Display Calibrations and Uncertainties. Christine Wall National Physical Laboratory

Similar documents
Introduction to metrology

Stray Light Rejection in Array Spectrometers

The Fundamentals of Moisture Calibration

MOY/SCMI/36 SPECIFICATION OF ACCURACY FOR A PRECISION CLINOMETER

Traceable UV-Vis Reference Materials

Introduction to the evaluation of uncertainty

Measurement method for the proficiency testing program

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012

Spectral and photopic studies for high intensity discharge (HID) lamps

Final Report 06 July 2006 Frank Wilkinson, Gan Xu, and Yuanjie Liu

Metrology Principles for Earth Observation: the NMI view. Emma Woolliams 17 th October 2017

Process Accuracy Ratio vs Process Uncertainty Ratio (Risk Mitigation: Calibration and the Customer s Process)

National Voluntary Laboratory Accreditation Program

UNCERTAINTY ANALYSIS FOR LABORATORY ACCREDITATION. Peter B. Crisp. Fluke Precision Measurement Ltd, 52 Hurricane Way, Norwich, UK

TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS

Meas ure ment: Uncertainty and Error in Lab Measurements

PRACTICAL UNCERTAINTY BUDGETS FOR SPECTRAL MEASUREMENTS OF LEDS

Uncertainty of Measurement A Concept

Humidity Calibration Solutions

Uncertainty and its Impact on the Quality of Measurement

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

Science of Stability 2017

Comparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576

This term refers to the physical quantity that is the result of the measurement activity.

Humidity Calibration Solutions

Are SQM-L measurements polarization dependent?

Measurement & Uncertainty - Basic concept and its application

Essentials of expressing measurement uncertainty

Upgrade of 5m-Bench System for Traceable Measurements of Tapes and Rules at SASO-NMCC Dimensional Laboratory

CALIBRATION. Calibration. General Principles & Theory, Equipment Considerations. Copyright Caltech 2014

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone

R SANAS Page 1 of 7

What is measurement uncertainty?

Uncertainty of Radionuclide Calibrator Measurements of 99

Morehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web: sales:

stage micrometers and calibration standards

Version 4.0 (09/2017) Version 3.0 (02/2015) Version 2.0 (03/2011) Version 1.0 (07/2007) EURAMET e.v. Bundesallee 100 D Braunschweig Germany

Document No: TR 12 Issue No: 1

CERTIFICATE OF CALIBRATION

Guidelines on the Calibration of Automatic Instruments for Weighing Road Vehicles in Motion and Measuring Axle Loads AWICal WIM Guide May 2018

Certificate of Accreditation to ISO/IEC 17025:2005

SAMPLE. Accuracy Calibration Certificate. Mettler Toledo 1900 Polaris Parkway Contact: Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC WI v.1.

What is a map? Understanding your map needs

Building your own Pizza-Box Spectroscope. *You will need to bring in a medium- sized sturdy cardboard pizza box, shoe box, or similar from home.

Accuracy Calibration Certificate

Loadcell Calibration - Evaluation of Uncertainties

SAMPLE. Accuracy Calibration Certificate. Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC ISO17025 WI v.1.0

CERTIFICATE OF CALIBRATION

Appendix C: Accuracy, Precision, and Uncertainty

Uncertainties associated with the use of a sound level meter

Know Your Uncertainty

JAB NOTE4. ESTIMATION OF MEASUREMENT UNCERTAINTY (Electrical Testing / High Power Testing) Japan Accreditation Board (JAB)

ROLE AND SIGNIFICANCE OF UNCERTAINTY IN HV MEASUREMENT OF PORCELAIN INSULATORS A CASE STUDY

Abstract. 1. Introduction

Recent Developments in Standards for Measurement Uncertainty and Traceability (An Overview of ISO and US Uncertainty Activities) CMM Seminar

ISO 376 Calibration Uncertainty C. Ferrero

Good Practice Guide No. 130

The steps to an uncertainty budget

Infrared Thermometer Calibration A Complete Guide

Calibration of Surface Contamination Monitors. UKAS Accredited Organisations Comparison. Final Report

Measurement & Uncertainty - Concept and its application

wavelength standards Jack Stone National Institute of Standards and Technology

Measurement Error PHYS Introduction

Characterisation & Use of Array Spectrometers

MEASUREMENT TRACEABILITY AND ERRORS RELATED TO FORCE MEASUREMENT

Measurement Error PHYS Introduction

Determination of particle size distribution Single particle light interaction methods. Part 4:

Accuracy Assessment. Bostomatic 32GS. for. Thomas Keating Ltd. March 2009

Uncertainty: A Reading Guide and Self-Paced Tutorial

Meeting future needs for Metrological Traceability A physicist s view

Appendix B: Accuracy, Precision and Uncertainty

AM Metrology at NPL. Stephen Brown. Tuesday 8 th December 2015

Teaching Measurement Uncertainty to Undergraduate Electronic Instrumentation Students*

A comparison of primary platinum-iridium kilogram mass standards among eighteen European NMIs

EXPERIMENT MEASUREMENT

Vocabulary of Metrology

Background Statement for SEMI Draft Document 4671 New Standard: MEASUREMENT METHOD FOR AMBIENT CONTRAST OF LIQUID CRYSTAL DISPLAYS

IMPORTANT FEATURES OF ESTIMATION & INTERPRETATION OF MEASUREMENT UNCERTAINTY. B. Bhattacharya, K. Ramani, P.H. Bhave

Traceability, validation and measurement uncertainty 3 pillars for quality of measurement results. David MILDE

Beginner's Guide to Measurement in Mechanical Engineering

2 Foundations of physics Calculation sheet. OCR Physics A. Determining uncertainty. Specification references. Learning outcomes

Introduction to 1118 Labs

Practical 1P4 Energy Levels and Band Gaps

Lab VI Light Emitting Diodes ECE 476

MEASUREMENT UNCERTAINTY PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE OCTOBER/NOVEMBER Prepared by MJ Mc Nerney for ENAO Assessor Calibration

Pre-Lab 0.2 Reading: Measurement

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement

New Thermometer Guidance Document from AASHTO Accreditation Program. Maria Knake Asphalt Binder ETG Meeting May 10, 2018

NATIONAL ASSOCIATION OF TESTING AUTHORITIES (NATA) REQUIREMENTS FOR ACCREDITATION OF ICP-MS TECHNIQUES

Renewal of the gage-block interferometer at INRIM

Chilled Mirror Hygrometry

Uncertainty in Humidity Measurement. Helping you make a better measurement.

Measurement and instrumentation. Module 1: Measurements & error analysis

Isothermal Technology Ltd

Calibration of instruments measuring broadband and spectral solar (UV) irradiance

One sided tests. An example of a two sided alternative is what we ve been using for our two sample tests:

Metallic materials Brinell hardness test. Part 3: Calibration of reference blocks

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Calibration Traceability Guidelines

METROLOGY IN CHEMISTRY MEETING. AZIZ MUKOTA Chemistry Laboratory Uganda National Bureau of Standards

Transcription:

Display Calibrations and Uncertainties Christine Wall National Physical Laboratory

Overview Why measure displays? What is measurement uncertainty? Measurement traceability NPL and UKAS What is an uncertainty, what is an error? Calibration of displays at NPL Calculating uncertainty budgets for displays Sample uncertainty budget

SAFETY COMMERCIAL ASTHETIC readability Brightness & contrast Avoid confusion of symbols Quality control Comparison between manufacturers Colour matching Customer preference

What is a measurement? Measurements tell us about a property of something. for example, how heavy or long something is. An instrument is needed to make a measurement e.g scales or a ruler. What is measurement uncertainty? Uncertainty of measurement is the doubt that exists between the result on the instrument and the true result. It is common to think that a well made clock is right but for every measurement even the most careful, there is always an element of doubt. The calculation of measurement uncertainties calculates the amount of doubt in the result.

Error or uncertainty Errors, are things that make you measurement wrong, but that you can correct for, where uncertainties cannot be corrected. There are two types of uncertainties called type A and B: Type A uncertainties affect the result randomly, so the more measurements you make and average, the closer you will be to the true value Type B uncertainties affect the result just the same, so by making more measurements you will not get closer to the result. For these you have to estimate the magnitude.

What is a traceable measurement? To ensure that a bolt bought in Buenos Aires will fit a nut from Norfolk we need to have international agreements on methods of measurement. Hence most countries have a National Metrology Institute (NMI) to hold national standards and ensure that they can trade with the rest of the world. The NMI s continually compare their measurement techniques, and agree standard methods of working to ensure that their measurements will agree. The NMI s then disseminate their techniques (and test artefacts) to companies and institutes in their own country. For example if you phone BT and set the time on your watch, BT s clock is traceable to NPL s atomic clock, and now so is your watch. Traceable measurements are required by ISO 9001 quality system. TRACEABILITY means your measurement can be referenced to a national standard.

NPL Richard the Lionheart, 1196 Throughout the realm there shall be the same yard of the same size and it should be made of Iron. 700 years later King George V opened the National Physical Laboratory (NPL), which is the United Kingdom s NMI. NPL maintains the national measurement system by holding national standards, These used to be physical a metre, a kilogram, but now are more often electrical or frequency based. Measurement is hard to avoid: your clothes come in agreed sizing, your food is measured for contaminants and weight, petrol volume, computer chip timing, electricity units, weather forecasting, TV's have 200+ measurements in manufacturing Many measurements in the UK are traceable to NPL, and hence international standards www.npl.co.uk

UKAS So NPL maintains the national measurement system, but we cant ensure every measurement is traceable to NPL which is where the United Kingdom s Accreditation Service UKAS come in. UKAS test the ability of companies, or other bodies to perform measurements accurately, and traceably to national standards. If they are, it provides an accreditation mark this is comparable to the kite mark for safety of products. UKAS is the sole national body recognised by government for the accreditation of testing and calibration laboratories, certification and inspection bodies. It is a not-forprofit company operating under a Memorandum of Understanding with the government through the Department of Trade and Industry The UKAS website can tell you if a lab is accredited, or where your nearest UKAS accredited lab is, you can guarantee that the measurements from a UKAS lab will be accurate to within the uncertainty they quote. www.ukas.co.uk

Current technique for display calibration 2. Measure a stable well characterised light source (calibrated by a UKAS accredited lab) 3. Make a correction file 5.Apply correction file to find calibrated result 4. Measure display 1. Select the instrument

The Contributions to Uncertainty Measurement instrument: 1.Wavelength uncertainty (monochromator, filters) 2.Stray light inside the instrument 3.Linearity of detector 4.Polarisation sensitivity of system (grating, optics, detector) 5.Chromatic aberration due to lenses (?) 6.Temperature sensitivity of detector Secondly there are effects due to the room of the test namely: 7. Stray light in the room (not from item under test) reaching the detector Thirdly the method of calibration (i.e. using a calibrated light source) 8. Inherited uncertainty from calibration 9. Output uniformity (position and angle, check alignment capability) 10.Output stability of lamp (check current stability using standard resistor) Finally: 11. Repeatability of the measurement

Uncertainty budget (luminance) Source of Uncertainty Luminace Type A or B Limiting Probability Divisor u 1 v i or v eff % value ± % distribution Inherited lamp uncertainty 1 B 2.0 normal = 2 2 1.00 Lamp uniformity 2 B 1.0 normal = 2 2.00 0.50 Linearity 3 B 0.0 - - 0.00 Bandwidth 4 B 0.3 rectangular = 3 1.73 0.17 Wavelength 5 B 2.4 rectangular = 3 1.73 1.39 Repeatibility 6 A 0.8 normal = 1 1 0.80 5 Combined Uncertainty normal 1.72 Expanded Uncertainty normal (k=2) 3.45 %

Summary 1) Think about what you want to measure 2) Consider anything that can affect the result, and list them 3) Where possible measure the effect to find the uncertainty 4) Calculate the uncertainty budget 5) Quote the result AND the uncertainty Still not sure: For the weak hearted read A beginners guide to uncertainty of measurement by Stephanie Bell from NPL. Not weak hearted? Try reading M3003- The uncertainty of measurement from UKAS. www.ukas.co.uk This is the definitive guide.