SILICON DRIFT DETECTORS FOR HIGH RESOLUTION, HIGH COUNT RATE X-RAY SPECTROSCOPY AT ROOM TEMPERATURE

Similar documents
Semiconductor Drift Detectors: Applications and New Devices

Detectors for High Resolution Gamma-ray Imaging Based on a Single CsI(Tl) Scintillator Coupled to an Array of Silicon Drift Detectors

VORTEX A NEW HIGH PERFORMANCE SILICON DRIFT DETECTOR FOR XRD AND XRF APPLICATIONS

Analysis of Background Events in Silicon Drift Detectors

Gamma-ray Spectroscopy with LaBr 3 :Ce Scintillator Readout by a Silicon Drift Detector

Semiconductor X-Ray Detectors. Tobias Eggert Ketek GmbH

Nuclear Instruments and Methods in Physics Research A

NEW X-RAY DETECTORS FOR XRF ANALYSIS. Jan S. Iwanczyk & Bradley E. Patt Photon Imaging, Inc., Northridge, CA 91324

Silicon Drift Detectors: Understanding the Advantages for EDS Microanalysis. Patrick Camus, PhD Applications Scientist March 18, 2010

A monolithic array of silicon drift detectors coupled to a single scintillator for γ-ray imaging with sub-millimeter position resolution

SDDs for Science Applications

Diffraction: spreading of waves around obstacles (EM waves, matter, or sound) Interference: the interaction of waves

Time-Resolved μ-xrf and Elemental Mapping of Biological Materials

BENEFITS OF IMPROVED RESOLUTION FOR EDXRF

Silicon Controlled Drift Detectors for Imaging Applications

Praktikum zur. Materialanalytik

Review of Semiconductor Drift Detectors

AEROSOL FILTER ANALYSIS USING POLARIZED OPTICS EDXRF WITH THIN FILM FP METHOD

LASER-COMPTON SCATTERING AS A POTENTIAL BRIGHT X-RAY SOURCE

XRF books: Analytical Chemistry, Kellner/Mermet/Otto/etc. 3 rd year XRF Spectroscopy Dr. Alan Ryder (R222, Physical Chemistry) 2 lectures:

NEW CORRECTION PROCEDURE FOR X-RAY SPECTROSCOPIC FLUORESCENCE DATA: SIMULATIONS AND EXPERIMENT

Key words: avalanche photodiode, soft X-ray detector, scintillation γ-ray detector, imaging device PACS: 07.85;95.55.A;85.60.D

CHARACTERIZING PROCESS SEMICONDUCTOR THIN FILMS WITH A CONFOCAL MICRO X-RAY FLUORESCENCE MICROSCOPE

PERFORMANCE OF A ROOM TEMPERATURE GAS PROPORTIONAL SCINTILLATION COUNTER IN X-RAY ANALYSIS OF METALLIC ALLOYS EXCITED WITH ALPHA PARTICLES

Silicon Drift Detectors for gamma-ray detection: 15 years of research (and collaboration between Politecnico and INAF-Milano)

Outline. Introduction, motivation Readout electronics, Peltier cooling Input J-FETsJ

DEVELOPMENT OF A NEW POSITRON LIFETIME SPECTROSCOPY TECHNIQUE FOR DEFECT CHARACTERIZATION IN THICK MATERIALS

ELECTRIC FIELD INFLUENCE ON EMISSION OF CHARACTERISTIC X-RAY FROM Al 2 O 3 TARGETS BOMBARDED BY SLOW Xe + IONS

SEM. Chemical Analysis in the. Elastic and Inelastic scattering. Chemical analysis in the SEM. Chemical analysis in the SEM

Radiation Effects nm Si 3 N 4

GEM: A new concept for electron amplification in gas detectors

Controlled Si-Drift Detectors

Energetic particles and their detection in situ (particle detectors) Part II. George Gloeckler

Special SLS Symposium on Detectors

Lecture 8. Detectors for Ionizing Particles

Lessons learned from Bright Pixels and the Internal Background of the EPIC pn-ccd Camera

Gamma and X-Ray Detection

Development of New MicroStrip Gas Chambers for X-ray Applications

arxiv:physics/ v2 [physics.ins-det] 18 Jul 2000

Performance of high pressure Xe/TMA in GEMs for neutron and X-ray detection

1842 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 5, OCTOBER M. Marisaldi, C. Labanti, H. Soltau, C. Fiorini, A. Longoni, and F.

ABNORMAL X-RAY EMISSION FROM INSULATORS BOMBARDED WITH LOW ENERGY IONS

X-Ray Radiation Channeling through Micro-Channel Plates: spectroscopy with a Synchrotron Radiation Beam

Amonolithic X/gamma-ray detector with an energy range

InAs avalanche photodiodes as X-ray detectors

RADIOACTIVE SAMPLE EFFECTS ON EDXRF SPECTRA

RADIOISOTOPE RH-101 AS X-RAY SOURCE FOR INSTRUMENTS ON SPACE MISSIONS

Chem 481 Lecture Material 3/20/09

Advances in the Micro-Hole & Strip Plate gaseous detector

FUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON EXCITATION

Semiconductor-Detectors

A COMPACT X-RAY SPECTROMETER WITH MULTI-CAPILLARY X-RAY LENS AND FLAT CRYSTALS

DEPFET sensors development for the Pixel Detector of BELLE II

X-ray spectrometry with Peltier-cooled large area avalanche photodiodes

X-ray induced radiation damage in segmented p + n silicon sensors

MT Electron microscopy Scanning electron microscopy and electron probe microanalysis

MONTE-CARLO MODELING OF SILICON X-RAY DETECTORS

Peter Fischer, ziti, Universität Heidelberg. Silicon Detectors & Readout Electronics

TRACE ELEMENT ANALYSIS USING A BENCHTOP TXRF- SPECTROMETER

Portable type TXRF analyzer: Ourstex 200TX

normalized occurence

Spectroscopy on Mars!

Solid State Detectors

CHEM-E5225 :Electron Microscopy X-Ray Spectrometry

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY

Semiconductor Detectors

IMPROVEMENT OF DETECTION LIMITS OF A PORTABLE TXRF BY REDUCING ELECTRICAL NOISE

Position Sensitive Germanium Detectors for the Advanced Compton Telescope

GLANCING INCIDENCE XRF FOR THE ANALYSIS OF EARLY CHINESE BRONZE MIRRORS

ULTRATHIN LAYER DEPOSITIONS A NEW TYPE OF REFERENCE SAMPLES FOR HIGH PERFORMANCE XRF ANALYSIS

Peter L Warren, Pamela Y Shadforth ICI Technology, Wilton, Middlesbrough, U.K.

Chemical Analysis in TEM: XEDS, EELS and EFTEM. HRTEM PhD course Lecture 5

STRESS ANALYSIS USING BREMSSTRAHLUNG RADIATION

EEE4106Z Radiation Interactions & Detection

The quantum e$ciency of pn-detectors from the near infrared to the soft X-ray region

Development of a Radiation Hard CMOS Monolithic Pixel Sensor

Development of High-Z Semiconductor Detectors and Their Applications to X-ray/gamma-ray Astronomy

08 - Miscellaneous and historical detectors

arxiv: v1 [physics.ins-det] 12 Jan 2016

Development and characterization of 3D semiconductor X-rays detectors for medical imaging

Advances in Field-Portable XRF

Position sensitive detection of thermal neutrons with solid state detectors (Gd Si planar detectors)

EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors

Lecture 2. Introduction to semiconductors Structures and characteristics in semiconductors. Fabrication of semiconductor sensor

Horst Ebel, Robert Svagera, Christian Hager, Maria F.Ebel, Christian Eisenmenger-Sittner, Johann Wernisch, and Michael Mantler

New perspectives in X-ray detection of concealed illicit materials brought by CdTe/CdZnTe spectrometric detectors

A t XFEL experiment, sensors should have. Capacitance calculations in p + n silicon pixel sensors using three dimensional TCAD simulation approach

GEM at CERN. Leszek Ropelewski CERN PH-DT2 DT2-ST & TOTEM

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Metallic magnetic calorimeters. Andreas Fleischmann Heidelberg University

PoS(TIPP2014)033. Upgrade of MEG Liquid Xenon Calorimeter. Ryu SAWADA. ICEPP, the University of Tokyo

JRA3: Technology development for high-time-resolution astronomy

Bolometry. H. Kroegler Assciazione Euratom-ENEA sulla Fusione, Frascati (Italy)

Development of Silicon Drift Detectors and recent applications*

Classification of Solids

Ground Calibration of X-ray CCD Detectors with Charge Injection for the X-ray Imaging Spectrometer on Astro-E2

Application of large-area avalanche photodiodes to energy-dispersive x-ray fluorescence analysis

physics/ Sep 1997

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

PERFORMANCE OF A BOREHOLE XRF SPECTROMETER FOR PLANETARY EXPLORATION

Transcription:

Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 53 SILICON DRIFT DETECTORS FOR HIGH RESOLUTION, HIGH COUNT RATE X-RAY SPECTROSCOPY AT ROOM TEMPERATURE P. Lechner* 1, C. Fiorini**, A. Longoni**, G. Lutz* 2, A. Pahlke* 3, H. Soltau* 1, L. Strüder* 3 * MPI Halbleiterlabor, Otto-Hahn-Ring 6, D-81739 München, Germany 1 PNSensor GmbH; 2 MPI für Physik; 3 MPI für extraterrestrische Physik ** Politecnico di Milano, Piazza L. da Vinci 32, I-20133 Milano, Italy ABSTRACT An advanced Silicon Drift Detector (SDD) for X-ray spectroscopy designed and fabricated at the MPI semiconductor laboratory has the input transistor of the amplifying electronics integrated on the detector chip. That way the total capacitance of the detector/amplifier system is very small (200 ff) guaranteeing excellent energy resolution (147 ev FWHM at 5.9 kev, 10 mm², -10 C) and extreme count rate capability (up to 10 6 incoming photons per sec). The elaborate process technology stands for a low leakage current level allowing operation at or close to room temperature. The new design of the Silicon Drift Detector Droplet (SD 3 ) offers an energy resolution of 128 ev (FWHM at 5.9 kev, 5 mm², -10 C) that conventional detectors only obtain at liquid nitrogen cooling. The concept of Multichannel SDDs provides large sensitive areas up to cm² without losing the energy resolution and count rate properties of the single SDD cells. INTRODUCTION The Silicon Drift Detector (SDD) is derived from the principle of sideward depletion [1, 2]. A volume of high-resistivity n-type silicon is fully depleted by reverse biased p + junctions covering both surfaces. Electrons generated within the depleted volume by thermal processes or by the absorption of ionizing radiation are forced to drift to a small-sized n + anode by an electric field with a strong component parallel to the wafer surface. In the original SDD design the electric field is achieved by segmentation of the p + regions on both surfaces to patterns of parallel strips and superposition of a voltage gradient [1]. The direction of the voltage gradient is such that the n + readout anode has the most positive potential and collects all electrons released within the depleted volume by the absorption of ionizing radiation or thermal generation. Originally SDDs have been designed and used as position sensitive detectors in particle physics where the measurement of the signal electrons drift time allows the reconstruction of one coordinate of the particle s interaction point. The second coordinate is given by an appropriate segmentation of the collecting anode(s). The SDD concept is of great flexibility in the choice of anode arrangements and drift directions. For instance two-dimensional (x, y)-position sensors with an active area of 4.2 x 3.6 cm 2 have been fabricated [3] as well as cylindrical (r, ϕ)-detectors with a diameter of 10 cm and an angular resolution of 1, realized by a radial drift field and 360 anodes placed along the edge of a 4-inch wafer [4].

This document was presented at the Denver X-ray Conference (DXC) on Applications of X-ray Analysis. Sponsored by the International Centre for Diffraction Data (ICDD). This document is provided by ICDD in cooperation with the authors and presenters of the DXC for the express purpose of educating the scientific community. All copyrights for the document are retained by ICDD. Usage is restricted for the purposes of education and scientific research. DXC Website www.dxcicdd.com ICDD Website - www.icdd.com

Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 54 SILICON DRIFT DETECTORS FOR X-RAY SPECTROSCOPY To optimize the SDD principle for X-ray spectroscopy the strip system on both surfaces is replaced by a large area p + junction on the side opposite to the n + anode. This non-structured p + junction is used as a homogeneous, thin radiation entrance window [5]. In addition backside illumination has a self-shielding effect: the radiation sensitive components, e.g. the readout structure, are placed on the non-irradiated surface and can only be hit by hard X-rays (> 10 kev), whose intensity is reduced by the absorption of the silicon bulk. A further improvement is the design of circular, concentric drift electrodes driving signal electrons to the small sized n + anode in the center of the device. The potential values of the drift rings are defined by an integrated voltage divider, i.e. only the inmost and the outmost rings must be biased externally [6]. With the SDD topology and the collecting anode s small physical size the detector contribution to the total readout capacitance is already minimized. An additional reduction is achieved by the integration of the first transistor of the amplifying electronics on the detector chip [7]. The integrated transistor is a single-sided n-jfet [8, 9] designed to operate on a fully depleted n-type bulk. It is placed in the center of the ring shaped collecting anode (figure 1). The anode is connected to the floating FET gate by a narrow metal strip, so that the change of the anode voltage caused by signal electrons can be easily measured as a modulation of the transistor current. The on-chip FET has an internal, self-adapting discharging mechanism [10], so that there is no need for an externally clocked reset pulse, and the whole system is working with dcvoltages only. The integration of the first FET not only minimizes the overall capacitance recommending the SDD for high-resolution, high-rate spectroscopy, but also makes it insensitive to a great extent against electronic pickup and microphony, i.e. noise induced by mechanical vibration. integrated FET S G D anode field strips V Figure 1 p+ n- silicon back contact path of electrons Circular SDD with uniform radiation entrance window. The entire thickness of the device is sensitive to radiation. The first FET of the amplifying electronics is integrated in the center of the ring-shaped anode. -150 back contact electron potential [V] -100-50 0 300 field strips Figure 2 Calculated distribution of the potential energy in an SDD. The simulation applies to the cross section of a detector as shown in figure 1. The arrow lines indicate the signal electrons drift directions. depth [µm] 200 100 anode 0 radius [mm] -1.0 0.0 1.0

Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 55 300 FWHM [ev] @ 5.9 kev, -15 C 280 260 240 220 200 180 160 shaping time 450 nsec 340 nsec 210 nsec 100 nsec 70 nsec 140 0 200 400 600 800 1000 incoming rate [kcps] Figure 3 Measured count rate dependent energy resolution of an SDD with analog pulse processing system [11] at a temperature of -15 C (5 F). With the shortest shaping time of 70 nsec the system is still able to operate at 10 6 incoming photons per sec with a throughput of 4 10 5 processed counts per sec. Figure 1 shows the topology of an SDD with cylindrical geometry, uniform entrance window, and integrated FET. Figure 2 shows the corresponding calculated potential energy distribution. SDDs of the type shown in figure 1 have been fabricated and qualified at the semiconductor lab of Max-Planck-Institute in Munich/Germany in large quantities in different shapes and sizes. Due to the advanced process technology used in the SDD fabrication the leakage current level is so low, typically 100 pa/cm 2 at full depletion of the 300 µm thick wafer, that the devices can be operated at room temperature or with moderate cooling by a single-stage thermoelectric cooler. The numbers quoted in the following reflect the typical performance of an SDD with a sensitive area of 10 mm 2 operated at a temperature of -10 C (14 F). The energy resolution in terms of full-width-at half-maximum (FWHM) at the Mn-K α line (5.9 kev) is 147 ev. Because of the small value of the total output capacitance of about 200 ff the SDD is operated at pulse shaping time constants of the order of 100 nsec (figure 3), whereas conventional systems with comparable spectroscopic quality require longer shaping times by about a factor of 100. This means that the SDD can be used at extremely high count rates far beyond the ability of other systems. Figure 3 demonstrates the count rate capability obtained with an analog pulse processing system with variable shaping time constant [11]. With a shaping time of 70 nsec the energy resolution is still 250 ev (FWHM at 5.9 kev) at 10 6 incoming photons per sec and a throughput of 4 10 5 processed counts per sec. For high energy resolution at lower count rates longer shaping time constants up to 450 nsec are selected. To prove their radiation hardness SDDs have been irradiated with a high flux of Mo-K α photons (17.5 kev). Photons of this energy have a maximum probability of transmission through the 300 µm thick silicon bulk and absorption in the radiation sensitive components (readout structure, integrated voltage divider) on the opposite surface. Irradiated devices reveal no change in energy resolution up to 10 13 absorbed photons [12]. At higher doses an increase of the leakage current is observed. That means that the SDD can be operated under worst case conditions for 30 continuous years at a constant rate of 10 4 photons per sec without any radiation damage effect. The radiation entrance window has been optimized for the detection of soft X-rays [13, 14]. The quantum efficiency is above 90% in the energy range from 300 ev up to 10 kev. On the low energy end the quantum efficiency is a function of the entrance window technology, on the high energy end it is given by the full depletion of the 300 µm thick silicon bulk. The quality of the entrance window is also reflected in the low energy background of SDD spectra (figure 5). The

Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 56 peak/background ratio defined as amplitude ratio of the Mn-Kα line relative to the average background at 1 kev is 3000. SDDs are found in a variety of scientific experiments and commercial systems. With their count rate capability they set new boundary conditions for electron microprobe analysis [15]. The fact that SDDs don t require expensive and inconvenient cooling by liquid nitrogen initiated new applications in X-ray spectroscopy. In combination with a micro-focus X-ray tube the SDD makes up a compact, portable spectrometer for XRF measurements in the field, i.e. independent of laboratory infrastructure, developed for the use in archeometry [16]. The ESA comet lander mission ROSETTA as well as NASA s 2003 Mars Exploration Rovers are equipped with an Alpha-Proton-X-ray-Spectrometer (APXS) including an SDD for PIXE analysis [17]. SD³ - A NEW SDD LAYOUT In the new Silicon Drift Detector Droplet (SD³) the readout structure, i.e. the collecting anode and the integrated FET, is no more placed in the center but at the edge of the sensitive area, where it can be easily shielded from direct irradiation by a circular collimator. The electric field that drives electrons towards the anode is generated by bow-shaped drift electrodes (figure 4). Compared to the standard circular device the SD³ layout is of great benefit in two respects [12]: Direct irradiation of the readout structure results in a partial loss of signal electrons to the integrated FET. Although the FET is small in area the effect is a significant contribution to the low energy background in SDD spectra. Therefore, with the SD³ geometry the peak/background ratio is increased from 3000 to 5000 using a collimator with = 2 mm (figure 5). The fact that in the SD³ geometry the signal electrons approach the anode from a given direction instead from all sides allows one to design the anode with different shape and size resulting in a reduction of the total output capacitance from 200 ff to 120 ff. The measured energy resolution of a 5 mm² SD³ is 128 ev (FWHM at 5.9 kev) at a detector temperature of - 10 C (14 F) and 124 ev at -20 C (-4 F). So the SD³ cooled by a thermoelectric Peltier element has a resolution superior to the best values reported for liquid nitrogen cooled Si(Li) detectors. Figure 4 Topology of the Silicon Drift Detector Droplet (SD³) with a sensitive area of 5 mm². The readout structure (black region) is placed at the edge of the device where it can be shielded from direct irradiation by a circular collimator. The arrow lines indicate the signal electrons drift directions.

Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 57 normalized counts 1 0.1 0.01 0.001 0.0001 circular SDD SD³ 0 2000 4000 6000 8000 energy [ev] Figure 5 Comparison of 55 Fe spectra recorded with two 5 mm² SDDs in conventional, circular geometry and in the new SD³ design, using a collimator with = 2 mm. The SD³ obtains an improvement in peak/background ratio (3000 5000) as well as in energy resolution (140 ev 128 ev FWHM at 5.9 kev, -10 C). MULTICHANNEL DRIFT DETECTORS To provide a large sensitive area without losing the energy resolution and count rate capability of the single SDD cell the concept of Multichannel Drift Detectors has been introduced [18]. A Multichannel SDD is a continuous, gapless arrangement of a number of SDDs with individual readout, but with common voltage supply, entrance window, and guard ring structure (figure 6). It allows not only to fill any large area (figure 6a), but it also implies the choice of any twodimensional shape according to the requirements of the experiment like a linear chain for diffractometry (figure 6b) or a closed ring (figure 6c). Multichannel SDDs are used at synchrotron light facilities as they are able to cope with the extremely high photon rates in fast counting applications like EXAFS [19] and X-ray holography [20]. The ring structure is the basic component of a compact XRF spectrometer: a sample is excited through an X-ray polycapillary fiber inserted through a laser-cut hole in the center of the detector chip, and the SDD-ring receives the fluorescence photons covering a large fraction of the solid angle around the sample (figure 7). For the readout of Multichannel SDDs the amplifier chip ROTOR (rotational trapezoidal readout) based on JFET-CMOS technology has been developed [21]. ROTOR is able to handle the random asynchronous event occurrence of a large number of SDDs with low read noise at count rates exceeding 10 5 counts per sec per channel. Figure 6 Different shapes of Multichannel SDDs: a.) hexagon, 19 x 5 mm²; b.) chain, 6 x 5 mm²; c.) ring, 12 x 5 mm².

Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 58 Figure 7 sample ring detector X-ray source with capillary fiber Principle of a compact XRFspectrometer using the SDD ring structure of figure 6. The sample is excited through an X-ray polycapillary fiber inserted through a laser-cut hole in the center of the detector chip. The SDD-ring receives the fluorescence photons covering a large fraction of the solid angle around the sample. REFERENCES [1] Gatti, E.; Rehak, P., Nucl. Instr. & Meth. A 225, 1984, 608-621 [2] Lechner, P.; Hartmann, R.; Holl, P.; et al., State-of-the-art silicon detectors for X-ray spectroscopy, these proceedings [3] Gatti, E.; Longoni, A.; Sampietro, M.; et al., Nucl. Instr. & Meth. A 273, 1988, 865-868 [4] Holl, P.; Rehak, P.; Ceretto, F.; et al., Nucl. Instr. & Meth. A 377, 1996, 367-374 [5] Kemmer, J.; Lutz, G.; Belau, E.; et al., Nucl. Instr. & Meth. A 253, 1987, 378-381 [6] Lechner, P.; Andricek, L.; Findeis, N.; et al., Nucl. Instr. & Meth. A 326, 1993, 284-289 [7] Lechner, P.; Eckbauer, S.; Hartmann, R.; et al., Nucl. Instr. & Meth. A 377, 1996, 346-351 [8] Radeka, V.; Rehak, P.; Rescia, S.; et al., IEEE El. Dev. Lett. 10 (2), 1989, 91-95 [9] Pinotti, E.; Bräuninger, H.; Findeis, N.; et al., Nucl. Instr. & Meth. A 326, 1993, 85-92 [10] Fiorini, C.; Lechner, P., IEEE Trans. Nucl. Sci. Vol. 46, No. 3, 1999, 761-764 [11] RÖNTEC GmbH, Berlin, Germany, X-flash product information, 1997 [11] Lechner, P.; Pahlke, A.; Soltau, H, accepted for publication in X-ray Spectrometry, 2003 [13] Hartmann, R.; Hauff, D.; Lechner, P.; et al., Nucl. Instr. & Meth. A 377, 1996, 191-196 [14] Hartmann, R.; Strüder, L.; Kemmer, J.; et al., Nucl. Instr. & Meth. A 387, 1997, 250-254 [15] Strüder, L.; Meidinger, N.; Stötter, D.; et al., Microsc. Microanal. 4, 1999, 622-631 [16] Longoni, A., Fiorini, C., Leutenegger, P.; et al., Nucl. Instr. & Meth. A 409, 1998, 395-400 [17] Rieder R.; Gellert, R.; Brückner J.; et al., The new Athena Alpha Particle X-Ray Spectrometer (APXS) for the Mars Exploration Rovers, accepted for publication in Journal of Geophysical Research - Planets, 2003 [18] Lechner, P.; Buttler, W.; Fiorini, C., et al. Proceedings of SPIE Vol. 4012, 2000, 592-599 [19] Gauthier, C.; Goulon, J.; Moguiline, E., et al. Nucl. Instr. & Meth. A 382, 1996, 524-532 [20] Adams, B.; Hiort, T.; Kossel, E.; et al., phys. stat. sol. (b) 215, 1999, 757-771 [21] Pullia, A.; Fiorini, C.; Gatti, E., IEEE Trans. Nucl. Sci., Vol. 45, 1998, 3179-3185