Commercialization of autonomous sensor systems for quantifying pco 2 and total inorganic carbon

Similar documents
INTRODUCTION TO CO2 MEASUREMENTS IN SEA WATER

DATA PRODUCT SPECIFICATION FOR PARTIAL PRESSURE OF CO 2 IN AIR AND SURFACE SEAWATER

PURIFICATION AND CHARACTERIZATION OF BROMOCRESOL PURPLE FOR SPECTROPHOTOMETRIC SEAWATER ALKALINITY TITRATIONS

Compact Hydrogen Peroxide Sensor for Sterilization Cycle Monitoring

ActiPix SDI300 Surface Dissolution Imaging System

INTRODUCTION TO CO2 CHEMISTRY

INTRODUCTION TO CO2 CHEMISTRY

Supporting Information for. Suppression of OH Generation from the Photo-Fenton Reaction in the Presence of α-pinene Secondary Organic Aerosol Material

ChromTech HPLC System

Chemometric Approach to the Calibration of Light Emitting Diode Based Optical Gas Sensors Using High-Resolution Transmission Molecular Absorption Data

Detection of trace contamination on metal surfaces using the handheld Agilent 4100 ExoScan FTIR

Nucleic Acid Quantitation in Microplates

HOW ADVANCED PYROMETERS INCREASE THERMAL PROCESS REPEATABILITY AND PRODUCT QUALITY

MICROFLUIDIC ANALYSER FOR ph IN WATER AND WASTEWATER

Temperature control for Varian Cary line of UV/Vis Spectrophotometers and the Eclipse Fluorometer

FIRST FIELD MEASUREMENTS AND DEVICE IMPROVEMENTS FOR MULTI SPECTRAL VOLUME SCATTERING METER (MVSM)

High Performance Liquid Chromatography

High Performance Liquid Chromatography

The investigation of the photokinetics of a platinum organoamine complex using the Cary 50/60

Gas exchange of CO 2 and O 2 in partially ice-covered regions of the Arctic Ocean investigated using in situ sensors

Advantages in performance of the RBR conductivity channel with Delrin TM /ceramic inductive cell.

- Acid-base equilibria - ph concept and scales - ph measurements - ph sensors, types - glass electrodes - Spectrophotometric sensors - Solid

SUPPORTING INFORMATION FOR. Spectrophotometric measurements of the carbonate ion concentration: aragonite

Reprint. Project Summary. Side by Side Evaluation of ChemScan UV-4100 and Hach (Dr. Lange) Optiquant at Littleton-Englewood Colorado

CONFOCHECK. Innovation with Integrity. Infrared Protein Analysis FT-IR

ph electrodes: what users should know

ONYX -MCE MULTI-CHANNEL OPTICAL FIBER PYROMETERS WITH ACTIVE EMISSIVITY COMPENSATION PRECISION TEMPERATURE MEASUREMENT FOR DEMANDING INDUSTRIAL

Multi-element process analyzer

Monitoring Nitrogen Trichloride in Chlorine Manufacture. solvent. O NaOH + 1 /2H 2

Precision and accuracy of protein size determination using the ActiPix TDA200 Nano-Sizing System

7. Timescale for hygroscopic conversion of calcite mineral particles through heterogeneous reaction with nitric acid

Magnetic Property Measurement System

Evaluation of spectrophotometric ph methods for freshwater measurements

A New Hyperspectral Spherical-Cavity Absorption Meter

Diquat 1,1 -ethylene-2,2 -bipyridium dibromide salt Paraquat 1,1 -dimethyl-4,4 -bipyridium dichloride salt Initial Preparation

Whitepaper. How to Measure Conductivity over a Wide Range and at Ultra-low Levels

Determination of the ph of sea water using the indicator dye m-cresol purple

DI 3500 Discrete Aoalvzer

Operating Instructions for Ammonium ISE Specifications

ELTECdata #138 Computing the Relationship of Pyroelectric Detector Output to Gas Concentration

Compact Knowledge: Absorbance Spectrophotometry. Flexible. Reliable. Personal.

Multiphase Oscillatory Flow Strategy for in Situ Measurement and Screening of Partition Coefficients

Thermal Resistance Measurement

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Pros and Cons of Water Analysis Methods

1 st European Union Science Olympiad in Dublin, Ireland. task B

Nepheloid Layer Measurements and Floc Model for OASIS

Ocean Bottom Seismometer Augmentation of the NPAL Philippine Sea Experiment

Optical density measurements automatically corrected to a 1-cm pathlength with PathCheck Technology

HPLC. High Performance Liquid Chromatography (HPLC) Harris Chapter 25

Evaluation of laser-induced breakdown spectroscopy (LIBS) as a new in situ chemical sensing technique for the deep ocean

Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste

S2 PICOFOX. Innovation with Integrity. Spectrometry Solutions TXRF

CHAPTER 19 PARACETAMOL + IBUPROFEN

Plop Plop, Fizz Fizz, Oh What A Relief It Is (Which Pain Reliever Works Fastest)

Ultraviolet-Visible and Infrared Spectrophotometry

The Use of Lagrangian Drifters to Measure Biogeochemical Processes and to Analyze Satellite Data Sets

On the possibility to create a prototype of laser system for space debris movement control on the basis of the 3-meter telescope.

Formaldehyde Fluorescent Detection Kit

Imagent for fnirs and EROS measurements

ISA QATAR : 90 th Seminar

Comparing the Microlab 600 to Volumetric Glassware and Air Displacement Pipettes. January 2013

Membrane Electrodes. Several types

Polymer analysis by GPC-SEC. Technical Note. Introduction

BioDrop DUO dsdna Application Note

The ROXI Colorimeter & Fluorimeter. Laboratory Application I. Colorimetric measurements via Beer s Law.

Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry

The Plus for your lab!

Exploiting the Interaction of Pyranine-3 with Poly(L-Lysine) to Mediate Nanoparticle Assembly: Fabrication of Dynamic ph-responsive Nanocontainers

Chapter content. Reference

Cork Institute of Technology. Summer 2005 Instrumental Analysis (Time: 3 Hours) Section A

Thick-wall, Liquid-Filled Quartz Capillaries for Scintillation and Wavelength Shifting Applications

BETTER DESIGN AND NEW TECHNOLOGIES IMPROVE LASER POWER MEASUREMENT INSTRUMENTATION

Supplementary Information. Back-Contacted Hybrid Organic-Inorganic Perovskite Solar Cells

*Correspondence to:

Determination of Total Bromine and Iodine Emission Spectrometric Method (ICP-OES) EuSalt/AS

Electronic Supplementary Information

Bayesian Methodology for Atmospheric Correction of PACE Ocean-Color Imagery

Experiment 1 (Part A): Plotting the Absorption Spectrum of Iron (II) Complex with 1,10- Phenanthroline

Backscattering and Polarization Properties of Marine Particles -- Instrument Development and Field Work

ISO-TECH SYSTEM, VARANASI Analysis by ph and ion selective electrodes

user manual blotting Hoefer PR648 Slot blot manifold um PR648-IM/Rev.F0/08-12

PROCEEDINGS OF SPIE. Shedding the light on spectrophotometry: the SpecUP educational spectrophotometer

Effects of Solvent Acidity on the Free-Radical-Initiated Synthesis of Methanesulfonic Acid from CH 4 and SO 3

Direct, Reliable, and Correct Oxygen Measurement

Burner Tubing Specification for the Turbulent Ethylene Non-Premixed Jet Flame

HARP Assessment of Uncertainty

Online Reaction Monitoring of In-Process Manufacturing Samples by UPLC

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive

Because light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency.

Chemistry Instrumental Analysis Lecture 23. Chem 4631

Self Assembly of Cyanine Dye on Clay Nanoparticles

Ultraviolet-Visible and Infrared Spectrophotometry

Spectroscopy in Transmission

2d-Laser Cantilever Anemometer

The Interaction Between Physics and Biogeochemistry in Nearshore Permeable Sediments

GENERAL INSTRUCTIONS GENERAL PREPARATION

Cyanine Dyes as Ratiometric Fluorescence Standards for the Far-Red Spectral Region

Rapid and precise calcium isotope ratio determinations using the Apex-ACM desolvating inlet system with sector-field ICP-MS in low resolution

Transcription:

Commercialization of autonomous sensor systems for quantifying pco 2 and total inorganic carbon PI: Michael DeGrandpre, PhD Dept. of Chemistry, University of Montana, Missoula, MT 59812 Phone: (406) 243-4227 Fax: (406) 243-4118 E-mail: Michael DeGrandpre@umontana.edu Co-PI: Jim Beck, MSME Sunburst Sensors, LLC, 1121 E. Broadway, Suite 114, Missoula, MT 59802 Phone: (406) 532-3246 Fax: (406) 543-2304 E-mail: jim@sunburstsensors.com Co-PI: Terry Hammar Woods Hole Oceanographic Institution, Woods Hole, MA 92543 Phone : (508) 289-2462 Fax: (508) 457-2189 E-mail: thammar@whoi.edu Co-PI: Andrew Dickson, PhD Scripps Institution of Oceanography, The University of California-San Diego 9500 Gilman Drive, La Jolla, CA 92093-0244 Phone: (858) 822-2990 Fax: (858) 822-2919 E-mail: adickson@ucsd.edu Award Number: OCE-052955 http://www.sunburstsensors.com OBJECTIVES This research, funded under 2004 s NOPP Topic 4B Sensors for Sustained, Autonomous Measurement of Chemical or Biological Parameters in the Ocean uses the NOPP funding to promote commercialization of the SAMI-CO 2, a sensor developed for autonomous measurements of the partial pressure of CO 2 (pco 2 ) in seawater. The SAMI-CO 2 was commercialized in 1999 through an exclusive license from the University of Montana to Sunburst Sensors, a company in Missoula, Montana (see sunburstsensors.com). Field deployments by DeGrandpre and others have demonstrated the excellent long-term stability predicted by the SAMI s well-understood theoretical response. The design, however, is complex and prone to failures, especially by customers who are not trained to operate the SAMI. Incremental changes in the design have improved reliability, but a full redesign is required to implement modern electronic and manufacturing technology. The new design will allow individual investigators to make pco 2 measurements reliably over long time periods in widespread ocean locations on many different ocean platforms.

PROGRESS Work has focused on completing the re-design. Two major hurdles have been to test the LED light source, and to isolate and eliminate the temporal drift of R CO2 to allow for varying sample intervals without re-calibration. LED LIGHT SOURCE The SAMI electronics were re-designed so that light emitting diodes (LED) can be used as the light source. A 621 nm LED (Lite-on Electronics, Inc.) and a 430-435 nm LED (Roithner Lasertechnik) were modulated at 512 Hz and 256 Hz, filtered with 620 and 435.8 nm bandpass filters (Intor), coupled to a 2 2 optical fiber bundle (Fiber Tech Optica) with 720-µm input bundles and 250-µm and 1000-µm output bundles that coupled to a reference photodiode and a black Delrin optical flow cell, respectively. Temperature dependence. Individual components of the LED SAMI were evaluated for temperature effects as follows. 1) The emission intensity and spectra were measured at temperatures between 5º and 35º C with the LEDs encased in an aluminum housing and immersed in a water bath; 2) intensity of light passing through the optical flow cell was measured at temperatures between 10º and 30ºC with the flow cell immersed in a water bath; and 3) measurements were taken with the LED SAMI electronics encased in the SAMI PVC housing which effectively thermostats the electronics over time periods <5 hours because of the time required for the air inside the housing to equilibrate to the temperature outside the housing. The conclusions from these experiments were that each component causes a temperature-dependent light signal. As shown in Figure 1, if the SAMI electronics are at the same temperature as the optical flow cell, the reference signal behaves similarly to the sample signals, and therefore subtracting the reference absorbance from the sample absorbance at each wavelength (indicated by corrected signals in Figure 1) largely removes the temperature dependence of the signal. However, when the electronics are enclosed in the SAMI housing, they remain at a relatively stable temperature while the flow cell is exposed to changing temperatures. In this case, the reference signal is very stable but the sample signal changes with temperature. The reference signal therefore cannot be used to correct for the temperature-dependence of the sample signal. The difference between the reference and sample temperatures could potentially be a problem in the field since the SAMI electronics must be enclosed in the housing to keep them dry. A possible solution is to use housing material that allows more rapid temperature equilibration. Absorbance precision. Absorbance precision was calculated as 1 standard deviation of the signals measured over 4 hours at a stable temperature. Precision was calculated from light signals measured every 10 seconds in a dry flow cell and every 5 minutes after pumping 50 µl of 5 10-5 mol kg -1 bromothymol blue (BTB) solution buffered at ph 7.4. As shown in Table 1, pumping the BTB solution caused about an order of magnitude increase in signal noise. However, even with the higher noise of the BTB solution the precision is close to that required to achieve an overall measurement precision of ±2 µatm pco 2, which is approximately 0.0016 AU at 436 nm and 0.0012 AU at 620 nm if signals at the two signals diverge; if the two signals covary the noise can be higher. LED SAMI calibration. The LED SAMI was immersed in a 20ºC water bath, molar absorptivities at 436 and 620 nm were measured, and the SAMI was calibrated 4 times. Three of the calibrations included 5 pco 2 levels, from ~200 600 µatm, while one calibration included

only 4 pco 2 levels, from ~200 450 µatm. Three of the calibrations started with the highest pco 2 level and ended at the lowest level, while one calibration started with the lowest level. Parts of the detection system were disassembled and re-assembled between the first and Figure 1. A) Raw and corrected (raw absorbance reference absorbance) absorbances from the sample signals, and B) absorbances from the reference signals, at 436 nm and 620 nm, for a solution of 5 10-5 mol kg -1 BTB solution, buffered at ph 7.4. Fifty microliters of BTB were pumped every 5 minutes and an absorbance reading was taken ~10 sec after the pump pulse. Table 1. Average signals and absorbance precision for the LED SAMI with a dry flow cell and with 5 10-5 mol kg -1 BTB pumping through the cell at 5-min intervals. Wavelength Gain Flow Cell Contents Average Signal Absorbance Precision 436 nm 2.4 10 7 air 22000 ± 6 0.00011 620 nm 2.4 10 7 air 38000 ± 5 0.00008 436 nm 9.4 10 6 BTB 5200 ± 21 0.0018 620 nm 9.4 10 6 BTB 4500 ± 14 0.0014

second calibrations. The LED SAMI calibration curves are compared to a theoretical calibration curve and 4 traditional SAMI calibration curves in Figure 2. All RCO 2 values were corrected to 19.98ºC. The LED SAMI curves are positively offset from the theoretical curve and show greater divergence from the theoretical curve at lower pco 2 than at higher pco 2. The difference between the LED SAMI and the traditional SAMI calibration curves could be due to different band width and center wavelength of the light sources. Figure 2. Comparison of calibration curves from 4 traditional SAMIs, 4 replicate calibrations from the LED SAMI, and a theoretical calibration curve. RCO 2 is calculated from the absorbances at 436 nm and 620 nm and the molar absorptivity values of the acid and base forms of BTB at the same wavelengths (DeGrandpre et al., 1995). R CO2 TEMPORAL DRIFT Recent studies have found that the responses of SAMI-CO 2 sensor (R CO2 ) differ significantly between the measurements based on 15-minute interval and those based on 60- minute interval. R CO2 is directly related to pco 2 of the sample solution and indicates how far away the indicator ph is from the indicator pk a. As shown in Figure 3, the difference between measurements at 15-minute and 60-minute intervals can be at least 0.02, which corresponds to 20 ppm CO 2. Further experiments have discovered that the acidic contamination from the flow cell material leads to higher R CO2 at a 60-minute interval. We have bubbled the same BTB indicator with the same concentration CO 2, and then measured R CO2 of the well-equilibrated indicator solution with a glass cuvet on a spectrophotometer. The response R CO2 was very stable (<0.005 for an hour) during the hourly measurement period. Several separate efforts have been made to minimize the acidic contamination from the cell material. First, NaCl was added to the BTB indicator to increase ionic strength and suppress ionic exchange between the cell surface and the indicator solution. Second, BTB indicator was dissolved in 10% methanol solution since methanol may help flush the contamination more efficiently. Third, the more soluble ph indicator Cresol Red (CR) replaced the current BTB

indicator to reduce the indicator adsorption on the cell surface. Fourth, the chemically inert material PTFE Teflon was used to make the flow cell. Fifth, both PTFE Teflon and white Delrin cells were silanized with SIGMACOTE (Sigma, Inc.) to suppress the ion exchange between the indicator and the OH group on the cell surface. Last, a Teflon AF-2400 based liquid core waveguide, which combines the membrane and flow cell into one unit due to its gas permeability and internal light reflection, was used on the test, as well. Unfortunately, none of the above efforts have ever eliminated the measurement errors between 15-minute and 60-minute intervals. Recently, we have changed the flow cell ID from 1000 μm to 600 μm for smaller internal volume and better flush efficiency. As shown in Figure 4, the difference has fallen within ±0.004 between 15-minute interval and 60-minute interval, corresponding to ~ 4 μatm error. Further efforts to eliminate the contamination problem are ongoing. 0.0900 0.0800 0.0700 0.0600 1 60 minutes ΔR CO2 0.0500 0.0400 5 0.0300 0.0200 0.0100 15 minutes 0.0000 0.00 0.10 0.20 0.30 0.40 0.50 0.60 0.70 0.80 0.90 Time (Days) Figure 3. Normalized responses of the SAMI-CO 2 sensor (ΔR CO2 at two different measurement intervals (15 and 60 minutes) with five measurements taken at each time interval (5 pump pulses). The flow cell is white Delrin Z- cell with 1000 μm optical pathlength. pco 2 is about 360 ppm and the silicone membrane is about 38" long with 0.02" ID and 0.037" OD. BTB indicator is 5.55 10-5 mol kg -1 with 8 10-5 mol kg -1 NaOH for the measurements. For each five measurements, R CO2 decreases from the first to the fifth measurement.

0.0900 0.0800 0.0700 0.0600 0.0500 ΔR CO2 0.0400 0.0300 15 minutes 0.0200 15 minutes 0.0100 60 minutes 0.0000 0.00 0.10 0.20 0.30 0.40 0.50 0.60 0.70 0.80 0.90 1.00 1.10 1.20 Time (Days) Figure 4. Normalized responses of the SAMI-CO 2 sensor (ΔR CO2 at two different measurement intervals (15 and 60 minutes) with one measurement plotted on the same scale as Figure 3. The flow cell is a white Delrin cell with 600 μm optical pathlength. pco 2 is about 360 ppm and the silicone membrane is about 22" long with 0.02" ID and 0.037" OD. BTB indicator is 5.6152 10-5 mol kg -1 with 8.047 10-5 mol kg -1 NaOH for the measurements (same as Figure 3 data). IMPACT AND APPLICATIONS Economic Development This research and development will help make the SAMI-CO 2 a robust, less-expensive means of measuring seawater pco 2. The high cost and complexity of the current design is a disincentive to sales, and widespread adoption within the oceanographic research community. This work may also be the springboard for moving the technology into other research areas, such as atmospheric CO 2 measurements and industrial applications.

References. DeGrandpre, M. D.; Hammar, T. R.; Smith, S. P.; Sayles, F. L. In situ measurements of seawater pco2. Limnology and Oceanography (1995), 40(5), 969-75.