METHOD 13B - DETERMINATION OF TOTAL FLUORIDE EMISSIONS FROM STATIONARY SOURCES (SPECIFIC ION ELECTRODE METHOD)

Similar documents
METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD)

NOTE: This method does not include all of the. specifications (e.g., equipment and supplies) and procedures

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

Chemistry 119: Experiment 6. Sampling and Analysis of a Solid Drain Cleaner

ASTM Designation: D Standard Test Method for Determination of Iodine Number of Activated Carbon

Tex-620-J, Determining Chloride and Sulfate Contents in Soil

METHOD 7 - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES. NOTE: This method does not include all of the

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows

GENERAL INSTRUCTIONS GENERAL PREPARATION

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

GETTING THE END POINT TO APPROXIMATE. Two hours

METHOD 9040B. ph ELECTROMETRIC MEASUREMENT

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

Operating Instructions for Ammonium ISE Specifications

S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) Fax: +44 (0)

1) A ph/mv meter or an ion meter, either line operated or portable.

Oxygen Demand, Chemical

GENERAL INSTRUCTIONS GENERAL PREPARATION

Standard Operating Procedure for GLNPO Specific Conductance: Conductivity Bridge

Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Use a pipet bulb when pipeting cyanide solutions, as these solutions are highly toxic.

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

PRETREATMENT TECHNICAL DATA SHEET A CHROME-FREE FINAL RINSE PRODUCT DESCRIPTION

Standard Test Methods for Fluoride Ion in Water 1

Percentage of Acetic Acid in Vinegar

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Total Carboxylic Acid Group Content Applicable Products: Carbopol * Polymers and Pemulen * Polymeric Emulsifiers

C = concentration as ppm sulfide V t = volume of titrant at endpoint V s = volume of standard used (10 ml)

Functional Genomics Research Stream. Lecture: February 17, 2009 Masses, Volumes, Solutions & Dilutions

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

To measure ph s in a variety of solutions and mixtures and to account for the results obtained.

# 12 ph-titration of Strong Acids with Strong Bases

Oxygen Demand, Chemical

STANDARD OPERATING PROCEDURE No. 52 STATIC ACID GENERATION (NAG) TEST

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

Titration with an Acid and a Base

Chromium, Total. Alkaline Hypobromite Oxidation Method 1 Method to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters)

POLYPHOSPHORIC ACID ASSAY

EXPERIMENT 6. Properties of Buffers INTRODUCTION

PRETREATMENT TECHNICAL DATA SHEET CHROMIUM-FREE ORGANIC PASSIVATING RINSE PRODUCT DESCRIPTION

METHOD 9012 TOTAL AND AMENABLE CYANIDE (COLORIMETRIC, AUTOMATED UV)

RAPID KJELDAHL BENCHNOTES

Ascorbic Acid Titration of Vitamin C Tablets

PRETREATMENT TECHNICAL DATA SHEET CHROMIUM-FREE ORGANIC PASSIVATING RINSE PRODUCT DESCRIPTION

Spectrophotometric Determination of Ferrocyanide in Effluents

perfection Guidebook perfection Combination Fluoride Electrode Successful Ion Measurement

Experiment 24. Chemical recycling of poly(ethylene) terephthalate (PET)

METHOD #: Approved for NPDES and SDWA (Ed. Rev. 1974, 1978) Fluoride, Total (Colorimetric, SPADNS with Bellack Distillation)

ALLOWAY. Method Outline. Standard Laboratory Method: Total Fluoride via Distillation. Distillation then Specific Ion. Date Issued: 01/04/82

Procedure for Body Fluid Unit Quality Control

COLE-PARMER LABORATORY NITRATE ION ELECTRODE INSTRUCTION MANUAL

Potentiometric measurement of ph

The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium ions in aqueous solutions.

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

Method 8326 (0.006 to mg/l Al 3+ ) Powder Pillows

Dynamar Polymer Processing Additives Parr Bomb Analytical Method for Determining Total Organic Fluorine Concentration in Polyethylene

DR/4000 PROCEDURE NITRATE. 2. The display will show:

Appendix. Abbreviations

Chloramine (Mono) and Nitrogen, Free Ammonia

The Cole-Parmer Nitrate Ion Electrodes are used to quickly, simply, accurately, and economically measure nitrate in aqueous solutions.

Periodicity of Properties of Oxides

E-TIN 34 PRE-DIP A Solderable Immersion Tin

COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

Chemistry Determination of Mixed Acids

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

Chromazurol S Method Method to 0.50 mg/l Al TNTplus 848

CHEM51LC PROJECT DETERMINATION OF DIASTEREOSELCTIVITY USING THERMODYNAMIC VERSUS KINETIC CONTROLLED REDUCTION PROCEDURES: A REDUCTION

DPD Test N Tube Method *

Chlorine, Free and Total, High Range

COLE-PARMER LABORATORY CYANIDE ION ELECTRODE INSTRUCTION MANUAL

User Manual. Thermo Scientific Orion Fluoride Ion Selective Electrode

Diquat 1,1 -ethylene-2,2 -bipyridium dibromide salt Paraquat 1,1 -dimethyl-4,4 -bipyridium dichloride salt Initial Preparation

METHOD 9200 NITRATE. 1.2 The applicable range of concentration is 0.1 to 2 mg NO -N per liter. 3 of sample.

Related concepts Electrolyte, electrical conductance, specific conductance, ion mobility, ion conductivity, conductometry, volumetry.

Approximate Volatile Acids by Titration

National standard of People s Republic of China

phoenix Electrode Company CARBON DIOXIDE GAS-SENSING ELECTRODE INSTRUCTION MANUAL 1. A ph/mv meter or an ion meter, either line operated or portable.

Determination of the K a Value and Molar Mass of an Unknown Weak Acid

Cyanide, colorimetric, pyridine-pyrazolone

Chemistry 119: Experiment 4. Titrimetric Determination of Acetic Acid Samples And Preparation of Youden Plot

Ascorbic Acid Titration of Vitamin C Tablets

Method to 0.50 mg/l NH 3 N Powder Pillows

METHOD 26 - DETERMINATION OF HYDROGEN HALIDE AND HALOGEN EMISSIONS FROM STATIONARY SOURCES NON-ISOKINETIC METHOD

PRETREATMENT TECHNICAL DATA SHEET MULTIMETAL IRON PHOSPHATE CONVERSION COATING PRODUCT DESCRIPTION

EXPERIMENT A7: VINEGAR TITRATION. Learning Outcomes. Introduction. Upon completion of this lab, the student will be able to:

METHOD 3520C CONTINUOUS LIQUID-LIQUID EXTRACTION

GENERAL INSTRUCTIONS GENERAL PREPARATION

Acids, Bases, Salts, and Buffers

(b) Write the chemical equation for the dissolution of NaOH in water.

EXPERIMENT 4. Identifying a Substance by Acid-Base Titration

Acid-Base Titration. M M V a

Scope and application: For wastewater, seawater, drinking water, surface water and process water.

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

#13 ph-titration of Weak Acids with Strong Bases

Name Period Date. Lab 9: Analysis of Commercial Bleach

Chlorine, Total. USEPA DPD Method 1 Method to mg/l as Cl 2 Chemkey Reagents. Test preparation. Before starting.

Method to 0.50 mg/l NH 3 N Powder Pillows

Mercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution

Ascorbic Acid Titration of Vitamin C Tablets

Transcription:

874 METHOD 13B - DETERMINATION OF TOTAL FLUORIDE EMISSIONS FROM STATIONARY SOURCES (SPECIFIC ION ELECTRODE METHOD) NOTE: This method does not include all of the specifications (e.g., equipment and supplies) and procedures (e.g., sampling and analytical) essential to its performance. Some material is incorporated by reference from other methods in this part. Therefore, to obtain reliable results, persons using this method should have a thorough knowledge of at least the following additional test methods: Method 1, Method 2, Method 3, Method 5, and Method 13A. 1.0 Scope and Application. 1.1 Analytes. Analyte CAS No. Sensitivity Total fluorides as Fluorine 7782-41-4 Not determined 1.2 Applicability. This method is applicable for the determination of fluoride (F - ) emissions from sources as specified in the regulations. It does not measure fluorocarbons, such as Freons. 1.3 Data Quality Objectives. Adherence to the requirements of this method will enhance the quality of the data obtained from air pollutant sampling methods.

875 2.0 Summary. Gaseous and particulate F - are withdrawn isokinetically from the source and collected in water and on a filter. The total F - is then determined by the specific ion electrode method. 3.0 Definitions. [Reserved] 4.0 Interferences. Grease on sample-exposed surfaces may cause low F - results because of adsorption. 5.0 Safety. 5.1 Disclaimer. This method may involve hazardous materials, operations, and equipment. This test method does not purport to address all of the safety problems associated with its use. It is the responsibility of the user of this test method to establish appropriate safety and health practices and to determine the applicability of regulatory limitations prior to performing this test method. 5.2 Corrosive Reagents. The following reagents are hazardous. Personal protective equipment and safe procedures are useful in preventing chemical splashes. If contact occurs, immediately flush with copious amounts of water at least 15 minutes. Remove clothing under shower and decontaminate. Treat residual chemical burn as thermal burn.

876 5.2.1 Sodium Hydroxide (NaOH). Causes severe damage to eye tissues and to skin. Inhalation causes irritation to nose, throat, and lungs. Reacts exothermically with limited amounts of water. 5.2.2 Sulfuric Acid (H 2 SO 4 ). Rapidly destructive to body tissue. Will cause third degree burns. Eye damage may result in blindness. Inhalation may be fatal from spasm of the larynx, usually within 30 minutes. May cause lung tissue damage with edema. 1 mg/m 3 for 8 hours will cause lung damage or, in higher concentrations, death. Provide ventilation to limit inhalation. Reacts violently with metals and organics. 6.0 Equipment and Supplies. 6.1 Sample Collection and Sample Recovery. Same as Method 13A, Sections 6.1 and 6.2, respectively. 6.2 Sample Preparation and Analysis. The following items are required for sample preparation and analysis: 6.2.1 Distillation Apparatus, Bunsen Burner, Electric Muffle Furnace, Crucibles, Beakers, Volumetric Flasks, Erlenmeyer Flasks or Plastic Bottles, Constant Temperature Bath, and Balance. Same as Method 13A, Sections 6.3.1 to 6.3.9, respectively. 6.2.2 Fluoride Ion Activity Sensing Electrode.

877 6.2.3 Reference Electrode. Single junction, sleeve type. 6.2.4 Electrometer. A ph meter with millivolt-scale capable of ± 0.1-mv resolution, or a specific ion meter made specifically for specific ion electrode use. 6.2.5 Magnetic Stirrer and Tetrafluoroethylene (TFE) Fluorocarbon-Coated Stirring Bars. 6.2.6 Beakers. Polyethylene, 100-ml. 7.0 Reagents and Standards. Unless otherwise indicated, all reagents are to conform to the specifications established by the Committee on Analytical Reagents of the American Chemical Society, where such specifications are available. Otherwise, use the best available grade. 7.1 Sample Collection and Sample Recovery. Same as Method 13A, Sections 7.1 and 7.2, respectively. 7.2 Sample Preparation and Analysis. The following reagents and standards are required for sample analysis: 7.2.1 Calcium Oxide (CaO). Certified grade containing 0.005 percent F - or less. 7.2.2 Phenolphthalein Indicator. Dissolve 0.1 g phenolphthalein in a mixture of 50 ml of 90 percent ethanol and 50 ml water. 7.2.3 Sodium Hydroxide (NaOH), Pellets.

878 7.2.4 Sulfuric Acid (H 2 SO 4 ), Concentrated. 7.2.5 Filters. Whatman No. 541, or equivalent. 7.2.6 Water. Same as Section 7.1.2 of Method 13A. 7.2.7 Sodium Hydroxide, 5 M. Dissolve 20 g of NaOH in 100 ml of water. 7.2.8 Sulfuric Acid, 25 Percent (v/v). Mix 1 part of concentrated H 2 SO 4 with 3 parts of water. 7.2.9 Total Ionic Strength Adjustment Buffer (TISAB). Place approximately 500 ml of water in a 1-liter beaker. Add 57 ml of glacial acetic acid, 58 g of sodium chloride, and 4 g of cyclohexylene dinitrilo tetraacetic acid. Stir to dissolve. Place the beaker in a water bath and cool to 20 EC (68 EF). Slowly add 5 M NaOH to the solution, measuring the ph continuously with a calibrated ph/reference electrode pair, until the ph is 5.3. Pour into a 1-liter volumetric flask, and dilute to volume with deionized, distilled water. Commercially prepared TISAB may be substituted for the above. 7.2.10 Fluoride Standard Solution, 0.1 M. Oven dry approximately 10 g of sodium fluoride (NaF) for a minimum of 2 hours at 110 EC (230 EF), and store in a desiccator. Then add 4.2 g of NaF to a 1-liter volumetric flask, and add enough water to dissolve. Dilute to volume with water.

879 8.0 Sample Collection, Preservation, Storage, and Transport. Same as Method 13A, Section 8.0. 9.0 Quality Control. 9.1 Miscellaneous Quality Control Measures. Section 8.0, 10.1 Quality Control Measure Sampling equipment leak-check and calibration Effect Ensure accurate measurement of stack gas flow rate and sample volume 10.2 Fluoride electrode Evaluate analytical technique, preparation of standards 11.1 Interference/recovery efficiency-check during distillation Minimize negative effects of used acid 9.2 Volume Metering System Checks. Same as Method 5, Section 9.2. 10.0 Calibration and Standardizations. 10.1. NOTE: Maintain a laboratory log of all calibrations. 10.1 Sampling Equipment. Same as Method 13A, Section 10.2 Fluoride Electrode. Prepare fluoride standardizing solutions by serial dilution of the 0.1 M fluoride standard solution. Pipet 10 ml of 0.1 M fluoride standard solution into a 100-ml volumetric flask, and make up to the mark with water for a 10-2 M standard solution.

880 Use 10 ml of 10-2 M solution to make a 10-3 M solution in the same manner. Repeat the dilution procedure, and make 10-4 and 10-5 M solutions. 10.2.1 Pipet 50 ml of each standard into a separate beaker. Add 50 ml of TISAB to each beaker. Place the electrode in the most dilute standard solution. When a steady millivolt reading is obtained, plot the value on the linear axis of semilog graph paper versus concentration on the log axis. Plot the nominal value for concentration of the standard on the log axis, (e.g., when 50 ml of 10-2 M standard is diluted with 50 ml of TISAB, the concentration is still designated "10-2 M"). 10.2.2 Between measurements, soak the fluoride sensing electrode in water for 30 seconds, and then remove and blot dry. Analyze the standards going from dilute to concentrated standards. A straight-line calibration curve will be obtained, with nominal concentrations of 10-4, 10-3, 10-2, 10-1 fluoride molarity on the log axis plotted versus electrode potential (in mv) on the linear scale. Some electrodes may be slightly nonlinear between 10-5 and 10-4 M. If this occurs, use additional standards between these two concentrations. 10.2.3 Calibrate the fluoride electrode daily, and check it hourly. Prepare fresh fluoride standardizing

881 solutions daily (10-2 M or less). Store fluoride standardizing solutions in polyethylene or polypropylene containers. NOTE: Certain specific ion meters have been designed specifically for fluoride electrode use and give a direct readout of fluoride ion concentration. These meters may be used in lieu of calibration curves for fluoride measurements over a narrow concentration ranges. Calibrate the meter according to the manufacturer's instructions. 11.0 Analytical Procedures. 11.1 Sample Loss Check, Sample Preparation, and Distillation. Same as Method 13A, Sections 11.1 through 11.3, except that the NOTE following Section 11.3.1 is not applicable. 11.2 Analysis. 11.2.1 Containers No. 1 and No. 2. Distill suitable aliquots from Containers No. 1 and No. 2. Dilute the distillate in the volumetric flasks to exactly 250 ml with water, and mix thoroughly. Pipet a 25-ml aliquot from each of the distillate into separate beakers. Add an equal volume of TISAB, and mix. The sample should be at the same temperature as the calibration standards when measurements are made. If ambient laboratory temperature fluctuates more than ± 2 EC from the temperature at which the calibration

882 standards were measured, condition samples and standards in a constant-temperature bath before measurement. Stir the sample with a magnetic stirrer during measurement to minimize electrode response time. If the stirrer generates enough heat to change solution temperature, place a piece of temperature insulating material, such as cork, between the stirrer and the beaker. Hold dilute samples (below 10-4 M fluoride ion content) in polyethylene beakers during measurement. 11.2.2 Insert the fluoride and reference electrodes into the solution. When a steady millivolt reading is obtained, record it. This may take several minutes. Determine concentration from the calibration curve. Between electrode measurements, rinse the electrode with water. 11.2.3 Container No. 3 (Silica Gel). Same as in Method 13A, Section 11.4.2. 12.0 Data Analysis and Calculations. Carry out calculations, retaining at least one extra significant figure beyond that of the acquired data. Round off figures after final calculation. 12.1 Nomenclature. Same as Method 13A, Section 12.1, with the addition of the following: M = F - concentration from calibration curve, molarity.

883 12.2 Average DGM Temperature and Average Orifice Pressure Drop, Dry Gas Volume, Volume of Water Vapor and Moisture Content, Fluoride Concentration in Stack Gas, and Isokinetic Variation. Same as Method 13A, Sections 12.2 to 12.4, 12.6, and 12.7, respectively. 12.3 Total Fluoride in Sample. Calculate the amount of F - in the sample using Equation 13B-1: F t ' K V t V d M A t Eq. 13B-1 where: K = 19 [(mg@l)/(mole@ml)] (metric units) = 0.292 [(gr@l)/(mole@ml)] (English units) 13.0 Method Performance. The following estimates are based on a collaborative test done at a primary aluminum smelter. In the test, six laboratories each sampled the stack simultaneously using two sampling trains for a total of 12 samples per sampling run. Fluoride concentrations encountered during the test ranged from 0.1 to 1.4 mg F - /m 3. 13.1 Precision. The intra-laboratory and interlaboratory standard deviations, which include sampling and analysis errors, are 0.037 mg F - /m 3 with 60 degrees of

884 freedom and 0.056 mg F - /m 3 with five degrees of freedom, respectively. 13.2 Bias. The collaborative test did not find any bias in the analytical method. 13.3 Range. The range of this method is 0.02 to 2,000 µg F - /ml; however, measurements of less than 0.1 µg F - /ml require extra care. 14.0 Pollution Prevention. [Reserved] 15.0 Waste Management. [Reserved] 16.0 Alternative Procedures. 16.1 Compliance with ASTM D 3270-73T, 91, 95 "Analysis for Fluoride Content of the Atmosphere and Plant Tissues (Semiautomated Method)" is an acceptable alternative for the distillation and analysis requirements specified in Sections 11.1 and 11.2 when applied to suitable aliquots of Containers 1 and 2 samples. 17.0 References. Same as Method 13A, Section 16.0, References 1 and 2, with the following addition: 1. MacLeod, Kathryn E., and Howard L. Crist. Comparison of the SPADNS- Zirconium Lake and Specific Ion Electrode Methods of Fluoride Determination in Stack Emission Samples. Analytical Chemistry. 45:1272-1273. 1973.

885 18.0 Tables, Diagrams, Flowcharts, and Validation Data. [Reserved]