Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Similar documents
Technical Procedure for Solid Phase Extraction of THC and THC-COOH for GC-MS Analysis

Technical Procedure for Blood Cannabinoid Liquid-Liquid Extraction (BCLLE) for Analysis by LC-MS/MS

Technical Procedure for General Laboratory Equipment

1. Preliminary qualitative analysis of unknown substances (liquid or solid).

Technical Procedure for ELISA Drug Screen Version 3 Toxicology Unit Effective Date: 02/15/2013. ELISA Drug Screen

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Psychoactive Drugs Urine LC-MS/MS Analysis Kit User Manual

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Procedure for Body Fluid Unit Quality Control

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

Sulfotepp impurities in Chlorpyrifos EC formulations

Technical Procedure for Measurement Assurance

Determination of Polymer Modifier in Asphalt

Uncontrolled Copy. SOP 109 Ethylene Glycol Screen by Gas Chromatography/Mass Spectrometry. Table of Contents. 1. Principle of Assay...

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF

PYRIPROXYFEN TECHNICAL

Toxicology Extraction Procedure for Base Drugs Using United Chemical Technologies Clean Screen Extraction Columns

TECHNICAL BRODIFACOUM

British American Tobacco Group Research & Development. Method - Determination of ammonia in mainstream smoke

H 3 CO H 3 CO S CH 3

HIPPURIC ACID and o-m-p- METHYLHIPPURIC ACIDS IN URINE BY UV CODE Z06010

Application Note. Author. Abstract. Xinlei Yang Agilent Technologies Co. Ltd Shanghai, China

VMA, HVA, 5-HIAA Urine LC-MS/MS Analysis Kit User Manual

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Prepared By: R. W. Waggoner, Jr. Effective Date: August 18, 2008

Appendix II- Bioanalytical Method Development and Validation

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

RESOLUTION OENO 33/2004 DETERMINATION OF SHIKIMIC ACID IN WINE BY HPLC AND UV-DETECTION

Draft Method proposal: determination of glucoheptonic acid (HGA) in fertilizers.

LAP-013. Procedure Title: Date: 6/28/96. Author: Raymond Ruiz and Tina Ehrman ISSUE DATE: 8/26/96 SUPERSEDES: 11/01/94.

Figure 1. Structures for Vitamin B2 and Vitamin B1.

637. Thiamethoxam. HPLC method

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7

Certified Reference Material - Certificate of Analysis Methylone, Primary Standard

HPLC Praktikum Skript

7. Stability indicating analytical method development and validation of Ramipril and Amlodipine in capsule dosage form by HPLC.

Uncontrolled Copy. SOP Ethylene Glycol Quantitation by Gas Chromatography/Mass Spectrometry. Table of Contents. 1. Principle of Assay...

British American Tobacco Group Research & Development. Method - Determination of phenols in mainstream cigarette smoke

Quantitative determination of cannabinoids in Hemp and Hemp products by High Performance Liquid Chromatography using a UV Detector

Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1

USGS Troy WSC Laboratory Ion Chromatography SOP EPA 300.0, Rev Jordan Road SOP. No. 1 Rev. No. 1.7 Troy, NY Date 01/25/18 Page 1 of 7

TECHNICAL TEMEPHOS. 1. Specification. Full specification WHO/SIT/19.R4 Revised 10 December Description

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

Diquat 1,1 -ethylene-2,2 -bipyridium dibromide salt Paraquat 1,1 -dimethyl-4,4 -bipyridium dichloride salt Initial Preparation

STABILITY INDICATING METHOD OF RELATED IMPURITIES IN VENLAFAXINE HYDROCHLORIDE SUSTAINED RELEASE TABLETS

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Chapter 4: Verification of compendial methods

METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY

maect PRODUCTION SOP M/1 JUNE 2007 version page: 1 Preparation of 2 liters of 10x concentrated PBS stock solution (phosphate buffered saline)

Certified Reference Material - Certificate of Analysis

ANALYTICAL METHOD PROCEDURES

Certified Reference Material - Certificate of Analysis

Method Development and Validation Of Prasugrel Tablets By RP- HPLC

Zearalenone Plate Kit PN 53018B

The analysis of organic acid content of additives, premix, feed, and water.

Method - Determination of aromatic amines in mainstream cigarette smoke

Chemistry Gas Chromatography: Separation of Volatile Organics

SOIL ORGANIC CONTENT USING UV-VIS METHOD

USP Method Transfer and Routine Use Analysis of Irbesartan Tablets from HPLC to UPLC

METHOD 3665 SULFURIC ACID/PERMANGANATE CLEANUP

Determination of Caffeine by HPLC

METHOD 8330A NITROAROMATICS AND NITRAMINES BY HIGH PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC)

TEMPLATE FOR AN EXAMPLE STANDARD TEST METHOD

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

METHADONE and EDDP in urine by GC-MS Code GC Method of Confirmation by GC-MS

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

Chemistry 3200 High Performance Liquid Chromatography: Quantitative Determination of Headache Tablets

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

25-OH VITAMIN D 3 AND 25-OH VITAMIN D 2 IN PLASMA BY UV / VIS FAST CODE Z19710

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID)

CHEMISTRY DEPARTMENT

Catecholamines Urine LC-MS/MS Analysis Kit User Manual

Volume 6, Issue 2, January February 2011; Article-015

Application Note PB 401. Improve HPLC Sample Preparation in an Analytical Laboratory A New Automated Sample Preparation Process

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009.

Tex-620-J, Determining Chloride and Sulfate Contents in Soil

1. A gas chromatograph equipped with flame ionization detection and a computing integrator.

Spectrophotometric Determination of Ferrocyanide in Effluents

CIPAC. CIPAC Free relevant impurities methods:

p-methoxyamphetamine Latest Revision: May 16, 2013

PAR Method Method to 2.0 mg/l Pb TNTplus 850

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

INTERNATIONAL CHEMISTRY TESTING. 258 Main Street- Suite 311- Milford, MA Tel: Fax:

IDENTIFICATION OF STEROIDS IN COSMETIC PRODUCTS BY TLC AND HPLC 1 02/12/2005 ACM 007 A. THIN LAYER CHROMATOGRAPHY (TLC)

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

2,5-DIMETHOXY-4-IODOPHENETHYLAMINE Latest Revision: August 22, 2005

Automated Sample Preparation in Quality Control of Eye-Drop Formulation

Transferring HPLC methods from the HP 1090 Series to the Agilent 1100 Series HPLC system

Agilent LC & LC-MS Cannabis Testing

Determination of Trace Cations in Power Plant Waters Containing Morpholine

CORESTA Recommended Method No. 86

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

FORMALDEHYDE IN URINE by UV Code Z65010

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

Impact factor: 3.958/ICV: 4.10 ISSN:

1-(2-METHOXYPHENYL)PIPERAZINE Latest revision: June 27, 2005

Transcription:

Technical Procedure for Concentration Determination of 1.0 Purpose This procedure specifies the required elements for the preparation and use of the Agilent 1100/1200 series High Performance Liquid Chromatograph (HPLC) to determine the concentration of methamphetamine in liquids. 2.0 Scope This procedure applies to the HPLC systems used in Drug Chemistry at the Raleigh and Western Regional locations of the State Crime Laboratory. 3.0 Definitions Performance verification The initial confirmation of the reliability of a previously or externally validated method or instrument. Quality control (QC) check Periodic confirmation of the reliability of equipment, instrumentation, and/or reagents. Reference Material Material sufficiently homogeneous and stable, with reference to specified properties, which has been established to be fit for its intended use in measurement or in examination of nominal properties. 4.0 Equipment, Materials and Reagents 4.1 Equipment 4.2 Materials Agilent 1100 Series Thermostatted Column Compartment Agilent 1100 Series Autosampler Agilent 1200 Series Vacuum Degasser Agilent 1200 Series Quaternary Pump Agilent 1100 Series Diode Array Detector Printer or other output device PC Data system with HPLC 2D Chemstation software, Version B.02 (or higher upgrade) Phenomenex Synergi Hydro-RP 4μ C18 Column, 150x4.6mm Solvent Reservoirs (or other interchangeable column) 2ml Autosampler vials with 11mm caps Filtration System with vacuum access Class A Volumetric Flasks Funnels Class A Graduated Cylinders Class A Volumetric Pipettes 0.45 μ filters 0.45 μ syringe filters Syringes 4.3 Commercial Reagents Phosphoric acid Triethanolamine HPLC grade water Page 1 of 8

HPLC grade acetonitrile 4.4 Primary Reference Materials - The following reference materials shall be selected from certified reference materials maintained by the laboratory. 5.0 Procedure d-methamphetamine HCl d-amphetamine (+/-)3,4-Methylenedioxyamphetamine (MDA) (+/-)3,4-Methylenedioxymethampetamine (MDMA) (+)-Pseudoephedrine (-)-Ephedrine 5.1 Reagents, Standards, and Controls - The reagents, standards, and controls may be prepared in any amount provided that the component ratios are kept constant. 5.1.1 Buffer Preparation 5.1.1.1 Add 22.5 ml of phosphoric acid and 22 ml of triethanolamine to 4 L of HPLC Grade Water. 5.1.1.1.1 The ph of this solution shall be between 2.2 and 2.3. The amount of phosphoric acid may be adjusted to achieve the desired ph. 5.1.1.2 Expiration date is one month from preparation if kept at room temperature. 5.1.1.3 If stored in the refrigerator, the buffer shall have the ph checked before use to ensure that it is still within the specification stated in 5.1.1.1.1. The expiration date on the buffer stored in the refrigerator will be one year after preparation. Buffer solution shall be used as the mobile phase in the instrument operation. 5.1.2 0.01 N Hydrochloric Acid (HCl) 5.1.2.1 Add 210 µl concentrated hydrochloric acid to approximately 200 ml of deionized water in a 250 ml volumetric flask. The volumes may be changed as long as the ratios remain constant. 5.1.2.2 Mix and dilute to volume with deionized water. 5.1.2.3 Expiration date: One year. 5.1.2.4 Refrigerate 5.1.2.5 QC Check: Test acidic to ph or litmus paper. Page 2 of 8

5.2 Standard Preparation 5.2.1 Prepare the following concentrations of standard solutions of methamphetamine. The calibration primary standard and the verification primary standard shall be sampled from the different suppliers or different lot numbers with a supplier available at the time of preparation. 5.2.2 Dilute each standard in a 0.01 N HCl solution. 5.2.3 The following weight to volume dilutions are recommendations which may be increased or decreased depending upon the volume of standard needed. However, the ratio of weight to volume shall remain constant. 5.2.4 Make a 20.0 mg/ml stock solution by weighing out 2.0 g into a 100 ml volumetric flask using an analytical balance. Dilute to 100 ml in a flask with a 0.01 N HCl solution. 5.2.5 0.40 mg/ml: Pipet 2.0 ml of the 20.0 mg/ml stock solution into a 100 ml flask. Dilute to 100 ml with a 0.01 N HCl solution. 5.2.6 0.20 mg/ml: Pipet 50.0 ml of the 0.40 mg/ml solution into a 100 ml flask. Dilute to100 ml with a 0.01 N HCl solution. 5.2.7 0.10 mg/ml: Pipet 50.0 ml of the 0.20 mg/ml solution into a 100 ml flask. Dilute to 100 ml with a 0.01 N HCl solution. 5.2.8 0.05 mg/ml: Pipet 50.0 ml of the 0.10 mg/ml solution into a 100 ml flask. Dilute to 100 ml with a 0.01 N HCl solution. 5.2.9 0.01 mg/ml: Pipet 20.0 ml of the 0.05 mg/ml solution into a 100 ml flask. Dilute to 100 ml with a 0.01 N HCl solution. 5.2.10 A 0.40 mg/ml verification standard per 5.2.1 shall be prepared and used to verify the calibration curve. Repeat 5.2.4 and 5.2.5 label solution as a verifier. 5.2.11 A 0.02 mg/ml verification standard per 5.2.1 shall be prepared and used to verify the calibration curve. Using the 0.40 mg/ml solution prepared in 5.2.10: pipet 5 ml of 0.40 mg/ml solution into a 100 ml flask. Dilute to 100 ml with a 0.01 N HCl solution. 5.2.12 The expiration date of the prepared standards shall be one year from the date prepared. All standards will be stored in the refrigerator. However, if a standard fails to produce a valid calibration curve, it shall be discarded and new standards shall be prepared. New standards shall provide accurate and reproducible data before they can be used in analysis (see 5.7). 5.3 System Check Solution Preparation 5.3.1 Weigh 5 mg (amount may be adjusted dependent on standard availability) of each of these commercial primary standards into a 50 ml flask and dilute with a 0.01 N HCl solution: pseudoephedrine, ephedrine, methamphetamine, amphetamine, MDA, and MDMA. Page 3 of 8

5.3.2 The expiration date of the system check solution shall be one year from the date prepared. The solution will be stored in the refrigerator. However, if a solution fails to produce valid results, it shall be discarded and a new solution shall be prepared. 5.3.3 The system check solution shall be verified before every run on the instrument. 5.4 Wash Solution 5.4.1 Mix 10 ml of acetonitrile and 10 ml of HPLC grade water to form a 50:50 acetonitrile: HPLC grade water solution. 5.5 Performance Verification for New Instrumentation 5.5.1 New high performance liquid chromatographs shall be installed by a manufacturer representative and shown to meet any manufacturer s requirements. 5.5.2 Performance verification shall be performed by the HPLC Key Operator on new high performance liquid chromatographs prior to being used for casework. 5.5.3 The performance verification shall include a successful calibration, system check solution run, verification standard runs, and any additional runs needed to show the repeatability, linearity, and selection of methamphetamine. 5.6 Instrument Maintenance and Shutdown 5.6.1 Record all maintenance in the maintenance log at the time it is performed. 5.6.2 When the HPLC is taken out of service (e.g., maintenance, malfunction, leaving the direct control of the Laboratory), correct operation shall be demonstrated by running the system check solution. 5.6.3 Suggested routine maintenance 5.6.4 Shutdown 5.6.3.1 This is a recommended maintenance schedule. Instrument use may alter the need for maintenance and shall be performed at the discretion of the HPLC Key Operator or designee. 5.6.3.2 Before and after each full run, the column shall be flushed with 100 % acetonitrile until a stability of the baseline is reached. 5.6.3.3 If the instrument is not in use for approximately three months, the buffer line shall be placed into a methanol reservoir. 5.6.4.1 If the instrument shuts down, this shut down shall be noted in the instrument log on the day this is noticed. After successfully bringing the instrument back on line with a stable baseline, an acceptable system check solution shall be completed before any calibration standards, verification standards, or case samples are analyzed. Page 4 of 8

5.7 Calibrations 5.7.1 The Agilent High Performance Liquid Chromatograph shall be calibrated with every run. Performance verification shall be performed prior to the analysis of any casework. 5.7.2 The first injection is the six component system check solution prepared in 5.3. 5.7.3 All compounds shall be effectively resolved from the methamphetamine peak, with the closest peak to methamphetamine at a resolution greater than 1.5. 5.7.4 Once the previous injection meets all specifications, the analysis of calibration standards, verification standards, and samples may begin. 5.7.5 The calibration shall consist of double injections of the 0.01, 0.05, 0.10, 0.20 and 0.40 mg/ml solutions. 5.7.6 The correlation coefficient produced by these standards shall be greater than 0.995. 5.7.7 The final injections are the 0.02 mg/ml and 0.40 mg/ml verification standards. These standards shall be entered into the sample table as a sample, not a standard. The result shall be within +/- 5 % of the true concentration of the standard. This verifier shall be considered a positive control. 5.7.8 Records 5.8 Sampling 5.7.8.1 The calibration data shall include the lot numbers of the calibrations standards and the system check solution standards. The chromatograms for each calibration standard shall have the coefficient of determination (r 2 ) values. 5.7.8.2 Record each calibration in the instrument log with the date, lot number of primary standards used (including those from the system check solution), the expiration date of the calibration run, and the operator initials. 5.7.8.3 Record any calibrations that do not meet the requirements in the instrument log. 5.8.1 Analysis samples shall be obtained as a representation from the entire sample. Sampling shall be performed as follows: 5.9 Instrument Parameters 5.8.1.1 For all liquids, a reasonable amount of sample shall be seized from the site by the processing Forensic Scientist, depending on the amount of liquid available at the scene, and the amount needed to perform qualitative analysis. Oven temperature: 55 C Flow rate: 0.8 ml/min Injection amount: 5.0 μl Page 5 of 8

Run time: approximately 17.5 minutes Detector: diode array detector 210 nm Bw10, reference 550 nm Bw 100 Mobile phase: 93 % buffer, 7 % acetonitrile Sample solvent: 0.01 N HCl Wash solution: 50 % HPLC grade acetonitrile, 50 % HPLC grade water 5.10 Application of Procedure on Evidence 5.10.1 Sample Preparation 5.10.1.1 For single-layered liquids, pipette a 1.0 ml aliquot of the liquid into a volumetric flask using a volumetric pipette. 100 ml flask size is suggested, but may be modified based on sample size and the ph of the sample solution. 5.10.1.2 Each layer of a two-layered liquid shall be analyzed separately. After separating the two-layered liquid, refer to 5.10.1 for sample preparation. 5.10.1.3 Two dilutions of each case sample or each liquid if two layers are present shall be prepared using a 0.01 N HCl solution as the dilution solvent. 5.10.1.4 The dilutions shall be mixed thoroughly before sampling to put on the instrument. For organic samples, a bi-layer will form once the 0.01 N HCl is added. The aqueous layer shall then be sampled out for analysis. 5.10.1.5 Ensure that an adequate amount of the dilution solvent has been added to produce an acidic solution. 5.10.1.6 Filter approximately 2 ml of each solution through a 0.45 μ syringe filter before injection. 5.10.2 Sample Injection 5.10.2.1 Place a 50:50 acetonitrile:water solution in the designated autosampler well for the instrument to perform self-cleaning on the needle and injection port after each injection. 5.10.2.2 Inject 5.0 μl of the blank solution (0.01 N HCl) before each sample injection, into the injector port. Observe the chromatogram for any interferences. This shall be a negative control. 5.10.2.3 Inject 5.0 μl of the filtered sample. Prepare each sample in duplicate. Analyze each preparation in duplicate. 5.10.2.4 When data collection is complete, observe the chromatograms for the following: The results shall be the concentration of methamphetamine hydrochloride in mg/ml. Page 6 of 8

5.10.2.5 Reporting The calculated concentrations shall differ no more than 3 % from one another. If any preparation falls outside of these limits, the dilution shall be prepared again and a new analysis shall be performed. 5.10.2.5.1 Include the calibration results with the reported r 2 values, the system check solution results showing the resolution of the peaks is greater than 1.5, the verification results, and the sample results into the FA case record. 5.11 Calculations - An average of the results of the four injections shall be calculated and that concentration will be the reported value. 5.11.1 For one-layered liquids, the concentration determined, as outlined in 5.10.2.4, shall be the final concentration reported. 5.11.2 For two-layered liquids, as stated in 5.10.1.2, each layer shall be worked separately. The concentration for each layer determined, as outlined in 5.10.2.4, shall be the final concentration reported. 5.12 Uncertainty of Measurement - See the Drug Chemistry Section Procedure for Uncertainty of Measurement. 6.0 Limitations - N/A 7.0 Safety 7.1 Use caution such as gloves and a fume hood when handling acetonitrile, triethanolamine, and phosphoric acid to avoid eye and skin contact. Use caution when handling the 0.01 N HCl solution as it has acidic properties. 8.0 References Chromatographic Quantitation of Methamphetamine. North Carolina State Bureau of Investigation Raleigh and Western Regional Laboratories, 2007. Weston, Robert. HPLC validation for use in quantification of methamphetamine in liquids. Oklahoma State Bureau of Investigation-Controlled Substances Unit, 2008. 9.0 Records FA System Case files HPLC Logbook HPLC Activity Log HPLC Maintenance Log 10.0 Attachments - N/A Page 7 of 8

Revision History Effective Date 09/17/2012 Version Number Reason 1 Original Document Page 8 of 8