Meeting the Challenges of Soil Analysis with the Avio 200 ICP-OES

Similar documents
Elemental analysis of river sediment using the Agilent 4200 MP-AES

ICP-OES Application Note Number 35

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors

Analysis of domestic sludge using the Agilent 4200 MP-AES

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

Using FIMS to Determine Mercury Content in Sewage Sludge, Sediment and Soil Samples

The Investigation of Fertilizer Analyses Using Microwave Digestion and the Agilent 720-ES

The Characterization of Nanoparticle Element Oxide Slurries Used in Chemical-Mechanical Planarization by Single Particle ICP-MS

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy

The Determination of Toxic, Trace, and Essential Elements in Food Matrices using THGA Coupled with Longitudinal Zeeman Background Correction

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Accurate Sizing and Precise Counting of 10 nm Gold Nanoparticles using the Enhanced Sensitivity of the NexION 2000 ICP-MS

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP

Determination of Total Bromine and Iodine Emission Spectrometric Method (ICP-OES) EuSalt/AS

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

OPTIMIZING METALS SAMPLE PREP

Test Method: CPSC-CH-E

U.S. EPA SW-846 Method 6010C Using the Prodigy7 High- Dispersion ICP Introduction. Application Note - AN1305. Experimental

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Analysis of trace elements in polymer samples to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES

Not Just for Paint Chips Anymore. High Throughput, Open Vessel Microwave Digestion of Matrices Requiring Moderate Conditions

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES

OES - Optical Emission Spectrometer 2000

Optimizing Analytical Performance in ICP-OES Applications

Quality Control Procedures for Graphite Furnace AAS using Avanta Software

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Determination of Phosphorus and Sulfur with Dynamic Reaction Cell ICP-MS

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Determinations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission

Application. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid. Authors. Introduction. Abstract.

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Analysis Repeatability of Trace and Major Elements in a Water Sample

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

Automated Determination of PPQ Levels of Thorium in High Purity Copper

Application note. Determination of exchangeable cations in soil extracts using the Agilent 4100 Microwave Plasma-Atomic Emission Spectrometer

Environmental Series - Trace contaminant analysis in brine using a Thermo Scientific icap 6000 Series Duo ICP

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

A Comparison of Microsecond vs. Millisecond Dwell Times on Particle Number Concentration Measurements by Single Particle ICP-MS

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

PROCEDURES. Pharmacopeial Forum 2 Vol. 36(1) [Jan. Feb. 2009]

High-throughput Quantification of DNA for NGS Library Prep with the Zephyr G3 Workstation and the VICTOR Nivo Plate Reader

INVESTIGATION OF ICP-OES ANALYSIS FOR DETERMINATION OF TRACE LEAD IN LEAD-FREE ALLOY

Samples compliant with the WATER

Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS

High-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application

110 Lowell Rd Hudson, NH USA Tel:

China Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS

Determination of Nutrients. Determination of total phosphorus. Extraktion with aqua regia, reflux method. Introduction

Determination of Trace Metals in Human Urine Using the NexION 300 ICP-MS

Innovative. Intuitive. Concentration (ppb) R = Versatile. Hydra II C Mercury Analyzer

Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3

Product Descriptions and Offerings

Brooklyn College Department of Chemistry

SP-ICP-MS Analysis of Size and Number Concentration in Mixtures of Monometallic and Bimetallic (Core-shell) Nanoparticles

Pharmacopeial Forum Vol. 36(1) [Jan. Feb. 2010] 1

Whole Tablet Measurements Using the Frontier Tablet Autosampler System

DISCLAIMER: This method:

Application of the Agilent 7900 ICP-MS with Method Automation function for the routine determination of trace metallic components in food CRMs

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

á233ñ ELEMENTAL IMPURITIES PROCEDURES

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS

PRINCIPLE OF ICP- AES

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY

Elemental Scientific. Spike Recovery. Elemental Scientific

Trace Analyses in Metal Matrices Using the ELAN DRC II

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Determination of Elements at Sub-ppb Concentrations in Naphtha Mixtures Using the NexION 300 ICP-MS

Analysis of Polyphenols in Saffron Petal Extracts with UHPLC/UV/MS

BATTERY INDUSTRY STANDARD ANALYTICAL METHOD

Atomic Spectroscopy AA/ICP/ICPMS:

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS

Analysis of Pharmaceuticals and Personal Care Products in River Water Samples by UHPLC-TOF

Mercury Determination in Lichen, BCR-482, Using the QuickTrace M-7600 CVAAS Mercury Analyzer

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ

Basic Digestion Principles

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

SeCtiOn 5 [STOCK AND CUSTOM] Assurance. Multi-Element CLP Standards

Fast, Quantitative Analysis of Residual Solvents in Cannabis Concentrates

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry

Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

Determination the elemental composition of soil samples

Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES

Instrumental Limitations for High Absorbance Measurements in Noise

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

Sample Preparation of Electronic Device Components for Hexavalent Chromium Analysis by IEC Method :2017

ICP-OES DETERMINATION OF IRON. Introduction:

(CATION EXCHANGE AND LN RESIN, WITH VACUUM BOX SYSTEM)

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

S.K. Sahoo 1 *, Z.S. Zunic 2, R. Kritsananuwat 1, H. Arae 1 and S. Mishra 1

1.1. This is a method for the separation and measurement of 228 Ra in water via its beta emitting 228 Ac daughter.

Shorten your Time-to-Result

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application

Transcription:

APPLICATION NOTE ICP-Optical Emission Spectroscopy Author: Nick Spivey PerkinElmer, Inc. Shelton, CT Meeting the Challenges of Soil Analysis with the Avio 200 ICP-OES Introduction Micronutrients contained within soil are the building blocks for the crops we eat, process, or feed to livestock. These micronutrients provide the core for proper growth and development of plants which are then consumed and passed up the food chain to eventually reach each of us. As such, the balance of micronutrients in soil is of global importance to growers, producers and consumers. Every soil varies in its basic elemental makeup depending on the local geology and whether the soil has been amended with fertilizers or other additions throughout its use. Without direct elemental analysis, there is no way to confidently understand the condition of the soil and make adjustments to the soil or to efficiently select the crop to be grown in the soil. The analysis of soils for elemental contents presents challenges during the sample preparation step. A common method for preparing a soil sample for inorganic elemental analysis involves digesting the soil sample in an acid that is heated to near-boiling to extract the elements for analysis. When using open vessels in heating blocks, this extraction method typically takes four hours or more to complete. The sample must then be centrifuged or filtered to remove solid particles prior to analysis. The use of a microwave digestion system can speed this up significantly by completing the acid digestion in less than 50 minutes. In addition, if desired, the microwave digestion system can deliver complete sample digestion with the use of appropriate acids, thereby providing an analysis of the total elemental content instead of only the extractable concentrations.

Inductively coupled plasma optical emission spectroscopy (ICP-OES) is generally favored in a multi-element analytical environment with detection capabilities appropriate for the quantification of micronutrients as demonstrated in this application. Flame atomic absorption (AA) systems, which provide cost savings, simplicity, and single-element analytical speed, can be an attractive alternative 1. However, measuring multiple elements by Flame AA requires each sample to be analyzed individually for each element, which can eliminate the speed advantage of Flame AA for multi-element analysis. This work will focus on the analysis of micronutrients in a selection of soil samples using a PerkinElmer Avio 200 ICP Optical Emission Spectrometer. Experimental Samples and Sample Preparation Soils vary greatly from region to region and can be vastly different even within a single large plot of land. To represent some of this variance, soil samples were taken from residential back yards and garden plots and also from commercial farms and pastures. The garden plot samples are interesting in that they represented welltended and highly amended soils and also have a high ratio of industrially composted bagged soil, as commonly available at home and garden stores. All soil samples represented healthy areas in current production and did not show any signs of depletion or disease prior to collection. To confirm the accuracy of the methodology, two certified reference materials were used: Soil Solution A and Soil Solution B (High Purity Standards, Charleston, South Carolina, USA). Both of these samples were analyzed directly, without dilution. A micronutrient extraction method was selected for sample preparation as this represents the most commonly used method for the analysis of soil samples. A mixture of hydrochloric acid and nitric acid was used for the extraction and, as expected, residue remained for every sample after digestion. Soil samples were prepared for analysis by closed-vessel microwave-assisted digestion using a PerkinElmer Titan MPS microwave digestion system. The digestion method, sample parameters, and reagents used are listed in Tables 1 and 2. Samples were weighed and placed into Titan MPS 75 ml Standard Pressure Digestion Vessels, and then the reagents and any sample spikes were added. The samples and reagents then sat open in the vessels for 10 minutes, which allows any early reactions to occur safely. After this time, the vessels were closed and placed into the Titan MPS system for heating and digestion. When the digestion had finished, the samples were transferred out of the digestion vessels by triple-rinsing with deionized (DI) water into sample vials and then diluted to the final volume. After dilution, the samples were set aside to allow the suspended particles to settle out of solution. This can also be achieved more rapidly by centrifuging the samples for around 10 minutes. Table 1. Titan MPS Digestion Method. Step Instrumental Conditions All analyses were performed on an Avio 200 ICP Optical Emission Spectrometer equipped with a PerkinElmer S10 Autosampler. The elements of interest and instrument conditions for the analysis of the soil samples are outlined in Tables 3 and 4. A Meinhard glass nebulizer was used with the standard baffled cyclonic spray chamber and a reduced sample uptake rate of 0.80 ml/min. When combined with the online flow of the internal standard, the total liquid flow rate for the nebulizer was 1 ml/min of solution. External calibration standards were created from stock standards and were diluted with DI water and trace-metal-grade hydrochloric and nitric acid to the final elemental concentrations listed in Table 5. The final acid concentration of the standards was approximately 10% of the acid blend to match the relatively high concentration of acid in the digested and diluted samples. Table 3. Avio 200 ICP-OES Instrumental Parameters. Parameter Temp ( C) Pressure Limit (bar) Ramp Time (min) Value Nebulizer Meinhard Glass Type K1 (Part No. N0777707) Spray Chamber Glass Cyclonic Baffled (Part No. N0791352) Sample Uptake Rate (ml/min) 0.80 RF Power (Watts) 1500 Nebulizer Gas (L/min) 0.70 Auxiliary Gas (L/min) 0.2 Plasma Gas (L/min) 8 Hold Time (min) Power Limit (%) 1 150 35 5 5 80 2 195 35 2 20 100 3 50 35 1 15 0 Table 2. Digestion Information. Parameter Reagents Used Initial Sample Weight Final Solution Weight (after dilution) Volume 6 ml of HCl (37%) and 3 ml of HNO 3 (70%) 1 g 50 g 2

Table 4. Method Parameters. Wavelength (nm) Plasma View Integration Range (sec) Al 308.215 Radial 0.1-2 Ba 233.527 Axial 0.1-5 Ca 317.993 Radial 0.1-2 Co 228.616 Axial 0.1-5 Cu 327.393 Axial 0.1-5 Fe 238.204 Radial 0.1-2 K 766.490 Radial 0.1-2 Mg 285.213 Radial 0.1-2 Mn 257.610 Radial 0.1-2 Na 589.592 Radial 0.1-2 Ni 231.604 Axial 0.1-5 P 178.221 Axial 0.1-5 S 181.975 Axial 0.1-5 V 292.464 Axial 0.1-5 Zn 206.200 Axial 0.1-5 Y (int std) 371.029 Radial 0.1-5 Y (int std) 371.029 Axial 0.1-5 Table 5. Calibration Standards. Std 1 Std 2 Std 3 Std 4 Std 5 Al 25 100 500 Ba 1 10 25 Ca 25 100 500 Co 1 10 25 Cu 1 10 25 Fe 25 100 500 K 25 100 500 Mg 25 100 500 Mn 1 10 25 Na 10 25 100 Ni 1 10 25 P 10 25 100 S 10 25 100 V 1 10 25 Zn 1 10 25 Analysis was performed using standard 2-point background correction with no other spectral correction formulas. Yttrium was included as an internal standard for all elements analyzed using an axial or radial plasma view where appropriate. Leveraging PerkinElmer s proven track record of ICP performance, the Avio 200 spectrometer benefits from a number of unique capabilities as well. The patented Flat Plate plasma technology delivers a robust plasma with zero maintenance and requires no cooling while using nearly half the argon plasma gas of helical load-coil systems. The entire sample introduction system and torch assembly are packaged into a single cassette that is simple to use and maintain. The Avio s Dual View capability allows automated axial or radial viewing, and teamed with its outstanding optical resolution, delivers a large linear dynamic range and exceptional stability and detection limits. All instrument control, sample analysis, and data processing were done through PerkinElmer s Syngistix software. Results and Discussion Calibration results are shown in Table 6. The excellent correlation for the calibration standards demonstrates the accuracy and precision of the Avio 200 ICP-OES. The independent calibration verification (ICV) recoveries ensure that the calibration is valid, and that the creation of the standards is accurate. Table 6. Calibration Results. Correlation Coefficient ICV (% Recovery) Al 0.99999 97 Ba 0.99999 98 Ca 0.99998 101 Co 0.99993 96 Cu 0.99988 96 Fe 0.99999 100 K 0.99992 97 Mg 0.99991 108 Mn 0.99999 102 Na 0.99985 96 Ni 0.99998 97 P 0.99986 98 S 0.99985 98 V 0.99995 99 Zn 0.99990 98 To verify the accuracy of the methodology, two certified reference material soil solutions were measured, with the results shown in Table 7. All concentrations are within 15% of the certified values for both samples, validating the accuracy of the method. The total sample-to-sample analysis time is less than five minutes for the measurement of these 15 elements and internal standards. By comparison, a typical Flame AA instrument would take over 30 minutes to sequentially analyze the same list of elements, not including the required extra calibration time. With the analytical methodology validated, the collected soil samples were prepared and analyzed. Using the Titan MPS digestion system, the samples were simply and quickly prepared for analysis, a major time saving over the typical process of open-vessel digestion. The results from the analysis of the soil samples are shown in Figure 1 and demonstrate the value of quantitative analysis in characterizing the samples. Some soils varied by as much as a factor of 10X in the concentration of specific elements, but otherwise showed surprising consistency. Since all the samples were taken from a relatively small geographic area, this is not entirely unexpected. 3

Table 7. Analysis of Soil Reference Solutions. Soil Solution A Soil Solution B Certified Experimental % Recovery Certified Experimental % Recovery Al 500 459 92 700 662 95 Ba 5 4.75 95 7.00 6.94 99 Ca 350 343 98 125 126 101 Co --- 0.027 --- 0.100 0.087 87 Cu 0.300 0.289 96 3.00 3.01 100 Fe 200 201 101 350 356 102 K 200 196 98 210 210 100 Mg 70 73 104 80.0 82.6 103 Mn 0.100 0.110 110 100 95.2 95 Na 70.0 63.8 91 100 92.5 92 Ni 0.300 0.287 96 0.20 0.20 100 P --- 6.72 --- --- 6.76 --- S --- 1.86 --- --- 2.03 --- V 0.100 0.096 96 0.800 0.772 97 Zn 1.00 1.02 102 70.0 68.9 98 Figure 1. Results from analyses of soil samples (logarithmic scale). With a large dynamic range and the Dual View capability of the Avio 200 ICP-OES, it was not necessary to make multiple sample dilutions for elements at high and low relative concentrations, which meant that every element could be measured in a single analysis at a single sample dilution. This resulted in increased productivity and higher sample throughput. Calibration curves covering the appropriate concentration range for each element, along with the use of calibration check solutions, ICVs, and sample spikes ensured accurate analysis. To assess any remaining matrix effects from the various soil samples after digestion, all samples were spiked prior to digestion with all elements at the levels shown in Table 8. The resulting spike recoveries appear in Figure 2 and show that all recoveries are within 10% of their true values. Effective digestion via the Titan MPS system eliminated the need for per-sample matrix matching or the use of a method of standard addition for accurate analysis. In all cases, the spike concentrations were selected to try to match or approach the expected concentration of the elemental concentration in the samples. This ensures that the spike signal is meaningful in relation to the sample signal for the purposes of analytical evaluation. 4

Table 8. Pre-Digestion Spike Levels (all units in mg/l). Al Ba Ca Co Cu Fe K Mg Mn Na Ni P S V Zn Conc. 5000 500 5000 50 100 5000 5000 5000 500 500 50 500 400 50 50 Figure 2. Spike recoveries in soil samples (% Recovery). The analytical results and spike recoveries for the samples, as well as confirmation of method accuracy in the measurement of CRMs, demonstrate an efficient method for the preparation and analysis of soil samples using a Titan MPS digestion system for sample preparation along with an Avio 200 ICP-OES for final sample analysis. Conclusion This work has demonstrated the ability of the Avio 200 ICP-OES to reliably and effectively analyze a variety of soil samples for an array of elements over a wide range of concentrations. With its extended capabilities, the Avio 200 ICP-OES provides greater multi-element sample throughput when compared to Flame AA, while allowing analysis of elements which are typically challenging for Flame AA (such as phosphorus and sulfur). Using the Titan MPS microwave digestion system simplified sample preparation while greatly increasing throughput and productivity for the laboratory when compared to hot plate or hot block digestions. The ability to completely digest the samples, if needed, is a capability unmatched by open-vessel digestion. The use of the Titan MPS system for sample preparation and the Avio 200 ICP-OES for analysis is an ideal combination for fast, simple, and accurate analyses of micronutrients in soil. Reference 1. Spivey, Nick, Analysis of Micronutrients in Fruit Juice Using FAST Flame Sample Automation for Increased Sample Throughput, Application Note, PerkinElmer, 2015. 5

Consumables Used This work has demonstrated the ability of the Titan MPS Avio 200 ICP-OES Microwave Sample Preparation System and Avio 200 ICP-OES to Component rapidly and accurately measure elements in a Part variety Number of sample Red/Red types PVC Pump which Tubing fall (drain) under the RoHS directive. 09908585 With the proper Black/Black choice PVC Pump of acids, Tubing a (sample) variety of different sample 09908587 types Orange/Green PVC Pump Tubing (internal standard) Autosampler Tubes Instrument Calibration Standard 2 Instrument Calibration Verification Standard Pure-Grade Phosphorus Standard (10,000 mg/l) Pure-Grade Sulfur Standard (1000 mg/l) Pure-Grade Yttrium Standard (10,000 mg/l) N0777110 B0193233 (15 ml) B0193234 (50 ml) N9301721 (125 ml) N9303953 (125 ml) N9303788 (125 ml) N9300139 (500 ml) N9303796 (125 ml) N9300154 (500 ml) N9303810 (125 ml) N9300167 (500 ml) Titan MPS Digestion System Component Consumables Kit for Standard 75 ml Digestion Vessels Rupture Disks for Standard 75 ml Digestion Vessels (25 pieces) Pressure Seal for Standard 75 ml Digestion Vessels (10 pieces) End cap Plug for Gas Containment Manifold Single Lip Seal Forming Tool for Standard 75 ml Digestion Vessels 8-Position Lip Seal Forming Tool for Standard 75 ml Digestion Vessels Part Number N3132000 N3132001 N3132002 N3134004 N3132015 N3132014 PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA P: (800) 762-4000 or (+1) 203-925-4602 www.perkinelmer.com For a complete listing of our global offices, visit www.perkinelmer.com/contactus Copyright 2016, PerkinElmer, Inc. All rights reserved. PerkinElmer is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 012881_01 PKI