DR/4000 PROCEDURE NITRATE. Using Powder Pillows

Similar documents
DR/4000 PROCEDURE NITRATE. 2. The display will show:

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ )

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls)

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

DR/4000 PROCEDURE. IRON, Total

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

DR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water.

DR/4000 PROCEDURE. HARDNESS, Total

Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials

DR/4000 PROCEDURE LEAD. 2. The display will show: 3. Fill a 250-mL graduated cylinder to the 250-mL mark with sample.

Nitrogen, Ammonia. Test Preparation. Powder Pillows Method Method 8155 Salicylate Method 1 Powder Pillows

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

Mercaptoacetic Acid Method*

NITRITE, Low Range (0 to mg/l NO 2- -N)

DR/4000 PROCEDURE SELENIUM. 4. Measure 100 ml of sample into a second 500-mL erlenmeyer flask (label the flask sample ).

Method to 0.50 mg/l NH 3 N Powder Pillows

Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters)

Nitrogen, Total Inorganic

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

Chromium, Total. Alkaline Hypobromite Oxidation Method 1 Method to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters)

Method to 0.50 mg/l NH 3 N Powder Pillows

Chloramine (Mono) and Nitrogen, Free Ammonia

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

CHLORINE, TOTAL (0 to 4.00 mg/l)

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

Chlorine, Free and Total, Low Range

Chlorine, Free and Total, High Range

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right.

Mercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution

Method 8326 (0.006 to mg/l Al 3+ ) Powder Pillows

Chlorine, Free and Total, High Range

NITROGEN, TOTAL, HR, Test N Tube

Powder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters)

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows

Persulfate Digestion Method Method to 40 mg/l N (HR) TNTplus 827

NITROGEN, AMMONIA, High Range, Test N Tube

Sulfide DOC Methylene Blue Method 1 (0 0.70, and mg/l) Method Test procedure

Cadion Method Method to 0.30 mg/l Cd TNTplus 852

Dimethylphenol Method Method to mg/l NO 3 N or 1.00 to mg/l NO

Chromazurol S Method Method to 0.50 mg/l Al TNTplus 848

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Quaternary Ammonium Compounds

NITROGEN, TOTAL, Test N Tube * Method 10071

PAR Method Method to 2.0 mg/l Pb TNTplus 850

DPD Test N Tube Method *

Scope and application: For wastewater, seawater, drinking water, surface water and process water.

Peracetic Acid (PAA) and Hydrogen Peroxide (H 2 O 2 )

Oxygen Demand, Chemical

Oxygen Demand, Chemical

Phenolphthalein and Total Alkalinity Method to 4000 mg/l as CaCO 3 Digital Titrator

DR/4000 PROCEDURE CYANIDE. 1. Press the soft key under HACH PROGRAM. Select the stored program for cyanide by pressing 1750 with the numeric keys.

Chlorphosphonazo Colorimetric Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow. Instrument Adapter Sample cell

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV

PAR Method Method to 2.0 mg/l Pb TNTplus 850

Salicylate Method Method HR (2 to 47 mg/l NH 3 N) TNTplus 832

Trihalomethane Formation Potential (THMFP)

SILICA, Low Range (0 to 1.60 mg/l)

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

p-dimethylaminobenzaldehyde Method*

Chlorine, Total. USEPA DPD Method 1 Method to mg/l as Cl 2 Chemkey Reagents. Test preparation. Before starting.

CHLORINE, FREE, Ultra-high Range ( mg/l Cl 2 ) Method 10069

Method (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows

DR/4000 PROCEDURE MERCURY. Phase 1: Sample Digestion must be done in a fume hood! Toxic gases may be produced.

TPH (Total Petroleum Hydrocarbons)

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

Manganese III Digestion Method * (with chloride removal)

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

Polychlorinated Biphenyls (PCB) in Soil

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

ALLOWAY METHOD OUTLINE

Experiment 18 - Absorption Spectroscopy and Beer s Law: Analysis of Cu 2+

Ammonia Gas Sensing Combination Electrode

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

METHOD 9200 NITRATE. 1.2 The applicable range of concentration is 0.1 to 2 mg NO -N per liter. 3 of sample.

MERCURY (0.1 to 2.5 µg/l) Method 10065

SIDE BY SIDE COMPARISON TABLE FOR HACH METHOD NUMBER COLORIMETRIC DETERMINATION OF CYANIDE

Ammonia Gas Sensing Combination Electrode

SILICA, Low Range SiLR

PRETREATMENT TECHNICAL DATA SHEET A CHROME-FREE FINAL RINSE PRODUCT DESCRIPTION

POCKET COLORIMETER II ANALYSIS SYSTEMS

METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC)

Sample Preparation LCW 909. Total Kjeldahl Nitrogen, TKN NB: NEW! LCW 909 GB

Experiment 5: Analysis of Nutrients in Natural Waters CH3600 / ESP 5090: Environmental Chemistry, Plymouth State University

Spectrophotometric Determination of Ferrocyanide in Effluents

L Mott SALT WATER AQUACULTURE. Model AQ-4/AQ-5 Code / test kit instruction manual

METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD)

PRETREATMENT TECHNICAL DATA SHEET CHROMIUM-FREE ORGANIC PASSIVATING RINSE PRODUCT DESCRIPTION

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

Introduction to Spectroscopy: Analysis of Copper Ore

Determination of Orthophosphate Ion

PRETREATMENT TECHNICAL DATA SHEET CHROMIUM-FREE ORGANIC PASSIVATING RINSE PRODUCT DESCRIPTION

Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter. Missouri State University.

Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,:

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

Transcription:

DR/4000 PROCEDURE Method 8171 Powder Pillows or AccuVac Ampuls Cadmium Reduction Method MR (0 to 5.0 mg/l NO 3 N) Scope and Application: For water, wastewater and seawater. The estimated detection limit for program numbers 2520 and 2525 are 0.1 and 0.1 mg/l NO 3 -N, respectively. Using Powder Pillows HACH PROGRAM 2 5 2 0 HACH PROGRAM: 2520 Nitrate MR ENTER 1. Press the soft key under HACH PROGRAM. Select the stored program number for mid range nitrate by pressing 2520 with the numeric keys. Press: ENTER Note: If samples cannot be analyzed immediately, see Sample Collection, Storage and Preservation following these steps. Adjust the ph of preserved samples before analysis. 2. The display will show: HACH PROGRAM: 2520 Nitrate MR The wavelength (λ), 400 nm, is automatically selected. Note: For best results, determine a reagent blank for each new lot of reagent as follows. Prepare a reagent blank by repeating steps 3 through 9, using deionized water as the sample. Zero the instrument on deionized water by pressing the soft key under ZERO. Insert the reagent blank and the blank value will be displayed. Correct for the reagent blank by pressing the soft keys under OPTIONS, (MORE), and then BLANK:OFF. Enter the reagent blank value and press ENTER. Repeat for each new lot of reagent. 3. Fill a sample cell with 10 ml of sample. 4. Add the contents of one NitraVer 5 Nitrate Reagent Powder Pillow (the prepared sample). Stopper. Nitrate_AVPP_Mid_CDR_Eng_4000.fm Page 1 of 8

, continued START TIMER START TIMER ZERO 5. Press the soft key under START TIMER. Shake the cell vigorously until the timer beeps in one minute. Note: A deposit of unoxidized metal will remain after the NitraVer 5 dissolves. The deposit will not affect results. Note: Shaking time and technique influence color development. For most accurate results, make successive tests on a 1.00 mg/l Nitrate Nitrogen Standard solution (listed under OPTIONAL REAGENTS AND STANDARDS). Adjust shaking times to obtain the correct result. 6. When the timer beeps, press the soft key under START TIMER. A 5-minute reaction period will begin. Note: An amber color will develop if nitrate nitrogen is present. 7. When the timer beeps, fill a second sample cell with 10 ml of sample (the blank). Place the blank into the cell holder. 8. Press the soft key under ZERO. The display will show: 0.0 mg/l NO 3 N Note: If you have entered a reagent blank correction, the display will show the correction. Note: For alternate concentration units, press the soft key under OPTIONS. Then press the soft key under UNITS to scroll through the available options. Press ENTER to return to the read screen. 9. Place the prepared sample into the cell holder. Close the light shield. The result in mg/l nitrate nitrogen (NO 3 N) will be displayed. Note: Measure sample within two minutes after timer beeps. Note: The result can be expressed as mg/l nitrate (NO 3 ). Press the soft keys under OPTIONS and then FORM: to scroll through the available options. Note: Rinse the sample cell immediately after use to remove all cadmium particles. Retain the spent sample for proper hazardous waste disposal for cadmium. Page 2 of 8 Nitrate_AVPP_Mid_CDR_Eng_4000.fm

, continued Using AccuVac Ampuls HACH PROGRAM 2 5 2 5 HACH PROGRAM: 2525 N, Nitrate MR AV ENTER 1. Press the soft key under HACH PROGRAM. Select the stored program number for mid range nitrate by pressing 2525 with the numeric keys. Press: ENTER Note: If samples cannot be analyzed immediately, see Sample Collection, Storage and Preservation following these steps. Adjust the ph of preserved samples before analysis. 2. The display will show: HACH PROGRAM: 2525 N, Nitrate MR AV The wavelength (λ), 400 nm, is automatically selected. Note: For best results, determine a reagent blank for each new lot of reagent as follows. Prepare a reagent blank by repeating steps 3 through 10, using deionized water as the sample. Zero the instrument on deionized water by pressing the soft key under ZERO. Insert the reagent blank and the blank value will be displayed. Correct for the reagent blank by pressing the soft keys under OPTIONS, (MORE), and then BLANK:OFF. Enter the reagent blank value and press ENTER. Repeat for each new lot of reagent. 3. Collect at least 40 ml of sample in a 50-mL beaker. Fill a NitraVer 5 Nitrate AccuVac Ampul with sample. Place a stopper over the tip of the ampul. Note: Keep the tip immersed while the ampul fills. 4. Press the soft key under START TIMER. A one-minute reaction period will begin. Invert the ampul repeatedly until the timer beeps. Wipe off any liquid or fingerprints. Note: Inversion rate can influence color development. For most accurate results, make successive tests on a 1.00 mg/l Nitrate Nitrogen Standard solution (listed under OPTIONAL REAGENTS AND STANDARDS.) Adjust inversion rate to obtain the correct result. Nitrate_AVPP_Mid_CDR_Eng_4000.fm Page 3 of 8

, continued START TIMER 5. When the timer beeps, press the soft key under START TIMER. A 5-minute reaction period will begin. Note: A deposit of unoxidized metal will remain after the NitraVer 5 dissolves. The deposit will not affect results. Note: An amber color will develop if nitrate nitrogen is present. 6. When the timer beeps, fill a zeroing vial (the blank) with at least 10 ml of sample. 7. Insert the AccuVac Ampul Adapter into the sample cell module by sliding it under the thumb screw and into the alignment grooves. Fasten with the thumb screw. 8. Place the blank into the cell holder. Close the light shield. ZERO 9. Press the soft key under ZERO. The display will show: 0.0 mg/l NO 3 N Note: If you have entered a reagent blank correction, the display will show the correction. Note: For alternate concentration units, press the soft key under OPTIONS. Then press the soft key under UNITS to scroll through the available options. Press ENTER to return to the read screen. 10. Place the AccuVac ampul into the cell holder. Close the light shield. Results in mg/l nitrate expressed as nitrogen (NO 3 N) will be displayed. Note: Measure sample within two minutes after timer beeps. Note: The results can be expressed as mg/l nitrate (NO 3 -). Press the soft keys under OPTIONS, then FORM: to scroll through the available options. Page 4 of 8 Nitrate_AVPP_Mid_CDR_Eng_4000.fm

, continued Interferences Table 1 Interfering Substances and Suggested Treatments Interfering Substance Chloride Ferric iron Nitrite ph Strong oxidizing and reducing substances Interference Level and Treatment Chloride concentrations above 100 mg/l will cause low results. The test may be used at high chloride concentrations (seawater) but a calibration must be done using standards spiked to the same chloride concentration. All levels All levels Compensate for nitrite interference as follows: a) Add 30-g/L Bromine Water dropwise to the sample in Step 3 until a yellow color remains. b) Add one drop of 30-g/L Phenol Solution to destroy the color. c) Proceed with Step 4. Report the results as total nitrate and nitrite. Highly buffered samples or extreme sample ph may exceed the buffering capacity of the reagents and require sample pretreatment. Interfere at all levels Sample Collection, Storage and Preservation Most reliable results are obtained when samples are analyzed as soon as possible after collection. If prompt analysis is impossible, store samples in clean plastic or glass bottles for up to 24 hours at 4 C. To preserve samples for longer periods, add 2 ml of concentrated sulfuric acid (H 2 SO 4 ) per liter and store at 4 C. Accuracy Check Before analysis, warm the sample to room temperature and adjust the ph to 7 with 5.0 N Sodium Hydroxide Standard Solution. Do not use mercury compounds as preservatives. Correct the test result for volume additions by dividing the total volume (acid + base + sample) by the original sample volume and multiplying the test result by this factor. Standard Solution Method To test accuracy, use a 1.00 mg/l Nitrate Nitrogen Standard Solution in place of the sample and perform the procedure as described. To adjust the calibration curve using the reading obtained with the 1.00 mg/l Nitrate Nitrogen Standard Solution, press the soft keys under OPTIONS, (MORE) then STD: OFF. Press ENTER to accept the default concentration, 1.00 mg/l NO 3 N. If an alternate concentration is used, enter the actual concentration and press ENTER to return to the read screen. See Section 1.5.5 Adjusting the Standard Curve for more information. Standard Additions Method a. Leave the unspiked sample in the sample compartment. Verify that the units displayed are in mg/l. Select standard additions mode by pressing the soft keys under OPTIONS, (MORE) and then STD ADD. b. Press ENTER to accept the default sample volume (ml), 25. (This is the volume to which standard addition aliquots are added.) c. Press ENTER to accept the default standard concentration (mg/l), 100. Nitrate_AVPP_Mid_CDR_Eng_4000.fm Page 5 of 8

, continued d. Press the soft key under ENTRY DONE. e. Fill three 25-mL graduated mixing cylinders with 25 ml of sample. f. Snap the neck off a Nitrate Nitrogen Voluette Ampule Standard, 100 mg/l NO 3 N. g. Use the TenSette Pipet to add 0.1, 0.2 and 0.3 ml of standard, respectively, to the three mixing cylinders. Stopper each and mix thoroughly. h. For analysis with AccuVac Ampuls, transfer solutions to dry, clean 50-mL beakers to facilitate filling of the ampules. For analysis with powder pillows, transfer only 10 ml of solution to the 10-mL sample cells. i. Analyze each standard addition sample as described above. Accept the standard additions readings by pressing the soft key under READ each time. Each addition should reflect approximately 100% recovery. j. After completing the sequence, the display will show the extrapolated concentration value and the best-fit line through the standard additions data points, accounting for matrix interferences. k. See Section 1.4.1 Standard Additions for more information. Method Performance Precision Standard: 3.0 mg/l NO 3 N Program 95% Confidence Limits 2520 2.9 3.1 mg/l NO 3 N 2525 2.9 3.1 mg/l NO 3 N For more information on determining precision data and method detection limits, refer to Section 1.5. Estimated Detection Limit Program EDL 2520 0.1 mg/l NO 3 N 2525 0.1 mg/l NO 3 N For more information on derivation and use of Hach s estimated detection limit, see Section 1.5.2. To determine a method detection limit (MDL) as defined by the 40 CFR part 136, appendix B, see Section 1.5.1. Page 6 of 8 Nitrate_AVPP_Mid_CDR_Eng_4000.fm

, continued Sensitivity Program Number: 2520 Portion of Curve Abs Concentration Entire Range 0.010 0.04 mg/l NO 3 N Program Number: 2525 Portion of Curve Abs Concentration Entire Range 0.010 0.04 mg/l NO 3 N See Section 1.5.3 Sensitivity Explained for more information. Calibration Standard Preparation To perform a nitrate calibration using the NitraVer 5 method, prepare calibration standards containing 1, 3, and 5 mg/l NO 3 N as follows: a. Into three different 100-mL Class A volumetric flasks, pipet 1.00, 3.00, and 5.00 ml of a 100-mg/L Nitrate Nitrogen Standard Solution using Class A glassware. b. Dilute to the mark with deionized water. Mix thoroughly. c. Using the NitraVer 5 Powder Pillow or AccuVac method and the calibration procedure described in the User-Entered Programs section of the DR/4000 Spectrophotometer Instrument Manual, generate a calibration curve from the standards prepared above. Summary of Method Cadmium metal reduces nitrates in the sample to nitrite. The nitrite ion reacts in an acidic medium with sulfanilic acid to form an intermediate diazonium salt. The salt couples with gentisic acid to form an amber colored solution. Safety Good safety habits and laboratory techniques should be used throughout the procedure. Consult the Material Safety Data Sheet for information specific to the reagents used. For additional information, refer to Section 1. Pollution Prevention and Waste Management Prepared samples will contain cadmium and must be disposed of according to Federal, State and local hazardous waste regulations. For information on pollution prevention and waste management, refer to Section 1. Nitrate_AVPP_Mid_CDR_Eng_4000.fm Page 7 of 8

, continued REQUIRED REAGENTS AND STANDARDS (Using Powder Pillows) Quantity Required Description per test Unit Cat. No. NitraVer 5 Nitrate Reagent Powder Pillows (for 10-mL sample).. 1 pillow... 100/pkg...21061-69 REQUIRED REAGENTS AND STANDARDS (Using AccuVac Ampuls) NitraVer 5 Nitrate Reagent AccuVac Ampul... 1 ampul... 25/pkg...25110-25 REQUIRED EQUIPMENT AND SUPPLIES (Using Powder Pillows) DR/4000 1-Inch Cell Adapter...1... each...48190-00 Sample Cells, matched pair, 1-inch, glass, with stoppers...2... pair...26126-02 REQUIRED EQUIPMENT AND SUPPLIES (Using AccuVac Ampuls) Beaker, 50-mL...1... each...500-41 DR/4000 AccuVac Ampul Adapter...1... each...48187-00 Sample Cell, 10-mL with cap (zeroing vial)...1... each...21228-00 Stopper, for 18-mm tube...1... 6/pkg...1731-06 OPTIONAL REAGENTS AND STANDARDS Bromine Water, 30-g/L...29 ml*...2211-20 Nitrate Nitrogen Standard Solution, 1.00-mg/L NO 3 N...500 ml...2046-49 Nitrate Nitrogen Standard Solution, 100-mg/L NO 3 N...500 ml...1947-49 Nitrate Nitrogen Standard Solution, 2-mL PourRite Ampule, 100-mg/L NO 3 N... 20/pkg...1947-20 Phenol Solution, 30-g/L...29 ml*...2112-20 Sodium Hydroxide Standard Solution, 5.0 N...1 liter...2450-53 Sulfuric Acid, ACS, concentrated...500 ml*...979-49 Water, deionized... 4 liters...272-56 OPTIONAL EQUIPMENT AND SUPPLIES AccuVac Snapper... each...24052-00 Cylinder, graduated, mixing, 25-mL... each...20886-40 Dropper, for 29-mL bottle... each...2258-00 DR/4000 Carousel Module Kit... each...48070-02 Flask, volumetric, 100-mL, Class A... each...14574-42 ph Paper, ph 1.0 to 11.0... 5/pkg...391-33 Pipet, serological, 2-mL... each...532-36 Pipet, TenSette, 0.1 to 1.0 ml... each...19700-01 Pipet Tips for 19700-01 TenSette Pipet... 50/pkg...21856-96 Pipet, volumetric, Class A, 1.00-mL... each...14515-35 Pipet, volumetric, Class A, 3.00-mL... each...14515-03 Pipet, volumetric, Class A, 5.00-mL... each...14515-37 Pipet Filler, safety bulb... each...14651-00 PourRite Ampule Breaker... each...24846-00 * Larger sizes available FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: In the U.S.A. Call toll-free 800-227-4224 Outside the U.S.A. Contact the HACH office or distributor serving you. On the Worldwide Web www.hach.com; E-mail techhelp@hach.com HACH COMPANY WORLD HEADQUARTERS Telephone: (970) 669-3050 FAX: (970) 669-2932 Hach Company, 1997-2003. All rights reserved. Printed in the U.S.A. 6/03 11ed Page 8 of 8 Nitrate_AVPP_Mid_CDR_Eng_4000.fm