each of the BTEXC constituents and temperatures. At 20 C, the RSD of recoveries were between 3.3 and 4.0; however, at 4 C improved recoveries of the constituents were achieved with RSD values of 2.7 to 3.5. Headspace. In order to investigate the effect of headspace, recovery experiments for each of the BTEXC components were performed using a C 18 column in an environment at a temperature of 4 C. This temperature provided recoveries of 79 to 91 of the constituents (Table 4). Parallel extractions of the components with and without headspace were performed. The results are presented Table 5. It can be seen that the incurred losses of the BTEXC constituents occurred in the range of 9 to 12 due to allowing a headspace. Since trace levels of BTEXC components could be found in water, or other real samples, the above losses (9 12) are quite significant. Thus, it was decided that the existence of a headspace increases the

1369 evaporative losses, which decreases the percentage recovery. This study was conducted while allowing no headspace in the sampling and at a temperature of 4 C. Acknowledgements The authors would like to thank the Bangladesh University grants commission for financial support to carry out this research. References 1. J. J. G. Klist, in Chemistry and Analysis of VOCs in the Environment, ed. H. J. T. Bloemen and J. Burn, 1993, Glasgow. 2. Environmental Quality Standards for Soil and Water, Netherlands Ministry of Housing, Physical Planning and Environment, 1991, Leidschendam. 3. L. A. Berrueta, B. Gallo, and F. Vicente, Chromatographia, 1995, 40, 474. 4. P. Loconto, LC-GC Int., 1991, 4, 10. 5. Z. Zhang and J. Pawliszyn, Anal. Chem., 1995, 67, 34. 6. I. S. Deans, C. M. Davidson, D. Littlejohn, and L. Brown, Analyst, 1993, 118, 1375. 7. W. A. Saner, J. R. Djadamec, R. W. Sager, and T. J. Kleen, Anal. Chem., 1979, 51, 2180. 8. C. L. Auther, J. Pawliszyn, and P. R. Belardi, J. High. Res. Chromatogr., 1992, 15, 741. 9. Z. Zhang and J. Pawliszyn, Anal. Chem., 1993, 65, 1843. 10. K. M. Meney, C. M. Davidson, and D. Littlejohn, Analyst, 1998, 123, 195. 11. M. J. Redondo, M. J. Ruiz, B. Boluda, and G. Font, Chromatographia, 1993, 36, 147. 12. R. A. Murray, Anal. Chem., 2001, 73, 1646. 13. B. C. D. Tan, P. J. Marriott, H. K. Lee, and P. D. Morrison, Analyst, 2000, 125, 469. 14. M. A. Stone and L. T. Taylor, Anal. Chem., 2000, 72, 1268. 15. L. Charles, L. S. Riter, and R. G. Cooks, Anal. Chem., 2001, 73, 5061. 16. M. A. Mottaleb, M. Ferdous, M. S. Islam, and M. A. Hossain, Anal. Sci., 1999, 15, 995. 17. M. A. Mottaleb and M. Z. Abedin, Anal. Sci., 1999, 15, 283. 18. M. A. Mottaleb, Mikrochim. Acta, 1999, 132, 31.