Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Similar documents
Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Sensitive Femtogram Determination of Aflatoxins B 1 , B 2 , G 1. and G 2. in Food Matrices using Triple Quadrupole LC/MS. Authors.

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

High-Throughput Protein Quantitation Using Multiple Reaction Monitoring

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Improved Automated Sample Preparation for the Analysis of 25-OH-Vitamin D3 by LC/MS/MS

Agilent ESI and APCI sources: for polar to non-polar compounds

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

Agilent 6400 Series Triple Quadrupole Users Workshop. MH QQQ Users workshop 2/21/2014 1

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

Taking Full Advantage of UHPLC with Agilent 6460A Triple Quadrupole MS. Outline

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

LC/MS/MS of Fungicides and Metabolites in Orange Juice with Agilent Bond Elut Plexa and Poroshell 120

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

SAMHSA-Compliant LC/MS/MS Analysis of Phencyclidine in Urine with Agilent Bond Elut Plexa PCX and Agilent Poroshell 120

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

LC/MS/MS of Fungicides and Metabolites in Apple Juice with Agilent Bond Elut Plexa and Poroshell 120

Ultrafast Analysis of Methadone and Metabolite EDDP in Urine by the Agilent RapidFire High-Throughput Mass Spectrometry System

Guide to Peptide Quantitation. Agilent clinical research

Application Note. Abstract. Authors. Environmental

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS)

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

materials analysis Solutions for Your Analytical Business Markets and Applications Programs

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses

+ butanal + 2-picolineborane. + acetic acid. + butanal + 2-picolineborane. + acetic acid

Determination of Chlorinated Acid Herbicides in Soil by LC/MS/MS Application Note

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM

Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Yohimbine Adulteration in Dietary Supplements and Botanicals

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

Accelerating the Metabolite Identification Process Using High Resolution Q-TOF Data and Mass-MetaSite Software

Detection of Trace Level Pharmaceuticals in Drinking Water by Online SPE Enrichment with the Agilent 1200 Infinity Series Online-SPE Solution

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer

Metabolomics Batch Data Analysis Workflow to Characterize Differential Metabolites in Bacteria

25-OH VITAMIN D 3 AND 25-OH VITAMIN D 2 IN PLASMA BY UV / VIS FAST CODE Z19710

Clinical Toxicology. Biomass Component Extraction: The uneaten cooked plant specimen was prepared for

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

Application Note. Author. Abstract. Food Safety. Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

PosterREPRINT COMPARISON OF PEAK PARKING VERSUS AUTOMATED FRACTION ANALYSIS OF A COMPLEX PROTEIN MIXTURE. Introduction

Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF. Application Note. Donghui Yi and Christine Miller Agilent Technologies

Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS

Accurate Mass Analysis of Hydraulic Fracturing Waters

Agilent 6400 Series Triple Quadrupole LC/MS/MS Users Session

Routine MS Detection for USP Chromatographic Methods

Supporting Information

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Application Note. Edgar Naegele. Abstract

The Agilent 1260 Infinity Analytical SFC System with Time-of-Flight Mass Spectrometric Detection

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Determination of pesticides in baby food by UHPLC/MS/MS using the Agilent 1290 Infinity LC system and the Agilent 6460 triple quadrupole LC/MS

Analysis of Cannabinoids and Amphetamines in Serum by RRLC/Triple Quadrupole Mass Spectrometry Using a Multimode Ion Source. Application.

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

MS n Analysis With Fast Polarity Switching in the Agilent 1100 Series LC/MSD Trap SL. Application Note. Christine Miller Agilent Technologies

Quantitation of Ethyl Glucuronide and Ethyl Sulfate in Urine using LC-MS/MS

LC/MS/MS Analysis of Pesticide Residues in Apples Using Agilent Chem Elut Cartridges

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

Accurate quantification of oleanolic acid and ursolic acid in traditional chinese medicine

Highly sensitive quantitative estimation of genotoxic impurities from API and drug formulation using LC/MS/MS

A Q-TOF Generated, Metabolomics- Specifi c LC/MS/MS Library Facilitates Identifi cation of Metabolites in Malaria Infected Erythrocytes

LC-MS/MS Analysis of Phytocannabinoids and their

AppNote 7/2014. Automating Liquid-Liquid Extractions using a Bench-top Workstation KEYWORDS ABSTRACT. Sample Preparation, Lab Automation, LC/MS/MS

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Application Note. Edgar Naegele. Abstract

Agilent G3212 GC-APCI Source

High-sensitivity, high-throughput quantitation of catecholamine metabolites in urine by LC/MS/MS for clinical research

LC walk-up system using the Agilent 1200 Series LC Method Development Solution and Agilent Easy Access software. Application Note

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Determination of urea in ultrapure water by IC-MS/MS

Rapid, Robust, and Sensitive Detection of 11-nor-D 9 - Tetrahydrocannabinol-9-Carboxylic Acid in Hair

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

770-9P. Sensitivity and Selectivity - A Case Study of LC/MS Enantioselective Resolution of Bupivacaine Using Vancomycin as a Chiral Stationary Phase

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Application Note LCMS-112 A Fully Automated Two-Step Procedure for Quality Control of Synthetic Peptides

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

25-OH VITAMIN D 3 IN PLASMA BY UV FAST CODE Z19610

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Transcription:

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System Application Note Authors Kevin McCann, Sameer Nene, Doug McIntyre, Edmond Neo, Dennis Nagtalon, and Dorothy Yang Agilent Technologies, Inc. Santa Clara, CA USA Abstract This application note demonstrates how the Agilent StreamSelect LC/MS system increases the throughput of traditional LC/MS analyses. By mirroring an LC method on two HPLC systems and coordinating the use of a single triple quadrupole mass spectrometer, up to twice the throughput can be achieved. Agilent MassHunter StreamSelect software requires minimal user input and is compatible with existing LC/MS methods. Based on the parameters provided, the software determines the most effi cient way of injecting and analyzing samples automatically. A previously developed analytical method for the analysis of 25-OH vitamin D2 and D3 by LC/MS/MS used a single high-performance liquid chromatography (HPLC) stream 1. The method was implemented on a StreamSelect LC/MS system to demonstrate an increase in LC/MS productivity while maintaining excellent robustness and reliability.

Introduction LC/MS/MS is ideally suited for the direct, rapid analysis of prepared biological samples. While analysis times can be shortened through appropriate LC method choices (gradient, flow rate, column packing, and so forth), a user is often only interested in a portion of the total data collected by a mass spectrometer. Typically, there is time during each chromatographic separation where no compounds of interest are being analyzed by the mass spectrometer, leaving the instrument under utilized for a large period of time. This application note describes the ability to increase mass spectrometer productivity through the use of the Agilent StreamSelect LC/MS system with an online sample cleanup option. The Agilent MassHunter StreamSelect control software is capable of orchestrating the timing of all HPLC components and coordinating the analytical usage of the mass spectrometer. In a previous application note 1, an analytical method was developed for the rapid, sensitive and accurate determination of 25-OH vitamin D2 and D3 in serum on the Agilent 6460 Triple Quadrupole LC/MS system equipped with Agilent JetStream (AJS) technology. This method was transferred to a StreamSelect LC/MS system, allowing for twice the sample throughput without any additional method development. Experimental Reagents and Standards Stock solutions of 25-OH D2, 25-OH D3, and their deuterated internal standards (IsoSciences) were prepared at 10 μg/ml in methanol (Honeywell) and stored at 20 C. Calibration standard solutions were prepared from these stock solutions at 1, 25, 75, 125, and 250 ng/ml in pooled serum. Deuterated 25-OH D2 and D3 internal standards were diluted to 1,000 ng/ml with methanol. Instrumentation The StreamSelect LC/MS system is completely integrated and comprised of a triple quadrupole mass spectrometer coupled to two HPLC systems, all controlled by a single software application. The HPLC systems used here were configured for online sample cleanup using one quaternary and one binary pump per system. The samples were loaded onto a trapping column where the analytes were retained and washed using the quaternary pump. The wash was sent to waste, reducing the amount of matrix introduced into the mass spectrometer. Shortly before the analytes eluted from the trapping column, a valve was switched and the binary pump was used to elute the analytes onto an analytical column where further chromatographic separation was performed. To set up the system, a previously collected standard data file was loaded into the MassHunter StreamSelect software. The data acquisition method was extracted from the data file and a window of interest was specified using the data file s chromatogram. Based on that information, the software automatically coordinated all timing related to running the HPLC system. Quaternary pump gradient Time (min) % B 0.00 50 1.00 90 2.30 90 2.40 98 3.90 98 3.91 50 Binary pump gradient Time (min) % B 0.00 85 3.20 85 3.21 98 3.80 98 3.81 85 Valve timing Time (min) Position 0.00 1 1.90 2 2.30 1 LC conditions Trapping column Agilent Eclipse Plus C18 Guard Column, 2.1 12.5 mm, 5 µm Analytical column Agilent Poroshell 120 EC-C18, 2.1 50 mm, 2.7 μm Column temperature 50 C Injection volume 10 µl Needle wash Flush port, 50:25:25 IPA:MeOH:H 2 O, 5 seconds Mobile phase A H 2 O + 0.1 % formic acid Mobile phase B MeOH + 0.1 % formic acid Flow rate 0.5 ml/min Stop time 5.0 minutes 2

MS/MS configuration and conditions Confi guration Agilent 6460 Triple Quadrupole Mass Spectrometer equipped with Agilent JetStream Ion mode Positive Drying gas temperature 275 C Drying gas flow 5 L/min Nebulizer pressure 45 psi Sheath gas temperature 325 C Sheath gas flow 11 L/min Capillary voltage 5,000 V EMV 200 V MRM transitions Compound Precursor Product Frag (V) Dwell (msec) CE (V) 25-OH Vitamin D3 401.3 383.2 106 50 4 25-OH Vitamin D3 401.3 365.3 106 50 4 25-OH Vitamin D2 413.3 395.3 106 50 4 25-OH Vitamin D2 413.3 355.2 106 50 4 25-OH Vitamin D3-d3 404.3 386.3 106 50 4 25-OH Vitamin D2-d3 416.4 398.3 106 50 4 Sample preparation 1. Combine 150 µl of sample with 15 µl of 1,000 ng/ml internal standard solution, and 150 µl of acetonitrile in an extraction tube. 2. Vortex for 30 seconds and let stand for 15 minutes at room temperature. 3. Add 750 µl of heptane and vortex for 30 seconds. 4. Centrifuge at 13,000 rpm for 5 minutes. 5. Transfer the organic layer (top) to a clean extraction tube. 6. Evaporate to dryness with nitrogen at room temperature. 7. Reconstitute in 200 µl of 75:25 methanol:0.1 % formic acid in water. 8. Transfer the samples to a 96-well plate for analysis. Results and Discussion In the previously developed analytical method, which used a single HPLC stream, it was observed that more than 50 % of the data collected by the mass spectrometer was not needed since the analytes of interest eluted during a narrow retention time window. The StreamSelect system mirrors components of a single stream system to provide a second stream, operating in parallel to the first. By loading the standard LC/MS method and specifying when the analytes of interest elute, the MassHunter StreamSelect software is able to determine the most efficient method of injecting and analyzing a batch of samples. By staggering injections on parallel streams and switching between the two streams at the appropriate time, the StreamSelect system can achieve up to twice the throughput of a standard LC/MS system (Figure 1). Stream 1 Stream 2 Staggered, overlapped Offset 0.0 2.5 5.0 7.5 10.0 12.5 15.0 17.5 20.0 22.5 25.0 Idle 0.0 2.5 5.0 7.5 10.0 12.5 15.0 17.5 20.0 22.5 25.0 Figure 1. The Agilent StreamSelect LC/MS System maximizes MS usage by running staggered analyses on two parallel LC systems and automatically coordinating MS analysis. 3

The two parallel LC systems displayed excellent agreement when comparing quantitative results (Figure 2) with an R 2 > 0.999 for both 25-OH vitamin D2 and D3. Deviations in retention time between the two streams were also minimal - well below 10 % RSD (Figure 3). Since both LC systems and the MS device were controlled by a single piece of integrated automation software, excellent reliability and robustness were also observed. If a system error occurs while a batch is running, the system will automatically reschedule samples to run on a single LC system without losing any samples. 1 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0 25-OH Vitamin D3 y = 0.0095*x + 0.014 R 2 = 0.9991 0 10 20 30 40 50 60 70 80 90 100 Concentration (ng/ml) 25-OH Vitamin D2 y = 0.0068*x + 0.0018 R 2 = 0.9992 0 10 20 30 40 50 60 70 80 90 100 Concentration (ng/ml) Figure 2. Combined calibration curves for 25-OH vitamin D2 and D3 across both LC systems; Stream 1 (green) and Stream 2 (blue). 4

Conclusions The Agilent StreamSelect LC/MS system was evaluated using a previously developed analytical method for the measurement of 25-OH vitamin D2 and D3 in serum. MassHunter StreamSelect software coordinates a completely integrated LC/MS system consisting of two parallel HPLC streams and a single triple quadrupole mass spectrometer. No special method development is required as the user selects a standard LC/MS method, defines when the analytes of interest elute, and the software automatically coordinates the analysis by setting all necessary timing. The use of the StreamSelect LC/MS system not only provides equivalent sensitivity, linearity, and reproducibility when compared to the previously developed method, but also increases productivity of a single triple quadrupole mass spectrometer. This application note illustrates the potential of the StreamSelect LCMS System as a solution for enhanced productivity in a wide range of LC/MS analyses. Counts 10 2 2 1.5 1 0.5 LC 1 Injection 001 RT = 2.583 LC 2 Injection 002 RT = 2.568 500 Continuous Injections LC 1 Injection 249 RT = 2.584 LC 2 Injection 250 RT = 2.573 Retention time RSD < 10 % LC 1 Injection 499 RT = 2.590 LC 2 Injection 500 RT = 2.570 Reference 1. R. Doyle, A. Szczesniewski, and K. McCann, Rapid Analysis of 25-OH Vitamin D in Serum Using an Agilent Triple Quadrupole LC/MS System with Automated Online Sample Cleanup, Agilent Publication 5991-2035EN (2013). 2 3 4 5 Acquisition time (min) Figure 3. Retention time reproducibility of 25-OH Vitamin D3 over 500 injections across both LC systems; Stream 1 (green) and Stream 2 (blue). 5

www.agilent.com/chem/streamselect For Research Use Only. Not for use in diagnostic procedures. This information is subject to change without notice. Agilent Technologies, Inc., 2014 Published in the USA, July 21, 2014 5991-2900EN