LANTHANUM BROMIDE SCINTILLATION DETECTOR FOR GAMMA SPECTROMETRY APPLIED IN INTERNAL RADIOACTIVE CONTAMINATION MEASUREMENTS

Similar documents
ENERGY PEAK STABILITY WITH COUNT RATE FOR DIGITAL AND ANALOG SPECTROMETERS

*Corresponding author,

A Prototype of LaBr3:Ce in situ Gamma-Ray Spectrometer for Marine Environmental Monitoring

Pulse height non-linearity in LaBr 3 :Ce crystal for gamma ray spectrometry and imaging

Preparatory experiments for cold-neutron induced fission studies at IKI

Quality Assurance. Purity control. Polycrystalline Ingots

REFERENCE SOURCES FOR THE CALIBRATION OF THE AUTOCORRELATION SINGLE-CRYSTAL SCINTILLATION TIME SPECTROMETER

COUNT RATE EFFECTS ON PEAK SHAPE AND THROUGHPUT IN DIGITAL AND ANALOG GAMMA RAY SPECTROMETRY

Calibration of a Whole Body Counter and In Vivo measurements for Internal Dosimetry Evaluation in Chile, Two years experience.

Measurement of activity of the predominant gamma-emitting radionuclides in activated components of a medical cyclotron plant

WM2014 Conference, March 2 6, 2014, Phoenix, Arizona, USA

Workshop on Understanding and Evaluating Radioanalytical Measurement Uncertainty November The Romanian underground laboratory

arxiv: v2 [physics.ins-det] 8 Feb 2013

Measurement of induced radioactivity in air and water for medical accelerators

EXPERIMENTAL DETERMINATION OF THE URANIUM ENRICHMENT RATIO

Non-Destructive Assay Applications Using Temperature-Stabilized Large Volume CeBr Detectors 14277

Alpha-Gamma discrimination by Pulse Shape in LaBr 3 :Ce and LaCl 3 :Ce

WM2018 Conference, March 18-22, 2018, Phoenix, Arizona, USA. PVT and LaBr3(Ce)-based Radon Express Analyzers 18164

Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the WBC

COMPARATIVE STUDY OF PIGE, PIXE AND NAA ANALYTICAL TECHNIQUES FOR THE DETERMINATION OF MINOR ELEMENTS IN STEELS

Sensitivity of the IRD whole-body counter for in vivo measurements in the case of accidental intakes

Nuclear Lifetimes. = (Eq. 1) (Eq. 2)

A coincidence method of thorium measurement

A gamma-gamma coincidence spectrometer for nuclear attribution and safeguard applications

Activities at the Laboratory of the Nuclear Engineering Department of the Polytechnic University of Valencia

Determination of the activity of radionuclides

Collimated LaBr 3 detector response function in radioactivity analysis of nuclear waste drums

Gamma-Spectrum Generator

Renewed whole-body counting chamber in STUK

RADEAGLET. Lightweight Handheld Radioisotope Identification Device

IdentiFINDER Digital Hand Held Spectrometer & Dose Rate Meter for Portable Applications

New Scintillators for the Border Monitoring Equipment

ELECTRONIC TECHNIQUES IN TIMING MEASUREMENTS FOR NUCLEAR STRUCTURE

Characterization of the Lanthanum Chloride Scintillation Detector

Sample Spectroscopy System Hardware

ɣ-radiochromatography

Recent advances and future perspectives of gamma imagers for scintimammography

List of Nuclear Medicine Radionuclides. Nuclear Medicine Imaging Systems: The Scintillation Camera. Crystal and light guide

Use of phoswich detector for simultaneous monitoring of high energy photon and its applications in in vivo lung counting

CURRENT CAPABILITIES OF THE IRD-CNEN WHOLE BODY COUNTER FOR IN VIVO MONITORING OF INTERNALLY DEPOSITED RADIONUCLIDES IN HUMAN BODY

MCNP analysis and optimization of a double crystal phoswich detector

GAMMA RAY SPECTROSCOPY FOR ARTIFICIAL CONTAMINATION EFFECTS EVALUATION IN LUMINISCENCE DATING OF ARTEFACTS FROM LOW DEPTH LAYERS IN SOUTHERN ROMANIA *

Efficiency. Calculations for Selected Scintillators. > Detector Counting. Efficiency. > Photopeak Efficiency of Various Scintillation Materials

THE COMPTON EFFECT Last Revised: January 5, 2007

RADON EQUILIBRIUM MEASUREMENT IN THE AIR *

Introduction to Environmental Measurement Techniques Radioactivity. Dana Pittauer 1of 48

Basic hands-on gamma calibration for low activity environmental levels

C2-05: Creation of national standards for some emerging pharmaceutical radionuclides to ensure the radioprotection of patients and medical staffs

Timing resolution measurement of a 3 Lanthanum Bromide detector

GROSS ALPHA, GROSS BETA AND GAMMA ACTIVITIES IN BOTTLED NATURAL MINERAL WATER FROM ROMANIA

Gamma background measurements in the Boulby Underground Laboratory

The SpecMAT project: An array of gamma-ray detectors around an active gas target

Gamma-ray Spectroscopy with LaBr 3 :Ce Scintillator Readout by a Silicon Drift Detector

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION

Monte-Carlo Simulation of Response Functions for Natural Gamma-Rays in LaBr 3 Detector System with Complex Borehole Configurations

Dual Isotope Imaging with LaBr3:Ce Crystal and H8500 PSPMT

Interactive Web Accessible Gamma-Spectrum Generator & EasyMonteCarlo Tools

arxiv: v3 [nucl-ex] 12 Jan 2012

SEQUENTIAL DETERMINATION OF AMERICIUM, PLUTONIUM AND URANIUM IN LIQUID EFFLUENTS FROM NUCLEAR POWER PLANTS

Uncertainty and detection limit in decontamination measurements of gamma emitting radio nuclides in unshielded environments Tom Koivuhuhta

STUDIES ON THE AEROSOL RADIOACTIVITY LEVEL AND AIR QUALITY AROUND NUCLEAR RESEARCH INSTITUTE AREA

Energy Calibration of Liquid Scintillation Counter to allow Semi-Qualitative Nuclide Identification in Water Samples

A. Faanu *, G. Emi-Reynolds, E. O. Darko, R. Awudu, E.T. Glover, O. K. Adukpo, F. 1

Using the LaBr 3 :Ce scintillation detector for mobile γ-spectrometry. Master of Science dissertation. Jonas Nilsson

Radiometric Measurements of Environmental Radioactivity

WM2013 Conference, February 24 28, 2013, Phoenix, Arizona, USA

X ray Imaging Android [XIA]

THE ANNUAL EFFECTIVE DOSE FROM NATURAL RADIONUCLIDES SOIL SURFACES OF UZHGOROD AREA

Gamma Spectroscopy. References: Objectives:

Radiation Detection and Measurement

Neutron Response of LaBr 3 :Ce and CsI:Na Phoswich Crystal Scintillators for PARIS

Characterization and Monte Carlo simulations for a CLYC detector

Engineering Physics, Tsinghua University, Beijing , China

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

Scintillation Detector

Measurement of Specific Activities of Some Biological Samples for Some Iraq Governorates

Michael G. Stabin. Radiation Protection and Dosimetry. An Introduction to Health Physics. 4) Springer

Acoustics and Ionising Radiation Formulation and Strategy. 13 November 2008 Alan DuSautoy

Developing DRF for 1 X1 LaBr 3 Detector. Sangkyu Lee

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies AUTOMATED QA/QC CHECK

Fundamentals of Radionuclide Metrology

Politehnica University of Bucharest, Romania, Received July 31, 2015

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

Single Channel Beta-Gamma Coincidence Detection of Radioactive Xenon Using Digital Pulse Shape Analysis of Phoswich Detector Signals

4- Locate the channel number of the peak centroid with the software cursor and note the corresponding energy. Record these values.

ß + RADIATION. Choose your application of. chromatography controlled. chemical purity TLC PET GABI. PET GABI 2x2 Nal (Tl) 0 1V output.

arxiv:nucl-ex/ v2 21 Jul 2005

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

Ultra High Quantum Efficiency PMT for energy resolution measurements of LaBr 3 (Ce) scintillation crystals

IAEA-SM-367/10/04/P SCREENING AND RADIOMETRIC MEASUREMENT OF ENVIRONMENTAL SWIPE SAMPLES

Radiation Dose, Biology & Risk

ORTEC AN34 Experiment 10 Compton Scattering

Physics Dept. PHY-503-SEMESTER 112 PROJECT. Measurement of Carbon Concentration in Bulk Hydrocarbon Samples

RPR 29 CYCLOTRON RADIOCHEMISTRY LABORATORY

High-efficiency and fast-timing scintillators

LISTA DE PUBLICATII A PROIECTULUI: PN-II-ID-PCE ,

Protection against environmental ionizing (mainly gamma)

High Purity Germanium Detector Calibration at ISOLDE

QuantumMCA QuantumNaI QuantumGe QuantumGold

Response function and linearity for high energy γ-rays in large volume LaBr3:Ce detectors

Transcription:

U.P.B. Sci. Bull., Series A Vol. 73, Iss. 4, 211 ISSN 1223-727 LANTHANUM BROMIDE SCINTILLATION DETECTOR FOR GAMMA SPECTROMETRY APPLIED IN INTERNAL RADIOACTIVE CONTAMINATION MEASUREMENTS Mirela Angela SAIZU 1, Gheorghe CATA-DANIL 2 Apariţia de noi tipuri de detectori cu scintilatori anorganici LaBr 3 (Ce) şi LaCl 3 (Ce), a generat necesitatea testării lor pentru diferite aplicaţii. Ca o alternativă a detectorului NaI(Tl), utilizat, în mod tradiţional, în sistemele gama de monitorizare a I-131, în tiroidă, a fost considerat pentru caracterizare, un detector LaBr 3 (Ce), de dimensiuni 1.5 x 1.5. Rezultatele au fost evaluate, urmarindu-se conturarea unui raspuns referitor la potrivirea unui astfel de detector pentru detecţia I-131 în tiroidă. Au fost obţinute spectre comparative pentru cei doi detectori, NaI(Tl) si LaBr 3 (Ce), în aceeaşi geometrie de măsură. By invention of the new types of nuclear detectors based on inorganic scintillators, LaBr 3 (Ce ) and LaCl 3 (Ce), it appeared the opportunity they to be tested in different applications. As an alternative to the traditional NaI(Tl) scintillation detector, used in gamma spectrometric systems for I-131 monitoring in thyroid, in the present paper, it was considered for testing, a 1.5 x 1.5 LaBr 3 (Ce) detector. Spectra were measured and analysed for two detectors, NaI(Tl) and LaBr 3 (Ce) of comparative size, in the same measuring geometry. Discussion of the results and conclusions on the suitability of the new detectors for I-131 monitoring are presented. Keywords: inorganic scintillators, thyroid monitoring, gamma spectrometry 1. Introduction In vivo monitoring of the I-131 retention in thyroid is performed by gamma spectrometric measurements in the Human Monitoring Laboratory-Whole Body Counter from IFIN-HH for people working in nuclear medicine department from hospitals and in radiopharmaceuticals production activities [1-3]. The accuracy of these measurements is important for radiation protection evaluations, in the process of estimation of the committed effective doses and radiological risk. For this reason, a sensitive detector appropriate to detect photons with energies of several hundreds kev of I-131 is necessary, in conditions of good shielding, 1 Senior Researcher, Life Sciences and Environmental Physics Department, National Institute for Physics and Nuclear Engineering Horia Hulubei (IFIN-HH), Măgurele, Romania, Ph.D.Student, University POLITEHNICA of Bucharest, e-mail: saizu_ang@yahoo.com 2 Prof. Physics Department, University POLITEHNICA of Bucharest and National Institute for Physics and Nuclear Engineering Horia Hulubei, Măgurele, Romania

12 Mirela Angela Saizu, Gheorghe Cata-Danil suitable collimation and calibration. The energy spectrum of I-131 decay contains the gamma lines of 8.2keV, 284.3keV, 364.5keV, 637keV and 723keV from which the 364.5keV one is dominant due to its high yield, of.812[4]. Taking into account the characteristics of I-131 energy spectra, room temperature detectors, namely, scintillation detectors based on NaI(Tl) crystals, with good efficiency and reasonable resolution, were considered appropriate for iodine monitoring in thyroid. These detectors were traditionally used, until now, in thyroid assessment gamma spectrometry systems [3]. In the last years, a new type of scintillation detector material LaBr 3 (Ce), available in small sizes, with better resolution, became attractive for different applications, as high energy physics research, medical imaging, non-destructive testing, geological exploration, etc. [5-6]. The present paper presents, according to our knowledge, the first attempt to explore the suitability of the LaBr 3 (Ce) detector for in vivo measurement of I- 131 activity retention in thyroid, by evaluating the performances of such a detector, in comparison with a classic NaI(Tl) detector, of comparative size. 2. Characteristics of LaBr 3 (Ce) detectors The properties that make the LaBr 3 (Ce) scintillator detector attractive for different applications based on gamma-ray spectrometry are: - very good energy resolution - very fast light output decay, enabling high count rate applications - high temperature stability - high gamma detection efficiency - operation at room temperature - promising technology for manufacturing crystal at larger sizes These properties compared to those of the NaI(Tl) scintillator detector are presented in Table 1 (data from Ref.[5]). Table 1 Comparison between the characteristics of LaBr 3 (Ce) and NaI(Tl) scintillators of comparative size. Ref.[5] Parameter LaBr 3 (Ce) NaI(Tl) Crystal density (g/cm 3 ) 5.29 3.67 Thickness for 5% attenuation of 662 1.8 2.5 kev gammas (cm) Light yield (photons/mev) 63 39 1/e decay time (ns) 26 25 Temperature coefficent of light output <.2 -.3 (%/C between C and 5C) Resolution 3-4% 6-7%

Lanthanum bromide scintillation detector [...] radioactive contamination measurements 121 Of particular relevance for the present work is the fact that LaBr 3 (Ce) scintillator crystal has an important peculiarity given by the presence of La- 138, a long-lived naturally occurring radioisotope of lanthanum. The La-138 has a radioactive half-life (T 1/2 ) of 1.5x1 11 years and it acts as a gamma and beta self-contaminant of the scintillator crystal. Its specific decay is presented in Fig.1 [4]. La-138 T 1/2 =1.35 x 1 11 y EC (66.4 %) 1436.8 kev β - (33.6%) 788.7 kev Ce-138 Ba-138 Fig.1 The La-138 decay scheme from Ref.[4]. See text for details. La-138 decays by EC-electron capture (66.4%) to Ba-138 and by betaminus decay (33.6%) to Ce-138., both being stable radioisotopes. During the electron capture decay of the La-138 to Ba-138, is produced the gamma ray of 1435.8 kev (1%) and the Ba K x-ray peak, at the 34.7 kev. The beta decay of La-138 to Ce-138, contributes itself with a beta continuum background with the endpoint energy of 255 kev and with the gamma ray of 788.7 kev (1%). This detector background radioactivity acts as a priory limiting factor in using this scintillation detector in low background measurements. It is the goal of the present paper to investigate this issue in details. 3. Experimental results The experimental system we used for the measurements was a standard gamma spectrometry one, based on a 1.5 x 1.5 LaBr 3 (Ce) detector, type BriLanCe-38 manufactured by Saint-Gobain Crystals[7]. Its associated electronics include the phototube high voltage power supply 556 ORTEC module, the CANBERRA 25 Preamplifier and analog and digital signal processing NIM

122 Mirela Angela Saizu, Gheorghe Cata-Danil modules. The data acquisition was performed using the dedicated multichannel analyzer emulation software, ORTEC- Gamma Vision 32 v-6 [8-11]. Radioactive standard point sources of Am-241(59.5keV), Ba- 133(32.9keV, 356keV), Cs-137(661.7keV) and Co-6(1173.2keV, 1332.5keV) were used for the calibration of the spectrometer, in the energy range of 3 kev 16 kev. The pulse height spectrum obtained with these calibration sources is shown in Fig.2. 4 LaBr3(Ce) detector 3 Counts 2 1 1 2 3 4 5 6 7 8 9 1 Energy [kev] Fig.2 Mixed radioactive source calibration spectra. See text for details. The energy calibration curve (Fig.3) can be described by a second order polynomial function. The excellent characteristics of the obtained fitting curve, namely, a correlation coefficient between experimental data and the output of the fitting curve, R=1, reveal the good linearity of the LaBr 3 (Ce) spectrometer in the energy range we have considered. The nominal energy resolution at the Cs-137 line energy of 661.7keV was measured for the LaBr 3 (Ce) detector and it was found a value of 3.3%. This result confirms the excellent resolution of lanthanum bromide scintillation detectors comparing with the value of 6.8% we measured for a 2 x 2 NaI(Tl) scintillation detector. This significant difference in energy resolutions between the two types of scintillation detectors is shown, in detail, in Fig.4. The energy resolution of LaBr 3 (Ce) detector is a characteristic that is considered for the evaluation of its suitability for the monitoring of I-131 in thyroid. The most important aspect on this evaluation is the effect of the La-138 decay lines on the background spectra, taking into account its decay scheme.

Lanthanum bromide scintillation detector [...] radioactive contamination measurements 123 LaBr3(Ce) detector Energy [kev] 18 16 14 12 1 8 6 4 2 E[keV] = 2E-5ch 2 + 1.397ch + 3.1995 R 2 = 1 1 2 3 4 5 6 7 8 9 1 Channels Fig.3 Energy calibration curve. R 2 is R-Square correlation coefficient between experimental data and the output of a fitted curve 6 5 4 Counts 3 2 1 NaI(Tl) LaBr3(Ce) 55 6 65 7 75 Energy [kev] Fig.4 Cs-137 spectra (detail) for LaBr3(Ce) and NaI(Tl) detectors In a qualitative analysis of the background spectra obtained with the thyroid gamma spectrometric system based on the 1.5 x 1.5 LaBr 3 (Ce) detector, in the energy interval 3keV- 16keV, there are two dominant peaks. The first peak, at the energy of 34.7keV, is due to the BaX-rays associated with the EC process. The second peak is obtained by the overlap of the 1435.8keV gamma ray,

124 Mirela Angela Saizu, Gheorghe Cata-Danil produced during the electron capture decay of the La-138 to Ba-138, with the 146.8keV gamma ray of K-4 from the natural background. In the same background spectra there are two energy intervals with specific characteristics. In the first one, 3 kev-7kev, could be distinguished the gamma rays of Ac-228, Tl-28, Pb-214 and Bi-214 from Th-232 and U-238 natural decay series. The second one has the energy range of 75 kev-15kev in which the background spectra shape is due to the gamma ray of 788.7keV in coincidence with a beta particle. From these two intervals, the first one, 3 kev 7 kev, is of interest for our purpose, taking into account the value of 364.5 kev of the main energy gamma line of I-131 used for the estimation of activity by spectrometric methods. LaBr3(Ce) detector 1 Counts 1 Bkg I-131 1 325 425 Energy [kev] Fig.5 Background and I-131 pulse height spectra measured with LaBr3(Ce) detector. See text for details A quantitative analysis of the pulse height background and I-131 spectra (Fig.5) was made in order to evaluate the suitability of the LaBr 3 (Ce) to be used for the detection of I-131 in thyroid. From the background spectra were determined the count rates (cps) for the energy interval 3keV -7keV and for the interval 325keV 425keV, corresponding to the region of interest of the I-131 peak. The pulse height spectra of I-131 was obtained with a volume source of 2ml solution of I-131 inserted in ORTEC neck phantom and placed at 1 cm distance from the LaBr 3 (Ce) detector[1]. Both spectra were considered to determine the Minimum Detectable Activity [MDA] for I-131 in the measuring conditions

Lanthanum bromide scintillation detector [...] radioactive contamination measurements 125 simulating the real monitoring conditions and thyroid geometry. Also, it was calculated the resolution of the LaBr 3 (Ce) detector at 364.5keV, the I-131 energy line of interest. All the results were declared in Table 2, together with those obtained, in the same measurement conditions, for the 2 x 2 NaI(Tl) scintillation detector. The uncertainties due to counting statistics are given in terms of one standard deviation (1σ). Table 2 Results obtained from background and I-131 spectra of LaBr 3 (Ce) and NaI(Tl) scintillators of comparative size. Parameter LaBr 3 (Ce) NaI(Tl) cps in the energy interval (3keV-7keV)- 9.17 ±.3cps 4.91 ±.2cps background spectra cps in the ROI of 364.5 kev I-131 peak 1.223 ±.4 1.114 ±.3 background spectra MDA [Bq] for 364.5keV of I-131 2.2 ±.1 15.1 ±.1 Resolution at 364.5 kev [%] I-131 spectra 3.7 ±.1 8.7 ±.1 It is noticed that the value of cps, in the energy interval (3keV-7keV), from LaBr3(Ce) background spectra is higher than that obtained with NaI(Tl) detector. This is due to the contribution to the background of the natural La-138 radioisotope, present in the Lanthanum crystal. The values of cps for the energy interval corresponding to the region of interest of the 364.5keV of I-131 peak present a slight difference. This result is important because it reveals that the energy interval considered is not affected, by the La-138 radionuclide with a consistent contribution. Also, we observe the high difference between the cps for the whole energy interval (3keV-7keV) and that corresponding to the 364.5keV of I-131 peak. This is due to the presence of natural radionuclides, namely, the Bi-27, Ac-228, Tl-28, Pb-214 and Bi-214, with important branching ratios in the energy interval 45keV-6keV. As expected, the different values for the MDA [Bq], calculated according to the formula from Ref.[12], at the 364.5keV line of I-131, for both detectors, demonstrate the dependence of the MDA on the background. The resolution of LaBr3(Ce) detector at the energy 364.5keV is, consistently, higher than the resolution of the NaI(Tl) detector. 6. Conclusions The scintillation detectors remain a very good choice of the room temperature detectors to be used for in vivo measurements. Tests were made on a 1.5 x 1.5 LaBr 3 (Ce) detector, using a standard NIM associated electronics. The results revealed the presence of the important effect of the long-lived naturally occurring radioisotope of lanthanum La-138 on the background spectra. Taking

126 Mirela Angela Saizu, Gheorghe Cata-Danil into account our specific application, there were determined relevant features of the detector, in the energy interval 325keV 425keV, corresponding to the region of interest of the 364.5keV line of I-131. They were compared with the same characteristics determined for a NaI(Tl) scintillation detector of comparative size, considered as a common room-temperature detector used for iodine monitoring in thyroid. From the quantitative analysis of the background spectra, for both detectors, there were obtained valuable results. These consist in similar values for cps in the energy interval of interest and for the MDA corresponding to the 364.5keV energy of I-131. The excellent energy resolution of the LaBr 3 (Ce) detector, significant higher than that of the NaI(Tl), was confirmed. Our detailed evaluation is based on promising results for our peculiar application. As a conclusion, we can recommend the lanthanum bromide scintillation detectors for in vivo measurement of I-131 activity retention in thyroid, by gamma spectrometry. Acknowledgements The work has been funded by the Sectoral Operational Programme Human Resources Development 27-213 of the Romanian Ministry of Labour, Family and Social Protection through the Financial Agreement POSDRU/6/1.5/S/19. R E F E R E N C E S [1] Mirela Saizu, NIPNE-HH Whole Body Counting Laboratory-Eight Years of Notified Activity. Statistics On Monitoring Data, Romanian Reports in Physics, Vol.61. Nr.4, (29), 662-668 [2] Mirela Saizu, Quality Assurance Programme Implemented in the Whole Body Monitoring Laboratory from IFIN-HH Romania, Radiation.Protection.Dosimetry Advanced Access, doi: 1.193/rpd/ncq372, 21 [3] Mirela Saizu, Quality Assurance Applied in Internal Dosimetry for Intake Estimation in a Case Study of I-131 Ingestion using In-Vivo Assessment Data, Proceedings of Third European IRPA Congress, Helsinki, Finland, 14-18 June 21 [4] A.A. Sonzogni, Nuclear Data Sheets 98, 515 (23) [5] A.Iltis, M.R.Mayhugh, P.Menge, C.M.Rosza, O.Selles and V.Solovyev, Lanthanum halide scintillators: Properties and applications, Nucl.Instr.Meth. A vol. 563 (26) 359-363 [6] A.Favalli, H.C.Mehner and F.Simonelli, Wide energy range efficiency calibration for a lanthanum bromide scintillator detector, Radiation Measurements vol. 43 (28) 56-59 [7] Saint-Gobain Crystals and Detectors, 12345 Kinsman Road Newberry, OH 4465, USA http://www.detectors.saint-gobain.com [8] *** CANBERRA Product Catalog, 21 [9] *** CANBERRA Genie TM 2 3.1, Operation Manual, 26 [1] *** ORTEC Product Catalog, 21 [11] *** ORTEC Maestro -32, Software User s Manual, 29 [12] ] *** ORTEC Gamma Vision V6 Users Manual, 21