Dosimetry: Electron Beams

Similar documents
Reference Dosimetry for Megavoltage Therapy Beams: Electrons

ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS *

Optimization of microionization chambers for small-field reference dosimetry

factors for NE2561 ionization chambers in 3 cm x 3 cm beams of 6 MV and 10 MV photons

ERRATA LIST AND UPDATES TO IAEA TRS-398 (2000) 1

The EPOM shift of cylindrical ionization chambers - a status report Hui Khee Looe 1, Ndimofor Chofor 1, Dietrich Harder 2, Björn Poppe 1

High-Energy Photon Beam Therapy Dosimetry with Ionisation Chambers

Referensdosimetri. Crister Ceberg Medical Radiation Physics Lund University Sweden

Monte Carlo Simulation concerning Particle Therapy

7. a XV-2 high spatial resolution lm detector (Kodak). Important parameters of these detectors are given in Table1. The ionization chambers and the di

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry

Chapter 9: Calibration of Photon and Electron Beams

Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water

NACP-02 perturbation correction factors for the NPL primary standard of absorbed dose to water in high energy electron beams

Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical Physics University of Wisconsin

Comparison between TG-51 and TRS-398: Electron Contamination Effect on Photon Beam Quality Specification.

Instrumentation for Verification of Dose

8/2/2012 UPDATING TG-51. When will it end? Part 1 - photon addendum. What are these updates? Photons: Electrons: More widespread revision required

) for Varian TrueBeam high-dose-rate therapy beams

An introduction to IAEA TRS-483

Updating reference dosimetry a decade after TG-51

M [scale units/s] of the system

Instrumentation for Verification of Dose

CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR

Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2

Chapter 10 Acceptance Tests and Commissioning Measurements

STANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS

FOREWORD In 1987 the IAEA published a Code of Practice entitled Absorbed Dose Determination in Photon and Electron Beams: An International Code of

Principles of applied dosimetry - illustrated by ionometry. Lesson FYSKJM4710 Eirik Malinen

Determination of Ambient Dose Equivalent at INFLPR 7 MeV Linear Accelerator

Study of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration

A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators

Code of Practice for the Absorbed Dose Determination in High Energy Photon and Electron Beams

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137

Metrological traceability and specific needs in: - IR measurement for radiation protection (RP) - IR measurement for radiotherapy (RT)

Small Field Dosimetric Measurements with TLD-100, Alanine, and Ionization Chambers

KCDB-Identifiers: EURAMET.RI(I)-K1.2 / EURAMET.RI(I)-K4.2 / EURAMET.RI(I)-K5.1. Pilot laboratory: BEV (AT) Andreas Steurer 1), Wilhelm Tiefenboeck

Use of a radioactive check device for redundancy check of ionization chambers

Chapiter VII: Ionization chamber

COMPARISON OF ABSORBED DOSE TO AIR CALIBRATION FACTORS FOR A PARALLEL PLATE IONIZATION CHAMBER*

General Overview of Gas Filled Detectors

N. E. Ipe*, K. E. Rosser, C. J. Moretti, J. W. Manning and M. J. Palmer

Efficiencies of Some Spherical Ion Chambers in Continuous and Pulsed Radiation: A Numerical Evaluation

Gerhard Stucki, Sandor VörösV

Determination of absorbed dose to water for 60Co by the scaling theorem. M. Boutillon and A.-M. Perroche (1)

A Thesis. entitled. Micro Ionization Chamber. Vidheesha Arora. Master of Science Degree in. Biomedical Sciences: Medical Physics

General characteristics of radiation dosimeters

Radiation Dosimetry. Electron interactions with matter. Important processes in radiotherapy. Contents. Alun Beddoe

Recombination factor dependancy of high and low dose pulsed accelerators for electron beam energies

IAEA TRS May 2001 (V.10A) PUBLISHED BY THE IAEA ON BEHALF OF IAEA, WHO, PAHO, AND ESTRO INTERNATIONAL ATOMIC ENERGY AGENCY IAEA

Air Kerma Primary Standard: Experimental and. Simulation Studies on Cs-137. J. Cardoso, L. Santos, C. Oliveira

Simulation Modeling in Dosimetry

ABSTRACT. Keywords: Megavoltage, dosimetry, TG51 protocol, TG21 protocol, parallel-plate chambers, crosscomparison. INTRODUCTION

B E A M Q U A L I T Y S P E C I F I C A T I O N I N K I L O-V O L T A G E R A D I O T H E R A P Y

Implementation of the IAEA-AAPM Code of Practice for the dosimetry of small static fields used in external beam radiotherapy

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved

ABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS

City University of Hong Kong

High dose-per-pulse electron beam dosimetry A model to correct for the ion recombination in the Advanced Markus ionization chamber

Ionization Chamber. Pocket dosimeter

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT

ABSORBED DOSE IN ION BEAMS: COMPARISON OF IONISATION- AND FLUENCE-BASED MEASUREMENTS

ELECTRON INTERACTIONS

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements

4.1b - Cavity Theory Lecture 2 Peter R Al mond 2011 Overview of Lecture Exposure (W/e)air Exposure Exposure and and and Air Air Kerma

NPL s progress towards absorbed dose standards for proton beams

Ionizing Radiation Dosimetry and Medical Physics

Comment on Proton beam monitor chamber calibration

Electron beam water calorimetry measurements to obtain beam quality conversion factors

Characterization of heavy charged particle fields using fluorescent nuclear track detectors

Chapter V: Cavity theories

Calibration of a pencil ionization chamber with and without preamplifier

PRIMARY STANDARDS of AIR KERMA for 60 CO and X-RAYS & ABSORBED DOSE in PHOTON and ELECTRON BEAMS. Malcolm McEwen

Current and Recent ICRU Activities in Radiation Protection Dosimetry and Measurements

Investigation of the standard temperature- pressure correction factor at low x-ray energies

Design, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors

Manipulation on charged particle beam for RT benefit.

Evaluation of different methods for determining the magnitude of initial recombination in ionization chambers

Composite field dosimetry

Small field dosimetry

Radiation Quantities and Units

Calibration of Radioprotection Instruments and Calibrated Irradiation: Characterization of Gamma Beam of 137 Cs and 60 Co

Radiation Dose, Biology & Risk

Ranges of Electrons for Human Body Substances

Small Field Dosimetry and IAEA/AAPM Protocol

ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER

Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory

R. C. Tailor and W. F. Hanson. Department of Radiation Physics. The University of Texas, M.D. Anderson Cancer Center. Houston, Texas 77030

Measurement of induced radioactivity in air and water for medical accelerators

Problem P7. Stéphanie Ménard. Dosimetry Department Fontenay-aux FRANCE IRSN QUADOS IRSN

Chapter 8 Electron Beams: Physical and Clinical Aspects

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

Progress in calculations of k Q for TG-51

Determination of contributions of scatter and distance error to the source strength of 192 Ir HDR brachytherapy source

Monte Carlo commissioning of photon beams in medical LINACS using wide-field profiles in a water phantom

Dosimetry and beam calibration

Measurements and comparisons for data of small beams of linear accelerators

Georgia Institute of Technology. Radiation Detection & Protection (Day 3)

Dosimetry. Introduction. General. CHAPTER 1

Accuracy, Calibration, Type Testing and Traceability General. Peter Ambrosi

Transcription:

ICTP SChool On MEdical PHysics For RAdiation THerapy: DOsimetry And TReatment PLanning For BAsic And ADvanced APplications 13-24 April 2015 Miramare, Trieste, Italy Dosimetry: Electron Beams G. Hartmann EFOMP & German Cancer Research Center (DKFZ) g.hartmann@dkfz.de

Content: 1. Dosimetry equipment 2. Calibration procedure 3. Correction factors 4. The radiation quality correction factor k Q : Determination & Calculation 5. Depth of measurement: at reference depth & at depth of maximum dose 6. Cross calibration

1. Dosimetry Equipment Ionization chambers Types of chambers used: q Cylindrical (also called thimble) chambers are used in calibration of: (Orthovoltage x-ray beams) Megavoltage x-ray beams Electron beams with energies of 10 MeV and above air-filled measuring volume central electrode

1. Dosimetry Equipment Ionization chambers Types of chambers used: q Parallel-plate (also called end window or plane-parallel) chambers are used : for the calibration of superficial x-ray beams for the calibration of electron beams with energies below 10 MeV for dose measurements in photon beams in the buildup region and surface dose front electrode air-filled measuring volume guard ring back electrode

1. Dosimetry Equipment Ionization chambers Cylindrical Chambers Plane Parallel Chambers Farmer-Chamber Roos-Chamber

1. Dosimetry Equipment

1. Dosimetry Equipment Electrometer, ioniation camber and radioactive check source

1. Dosimetry Equipment Electrometer plus connectors From the PTW Catalogue: "Ionizing Radiation Detectors" "The following overview of connecting systems facilitates the identification of a variety of adequate connectors"

1. Dosimetry Equipment Phantoms Water Phantoms Solid Phantoms

1. Dosimetry Equipment Phantoms q q q Pease note: Water is always recommended in the IAEA Codes of Practice as the phantom material for the calibration of megavoltage photon and electron beams. The phantom should extend to at least 5 cm beyond all four sides of the largest field size employed at the depth of measurement. There should also be a margin of at least 5 g/cm 2 beyond the maximum depth of measurement except for medium energy X rays in which case it should extend to at least 10 g/cm 2.

1. Dosimetry Equipment Phantoms for measurements Solid (plastic) phantom: Please note: In spite of their increasing popularity, the use of plastic phantoms is strongly discouraged for reference measurements. In general such measurements are responsible for the largest discrepancies in the determination of absorbed dose for most beam types.

1. Dosimetry Equipment Phantoms for measurements Solid (plastic) phantom: Several disadvantages because a plastic phantom requires: scaling of depth: z w = z pl c pl scaling of dosimeter reading M Q,pl : M Q = M Q,pl h pl where c pl is a depth scaling factor h pl is a fluence scaling factor

1. Dosimetry Equipment Phantoms for measurements Values from TRS 398 for c pl and h pl Note: The high uncertainty associated with h pl is the main reason for avoiding the use of plastic phantoms.

2. Calibration procedure General formula D = M N k w,q Q D,w,Q Q,Q o o M QO N DwQ,, O k Q,Qo is the chamber reading in beam of quality Q and corrected for influence quantities to the reference conditions used in the standards laboratory. is the water dose calibration coefficient provided by the standards laboratory for reference beam quality Q o. is a factor correcting for the differences between the reference beam quality Q o and the actual user quality Q.

2. Calibration procedure Positioning of the ionization chamber in water Positioning can be defined as the adjustment of the reference point of a chamber with respect to the measuring depth. Positioning of the reference point of a cylindrical chamber according to the International Code of Practice of the IAEA, TRS 398: Beam calibration Purpose Depth dose measurement Co-60 at measuring depth 0.6 r deeper than measuring depth HE photons at measuring depth 0.6 r deeper than measuring depth HE electrons 0.5 r deeper than measuring depth 0.5 r deeper than measuring depth

2. Calibration procedure Positioning of the ionization chamber in water Positioning of the reference point of a plane parallel chamber according to the International Code of Practice of the IAEA, TRS 398: Co-60 HE photons HE electrons Purpose Beam calibration Depth dose measurement always at measuring depth

Positioning for high energy electrons depth of measurement plane-parallel chamber cylindrical chamber

3. Correction factors q If the chamber is used under conditions that differ from the reference conditions, then the measured charge must be corrected for the influence quantities by socalled influence correction factors k. q The three most import correction factors are: k T,P for air density k pol for polarity effects k sat for missing saturation effects

4. The beam quality correction factor q Frequently, the reference quality Q o used for the calibration of ionization chambers is the cobalt-60 gamma radiation and the symbol k Q is then normally used to designate the beam quality correction factor: = =

Determination of radiation quality correction factor k Q Beam quality index

Determination of the quality index for HE electrons Definition of the quality parameter Q for HE photons The quality parameter used for megavoltage electron beam specification is commonly based upon the half-value depth in water, R 50 The unit of R 50 is gcm -2

Determination of the quality index for HE electrons Definition of the quality parameter Q for HE photons according TRS 398: R 50 is measured with - a constant SSD of 100 cm - a field size at the phantom surface of at least 10 cm x 10 cm for R 50 7 g cm -2 (E 0 < 16 MeV) at least 20 cm x 20 cm for R 50 > 7 g cm -2 (E0 16 MeV).

Determination of the quality index for HE electrons Measurement of R 50 : Problem: The measurement with an ionization chamber yields an ionization-depth curve (dose in air), not a dose-depth curve. Dose in water would be: D P = D s p w( ) air wair, s w air, however, is dependent on energy, and hence on the depth PDD 100 80 60 40 20 0 depthionization curve depth / cm 18 MeV depthdose curve 0 2 4 6 8 10 12 14 16

Determination of the quality index for HE electrons Solution of this problem: The half-value of the depth-dose distribution in water R 50 can be obtained directly from measured depth ionization curves using: R 50 = 1.029 R 50,ion - 0.06 g cm -2 (R 50,ion 10 g cm -2 ) R 50 = 1.059 R 50,ion - 0.37 g cm -2 (R 50,ion > 10 g cm -2 ) As an alternative to the use of an ionization chamber, other detectors (for example diode, diamond, etc.) may be used to determine R 50. In this case the user must verify that the detector is suitable for depthdose measurements by test comparisons with an ionization chamber at a set of representative beam qualities.

The values k Q tabulated in TRS 398 have been obtained by calculation. Calculation of k Q, 0 0 0, w air w air Q Q Q Q Q Q Q W s p e k W s p e =

5. Reference depth for HE electrons A further reference condition for HE elctrons: k Q The values of are valid only if the calibration measurement is performed at the reference depth z ref z ref is energy dependent, and obtained by: z ref = 0.6 R 50-0.1 g cm -2 (R 50 in g cm -2 ) This depth is close to the depth of the absorbed-dose maximum z max at beam qualities R 50 < 4 g cm -2 (E 0 <10 MeV), but at higher beam qualities is deeper than z max.

Absorbed dose at z max for HE electrons Frequently, the basic output for an electron beam is wanted to be obtained at z max. This again requires the determination of a depth dose curve. A depth dose curve has to be converted from a measured depth ionization curve. The conversion is performed by multiplying the depth ionization curve with the depth dependent water to air stopping power ratio adjusted to the beam quality of the electron beam.

Absorbed dose at z max for HE electrons This is the depth dependent water to air stopping power ratio adjusted to the beam quality of the electron beam: Δ a + bx + cx + dy s ( ) wa, z = 2 3 1 + ex + fx + gx + hy 2 with x = ln(r 50 /cm), and y = z / R 50 a = 1,0752 b = -0,50867 c = 0,08867 d = -0,08402 e = -0,42806 f = 0,06463 g = 0,003085 h = -0,1246

6. Cross calibration in electron beams Concept q What is a cross-calibration of an ionization chamber? Cross-calibration refers to the calibration of a user chamber by direct comparison in a suitable user beam against a reference chamber that has previously been calibrated. A particular example is the cross-calibration of a plane-parallel chamber for use in electron beams against a reference cylindrical chamber calibrated in 60 Co gamma radiation. Despite the additional step, such a cross-calibration generally results in a determination of absorbed dose to water using the plane-parallel chamber that is more reliable than that achieved by the use of a plane-parallel chamber calibrated directly in 60 Co The main reason is: problems associated with the p wall correction for plane-parallel chambers in 60 Co, entering into the determination of k Q, are avoided.

Uncertainty of Calibration for High Energy Electrons (from the International Code of Practice TRS 398) cylindrical plane-parallel

6. Cross calibration in electron beams Cross-calibration procedure q The highest-energy electron beam available should be used; E o > 16 MeV is recommended. Note: This is now the calibration quality! q The reference chamber and the chamber to be calibrated are compared by alternately positioning each at the reference depth z ref in water q The calibration factor N in terms of absorbed DwQ,, cross dose to water for the chamber under calibration, at the cross-calibration quality Q cross, is then given by: ref M x Qcross ref ref = DwQ,, x DwQ,, Q, Q M Q N N k cross o cross o cross x

6. Cross calibration in electron beams Cross-calibration procedure Such equations require some exercise for reading. ref M x Qcross ref ref = DwQ,, x DwQ,, Q, Q M Q N N k cross o cross o However, when applied to an example, they can be translated cross Example: chamber to be cross calibrated: plane-parallel Roos chamber cross calibrated against: cylindrical Farmer chamber cross calibration performed at an electron energy of 18 MeV Farmer M Roos 18MeV Farmer Farmer Dw,,18MeV = Roos Dw, 18MeV M18MeV N N k

6. Cross calibration in electron beams Cross-calibration procedure q Subsequent use of a cross-calibrated chamber The cross-calibrated chamber with calibration factor may be used subsequently for the determination of absorbed dose in a user beam of quality Q using the basic equation: The values for x x x wq, = Q DwQ,, QQ, D M N k x QQ, k k x QQ, x QQ, cross cross = cross are derived using the procedure: x Q k x k QQ, int x Q, Q cross int x DwQ,, cross where k and k are given in TRS 398, Table 19. int cross, Q int cross N

Summary: Beam Calibration of Electron Beams TRS 398 1) Cylindrical chambers are used in the calibration of electron beams at energies of 10 MeV and above; Parallel-plate chambers are used below 10 MeV 2) The mother of any calibration equation is: 3) The most important correction factors to be applied to the measured charge are: k T,P for air density k pol for polarity effects k sat D = M N k w,q Q D,w,Q Q,Q for missing saturation effects o o

Summary: Beam Calibration of Electron Beams TRS 398 4) Quality correction factors are tabulated in TRS 398. k Q can be calculated as: k Q = W e W e Q s Q pq Q 0 s w, air Q 0 w, air p Q 0 5) Measurement have to be performed at energy dependent reference depths: z ref = 0.6 R 50-0.1 g cm -2 (R 50 in g cm -2 ) 6) Cross calibration is used for plane-parallel chambers in electron dosimetry to reduce the uncertainty of the resultant absorbed dose to water