AppNote 7/2003. Coupling Retention Time Locked Methods and Libraries to Automated SPME or SBSE for Analysis of Flavors and Fragrances KEYWORDS

Similar documents
AppNote 7/2003. Coupling Retention Time Locked Methods and Libraries to Automated SPME or SBSE for Analysis of Flavors and Fragrances

AppNote 6/2004. Thermal Desorption GC Analysis of High Boiling, High Molecular Weight Hydrocarbons SUMMARY

AppNote 8/2008. A New Purge Tool for Use with Automated Headspace Analysis KEYWORDS ABSTRACT

AppNote 14/2012. Thermal Gravimetric Analysis/Mass Spectrometry Simulation using the GERSTEL Automated Pyrolyzer KEYWORDS ABSTRACT

AppNote 2/2002. Discrimination of Soft Drinks using a Chemical Sensor and Principal Component Analysis KEYWORDS ABSTRACT

AppNote 1/2004. Use of a Mass Spectral Based Chemical Sensor to Discriminate Food and Beverage Samples: Olive Oils and Wine as Examples KEYWORDS

AppNote 4/2008. Automated Dynamic Headspace Sampling of Aqueous Samples Using Replaceable Adsorbent Traps KEYWORDS ABSTRACT

AromaOffi ce: Application of a Novel Linear Retention Indices Database to a Complex Hop Essential Oil

TechNote 5/2002 KEYWORDS ABSTRACT. SBSE, Twister, Thermal Desorption

AppNote 4/1998. New Developments in Multi-Dimensional Capillary Gas Chromatography KEYWORDS ABSTRACT

AppNote 5/2008. Automation of Sample Preparation Steps using a Robotic X-Y-Z Coordinate Autosampler with Software Control KEYWORDS ABSTRACT

AppNote 15/2012. Large Volume Injection with Solvent Venting - Application to Trace Detection of Analytes in Water INTRODUCTION

AppNote 3/2014. Analysis of Aroma Compounds in Edible Oils by Direct Thermal Desorption GC/MS using Slitted Microvials KEYWORDS ABSTRACT

AppNote 2/2000. Stir Bar Sorptive Extraction (SBSE) applied to En vi ron men tal Aqueous Samples INTRODUCTION

AppNote 1/2007. Automated Dynamic Headspace Sampling using Replaceable Sorbent Traps KEYWORDS ABSTRACT

AppNote 8/2002 KEYWORDS ABSTRACT. Electronic Nose, Fast GC, Headspace Analysis, Chemometrics,

AppNote 5/2002. Characterization of a Mass Spectrometer based Electronic Nose for Routine Quality Control Measurements of Flavors KEYWORDS ABSTRACT

AppNote 3/2007 KEYWORDS INTRODUCTION. Beverage, Detector, Headspace, Quality Control

AppNote 2/2001. Stir Bar Sorptive Extraction: Enhancing Selectivity of the PDMS Phase KEYWORDS ABSTRACT

AppNote 6/2010. part a. Alternative Procedure for Extraction and Analysis of PAHs in Seafood by QuEChERS-SBSE-GC-MS KEYWORDS ABSTRACT

Analysis of Food Samples using Thin Film Solid Phase Microextraction (TF-SPME) and Thermal Desorption GC/MS

AppNote 4/2005. A Selectable Single or Multidimensional. Fraction Collection and Dual Detection for Trace Analysis of Complex Samples KEYWORDS

AppNote 2/2004. Analysis of Packaging Materials using a Mass Spectral Based Chemical Sensor KEYWORDS ABSTRACT

AppNote 2/1996. Andreas Hoffmann Gerstel GmbH & Co.KG, Eberhard-Gerstel-Platz 1, D Mülheim an der Ruhr, Germany

Off-gassing of Rubber Particles Used for Athletic Fields using Automated Dynamic Headspace Sampling

AppNote 10/2012. Large Volume Injection with GC/MS Multi Column Switching for Direct Determination of Ethyl Carbamate in Alcoholic Beverages KEYWORDS

AppNote 6/2012. Flavor and Fragrance Analysis of Consumer Products - Dynamic Headspace Compared to Some Traditional Analysis Approaches KEYWORDS

Thermal Desorption Unit TDU

AppNote 1/2011 KEYWORDS ABSTRACT

AppNote 8/2009. Fragrance Profi ling of Consumer Products using a Fully Automated Dynamic Headspace System KEYWORDS ABSTRACT

AppNote 7/2002. Classifi cation of Food and Flavor Samples using a Chemical Sensor KEYWORDS ABSTRACT

AppNote 2/2007. Automated Solid Phase Microextraction using the GERSTEL MPS Prepstation and MAESTRO Software KEYWORDS ABSTRACT

Screening of Pesticide Residues in Water by Sequential Stir Bar Sorptive Extraction-Thermal Desorption with GC/MS

AppNote 2/2012. Determining Phenolic Compounds in Whisky using Direct Large Volume Injection and Stir Bar Sorptive Extraction KEYWORDS ABSTRACT

AppNote 6/2014. Automated Online Desorption and Analysis of DNPH Derivatives of Airborne Aldehydes and Ketones KEYWORDS ABSTRACT

AppNote 8/2000. Volatile Organic Compounds from Adhesives and their Contribution to Indoor Air Problems KEY WORDS ABSTRACT

AppNote 9/2012. Solid Phase Micro-Extraction (SPME) Coupled with Selectable 1 D/ 2 D GC-MS for the Determination of Food Product Flavors KEYWORDS

AppNote 7/2014. Automating Liquid-Liquid Extractions using a Bench-top Workstation KEYWORDS ABSTRACT. Sample Preparation, Lab Automation, LC/MS/MS

AppNote 3/2012. Rapid Automated Screening of Extractable Compounds in Materials for Food Packaging, Medical or Technical Purposes KEYWORDS ABSTRACT

AppNote 8/2012. Rapid Automated Extraction and Confi rmation of Buprenorphine and Norbuprenorphine in Urine by DPX-LC/MS/MS KEYWORDS ABSTRACT

AppNote 1/2013. Determination of Barbiturates and 11-Nor-9-carboxy- 9 -THC in Urine using Automated Disposable Pipette Extraction (DPX) and LC/MS/MS

AppNote 8/2011 KEYWORDS ABSTRACT INTRODUCTION

AppNote 2/1998. Direct Injection of Distilled Spirits with PTV Matrix Removal: The Perfect Splitless Injection? KEYWORDS INTRODUCTION

AppNote 12/2008 KEYWORDS ABSTRACT

AppNote 3/1998 KEYWORDS ABSTRACT. Beverage, Food & Flavor, PTV, Large Volume Injection, Headspace, Thermal Desorption.

AppNote 2/2013. A High Throughput Automated Sample Preparation and Analysis Workfl ow for Comprehensive Toxicology Urine Screenings using LC/MS/MS

AppNote 5/2001. The Analysis of the Bitter and Other Flavour Compounds in Beer and Wort by Stir Bar Sorptive Extraction (SBSE) Followed by HPLC

KEYWORDS ABSTRACT AppNote 6/2002

AppNote 12/2012. Comparison of EG-Silicone-SBSE and Derivatization-PDMS-SBSE for the Analysis of Phenolic Compounds and Off-fl avors in Water KEYWORDS

AppNote 3/2011. Using Three Types of Twister Phases for Stir Bar Sorptive Extraction of Whisky, Wine and Fruit Juice KEYWORDS ABSTRACT

AppNote 2/2000. Stir Bar Sorptive Extraction (SBSE) applied to Environmental Aqueous Samples

AppNote 2/1999. Automated Sample Fractionation and Analysis Using a Modular LC-GC System

AppNote 7/2005. Detection of Packaging Emissions using a Flexible Headspace Sampler Combined with a Multi Sensor System and a Separation Unit KEYWORDS

Selective Sample Enrichment of Gaseous Samples using a Cooled Injection System: Trace Determination of Sulphur Components in Natural Gas

The Focus Robotic Sample Processor as a Tool for the Multiple Analysis of Samples using Complementary Techniques

AppNote 8/2001 KEYWORDS ABSTRACT. PAH, SPME, SBSE, GERSTEL Twister, Thermal desorption, GC-MS, PCB

GERSTEL MultiPurpose Sampler MPS. Versatile autosampler and sample preparation robot

Determination of Volatile Substances Proof of Food Adulteration

AppNote 6/2010. part b. High Throughput Method for the Determination of PAHs in Seafood by QuEChERS-SBSE-GC-MS SCOPE PRINCIPLE

AppNote 5/1994. The Use of a Multi Purpose Sampler for Headspace GC-MS Analysis of Volatile Organic Compounds in Human Urine

GERSTEL Thermal Desorption System TDS 3 The GERSTEL TDS 3 is a flexible multi-functional thermal desorption system for highly sensitive and accurate d

Analysis of Terpenes in Cannabis Using the Agilent 7697A/7890B/5977B Headspace GC-MSD System

AppNote 11/2012. Fully Automated Analysis of Doping Substances in Equine Urine by Disposable Pipette Extraction (DPX) coupled to GC/QqQ-MS KEYWORDS

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Discovering the Scent of Celery: HS-SPME, GC-TOFMS, and Retention Indices for the Characterization of Volatiles

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

U.S. EPA Method 8270 for multicomponent analyte determination

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Characterisation of allergens in cosmetics by GC GC TOF MS with Select-eV variable-energy electron ionisation

Expectations for GC-MS Lab

GAS CHROMATOGRAPHY MASS SPECTROMETRY. Pre-Lab Questions

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Application. Gas Chromatography February Introduction

Determination of Formaldehyde and VOCs in Wood-based Products using an Automated Micro-Scale Chamber

AppNote 2/2015 KEYWORDS ABSTRACT

anthracene Figure 1: Structures of Selected Polyaromatic Hydrocarbons (PAHs)

AppNote 2/1996. Andreas Hoffmann Gerstel GmbH & Co.KG, Aktienstrasse , D Mülheim an der Ruhr, Germany

Multi-residue analysis of pesticides by GC-HRMS

AppNote 7/2013. Comprehensive Automation of the SPE- GC/MS Analysis of Opioids, Cocaine and Metabolites from Serum and Other Matrices KEYWORDS

Gas Chromatography (GC)

GAS CHROMATOGRAPHY (GC)

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Enhanced Simultaneous Total ion and Selective Multi-Ion GC/MS Using a Combination of New Software and Low-Bleed Columns

Sulfotepp impurities in Chlorpyrifos EC formulations

Analysis of BTEX in Natural Water with SPME

Multi-residue GC-MS analysis. Richard Fussell CSL York, UK

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace

Determination of volatiles in Food sample preparation strategies and techniques Dr Kathy Ridgway

Determination of Organochlorine Pesticides and Polychlorinated Biphenyls Using GC/MS/MS Operated in the MRM Mode

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Automated and accurate component detection using reference mass spectra

Yokogawa Analytical Systems Inc., Nakacho, Musashino-shi, Tokyo , Japan.

Author. Abstract. Introduction

Highly Sensitive and Rugged GC/MS/MS Tool

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

GC/MS Application Note

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Transcription:

AppNote 7/2003 Coupling Retention Time Locked Methods and Libraries to Automated SPME or SBSE for Analysis of Flavors and Fragrances Vanessa R. Kinton, Edward A. Pfannkoch, Jacqueline A. Whitecavage, John Thorp Gerstel, Inc., 701 Digital Drive, Suite J, Linthicum, MD 21090, USA KEYWORDS Flavors, gas chromatography, mass spectrometry, retention time locking, SPME, SBSE ABSTRACT Analysis of foods, flavors and fragrances is commonly done by capillary gas chromatography coupled to a range of detectors depending on the particular needs of the analysis. Using Retention Time Locked (RTL) methods can facilitate the transfer from one instrument to another, particularly if different detectors (eg. MSD and FID) are used. Although liquid or headspace injection methods are often used for these analyses, difficult and time-consuming sample preparation steps, or lack of sensitivity are commonly encountered. Automated Solid Phase Microextraction (SPME) or Stir Bar Sorptive Extraction (SBSE) can often provide substantial improvements to existing methods. In this contribution, we illustrate the use of RTL methods and the Agilent RTL Flavor Library when the GERSTEL MPS 2 autosampler is used to automate either SPME or SBSE as sample extraction and introduction technique. Procedures are given for calibrating the GC method using

either technique. RTL methods created for both sample introduction techniques could be moved from a GC- FID to a GC-MSD system with different column head pressure requirements. Calibrated RTL SPME or SBSE methods were then used to analyze several sample types that otherwise would pose significant matrix interference or detection limit issues. Example chromatograms with identifications based on the RTL flavor library are given. INTRODUCTION Analysis of flavor and fragrance composition can be challenging. Flavors and fragrances are usually complex mixtures that may contain many constituents that could include terpenes, diterpenes, sesquiterpenoids, etc. Common techniques used to analyze these types of samples are gas chromatography (GC) coupled to either flame ionization detector (FID) or mass spectrometer (MS). GC/MS is widely used to separate and tentatively identify flavor constituents. There are several techniques that introduce flavor components into GC instruments. These include static headspace or enrichment techniques such as solid phase microextraction (SPME) or, more recently, stir bar sorptive extraction (SBSE). Regardless of the method used for sample introduction the identity obtained with the mass spectrometer for the flavor components is considered tentative. In order to have full confidence of the identity of a compound, standards are necessary to validate the MS findings. Identification of flavor components with an MS detector is also difficult because usually they are mixtures of compounds with similar mass fragmentation. Since isomers give similar spectra, retention indices are commonly used to complement the information obtained with the MS. There are published libraries with retention indices for flavor and fragrances, but these libraries depend on the column type, linear velocity of the carrier gas, oven profile, etc. Due to the dependence of these experimental conditions, the retention indices can be difficult to reproduce. A technique known as retention-time-locking (RTL) can compensate for the disadvantages mentioned before of using GC/MS. The ability to lock a GC/MS method and subsequently running samples under locked conditions has the advantage of providing complementary information to the identification obtained using a mass spectral library. When a GC/MS method is locked, the identification of the individual components is carried out with the mass spectral profile and also by its retention time. The extra information is particularly useful when isomers with similar mass spectra are encountered. In this study, an RTL flavor database created by David et al [1]. was used to analyze different beverage samples. The combination of the RTL capabilities along with a diversity of sample introduction into a GC creates a versatile way of analyzing flavor and fragrances samples. EXPERIMENTAL Instrumentation. All analyses were performed on a GC (6890, Agilent Technologies) with mass selective detection (MSD). The instrument was equipped with a Thermal Desorption unit with autosampler (TDS 2 & TDS A, Gerstel), a Multipurpose sampler with SPME capability (MPS 2, Gerstel) and a PTV inlet (CIS 4, Gerstel). Analysis Conditions. Column: 30m HP-5 (Agilent), d i = 0.25mm, d f = 0.25mm Pneumatics: He, P i = 10.6 psi (flavors), P i = 20.3 psi (pesticides) constant pressure Oven: 60 C, 3 C/min, 240 C for flavors 70 C (2 min), 25 C/min, 150 C, 3 C/min, 200 C, 8 C/min, 300 C (5 min) for pesticides Twister desorption (SBSE). TDS 2 splitless, 20 C, 60 C/min, 250 C (5 min) PTV solvent vent (50 ml/min), split ratio 20:1-120 C (0.2 min), 12 C/s, 280 C (3 min) MPS 2 / SPME Fibers 100μm PDMS Equilibration 60 C (10 min) Extraction 60 C (10 min) PTV 0.8 min splitless isothermal at 250 C Sample Preparation. Grapefruit juice. 0.1 ml of undiluted sample was transferred to a 20 ml headspace vial (SPME). Orange juice. 10 ml of undiluted orange juice was transferred to a 10ml headspace vial. The sample was AN/2003/07-2

spiked with 10 ug/l (Restek #32032 Pesticide Evaluation Mix). A Twister was added and the sample stirred at room temperature for 90 minutes. After extraction the Twister was removed, rinsed, dried and placed into a thermal desorption tube for analysis (SBSE). Retention Time Locking. Identification of the flavors and pesticides was performed automatically using the Agilent Screener software in combination with the Agilent Retention Time Locking (RTL) library for flavors. For comparison purposes, mass spectral identifications were carried out using the Wiley 138 mass spectral library. The locking compound used for the flavors was hexadecane, the inlet pressure was adjusted to give a retention time of 31.535 minutes. The locking compound used for the pesticides was Aldrin, the inlet pressure was adjusted to give a retention time of 18.530 minutes. Hexadecane is part of the 400 compounds included in the flavor RTL database with an elution time of 31.535 minutes. Our system (Figure 1) included a programmable temperature vaporizer (PTV) inlet that includes a septumless head injector (CIS 4, Gerstel). The GERS- TEL septumless head (SLH) on the CIS 4 eliminates the problem of inlet septum coring common to most inlets used with the blunt-tipped SPME needles. After some preliminary runs, the method was locked and 5 runs were automatically created using Agilent s ChemStation data acquisition software (version DA.00.01, Agilent Technologies). Figure 2 shows the 5 runs used for locking the flavor method. The RTL algorithm creates an RTL curve using the inlet pressure and the locking compound retention time. In this study, the measured retention time was 31.587 minutes and the locked retention time with a constant pressure of 10.42 psi was 31.535 minutes (obtained from the RTL database). RESULTS AND DISCUSSION The experimental conditions used in this study are similar to the ones used by David et al [1]. In their study, they replicated the parameters used by Adams [2] in which he publishes retention indices databases useful in the flavor industry. Since all the experimental conditions are the same, the spectra and retention data are valid for all studies. In order to use the RTL flavor database, we decided to use hexadecane to lock our GC/MS method. 32.302 1 Time--> 10.00 20.00 30.00 31.937 1 Time--> 10.00 20.00 30.00 31.587 1 Time--> 10.00 20.00 30.00 1 31.251 Time--> 10.00 20.00 30.00 1 30.929 Figure 1. GC/MS system used in the study. SPME autosampler coupled with a septumless head in the front injector and Thermal Desorption Autosampler coupled with a PTV injector in the back injector. Time--> 10.00 20.00 30.00 Figure 2. Retention time locking setup using hexadecane as the locking compound. The 5 individual runs were created automatically by using the Agilent s RTL method acquisition software. AN/2003/07-3

We decided to examine a grapefruit juice sample (purchased at a local store) using solid phase microextraction. Figure 3 shows the TIC obtained after sampling the grapefruit juice headspace with a PDMS SMPE fiber. Table 1 shows the identities obtained for some compounds using Wiley 138 mass spectral database and the flavor RTL database. 3000000 6 000 Time--> 1 2 3 4 5 7 8 9 10 10.00 20.00 30.00 Figure 3. Total ion chromatogram obtained after sampling the headspace above 100 ul of grapefruit juice with a SPME fiber using an automated sampler. Data was acquired under locked conditions using hexadecane as the locking compound. Table 1. Identification of some compounds by sampling grapefruit juice with SPME fiber. If more than one match was found with the Wiley library with a quality match of over 90% an * was placed beside the match. Peak # Compound Wiley RTL Screener 1 α-pinene X * X 2 Myrcene X X 3 δ-3-carene X * X 4 Cymene X * X 5 Linalool X * X 6 α-terpineol X X 7 cis-geraniol X * X 8 Carvone X * X 9 Geraniol X * X 10 Valencene X * X Inspection of Table 1 indicates that the additional information obtained while running the juice sample under locked conditions is complementing the mass spectral information. For example, α-pinene was easily identified using the RTL database while the Wiley database gave three matches all with match qualities of over 90%. Figure 4 shows the library matches for the peak eluting at 5.14 minutes in the grapefruit juice. It can be seen in this figure that the mass spectra profiles are very similar for all the possible matches. In this particular case, the information provided by using the RTL screener clearly identifies the peak eluting at 5.14 minutes as α-pinene. AN/2003/07-4

7000 93 A 8000 #8160:.alpha.-Pinene, (-)- CAS# 80-56-8 MQ 94 93 B 5000 3000 77 1000 105 121 136 77 105 121 136 m/z--> 40 60 80 100 120 140 8000 93 #8161: Bicyclo[3.1.1]hept-2-ene, 3,6,6-trimethyl- CAS# 4889-83-2 MQ 94 C m/z--> 40 60 80 100 120 8000 #8096:.gamma.-Terpinene CAS# 99-85-4 MQ 90 93 D 77 105 121 136 77 105 121 136 m/z--> 80 100 120 m/z--> 40 60 80 100 120 Figure 4. Possible library matches using Wiley 138 for peak eluting at 5.14 minutes from TIC of Figure 3. A) Scan of peak eluting at 5.14 min in grapefruit juice; B) Wiley 138 library match #1; C) Wiley 138 library match #2; D) Wiley library match #3. [Note: By using the RTL database, it was concluded that the peak eluting at 5.14 min was α-pinene (spectrum displayed in B)] We also decided to inspect the pesticide RTL database [3] using a GC/MS method. For this example, we inspected an orange juice sample spiked with pesticides. As seen in Figure 5, the combination of stir bar sorptive extraction (SBSE) coupled with a PTV and thermal desorption system can provide a complex totalion-chromatogram profile. It can also be seen in this figure that even at 10 ppb levels the pesticides were detected and screened with the RTL pesticide database (Figure 6). AN/2003/07-5

000 3000000 000 Time--> 10.00 1 3 2 20.00 30.00 Figure 5. Total ion chromatogram obtained after sampling spiked-orange juice with SBSE. The pesticides were spiked at 10 ppb levels. Peak identities are 1) Aldrin; 2) Endrin; 3) p,p DDT. 8000 Ion 263.00 (262.70 to 263.70) Ion 66.00 (65.70 to 66.70) Ion 265.00 (264.70 to 265.70) Ion 261.00 (260.70 to 261.70) 18.50 Time--> 17.50 18.00 18.50 19.00 19.50 66 263 (265) Aldrin 18.495min (-0.033) 91 response 22047 Ion Exp% Act% 220 293 329 263.00 100 100 66.00 86.70 97.02 265.00 66.70 67.44 261.00 64.40 69.12 m/z--> 100 200 300 Figure 6. Identification of Aldrin by the RTL screener in spiked orange juice. AN/2003/07-6

CONCLUSIONS By using enrichment techniques such as SPME and/or SBSE flavor constituents were easily detected using a GC/MS. Also, the use of an RTL method added meaningful retention time information supporting analytes identification. The use of automated SPME or SBSE is compatible with RTL databases and does not adversely affect the ability to lock methods. The overall result of the combination of hardware and software created a powerful tool in screening flavor compounds quickly and accurately. REFERENCES [1] David, F. et al. Analysis of Essential Oil Compounds Using Retention Time Locked Methods and Retention Time Databases. Agilent Application Note 5988-6530 EN, May 2002. [2] Adams, R. P. Identification of Essential Oil Components by Gas Chromatography-Mass Spectrometry; Allured Publishing Corporation: Illinois, 1995. [3] Sandra, P et al. J. Chromatogr. A. 1000 (2003) 299-309. AN/2003/07-7

GERSTEL GmbH & Co. KG Eberhard-Gerstel-Platz 1 45473 Mülheim an der Ruhr Germany +49 (0) 208-7 65 03-0 +49 (0) 208-7 65 03 33 gerstel@gerstel.com www.gerstel.com GERSTEL Worldwide GERSTEL, Inc. 701 Digital Drive, Suite J Linthicum, MD 21090 USA +1 (410) 247 5885 +1 (410) 247 5887 sales@gerstelus.com www.gerstelus.com GERSTEL AG Wassergrabe 27 CH-6210 Sursee Switzerland +41 (41) 9 21 97 23 gerstelag@ch.gerstel.com www.gerstel.ch GERSTEL K.K. 1-3-1 Nakane, Meguro-ku Tokyo 152-0031 SMBC Toritsudai Ekimae Bldg 4F Japan +81 3 5731 5321 +81 3 5731 5322 info@gerstel.co.jp www.gerstel.co.jp GERSTEL LLP Level 25, North Tower One Raffles Quay Singapore 048583 +65 6622 5486 +65 6622 5999 SEA@gerstel.com www.gerstel.com GERSTEL Brasil Av. Pascoal da Rocha Falcão, 367 04785-000 São Paulo - SP Brasil +55 (11)5665-8931 +55 (11)5666-9084 gerstel-brasil@gerstel.com www.gerstel.com.br Information, descriptions and specifications in this Publication are subject to change without notice. GERSTEL, GRAPHPACK and TWISTER are registered trademarks of GERSTEL GmbH & Co. KG. Copyright by GERSTEL GmbH & Co. KG Awarded for the active pursuit of environmental sustainability