CALIBRATION MECHANISM FOR INFRARED SPACE TELESCOPE. Patrice Kerhousse

Similar documents
EUCLID M2 MIRROR MECHANISM

GAIA M2M POINTING MECHANISM QUALIFICATION

Equal Pitch and Unequal Pitch:

EXPERIMENTAL RESEARCH REGARDING TRANSIENT REGIME OF KINEMATIC CHAINS INCLUDING PLANETARY TRANSMISSIONS USED IN INDUSTRIAL ROBOTS

SAMPLE CANISTER CAPTURE MECHANISM FOR MSR: CONCEPT DESIGN AND TESTING RESULTS INCLUDING 0-G ENVIRONMENT

Linear guide drives. Synchronous shafts The use of synchronous shafts enables several linear axes to be operated with one drive.

A MIRROR CONTROL MECHANISM FOR SPACE TELESCOPE

Piezoelectric Actuators and Future Motors for Cryogenic Applications in Space

Precision Ball Screw/Spline

FOUR-POINT CONTACT SLEWING RINGS - without gear [ O ]

A METHOD FOR EVALUATION OF THE CHAIN DRIVE EFFICIENCY

INVESTIGATION OF BALL SCREWS FOR FEED DRIVE 1. INTRODUCTION

QinetiQ Proprietary BepiColombo TDA

Outline. Theoretical estimate of the diffraction pattern on the primary mirror plane

OBSERVATIONS ON THE INFLUENCE OF LAUNCH VIBRATION ON BEARING TORQUE AND LUBRICANT PERFORMANCE

T20WN. Data Sheet. Torque transducers. Special features. Installation example with bellows couplings. B en

THE SCANNING MECHANISM FOR ROSETTA / MIDAS FROM AN ENGINEERING MODEL TO THE FLIGHT MODEL

E D D Y C U R R E N T D A M P E R A P P L I C AT I O N D ATA

Selection Calculations For Linear & Rotary Actuators

EQUIPMENT INTERFACE LOAD CHARACTERIZATION IN ACOUSTICS

SKF actuators available for quick delivery. Selection guide

RGS08 Linear Rails: RGS08 Non-Motorized With and Without Guide Screw. RGS08 Non-Motorized Linear Rails

Catapult tests for microgravity characterization of the MICROSCOPE accelerometers. Manuel Rodrigues On behalf ONERA & ZARM team

DISTRIBUTION LIST. Others original copies Name amount. Lens Research & Development 1x Uittenhout, J.M.M. 1x DOCUMENT CHANGE RECORD

Cam Roller Guides. The Drive and Control Company. Service Automation. Electric Drives and Controls. Linear Motion and Assembly Technologies

VOLUMEC. Valve Position Indicator 5

Cold Gas Thruster Qualification for FORMOSAT 5

ROLLER BEARING FAILURES IN REDUCTION GEAR CAUSED BY INADEQUATE DAMPING BY ELASTIC COUPLINGS FOR LOW ORDER EXCITATIONS

KNIFE EDGE FLAT ROLLER

Development of a probe traversing system for an open test section wind tunnel

DRS 61: Incremental encoders, number of lines and zero pulse width freely programmable DRS 60: Incremental Encoders with Zero-Pulse-Teach

The basic dynamic load rating C is a statistical number and it is based on 90% of the bearings surviving 50 km of travel carrying the full load.

SC125MS. Data Sheet and Instruction Manual. ! Warning! Salem Controls Inc. Stepper Motor Driver. Last Updated 12/14/2004

ALCATEL SPACE PLASMA PROPULSION SUBSYSTEM QUALIFICATION STATUS

Mounts for large lens in cooled environment

Product description. Compact Modules. Characteristic features. Further highlights

Intelligent Control of a SPM System Design with Parameter Variations

Nonlinear Rolling Element Bearings in MADYN 2000 Version 4.3

Actuators with ball monorail guidance system and ball screw drive MKUVE15-KGT, MKUVE20-KGT

High Power Electric Propulsion Pointing Mechanisms (HP EPPM) Technical Development Activities (TDA s) strategy, status and planning

AP Physics B Summer Assignment

Dimensional drawing face mount flange radial. R = bending radius min. 40 mm. Dimensional drawing face mount flange axial

Circular Motion and Centripetal Force

Coating K = TFE wear resist, dry lubricant Kerkote X = Special coating, (Example: Kerkote with grease)

R-Plus System Frontespizio_R_PlusSystem.indd 1 11/06/ :32:02

SPICE SDM: INNOVATIVE APPROACHES FOR LINEAR MOTION AND HEAT MANAGEMENT

1. INTRODUCTION ABSTRACT

Selection Calculations For Motorized Actuators

DESIGN AND QUALIFICATION OF THE CRYOGENIC COVER MECHANISM FOR THE MID-INFRARED INSTRUMENT ON THE JAMES WEBB SPACE TELESCOPE

UNI EN ISO 9001:2008 UNI EN ISO 14001:2004 BS OHSAS 18001:2007 EC DIRECTIVE 2014/34/EU (ATEX) CERTIFIED MANAGEMENT SYSTEM

Massachusetts Institute of Technology Department of Electrical Engineering and Computer Science Electric Machines

Nanomotion Precision Piezo Ceramic Motors

Flight Demonstration of Electrostatic Thruster Under Micro-Gravity

Robotics I. Test November 29, 2013

Mathematical model of electromechanical system with variable dissipativity

Development of Microwave Engine

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5

X-LRT-EC Series: High-load, closed-loop stages with built-in stepper motors, motor encoders, controllers, and dust covers

Glossary Innovative Measurement Solutions

X-LRQ-SV2 Series: High vacuum motorized linear stages with built-in controllers

Jerad P. 10/1/2015. Motor Thermal Limits on Torque Production (Frameless Motor Level)

ADM-Aeolus Science and Cal/Val Workshop

OWL: Further steps in designing the telescope mechanical structure and in assessing its performance

Mechatronics. MANE 4490 Fall 2002 Assignment # 1

VIRTUAL TESTING SIMULATION TOOL FOR THE NEW QUAD HEAD EXPANDER ELECTRODYNAMIC SHAKER

EVALUATION OF THE TORQUE PERFORMANCE OF A RANGE OF SPACE-COMPATIBLE OILS & GREASES AS A FUNCTION OF TEMPERATURE, SPEED & LUBRICANT QUANTITY

Physics. Student Materials Advanced Higher. Tutorial Problems Mechanics HIGHER STILL. Spring 2000

RGS06 Linear Rails: RGS06 Non-Motorized With and Without Guide Screw. RGS06 Non-Motorized Linear Rails

ELECTROMECHANICAL DESIGN OF A DRIVE MECHANISM FOR THE ROTATION OF THE SPECTROMETER WHEEL INSIDE MEDET EXPERIMENT

Replacement of Grid Coupling with Bush Pin Coupling in Blower

PHYSICS 9646/02. NANYANG JUNIOR COLLEGE Science Department JC 2 PRELIMINARY EXAMINATION Higher 2. Candidate Name. Tutor Name.

Wheel and Axle. Author: Joseph Harrison. Research Ans Aerospace Engineering 1 Expert, Monash University

AE / AER Series. AER Series

LINE TECH Linear Modules. Ready to built-in linear modules with drive

2. To study circular motion, two students use the hand-held device shown above, which consists of a rod on which a spring scale is attached.

Texas A & M University Department of Mechanical Engineering MEEN 364 Dynamic Systems and Controls Dr. Alexander G. Parlos

The Design and Development of a Separation System for a Low-Cost Spherical Satellite

Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1

SET C PAPER 2 QUESTION

CS Series. Capacitance Type. Type. Proximity Sensor. Features

Lecture 19. Measurement of Solid-Mechanical Quantities (Chapter 8) Measuring Strain Measuring Displacement Measuring Linear Velocity

Advanced Higher Physics. Rotational motion

Two Dimensional Rotational Kinematics Challenge Problem Solutions

FEASIBILITY STUDY ON A LARGE CHOPPER DISC FOR A TOF SPECTROMETER

Generation X. Attitude Control Systems (ACS) Aprille Ericsson Dave Olney Josephine San. July 27, 2000

TE 75R RESEARCH RUBBER FRICTION TEST MACHINE

Influence of electromagnetic stiffness on coupled micro vibrations generated by solar array drive assembly

Chapter 9-10 Test Review

ADCSS 2017: Sodern presentation

Measurements in Mechatronic design. Transducers

Quanser NI-ELVIS Trainer (QNET) Series: QNET Experiment #02: DC Motor Position Control. DC Motor Control Trainer (DCMCT) Student Manual

RV SeRieS S P E C I F I C A T I O N S

ScienceDirect. Finite element analysis and optimization of flexure bearing for linear motor compressor

Technical Guide for Servo Motor Selection

Linear Shaft Motor Sizing Application Note

This equation of motion may be solved either by differential equation method or by graphical method as discussed below:

Vacuum Capacitor Lifetime

Revision 1.0, January 2014 MICRONIX USA, LLC

THE EUCLID VIS READ-OUT SHUTTER UNIT: A LOW DISTURBANCE MECHANISM AT CRYOGENIC TEMPERATURE

JRE SCHOOL OF Engineering

Transcription:

CALIBRATION MECHANISM FOR INFRARED SPACE TELESCOPE Patrice Kerhousse Alcatel space industries Boulevard du Midi, BP 99-06156 Cannes la Bocca Cedex Telephone : 33-(0)4-92-92-72-38 / Fax : 33-(0)4-92-92--20 E-mail : Patrice.Kerhousse@space.alcatel.fr ABSTRACT In order to achieve high infrared optical performances of the HELIOS2 telescope, a Calibration Mechanism, called MECALIB, is developed by ALCATEL. The specific kinematic architecture of MECALIB is based on an eccentric shaft, driven by a stepper motor, linked to a push rod that drives the rotative mirror mounting. This architecture presents the following main advantages: as the two pointing positions are close to the upper dead point of the mechanism, the reduction ratio is high enough to achieve the positioning accuracy during on ground tests (no need of zero G tools). the reduction ratio is highly non-linear so, even if the stepper motor is driven with full steps, the specified angle between the two pointing positions can be accurately adjusted by rotation of the motor housing. only preloaded ball bearings are used in the mechanism to achieve the function without any backlash. The development tests demonstrated that the MECALIB design meets all the requirements but a failure occurred during lifetime tests : the resistive torque increased drastically at the third of the specified lifetime. It has been investigated that the failure was due to the ball bearings MoS2 degradation. The design was modified to use fluid lubricated ball bearings. The qualification tests are in progress. 1. INTRODUCTION Because of unstability of usual space infrared detectors used in infrared telescopes, a periodic calibration of them is necessary to achieve high optical performances. ALCATEL is in charge of the development, qualification and delivery of two flight models of a Calibration Mechanism (MECALIB). The main function of the MECALIB is to allow the in-orbit calibration of the detectors by pointing a mirror at 2 black bodies. This mechanism is to be used on the HELIOS2 telescope. The paper will present the design of the MECALIB, its development test results, and its performances and will point out the ball bearings failure that occurred during development life test. The novelty feature of the mechanism will be highlighted : the kinematic architecture that allows high positioning accuracy on the two pointing positions, with very simple mechanism design and drive electronics. 2. MAIN SPECIFICATIONS The main specifications which were the basis of the MECALIB design are detailed in the following table : REQUIREMENT SPECIFICATION Pointing accuracy : θx, θy, θz Pointing reproducibility : θx, θy, θz Cycles duration : -> -> -> Launch loads : quasi-static sine vibrations random vibrations Stiffness : launch in orbit Mass : < 2 mm < 5 < 0,1 mm < 2,5 < 5 s < 0,7 s < 1,3 s 150 g 30 g g (on internal parts) > 300 Hz > 10 Hz < 4 kg

3. MECHANISM DESCRIPTION MECALIB is mainly composed of: a structure made of titanium a calibrating mirror a mirror mounting with two pairs of ball bearings an actuator with a stepper motor a speed reduction unit a detection system by magneto-resistors a hold-down and release device The drawing thereafter shows the overall architecture of the MECALIB. Actuator The actuator is based on a stepper motor ( steps) whose rotor is directly mounted on the shaft that drives the speed reducer. The shaft is guided by a pair of preloaded ball bearings. The speed reducer is made of an eccentric located on the motor shaft that drives a push rod. The push rod is linked to the mirror mounting. Detection system A position status is sent for each in-orbit functional position (,, ) in order to validate the good execution of each command. The detection device is based on 6 magneto-resistors (MR), 3 nominal and 3 redundant, which resistance changes when a steel tooth, mounted on the motor shaft approaches the MR. Hold-on and release device This device is based on a Shape Memory Alloy (SMA) component that breaks the tightening rod. To minimise the development costs, the SMA component is fully recurrent of the Refocusing Mechanism one, that explains the large size of the component compared to the mechanism. The SMA component has a 90.000 N capability when the rod breaking force is 13.000 N. The main advantage of this technology are : high reliability due to the use of a physic behaviour multi-working possibility.

4. KINEMATIC ARCHITECTURE Ymmc The main kinematic parameters are presented in the scheme hereunder : θ m=224,4 (PMH) (Motor shaft axis) θ m=0 (PMB) 41,4 θ m=183 C 7,2 θ m=31,8 Xmmc (Eccentric axis) 144 θ m=39 Cinématique (Angles) Angles ( ) Angles 1 20 0 θ1 θ2 θ3 30 50 70 90 110 130 150 170 1 190 210 220 230 θm = motor angle θ1 = mirror mounting angle θ2 = angle between mirror mounting and push rod θ3 = angle between shaft axis and push rod This architecture (location of the different rotation axes, eccentric radius, push rod length) is optimised in order to fulfil the following criteria : the angular stroke must be sufficient to hold the mirror out of the field of view during the sleeping mode. in the stowed position, the motor shaft must be stiffly held. the reduction ratio in the pointing positions must be sufficient to provide the specified pointing accuracy without a zero G device. all positions of the MECALIB must coincide with a full step of the motor to simplify the drive electronics. The main positions of the MECALIB are the two pointing positions () and (), the sleeping position (), the stowing position (), the upper dead point (PMH) and the lower dead point (PMB). The complete stroke of MECALIB is from () to () located near (PMH) and needs 107 of the steps of the stepper motor. The location of the different MECALIB positions is detailed in the following figure. The stowing position () was chosen at a motor angle of 31.8 to provide enough stiffness during launch. Taking into account this geometric architecture, the following graphs show the evolution of the different angles and reduction ratios function of the motor angle. Reduction Ration Rapports de Réductions 1.4 1.2 1 0.8 0.6 0.4 0.2 0-0.2-0.4 30 50 70 Motor Angle θm ( ) Rotation Moteur (θm) Cinématique (Rapports de Réduction) 90 110 130 150 170 1 190 210 220 230 Rotation Moteur (θm) Motor Angle θm ( ) dθ1/ dθm dθ2/ dθm dθ3/ dθm The MECALIB kinematic architecture presents the following advantages : as the pointing position () is located at the upper dead point, the angle between the two pointing position (4 10 ) can be easily adjusted by rotation of the motor housing. This rotation has a negligible effect on the () position because of the very high reduction ratio, but it is the opposite for the () position that has a reduction ratio of 1:6. After this adjustment, the absolute pointing accuracy is achieved by adjustment of the three shims located at MECALIB structure interface plane. all positions (,,, ) are reached with the specified accuracy on full steps of the motor in open loop. With this configuration, the pointing accuracy does not depend on the motor coils voltage and current stability, that allows a very simple drive electronics. only preloaded ball bearings are used in the mechanism to achieve the function without any backlash (widely flight proven technology).

5. MECALIB KEY PERFORMANCE DATA Pointing accuracy : θx, θy, θz Pointing reproducibility : θx, θy, θz Cycles duration : -> -> -> Stiffness : launch in orbit Mass : < 0.7 mm < 4,4 (worst case 1 g) <1.5 < 0,04 mm < 2,5 (worst case 1 g) < 1 < 5 s < 0,7 s < 1,3 s > 291 Hz > 20 Hz = 4.020kg max 7. BALL BEARINGS FAILURE After visual inspection in ALCATEL, some ball bearings were sent back to the supplier and to the French National Space Agency (CNES) in order to proceed to fine measurements and Electronic Scanning Microscope analyses. The measurements and analyses showed that all the ball bearings had the same damages : the inner and outer rings present a large track without MoS2 in the area of balls contacts. a high quantity of titanium particle were found on the tracks The following figures present the ball bearings tracks observations and their microscope analyses. Track without MoS2 6. DEVELOPMENTS TESTS These tests have been performed on a development model fully representative of a flight model except for the mirror (replaced by a dummy mass) and the cleanliness class (class.000 instead of class ). The test sequence was the following : MoS2 coating DM Inner ring view (CNES photograph) VIBRATIONS TESTS LIFETIME TEST Main results The specifications in terms of accuracy, reproducibility, release of the mirror, duration of the cycle, motorization margin have been verified in the initial functional test. The vibration tests and the functional tests after vibrations have been performed also nominaly. Nonetheless, a very high degradation of the motorization margin appeared during the lifetime test after.000 cycles (112.000 specified). After inspection, it was found that this degradation was caused by the same kind of damages on all the ball bearings. All the ball bearings were of the same TiC/MoS2 technology. Microscope track view(cnes photograph) Track elements analysis (CNES data)

These measurements demonstrate that the MoS2 coating had completely disappeared from the area of balls contact. A series of two lifetime tests were performed on new ball bearings in order to determine if the degradation was initiated by vibrations damages or because of humidity during the DM lifetime tests. The two additive lifetime tests (one under Nitrogen and the other one under 10-8 Torr vacuum) gave nearly the same results : the ball bearings were not able to achieve even the half specified lifetime. But as the vacuum test was stopped when the resistive torque just began to increase, it has been identified that the MoS2 was responsible for the degradation : the same large track without MoS2 as in the previous ball bearings was found. there was no titanium particles (that were coming from the balls TiC coating in the previous ball bearings) on the track. As no anomaly was found in the MoS2 coating process and on the ball bearings coatings, and because of the MECALIB schedule criticality, it has been decided to replace the TiC/MoS2 technology by the MAPLUB fluid lubrication developed by CNES for Qualification and Flight models. In parallel investigations are still in progress to understand the reason why the MoS2 coating failed so early compared to demonstrated capabilities on previous programmes. 7. QUALIFICATION TESTS These tests will be performed on a fully representative flight model of the MECALIB. The test sequence is the following: The qualification tests will be achieved by October 1999. 8. CONCLUSION The development tests performed on MECALIB demonstrate that the design meets all the functional requirements and withstands the mechanical loads. In particular, the novelty of the kinematic architecture and its associated pointing adjustment philosophy was fully validated, with a pointing accuracy less than 1 in the orbital configuration. However, the failure that occurred during the MECALIB life test demonstrates that, even if the design is very simple and only uses widely flight proven technologies, it is necessary to verify all the requirements on a Development Model as early as possible during the studies. QM VIBRATIONS TESTS THERMAL CYCLES LIFETIME TEST VISUAL INSPECTION