Application Note. Abstract. Authors

Similar documents
Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS

Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Elemental analysis of river sediment using the Agilent 4200 MP-AES

China Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Analysis of domestic sludge using the Agilent 4200 MP-AES

Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode

Application. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid. Authors. Introduction. Abstract.

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

ICP-OES Application Note Number 35

High-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

On-line isotope dilution analysis with the 7700 Series ICP-MS: Analysis of trace elements in high matrix samples

Application of the Agilent 7900 ICP-MS with Method Automation function for the routine determination of trace metallic components in food CRMs

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS

Meeting the Challenges of Soil Analysis with the Avio 200 ICP-OES

Real World Analysis of Trace Metals in Drinking Water Using the Agilent 7500ce ICP-MS with Enhanced ORS Technology. Application. Authors.

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy

Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode

Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

Direct Measurement of Trace Metals in Edible Oils by 7500cx ICP-MS with Octopole Reaction System

Samples compliant with the WATER

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS

Application note. Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ

Performance of the Agilent 7900 ICP-MS with UHMI for high salt matrix analysis

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry

Agilent ICP-MS. Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis

Implementing EPA Method 6020 with the Agilent ICP-MS Portfolio

The Investigation of Fertilizer Analyses Using Microwave Digestion and the Agilent 720-ES

Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS)

Rapid Sample Assessment and Simplified Method Development with IntelliQuant

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS)

Following documents shall be used for reference on quantities, units, prefixes and other technical vocabulary in this document:

Elemental Scientific. Spike Recovery. Elemental Scientific

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP

Agilent 8900 Triple Quadrupole ICP-MS

Agilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology

OPTIMIZING METALS SAMPLE PREP

Accurate Quantification of Cadmium, Chromium, Mercury and Lead in Plastics for RoHS Compliance using the Agilent 7500ce ICP-MS

The Determination of Toxic, Trace, and Essential Elements in Food Matrices using THGA Coupled with Longitudinal Zeeman Background Correction

Rapid, simple, interference-free analysis of environmental samples using the XSERIES 2 ICP-MS with 3 rd generation CCT ED

Trace Analyses in Metal Matrices Using the ELAN DRC II

Determination of Elements at Sub-ppb Concentrations in Naphtha Mixtures Using the NexION 300 ICP-MS

Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES

Agilent ICP-MS Journal September 2009 Issue 39

hydrideicp Hydride Generation System

Analysis of trace elements in polymer samples to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Rapid determination of five arsenic species in polished rice using HPLC-ICP-MS

Elemental Scientific. Stability. Elemental Scientific

Agilent Technologies at TIAFT 2013

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

Automated Determination of PPQ Levels of Thorium in High Purity Copper

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

U.S. EPA SW-846 Method 6010C Using the Prodigy7 High- Dispersion ICP Introduction. Application Note - AN1305. Experimental

Validating performance of an Agilent 7800 ICP MS for USP <232>/<233> & ICH Q3D/Q2(R1): Elemental Impurity Analysis in Pharmaceuticals

Single particle analysis using the Agilent 7700x ICP-MS

OES - Optical Emission Spectrometer 2000

Using FIMS to Determine Mercury Content in Sewage Sludge, Sediment and Soil Samples

Hands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Fergus Keenan Shona McSheehy Julian Wills

The New Agilent 7700 Series ICP-MS

Analysis of Saponins in Spine Date Seed using Agilent ZORBAX Solvent Saver HT (1.8 µm) Columns

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Examples of Method Validation Studies Conducted in Different Economies

Application note. Determination of exchangeable cations in soil extracts using the Agilent 4100 Microwave Plasma-Atomic Emission Spectrometer

Analytical Method Determination of airborne metal concentrations by ICP-MS Sampling using Solu-Sert filter capsules

Transcription:

A Validated Method for Direct Analysis: 13 Trace Elements Determination in Tea Infusions using the Agilent 7700x ICP-MS with Integrated Sample Introduction System in Discrete Sampling (ISIS-DS) Application Note Authors Raquel F. Milani, Leandro I. Peron, Elisabete S. Saron, Marcelo A. Morgano Instituto de Tecnologia de Alimentos, Campinas SP, Brasil Fabio F. Silva Agilent Technologies, Brasil, São Paulo, SP, Brasil Abstract The Integrated Sample Introduction System Discrete Sampling (ISIS-DS) accessory increases the productivity of the Agilent 7700x ICP-MS, allowing direct introduction of complex samples with no interface damage. This application note describes a method for the determination of 13 elements in tea infusion by direct analysis, using an ISIS-DS system coupled to a 7700x ICP-MS operating in three different cell modes: No gas, He, and High Energy Helium (HEHe). The method shows high productivity, good stability and low detection limits. Solange Cadore Universidade Estadual de Campinas, Campinas, SP, Brasil

Introduction Tea, when prepared by infusion of leaves, flowers, or roots, has generated significant scientific interest due to its increased consumption, antioxidant activity, and presence of some micronutrients, such as minerals, flavonoids, and catechins [1]. Studies have shown a presence of potentially toxic and cumulative substances in herbs such as inorganic contaminants [2]. These contaminants originate from different aspects of the herbal tea manufacturing process and include sources such as soil and water, fertilizers, and airborne industrial emissions [3]. Depending on the concentration, potentially toxic elements can cause damage to human health ranging from liver and kidney dysfunctions to carcinogenesis. However, not all contaminants remain in the tea infusion. Most residual toxins are present in very low levels, lowering the risk associated with tea ingestion. One of the most common methods to assess trace multi-element concentrations in tea infusion samples is ICP-MS elemental determination after sample digestion. This method requires microwave digestion systems and reagents such as concentrated acids and oxidant agents. The whole process can achieve dilution factors of 10 100 times. This approach requires long run times and excess reagents, often resulting in poor method detection limits. Conversely, direct analysis also compromises performance parameters, such as precision, long-term stability, residue build-up, and increased maintenance. The Agilent Integrated Sample Introduction System (ISIS) accessory for 7700x Series ICP-MS incorporates the Discrete Sampling (DS) mode for direct analysis. This provides the advantage of accurate sample use, avoiding unnecessary sample exposure by the ICP-MS sample introduction system. In addition, DS mode has the ability to maintain instrument performance and provide high speed analysis when combined with the unique capabilities of ORS 3 and He as collision gas. The synergy between a collision cell that operates effectively with an inert cell gas (helium) and discrete sampling results in excellent performance as measured by detection limits, precision, and throughput. This application note demonstrates the benefits of combining helium collision mode and discrete sampling to develop and validate an easy-to-use, polyatomic interference-free method for the determination of 13 elements in tea infusion. 2

Experimental A sample loop The Agilent 7700x ICP-MS System (G3281A) was fitted with an Agilent I-AS Autosampler (G3160B) with a standard sample introduction system, which includes a MicroMist glass concentric nebulizer, quartz double pass spray chamber, connector tube with dilution port, a quartz torch with a 2.5-mm id and standard nickel cones. Discrete Sampling (DS) is performed with the ISIS accessory (G4911B) set up with one peristaltic pump and one 6-port switching valve. It is coupled to the ICP-MS to deliver only the required amount of sample, achieving productivity gains, and further improving long-term matrix tolerance while reducing run time and sample exposure as much as possible. A preconfigured ISIS tubing kit for ISIS-DS is available (p/n G4911-68202) for easy setup and convenience. Figure 1 shows a schematic of how the ISIS-DS system works. While there is a continuous flow of the internal standards and carrier solutions, the sample fills the loop with the 6-port valve in the load position. The valve then switches to the injection position, sending the carrier solution through the sample loop and pushing the sample into the nebulizer. sample B 6-port valve P2 carrier 6-port valve P2 carrier ISTD sample loop ISIS P1 ISIS P1 waste waste ISTD mixing tee ISTD mixing tee to nebulizer to nebulizer A detailed description of the ISIS-DS operation can be found in previous publications [4,5]. The 7700x tuning was based on a preset method for food and clinical applications using ICP-MS MassHunter software. No Gas, He, and HEHe modes were used in this method and the tuning conditions can be found in Table 1. For calibration, all 13 elements were divided into two groups; one in the lower range (0, 0.1, 0.5, 1, 5, 10, and 100 µg/l) including As, Cd, Cr, Cu, Hg, Ni, Pb, and Se, and the other in the higher range (0, 10, 100, 250, 500, 1,000, and 2,000 µg/l) including Al, Ba, Fe, Mn, and Zn. Both analytical curves were made in HNO 3 0.2% v/v. The method did not require matrix matching or recalibrations. A Sc, Ge, and Y solution containing 250 µg/l of each element was used as online internal standard. After initial calibration, a sequence with 10 analytical blanks was run to find standard deviation (s) and calculate detection limits (DL = 3 s) and quantification limits (QL = 10 s). Repeatability was calculated using coefficient of variation (CV with n = 16). Due to the lack of certified reference materials for this matrix, accuracy was assessed through recovery of the analytes after spikes. In both cases (precision and accuracy), the infusion samples were spiked at different concentration levels: As, Cd, Cr, Cu, Hg, Ni, Pb, and Se at 10 and 50 µg/l; Al, Ba, Fe, Mn, and Zn at 250 and 500 µg/l. sample Figure 1. Table 1. ISTD ISIS operation at DS mode: Valve in load position (A) and injection position (B). Agilent 7700x ICP-MS and ISIS-DS Operating Conditions Agilent 7700x parameters RF power Carrier gas flow 1,550 W 1.1L/min Number of elements/internal standards 13/3 Replicate/peak pattern/sweeps 3/3/100 Reading time Cell tuning modes Shift time between modes(no Gas/He/HEHe) He cell flow on HEHe mode Kinetic energy discrimination on HEHe mode He cell flow on He mode Kinetic energy discrimination on He mode Agilent ISIS-DS parameters Loop volume and size Uptake time Acquisition delay Rinse time 60 seconds No Gas/ He/ HEHe 5 seconds 10 ml/min 7 V 5 ml/min 5 V 150/30 (µl/cm) 20 seconds 20 seconds Pre-emptive rinse during acquisition 3

Method performance was verified using four varieties of tea as detailed in Table 2. For each variety, the study tested three different brands found in a local market (2013 Campinas, SP - Brazil), and for each brand, three infusions were prepared. Infusion preparation was done by boiling 200 ml of deionized water (18.2 MW.cm) with 1.5 g of sample (equivalent to a commercial tea bag) for 3 minutes. After cooling, the infusions were acidified (HNO 3 0.2% v/v) and filtered through a 250-µm polymeric membrane and transferred to the autosampler vials. Table 2. Tea Samples Used in Method Application Lemon balm Leaves of Melissa officinalis, L. Fennel Fruits of Pimpinella anisum, L. Flowers and fruits Apple Fruit - Pyrus malus L. Flowers Hibiscus - Hibiscus sabdariffa, L. Fruits and Flowers Wild Rose - Rosa canina, L. Orange Peel - Citrus aurantium, L. Lemon Peel - Citrus Limonum, (L.) Osbeck leaves Stevia - Stevia rebaundiana, Bert Red Stems and leaves of Camellia sinensis (L.) Kuntze Results and Discussion Routine methods for this application require high througput, low DLs, and low CV. All these figures of merit are discussed below Productivity and stability In an analysis process, one of the most time-consuming steps is sample preparation. The replacement of conventional digestion by direct sample introduction dramatically reduces the analysis time. Some standard features of the Agilent 7700x ICP-MS such as the MicroMist nebulizer, Peltier-cooled double pass spray chamber, High Matrix Introduction (HMI) system, wide diameter (2.5-mm id) torch injector, and solid state RF (27 MHz) generator are important to ensure a robust performance in challenging matrices such as those used in this study. These features result in a very robust plasma with CeO + /Ce + ratios lower than 1.5%. Even with this robust plasma performance, a long routine with direct tea sample injections can cause cone buildup and internal standard instability. This requires maintenance stops for cleaning or replacement of some sample introduction parts. With ISIS-DS, it is possible to overcome these problems by reducing ICP-MS sample exposure by 3 to 4 times, resulting in increased overall system stability. It is possible to have an estimate of the system stability from the 2-hour internal standard profile run shown in Figure 2. ISTD recovery % (outlier setting) 150 100 50 0 45 Sc (no gas) 45 Sc (He) 45 Sc (HEHe) 72 Ge (no gas) 72 Ge (He) 72 Ge (HEHe) 89 Y (no gas) 89 Y (He) 89 Y (HEHe) Figure 2. Internal standard recoveries normalized to the calibration blank for all samples. 4

Comparing total analysis time per sample (sample uptake, stabilization, reading, and rinse times), DS-ICP-MS achieved a 30% cycle time reduction over conventional ICP-MS (197 seconds to 136 seconds). This analysis time reduction was achieved by ISIS at DS mode only. The ORS 3 contribution must also be considered, because it can resolve almost every polyatomic interference using just He in different modes (regular He mode and HEHe mode), and its small volume allows the shift between No Gas, He, and HEHe modes in approximately 5 seconds. The analyst can then choose more than one cell mode in the same method with no significant increase in analysis time. Regarding the food quality control, some elements can be toxic contaminants such as Al, As, Ba, Cd, Hg, Ni, and Pb; nutrients such as Fe, Mn, and Zn; or even both such as Cu, Cr, and Se. Usually, the contaminants are present at low levels in a sample, while nutrients can be present at higher levels. This is a challenge faced by food analysts, because it is difficult to combine different concentration ranges in a single method or even in a single instrument. The Agilent 7700x ICP-MS dual stage detector can obtain calibration curves for all these elements in different ranges with linearity (R 2 ) greater than 0.9999. Detection limits, precision, and recovery Direct analysis of tea samples minimized sample handling, thereby decreasing the risk of contamination and hazardous situations with chemical handling. In addition, this method does not require dilution, thereby improving detection limits. The fast shift between cell modes provides the opportunity for the user to analyze with three different tune modes taking the best of each one. For example, 52 Cr +, which has a strong interference from 12 C 40 Ar +, but in this application Cr could be easily quantified in He mode, while 202 Hg +, which has no polyatomic interference observed in tea infusion analysis could be analyzed in No Gas mode. This feature lowers detection limits. Another challenge in lowering detection limits is the potential for contamination in noncontrolled environments. This affects elements such as Al, Cu, Mn, and Zn. Table 3 lists DLs, QLs, and other performance measures as well as the best analysis conditions for the analytes studied. Using the conditions proposed by this method, it was possible to keep the precision values less than 20%, and recoveries in the two levels between 82 and 121% Table 3. Analysis Conditions and Performance Measures Limits (µg/l) Precision Recovery (%) Isotope Cell ISTD DL (3s) QL (10s) (%) Level 1 Level 2 27 Al He 72 Ge 6 20 13 103 ± 5 99 ± 3 52 Cr He - 0.09 0.29 17 85 ± 1 91 ± 1 55 Mn He 72 Ge 0.11 0.38 11 87 ± 4 89 ± 6 56 Fe HEHe 72 Ge 0.66 2.2 8 85 ± 4 92 ± 3 60 Ni He - 0.63 2.1 10 82 ± 1 89 ± 1 63 Cu - 72 Ge 0.51 1.7 2 95 ± 5 92 ± 1 66 Zn He 72 Ge 10 35 15 95 ± 6 95 ± 3 75 As HEHe 89 Y 0.14 0.46 4 96 ± 2 104 ± 2 80 Se HEHe - 0.15 0.50 3 106 ± 2 110.5 ± 0.4 111 Cd - 72 Ge 0.016 0.053 4 99 ± 1 100 ± 1 138 Ba He 72 Ge 0.60 2.0 12 110 ± 8 111 ± 3 202 Hg - 45 Sc 0.10 0.34 9 111 ± 4 121.0 ± 0.4 206 Pb HEHe 72 Ge 0.12 0.39 6 112 ± 3 113 ± 2 5

Sample analysis After optimizing analysis conditions and evaluating the method performance, three samples of four selected varieties of tea were analyzed in triplicate to verify the performance of the method while running real samples. Table 4 shows the results. The data show the method capability, using the discrete sampling system ISIS-DS (without sample dilution), for quantifying extremely low concentrations of several elements, including those that present risk to human health (such as As, Cd, Cr, Hg, and Pb). Results showed a large variation in concentration values when comparing different varieties of tea infusions. It was determined that the infusion of red tea had the highest concentrations of Al, As, Cr, Mn, Ni, and Pb. However, no sample presented values higher than the maximum established in Resolution 12/11 where the stated limits are: 600 µg As/L, 600 µg Pb/L, and 400 µg Cd/L [7]. Table 4. Average Result and Concentration Range of the Tea Infusions Analyzed Average result and range (n = 3, µg/l) Element Lemon balm Fennel Flowers and fruits Red Al ND < 20 32 (31 34) 28 (24 32) 954 (882 1,033) As ND < 0.46 ND < 0.46 0.18 (ND < 0.46 0.53) 0.76 (ND < 0.46 1.8) Ba 16 (8.2 21) 7.7 (7.0 8.1) 103 (89 111) 15 (13-20) Cd 0.093 (ND < 0.053 0.21) ND < 0.053 0.021 (ND < 0.053 0.063) ND < 0.053 Cr ND < 0.29 ND < 0.29 0.33 (ND < 0.29 0.53) 0.66 (0.58 0.74) Cu 7.7 (5.2 9.2) 21 (19 22) 9.8 (6.1 13) 6.3 (4.5 9.7) Fe 18 (11 24) 53 (27 70) 91 (70 102) 65 (47 81) Hg ND < 0.34 ND < 0.34 ND < 0.34 ND < 0.34 Mn 509 (157 896) 56 (47 62) 405 (357 434) 1020 (951 1,131) Ni ND < 2.1 ND < 2.1 3.2 (2.1 4.5) 19 (16 22) Pb 0.43 (ND < 0.39 0.86) ND < 0.39 0.22 (ND < 0.39 0.65) 0.78 (ND < 0.39 1.3) Se ND < 0.50 ND < 0.50 ND < 0.50 ND < 0.50 Zn 21 (ND < 35 62) 42 (37 46) ND < 35 30 (ND < 35-45) *ND = Non Detected 6

Conclusion This application note describes the development and validation of an analytical methodology for the determination of the total concentration of 13 potentially toxins and micronutrients in tea infusions using a direct analysis approach with ISIS-DS. This method provides significant benefits in terms of reduced cycle time and sample handling. Trace amounts for all 13 chemicals were found to be below published regulatory limits [7]. This method can be used to monitor tea infusions with good precision and accuracy and can also be adapted for other drinks such as coffee and fruit juices. References 1. R. Bunkova, I. Marova, M. Nemec Antimutagenic properties of green tea Plant Foods for Human Nutrition 60: 25-29, 2005. 2. T. Karak, R.M. Bhagat Trace elements in tea leaves, made tea and tea infusion: A review Rood Research International 43: 2234-2252, 2010. 3. W.Y. Han, M.A.L.F. Shi Yz, J.Y. Ruan Arsenic, cadmium, chromium, cobalt, and copper in different types of Chinese tea Bulletins of Environmental Contamination and Toxicology 75: 272-277, 2005. 4. S. Wilbur and C. Jones Maximizing Productivity in High Matrix Samples using the Agilent 7700x ICP-MS with ISIS Discrete Sampling: EPA 6020A Compliant Analysis in Less Than 2 Minutes per Sample Agilent Technologies publication 5990-5437EN 5. S. Wilbur, T. Kuwabara, and T. Sakai, HighSpeed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS) Agilent Technogies publication 5990-3678EN 6. FAO/WHO Summary Evaluation Performed by the Joint FAO/WHO Expert Committee on Food Additives (JECFA). Maastricht, The Netherlands, 2012. 7. Limites Máximos de Contaminantes Inorgânicos em Alimentos, Regulamento Técnico MERCOSUL, Resolução 012/2011. Assunção, Paraguai, 2011. For More Information These data represent typical results. For more information on our products and services, visit our Web site at www.agilent.com/chem. 7

www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 2013 Printed in the USA October 14, 2013 5991-3252EN