Development And Testing of a Thermoluminescent Dosemeter for Mixed Neutron-Photon-Beta Radiation Fields

Similar documents
A PROPOSED FOUR-ELEMENT NEUTRON-PHOTON-BETA TL DOSEMETER * Stanford Linear Accelerator Center, MS 48, P.O. Box 4349, C. S. Sims and A. B.

Characterization of Low-Energy (6-30 kev) Response of Polish TLDs (MTS-N, MCP-N) with Synchrotron Radiation and Determination of Some TLD Quantities.

Energy Response Characteristics of Several Neutron Measuring Devices Determined By Using the Scattered Neutron Calibration Fields of KAERI

THE ACTIVE PERSONNEL DOSIMETER - APFEL ENTERPRISES SUPERHEATED DROP DETECTOR*

PERSPECTIVES OF PERSONNEL EXTERNAL DOSIMETRY AT STANFORD LINEAR ACCELERATOR CENTER

A COMPUTER PROGRAM FOR THE DECONVOLUTION OF THERMOLUMINESCENCE GLOW CURVES K. S. Chung 1,, H. S. Choe 1, J. I. Lee 2, J. L. Kim 2 and S. Y.

Neutron Dosimetry with Ion Chamber-Based DIS System

Performance Test of the Electronic Personal Neutron Dosemeter in Neutron Fields Simulating Workplaces of MOX Fuel Fabrication Facilities

PERSONNEL DOSE EQUIVALENT MONITORING AT SLAC USING LITHIUM-FLUORIDE TLD S*

Radiation Protection Dosimetry (2007), Vol. 126, No. 1 4, pp Advance Access publication 23 May 2007

Chapter VIII: Photographic films

A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators

A new neutron monitor for pulsed fields at high-energy accelerators

External Dosimetry at NPPs; NVLAP, Noble Gas, and TEDE

Jeancarlo Torres, CHP. Naval Surface Warfare Center Carderock Division

An extremity (wrist) dosemeter based on the Landauer InLight TM whole body dosemeter

UNIFORMED SERVICES UNIVERSITY OF THE HEALTH SCIENs:::ES F. EDWARD HEBERT SCHOOL OF MEDICINE 4301 JONES BRIDGE ROAD BETHESDA, MARYLAND

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements

Radiation Protection Dosimetry (2006), Vol. 118, No. 3, pp Advance Access publication 6 October 2005

Energy response for high-energy neutrons of multi-functional electronic personal dosemeter

Neutron and/or photon response of a TLD-albedo personal dosemeter on an ISO slab phantom

Journal of Radiation Protection and Research

Dosimetric Quantities and Neutron Spectra Outside the Shielding of Electron Accelerators

The development of new generation electronic personal dosimeters

Activities of the neutron standardization. at the Korea Research Institute of Standards and Science (KRISS)

Monte Carlo Calculations Using MCNP4B for an Optimal Shielding Design. of a 14-MeV Neutron Source * James C. Liu and Tony T. Ng

RESULTS OF UNDERGROUND DOSE RATES USING TL DETECTORS AND PRELIMINARY TL CHARACTERISTICS OF IRRADIATED SALT ROCK *

A N Singh, S N Singh & L L Singh

Neutron Detector Gamma Insensitivity Criteria

Calculations of Photoneutrons from Varian Clinac Accelerators and Their Transmissions in Materials*

CARBOXEN 564 A NEW CARBON MOLECULAR SIEVE SORBENT USED FOR RADON AND THORON MONITORING

Commissioning of the Beta Secondary Standard (BSS2)

COMPARATIVE STUDY OF PIGE, PIXE AND NAA ANALYTICAL TECHNIQUES FOR THE DETERMINATION OF MINOR ELEMENTS IN STEELS

Radionuclide Imaging MII Detection of Nuclear Emission

Recent Activities on Neutron Standardization at the Electrotechnical Laboratory

Laboratoire National Henri Becquerel (CEA/LIST/LNHB), France (2) ENEA-Radiation Protection Institute, Bologna, Italy (3)

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved

Neutron Spectrometry in Mixed Fields: Characterisation of the RA-1 Reactor Workplace

CHARACTERIZATION OF A RADIATION DETECTOR FOR AIRCRAFT MEASUREMENTS

SLAC-PUB Submitted to Radiation Protection and Dosimetry. Work supported by Department of Energy contract DE-AC02-76SF00515

OSL in Household Salt (NaCl) for Environmental, Occupational and Medical Dosimetry

Multisphere Neutron Spectrometric System with Thermoluminescent Dosemeters: Sensitive Improvement

Update on Calibration Studies of the Canadian High-Energy Neutron Spectrometry System (CHENSS)

Personal Dose Monitoring System

Light ion recoil detector

Researchers at the University of Missouri-Columbia have designed a triple crystal

Acronyms, Abbreviations, and Symbols Foreword to the First Edition Foreword to the Second Edition Preface to the First Edition Preface to the Second

EURADOS Intercomparison exercises on whole body and extremity dosemeters ( ) Outline and Results

Distinguishing fissions of 232 Th, 237 Np and 238 U with beta-delayed gamma rays

IRPA th International Congress of The International Radiation Protection Association

Thermoluminescence Properties of Local Feldspar from Gattar Mountain Area

Investigation of pulse shapes and time constants for NaI scintillation pulses produced by low energy electrons from beta decay

Development of a Passive Dosimeter for Life Science Experiments in Space (PADLES) in NASDA

Fast-Neutron Production via Break-Up of Deuterons and Fast-Neutron Dosimetry

Recent Activities on Neutron Calibration Fields at FRS of JAERI

Study Some of Optical and Thermoluminescence Properties of Muscovite Mica Exposed to Ultraviolet Radiation

GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY

Development of a Characterized Radiation Field for Evaluating Sensor Performance

City University of Hong Kong

Measurements of Radiation Doses Induced by High Intensity Laser between and W/cm 2 onto Solid Targets in LCLS MEC Instrument

Radiation Detection. 15 th Annual OSC Readiness Training Program.

Calibration of Radioprotection Instruments and Calibrated Irradiation: Characterization of Gamma Beam of 137 Cs and 60 Co

Atoms, Radiation, and Radiation Protection

Geant4 simulation of SOI microdosimetry for radiation protection in space and aviation environments

MEASUREMENT AND DETECTION OF RADIATION

Department of Radiation Protection, Nuclear Science Research Institute, Japan Atomic Energy Agency

SIMULATION OF COMPETING FADING AND IRRADIATION EFFECTS IN THERMOLUMINESCENCE (TL) DOSIMETRY: 1 st ORDER KINETICS

Comparison of some human tissues and some commonly used thermoluminescent dosimeters for photon energy absorption

Accuracy, Calibration, Type Testing and Traceability General. Peter Ambrosi

SCINTILLATION DETECTORS AND PM TUBES

Measurements with the new PHE Neutron Survey Instrument

Since the beam from the JNC linac is a very high current, low energy beam, energy loss induced in the material irradiated by the beam becomes very lar

Determination of Ambient Dose Equivalent at INFLPR 7 MeV Linear Accelerator

Recent activities in neutron standardization at NMIJ/AIST

A quantitative kinetic model foral 2 O 3 :C: TL response to ionizing radiation

RADIOACTIVITY IN THE AIR

CONSIDERATION ON THE H p(10) AND H*(10) SECONDARY STANDARD CHAMBER CHARACTERISTICS

Contents. Charged Particles. Coulomb Interactions Elastic Scattering. Coulomb Interactions - Inelastic Scattering. Bremsstrahlung

Dosimetry for space radiation in ISS lifescience experiments using PADLES system

Chapter V: Cavity theories

Digital two-parametric processing of the output data from radiation detectors

RADIATION SAFETY TRAINING SEALED SOURCES

REALISTIC NEUTRON SPECTRA FOR RADIATION PROTECTION AND OTHER APPLICATIONS AT AERI, BUDAPEST

1 of 5 14/10/ :21

Characterization Survey Techniques and Some Practical Feedback

Unit 2. Instrumentation. Experts Teaching from Practical Experience

Contribution of the Nuclear Physics Institute of the AS CR to the BEOBAL Project

Bonner Sphere Spectrometer. Cruzate, J.A.; Carelli, J.L. and Gregori, B.N.

František SPURNÝ, Iva JADRNÍCKOVÁ. Nuclear Physics Institute Department of Radiation Dosimetry, Academy of Sciences of the Czech Republic, Prague

Evaluation and Measurements of Radioactive Air Emission and Off-Site Doses at SLAC

A Selected Data on Radionuclides of Health Physics Interest

Some problems to be solved for automatic analysis of a CR-39 nuclear track detector in space radiation

SECTION 8 Part I Typical Questions

Keywords: experimental verification; LiF:Mg,Cu,P; lithium fluoride; radial dose distribution

Safety Training for Radiation Workers at ICRR, Univ. of Tokyo. April, 2017

Bubble Detector Characterization for Space Radiation

Radiation Protection Dosimetry (2007), Vol. 126, No. 1 4, pp Advance Access publication 11 May 2007

ROC7080: RADIATION THERAPY PHYSICS LABORATORY LAB TWO : THERMO LUMINESCENT DOSIMETRY

Calculation of Bubble Detector Response Using Data from the Matroshka-R Study

Portable gamma and thermal neutron detector using 6 LiI(Eu) crystals

Transcription:

SLAC-PUB-7901 August 1998 Development And Testing of a Thermoluminescent Dosemeter for Mixed Neutron-Photon-Beta Radiation Fields John J. Jummo and James C. Liu Presented at 12th International Conference on Solid State Dosimetry, 7/5/98 7/10/98, Burgos, Spain Stanford Linear Accelerator Center, Stanford University, Stanford, CA 94309 Work supported by Department of Energy contract DE AC03 76SF00515.

DEVELOPMENT AND TESTING OF A THERMOLUMINESCENT DOSEMETER FOR MIXED NEUTRON-PHOTON-BETA RADIATION FIELDS SLAC-PUB-7901 John J. Zummo and James C. Liu Genentech, Inc., 1 DNA Way, South San Francisco, CA 94080, USA SLAC, MS 48, P.O. Box 4349, Stanford, CA 94309, USA ABSTRACT A new four-element thermoluminescent (TL) dosemeter and dose evaluation algorithm have been developed and tested to better characterize personnel exposure in mixed neutron-photon-beta radiation fields. The prototype dosemeter is based on a commercially available TL card (with three LiF-7 chips and one LiF-6 chip) and modified filtration elements. The new algorithm takes advantage of the high temperature peak characteristics of the LiF-6 element to better quantify the neutron dose component. The dosemeter was tested in various radiation fields, consisting of mixtures of two radiation types typically used for dosemeter performance testing, as well as mixtures of three radiation types to simulate possible exposure conditions. The new dosemeter gave superior performance, based on the tolerance levels, when using the new algorithm as compared to a conventional algorithm that did not use the high temperature peak methodology. The limitations and further improvements are discussed.

OBJECTIVE The objective of this work was to characterize the response of the prototype dosemeter and develop an algorithm utilizing the high temperature peaks of TLD-600 to quantify the neutron component of the field (1). The dosemeter performance was tested in various mixed fields and the algorithm was compared to an algorithm that did not utilize the high temperature peaks. DOSEMETER DESIGN The Harshaw 7776-1161 card was used. The standard Harshaw 8814 dosemeter holder was modified to give a configuration as shown in Table 1. This dosemeter design is comparable to the dosemeter design of thick element 4 previously described (1), except that the thin element in position 3 is 0.15 mm versus 0.09 mm thick. The thinner chip is not currently routinely supplied in the standard dosemeter due to manufacturing and performance issues. Individual element correction coefficients were determined and applied to normalize the response of the chips to the mean of the population. IRRADIATION AND PROCESSING A total of 192 dosemeters were exposed at Pacific Northwest Laboratory to a series of American National Standards Institute (ANSI) sources (2). A total of 11 single source, 29 dual source and 6 triple source exposures were performed at dose ratios up to 3:1. The dosemeters were read on a Harshaw 8800 automated hot gas reader using a linear time-temperature profile from 50 C to 300 C, heating at 25 C/second. The glow curve data in nc was used, taking channels 96 through 145 as the main dosimetric peaks 3, 4, and 5, and channels 146 through 200 as the high temperature peaks 6 and 7 of the TLD-600 element. This followed established methodology (3). Channels 89 through 138 were taken as the main dosimetric peaks for the thin element in position 3, since the thinner chip heated at a faster rate than the thicker chips.

ALGORITHM DEVELOPMENT The single source exposures were used to determine response functions for the four TL elements to different radiations. The latest ANSI exposure to dose conversion factors (2) were used. An unpublished conversion factor was used for 241 Am (4). In the following discussion, sensitivities, responses and doses will be designated by S, R and H, respectively. Photons, betas and neutrons will be designated as p, b and n. Deep and doses will be designated as d and s. Th is the response of the high temperature peaks of element 4 (peaks 6 and 7) and Tl is the response of the low temperature peaks of element 4 (peaks 3, 4 and 5). The K value is defined to be the TL signal ratio between peaks 6-7 and peaks 3-7. In the algorithm utilizing the high temperature peak, the presence of neutrons is determined by calculating the K value from element 4. If the K value is less than 0.03, which theoretically corresponds to a photon to neutron dose ratio of 100 to 1, it is assumed that the field has no neutrons. The photon energy is determined by either the element 4/2 ratio if no neutrons are present, or the element 1/2 ratio if neutrons are present. The Cd-filtered element 4 gives a better photon discrimination than the Cu-filtered element 1. The Cu filtration is too thin to be useful for photons above 120 kev. However, the Cu filter must be used for photon energy determination when any neutrons are present. The following equations are used, as derived in a previous work (1) : Hn = {[(1-Kp)Th - KpTl] / (0.146 - Kp)} / S4n (1) R4p = (-0.854 Th + 0.146 Tl) / (0.146 - Kp) (2) Hpd = R2/S2pd (3) R3b = R3 - Hps S3ps (4) Hb = R3b/S3b (5) In the conventional algorithm that does not utilize the high temperature peak, the photon energy is based on the Cu-filtered element 1. With equations 3-5 above, the following equation is also used: Hn = (R4 - Hpd S4pd) / S4n (6)

In these algorithms, the constant values shown in Table 2 are used. Note that S4n for the new algorithm includes the response from peaks 3-7, but only peaks 3-5 for the conventional algorithm. These values are constant with dose over the two orders of magnitude tested, from 0.03 to 15 msv. RESULTS AND DISCUSSION The lower limit of detection (LLD) was calculated for each set of dosemeters following ANSI methodology (2). The LLD for all radiation fields tested for the new algorithm was < 40 msv and 20 msv, compared to < 40 msv and 10 msv for the conventional algorithm, after 3 weeks of storage. To evaluate the performance, the tolerance levels were calculated as defined by ANSI (2). A maximum level of 0.35 is considered to be acceptable performance for a personnel dosimetry system. Table 3 shows that the new algorithm has better overall tolerance levels than the conventional algorithm for the tested irradiation fields. As shown in Table 4, when considering only the neutron component of the mixed field cases, the performance of the new algorithm is markedly better. The new algorithm passed all mixed field exposures, while the conventional algorithm failed in two of the mixed field exposures. It is also noteworthy that the standard deviations of the tolerance levels are smaller for the new algorithm that indicates a more consistent performance over the mixed radiation fields that were tested. The dosemeter design and algorithm are limited in two aspects, just as other 4-element dosemeters are, in that the energy of the beta and neutron components of the mixed field need be known to give accurate results. To test if the algorithm would be usable if the beta energy was unknown, the algorithms were modified by using an average beta sensitivity value and the results are shown in Table 5. As expected, larger (but still acceptable) tolerance levels are found in the dose estimations only. This beta energy limitation of many dosemeter systems is well known (5). Another standard algorithm (6) takes advantage of a logarithmic relationship between the ratio of elements 2 and 3 and dose. This is possible when there is much less filtration over element 2. Our dosemeter design effectively stopped all the betas from reaching element 2. The other limitation, the energy of the neutron field, is more problematic. There is an order of magnitude difference in the sensitivities of element 4 to moderated versus unmoderated neutrons. It is known that for any

albedo dosemeter to give accurate results, the neutron field must be characterized by other methods, e.g. Bonner Sphere spectroscopy or 9 to 3 ratio. To further test the algorithms in more rigorous mixed field conditions, mathematical calculations were made using average element responses. Four radiation sources (x-ray, beta, neutron, and high energy photon) at various mixture ratios of up to 5:1:1:1 were simulated. Since these are single calculated exposures, the tolerance level could not be calculated. The performance quotient was used as a measure of the accuracy of the algorithms and the results are shown in Table 6. The new algorithm is also superior to the conventional algorithm in these simulations, especially in the cases where there is a significant 137 Cs dose. This is due to the fact that the conventional algorithm underestimated the photon energy, thus underestimating the dose. The Cu filter is not thick enough to differentiate photons above 120 kev. In the new algorithm, the Cd filtered element, combined with the Kp methodology, properly identified the photon energy. The dosemeter and new algorithm were designed for mixed field use. While standard ANSI testing methods only use mixtures of two fields at a maximum 3/1 dose ratio, the algorithm gave superior results in actual 3 field mixtures as well as simulated 4 field mixtures at dose ratios up to 5/1. A dosemeter with approximately 300 mg/cm 2 over element 2 could be used to account for the beta energy issue. The algorithm would have to be modified somewhat to account for the additional beta response, but the basic concept of the high temperature peak method would not change. A thicker Cu filter could be used to achieve better photon discrimination. It would also be useful to test the dosemeter and algorithm under actual field conditions. Laboratory testing is idealized, and not subject to the variables such as geometry and ratios of mixtures that are encountered in the field. References 1. Liu, J. C., Sims, C. S., and Ahmed, A. B. A Proposed Four-Element Neutron-Photon-Beta Thermoluminescence Dosimeter. Health Phys. 63(3) 316-323 (1992). 2. Health Physics Society, HPS N13.11 American National Standard for Dosimetry-Personnel Dosimetry Performance-Criteria for Testing (1993).

3. Liu, J. C. and Sims, C. S. Mixed Field Personnel Dosimeter, Part I: The High Temperature Peak Characteristics of the Reader-Annealed TLD-600. Stanford CA: Stanford Linear Accelerator Center SLAC-Pub- 5340 (1991). 4. Chris Soares, National Institute of Standards and Technology, private communication. 5. Horowitz, Y.S., Hirning, C.R., Yuen, P., and Aikens, M. Beta Ray Spectroscopy Based On A Plastic Scintillation Detector/Silicon Surface Barrier Detector Coincidence Telescope. Nucl. Instr. and Meth. in Phys. Res. A 338 522-533 (1994). 6. Moscovitch, M. Dose Algorithms For Personnel Thermoluminescence Dosimetry Radiat. Prot. Dosim. Vol 47 No 1/4 373-380 (1993).

Table 1. Dosemeter design showing elements and filtration. Position Material Thickness (mm) Original Filtration (mg.cm -2 ) Added Filtration (mg.cm -2 ) Total Filtration (mg.cm -2 ) 1 TLD-700 0.38 242 ABS + 91 Cu 567 Acrylic 900 2 TLD-700 0.38 106 ABS + 894 ---- 1000 PTFE 3 TLD-700 0.15 17 Mylar ---- 17 4 TLD-600 0.38 300 ABS 713 Cd 1013 Table 2. Values used as constants in equations 1-6. New Algorithm Conventional Algorithm S4n (mod Cf-252) 34.82 nc/sv 29.61 nc/sv S4n (bare Cf-252) 4.18 nc/sv 3.54 nc/sv S3b (Tl-204) 0.78 nc/sv 0.78 nc/sv S3b (Sr-90/Y-90) 1.57 nc/sv 1.57 nc/sv Kn 0.146 Table 3. Tolerance levels for all tested fields. Tolerance Level New Algorithm Conventional Algorithm Sum Average Standard Deviation Range 4.03 3.20 0.076 0.060 0.042 0.044 0.01-0.17 0.00-0.17 4.26 3.53 0.080 0.067 0.053 0.051 0.02-0.30 0.00-0.19

Table 4. Tolerance levels for neutron dose estimation. Tolerance Level New Algorithm Conventional Algorithm Sum 1.73 3.15 Average 0.10 0.17 Standard Deviation 0.06 0.16 Range 0.02-0.23 0.04-0.56 Table 5. Tolerance levels for all tested fields using an average beta sensitivity. Tolerance Level New Algorithm Conventional Algorithm Sum Average Standard Deviation Range 7.01 3.20 0.132 0.060 0.081 0.044 0.01-0.35 0.00-0.17 7.19 3.53 0.136 0.067 0.085 0.051 0.02-0.35 0.00-0.19

Table 6. Performance quotients for the new and conventional algorithm from simulated irradiation ratios for,, neutron and beta dose. Mixture Algorithm s d n b M30/ 204 Tl/ 252 Cf(D 2 O)/ 252 Cf/ 137 Cs 1/1/1/0/1 new 0.00 0.00-0.11 0.00 conv 0.00-0.14-0.15 0.00 1/1/0/1/1 new 0.00 0.00-0.04 0.00 conv 0.00-0.14-0.03 0.00 3/1/1/0/1 new 0.00 0.00-0.11 0.00 conv 0.00-0.07-0.15 0.00 1/3/1/0/1 new 0.00 0.00-0.11 0.00 conv 0.00-0.14-0.15 0.00 1/1/3/0/1 new 0.00 0.00-0.11 0.00 conv 0.00-0.14-0.15 0.00 1/1/1/0/3 new 0.00 0.00-0.11 0.00 conv 0.00-0.20-0.15 0.00 5/1/1/0/1 new 0.00 0.00-0.11 0.00 conv 0.00-0.05-0.15 0.00 1/5/1/0/1 new 0.00 0.00-0.11 0.00 conv 0.00-0.14-0.15 0.00 1/1/5/0/1 new 0.00 0.00-0.11 0.00 conv 0.00-0.14-0.15 0.00 1/1/1/0/5 new 0.00 0.00-0.11 0.00 conv 0.00-0.21-0.15 0.00 3/1/0/1/1 new 0.00 0.00-0.04 0.00 conv 0.00-0.07-0.03 0.00 1/3/0/1/1 new 0.00 0.00-0.04 0.00 conv 0.00-0.14-0.03 0.00 1/1/0/3/1 new 0.00 0.00-0.04 0.00 conv 0.00-0.14-0.03 0.00 1/1/0/1/3 new 0.00 0.00-0.04 0.00 conv 0.00-0.20-0.03 0.00 5/1/0/1/1 new 0.00 0.00-0.04 0.00 conv 0.00-0.05-0.03 0.00 1/5/0/1/1 new 0.00 0.00-0.04 0.00 conv 0.00-0.14-0.03 0.00 1/1/0/5/1 new 0.00 0.00-0.04 0.00 conv 0.00-0.14-0.03 0.00 1/1/0/1/5 new 0.00 0.00-0.04 0.00 conv 0.00-0.21-0.03 0.00