Application Note XSP-AN-005. Determination of Ultra Low Chlorine (and Sulfur) in Oil. X-Supreme8000. Key Benefits

Similar documents
S8 TIGER Series 2 for ASTM D 6443

Pulsar. Delivering NMR to your benchtop

The RX-6000 Sulfur in Oil Analyzer For the analysis of Sulfur in Oil and other elements from S to U

Multi-element process analyzer

Elemental analysis by X-ray f luorescence. High performance EDXRF elemental analyzer

NMR and Core Analysis

Defining quality standards for the analysis of solid samples

AZtec LayerProbe TM. Layer thickness and composition characterisation in the SEM. ...the ultimate non-destructive, high-resolution solution.

On-line XRF analysis of multiple elements in liquid process streams C-QUAND

Thermo Scientific ICP-MS solutions for the semiconductor industry. Maximize wafer yields with ultralow elemental detection in chemicals and materials

Which Optistat is right for you?

Reprinted from February Hydrocarbon

Elemental analysis by X-ray f luorescence. Sequential benchtop WDXRF spectrometer

Simplifying the Process: Automated USP 643 / EP

Agilent Cary 630 FTIR sample interfaces combining high performance and versatility

Cap = Capo x A x B x C x D x E

Pittcon External Seminar Template X-ray Analysis in Petrochemical Industry: Challenges and Solutions

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

BOXA-II ON-STREAM X-RAY FLUORESCENCE ANALYZER. Technical Description Version 2 BGRIMM AUTOMATION

Application Note 12: Fully Automated Compound Screening and Verification Using Spinsolve and MestReNova

The DialPath Solution an Easier Way to Analyze Liquids By FTIR. Yanqia Wang, Ph.D. Application Engineer - FTIR Agilent Technologies May 7, 2015

The Plus for your lab!

Test Method Development and Validation as Pertaining to Microtrac Particle Size Measuring Instruments

Measurement & Analytics Measurement made easy. MB3600-CH70 FT-NIR polyol analyzer Pre-calibrated for OH value determination

Analysis of domestic sludge using the Agilent 4200 MP-AES

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

Ideal and powerful solution

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Elemental analysis of river sediment using the Agilent 4200 MP-AES

Level. Automate your processes Automate your lab Innovate with. sensing. EasyPREP. sample handler. Automated Sample Preparation.

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors

University Wet Chemistry Analyzer Packages

Chemical Analysis. Low Level Oxygenates Analyzer. Trace Analysis of Oxygenates in Hydrocarbon Matrices. Gas Chromatography.

Explorative Study of NMR Drilling Fluids Measurement

EFFECT OF CALIBRATION SPECIMEN PREPARATION TECHNIQUES ON NARROW RANGE X-RAY FLUORESCENCE CALIBRATION ACCURACY

Accurate and Precise Sulfur Analysis using the Thermo Scientific FlashSmart Elemental Analyzer

110 Lowell Rd Hudson, NH USA Tel:

POWER + FLEXIBILITY IN ELEMENTAL ANALYSIS

792A AC/DC Transfer Standard

HPLC Sample Preparation

Sample preparation for X-ray fluorescence analysis VII. Liquid sample

FrelTec GmbH. Thick Film Chip Resistor SMD Low Ohm

The. Family of Mercury Analyzers. Hydra II AA Atomic Absorption Detection Liquid Samples

Maintenance with OQ. ZEEnit 700. Operational Qualification OQ within the scope of a maintenance for the. Atomic Absorption Spectrometer

The Vaisala AUTOSONDE AS41 OPERATIONAL EFFICIENCY AND RELIABILITY TO A TOTALLY NEW LEVEL.

881 Compact IC pro / 882 Compact IC plus for ion chromatography. Intelligent. Reliable. Precise.

Burning Through the Confusion: See How Combustion IC Provides Superior Halide Analysis

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Introduction to Fourier Transform Infrared Spectroscopy

Drug Testing Solutions INDIKO DRUG TESTING ANALYZER. 1 P F

ISO INTERNATIONAL STANDARD

Introduction to Fourier Transform Infrared Spectroscopy

Corn Sugar (crude & refined) Analysis F MOISTURE (Karl Fischer)

Dual N-channel field-effect transistor. Two N-channel symmetrical junction field-effect transistors in a SOT363 package.

HYPHENATED TECHNOLOGY GUIDE

881 Compact IC pro 882 Compact IC plus. Compact ion chromatography for intelligent, reliable and precise analysis

On the use of Deming and quadratic calibrations in analytical chemistry

Peter L Warren, Pamela Y Shadforth ICI Technology, Wilton, Middlesbrough, U.K.

CCME Reference Method for the Canada-Wide Standard for Petroleum Hydrocarbons (PHC) in Soil - Tier 1 Method

ALPHA. Innovation with Integrity. The very compact and smart FTIR spectrometer FTIR

S2 PUMA Spectrometry Solutions. Innovation with Integrity EDXRF

SIFT-MS SELECTED ION FLOW TUBE MASS SPECTROMETRY TECHNOLOGY OVERVIEW \ SYFT TECHNOLOGIES

-202mA. Pin 1 D1. Diode. Part Number Case Packaging DMC21D1UDA-7B X2-DFN ,000/Tape & Reel

AEROSOL FILTER ANALYSIS USING POLARIZED OPTICS EDXRF WITH THIN FILM FP METHOD

Technical Note A Quantification and NIST Traceability of Micromatter Thin Film Standards for XRF Analysis. Version March 16, 2017

CHEMISTRY WORKSTATION. The power of proven quality GIVING YOU PEACE OF MIND

Elemental analysis by X-ray fluorescence. High performance, direct excitation EDXRF elemental analyzer

Citric Acid Analysis L-6-1. MOISTURE (Karl Fischer)

Cracking. 191 minutes. 186 marks. Page 1 of 27

HFM 100 Series. Thermal Conductivity Meter for measurement of insulation and construction materials.

Agilent Oil Analyzer: customizing analysis methods

The Claviature of Gas Analysis

In-Situ Analysis of Traces, Minor and Major Elements in Rocks and Soils with a Portable XRF Spectrometer*

RamanStation 400: a Versatile Platform for SERS Analysis

Natural Products. Innovation with Integrity. High Performance NMR Solutions for Analysis NMR

ISO INTERNATIONAL STANDARD

Meeting the Challenges of Soil Analysis with the Avio 200 ICP-OES

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

Power Resistor Thick Film Technology

Application Note S/SL

Ruggedized Electrical Double Layer Energy Storage Capacitors

Altitude influence of elemental distribution in grass from Rila mountain. Dr. E. Nikolova, Dr. A. Artinyan, S. Nikolova INRNE - BAS XRF Laboratory

Agilent Cary 630 FTIR Spectrometer Supporting Organic Synthesis in Academic Teaching Labs

Everyday NMR. Innovation with Integrity. Why infer when you can be sure? NMR

Analytical Instrumentation

Precise and Accurate whole Blood Dispensing with Multipette E3x and Combitips advanced

INTERNATIONAL STANDARD

Agilent G2350A Atomic Emission Detector (AED)

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

SFT3000S Measurement of Sn-Cu Coating

Food Safety and Quality Management System

Specific Accreditation Guidance

4-bit magnitude comparator

N-channel TrenchMOS ultra low level FET. Higher operating power due to low thermal resistance Interfaces directly with low voltage gate drivers

N-channel TrenchMOS logic level FET

I hope you aren't going to tear open the package and count the nails. We agree that. mass of nails = 1340 g g = g

1. Breakthrough quantitative analysis

PSMN B. N-channel TrenchMOS SiliconMAX standard level FET. High frequency computer motherboard DC-to-DC convertors

CONCENTRATED VOLUMETRIC & VOLUMETRIC SOLUTIONS

Transcription:

Determination of Ultra Low Chlorine (and Sulfur) in Oil Application Note XSP-AN-005 Key Benefits The X-Supreme8000 can be used as part of a quality control procedure to ensure petroleum products are manufactured to customer specifications, allowing optimisation of the production process with resultant cost savings. Recently the possible presence of chlorine in crude oils has become a concern in the industry as chlorine in the oil can potentially cause corrosion damage in the refining process. The request was to develop an additional rapid, simple to use, analytical method on the X-Supreme to detect and quantify Chlorine. The resultant calibration demonstrates good performance to detect Chlorine from ppm s to high % s levels even when the oil contains high % levels of sulphur. This allows verification of the Chlorine to confirm product specification and optimisation of the refining process. Analysis using Energy Dispersive X-Ray Fluorescence (EDXRF) Within the Petroleum industry, routine 24/7 quality control analysis for sulphur in a range of petroleum products, e.g. diesel fuel, kerosene, jet fuel, naptha, crude oil, biodiesel etc. is often carried out using simple to use benchtop Energy Dispersive X-ray Fluorescence (EDXRF) analysis such as the Lab-X3500 and X-Supreme8000 instruments. The X-Supreme8000 is the latest benchtop EDXRF analyser from Oxford Instruments offering field proven reliability with simple methods to deliver trusted, reliable accurate analysis. The X-Supreme s compactness and robustness makes it ideal for location either in laboratories or in production sites for twenty-four-hour operation. The X-Supreme includes a tenposition autosampler to enable simple and unattended multiple analysis. This application can now be extended to include Chlorine determination in crude oils offering the same benefits of analysis by EDXRF including: No sample preparation, i.e. the liquid sample is poured directly into the sample cell with no weighing or volumetric measurements required Non-destructive analysis so that the sample can be retained for any future investigation Rapid analysis with provisional results available five seconds after measurement starts and continuously updated until the end of the measurement period allowing a rapid sample assessment to be made X-Supreme8000 1

Determination of Ultra Low Chlorine (and Sulfur) in Oil 1. Sample Preparation and Calibration Standards Key Benefit For highest level of accuracy and traceability, an empirical calibration was derived using a series of commercially available oil standards containing Chlorine and Sulfur. The oil samples were poured directly into Oxford Instruments sample cells (P/No L242) fitted with Poly-M sample film. The quantity of sample is not critical providing its depth is at least two millimetres. However it is advisable to be consistent and fill the sample cup to the internal line (approximately 13ml). A secondary safety window is also fitted with Poly-M film so in the unlikely event of sample leakage the secondary safety window will retain the oil and prevent possible instrument contamination with any associated downtime. Note: Before filling the cup ensure it is free or any hair (hair contains approx. 5% sulphur) and if present in the cup or on the film, will increase the sulphur result on samples. Two optimised calibrations were derived, one covering the ultra low of Chlorine (0-200ppm) and the other for higher levels of chlorine (0-1%) with details shown in Table 1. Standard Ultra low Cl and S method (Standards P/No 54- CM0031) Chlorine ppm Sulfur Wt% Cl and S method (Standards P/No 54- CM0003) Chlorine Wt% Sulfur Wt% #1 0 0.000 0.00 0.00 #2 125 0.100 0.80 0.10 #3 10 0.602 0.50 0.60 #4 150 0.201 0.30 0.20 #5 175 0.700 0.20 0.70 #6 25 0.300 0.60 0.30 #7 200 0.400 0.10 0.40 #8 100 0.500 0.40 0.50 #9 50 0.801 1.00 0.80 #10 75 1.000 0.00 1.00 Table 2: Details of calibration standards for ultra low Cl and S method and for Cl and S method Calibration A B Chlorine 0-200 ppm 0-1% Standard set P/No 54-CM0031 54-CM0003 Number of standards Measurement Time (seconds) 10 10 240 50 Figure 1: Spectrum scan of S: 0.100% and Cl: 125ppm in mineral oil sample Table 1: Calibration standards details 2

Application Note XSP-AN-005 2. Calibration Details Measuring the oil calibration standards shown in Table 2, a regression was obtained as shown in Figures 2 and 3. Key Benefit Excellent correlation demonstrates high accuracy of analysis for both low and high Cl s in crude oil. Routine measurements gives accurate Cl values so any potential corrosion issues can be quickly quantified and minimised/prevented. Figure 2: X-ray calibration (regression line) of Chlorine (0 to 200ppm) showing given versus calculated Figure 3: X-ray calibration (regression line) of Sulfur (0 to 1.0 wt%) showing given versus calculated 3

Determination of Ultra Low Chlorine (and Sulfur) in Oil 3. Results For the low Chlorine method the calibration standards number 3 and 10 were measured ten times to demonstrate instrument repeatability (stability) and results are shown in Table 3. Standard Number 3 Standard Number 10 Measurement Chlorine ppm Sulfur Wt% Chlorine ppm Sulfur Wt% #1 11 0.587 71 1.007 #2 11 0.588 73 1.007 #3 9 0.588 71 1.009 #4 9 0.588 74 1.007 #5 9 0.587 72 1.008 #6 11 0.588 72 1.007 #7 10 0.587 73 1.008 #8 10 0.586 73 1.007 #9 11 0.586 73 1.007 #10 11 0.587 72 1.006 Mean of 10 (Wt %) 10 0.587 72 1.007 Certified 10 0.602 75 1.000 Standard Deviation (1 ) 0.7 0.001 0.8 0.001 Standard Deviation (2 ) 1.4 0.002 1.6 0.002 Table 3: Standard 3 and 10 repeat measurement results using 240 second measurement time each 4

Application Note XSP-AN-005 Typical calibration performance for sulfur and ultra low chlorine in oil method is listed in Table 4. Analyte Calibration range Standard error of calibration Theoretical limit of detection* (3 ) Guaranteed limit of detection* Precision (95% confidence- 2 ) Total analysis time (seconds) Cl 0 200ppm 1.50ppm <1 ppm <1.5 ppm 0.65ppm S 0 1.0 wt% 0.0103 wt% N/A* N/A* 0.002 wt% 240 Table 4: Typical calibration performance for sulfur and ultra low chlorine in oil method (S 0-1.0 wt%; Cl 0-200ppm), n/a = not applicable due to high Analyte Calibration range Standard error of calibration Guaranteed limit of detection* (3 ) Precision (95% confidence-2 ) Total analysis time (seconds) Cl 0 1.0 wt% 0.007 wt% N/A* 0.002 wt% S 0 1.0 wt% 0.008 wt% N/A* 0.003 wt% 50 Table 5: Typical calibration performance for sulfur and chlorine in oil method (Both S and Cl 0-1.0 wt%); n/a = not applicable due to high The precision was calculated from the results of ten measurements on a number of the calibration standards that represented suitable mid-range s for the elements. Conclusion The X-Supreme8000 can determine sulfur and chlorine in oil samples with ease and typical calibration performance shown in Tables 4 and 5. The excellent performance, versatility, ease of use, speed, and cost-effectiveness of this technique make the EDXRF spectrometer the analytical tool of choice for fuel analysis, from the lower detection limits to high levels. 5

Instrument Configuration These methods (P/No XSMS-06A.(Cl) & XSMS-06B.(Cl)) are specific to an X-Supreme which must be fitted with a Titanium target X-ray tube. The instrument plus accessories required to measure Cl and S in oil are listed below. Item Description P/No 1 X-Supreme fitted with Ti tube and SDD 54-B-ZX0008 2 Helium Gas Purge ZX08 3 Liquid Sample Accessories 54-TXPAK1 4 Poly-M sample film 54-L77 5 Method Template XSMS-06A.(Cl) & XSMS-06B.(Cl) 6 Setting Up Sample (SUS) for low Cl and S 54-SU-S40D 7 Setting Up Sample for low S & Cl. Light mineral Oil, white former known as Low Viscosity Mineral Oil (LVMO) ( S <0.2mg/kg) 54-CM0038 Note: These methods are pre-loaded but purchase of calibration standards and measurement are not included in the standard package. X-Supreme8000: Instrument application packages P/No XSP-Sulfur or XSP-Fuels If either the XSP-Sulfur or the XSP-Fuels application packages are ordered then the method templates XSMS-06A.V1 & XSMS- 06B.V1 are included. Note: These methods are pre-loaded but purchase of calibration standards and measurement are not included in the standard package. For more information visit: www.oxford-instruments.com/x-supreme This publication is the copyright of Oxford Instruments plc and provides outline information only, which (unless agreed by the company in writing) may not be used, applied or reproduced for any purpose or form part of any order or contract or regarded as the representation relating to the products or services concerned. Oxford Instruments policy is one of continued improvement. The company reserves the right to alter, without notice the specification, design or conditions of supply of any product or service. Oxford Instruments acknowledges all trademarks and registrations. Oxford Instruments plc, 2013. All rights reserved. Part no: XSP-AN-005