IRS-TR 05001: Detector Substrate Voltage and Focal Plane Temperature of the LH module

Similar documents
IRS-TR 04001: Photoresponse and Read Noise Trends

Observing with the Infrared Spectrograph

The IRS Flats. Spitzer Science Center

Detectors for IR astronomy

Reduction procedure of long-slit optical spectra. Astrophysical observatory of Asiago

NICMOS Status and Plans

THE INFRARED SPECTROGRAPH ON THE SPITZER SPACE TELESCOPE

SPITZER SPACE TELESCOPE

Spitzer Space Telescope

Mitigation of H2RG persistence with image illumination Brian A. McLeod* a, Roger Smith b

arxiv:astro-ph/ v1 7 Jun 2004

COS Cycle 17: 20 programs 149 external orbits 446 internals FUV Detector Sensitivity Monitor 36 0 Oct. 3, 2010

WFC3 IR Blobs, IR Sky Flats and the measured IR background levels

Grand Canyon 8-m Telescope 1929

estec NIRSpec Technical Note NTN Spectral intensity of the MSA glow sources Abstract:

Announcement of Opportunity AKARI (ASTRO-F)

Normal and unusual transient events in IRAC images

NAOYUKI TAMURA Subaru Instrument Astronomer Subaru Telescope, NAOJ

A Fast Algorithm for Cosmic Rays Removal from Single Images

C. Watson, E. Churchwell, R. Indebetouw, M. Meade, B. Babler, B. Whitney

arxiv:astro-ph/ v1 12 Nov 2003

Lab 4: Stellar Spectroscopy

Large Area Imaging Survey of Near-Infrared Sky with Korean Compact Space Telescopes

1 Lecture, 2 September 1999

Spitzer Space Telescope Calibration Strategy: The Use of Asteroids

Ivan Valtchanov Herschel Science Centre European Space Astronomy Centre (ESAC) ESA. ESAC,20-21 Sep 2007 Ivan Valtchanov, Herschel Science Centre

FMOS. A Wide-field Multi-Object Infra-red Spectrograph for the Subaru Telescope. David Bonfield, Gavin Dalton

The VLT dealing with the Atmosphere, a Night Operation point of view

Near-IR Background Fluctuation Results from the Cosmic Infrared Background Experiment


Landsat-8 Operational Land Imager (OLI) Initial On-Orbit Performance

Scattered Light from the Earth Limb Measured with the STIS CCD

FORCAST: Science Capabili2es and Data Products. William D. Vacca

Lecture 8. October 25, 2017 Lab 5

PACS Wavelength Switching AOT release note

Scientific Capability of the James Webb Space Telescope and the Mid-InfraRed Instrument

Overview of Hyperion On-Orbit Instrument Performance, Stability, and Artifacts

Comet Measurement Techniques. Karen Meech Institute for Astronomy Session 27 1/18/05

Calibration Goals and Plans

WFC3 TV2 Testing: IR Channel Read Noise

LCO Global Telescope Network: Operations and policies for a time-domain facility. Todd Boroson

Common questions when planning observations with DKIST Jan 30, 2018

PHOTOMETRIC CALIBRATIONS FOR THE SIRTF INFRARED SPECTROGRAPH

Detection of Artificial Satellites in Images Acquired in Track Rate Mode.

Study of Physical Characteristics of High Apogee Space Debris

The Stratospheric Observatory for Infrared Astronomy (SOFIA)

WFC3 TV2 Testing: UVIS-2 Dark Frames and Rates

Data Reduction - Optical / NIR Imaging. Chian-Chou Chen Ph319

Measuring Radial Velocities of Low Mass Eclipsing Binaries

Introduction of near-infrared (NIR) spectroscopy. Ken-ichi Tadaki (NAOJ)

1 The Preliminary Processing

Data Processing in DES

624 SNPP VIIRS Solar Diffuser BRDF Degradation Trend Changes in Early Evan Haas* and Frank De Luccia The Aerospace Corporation, El Segundo CA

James Webb Space Telescope Cycle 1 Call for Proposals

THE OBSERVATION AND ANALYSIS OF STELLAR PHOTOSPHERES

Performance of the NICMOS ETC Against Archived Data

Fringe Correction for STIS Near-IR Long-Slit Spectra using Contemporaneous Tungsten Flat Fields

Overview of Measured SSI and Its Variability

ABSTRACT 1. INTRODUCTION 2. CALIBRATION CHANGES

WISE Science Data System Single Frame Position Reconstruction Peer Review: Introduction and Overview

The Cornell Atlas of Spitzer Spectra (CASSIS) and recent advances in the extraction of complex sources

WFPC2 Dark Current vs. Time

Space-Based Visible End of Life Experiments

Extraction of Point Source Spectra from STIS Long Slit Data

Early-Science call for observing time with SAM-FP

Announcement of Opportunity for Open Time. SPIRE PACS Parallel Mode Observers' Manual

An Algorithm for Correcting CTE Loss in Spectrophotometry of Point Sources with the STIS CCD

Lecture Outlines. Chapter 5. Astronomy Today 8th Edition Chaisson/McMillan Pearson Education, Inc.

First observations of the second solar spectrum with spatial resolution at the Lunette Jean Rösch

Exploring Data. Keck LRIS spectra. Handbook of CCD Astronomy by Steve Howell Chap. 4, parts of 6

WISE Science Data System Frame Co-addition Peer Review: Introduction and Overview

COS FUV Focus Determination for the G140L Grating

Jean-Baptiste Camille COROT ( ) Woman with a pearl (1869) Interrupted reading (1870)

Spitzer Space Telescope (Space InfraRed Telescope Facility)

The Challenge of AZ Cas-Part 1. John Menke Barnesville, MD Abstract

Pan-Planets. A Search for Transiting Planets Around Cool stars. J. Koppenhoefer, Th. Henning and the Pan-PlanetS Team

Future X-rayX Spectroscopy Missions. Jan-Willem den Herder

Studies of diffuse UV radiation

Impacts of focus on aspects of STIS UV Spectroscopy

Selection of stars to calibrate Gaia

Mid-Infrared Astronomy with IRAIT at Dome C: performances and simulations

Calibration of Ocean Colour Sensors

TIPS-JIM Meeting 13 October 2005, 10am, Auditorium

Chapter 6 Telescopes: Portals of Discovery. Agenda. How does your eye form an image? Refraction. Example: Refraction at Sunset

The Kepler Exoplanet Survey: Instrumentation, Performance and Results

Thermal And Near infrared Sensor for carbon Observation (TANSO) onboard the Greenhouse gases Observing SATellite (GOSAT) Research Announcement

Detecting Cosmic Rays in Infrared Data

Foundations of Astronomy 13e Seeds. Chapter 6. Light and Telescopes

Photometric Studies of GEO Debris

Atmospheric Extinction

AST 101 Intro to Astronomy: Stars & Galaxies

EXPOSURE TIME ESTIMATION

Keck/Subaru Exchange Program Subaru Users Meeting January 20, 2011

The distribution of electron energy is given by the Fermi-Dirac distribution.

Absolute Flux Calibration for STIS First-Order, Low-Resolution Modes

Astronomical Techniques I

MID INFRARED ASTRONOMY TECHNIQUES, AND DATA

New On-Orbit Sensitivity Calibration for All STIS Echelle Modes

How to calibrate interferometric data

SMOV Absolute Flux Calibration of the COS FUV Modes

Transcription:

Infrared Spectrograph Technical Report Series IRS-TR 05001: Detector Substrate Voltage and Focal Plane Temperature of the LH module Daniel Devost (1), Jeff van Cleve (2), G. C. Sloan (1), Lee Armus (3), Jim Ingalls (3), Phil Appleton (3) 17 May, 2005 Abstract We present the results for campaigns IRS-N2, IRS-12 and IRS-18 of the detector properties analysis procedure for the Long-High module. From campaign IRS-N2 to campaign IRS-18, the figure of merit of the current voltage and temperature settings has dropped from 8% below the maximum figure of merit to 58% below the maximum. Also, the number of bad pixels in the LH module has quadrupled. The probability of having a bad pixel in any 6-pixel window is now greater than 50%. We recommend to change the detector substrate voltage from 3.0 V to 3.4 V and to leave the focal plane temperature to 4.4 K. This change will lower the bad pixel count but a loss in sensitivity per pixel especially at λ > 35 µm will also occur. A minimum of 10 hours of telescope time and 5-7 weeks of analysis are required to bring the system to the same level of calibration that we have now. (1) Infrared Spectrograph Science Center, Cornell University; (2) Ball Aerospace, Boulder, CO; (3) Spitzer Science Center, Caltech

1 Introduction After the initial suite of diagnostics performed on the Infrared Spectrograph (IRS) on the Spitzer Space Telescope done during In-Orbit Checkout (IOC), the operating parameters for all four modules were changed prior to campaign IRS-K2. These changes were done based on a Figure Of Merit (FOM) calculated for each module using data that were gathered during Campaigns IRS-E, IRS-F, and IRS- H. The FOM was calculated by taking into consideration the sensitivity, noise, and number of bad pixels at a number of predefined Voltages and Temperature Settings (VTS). As a result of this analysis, the detector substrate voltage (V det ) and the focal plane temperature were adjusted to get optimal performance from the detectors. Table 1 lists those values for each module. Throughout this document, we will use V det as the reference to the voltage measurements. This quantity is related to the bias voltage (V bias ) by: V det + V bias = 5.0 Volts (1) During Campaign IRS-N of IOC, the Sun produced one of the most powerful solar flares ever measured. It was then decided to create a special campaign, IRS- N2, to probe the extent of the damage from this event. The same FOM analysis done earlier during IOC was performed once again on all the modules. One of the main conclusions of the analysis was that the sensitivity of the detectors had dropped by about 20% and that the number of bad pixels had increased significantly on all modules. At this time, though, the total merit of the operating VTS was only 8% below the maximum merit (see Table 4) 1. Even with the increased number of bad pixels, the improvement introduced by changing the operating voltages and temperatures was deemed insignificant compared to the complications it would produce in the calibration pipeline for the IRS. As a result, it was decided to continue operating all the modules at the voltages and temperatures set before Campaign IRS-K2 (Table 1). The data necessary to perform the FOM analysis were obtained once again during Campaign IRS-12 as part of the mandatory periodic detector health assessment. It was found that the Long-High (LH) detector had degraded. The probability of having a bad pixel in any 6-pixel window was now around 40% compared to 25% during Campaign IRS-N2. We also found that the LH VTS 1 The Tables giving the figure-of-merit results for the different VTS values list the settings in order of figure of merit. 2

TABLE 1 VOLTAGE AND TEMPERATURE SETTINGS, 2003 NOVEMBER Module V det (V) T (K) Short-Low 3.0 6.2 Short-High 3.0 6.2 Long-Low 3.2 4.4 Long-High 3.0 4.4 merit was 32% below the maximum merit function (see Table 3). After discussions on these numbers and their impact on the instrument performance and on the data reduction process, the instrument team decided to leave the bias voltage and focal plane temperature of the LH array unchanged. Another powerful Solar storm hit Spitzer just after Campaign IRS-17. It was decided to perform the FOM analysis again during the course of Campaign IRS- 18 to see how the detectors were affected by this storm. This paper presents the analysis of the LH dataset and compares it with the datasets from Campaigns IRS-N2 and IRS-12. The calculation of the FOM has been done using the same analysis plan and software used during Campaigns IRS-E, IRS-F, IRS-H, IRS-N2 and IRS-12. 2 Analysis The calculation of the FOM for each module is done on datasets collected with the detector responsivity Astronomical Observing Request (AOR). The same AORs were run during Campaigns IRS-N2, IRS-12 and IRS-18. The detector responsivity AOR gathers data that we use to examine the variation of detector noise as a function of detector voltage, focal plane temperature, and ramp time. There are 3 values of V sub (+ one repeat of the default voltage, which is discarded), 3 temperatures, and 4 ramp times, for a total of 40 exposures. The detector responsivity AOR collects data to examine the variation of the photoresponse (ηg), the gain dispersion product (βg) and the detector noise as a function of voltage, temperature and ramp time. There are 5 STIM settings, 3 3

voltages, and 3 temperature for a total of 45 exposures. All the data is acquired in the Sample Up the Ramp (SUR) acquisition mode and the Sample Reset Sample (SRS) acquisition mode. The data collected during the performance and responsivity AORs are used to derive the number of bad pixels, the Bright source FOM (BFOM), the Faint source FOM (FFOM) and the Total FOM (TFOM) for each combination of bias voltages and focal plane temperatures. The BFOM is given by: while the FFOM is: The TFOM is given by: BFOM = n 6 ηg good βg (2) FFOM = n 6 ηg good. (3) N d TFOM = BFOM FFOM (4) In these equations, n good is the fraction of pixels which are good and N d is the dark noise for the second longest ramp time permitted by science Astronomical Observing Templates (AOT). The calculations of pixel badness criteria are outlined in detail in the data analysis procedure for these activities, which may be downloaded from the IRS IOC Web site http://isc.astro.cornell.edu/tech/ioc/. The 6th power of the fraction of good pixels gives the probability that there are no bad pixels in an extracted point source comprised of 6 pixels. 3 Results 3.1 FOM analysis Tables 2, 3, and 4 give summaries of the LH results obtained from the analysis of the data from Campaigns IRS-18, IRS-12, and IRS-N2, respectively. Figure 1 shows how the number of bad pixels increases with detector voltage. These images are darks taken during the IRS long-wavelength detector responsivity AOR. 3.2 Sensitivity per wavelength element A change to the voltage has repercussions on the number of bad pixels and the sensitivity. The FOM analysis results presented here incorporate the detector photoresponse (ηg), but not the sensitivity as a function of wavelength. 4

TABLE 2 CAMPAIGN IRS-18 V det (V) T (K) n 6 good BFOM FFOM TFOM ( 10 6 ) 3.4 4.4 0.734 23413 371.2 8.7 3.4 5.0 0.733 23100 374.4 8.6 3.4 3.8 0.734 20419 360.8 7.4 3.8 5.0 0.906 9558 504.8 4.8 3.8 4.4 0.906 8676 504.8 4.4 3.0 5.0 0.464 14298 254.2 3.6 3.0 4.4 0.464 14703 246.1 3.6 3.0 3.8 0.464 14783 237.3 3.5 3.8 3.8 0.907 5417 485.9 2.6 TABLE 3 CAMPAIGN IRS-12 V det (V) T (K) n 6 good BFOM FFOM TFOM ( 10 6 ) 3.4 4.4 0.822 26597 285.3 7.6 3.4 5.0 0.822 26139 287.9 7.5 3.4 3.8 0.822 22948 277.4 6.4 3.0 4.4 0.600 19058 265.5 5.1 3.0 5.0 0.600 18407 274.7 5.1 3.0 3.8 0.600 19472 256.0 5.0 3.8 5.0 0.929 10137 360.0 3.6 3.8 4.4 0.929 9323 360.3 3.4 3.8 3.8 0.930 6033 347.4 2.1 5

Figure 1 Comparison of LH darks at different voltages. Upper left: V det =3.0 V. Upper right: V det =3.4 V. Lower left: V det =3.8 V. The temperature for all the slope images presented here is 4.4 K. These images have been cleaned from cosmic ray signatures and all have the same ramp times. 6

TABLE 4 CAMPAIGN IRS-N2 V det (V) T (K) n 6 good BFOM FFOM TFOM ( 10 6 ) 3.4 5.0 0.879 24543 471.0 11.6 3.4 4.4 0.879 24583 466.2 11.5 3.0 4.4 0.748 22190 480.5 10.7 3.0 5.0 0.747 20684 497.9 10.3 3.0 3.8 0.748 22043 461.4 10.2 3.4 3.8 0.879 20995 451.1 9.5 3.8 5.0 0.941 6857 415.0 2.8 3.8 4.4 0.942 6166 415.7 2.6 3.8 3.8 0.942 3863 397.7 1.5 In order to have a better understanding of the impact of a voltage change on sensitivity, we used the test results from the University of Rochester Group (URG) on antimony-doped silicon (Si:Sb) flight detectors. The report shows that the detector quantum efficiency (DQE) of Si:Sb detectors decreases when V det is raised. Figure 2 shows the relative signal-to-noise ratio (RSNR) as a function of V det and wavelength for the LH detector. We can see that the RSNR drops rapidly after 30 µm and that it goes below 3 at 36 µm when V det is 3.0 V, at 33.5 µm for a V det of 3.4 V and at 31.5 µm at V det of 3.8 V. Also, the photoresponse degrades significantly at wavelengths greater than 35 µm. This issue was addressed by the URG s analysis. They found that sensitivity decreases roughly linearly with wavelength between 30 µm and a cut-off wavelength, defined by some minimum DQE (5% or 10% are typical). The URG concluded that V det < 3.0 V is required to give 10% DQE at 36 µm. Table 5 shows how the cut-off values vary as a function of V det. The current VTS for LH is 3.0V/4.4K (Table 1). For this particular setting, the Campaign IRS-N2 data (Table 4) gave a TFOM close to the maximum TFOM. The Campaign IRS-12 analysis (Table 3) indicates that the current VTS is lower than the maximum TFOM by about 32%. At the time, the LH detector was judged healthy and the decision was taken to leave its voltage and temperature unchanged. Also, the results from Campaign IRS-N2 and IRS-12 showed that it took nearly 7

Figure 2 Relative SNR for different voltages as reported by the URG. 8

TABLE 5 CUT-OFF WAVELENGTHS V det (V) 10% DQE λ (µm) 5% DQE λ (µm) 3.0 35.1 37.6 3.4 33.5 36.3 3.8 31.5 34.1 a year to degrade the LH detector and letting the detector degrade at this rate for another year seemed reasonable. 4 Discussion From Campaign IRS-12 to Campaign IRS-18, the VTS TFOM has degraded from 32% to 58% below maximum. Also, the number of bad pixels has nearly doubled (Tables 2 and 3). This degradation is similar to what was observed from Campaigns IRS-N2 to IRS-12. However, 11 months separated Campaign IRS-N2 from Campaign IRS-12 while only 4 months separated Campaign IRS-12 from Campaign IRS-18. The Solar storm that happened after Campaign IRS-17 seems to have done as much damage as 11 months of exposure to low-level Solar activity. It is also important to notice that the probability of having a bad pixel in any 6-pixel window (1 n 6 good ) has become greater than 50% for the first time in Campaign IRS-18. Increasing the value of V det to 3.4 V would maximize the operating VTS (Table 2). The operating temperature has little effect on the TFOM. Also, an operating value of 3.4 V would bring the probability of finding a bad pixel in a 6-pixel window back to 27%. Nevertheless, a change of voltages impacts the sensitivity of the detector. If we assume that a SNR of 3 represents a detection limit, then a source of constant f λ estimated to have a SNR of 4.5 at 28.15 µm will be undetected at wavelengths greater than 36 µm at the present voltage settings (V det =3.0 V; Figure 2). If V det is raised to 3.4 volts, then the same source will have a SNR of 4.1 (at 28.15 µm) and will be undetected at wavelengths longer than 33.5 µm. At 3.8 V, the source SNR will be 3.4 and it will be undetected at 31.5 µm and after. 9

5 Required Calibration and Analysis A change in operating parameters will require new calibration observations in order to characterize the behavior of the LH detector. These observations include, but may not be limited to: 1. SUR and SRS Stim data to characterize the droop and linearity coefficients at the new bias setting. We estimate that 1 2 hrs are required to collect these data. 2. Bright star spectral maps and deep zodiacal spectra to characterize the flat field at the new bias settings. Two super-sampled (1/3-pixel steps) spectral mapping observations of standard stars are required to produce slit-filling spectrophotometric observations. The stars ξ Dra (HR 6688) and 35 Dra (HR 6705) are used for this. One star provides good coverage of the blue responsivity, and the other is optimized for the red. Deep zodiacal observations are used to correct for uneven illumination along the slit. We estimate that 7 8 hrs are required to collect these data. In addition, we will need to build up a set of darks and standard star (photometric) observations to fully calibrate the LH module at the new bias setting. The standard star observations will need to be obtained over a number of IRS campaigns. It is expected that the wavelength solution will not be affected by the bias change. It will be necessary to either suspend science operations with the LH module, or embargo science data taken with the LH module, while these calibration observations, and the subsequent analysis and derivation of coefficients for the pipeline, are performed. We estimate that this process (inclusive of the observations and assuming a two-week IRS campaign) will take approximately 5 7 weeks, excluding the time for observations of multiple (photometric) standard stars for which we have accurate models. Once the pipeline has been validated with the new coefficients, science observations can begin in parallel with ongoing photometric calibration. 6 Recommendation The FOM analysis of Campaign IRS-18 data shows that raising V det would bring the operating VTS to the maximum TFOM. It would also cut the number of bad 10

pixels by a factor of approximately two. Tables 2, 3 and 4 show that the merit for the 3.0 V (V det ) voltage has degraded since Campaign IRS-N2. Based on that, we recommend that the operating parameters for the Long-High module on the Infrared Spectrograph aboard Spitzer be modified as a compromise between perpixel sensitivity and bad-pixel count. Our estimate for best performance shows that the detector substrate voltage (V det ) should be changed from 3.0 V to 3.4 V, and the focal plane temperature should be left at 4.4 K. A Appendix Voltage change procedure V bias is set by setting V det ; the two quantities are related by V bias + V det = 5.0 V. The seting is accomplished by changing the patchable constant CD DAC SUB DAC when a CESISETVLT command is issued. (The detector substrate voltage is referred to in some documentation as V detsub.) The focal plane temperature is changed by modifying the CESITRMON variable in the READY to OBSERVE block, not by changing a patchable constant. 11