Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Similar documents
LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Ultrafast Analysis of Methadone and Metabolite EDDP in Urine by the Agilent RapidFire High-Throughput Mass Spectrometry System

Taking Full Advantage of UHPLC with Agilent 6460A Triple Quadrupole MS. Outline

Accelerating the Metabolite Identification Process Using High Resolution Q-TOF Data and Mass-MetaSite Software

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Sensitive Femtogram Determination of Aflatoxins B 1 , B 2 , G 1. and G 2. in Food Matrices using Triple Quadrupole LC/MS. Authors.

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Ultrafast Analysis of Metabolic Stability Assays Using Agilent RapidFire High-resolution MS

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

The Agilent 1260 Infinity Analytical SFC System with Time-of-Flight Mass Spectrometric Detection

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Developing a fast, generic method for rapid resolution LC with quadrupole MS detection. Application Note

Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Application Note. Edgar Naegele. Abstract

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

Agilent ESI and APCI sources: for polar to non-polar compounds

SAMHSA-Compliant LC/MS/MS Analysis of Phencyclidine in Urine with Agilent Bond Elut Plexa PCX and Agilent Poroshell 120

High-Throughput Protein Quantitation Using Multiple Reaction Monitoring

Pharmaceutical Impurity Identification and Profiling Using Agilent Q-TOF LC/MS Combined with Advanced MassHunter Data Processing Software

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

materials analysis Solutions for Your Analytical Business Markets and Applications Programs

Application Note. Abstract. Authors. Environmental

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS

Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Yohimbine Adulteration in Dietary Supplements and Botanicals

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM

Detection of Trace Level Pharmaceuticals in Drinking Water by Online SPE Enrichment with the Agilent 1200 Infinity Series Online-SPE Solution

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

Computer-assisted analysis of complex natural product extracts

Quantitation of Cystine and Identification of Related Metabolites in White Blood Cells Using a High Resolution Accurate Mass LC/MS Approach

Application Note. Edgar Naegele. Abstract

Application Note. David A. Weil Zoltan Timar Michael Zumwalt Edgar Naegele. Abstract

Measuring Drug-to-Antibody Ratio (DAR) for Antibody-Drug Conjugates (ADCs) with UHPLC/Q-TOF

Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system

Agilent 6400 Series Triple Quadrupole Users Workshop. MH QQQ Users workshop 2/21/2014 1

LC/MS/MS of Fungicides and Metabolites in Apple Juice with Agilent Bond Elut Plexa and Poroshell 120

Application Note. Author. Abstract. Food Safety. Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Determination of pesticides in baby food by UHPLC/MS/MS using the Agilent 1290 Infinity LC system and the Agilent 6460 triple quadrupole LC/MS

Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF. Application Note. Donghui Yi and Christine Miller Agilent Technologies

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.

LC/MS/MS of Fungicides and Metabolites in Orange Juice with Agilent Bond Elut Plexa and Poroshell 120

Separation of Large and Small Peptides by Supercritical Fluid Chromatography and Detection by Mass Spectrometry

Rapid, Reliable Metabolite Ratio Evaluation for MIST Assessments in Drug Discovery and Preclinical Studies

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

Metabolomics Batch Data Analysis Workflow to Characterize Differential Metabolites in Bacteria

Accurate Mass Analysis of Hydraulic Fracturing Waters

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Clinical Toxicology. Biomass Component Extraction: The uneaten cooked plant specimen was prepared for

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS)

Determination of Chlorinated Acid Herbicides in Soil by LC/MS/MS Application Note

An Evaluation of Various High-Resolution, Accurate-Mass Scan Modes for In Vitro Drug Discovery Screening

Supporting Information

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Analysis of Cannabinoids and Amphetamines in Serum by RRLC/Triple Quadrupole Mass Spectrometry Using a Multimode Ion Source. Application.

Confirmation of In Vitro Nefazodone Metabolites using the Superior Fragmentation of the QTRAP 5500 LC/MS/MS System

Conc n of A OAs present (um) Conc n of B OAs present (um)

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

A Q-TOF Generated, Metabolomics- Specifi c LC/MS/MS Library Facilitates Identifi cation of Metabolites in Malaria Infected Erythrocytes

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Analysis of Polyphenols in Saffron Petal Extracts with UHPLC/UV/MS

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

The Agilent 6495 Triple Quadrupole LC/MS: Peptide Quantitation Performance

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

Ultrasensitive EPA Method 1694 with the Agilent 6460 LC/MS/MS with Jet Stream Technology for Pharmaceuticals and Personal Care Products in Water

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

LC walk-up system using the Agilent 1200 Series LC Method Development Solution and Agilent Easy Access software. Application Note

Liquid Chromatograph Mass Spectrometer LCMS-8040 C146-E188A

Screening of Pharmaceuticals and Personal Care Products in Effluents and Surface Waters with LCMS Q-TOF

Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Mixed-Mode, Weak Anion-Exchange, Solid-Phase Extraction Method for the Extraction of Niflumic Acid from Human Plasma

Transcription:

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS Application Note Drug Discovery Authors Srividya Kailasam Agilent Technologies Bangalore India Kerry Frost Cyprotex Discovery Limited % Analyte remaining 14 12 1 8 6 4 2 Eucatropine 2 4 6 8 1 12 14 Time (min) Abstract This Application Note demonstrates the use of rapid resolution liquid chromatography in conjunction with tandem mass spectrometry to study the plasma stability of pharmaceutical compounds belonging to four different drug classes. The Agilent 12 Series Rapid Resolution LC system (RRLC) provided fast separation of compounds and internal standard in all samples. The Agilent MassHunter Optimizer software identified compound-specific MS parameters such as the most abundant product ions, optimal fragmentor voltages and collision energies. Taken together, robust methods for a larger set of compounds could be developed within a short period of time.

Introduction Compounds with poor stability characteristics tend to have low bioavailability. Certain functional groups such as amides, esters or lactones make compounds favorable targets for plasma enzymes such as hydrolases and esterases. Knowledge of plasma stability profiles of New Chemical Entities (NCEs) helps in modifying the molecular structure to improve physicochemical properties and in selecting more suitable molecules for subsequent development. In this study rapid resolution LC/MS- MS was used to study the plasma stability of ten pharmaceutical compounds. Four beta blockers (atenolol, labetalol, metoprolol and propranolol), one calcium channel blocker (verapamil), one serotonin receptor agonist (buspirone) and four antidepressants (nefazodone, nortriptyline, imipramine and trimipramine) were analyzed using reserpine as the internal standard. Eucatropine, an ophthalmic drug, was used as the control and analyzed using metoprolol as the internal standard. Experimental Chemicals Atenolol, labetalol, metoprolol, propranolol, buspirone, nefazodone, nortriptyline, imipramine, trimiparmine, verapamil, eucatropine and reserpine were purchased from Sigma-Aldrich, Bangalore, India. HPLC-grade acetonitrile and methanol were purchased from Lab-Scan. Formic acid was purchased from Fluka. Stock solutions and suitable dilutions of target analytes were prepared in methanol for method development. Sample preparation Plasma was adjusted to ph 7.4 with either.1 M hydrochloric acid or.1m sodium hydroxide. Test compounds were dissolved in DMSO to a final concentration of 1 mm and then diluted to 4 µm in DMSO. Incubations were carried at a test compound concentration of 1 µm with a final DMSO concentration of 2.5%. 195 µl of plasma was added to the wells of a 96- well plate before spiking them with the 5 µl of test compound. The spiked plasma samples were incubated at 37 C for 2 hrs. Reactions were terminated at, 15, 3, 6 and 12 min by adding 4 µl of methanol containing reserpine as an internal standard. Simultaneously, plasma samples containing eucatropine (control compound) were terminated by adding 4 µl of methanol containing metoprolol as an internal standard. In addition, matrix blanks were prepared by adding methanol solutions containing internal standards to plasma samples without any of the analytes or control com- pounds. The sample plate was centrifuged at 25 rpm for 45 min at 4 C and the supernatant was transferred to a fresh 96-well plate for analysis by LC/MS-MS. Instrumentation We used an Agilent 12 Series Rapid Resolution LC system which included a micro vacuum degasser, binary pump, high-performance autosampler with thermostat and a thermostatted column compartment coupled to an Agilent 646 Triple Quadrupole Mass Spectrometer. The mass spectrometer operated in ESI mode with Agilent's Jet Stream Technology. The Agilent MassHunter Workstation (version B.2.) was used for data acquisition and MassHunter Quantitative analysis software (version B.1.4) was used for data analysis. We employed the Agilent MassHunter Optimizer software (B.2.) for optimization of two important MS parameters: the fragmentor voltage and collision energies. Optimizer also provided the most abundant MRM transitions used in this study. 2

Method details LC conditions used for all analytes and control are presented in Table 1. Parameter Mobile phase Flow rate Detail A: Water (.1% Formic acid) B: Acetonitrile (.1% Formic acid).5 ml/min Gradient conditions min 2 % B.2 min 5 % B.3 -.5 min 7 % B.6-1.2 min 95 % B 1.25-1.5 min 2 % B Run time 1.5 min Post time 1 min Injection volume 1 µl Needle wash 1 sec wash port with 1:1 methanol:water mix Column Agilent Zorbax Eclipse XDB - C18, 2.1 5 mm, 1.8 µm particle size (p/n 9277-92) Column temperature 5 C Table 1 LC Conditions. Reserpine was used as the internal standard for the analytes under study and metoprolol was used as the internal standard for eucatropine which was the control for the study. A generic set of MS conditions as shown in Table 2 were applied for ionization of all compounds including the control and the two internal standards. Parameter Detail Ionization mode ESI-positive with Agilent Jet Stream Technology Drying gas temperature 35 C Drying gas flow rate 1 L/min Nebulization gas pressure 2 Psi Capillary voltage 2 V Sheath gas temperature 4 C Sheath gas flow rate 11 L/min Nozzle voltage V Q1 resolution Unit Q2 resolution Unit Table 2 MS conditions. 3

The MassHunter Optimizer identified compound-specific MS parameters such as optimal fragmentor voltages, most abundant product ions for the selected precursor ion, and associated collision energies. The MRM transitions used in the study are presented in Table 3. A 5-ms dwell time used for all transitions provided more than 5 data points across all peaks. Compound Empirical Precursor Ion Product Ion(s) Fragmentor Collision energy (ev) Formula Quantifier, Qualifier voltage (V) Quantifier, Qualifier Atenolol C 14 H 22 N 2 O 3 267.2 145, 19 115 25, 13 Labetolol C 19 H 24 N 2 O 3 329.2 311.1, 294.1 15 5, 13 Metoprolol C 15 H 25 NO 3 268.2 116.1, 74.1 115 13,17 Propranolol C 16 H 21 NO 2 26.2 116.1,183 11 13,13 Buspirone C 21 H 31 N 5 O 2 386.3 122, 15 164 29,29 Verapamil C 27 H 38 N 2 O 4 455.3 165, 33.1 158 25, 21 Nefazodone C 25 H 32 ClN 5 O 2 47.2 274.1, 246.1 15 25,33 Nortriptyline C 19 H 21 N 264.2 233.1, 9.9 8 1, 2 Imipramine C 19 H 24 N 2 281.2 86.1, 58.1 11 13, 39 Trimipramine C 2 H 26 N 2 295.2 1.1, 58.1 15 9,41 Eucatropine C 17 H 25 NO 3 292.2 58.1 12 21 Reserpine C 33 H 4 N 2 O 9 69.38 195.1 15 41 Table 3 MS/MS parameters. Data analyses Five replicate injections were performed for each sample. The mean values of relative responses (analyte concentration/internal standard concentration) at different time points were calculated. The mean relative response at each time point was normalized with respect to mean relative response at min (T ). The normalized values (as percentages) were plotted against exposure time to obtain a plasma stability profile. Results and discussion This screening assay identifies any change in plasma concentration relative to the initial concentration at min. Therefore, it is not necessary to determine the exact concentration in the samples. The percentage of parent remaining at each time point relative to the -min sample is calculated. A compound is considered stable in plasma if the percentages at each time point are within ± 15% of the -min sample. 4

Figure 1A shows two overlaid peaks corresponding to the quantifier and qualifier transitions of one of the target compounds, nortriptyline, along with a peak representing the transition for the internal standard, reserpine. Figure 2A shows MRM transition peaks of eucatropine and its internal standard, metoprolol. A short run time of 1.5 min enables fast analyses of multiple samples. The selectivity of the technique can be seen in Figures 1B and 2B which are the MRM chromatograms of sample and control blanks respectively. The most abundant peak(s) in the two figures are of the two internal standards, reserpine and metoprolol (quantifier and qualifier) respectively. No significant peaks are seen for control or target compounds. Counts 1 3 + MRM (264.21 -> 9.939) Nortriptylinemin-r1.d 6 5 4 3 2 1 A.1.5 1 1.5 Acquisition Time (min) Counts 1 3 + MRM (268.21 -> 116.9961) Plasma_Bl_3-r4.d 2 1.5 1.5 B.1.5 1 1.5 Acquisition Time (min) Figure 1A MRM chromatogram of a study sample showing nortriptyline and IS (reserpine) peaks Figure 1B Plasma blank of samples containing only (IS) reserpine. 1 4 + MRM (268.21 -> 74.9961) Eucatropinemin-r2.d 1 4 + MRM (268.21 -> 116.9961) Plasma_Bl_7-r4.d 2 1 Counts 1 A Counts.5 B.1.5 5 1 1.5 Acquisition Time (min).1.5 1 1.5 Acquisition Time (min) Figure 2A MRM chromatogram of control showing eucatropine and (IS) metoprolol peak. Figure 2B Plasma blank of control containing only (IS) metoprolol. 5

Figure 3 depicts the plasma stability of the control compound eucatropine along with plasma stabilities of two of the target compounds. The developed method was able to successfully track the known stability characteristic of eucatropine [1]. Nortriptyline undergoes degradation in plasma, while metoprolol was found to be stable for up to two hours in plasma. Fluctuations in plasma concentrations after different contact times, despite excellent injection precision (CV <15%), was similar to what has been reported [2]. Deviation in concentration within ± 2% is acceptable for a screening assay. The plasma stabilities for all the target analytes and control compounds are shown in Table 4. Compound Atenolol Labetolol Metoprolol Propranolol Buspirone Verapamil Nefazodone Nortriptyline Imipramine Trimipramine Eucatropine Table 4 Plasma stability characteristics. Plasma Stability Unstable Unstable Unstable 14 14 12 Eucatropine 12 Nortriptyline % Analyte remaining 1 8 6 4 % Analyte remaining 1 8 6 4 2 2 2 4 6 8 1 12 14 Time (min) 2 4 6 8 1 12 14 Time (min) 14 12 Metoprolol % Analyte remaining 1 8 6 4 2 2 4 6 8 1 12 14 Time (min) Figure 3 Stability curves for three analytes. 6

Conclusion We developed an LC method along with a set of mass spectrometric conditions to assess plasma stability of ten pharmaceutical compounds. The MassHunter Optimizer software identified the most intense product ions for the selected precursor ion. The software also provided the most suitable fragmentor voltages and collision energies. Therefore, the mass spectrometric method could be developed quickly for a standard compound. We could show the suitability of the proposed method for estimating plasma stability of compounds belonging to different classes. References 1. Di, L., Kerns, E. H., Hong, Y. and Chen, H., "Development and application of high throughput plasma stability assay for drug development", International Journal of Pharamaceutics, 297 (25) 11-119. 2. http://www.cyprotex.com/assay_repor ts/cloe_screen_plasma_stability.pdf, Cyprotex, Cloe Screen Plasma Stability flyer 7

www.agilent.com/chem/qqq Agilent Technologies, Inc., 29 October 1, 29 Publication Number 599-463EN