HPLC Analysis of Non-volatile Analytes Using

Similar documents
HPLC Analysis of Nonvolatile Analytes Using Charged Aerosol Detection

Selecting Detectors for Compounds with No Optical Absorbance

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD)

Determination of the Composition of Natural Products by HPLC with Charged Aerosol Detection. Introduction. Black Cohosh. Corona Detector Parameters

Charged Aerosol Detection: Factors for consideration in its use as a generic quantitative detector

Evaporative Light Scattering Detector (ELSD)

No Chromophore? No Problem! Universal HPLC Detection. The world leader in serving science

Agilent 385-ELSD Evaporative Light Scattering Detector

Utilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography. Tips and Techniques to Optimize ELSD and MS based Purification

Agilent 385-ELSD Evaporative Light Scattering Detector

High Pressure/Performance Liquid Chromatography (HPLC)

Charged Aerosol Detection 101

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

penta-hilic UHPLC COLUMNS

Quantitation of Surfactants in Samples by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Varian 385-LC. ELSD for Low Molecular Weight Compounds

Agilent 1260 Infinity Evaporative Light Scattering Detector (ELSD)

Varian 380-LC. ELSD for Routine HPLC and GPC Analysis. Key Benefits

Routine MS Detection for USP Chromatographic Methods

A Novel LC-ELSD Method for the Quantification of 2-Deoxy D-Glucose Using HILIC Chromatography

penta-hilic UHPLC COLUMNS

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

Cation Exchange HPLC Columns

Chemistry Instrumental Analysis Lecture 37. Chem 4631

Selecting an LC/MS Interface Becky Wittrig, Ph.D.

Introduction to Pharmaceutical Chemical Analysis

Analysis - HPLC A.136. Primesep 5 µm columns B.136

Introduction. Chapter 1. Learning Objectives

Address correspondence to M.C. Allgeier. HPLC effluent. Nebulizer. Nitrogen. Heated drift tube. Data acquisition. Analyte.

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution

Effects of Aqueous Sample Content and Aqueous Co-Solvent Composition on UPC 2 Separation Performance

for Acclaim Mixed-Mode HILIC-1 Column

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

Use of Charged Aerosol Detection as an Orthogonal Quantification Technique for Drug Metabolites in Safety Testing (MIST)

LC-MS Based Metabolomics

Pharmaceutical Analysis of API and Counter Ions in Complex Formulations in a Single Injection

MODERN HPLC FOR PRACTICING SCIENTISTS

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Enhancement of Linearity and Response in Charged Aerosol Detection

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA

Advantages of polymerbased. an alternative for ODS

Discover what you re missing

Ch.28 HPLC. Basic types of Liquid Chromatography Partition (LLC) Adsorption (LSC) Ion Exchange (IC) Size Exclusion (SEC or Gel Chromatography)

Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application

The Theory of HPLC. Quantitative and Qualitative HPLC

with diode array detection

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Polymer analysis by GPC-SEC. Technical Note. Introduction

GPC/MS. Analysis of PET

C18 Column. Care & Use Sheet

Supporting Information

Thermo Scientific. Acclaim Trinity P2. Column Product Manual. P/N: October Part of Thermo Fisher Scientific

HPLC Praktikum Skript

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

TANDEM MASS SPECTROSCOPY

Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Online Reaction Monitoring of In-Process Manufacturing Samples by UPLC

Corona Veo the Next Generation Charged Aerosol Detector

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

Tips & Tricks GPC/SEC: Quantify and Get More Than Molar Mass Averages

Luminescence transitions. Fluorescence spectroscopy

Developing a fast, generic method for rapid resolution LC with quadrupole MS detection. Application Note

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

Charged Aerosol Detection and Evaporative Light Scattering Detection Fundamental Differences Affecting Analytical Performance

Assay Transfer from HPLC to UPLC for Higher Analysis Throughput

Chromatography. Gas Chromatography

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Open Column Chromatography, GC, TLC, and HPLC

DEMONSTRATING SUPERIOR LINEARITY: THE ACQUITY UPLC PHOTODIODE ARRAY DETECTOR

Luna 2.5 µm C18(2)-HST. Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Saccharides and Organic Acids

CHEM 429 / 529 Chemical Separation Techniques

Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS

Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Live Webinar : How to be more Successful with your ACQUITY QDa Detector?

A RP-HPLC METHOD DEVELOPMENT AND VALIDATION OF PARA- PHENYLENEDIAMINE IN PURE FORM AND IN MARKETED PRODUCTS

M M e M M H M M H. Ion Sources

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

Value of Eclipse Plus C18 and ph Selectivity to Analyze Amine-Containing Drugs

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Analysis of pharmaceuticals using HPLC with evaporative light scattering detection

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Acclaim HILIC-10 Columns

Welcome to our E-Seminar: Choosing HPLC Columns for Faster Analysis Smaller and Faster

HPLC. High Performance Liquid Chromatography (HPLC) Harris Chapter 25

Direct Analysis of Surfactants using HPLC with Charged Aerosol Detection

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Shodex TM ODP2 HP series columns

Introduction of Hitachi Chromaster 5610 MS Detector for High Performance Liquid Chromatograph

Transcription:

HPLC Analysis of Non-volatile Analytes Using Charged Aerosol Detection Paul H. Gamache, Ryan S. McCarthy, Scott M. Freeto, Darwin J. Asa, Matthew J. Woodcock, Katherine Laws and Roderic O. Cole ESA Inc., Chelmsford, Massachusetts, USA Millennium Pharmaceuticals Inc., Cambridge, Massachusetts, USA A new detection method based upon aerosol charging was examined for its applicability and performance with high performance liquid chromatography (HPLC). Our results demonstrate universal detection of nonvolatile analytes with response magnitude that is independent of analyte chemical properties, four orders of magnitude dynamic range, low nanogram, lower limits of detection and % relative standard deviation response variability. Broad applicability was demonstrated for a range of methods including those using gradient elution, reversed phase, hydrophilic interaction, and ion chromatography; normal and narrow bore column formats; and in combination with other detectors (for example, UV detectors, evaporative light-scattering detectors and mass spectrometers). There continues to be strong demand for improvements in sensitivity, selectivity, throughput, qualitative content and many other performance characteristics of high performance liquid chromatography (HPLC). A major need is for methods that provide both universal detection and quantitative analysis. It is widely recognized that no single HPLC detector is capable of distinguishing all possible analytes from a given chromatographic eluent and the term universal is often used to describe detection of a diverse range of analytes. A primary goal for most analyses that seek universal detection is to obtain a consistent relationship between the magnitude of response and quantity injected for a range of analytes. This consistency of response factors allows the use of global mathematical relationships to estimate quantity (for example, use of parent drug response factor to quantify metabolites and degradants). This characteristic is useful in many applications, where it is impractical or impossible to use individual standards to calibrate the response for each analyte such as drug library QC, pharmaceutical impurity testing, complex lipid analyses and many applications requiring mass balance assessment. Currently, performing such an analysis is difficult with available technologies. LC GC Europe 8(6) (00)

Several techniques are employed for universal detection that are based upon measuring bulk properties or generic attributes among diverse analytes. These techniques include refractive index (RI), low-wavelength UV, evaporative light-scattering detection (ELSD) and chemiluminescent nitrogen detection. These techniques are in contrast to those geared towards selective detection, which is typically based upon measurement of specific properties or reactivity of analytes (for example, fluorescence at a specific wavelength, oxidation at a specific potential and mass-to-charge ratio [m/z]). It should be noted that for most detectors, selectivity and universality are mutually exclusive. Universal detectors often require very selective sample prepurification and analytical separation techniques for analysis of biological matrices. Various characteristics of commercially available instruments employed for universal detection have been compared previously. RI, while widely used, has significant limitations in sensitivity and is not compatible with gradient elution. Lowwavelength UV provides higher sensitivity and improved gradient compatibility. However, detection is limited to s, and response magnitude depends upon molar absorbtivity, which can vary by orders of magnitude even among analogous structures. Although UV detection remains the primary technique for many HPLC analyses, it is unable to see compounds that lack a sufficient UV such as many underivatized amino acids, carbohydrates, lipids, polymers, surfactants, drug substances and natural products. ELSD has become widely used alone or to complement absorbance and mass spectrometry (MS) detectors. ELSD can see compounds lacking a provided they have low volatility. Also, ELSD response magnitude is often less dependent on analyte chemical properties than MS or UV. ELSD, however, has significant limitations in precision, sensitivity, dynamic range and the nature of calibration curves.,, Chemiluminescent nitrogen detection is a relatively new technology for nitrogen-containing molecules and, while less universal, can detect a broad range of pharmaceutical compounds. It has been reported that this detection technique can provide a more linear response and lower limit of detection (LOD) than ELSD, but with poorer precision, higher maintenance and no compatibility with nitrogen-containing mobile phases such as those with acetonitrile and triethylamine. The demand for improved methods continues to drive novel approaches toward universal HPLC detection such as those involving condensation nucleation light scattering and inductively coupled plasma MS. 6 Figure : Conceptual diagram of CAD. Gas inlet HPLC Split Liquid inlet Column Nebulizer Impactor Reagent gas stream Analyte stream Droplets Excess ions trapped Large droplets waste Particles Ions Corona charger Charged analyte particles A new HPLC detection method has been developed based upon charged aerosol detection (CAD). This technique is fundamentally different from that of other detectors and is based upon the coupling of HPLC with widely used electrical aerosol analyser technology.,7 0 A similar HPLC detection method, termed aerosol charge detection, was described by Dixon and Peterson. The detection principle involves the charging of aerosol particles via corona discharge with subsequent electrometer-based measurement and thus has some commonality with atmospheric-pressure chemical ionization (APCI) MS. However, CAD operates by detecting charged particles that have a selected range of mobility rather than by measuring individual gas-phase ions that are differentiated based upon m/z. Furthermore, previous studies have shown that the signal obtained with electrical aerosol technology depends primarily upon particle size across a wide range and does not depend significantly upon individual analyte properties. 9 This more generic principle can thus theoretically provide consistent interanalyte response factors and complement atmospheric pressure ionization MS techniques such as electrospray and APCI. Described in the following sections is our initial evaluation of this technique. We tested the general applicability of CAD using various HPLC systems and modes of separation and for a diversity of analytes. Particular emphasis was on detection of Table : Peak area response for two CAD devices from flow injection analysis of 0 µg of each compound with 0% aqueous methanol as mobile phase and diluent. Values are the average from three injections. Compound Response Response Comment CAD # CAD # Albumin.90.87 MW 6000 Da Amitriptyline..9 Weak base Caffeine.88.9 Weak base Dextrin.8.78 Polymer, no UV Estradiol.87.9 Neutral Fructose.9.0 Polar, no UV Glucose.99.0 Polar, no UV Glutathione.96.0 Zwitterion Homocysteine.97.07 Zwitterion Lactose.89.0 Polar, no UV Methionine.09.7 Zwitterion Naphthol Not detected Not detected Neutral, volatile Nortriptyline.0. Weak base Propranolol.6. Weak base Salicylic acid..6 Weak acid Sucrose.98.0 Polar, no UV Umbelliferone.9.00 Neutral Filter electrometer Signal output Range.8.0.78. Mean.87.96 % RSD 6.% 8.% www.lcgceurope.com

non-s and to assess dependence of response upon analyte chemical properties and chromatographic variables. We also evaluated CAD for its use, in parallel, with MS and photodiode-array detection for high-throughput analysis of diverse small-molecule pharmaceutical library compounds using typical methods for quality control and for support of synthesis. Experimental A diagram of the CAD device is shown in Figure. The detection method involves pneumatic nebulization of HPLC column eluent, exclusion of the largest droplets via an impactor and ambient T solvent evaporation to form an aerosol of analyte particles. A turbulent jet of positive ions, formed by passing a secondary stream of gas by a corona discharge and through an orifice, is mixed with the opposing analyte aerosol stream. 8 In this process, charge is transferred diffusionally to analyte particles. Excess positive ions are trapped by a weak electric field and charged particles impinge on a conductive filter, which transfers the charge to an electrometer for signal transduction. Several CAD devices (Corona CAD, ESA Inc., Chelmsford, Massachusetts, USA) were evaluated with a variety of HPLC systems, including Agilent 00LC (Agilent Technologies, Wilmington, Delaware, USA) and ESA HPLC systems. Figure : LC PDA MS ELSD CAD system. Column: mm.6 mm, µm d p CapCell Pak MG C8 (Shiseido, Tokyo, Japan); binary gradient: 00% solvent A to 00% B in either or 0 min; solvent A: 98: (v/v) water isopropanol and solvent B: 7: (v/v) acetonitrile methanol, each containing 0 mm ammonium acetate; temperature: ambient; flow-rate:. ml/min. Peak area (x 0 6 ) Pump and autosampler.0..0..0 0. Column Glucose Caffeine Propranolol Chlorpromazine Gentistic acid Amitriptylline PDA Flow split MS CAD ELSD Figure : Peak area CAD response from FIA of seven compounds using three different isocratic conditions showing response proportionality to aqueous organic eluent composition. Percentage composition refers to solvents A and B described for Figure. Nitrogen gas (ESA nitrogen generator or cylinder), regulated at psi was introduced to the detector and the resultant gas flow-rate was regulated automatically and monitored by the CAD device. The only parameter that required user input was response range, which typically was set to 00 pa full scale with V analogue output to the Agilent ChemStation or EZChrom Elite for ESA chromatography data systems. Initial studies of detection scope and response factors involved flow-injection analysis (FIA) of compounds representing a wide diversity of chemical properties. Precision, sensitivity, dynamic range and nature of calibration curves were evaluated for a variety of compounds and chromatographic conditions and included some comparison with UV and ELSD. Specific conditions are described with results. Results and Discussion To study the chemical scope of CAD and dependence of response on analyte nature, we performed FIA of 7 chemically diverse compounds. Results and conditions, summarized in Table, show that all analytes, except for the volatile compound naphthol, were detected by CAD. Using these conditions, naphthol was detected by UV but did not respond with low-temperature ELSD (data not shown). These results suggest that CAD response depends upon analyte volatility. This is as expected based upon similar observations with other techniques that involve particle detection subsequent to solvent evaporation (for example, ELSD). Peak area from FIA of a constant mass (0 µg) of each analyte ranged from.78. 0 6 units across two devices with 6. and 8.% RSD within each instrument (Table ). Because these analytes represent a wide range of molecular weights, polarity, acid base characteristics, and included several nons, these results demonstrate that response magnitude does not depend significantly upon analyte chemical structure or properties. This is consistent with the described independence of electrical aerosol measurement on the nature of analytes. 9 The detection scope was characterized further for 7 compounds from Millennium Pharmaceuticals (Cambridge, Figure : Gradient reversed-phase separation and detection of diverse compounds with CAD. Column, mobile phases, flow-rate and temperature were the same as in Figure. Gradient: 0 00% B in min. Sample: µg, 0 replicate injections. 00 80 60 0 0 00 80 60 0 0 0 0.0 0..0..0..0..0..0 0.0 0 0 0 0 0 0 60 70 80 Percent organic Peaks: glucose, caffeine, propranolol, chlorpromazine. LC GC Europe 8(6) (00)

Massachusetts, USA) small molecule library. Compound molecular masses ranged from 6 to 68 Da and included a wide range of chemical structures and properties from several drug discovery programs. CAD was used in combination with photodiode-array detection, single-quadrupole MS and ELSD, as shown in Figure. Using the described conditions, the same 7 compounds were detected by both CAD and ELSD. The only two compounds that did not respond were those with the lowest MW (6 and 78 Da). This provides further evidence of the dependence of response on analyte volatility and demonstrates that CAD, which uses ambient temperature evaporation, has a similar detection scope to low-temperature ELSD. CAD response factors (area/micrograms) were highly variable with 0% RSD for the 7 library compounds detected. Because our FIA results demonstrated that there was no significant dependence of response upon chemical structure, we therefore studied the dependence of response upon analytical conditions. We performed FIA of six compounds of known purity using various mixtures of solvents A and B, as described in Figure. The results demonstrate that both ELSD (not shown) and CAD response were dependent upon eluent composition. Figure shows that CAD response was directly proportional to eluent percentage organic content, and that the magnitude of this effect was very similar for all compounds. These results are consistent with the theory of nebulization, specific droplet size selection (for example, impactor, syphon) and evaporation. This effect, in which higher solvent volatility favours delivery of analyte for subsequent detection, has been observed widely for many techniques that involve solvent evaporation (for example, atmospheric pressure ionization [API] MS, ELSD). This is further demonstrated in Figures and, which show the response obtained for a constant mass of several compounds separated by gradient reversed-phase and hydrophilic interaction chromatography (HILIC), respectively. As shown, peak area either increases or decreases proportionally with retention time as a consequence of a corresponding increase or decrease in percentage organic content during gradient elution. As shown in Figures, the dependence of CAD response upon eluent composition was predictable and uniform among diverse analytes. Based upon these observations, our ongoing efforts are to examine specifically the use of empirical and computational approaches as a means of normalizing CAD response. These approaches will be applied to a broader scope of Millennium Pharmaceuticals compound library (that is, more compounds, greater chemical diversity) to assess the practical capabilities of CAD to provide uniform response factors (that is, accurate quantification) across numerous Figure 6: Isocratic HILIC separation and CAD of simple carbohydrates. Column: 0 mm.6 mm, µm d p Asahipak NH P (Shodex); mobile phase: 70:0 (v/v) acetonitrile water; flow-rate:.0 ml/min; temperature: C; sample: 0 ng. 0 8 6 Peaks: fructose, glucose, sucrose, lactose. 0 6 7 8 9 0 Figure : CAD of oligosaccharides using gradient HILIC. Column: 0 mm.6 mm, µm d p Asahipak NH P (Shodex); mobile phase A: acetonitrile; mobile phase B: water; gradient: 70 0% A in 0 min; flow-rate:.0 ml/min; temperature: C; sample: 00 ng. 0 0 0 0 0 6 8 0 6 8 Peaks: DP, DP, DP, DP, DP, 6 DP6, 7 DP7. 6 7 Figure 7: Isocratic ion chromatographic separation and CAD of anions. Column: 00 mm. mm, 0 µm d p PRP-X00 (Hamilton, Reno, Nevada); mobile phase: water with 60 mm ammonium hydroxide and 00 mm formic acid (isocratic); flow-rate: ml/min; temperature: ambient. 7 0 00 7 0 0.0 0 6 7 8 9 0 Peaks: phosphate, chloride, bromide, nitrate, sulphate. www.lcgceurope.com

analytical runs and in support of several ongoing drug discovery programs. The chromatograms in Figures 8 show several examples of the use of CAD for a variety of separations. This includes its use with several solvents, additives and conditions including gradient elution (Figures and ), reversed phase (Figures and 8), hydrophilic interaction (Figures and 6), and ion chromatography (Figure 7). As expected, our data shows that CAD has similar requirements for mobile phase volatility as API MS and ELSD. In addition to that shown, the detection technique was also compatible with many other solvents and modifiers including acetone, methylene chloride, hexane, triethylamine and trifluoroacetic acid. Figure 8 shows the use of CAD with several HPLC column formats. Using a single nebulizer design and the described conditions, peak width increased and response decreased only when using a column inner diameter less than mm. These results are in accordance with our measurements of an effective detector volume of approximately µl. This demonstrates good compatibility with narrow and normal bore chromatography, and further study is required to address lower volume formats. Figure is an overlay of 0 replicate injections in which response variability (percent RSD) was.6,.7,. and.8 for glucose, caffeine, propranolol and chlorpromazine, respectively. Variability obtained with ELSD, using these same conditions, was.,.6,.9 and. %RSD. for these compounds. This twofold or more improvement in precision over ELSD has been observed consistently in several applications in which we have compared CAD with ELSD. The lower limit of detection for the previously mentioned analytes was estimated using serially diluted standards and determined as the lowest mass injected that yielded a peak-to-peak signal-to-noise ratio of or greater. The limit of detection ranged from to 0 ng for CAD and 00 to 0 for ELSD. Figure 6 provides an example of CAD sensitivity for simple sugars. The 0-fold or better improvement in LOD obtained with CAD has been consistently observed again in several comparison studies with ELSD. Figure 9 shows that a dynamic range of at least four orders of magnitude (typically 0 ng to 00 µg) was obtained with CAD with very similar calibration curves (that is, response factors) for all five phenolic compounds. CAD response was nonlinear and can be represented by y ax b where y peak response (height or area), x analyte mass and b the exponential response factor (that is, sensitivity). The slope determined from a log log plot of response versus concentration for these five phenolic compounds ranged from 0.8 to 0.0. This response behaviour is in substantial agreement with both theory and empirical data on pneumatic nebulization and aerosol charging. 8 Because CAD and low-temperature ELSD use comparable techniques for aerosol formation, the observed improvements in precision, sensitivity and dynamic range most likely result from differences in particle detection technology. It is widely recognized that light-scattering efficiency is dependent upon particle diameter and involves exponential changes between scattering regimes (that is, Rayleigh, Mie, refraction, reflection). This has been shown to underlie the dramatic drop-off in response, limited sensitivity, and high response variability often observed with ELSD., In contrast, the response obtained with electrical aerosol detection technology used with CAD has been shown to have unimodal efficiency over a broad range of particle diameters.,8 These performance These performance characteristics can provide significant advantages in many applications, including those requiring detection of diverse compounds and trace analyses. characteristics can provide significant advantages in many applications including those requiring detection of diverse compounds and trace (for example, impurity, metabolite) analyses. Conclusions The described studies demonstrate that charged aerosol detection applies to a broad range of HPLC separations and appears to possess a number of very desirable characteristics. These include: universal detection of nonvolatile analytes, response independent of chemical properties, a wide dynamic response range with high sensitivity, good precision for a wide Figure 8: CAD response for a constant mass of µg of caffeine analysed with.6,.0,.0, and. mm i.d. column formats. Columns: 7 mm long, µm d p CapCell Pak C8 (Shiseido, Tokyo, Japan); mobile phase: 0% aqueous methanol. Flow-rate was scaled proportionally with column volume (for example,. ml/min for the.6 mm i.d. column and 0.8 ml/min for the. mm i.d. column). Injection volumes were 0, 0, 6 and 8 µl, respectively, for the.6,.0,.0 and. mm i.d. columns. Peak height ( 0 6 ) 90 80 70 60 0 0 0 0.0.6.0. 0..0..0..0. Figure 9: Dynamic response range of CAD. 600000 00000 00000 00000 00000 00000 0 0 00 00 600 800 000 00 0 0 0000 0000 60000 80000 00000 0000 Mass injected (ng) -HPAC Chlorogenic acid -HPAC Gallic Chlorogenic acid acid Gallic Gentisic acid Gentisic acid acid Protocatechuric acid acid 6 LC GC Europe 8(6) (00)

diversity of analytes, and simple and reliable operation. The magnitude of CAD response was found to be proportional to eluent percentage organic content. Additional study is therefore required to determine if the observed predictable nature of this dependency can provide a means of normalizing response for accurate quantification of unknowns. The applicability of this technique to biological matrices also requires additional study. Acknowledgments The authors would like to thank S.L. Kaufman, TSI Inc. (Shoreview, Minnesota, USA) for his fundamental work with electrical aerosol technology, its initial coupling with liquidphase techniques, and for his valuable technical insight. References. M.C. Allgeier, M.A. Nussbaum and D.S. Risley, LC GC, (), 76 (00).. K. Petritis, C. Elfakir and M. Dreux, J. Chromatogr., A, 96, 9 (00).. D.A. Yurek, D.L. Branch, and M.S. Kuo, J. Comb. Chem.,, 8 8 (00).. R.W. Dixon, and D.S. Peterson, Anal. Chem., 7, 90 97 (00).. J.A. Koropchak et al., Adv. Chromatogr., 0, 7 (000). 6. P.A. Lawless, J. Aerosol Sci., 7, 9 (996). 7. A. Medved et al., J. Aerosol Sci.,, S66 S67 (000). 8. B.Y.H. Liu and D.Y.H. Pui, J. Aerosol Sci., 6, 9 (97). 9. R.C. Flagan, Aerosol Sci. Technol., 8 (998). 0. A.J. Alpert, J. Chromatogr., 99, 77 96 (990). Paul H. Gamache is Director, applications, with ESA Inc., Chelmsford, Massachusetts, USA. Ryan S. McCarthy is an applications specialist with ESA Inc., Chelmsford, Massachusetts, USA. Scott M. Freeto is Applications Manager with ESA Inc., Chelmsford, Massachusetts, USA. Darwin J. Asa is Marketing Manager with ESA Inc., Chelmsford, Massachusetts, USA. Matthew J. Woodcock, is a research associate, Millennium Pharmaceuticals, Cambridge, Massachusetts, USA. Katherine A. Laws is a research investigator with Millennium Pharmaceuticals, Cambridge, Massachusetts, USA. Roderic O. Cole is Director, analytical sciences, with Millennium Pharmaceuticals, Cambridge, Massachusetts, USA. www.lcgceurope.com 7