Cryogenic Target Handling System Operations Manual Volume IV CTHS Description Chapter 5: Tritium Fill Station (TFS)

Similar documents
Advanced Tritium Recovery System

Performance of the Tritium Removal Systems at LLE

Cryogenic Target Handling System Operations Manual Volume IV CTHS Description Chapter 12: Tritium Removal System (TRS)

Tritium Inventories and Tritium Safety Design Principles for the Fuel cycle of ITER

High-Pressure Volumetric Analyzer

Manufacturing Process of Self-Luminous Glass Tube (SLGT) Utilizing Tritium Gas: Design of Tritium Handling Facilities

Microscopy Cryostat System

On the Emissivity of Silver Coated Panels, Effect of Long Term Stability and Effect of Coating Thickness

3.1 Tritium Plant and Detritiation

Tritium Fuel Cycle Safety

The Vacuum Case for KATRIN

UPDATE OF ITER ISS-WDS PROCESS DESIGN 2 TW6-TTFD-TPI-55 (EFDA/ )

12 Moderator And Moderator System

The Preparative Gas Chromatographic System of the JET Active Gas Handling System Tritium Commissioning and Use during and after DTE1

10.1 Functions, Basic Configuration and System Boundaries

Preparations for proof-of- principle D-T IEC experiments

WM 05 Conference, February 27 March 3, 2005, Tucson, AZ DEVELOPMENT OF A STORAGE VESSEL AND SHIPPING PACKAGE FOR TRITIUM

Tritium Safety of Russian Test Blanket Module

GA A22900 STATUS OF THE DESIGN AND TESTING OF THE OMEGA CRYOGENIC TARGET SYSTEM (OCTS)

SOME CONSIDERATION IN THE TRITIUM CONTROL DESIGN OF THE SOLID BREEDER BLANKET CONCEPTS

SOUTHERN NUCLEAR COMPANY VOGTLE ELECTRIC GENERATING PLANT UNITS 1 AND 2 NRC DOCKET NOS AND

Titanium for Long-Term Tritium Storage (U)

Studsvik Nuclear AB, Sweden

STATE OF COLORADO DESIGN CRITERIA FOR POTABLE WATER SYSTEMS WATER QUALITY CONTROL DIVISION. Price: $5.00. Revised March 31, 1997

TRITIUM RECOVERY FROM WASTE USING A PALLADIUM MEMBRANE REACTOR

Vacuum. Residual pressure can thwart the best cryogenic design. Each gas molecule collision carries ~kt from the hot exterior to the cold interior.

Figure 1 Enlargement of Powdered Activated Carbon by microscope.

An Overview of SRNL Tritium Activities

MELCOR Analysis of Helium/Water/Air Ingress into ITER Cryostat and Vacuum Vessel

ESS TARGET STATION AN OVERVIEW OF THE MONOLITH LAYOUT AND DESIGN

3. Chemical Hygiene Plan: Laboratory Standard Operating Procedures. A. Laboratory Specific Information and Signatures

Complex Compounds Background of Complex Compound Technology

PRODUCTION OF URANOUS NITRATE WITH HYDRAZINE NITRATE AS REDUCING AGENT IN PRESENCE OF ADAMS CATALYST FOR URANIUM RECONVERSION PURPOSES

Thermal Coatings for In-vacuum Radiation Cooling LIGO-T C R. Abbott, S. Waldman, Caltech 12 March, 2007

Two mark questions and answers UNIT I BASIC CONCEPT AND FIRST LAW SVCET

Overview of Control System Design

HEAT TRANSFER CAPABILITY OF A THERMOSYPHON HEAT TRANSPORT DEVICE WITH EXPERIMENTAL AND CFD STUDIES

Cryogenics for bolometric interferometry: The BRAIN experiment

TECHNICAL NOTE related to the Market survey on Water Holding Tanks for the Water Detritiation System (WDS) Prepared by Reviewed by Approved by ---

THE CERN CRYOGENIC TEST FACILITY FOR THE ATLAS BARREL TOROID MAGNETS

GA A22714 THE DESIGN OF THE OMEGA CRYOGENIC TARGET SYSTEM

VACUUM PUMPING METHODS

Experience with Tritium Evolution During Irradiation of MSRE Flibe in the MITR

TECHNICAL MANUAL 820 LX / 910 LX / 1300 LX

Tritium Control and Safety

Tritium Inventories and Containment Philosophy for the Fuel cycle of ITER

CRYOGENIC SYSTEMS FOR INERTIAL FUSION ENERGY. CEA-Grenoble France DSM/DRFMC/SBT

EVIDENCE FOR PRODUCTION OF TRITIUM VIA COLD FUSION REACTIONS IN DEUTERIUM GAS LOADED PALLADIUM

CHLORINE RECOVERY FROM HYDROGEN CHLORIDE

Be prepared to discuss the quantitative comparison method in the oral exam.

1/54 Circulation pump, safety valve, expansion vessel

OVERVIEW OF THE TRITIUM-IN-AIR MONITORING SYSTEM OF CERNAVODA NPP U1 ROMANIA MODERNIZATION AND IMPROVEMENT PROJECT

Magnetic Property Measurement System

Vacuum Pumps. Two general classes exist: Gas transfer physical removal of matter. Mechanical, diffusion, turbomolecular

DEPOSITION OF THIN TiO 2 FILMS BY DC MAGNETRON SPUTTERING METHOD

Modification In Charging Composition In Order To Arrive At Desired Circulation Composition In The Context Of Sorption Compressor Based J-T Cooler

CEA Saclay Seminar. Cryogenic Research for HTS Transmission Cables in Korea

Chemistry Instrumental Analysis Lecture 27. Chem 4631

SafeN130 Movable Apparatus for Radioactive Noble Gases Removing

Flight Acceptance Testing of the Two JPL Planck Sorption Coolers

Chemical Health and Safety General Program

Exercise 4-3. Titration of Weak Acids EXERCISE OBJECTIVE DISCUSSION OUTLINE. The 5% rule DISCUSSION

FVBSS fixed volume bottle

SORB-AC Getter Wafer Modules and Panels. We support your innovation

Hazard Communication & Chemical Safety. Based on OSHA Standard

Summary Tritium Day Workshop

The UK Tritium Users' Group

SECTION 8 Part I Typical Questions

costech instruments ECS 4010 Elemental Combustion System CHNS-O

The vacuum insulated transfer lines for CMS CO 2 cooling: performances and lessons learnt

Product Description and Specification

Report Liquid Hydrogen Target for the COMPASS Experiment

Noble Gas Control Room Accident Filtration System for Severe Accident Conditions (N-CRAFT)

Welcome to the 2015 Radiation Safety Refresher Training session for sealed source users. As a radiological worker, training concerning the safety

Safety Issues Related to Liquid Metals

Successful Development of a New Catalyst for Efficiently Collecting Tritium in Nuclear Fusion Reactors

SAFETY TRAINING LEAFLET 08 ACETYLENE, CALCIUM CARBIDE, LIME SLUDGE AND PURIFYING MATERIALS

SOPHISTICATED DESIGN, INTUITIVE OPERATION, RESEARCH-GRADE RESULTS

CHAPTER 5 CONVECTIVE HEAT TRANSFER COEFFICIENT

Experimental Investigation of Hybrid System Pulse Tube and Active Magnetic Regenerator

PowerPoint Presentation by: Associated Technical Authors. Publisher The Goodheart-Willcox Company, Inc. Tinley Park, Illinois

Data acquisition system for optimization and control of the processes from an isotopic exchange column

Monitoring Flammable Vapors and Gases in Industrial Processes

New Test Results For Physical Separation Of Tritium From Noble Gases And It s Implications For Sensitivity And Accuracy In Air And Stack Monitoring

Methods to identify and locate spent radiation sources

Lecture 4. Ultrahigh Vacuum Science and Technology

Hydrogen & Chlorine monitoring for the Chlor-Alkali Industry

14.5 K Hydrogen Sorption Cooler: Design and Breadboard Tests

COVENANT UNIVERSITY NIGERIA TUTORIAL KIT OMEGA SEMESTER PROGRAMME: MECHANICAL ENGINEERING

Separations and Reactors Design Project. Production of Allyl Chloride

Cryogenic and Electrical Test Results of a 30 M HTS Power Cable

ORTEC. SLP Series Silicon Lithium-Drifted Planar Low-Energy X Ray Detector Product Configuration Guide

Sampling. Information is helpful in implementing control measures for reducing pollutant concentration to acceptable levels

Solar Flat Plate Thermal Collector

NON-METHANE ORGANIC CARBON ANALYZER (NMOC Method 25)

Tritium Transport Modelling: first achievements on ITER Test Blanket Systems simulation and perspectives for DEMO Breeding Blanket

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds

Chem 481 Lecture Material 4/22/09

IL H DESIGN FOR KRYPTON-85 ENRICHMENT BY THERMAL DIFFUSION

Parts Manual R410A Models

Transcription:

CHATER 5: TRITIUM FILL STATION (TFS) REVISION B (ARIL 2004) S-AA-M-31 Cryogenic Target Handling System Operations Manual Volume IV CTHS Description Chapter 5: Tritium Fill Station (TFS) Table of Contents 5.0 INTRODUCTION... 1 5.0.1 Equipment Overview... 2 5.0.2 Overview of Room-Temperature Fill... 2 5.0.3 Overview of Cryogenic, High-ressure Fill... 4 5.0.4 Tritium Fill Station Summary... 4 5.1 EQUIMENT OVERVIEW... 4 5.1.1 Uranium Bed... 4 5.1.2 Assay Volume... 5 5.1.3 Target Filling System... 5 5.1.3.1 Closed Cycle Cryogenic Refrigeration System... 6 5.1.3.2 Condensation Tube... 6 5.1.3.3 Contact Tube... 6 5.1.4 Vacuum umps... 7 5.2 TFS GLOVEBOX CONTROL SYSTEM... 7 5.2.1 Glovebox urification System (GS)... 9 5.2.2 rocess Loop Getters... 9 5.2.3 Glovebox ressure Control System... 10 5.2.4 Glovebox Bubbler ressure Control... 10 5.2.5 Overhoff Tritium Air Sampler... 10 5.2.6 Glovebox... 11 5.2.7 Glovebox Cleanup System Instrumentation... 11 i

CRYOGENIC TARGET HANDLING SYSTEM VOLUME IV 5.3 ROCESS OVERVIEW... 13 5.3.1 Initialization and reparation... 13 5.3.2 Driving DT from the U-Bed to the Condensation Tube... 13 5.3.3 Expanding the DT into the Assay Volume... 14 5.3.4 Condensing the DT Inventory Required for a Room-Temperature Fill... 14 5.3.5 Returning the Unused DT to the Cold U-Bed... 15 5.3.6 Target Filling... 15 5.3.7 Returning DT to the Uranium Bed... 16 5.4 TRITIUM MANAGEMENT AND MONITORING... 16 5.4.1 Tritium Monitors... 17 5.4.2 Tritium Activity... 17 5.5 CONTROLS... 18 ii

CHATER 5: TRITIUM FILL STATION (TFS) REVISION B (ARIL 2004) AGE 1 Chapter 5 Tritium Fill Station (TFS) 5.0 INTRODUCTION LLE and Ontario Hydro have collaborated since FY86 to construct a facility at LLE to fill polymer shells with DT. The Tritium Fill Station (TFS) was delivered during FY91 and reconfigured to fill stalk-mounted OMEGA targets during FY95. The TFS is used for room-temperature, low-pressure fills and in conjunction with the DTHS for cryogenic, high-pressure fills. For room-temperature, low-pressure fills, a maximum of 18 targets, divided into three batches of six, can be filled with DT in a fill cycle. Each batch can be filled to a different pressure, up to a maximum of 150 atm. The required fill pressure, the thickness of the target wall, and the permeation time constant of the Al barrier layer determines the length of the cycle. Typically, 1-mm-diam targets with 10-µm walls are filled to 10 atm in 1 day. Thicker-walled targets are filled to higher pressures in two days. For cryogenic, high-pressure fills, the TFS is used to supply tritium to the DTHS to fill a maximum of four targets. The high-pressure DT fill process is described in Chap. 13. These highpressure fills require 18 h to pressurize targets and 2 days to cool targets. As illustrated in Fig. 5.0-1, the tritium-handling equipment is housed entirely in a helium glovebox that has an integral tritium-decontamination unit and a backup tritium scrubber. The system is computer controlled throughout the multiple-day filling sequence. All the safety devices are passive fail-safe components. Figure 5.0-1 The Tritium Fill Station process equipment can be seen through the three panels. The panel on the right houses the U-beds and pumps; the middle panel contains the contact tube and condensation cell. The left panel is used as a pass box to bring equipment or targets in and out of the TFS. The panel on the left also provides cold storage for room-temperature, low-pressure-filled targets. G6041

AGE 2 CRYOGENIC TARGET HANDLING SYSTEM VOLUME IV: CTHS DESCRITION 5.0.1 Equipment Overview The major components in the Tritium Fill Station are shown schematically in Fig. 5.0-1. A brief description of these components follows: 1. Uranium beds: The uranium beds absorb tritium at ambient temperatures. The tritium is released when the U-bed is heated to 430 C. 2. Calibrated volume: This 7.2-liter vessel is used to expand the tritium inventory. The vessel has pressure and temperature gauges so accurate assays can be performed. 3. Condensation cell: The condensation cell is a vacuum-jacketed vessel that has a cold head to bring a cold finger to cryogenic temperatures. The tritium is condensed on the cold finger. 4. Contact tube: This tube contains the targets and is where the target filling takes place when the TFS is used to fill ambient pressure targets. 5. Normetex pump: This pump is used for circulation and evacuation procedures. 6. Metal bellows pump: This pump serves as the backing pump for the Normetex pump. It exhausts into the glovebox purification system. 7. Glovebox purification system: This system removes impurities, moisture, and tritium from the glovebox and process loop. 8. TFS glovebox: The glovebox is filled with He and is a barrier to outside moisture. It is also designed to contain any tritium that may be accidentally released. The glovebox contains several glove ports that are used to maintain the equipment and transfer targets in and out of the TFS. 5.0.2 Overview of Room-Temperature Fill The maximum permissible tritium inventory at LLE is 1 gram of tritium (10,000 Ci). This tritium is reversibly stored in one of two uranium beds. At room temperature, the U-beds absorb deuterium/tritium (by converting DT to uranium hydride), and the tritium is released when the U-bed is heated to 430 C. To improve the efficiency of DT removal, it is condensed in the condensation cell to minimize back pressure. The DT is transferred from the U-bed to the condensing cell and then to the assay volume. The assay volume is a 7.2-liter volume that has two pressure gauges and temperature gauges that are used to assay the contents. A portion of the DT is then recondensed in the condensation cell. The DT that is not condensed is returned to the cold U-bed. The condensation cell temperature is then raised to evaporate the DT and increase the pressure. The DT is then transferred to the contact cell where the targets reside. After the fill has been completed, the DT is transferred to the second U-bed (referred to as the cold U-bed ). The residual DT left in the process line is removed by circulating the residual DT over the cold U-bed using the Normetex pump. The system is then evacuated.

CHATER 5: TRITIUM FILL STATION (TFS) REVISION B (ARIL 2004) AGE 3 V22 V21 V4 Normetex pump V6 V61 V3 V5 T1B V19 T1A V2 V1 Calibrated volume T1A T1B V60 V20 V18 V301 V301 V305 V16 V14 Metal bellows pump V7 V11 V50 V57 V55 V59 V16 U-bed #1 V14 V13 V56 V58 Ion gauge V54 V17 V15 V53 V12 V9 V60 G6042 V17 V15 To the TFS glovebox U-bed #2 Room 157 urification system V102 V101 V309 V10 V10 Condensation tube Contact tube Figure 5.0-1 The process and instrument diagram for the Tritium Fill Station. V62 V2 V1 To DTHS

AGE 4 CRYOGENIC TARGET HANDLING SYSTEM VOLUME IV: CTHS DESCRITION 5.0.3 Overview of Cryogenic, High-ressure Fill The TFS supplies DT for the cryogenic, high-pressure fills. The DT is first transferred to the condensation cell by heating the U-bed and condensing the DT in the condensation cell. The pressure remains very low at the end of the heating cycle because the DT gas is condensed. This allows all of the DT to be driven off the hot U-bed in one heating. The DT is then expanded to the assay volume, and an assay is taken. After the assay, the DT is transferred to the DTHS (condensed in DTHS condensation tube). The DTHS will return portions of the inventory during the high-pressure fill cycle. Each time DT is returned, the DT is assayed and then transferred to the cold U-bed. The residual DT is then circulated over the cold U-bed and the TFS is evacuated. The high-pressure fill cycle is described in more detail in another chapter. 5.0.4 Tritium Fill Station Summary The following is an overview of the TFS: The TFS can simultaneously fill 18 capsules with up to 60 atm of DT at room temperature. The TFS can contain up to 1 g (10,000 Ci) of tritium mixed with an equimolar ratio of deuterium. A typical 1-mm target filled to 20 atm contains about 10 mci of DT. Filling operations are accomplished in gloveboxes to meet environmental, health, and safety regulations. 5.1 EQUIMENT OVERVIEW As originally constructed, the Tritium Fill Station (TFS) consisted of the receiving and inventory loop, pumping system, DT storage loop, capsule-charging loop, piping and valving, glovebox, ancillary systems, glovebox purification system, electrical and instrumentation systems, glovebox atmosphere monitors, control system, and the supporting facility. In 1995 the TFS was modified with the addition of a glovebox annex to allow for the handling and coating of filled targets, modification of the process loop to allow for additional cryogenic filling capabilities, replacement of the contact tube to allow filling of mounted targets, and installation of a cryogenic cooler. The TFS is a very sophisticated piece of equipment and will be discussed in more detail in subsequent sections. 5.1.1 Uranium Bed Long-term storage of DT in the TFS is provided by a pair of uranium storage beds (U-beds shown in Fig. 5.1-1) to limit leaks and outgassing, as well as reduce the possibility of accidental releases. At ambient temperature, hydrogen isotopes form uranium hydrides within the beds. One storage bed provides adequate storage capacity for the entire TFS inventory, but a second bed has been provided to serve as a backup. The prime function of the uranium storage bed is to store tritium and deuterium in a stable form to minimize the possibility of releases from the system. A secondary function of the bed is to act as a helium filter. Hydrogen isotopes and impurities such as O 2, N 2, H 2 O, and hydrocarbons form chemical compounds with uranium, but helium is left in a gaseous form. After the hydrogen isotopes have been absorbed, residual helium can be evacuated from the process systems.

CHATER 5: TRITIUM FILL STATION (TFS) REVISION B (ARIL 2004) AGE 5 The beds are enclosed in a rechargeable vacuum jacket to maximize thermal efficiency and reduce outer-skin temperature to 50 C. Heating the U-bed to 400 C desorbs the elemental D and T that are chemically bonded to the uranium. Figure 5.1-1 Tritium is stored in uranium beds. The U-beds absorb tritium at room temperatures and releases the tritium at ~400 C. U-bed T1305 5.1.2 Assay Volume The major component of the receiving and inventory loop is a 7.13-liter calibrated volume. Tritium and deuterium are loaded into the loop through the calibrated volume, where they are assayed using high-accuracy temperature- and pressure-sensing equipment. 5.1.3 Target Filling System The target filling system is comprised of four major components: 1. A closed-cycle cryogenic refrigeration system, 2. A condensation tube, 3. A pressure-reducing mechanism, and 4. A contact tube.

AGE 6 CRYOGENIC TARGET HANDLING SYSTEM VOLUME IV: CTHS DESCRITION 5.1.3.1 Closed-Cycle Cryogenic Refrigeration System The closed-cycle refrigeration system is used to cool a condensation tube to 10 K and (by cryopumping action) to evacuate DT from the uranium getters or calibrated volume to the condensation tube. The refrigeration system consists of A helium compressor located outside the glovebox, A cryo-refrigerator (cold head) located inside glovebox A, A vacuum shroud, and A helium-transfer line between the compressor and the cryo-refrigerator. The refrigeration system is a closed-cycle helium refrigeration system that utilizes the Gifford McMahon refrigeration cycle. The cold head achieves the Gifford McMahon cycle by expanding highpressure helium to low pressure. This is accomplished through a reciprocating expansion device that allows the helium to expand in two stages: first to approximately 50 K and then to 10 K. Helium is then returned to the compressor through a second transfer line. The cold head is enclosed in a vacuum jacket to permit the system to achieve cryogenic temperatures by minimizing heat absorbtion from the surrounding areas. The system requires a vacuum of 60 mtorr to provide adequate insulation. 5.1.3.2 Condensation Tube Deuterium and tritium are transferred from the uranium getters or assay volume to the condensation tube (shown in Fig. 5.1-2). The tube is isolated from the other TFS loops, and the DT is cryogenically pumped to the condensation tube, where it is condensed in solid form. When the transfer is complete, the tube is heated to room temperature, generating DT pressures up to 23,800 ka. Figure 5.1-2 The condensation tube is encased in a vacuum jacket that provides thermal isolation. The cold-finger temperature is lowered to 10 K, and tritium is condensed on the cold finger. T1306 Condensation tube 5.1.3.3 Contact Tube The targets are charged in the contact tube a long tube sealed at one end with a welded end cap. The other end is fitted with a male VCR gland for connection to the charging loop by a VCR connector. Eighteen targets can be filled at one time (six per rack three racks).

CHATER 5: TRITIUM FILL STATION (TFS) REVISION B (ARIL 2004) AGE 7 5.1.4 Vacuum umps The vacuum-pumping loop consists of a Normetex scroll vacuum pump for generating vacuums in the 5-mtorr range. This is backed by a two-stage metal bellows, single containment pump. The pumps (shown in Fig. 5.1-3) are used to evacuate the various process loops prior to transfer of DT. The Normetex pump is also used to circulate hydrogen back to the U-beds to enhance gettering. Normetex scroll pump Figure 5.1-3 The Normetex pump is used to evacuate the process lines. The metal bellows pump is the backing pump. Metal bellows pump G6040 5.2 TFS GLOVEBOX CONTROL SYSTEM The inert helium atmosphere within the TFS glovebox is maintained by a glovebox control system. The purpose of this system is as follows: 1. To control the glovebox pressure, 2. To monitor and trap residual tritium in the glovebox, 3. To monitor and reduce moisture, 4. To reduce organic impurities, 5. To monitor and reduce TFS process line tritium discharge, and 6. To monitor tritium discharged to the stack. There are six subsystems (illustrated in Fig. 5.2-1) required to achieve the objectives listed above: 1. Glovebox urification System (GS): This system takes effluent from the TFS glovebox, reduces the moisture, removes impurities, and reduces the tritium content. The effluent is then returned to the TFS glovebox system. 2. rocess Loop Getters: This system takes effluent from the process line evacuation pumps, removes impurities, and reduces the tritium content. The process effluent is sent to the GS system for additional processing and is finally discharged to the TFS glovebox.

AGE 8 CRYOGENIC TARGET HANDLING SYSTEM VOLUME IV: CTHS DESCRITION 3. Glovebox rimary ressure Control System: This system controls the glovebox pressure +1/2 in. of H 2 O. This system requires He pressure and power to operate. 4. Glovebox Bubbler ressure Control: This system passively controls the glovebox pressure to +8 in. of H 2 O of setpoint. This can operate without power or He pressure and serves as overpressure protection of the glovebox. 5. Overhoff Tritium Air Sampler: This system samples the exhaust stack stream to verify compliance with environmental release limits. 6. TFS Glovebox: A plexiglass enclosure pressurized with He. External stack Vacuum relief bubbler Glovebox bubbler pressure control Cleanup system return rocess line metal bellows pump exhaust Glovebox Cleanup system input Overhoff tritium air sampler Overpressure relief bubbler ZrFe ZB-501 TZ Ni NB-501 TZ Glovebox inlet Helium 10 psig Helium supply to glovebox TMT 203 rocess loop getters TMT 201 Glovebox pressure control Glovebox primary pressure control system Glovebox exhaust TMT 202 ZrFe ZB-201 TZ NB-201 Ni DT 202 TZ Drier DT 201 TZ Glovebox recirculation pump Glovebox purification system G6043 Figure 5.2-1 The TFS glovebox control system is comprised of six subsystems: (1) glovebox bubbler pressure control, (2) glovebox, (3) process loop getters, (4) glovebox purification system, (5) glovebox primary pressure control system, and (6) Overhoff Tritium Air Sampler.

CHATER 5: TRITIUM FILL STATION (TFS) REVISION B (ARIL 2004) AGE 9 5.2.1 Glovebox urification System (GS) The inert helium atmosphere is maintained by a Glovebox urification System (GS). The glovebox helium is continuously circulated through the reactor to extract impurities, including hydrogen oxides, nitrogen, hydrogen, oxygen, hydrocarbons, and moisture. Separate moisture and tritium monitoring systems are provided to monitor the glovebox atmosphere. The basic components of the GS are summarized below and shown in Fig. 5.2-2: 1. Molecular sieve: removes moisture 2. Nickle catalyst: cracks hydrogen oxides and hydrocarbons 3. SAES (ZrFe) metal getter: removes gaseous tritium, nitrogen, and hydrogen 4. Tritium monitors 5. Glovebox recirculation pump Molecular sieve Nickel cracker SAES getter Figure 5.2-2 The Glovebox urification System s main components include the DT monitors, the SAES (ZrFe) getter, the nickel cracker, and the molecular sieve bed. T1311 Inlet DT monitor Outlet DT monitor 5.2.2 rocess Loop Getters The process loop getters are used only when the process lines are being evacuated. The process effluent is directed through the reactors to extract impurities, including hydrogen oxides, hydrogen, oxygen, and hydrocarbons. The basic components of the rocess Loop Getter are summarized below and shown in Fig. 5.2-2: 1. Nickle catalyst: cracks hydrogen oxides and hydrocarbons. 2. SAES (ZrFe) metal getter: removes tritium and hydrogen 3. Tritium monitor

AGE 10 CRYOGENIC TARGET HANDLING SYSTEM VOLUME IV: CTHS DESCRITION 5.2.3 Glovebox ressure Control System Glovebox pressure is maintained 1/2 in. of H 2 O of atmospheric pressure by a system that exhausts helium to the stack from the glovebox on overpressure conditions, or supplies fresh helium when the pressure is too low (see Fig. 5.2-1). A check valve on the cleanup system return line ensures that glovebox He does not backflow to the stack in an overpressure condition. 5.2.4 Glovebox Bubbler ressure Control The bubbler pressure control system is used to protect against extreme pressure conditions such as a failure of the primary pressure control system. This system will limit the pressures when the glovebox pressure deviates 8 in. of H 2 O from setpoint. 5.2.5 Overhoff Tritium Air Sampler An Overhoff Tritium Air Sampler (see Fig. 5.2-3) collects liquid scintillation counter samples from the exhaust stack stream to verify compliance with release limits. To date, the average tritium concentration released to the environment is below the DEC permit limit (2.2 Ci/year) established by the NYS Department of Environmental Conservation. Figure 5.2-3 The Overhoff Tritium Air Sampler collects tritium samples from the exhaust stack stream. The four liquid sample vials can be seen. G6010 Liquid scintillation sample vials A pump controlled by an adjustable constant flow controller draws a representative sample from the exhaust line and passes it through liquid bubbler scintillation vials separated by a high-efficiency (typically 99%) catalytic oxidizer. The first two vials in the process stream collect the oxide form of tritium, and the second two vials collect elemental tritium after it is converted to the oxide form by the catalytic oxidizer. Each vial has a >95% collection efficiency. The vials are located on the front panel of the sampler for ready detachment and counting by the liquid scintillation counter. The system is calibrated to determine the percentage of tritium collected by the vials to allow accurate measurements of tritium. Typically, a collection time of 8 h will yield detection thresholds to <10 9 Ci/ml. Increasing the collection time results in even-lower detection thresholds.

CHATER 5: TRITIUM FILL STATION (TFS) REVISION B (ARIL 2004) AGE 11 5.2.6 Glovebox The glovebox is fabricated of stainless steel with plexiglass windows on the front and back. Materials are transferred into or out of the glovebox through an airlock. Glove ports are provided on the front and back of the glovebox to permit access for operation and maintenance (as can be seen in Fig. 5.2-4). Figure 5.2-4 The gloves allow the operator to load or unload targets and perform maintenance on the process equipment. Gloves T1304 5.2.7 Glovebox Cleanup System Instrumentation The instrumentation rack for the glovebox cleanup system is shown in Fig. 5.2-5. All of these instruments can be located on the process diagram shown in Fig. 5.2-1. 1. Cleanup System Inlet Monitor (TMT-201): This monitor measures the tritium activity of the Glovebox Cleanup System inlet. This is the tritium activity level of the TFS glovebox. A high activity level will initiate an emergency stop. 2. rocess Loop Outlet Monitor (TMT-203): This monitor measures the tritium activity of the TFS process loop outlet (after the process loop getters). This shows the tritium activity level of the TFS process loop before it enters the Glovebox Cleanup System. 3. Cleanup System Outlet Monitor (TMT-202): This monitor measures the tritium activity of the Glovebox Cleanup System outlet before it is returned to the TFS glovebox. 4. Glovebox Getter Heater Controllers (ZB-201 and NB-203): These units control the temperature of the glovebox getters.

AGE 12 CRYOGENIC TARGET HANDLING SYSTEM VOLUME IV: CTHS DESCRITION 5. Glovebox Getter Over Temperature Controllers: These units monitor the temperature of the glovebox getters and disrupt the power to the heaters. 6. rocess Loop Getter Heater Controllers (ZB-501 and NB-501): These units control the temperature of the process loop getters. 7. Moisture Sensors (DT-201 and DT-202): There are two moisture sensors in the glovebox cleanup system; one sensor measures the moisture level at the drier inlet and the other sensor measures the dryer outlet moisture level. A high moisture level will initiate a emergency stop. Glovebox getter over temperature controllers rocess loop outlet tritium monitor Cleanup system inlet tritium monitor (TMT-201) Cleanup system outlet tritium monitor (TMT-202) Glovebox getter heater controllers rocess loop getter bed heater controllers G6227 Figure 5.2-5 The Glovebox Cleanup System instrumentation rack.

CHATER 5: TRITIUM FILL STATION (TFS) REVISION B (ARIL 2004) AGE 13 5.3 ROCESS OVERVIEW During the loading operation, up to 10,000 Ci of tritium are retrieved from the uranium bed, assayed, and transferred to the high-pressure loop of the system where actual loading takes place. The different modes of operation during a typical loading cycle are described below. 5.3.1 Initialization and reparation The system is evacuated ~5 mtorr, venting the effluent to the glovebox purification system to remove any DT. The contact cell containing the mounted capsules is connected to the TFS, and the system is evacuated. The high-pressure loop is then isolated. 5.3.2 Driving DT from the U-Bed to the Condensation Tube The contact cell containing the mounted capsules is connected to the TFS and the system is evacuated. The condensation tube (at 10 K) is connected to the U-beds. The DT inventory is driven off the U-bed by energizing the bed heaters and condensed into the condensation tube. The DT pressure is kept to a minimum since the gas is condensed as it is driven off the U-bed (see Fig. 5.3-1). Assay volume U-bed #1 Scroll pump Figure 5.3-1 The uranium bed is heated to condense the DT into the condensation tube. U-bed #2 T Metal bellows pump Condensation tube Contact cell G6103a

AGE 14 CRYOGENIC TARGET HANDLING SYSTEM VOLUME IV: CTHS DESCRITION 5.3.3 Expanding the DT into the Assay Volume The assay volume is connected to the condensation tube. The DT is then expanded into the assay volume by heating the condensation tube to ~40 K (see Fig. 5.3-2). Assay volume U-bed #1 Scroll pump U-bed #2 T Metal bellows pump Condensation tube Contact cell G6103b Figure 5.3-2 The DT is expanded into the assay volume by heating the condensation tube. 5.3.4 Condensing the DT Inventory Required for a Room-Temperature Fill The room-temperature fills require only a fraction of the DT inventory. The required amount is condensed back to the cold finger by cooling the condensation tube to ~10 K (see Fig. 5.3-3). Assay volume U-bed #1 Scroll pump U-bed #2 T Metal bellows pump Condensation tube Contact cell G6103c Figure 5.3-3 The pressure of the DT is increased by first condensing it, confining it, and then reheating it.

CHATER 5: TRITIUM FILL STATION (TFS) REVISION B (ARIL 2004) AGE 15 5.3.5 Returning the Unused DT to the Cold U-Bed The cold U-bed is connected to the assay volume and the unused DT is pumped to the cold U-bed (see Fig. 5.3-4). Assay volume U-bed #1 Scroll pump U-bed #2 T Metal bellows pump Condensation tube Contact cell G6103d Figure 5.3-4 The DT not required for the fill is absorbed by the cold U-bed. 5.3.6 Target Filling The condensation tube is heated to return it to room temperature, thereby increasing the pressure of the DT. The DT is transferred to the contact cell by sequentially opening and closing valves at each end of a trapped volume. The capsules are therefore subject to a gradual increase in pressure until the desired pressure in the capsules is attained (see Fig. 5.3-5). Assay volume U-bed #1 Scroll pump U-bed #2 T Metal bellows pump Condensation tube Contact cell G6103e Figure 5.3-5 The condensation cell is brought up to room temperature, which releases and pressurizes the DT. The DT is then transferred to the contact cell, where the targets reside.

AGE 16 CRYOGENIC TARGET HANDLING SYSTEM VOLUME IV: CTHS DESCRITION 5.3.7 Returning DT to the Uranium Bed After charging, the DT is transferred from the high-pressure loop to the uranium storage beds, where the remainder of the DT inventory is stored. The high-pressure loop is evacuated and backfilled with helium, and the contact cell and capsules are removed from the TFS for post-filling processing (see Fig. 5.3-6). Assay volume U-bed #1 Scroll pump U-bed #2 T Metal bellows pump Condensation tube Contact cell G6103f Figure 5.3-6 After the fill has been completed, the DT is pumped back to the cold U-bed. 5.4 TRITIUM MANAGEMENT AND MONITORING Under normal operation, the TFS completes up to one loading cycle every 3 days. The actual tritium content of each polymer capsule will be variable dependent upon the experimental program. A reference value of 12.5 mci is assumed for gaseous targets, and a reference value of 0.35 Ci is assumed for cryogenic targets. The TFS was designed to achieve practically zero leakage of tritium from system components located in the glovebox. The glovebox serves as a second barrier to prevent tritium leakage into the atmosphere. Any tritium that enters the glovebox will be collected in the Glovebox urification System. As a third barrier to people within the UR/LLE building, the room housing the TFS is kept at a negative pressure to ensure zero leakage to adjacent spaces. To ensure safety, accidents that could affect either the general environment and population or individual UR/LLE employees were analyzed. The analysis concluded that there was no accident that would put either the general population or environment at risk. An accident that releases the entire inventory of tritium to the TFS room and then from there to the environment in one hour is the worstpossible accident. This accident would result in a 6-µrem exposure to the general public and <60-mrem exposure to an operator. The TFS Radiological Safety Assessment evaluates the probability of this accident to be very remote since it would require at least dual failures (i.e., failure of the TFS system resulting in release to the glovebox and the simultaneous failure of the glovebox).

CHATER 5: TRITIUM FILL STATION (TFS) REVISION B (ARIL 2004) AGE 17 5.4.1 Tritium Monitors Tritium levels are monitored continuously inside the glovebox by a tritium monitor integral to the TFS. If this monitor detects excessive tritium, the TFS is automatically shut down and the tritium is returned to the uranium beds. 5.4.2 Tritium Activity Normal glovebox concentrations typically remain <5 mci/m 3. Stack effluent concentrations average <0.05 µci/m 3 for each year. Tritium activity trended against process operations can be seen in Fig. 5.4-1. Box activity (mci/m 3 ) T1646 100 10 1 rocessed targets Cleanup system removed for regen Shield exposed Removed cells Removed cells Cleanup system removed for regen eriod covered: 1/24 to 8/4/97 0.1 0 50 100 150 200 rocessed targets Elapsed time from 1/24/97 (days) rocess loop evacuted Cryostorage opened Figure 5.4-1 The process operation impacts the tritium activity within the glovebox.

AGE 18 CRYOGENIC TARGET HANDLING SYSTEM VOLUME IV: CTHS DESCRITION The getter system has proven to be very effective. In one incident where 450 Ci was released into the glovebox, the concentration was reduced to about 100 mci/m 3 after 1 day. Tritium in the stack effluent was measured for weeks following the incident. This can be seen in Fig. 5.4-2. Box atmosphere activity (Ci/m 3 ) 10 2 10 1 10 0 10 1 t = 8.3 h Instrument automatic scale change 17 18 April 2000 t = 4.2 h T1647 0 5 10 15 20 25 Friday Time (h) 1:20 pm Figure 5.4-2 The glovebox purification control system was very effective in processing a 450-Ci release in the glovebox. 5.5 CONTROLS All operations in the TFS are manually controlled and initiated, with the exception of the emergency stop and the target-filling sequence, which is manually initiated, but automatically controlled by a programmable logic controller located in a control console adjacent to the glovebox. In addition to the LC, the panel contains an LCD touch panel for control of the valves, instrumentation readouts, alarms, etc. Most process-related instrumentation readouts are located on the controls console.