Mass Spectrometry. Fundamental LC-MS. Solvents, Buffers and Additives

Similar documents
The Theory of HPLC. Ion Pair Chromatography

Chemistry Instrumental Analysis Lecture 37. Chem 4631

M M e M M H M M H. Ion Sources

Electrospray Ion Trap Mass Spectrometry. Introduction

Reversed Phase Solvents

LC and LC/MS Column Selection Flow Chart

Determination of trace anions in concentrated hydrofluoric acid

HPLC solvents and mobile phase additives

LC Technical Information

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD)

LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC

Selecting an LC/MS Interface Becky Wittrig, Ph.D.

Packings for HPLC. Packings for HPLC

Shodex TM ODP2 HP series columns

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

penta-hilic UHPLC COLUMNS

Mass Spectrometry. Fundamental LC-MS. Electrospray Ionisation Instrumentation

Analysis - HPLC A.136. Primesep 5 µm columns B.136

Mass Spectrometry. Fundamental LC-MS. Atmospheric Pressure Chemical Ionisation (APCI)

Mass Spectrometry. Electron Ionization and Chemical Ionization

ProPac WCX-10 Columns

Chromatography. Gas Chromatography

Comparison of different aqueous mobile phase HPLC techniques

Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS

Too Polar for Reversed Phase What Do You Do?

Product Bulletin. ACE LC/MS and Rapid Analysis HPLC Columns. HPLC Columns

Mechanisms of retention in HPLC

API Optimization, Source Design, and Other Considerations of LC/MS

Simultaneous Determination of Paraquat and Diquat in Environmental Water Samples by HPLC-MS/MS

Dionex IonPac AS28-Fast-4µm column

High Performance Liquid Chromatography

High Performance Liquid Chromatography

Utilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography. Tips and Techniques to Optimize ELSD and MS based Purification

HPLC Winter Webinars Part 2: Sample Preparation for HPLC

HICHROM. Chromatography Columns and Supplies. LC COLUMNS SIELC Primesep. Catalogue 9. Hichrom Limited

LC MS analysis of metabolites. Basis of Chromatography

Hints for Strong Ion Exchange Resins

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Protein separation and characterization

penta-hilic UHPLC COLUMNS

Lecture 15: Introduction to mass spectrometry-i

Chemistry Instrumental Analysis Lecture 28. Chem 4631

for Acclaim Mixed-Mode HILIC-1 Column

LC-MS Based Metabolomics

Mass Spectrometry (MS)

Webinar: How to be More Successful with your ACQUITY QDa Detector Eric van Beelen

High Pressure/Performance Liquid Chromatography (HPLC)

Open Column Chromatography, GC, TLC, and HPLC

Method Development in Solid Phase Extraction using Non-Polar ISOLUTE SPE Columns for the Extraction of Aqueous Samples

Routine MS Detection for USP Chromatographic Methods

Experiment 1: Thin Layer Chromatography

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics

Dilution(*) Chromatography

770-9P. Sensitivity and Selectivity - A Case Study of LC/MS Enantioselective Resolution of Bupivacaine Using Vancomycin as a Chiral Stationary Phase

for Acclaim Trinity TM P1

Fall 2012 Due In Class Friday, Oct. 19. Complete the following on separate paper. Show your work and clearly identify your answers.

Instrumental Chemical Analysis

Zirconia-Based Phases as a Powerful Complement to Silica-Based Phases for LC and LC-MS under Non-extreme Mobile Phase Conditions

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS)

High Performance Liquid Chromatography

CHROMATOGRAPHY. The term "chromatography" is derived from the original use of this method for separating yellow and green plant pigments.

HPLC Background Chem 250 F 2008 Page 1 of 24

Introduction to Pharmaceutical Chemical Analysis

Chapter content. Reference

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer

Abstract: An minimalist overview of chromatography for the person who would conduct chromatographic experiments, but not design experiments.

HICHROM. Chromatography Columns and Supplies. LC COLUMNS SeQuant ZIC-HILIC. Catalogue 9. Hichrom Limited

FAIMS Technology at a Glance

Cation Exchange HPLC Columns

High Performance Liquid Chromatography

Developing a fast, generic method for rapid resolution LC with quadrupole MS detection. Application Note

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

Novel LC/MS-Compatible Stationary Phase with Polar Selectivity

Rapid Post-Blast Inorganic Explosive Analysis by Suppressed Ion Chromatography

The Theory of HPLC. Quantitative and Qualitative HPLC

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

FAQ's. 1 - Which column would be the most appropriate for my application?

Chapter 5. Irbesartan. Page no. 64 to 80

Theory and Instrumentation of GC. Chromatographic Parameters

1) Which of the following represents the breaking of a noncovalent interaction? Topic: The Nature of Noncovalent Interactions

Acclaim Mixed-Mode WCX-1

LC-MS compatible Separation of the Fungicide Spiroxamine

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

Theory of Headspace Sampling

The Secrets of Rapid HPLC Method Development. Choosing Columns for Rapid Method Development and Short Analysis Times

Introduction. Chapter 1. Learning Objectives

Ion Chromatography. Anion Exchange. Chromatography Ion Exchange Theory. Dr. Shulamit Levin

HPLC analysis of VFAs, Furfural and HMF

TANDEM MASS SPECTROSCOPY

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment

Thermo Scientific. Acclaim Trinity P2. Column Product Manual. P/N: October Part of Thermo Fisher Scientific

Yuhui Yang, An Trinh, Michael Ye, and Tom Henderson 595 North Harrison Road, Bellefonte, PA T HKB

LC800. Smart HPLC. Until now UHPLC From now Smart HPLC

Quantitative analysis of small molecules in biological samples. Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB.

1. Ion exchange chromatography

LC III: HPLC. Originally referred to as High-Pressure Liquid Chromatography. Now more commonly called High Performance Liquid Chromatography

columns Acclaim Mixed-Mode WCX-1 for Separating Basic Molecules

HILIC Method Development in a Few Simple Steps

Transcription:

Mass Spectrometry Fundamental LC-MS Solvents, Buffers and Additives i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference manual.

Aims and bjectives Aims and bjectives Aims Examine common additives currently used in conventional LC/MS Explain the role of solvents, buffers and eluent ph in signal optimisation Explain the use of volatile ion pair reagents for successful electrospray bjectives At the end of this Section you should be able to: Identify which additives are suitable for ESI and APCI Review and discuss the most important parameters that need to be considered when selecting additives for LC/MS List source considerations that need to be taken into account for sample and eluent types in LC-(API)MS Explain technical solutions that increase the compatibility of non-voltile components with API interfaces

Content Introduction 3 LC-API Compatibility 4 ESI Solution Chemistry 4 ESI Eluent Solvent 6 Solvent Viscosity in ESI 7 rganic Modifiers in ESI 8 Eluent Solvent Positive ESI 9 Eluent ph Positive ESI 10 Eluent ph egative ESI 11 ESI Positive/egative Ion Mode 13 ESI Reagents for ph Control 15 Sources for non-volatile Systems 17 Z-Spray and non-volatile Systems 19 ESI Test compound infusion 20 ESI Buffer Choice and Concentration 22 ESI Buffer Systems 24 ESI Ion Pair Reagents verview 25 ESI Ion Pair Reagents Considerations 27 Cationisation vs. Anionisation in ESI 28 Pre and Post Column Addition in ESI 30 Post Column Addition in ESI 31 ESI Adduct Formation 33 ESI Considerations 35 APCI Solvent Choice verview 35 APCI Solvent Choice Mechanisms 37 Buffers for APCI verview 38 Buffers for APCI Concentration 39 Ion pairing reagents in APCI 41 APCI Considerations 42 References 42 Crawford Scientific www.chromacademy.com 2

Introduction Solvents, buffers and additives commonly used in conventional LC, in which detection relies on analytes being in solution, may not be compatible with LC-MS analysis. Mass spectrometric detection relies on desolvation to produce gas phase analyte ions with complete removal of other eluent components. An efficient process will result in optimised analyte response and increased sensitivity. ESI favours ions in solution. In contrast, reversed-phase liquid chromatographic conditions generally attempt to suppress ionisation (by increasing the analyte hydrophobicity), and hence gain retention on the hydrophobic stationary phase. [1] ormalphase eluents are less polar, do not support ions in solution and are not compatible with ESI-MS. APCI is compatible with non-polar mobile phases. on-polar analytes are usually more volatile and therefore good candidates for APCI that relies on the ability to quickly volatilise eluent systems. APCI is most compatible with neutral analytes that are able to accept or donate protons in the gas phase. LC/MS system Crawford Scientific www.chromacademy.com 3

LC-API Compatibility The compatibility of the various modes of HPLC with ESI and APCI is governed by the fact that each mode relies on different ion production mechanisms. This coupled with the ability to perform each technique in positive and negative ion mode means that most conventional LC methods can be adapted to LC-MS. HPLC type compatibility with ESI and APCI can be summarized as follows (for more information regarding the separation mechanism on each HPLC mode you can refer to the HPLC Channel): [1] ormal Phase chromatography is compatible with all API interfaces; however, much of its analytical domain is covered by GC-MS. Reverse Phase Chromatography, there is no question that this is the most widely separation technology in LC-MS. It is compatible with all API interfaces. Size Exclusion/Gel Filtration chromatography is highly compatible with ESI-MS, aqueous mobile phases and samples such as proteins, which carry a charge, are excellent matches. Gel Filtration chromatography is not compatible with APCI, the samples are non-volatile and the buffers may cause source fouling problems. Size Exclusion/Gel Permeation chromatography is not compatible with ESI-MS due to the highly organic mobile phases used. Ion Exchange Chromatography presents low compatibility with API ionisation techniques, ionic compounds require very high temperatures to get into the gas phase, furthermore the high ionic strength of the mobile phase can suppress ion production. Table 1. HPLC compatibility with ESI and APCI. HPLC TYPE MDE ESI APCI Reversed Phase High Moderate ormal Phase Low High Size Exclusion/Gel filtration High Low Size Exclusion/Gel permeation Low High Ion Exchange Low Low ESI Solution Chemistry The primary considerations for the solution to be analysed using LC-(ESI)MS are: The eluent flow rate and chemistry The sample pka Solution ph Solution conductivity (governed by the type and concentration of buffer salts, and other mobile phase additives such as ion pairing reagents or ph modifiers) The surface tension of the eluent The factors listed below need to be considered when transferring existing LC methods or when developing methods for LC-(ESI)MS: Solvents used Buffer salts used and the concentration of buffers ph of the eluent and ph adjusting species Ion pairing reagents Crawford Scientific www.chromacademy.com 4

Electrospray ionisation is highly compatible with analytes possessing the following characteristics: Moderate to highly polarity Up to 100,000 Dalton Ionise in solution perhaps with multiple charge Electrospray ionisation favours the analyte in the ionised form, in eluent solution prior to introduction into the API interface. i Important: The ESI interface Electrospray API interfaces rely mainly on desorbation to achieve gas phase ions. The analyte must be charged in the eluent solution for highest sensitivity which can present problems when using reverse phase HPLC for the separation. All HPLC eluent components should be chosen to: Increase volatility Lower the ionic strength Achieve a ph at which the analyte is ionised in solution Avoid ion-pairing reagents (which can suppress the analyte signal) Avoid the use of involatile buffers (which foul the source and compromise voltages applied to the API source) Crawford Scientific www.chromacademy.com 5

ESI Eluent Solvent Many typical HPLC solvents are compatible with atmospheric pressure ionisation techniques. A good electrospray solvent will easily support ions in solution, this generally requires that the solvent has some dipole moment (polarity). on-polar solvents (such as toluene) do not easily support ions in solution, cannot be used for ESI. [2,3,4] Although water is an excellent solvent for ions, its lower solvation energy (with polar analytes), and high surface tension, make desolvation and ion desorption more difficult, often requiring higher drying gas temperatures, higher nebulising gas flow rates and/or higher nebulising voltages. on-polar solvents like benzene, hexane, cyclohexane and carbon tetrachloride are unsuitable for electrospray without addition of secondary electrolyte species. However, this problem can be overcome by a 10% addition of polar solvent (usually methanol or isopropanol), or a buffer electrolyte. Reducing the effects of surface tension, viscosity and heat of vaporisation will promote high sensitivity in ESI. A study of acids, bases, amphoteric and polar neutral drug compounds (see below), showed that in ESI(+), methanol generally gave stronger signals than acetonitrile. [5] Responses were typically 10-20% higher, however some compounds showed 50-fold increases in sensitivity when ammonium formate was used as the buffer electrolyte, possibly due to changes in the degree of analyte ionisation combined with the higher ability of the solution to support ions in solution. Interestingly, these increases in sensitivity when using methanol rather than acetonitrile were not seen in negative ion mode. i Important: Important: As all values are greater than 1, this suggests that in all cases the instrument response using methanol as the eluent organic modifier was greater than the comparable experiment using acetonitrile as the organic modifier Crawford Scientific www.chromacademy.com 6

Table 2. Analytes of interest. ame Structure Molecular Formula Caffeine (194.19 Da) C 8 H 10 2 Paracetamol (151.17 Da) H C 8 H 9 2 Tryptohan (204.23 Da) H H C 11 H 12 2 2. H H 2 Tolbutamide (270.35 Da) C 12 H 18 2 3 S S H Indomethacin (357.79 Da) Cl C 19 H 16 Cl 4 H Solvent Viscosity in ESI Solvents having higher viscosity tend to be less volatile and have higher surface tension. [6] A high viscosity solvent will form larger, less efficiently charged droplets at the capillary tip and will require higher interface drying gas temperatures to efficiently de-solvate the electrosprayed eluent droplets. Desolvation is a critical step leading to desorption of ions. Solvents must be able to efficiently evaporate from the droplet to reduce its overall size. The ions should also be solvophobic to ensure the surface of the droplet has a high density of surface ions leading to efficient Rayleigh droplet jet fission. In general the eluent system should be designed with minimum surface tension, minimum solvation energy and reduced viscosity for optimum sensitivity in ESI-MS. ptimising these parameters will result in correct droplet nebulisation, promote droplet desolvation and aid ion evaporation from the de-solvated droplets. Crawford Scientific www.chromacademy.com 7

i Increasing eluent methanol content leads to more efficient droplet desolvation. This in turn leads to an increase in the gas phase ions available for sampling into the mass analyser and hence an increase in ion abundance. rganic Modifiers in ESI Higher percentages of organic modifier within LC eluent systems tend to result in better ESI sensitivity due to lowering surface tension and solvation energies for polar compounds. Further, the increased volatility of the organic modifier allows for more efficient desolvation of the electrosprayed droplet. In the example below, [7] the signal intensity of Penicillin G against % organic modifier in the electrosprayed eluent are shown. In this example the absolute response varies between 75 and 110% relative intensity as the percentage organic modifier (in this case acetonitrile) is altered this may result in robustness issues if the eluent system is not accurately prepared for the analysis even in isocratic mode! Response of penicillin G with varing percentages of acetonitrile. Crawford Scientific www.chromacademy.com 8

Further, the absolute analyte response MAY change when comparing early and later eluting species in gradient HPLC analysis. The changing nature of the eluent system (usually containing a greater percentage of an eluent component more suitable for ESI- MS i.e. the organic modifier), may result in improved response for later eluting analytes. With this in mind it is not good practice to directly compare the response of analytes with widely differing retention times in gradient HPLC analysis. Typical examples of this type of analysis include area normalisation approaches to pharmaceutical stability analysis or quantification of pesticide residues in river water extracts etc. This is highlighted in the figure below where the relative response ratio of ethanol to heptanol is 0.72 whereas all analytes are present in the sample solution at equal concentration! Analysis of homologous alcohols (equal concentrations). The measured ethanol(c 2 ):heptanol(c 7 ) ratio is 0.72 (true value 1) Eluent Solvent Positive ESI ESI-MS requires the analyte molecule is ionised in solution to obtain the highest sensitivity. This however, has implications with analyte retention in reverse phase HPLC separations. Although ionisation can be induced (through tribo-charging and electrostatic effects), analytes sprayed in the ionic form give rise to more robust, reproducible and predictable MS response. Positive ion mode electrospray involves the creation of [M+H] + ions (for basic species) or [M-H] - ions (for acidic species) in solution. The generation of these ions is a function of the sample pka and the eluent solution ph. A basic analyte will become protonated (and positively charged) if the eluent solution is acidic (below the pka of the molecule) and an acidic analyte will loose a proton to form an anion when the eluent ph is above the molecule pka. Ionisation is an equilibrium process Crawford Scientific www.chromacademy.com 9

and sample sensitivity depends on the degree of ionisation of the analyte. Knowing the pka of the analyte functional groups will allow the user to predict the required eluent solution ph for maximum sensitivity ESI-MS. i Production of charged droplets Eluent ph Positive ESI The effect of changing eluent ph in positive ESI-MS is illustrated below with phenytoin (an anticonvulsant drug, 252.3 Dalton). [8] If the ph of the eluent solution is adjusted to match the pka of the basic functional group, 50% of the analyte molecules will be ionised at any one instant. Using an eluent ph equal to the pka is not ideal as retention times can be variable and affected by small changes in the eluent ph. By raising the eluent ph by 2 units the basic functional group will be 99.5% non-ionised and although it s retention time will increase, the signal intensity will decrease. Conversely, by lowering the eluent ph 2 units below the pka, the basic group will be 99.5% ionised and will show good ESI-MS positive mode sensitivity, although it s retention time will be considerably shortened. Compounds that have functional groups that protonate readily such as basic nitrogens show good sensitivity. Those that are polar but contain no basic nitrogen atoms show moderate sensitivity. Hydrocarbons have poor positive mode ESI-MS sensitivity due to the lack of ionisable or polarisable functional groups. Crawford Scientific www.chromacademy.com 10

i Eluent ph effect in the analysis of a basic sample ote how in the equilibrium R 3 H2 R3 H H When ph increases (H - addition) decreases dissociation When ph decreases (H 3 + addition) increases dissociation Eluent ph egative ESI Solution chemistry for negative ion analysis involves the creation of [M-H] - ions in solution and the generation of these ions in solution is again a function of the sample pka and the eluent solution ph. It is important to establish the pka of the various acidic functional groups within the analyte molecule to allow prediction of the eluent ph that will give rise to the highest ESI-MS sensitivity. The sample will loose a proton in solution when the eluent ph is increased and become negatively charged. The ph effect in negative ESI-MS is illustrated opposite with indomethacin (antiinnflamatory & pain killer, 345.5 Dalton). [9] By raising the eluent ph, two ph units above the pka, the acidic group will be 99.5% ionised and will show good ESI-MS negative Crawford Scientific www.chromacademy.com 11

mode sensitivity, however it s rerention time in reverse phase HPLC will be significantly reduced. By reducing the eluent ph below the analyte pka, the retention time of the analyte will increase but the signal intensity will be significantly reduced. For negative ionisation, analytes with functional groups that deprotonate readily, such as carboxylic or sulfonic acids, show the best sensitivity. Analytes that are polar but contain no acidic groups show less sensitivity. i Effect of ph on retention time ote how in the equilibrium HAc H Ac 2 H3 When ph increases (H - addition) dissociation increases When ph decreases (H 3 + addition) dissociation decreases Crawford Scientific www.chromacademy.com 12

ESI Positive/egative Ion Mode For ESI positive ion mode, analytes with functional groups that protonate readily, such as amines show good sensitivity. For ESI negative ion mode, analytes with functional groups that deprotonate readily, such as carboxylic or sulfonic acids, show the best sensitivity. Analytes presenting polar but no highly ionisable functional groups can be analysed in either positive or negative ion mode (according to their ability to gain or loose a proton respectively), in both cases sensitivity is moderate. Certain analytes can be monitored in both ion modes (positive and negative). The ph of the solution will determine which mode and degree of protonation/deprotonation that is observed. Knowing the pka of the analyte s functional groups will allow the user to predict the required eluent solution ph for maximum sensitivity. A basic analyte will become protonated if the eluent ph is kept below its pka, in a similar manner an acidic analyte will protonate if the eluent ph is maintained above its pka. The amino acid lysine presents three functional groups (one carboxylic acid and two amines), according to the ph of the medium, this molecule can hold different charge states (+2, +1, 0, or -1) and it can be analysed under different ESI ionisation modes (positive and negative). Lysine acid base equilibrium Crawford Scientific www.chromacademy.com 13

i Lysine acid base equilibrium (continuation) Crawford Scientific www.chromacademy.com 14

ESI Reagents for ph Control Many conventional LC separations use buffers and additives to control ph and hence ionisation of analytes. In LC-MS, non-volatile eluent components should be avoided. Such components are deposited on critical surfaces (contamination), which in turn adversely affect the electrolytic formation of charged droplets, leading to reduced sensitivity and limit the operational lifetime between cleaning. If the interface is allowed to coat with involatile salt species, over time the signal intensity will reduce due to reduced apparent potential differences and non-homogenous fields within the API interface. Below, the ESI(+)-MS analysis of beta blockers (a class of drugs of particular importance for the management of cardiac arrhythmias and cardio-protection after myocardial infarction) is presented. Here the use of an involatile buffer (potassium phosphate) [11, 12] drastically reduces signal intensity. i ESI(+)-MS analysis of beta blockers. Eluent system: 20%MeC + 80% Potassium phosphate (ph = 3.0, 20mM), 1.0mL/min on a ZRBAX SB C18, 2.1mm 50mm, 1.8μm Crawford Scientific www.chromacademy.com 15

Table 3. Analytes of interest. Label Structure Molecular Formula A Atenolol (266.34 Da) H B adolol (309.41 Da) H H H 2 H H H H C Pindolol (248.32 Da) H H D Timolol,(316.42 Da) H H H S. H H E Metoprolol (267.36 Da) H F xprenolol (265.35 Da) H H H G Labetalol (364.90 Da) H H 2 H H H Propanolol (259.35 Da) H H I Alprenolol (239.35 Da) H H Crawford Scientific www.chromacademy.com 16

Sources for non-volatile Systems Contamination and blocking of the electrospray capillary or any ion transfer tubes will also cause disruption to the chagrining of the component, but these components can also generally be restored by cleaning. Instrument manufacturers have produced interfaces that can increase the compatibility of non-voltile components with API interfaces. These include: 1. Turbo-ionspray from Applied Biosystems / MDS-Sciex (Foster City, CA, USA) i 2. rthogonally sprayed sources such as the Z-Spray source from Micromass (Manchester, UK) Crawford Scientific www.chromacademy.com 17

3. Sources employing rthogonal Spraying and Curtain gas technology such as the 59987A Source from Agilent Technologies (Palo-Alto, CA, USA) 4. Continuously washed sources such as the Aqa from Thermo Finnigan (Riviera Beach, Ca, USA) Crawford Scientific www.chromacademy.com 18

Z-Spray and non-volatile Systems Using an orthogonally sprayed source (Z-spray, Waters, Manchester, UK), the non-volatile buffer salts are deposited onto a baffle plate and are not directly deposited on the critical electrode surfaces or within the sampling plate orifice. This allows for greater instrument operating periods between source cleaning. Table 4. Components of the Hanks Balanced salt solution. Hanks balanced salt solution Concentration Calcium Chloride 1.0 mm Potassium Chloride 5.0 mm Sodium Chloride 138.0 mm Magnesium Sulphate 1.0 mm Sodium Bicarbonate 4.0 mm Phosphates 1.0 mm Glucose 5.0 mm Total 10.0 g/l Probenecid is a uricosuric drug (substances that increase the excretion of uric acid), it is primarily used in treating gout or hyperuricemia. Probenecid presents limited solubility in HPLC reverse phase solvents; however, it can be solubilised in acetonitrile with the Hanks balanced salt solution. [13] ESI(+)-MS determination of probenecid on a Xterra C18, 2.1mm 20mm, 2.5μm Crawford Scientific www.chromacademy.com 19

The experiment was designed to test the ruggedness of the Z-Spray orthogonal source and it was found that after 30 injection cycles, the relative response remains almost constant whilst under the same conditions the relative response using an on-axis electrospray decreases by 70-80%. ESI(+)-MS determination of probenecid on a Xterra C18, 2.1mm 20mm, 2.5μm Crawford Scientific www.chromacademy.com 20

ESI Test compound infusion In order to study the effects of varying buffers and additives on the ESI-MS process and to optimise the interface parameters, infusion devices may be employed to study the instrument signal over extended periods of time. Infusion is the continuous delivery of sample solution into the API interface via a syringe pump or from a manual injection valve. Infused data generally provide high sensitivity due to the ability to collect data with slow MS scanning speeds over extended periods of time. Infusion requires pure samples or mixtures (usually not more than five components). ften, problems of altered relative response (ion suppression) are observed when infusing solutions containing more than one standard material. The sample size has to be large due to the volumes consumed during the constant infusion process unless nanoelectrospray sources are available. With manual injection, the baseline signal rises as the sample enters the mass spectrometer, followed by a steady state as the material is slowly introduced from the loop. The signal then falls exponentially as the sample is exhausted and the injection valve is emptied. Depending upon the type of injection valve used and the loop size, several minutes of data should be obtainable. i Infusion pump Crawford Scientific www.chromacademy.com 21

ESI Buffer Choice and Concentration In ESI-MS the buffer salts present within the eluent solution and in the sample matrix may interfere with crucial API processes. When choosing an appropriate volatile buffer the nature of the buffer species should be considered. In both positive and negative ion mode an increase in the buffer concentration lead to a decrease in the analyte signal. [5] However, this effect is compound dependant, with some analytes showing only a small loss in response. This effect is presented opposite, note that compounds such as indomethacin and caffeine, showed a 100-fold decrease in response from 1 to 100 mm ammonium acetate buffer concentration, but some others like lidocaine remains almost constant. At higher buffer ion concentrations the competition for surface sites on the electrosprayed droplet favours the higher concentration species and analyte species become depleted. In negative ion mode formic acid and ammonium formate performed equally well, but trifluoroacetic acid (TFA) suppressed all of the sample signals, making it unsuitable for ESI/MS (the results are not shown). In positive ion mode, the analyte response decreased across the series (10 mm eluent solutions) in the order: Formic acid (100%) > ammonium formate (60-85%) > TFA (30-50%). [5] ESI analysis of some pharmacologically active species Crawford Scientific www.chromacademy.com 22

These results suggest that the concentration of acetate and formate buffer additives should be kept to a minimum when good MS response is required. However, this should be balanced with the chromatographic requirements of the system (highly base deactivated columns may not be able to produce good chromatography at buffer concentrations as low as 1 mm). Formic acid may be a better overall choice of eluent electrolyte, as the relative independence of the response with respect to concentration is advantageous. Table 5. Analytes of interest. ame Structure Molecular Formula Molecular weight (Da) Caffeine C 8 H 10 2 194.19 Paracetamol H C 8 H 9 2 151.17 Warfarine H C 19 H 16 4 308.33 H Lidocaine H C 14 H 22 2 234.34 Tolbutamide C 12 H 18 2 3 S 270.35 S H Indomethacin Cl C 19 H 16 Cl 4 357.79 H Crawford Scientific www.chromacademy.com 23

ESI Buffer Systems When choosing an appropriate buffer system, the results from the previous experiment need to be considered in a practical context. [5] Eluent buffer species will compete for surface sites on the electrosprayed droplets. Droplet surface residence of the analyte is crucial to the ion evaporation process, and as such the ability of the buffer species to reside at the droplet surface should be minimized. This can be achieved by using buffer species that have low solution mobility. Ideally these buffer systems should be hydrophilic and present low molecular weight and high charge state. This will ensure that the buffer ions present have a large sphere of hydration, which hampers their ability to move through solution to the droplet surface. By ensuring that these criteria are met it is likely that the analyte species will be able to more efficiently migrate to the droplet surface and therefore analyte sensitivity will be optimised. Signal (ion) suppression is experienced when some additives are present in the eluent system. TFA (trifluoroacetic) is widely used in reverse-phase liquid chromatography as an ion pair reagent with basic compounds. The resulting ion pair appears neutral to the stationary phase and the chromatographic peak elutes with satisfactory retention. i Migration to the droplet surface Sphere of hydration comparison (left buffer ion, right analyte ion) The mechanism of ion suppression with TFA involves the [CF 3 C] - anion pairing with the [M+H] + cation in solution. Even if the ion-paired analyte molecule is ejected into the gas phase, it will be uncharged and therefore will not give rise to a signal in the mass spectrometer effectively quenching or suppressing the analyte signal. Crawford Scientific www.chromacademy.com 24

Ion pair formation ESI Ion Pair Reagents verview Ion-pair reagents may be necessary for the separation of ionic sample components. The charged sample ion forms a complex with the oppositely charged ion-pair reagent resulting in greater retention and separation. The challenge for the analyst is to find an ion-pair system that does not interfere with the ion formation process but still yields acceptable chromatographic behaviour. Unfortunately, the most common ion-pair reagents are non-volatile and include alkyl sulfonates and tetra-alkyl ammonium salts. API LC-MS requires that the analyte is ionic in solution (ESI) or that the analyte accepts or donates a proton in the gas phase, (APCI). Ion-pairing reagents can interfere with the ionisation process due to charge neutralisation in ESI-MS and by agglomerating analyte and altering proton affinities in APCI-MS. The ESI signal for some biochemical active compounds is presented below. The signal for norepinefrine improve by the presence of the volatile ion-pairing reagents (VA and PFHA). Ionisation of erythromycin also improves when PFHA is present. HSA causes significant suppression for all analytes because the non-volatile ion-pair hinders the escape of the analyte from the droplets during electros pray ionisation. Crawford Scientific www.chromacademy.com 25

i Ion pair reagent effect in ESI (refer to tables 6 and 7) Table 6. Analytes of interest. Compound ame/property Structure Molecular Formula Molecular weight (Da) A Erythromycin (EYR): Low volatility C 37 H 67 13 733.94 H H H H B C Mitowazone (MGBG): Moderate volatility orepinephrine (R): High volatility H 2 H H H H H H C 5 H 12 8 184 H 2 H 2 C 8 H 11 3 169 H D Tyramine (TYR): High volatility H 2 H H C 8 H 9 137.18 AmAc Ammonium acetate CH 3 CH 2 C 2 H 5 2 75 PFHA Perfluoro heptanoic CF 3 (CF 2 ) 5 CH C 7 F 13 2 H 364 acid HSA Heptane sulphonic CH 3 (CH 2 ) 6 S 3 H C 7 H 16 3 S 180 acid VA Valeric acid CH 3 (CH 2 ) 3 CH C 5 H 10 2 102 Crawford Scientific www.chromacademy.com 26

Table 7. Ion pairing reagents. Label Compound Structure Molecular Formula Molecular weight (Da) AmAc Ammonium acetate CH 3 CH 2 C 2 H 5 2 75 PFHA Perfluoro heptanoic acid CF 3 (CF 2 ) 5 CH C 7 F 13 2 H 364 HSA Heptane sulphonic acid CH 3 (CH 2 ) 6 S 3 H C 7 H 16 3 S 180 VA Valeric acid CH 3 (CH 2 ) 3 CH C 5 H 10 2 102 Weak and strong volatile ion-pair reagents such as valeric acid (VA) and perfluoroheptanoic acid (PFHA) can replace nonvolatile strong ion-pair reagents such as sodium heptane sulfonic acid (HSA). A study of four pharmaceutically active compounds highlights the impact of the ion-pair reagent on MS response and separation as a percent of a control (ammonium acetate). [9] The ESI signal for norepinephrine improves by the presence of the volatile ion-pair reagents (VA and PFHA). Ionisation of erythromycin also improves when PFHA is present. HSA causes significant suppression for all analytes because the nonvolatile ionpair hinders the escape of analyte from the droplets during electrospray ionisation. Positively charged nonvolatile ion-pair reagents such as tetrabutylammonium hydroxide can be replaced with tributylamine i Ion pair reagent effect in ESI (refer to tables 6 and 7) ESI Ion Pair Reagents Considerations Weak and strong volatile ion-pair reagents such as valeric acid (VA) and perfluoroheptanoic acid (PFHA) can replace nonvolatile strong ion-pair reagents such as sodium heptane sulfonic acid (HSA). A study of four pharmaceutically active compounds highlights the impact of the ion-pair reagent on MS response and separation as a percent of a control (ammonium acetate). [14] The ESI signal for norepinephrine improves in the presence of the volatile ion-pair reagents (VA and PFHA). Ionisation of erythromycin also improves when PFHA is present. HSA causes significant suppression for all analytes because the nonvolatile ionpair hinders the escape of analyte from the droplets during electrospray ionisation. Crawford Scientific www.chromacademy.com 27

Positively charged nonvolatile ion-pair reagents such as tetrabutylammonium hydroxide can be replaced with tributylamine i Ion pairing reagent effect on signal response (please refer to tables 6 and 7 for more information). Cationisation vs. Anionisation in ESI eutral molecules that have any propensity for hydrogen bonding will form adduct ions with ammonium or alkali metal ions The detection of neutral compounds can be enhanced in electrospray mode by the addition of cationisation/anionisation reagents. Sodium acetate or ammonium acetate added in 20-50 mm amounts can significantly increase the signal intensity. ote that additional heptafluorobutyric acid can also increase the signal. Cationisation of certain molecules can also be achieved by adding certain organometallic salts, especially lithium, sodium, potassium, rubidium, and caesium; yielding species of the form [Analyte+Metal] +. [Analyte+2Metal] 2+, etc. Alcoholic solutions of the above salts are usually prepared at a concentration of approximately 1mg/mL and then mixed with the eluent system. [15] Crawford Scientific www.chromacademy.com 28

Cationisation in ESI Tetraethyl or tetrabutyl ammonium hydroxide had been used to increase signal response in ESI(-). In negative ion mode, signal intensity can be enhanced by the presence of certain counter ions (like sodium or ammonium) in the eluent system. The increased response is due to at least two factors: improved electrospray formation and increased ionization of the weak acid due to the higher ph. Crawford Scientific www.chromacademy.com 29

Anaionisation in ESI The ESI response of propionate is increased three-fold when the counterion is sodium, relative to the response when the counterion is mostly H +. If the counterion is H 4 the signal response is increased six-fold. Pre and Post Column Addition in ESI The increase in response of an analyte in ESI-MS can occur for a variety of reasons. Typically, this is due to an increase in the concentration of the analyte in the chromatographic peak, such as injecting increasing amounts of analyte; however, response can also be increased by enhancing the generation of gas phase analyte ions. This can be accomplished by manipulating various ion source parameters or through changes in mobile phase composition and flow rate. The chemical composition of the mobile phase plays an important role both in the chromatographic separation and in the ESI ionisation process. The use of an auxiliary liquid could help not only in the generation of ions (under ESI conditions) but it can also help to optimise the HPLC separation. Two main strategies have been developed: Crawford Scientific www.chromacademy.com 30

Pre-column addition is mainly used to improve the quality of the HPLC separation. Here, the auxiliary liquid is mixed with the eluent system before it enters the HPLC column. Post-column addition is mainly used to improve the ionisation process. Here, the auxiliary liquid is mixed with the eluent system after it leaves the HPLC column. Pre and post column addition allows the analyst to independently optimise separation and ionisation processes. Pre and post column addition Post Column Addition in ESI Post-column addition allows the analyst to optimise the electrospray ionisation conditions [10, 16] (ph, ionic strength, etc) without compromising the quality of the HPLC separation. Buffers can be used to adjust and keep the correct ph for the ionisation process to takes place. Certain organic solvents (like methanol) can be used to reduce ionic strength, which in turn facilitate the ion evaporation process. The addition of anionisation/cationisation reagents will increase signal response. Some widely used volatile reagents (such as trifluoroacetic acid TFA) form very strong ion pairs with the analyte (preventing ionisation). The addition of certain species (like weak volatile acids) can minimise these problem. Typically, the reagent stream flow rate is 10-50% of the column separation flow rate. There are some disadvantages inherent to post-column addition: it requires an extra pump and can broaden chromatographic peaks due to the dead volumes introduced; analyte concentration is also lowered, which may reduce the sensitivity of the analysis. Crawford Scientific www.chromacademy.com 31

With the advent of new HPLC stationary phases that can better retain ionic species, the use of post column addition is becoming less important. Post column addition TFA form very strong ion pairs with the analyte preventing ionisation: CF C M H] [ M H CF C ] 3 [ 3 Apffel et al proposed the post-column addition of a highly concentrated weak acid (like formic, acetic, propionic, butyric and valeric acids) of the form RCH which would compete with TFA (strong acid) to pair with the analyte. As the weak acid is present in a high concentration, the competition between TFA and the weak acid would be driven towards the deprotonation of the weak acid: [10] CF3 C RCH 3 CF CH RC Consequently, the protonated TFA (CF 3 CH) could be evaporated from the droplet and the analyte would form ion pairs with the new anion (RC ). As these ion pairs would be weaker than those previously formed with the TFA anion, the presence of the analyte in its protonated form would be promoted. RC [ M H] [ M H RC ] The ion pair analyte-tfa is replaced with another ion pair (formed between the analyte and the weak acid) which is much easier to dissociate when the analyte is liberated into the gas phase. Eprosartan is an angiotensin II receptor antagonist drug (group of pharmaceuticals which modulate the renin-angiotensin-aldosterone system) used for the treatment of high blood Crawford Scientific www.chromacademy.com 32

pressure. The diagram below shows the ESI analysis of an eprosartan extract from a whole blood sample the mobile phase contains TFA. ESI(+)-MS determination of eprosartan, eluent system: 200μL/min 40%(0.2%TFA in water) + 60%(0.2%TFA in acetonitrile), 200 μl/min. Column: Vydac, 250 2.1mm, 5μm The increase in signal intensity with the addition of increasing amounts of organic acid is noteable. The TFA is replaced as the ion pair with the analyte molecule by the weak organic acid this ion pair dissociates effectively as the ion pair is librated into the gas phase, giving rise to an analyte ion as opposed to the ion-pair. ESI Adduct Formation When a molecule is analysed by API techniques, typically, the predominant specie in the mass spectrum is the protonated molecular ion (pseudomolecular ion): [M+H] + acidic conditions (positive ion mode) [M-H] - basic conditions (negative ion mode) ther pseudomolecular ions may be produced and even predominate, depending upon [17, 18, 19, 20] the mobile phase additives, and impurities. The analytical process includes recognition of target masses (i.e. for analysis of combinatorial library samples), it is possible to reject compounds whose mass does not match the target due to adduct formation. Computer algorithms are often programmed to search the spectrum obtained for the target mass of interest and also for the common adducts to avoid this situation. Crawford Scientific www.chromacademy.com 33

Common sources of sodium and potassium within the interface are leachates from laboratory glassware. To avoid problems with adduct formation, de-sodiated and depotassiated glassware is available. Table 8. Common adducts observed in ESI positive ion mode bserved Explanation Mass [M+H] + Protonation. M + 1 [M+H+H 4 ] + Mainly seen when using CH 3 H 4 M + 18 [M+H+nH 2 ] + M + H 2 adduct M + 1 + 18 n [M+H+H 2 ] + M + H 2 adduct M + 19 [M+H+a] + M + a adduct M + 24 [M+H+K] + M + K adduct. M + 40 [M+H+CH 3 C] + Mainly when using acetonitrile M + 42 [2M+H] + Analyte dimerisation. 2M + 1 [M+H+CH 3 C] + In presence of CH 3 C M + 42 [M+H+CH 3 C+nH 2 ] + Adduct of CH 3 C and H 2 M + 42 + 18 n Table 9. Common adducts observed in ESI negative ion mode bserved Explanation Mass [M-H] - Deprotonation. M - 1 [M-H-nH 2 ] - Deprotonation and dehydration M 1-18n [M+Cl] - / (M+35.5) Ion attachment. M + 35.5 [M-2H+a] - M + a adduct M + 21 [M-H-C 2 ] - Carbon dioxide looses. M - 45 Crawford Scientific www.chromacademy.com 34

ESI Considerations Table 10. ESI Considerations ESI Consideration 1. Choose samples and matrices that are suitable for ESCI-MS analysis 2. Suitable solvents are polar with low surface tension, high conductivity and low heat of solvation. 3. Keep ph same as for LC-UV methods, switch to more volatile buffers and be aware that eluent ph may have to be adjusted for instrument response. 4. When choosing buffer salts, ensure that they are volatile and avoid species that are known to suppress the signal from analyte ions due to disruption of proton transfer reactions. Avoid buffers that are known to strongly ion pair (like TFA) 5. When transferring HPLC methods to MS detection 6. When choosing ion-pairing reagents, ensure that you can optimize signal response by using post-column addition 7. Eliminate buffer salts and ion-pairing reagents that are known to form adducts 8. If analyte retention is a problem choose a new HPLC column. Further Explanation Moderate to highly polar Up to 100000 Da Ionise in solution perhaps with multiple charge Typical examples include methanol, isopropanol, acetonitrile, etc Substitute phosphates, sulphates, borates, and citrates with ammonium acetate, formate or hydroxide, formic or acetic acid or trifluoroacetic acid (TFA) Volatile buffers are essential to give extended periods between source cleaning. Choose buffers with low molecular weight that are either polar or ionized and capable of holding multiple charges. In this way the analyte species will more easily occupy droplet surface sites giving high concentration of gas phase ions. Initially chose to keep the ionic strength of eluent the same, however, this may have to be reduced if significant signal suppression is observed Use volatile ion pairing reagents initially For example those containing a, K, etc. ote that you can choose HPLC columns that are able to separate ionic species. APCI Solvent Choice verview In general, most common LC solvents are compatible with APCI techniques. It is preferable to have neutral molecules eluting into the interface for APCI as the ionisation process is initiated by a discharge from the eluent vapour. Solvent choice for droplet formation and vaporisation is not as critical in APCI-MS than ESI-MS. Adjustment of nebulising gas flow and vaporisation temperatures overcome factors such as surface tension and volatility that are crucial in ESI-MS, however, higher sensitivity is expected when employing high volatility eluent components. [21] Vaporisation favours neutral species due to their increased volatility compared to charged species. Therefore, eluent ph is normally adjusted to give the neutral analyte species in solution exactly the opposite to the situation encountered with ESI-MS. This is favourable in terms of analyte retention, with neutral species being well retained on reverse phase HPLC stationary phase materials. Crawford Scientific www.chromacademy.com 35

i The order of events for ion formation in APCI is as follows: Eluent containing the neutral analyte is nebulised to form droplets Droplets are vaporised Electrons from a corona electrode ionise the gas-phase eluent Analyte molecules are ionised in a series of charge transfer reactions with the eluent ions APCI(+)-MS determination of phenitioin. Mobile phase 700μL/min of 60% acetonitrile + 40%(5mM ammonium buffer). Column Inertsil DS-3 (50 3mm, 5μm) Crawford Scientific www.chromacademy.com 36

APCI(+)-MS determination of indomethacin. Mobile phase 300μL/min of 60% acetonitrile + 40%(5mM acetic acid buffer). Column LiChrosper RP-18 (125 3mm, 5μm) APCI Solvent Choice Mechanisms ptimising proton transfer reactions in the solvent plasma around the APCI corona electrode is critical in APCI-MS. A free electron discharge from the corona electrode ionises the solvent molecules and the choice of solvent becomes critical when considering the eluent charging reactions, ion discharge and the production of analyte ions. The mode of ionisation depends on the solution chemistry and ion polarity (positive or negative). Common ionisation mechanisms are: 1. Proton abstraction (for acidic analytes): H - + H + 2. Electron capture (for halogenated and aromatic analytes): Br + e - Br- Crawford Scientific www.chromacademy.com 37

3. Charge exchange + + :H+ H 2 H 3 + 4. Proton or adduct addition + H + H 2 H 3 + In considering the sensitivity of the APCI technique for a particular analyte the relative proton affinities for the solvents and analyte species involved must be balanced. Buffers for APCI verview Several factors need to be considered when choosing buffer salts for APCI-MS analysis. In general in-volatile buffer species should be avoided. In-volatile buffers reduce the sensitivity of the interface through coating of the crucial interface surfaces in exactly the same way as described for ESI-MS, however, other considerations in APCI-MS when using in-volatile buffer species include: The effect of the buffer species on the volatility of the eluent The effect of the buffer on critical proton transfer reactions in the gas phase at the corona electrode The chemical nature of the eluent system is crucial for good APCI-MS results. A sample containing caffeine, lincomycin and sulfachloropyridazine was analysed under APCI-MS conditions. The separation was performed on a a ZRBAX ECLIPSE XDB C8 2.1 50 mm, 5μm (30 o C). The effect of using different eluent systems is described below. [22] Crawford Scientific www.chromacademy.com 38

i When sodium phosphate (50 mm) is added to the eluent system, higher molecular weight species are suppressed as the volatility of the droplet is much reduced and these species become entrained into the solid residues formed prom the phosphate species as the volatile eluent evaporates in the nebuliser tube. Buffers for APCI Concentration Whenever possible, concentrations of buffer salts in the HPLC eluent should be kept to a minimum to avoid signal suppression via disruption of the proton transfer processes, reduced droplet or analyte volatility. The results presented below, show the effect of buffer concentration (ammonium acetate 0 350 mm), on signals from caffeine and reserpine. [22] At lower concentrations the ammonium acetate acts to promote the proton transfer process for reserpine, however at increasing concentrations the volatile ammonium acetate makes it increasingly more difficult to volatilise reserpine from the eluent droplet. Caffeine is a weaker base than ammonia; therefore as the concentration of buffer in the eluent is increased in the gas phase it competes strongly for the proton ions in the gas plasma around the corona electrode and suppresses the caffeine signal. Crawford Scientific www.chromacademy.com 39

i Experimental conditions Mobile phase - ammonium acetate (0-350 mm) methanol Flow rate - 0.6 ml/min. Inj. Vol. - 1 ml of reserpine (84 ng) or caffeine (125 ng) SIM Ions - 195.2 and 609.3 Drying gas - 350 o C, 5L/min. ebuliser gas -60 psi Capillary - 4000V Vaporiser - 400 o C Table 11. Analytes of interest. Compound Structure High/low buffer concentration Caffeine C 8 H 10 2 (194.19 Da) Low Buffer Concentration [ M H] H M H 3 4 High Buffer Concentration H3 H 2 M [ M H] ( n 1) H 2 Reserpine C 33 H 40 2 9. (608.69 Da) H Low Buffer Concentration o H Ac M] [ H Ac] [ M ] 3 [ 2 H High Buffer Concentration [ M H] H M H 3 4 Crawford Scientific www.chromacademy.com 40

Ion pairing reagents in APCI When using ion-pairing reagents with LC-APCI-MS the possibility of volatility and proton transfer disruption also needs to be considered. Although it may be considered beneficial to add ion-pairing reagent to the eluent to ensure that the analyte is in a neutral form, the choice of ion-paring reagent can have profound effects on the sensitivity of the analysis, which is highly analyte specific. The results shown opposite indicate the differences obtained when using, weak and strong ion pairing reagents as well as volatile and nonvolatile reagents. [9] orepinephrine, which is the most volatile analyte, shows the most marked improvement in sensitivity. VA and PFHA assist in the ionization process by acting as strong gas phase acids and hence increase the ionization efficiency and signal obtained in positive mode APCI-MS. Erythromycin is the least volatile of the analytes and its signal is strongly suppressed by HSA, which traps the analyte in the dried droplets as it agglomerates on droplet desolvation. The ability to strongly ion pair is often detrimental to the APCI-MS signal as is seen in the case of MGBG and PFHA. The pair formed is so strongly associated that it is not disrupted and shows very low proton affinity. PFHA: CF 3 (CF 2 ) 5 CH Strong ion pair/volatile HSA: CH 3 (CH 2 ) 5 S 3 H Strong ion pair/non-volatile VA: CH 3 (CH 2 ) 3 CH Weak ion pair/volatile i Ion pairing reagent effect on APCI signal response (please refer to tables 6 and 7 for more information). Crawford Scientific www.chromacademy.com 41

APCI Considerations Table 12. APCI considerations APCI Consideration 1. Choose samples and matrices that are suitable for APCI-MS analysis 2. Volatile solvents are preferred for APCI analysis as volatility of the eluent is critical. 3. Keep ph same as for LC-UV methods, switch to more volatile buffers at concentrations under 100mM 4. When choosing buffer salts, ensure that they are volatile and avoid species that are known to suppress the signal from analyte ions due to disruption of proton transfer reactions 5. When transferring HPLC methods to MS detection 6. When choosing ion-pairing reagents, ensure that these species are volatile and present in low enough concentrations to ensure proton transfer reactions are not disrupted 7. Eliminate buffer salts and ion-pairing reagents that are known to form adducts Further Explanation Examples of analytes suitable for APCI include PAHs, PCBs, fatty acids, phthalates, and glycerides Samples that contain heteroatoms such as benzodiazepines and carbamates also work well When developing methods, eluent systems containing acetonitrile should be fully evaluated as charge transfer reactions are possible that may reduce the sensitivity of the instrument Substitute phosphates, sulphates, borates, and citrates with ammonium acetate, formate or hydroxide, formic or acetic acid or trifluoroacetic acid (TFA) Be aware of the relative proton affinities of the buffer and analyte species. Volatile buffers are essential to give extended periods between source cleaning and ensure the formation of gas phase analyte species around the formation of gas phase analyte species around the corona discharge electrode Initially chose to keep the ionic strength of eluent the same, however, this may have to be reduced if significant signal suppression is observed Replace non-volatile ion-pairing species with heptafluorobutyric acid (HFBA) or tetrabutylammonium hydroxide (TBAH) For example those containing a, K, etc. References 1. The Theory of HPLC and Instrumentation of HPLC from the HPLC Channel. 2. D. Temesi, and B. Law. LC-GC International. (175-180), March 1999. 3. Perry, R.H.; Green, D.W. (1997). Perry's Chemical Engineers' Handbook (7th Edition). McGraw-Hill. 4. CRC Handbook of Chemistry and Physics. 85 th edition 2004-2005. 5. Hulya Yilmaz. Turk. J. Phys. 26 (2002), 243-246. 6. Christian G. Huber, Andreas Premstaller, Gerald Kleindienst. Evaluation of volatile eluents and electrolytes for high-performance liquid chromatography electrospray ionization mass spectrometry and capillary electrophoresis electrospray ionization mass spectrometry of proteins II. Capillary electrophoresis Journal of Chromatography A, 849 (1999) 175 189 Crawford Scientific www.chromacademy.com 42