New Stationary Phases for Solid-Phase Extraction. Pranathi R. Perati, Rosanne Slingsby, and Carl Fisher Thermo Fisher Scientific, Sunnyvale, CA, USA

Similar documents
High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

A Total Solution for Explosives Analysis by Reversed-Phase HPLC with a Parallel HPLC System

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

HILIC Method Development in a Few Simple Steps

Monitoring Protein PEGylation with Ion Exchange Chromatography

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Thermo Scientific. Anion-Exchange Column

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency

[ Care and Use Manual ]

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

A Platform Method for Pharmaceutical Counterion Analysis by HPLC

Determination of Verapamil Hydrochloride Purity Using the Acclaim PA Column

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Hydrophilic Interaction Liquid Chromatography: Method Development Approaches

Determination of Aniline and Nitroanilines in Environmental and Drinking Waters by On-Line SPE

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

Dionex IonPac AS28-Fast-4µm column

Quantitation of Surfactants in Samples by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Simultaneous Determination of Aromatic Amines and Pyridines in Soil

Determination of Benzo(a)pyrene in Sausage and Preserved Ham

Simultaneous Determination of Pharmaceutical Peptides and Acetate Counterions by HPLC Using a Mixed-Mode Weak Anion-Exchange Column

Determination of trace anions in concentrated hydrofluoric acid

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

Thermo Scientific Dionex IonPac AS23 Anion-Exchange Column

Keywords Haloacetic acids, Water analysis, 2-D ion chromatography, ICS-3000

Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Determination of Carbonyl Compounds in Workplace Air

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites

Determination of the Composition of Natural Products by HPLC with Charged Aerosol Detection. Introduction. Black Cohosh. Corona Detector Parameters

of mass spectrometry

Supported Liquid Extraction (SLE) Guide and FAQ s. User Guide

columns IonPac AS17-C Anion Exchange Column

Quantitation and Characterization of Copper and Nickel Plating Bath Additives by Liquid Chromatography

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Advances in Sample Preparation for Accelerated Solvent Extraction

Determination of Inorganic Anions in Drinking Water by Ion Chromatography

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

columns IonPac SCS 1 Silica Cation Separator

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System

Semi-Targeted Screening of Pharmaceutically- Related Contaminants in the Thames Tideway using LC-HRMS

Metoprolol and Select Impurities Analysis Using a Hydrophilic Interaction Chromatography Method with Combined UV and Charged Aerosol Detection

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

Utility of the Charge Detector in Ion Chromatography Applications

Simultaneous Determination of Paraquat and Diquat in Environmental Water Samples by HPLC-MS/MS

columns IonPac AS22 Anion-Exchange Column

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Efficient Separations Obtained by Using Smaller Particle Size Analytical Columns with a High-Pressure Ion Chromatography System

EVOLUTE WCX Columns for Solid Phase Extraction of Quaternary Amines and Strongly Basic Compounds from Aqueous Samples

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

Determination of Trace Cations in Power Plant Waters Containing Morpholine

columns Acclaim Mixed-Mode WCX-1 for Separating Basic Molecules

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

Expanded capabilities in ammonium and amine detection

Charged Aerosol Detection and Evaporative Light Scattering Detection Fundamental Differences Affecting Analytical Performance

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

Determination of Methylamine in Drug Products

Thermo Scientific Accucore XL HPLC Columns. Technical Manual

ProPac WCX-10 Columns

Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution

anion-exchange column

columns IonPac AS18 Anion-Exchange Columns

Method Development for a Simple and Reliable Determination of PCBs in Mineral Insulating Oil by SPME-GC-ECD

Thermo Scientific Dionex IonPac AS25 Anion-Exchange Column. Minutes

Method Development in Solid Phase Extraction using Non-Polar ISOLUTE SPE Columns for the Extraction of Aqueous Samples

Fast methods for the determination of ibuprofen in drug products

Analytical Methods to Qualify and Quantify PEG and PEGylated Biopharmaceuticals

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection

Thermo Scientific HyperSep Dispersive SPE Products. Efficient sample preparation and clean-up using the QuEChERS Method

New Multi-Collector Mass Spectrometry Data for Noble Gases Analysis

Mixed-Mode, Weak Anion-Exchange, Solid-Phase Extraction Method for the Extraction of Niflumic Acid from Human Plasma

Monitoring Protein PEGylation with Ion Exchange Chromatography

Analysis - HPLC A.136. Primesep 5 µm columns B.136

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

Exploring Mixed-Mode Chromatography: Column Chemistry, Properties, and Applications

Transcription:

New Stationary Phases for Solid-Phase Extraction Pranathi R. Perati, Rosanne Slingsby, and Carl Fisher Thermo Fisher Scientific, Sunnyvale, CA, USA

Overview Purpose: To develop five new polymer-based stationary phases for large volume SPE, which include a hydrophilic reversed phase (HRPHS), strong anion exchange (SAX), strong cation exchange (SCX), weak anion exchange (WAX), and weak cation exchange (WCX). Methods: Large volume samples are pre-concentrated on a Thermo Scientific Dionex SolEx SPE cartridge, using Thermo Scientific Dionex AutoTrace 280 Solid-Phase Extraction instrument followed by HPLC analysis. Recovery ypercentages are calculated with the separation and quantification of analytes. Results: Recoveries are all greater than 95% for all the analytes. Successful recovery of analytes from matrices are shown here by automated SPE without laborious sample preparation. Introduction Five new polymeric stationary phases were developed for solid phase extraction (SPE) sample preparation of large sample volumes. The phase chemistries include a hydrophilic reversed phase material as well as mixed mode, strong and weak, anion and cation exchange. Base particles are derivatized with functional groups to achieve a useful balance between reversed phase and ion exchange capacities. The hydrophilic reversed phase material bears specific functionality to retain polyphenolic and azocontaining species through a charge transfer mechanism and H-bonding. This allows the retention of such species as humic acids along with the ability to elute other organic species using conventional reversed phase techniques. These new resins are formatted to be compatible with sample volumes larger than 20 ml in automated SPE instrumentation. TABLE 1. Recommended d applications of the five stationary ti phases. Cartridge Type HRPHS SAX Functional Group Divinylbenzene e e (DVB)/ polyvinylpyrrolidone (PVP, high) DVB-sulfonate Example Applications General e reversed e phase, polyphenolics, azo compounds Ionizable (weak) anions, carboxylic acids SCX DVB-quaternary ammonium Ionizable (weak) cations, amines WAX WCX DVB-amine DVB-carboxylate Permanent anions, inorganic and organic, sulfonates, sulfates Permanent cations, inorganic and organic, quaternary ammonium Two applications will be shown. The targets include triclosan, a popular but very hydrophobic bacteriocide, and 14 explosives regulated in U.S. EPA Method 8330. The matrices include tap water and soap water. Methods Sample Preparation TABLE 2. Sample preparation conditions. Sample Prep Conditions Sample SPE barrel Method A: Analysis of Explosives 100 ml, 2 ppb 14 explosive standard in tap water Method B: Analysis of Triclosan 100 ml 1/3000 liquid hand soap containing triclosan in Sunnyvale tap water 6 ml cartridge, 500 mg HRPHS 6 ml cartridge, 200 mg HRPHS Loading flow rate 5 ml/min 10 ml/min Elution solvent 5 ml of acetronitrile 5 ml of methanol 2 New Stationary Phases for Solid-Phase Extraction

Sample Preparation Instrument: Dionex AutoTrace 280 Solid-Phase Extraction instrument Liquid Chromatography: Thermo Scientific Dionex UltiMate 3000 HPLC system including: DGP 3600M Dual-Gradient Micro Pump SRD 3600 Integrated Solvent and Degasser Rack, 6 Channels TCC-3000 Thermostatted Column Compartment WPS-3000 Wellplate Sampler equipped with 5 μl loop PDA-3000 Photodiode Array Detector Data Analysis Software: Thermo Scientific Dionex Chromeleon Chromatography Data System, 6.80 SP4 and higher FIGURE 1. Dionex AutoTrace 280 Solid-Phase Extraction instrument; a Dionex SolEx SPE cartridge and the fluid connections are shown. TABLE 3. Liquid chromatography conditions for Methods A and B. Analytical Conditions Column Method A: Analysis of Explosives Thermo Scientific Acclaim Explosives columns E2, 3µm, 3.0 250 mm Method B: Analysis of Triclosan Flow rate 0.3 ml/min 0.21 ml/min Acclaim columns 120, 2.1 50mm Eluent 48/52 v/v MeOH/H 2 O 65/35 v/v acetonitrile/h 2 O Gradient Isocratic Isocratic Temperature 25 C 25 C Detection UV 254 nm UV 254 nm Results Method A: Automated solid-phase extraction of 14 explosives in tap water based on U.S. EPA Method 8330 using hydrophilic reversed-phase cartridge followed by HPLC with UV detection Land and groundwater have been found to be laden with explosive residues such as 2,4,6-trinitrotoluene (TNT) and associated nitroamine impurities. Target compounds include nitroaromatics, nitramines, and nitrate esters, which present health concerns due to their carcinogenic, mutagenic, and toxic effects. The U.S. EPA Method 8330, first introduced in November 1990, is the most common method for the analysis of explosives. Used here are the Dionex AutoTrace 280 Solid-Phase Extraction instrument and the Dionex SolEx SPE HRPHS cartridge to concentrate the explosives, followed by HPLC analysis using Acclaim Explosives E2 column. Thermo Scientific Poster Note PN71398-EN 1114S 3

The HRPHS SPE phase was packed into a 6mL barrel with a bed weight of 500 mg. In the SPE steps, HRPHS sorbent is preconditioned by methanol followed by water. After drying under nitrogen for 5 min, 100 ml of sample was loaded at a flow rate of 10 ml/min. Hydrophilic matrix was removed by water rinse of the cartridge. The concentrated explosives were eluted in the final step by an organic solvent such as acetonitrile. The separation of 14 explosives cyclotetramethylenetetraanitramine (HMX), cyclotrimethylenetrinitramine(rdx),1,3,5-trinitrobenzene,1,3-dinitrobenzene, nitrobenzene, 2,4,6-trinitrobenzene, tetryl, 2,6-dinitrotoluene, 2,4-dinitrotoluene, 2- nitrotoluene, 4-nitrotoluene, 3-nitrotoluene, 4-amino-2,6-dinitrobenzene, 2-amino-4,6- dionitrotoluene are shown in the chromatogram in Figure 2. Most of the explosives are well separated with the exception of RDX and 1,3,5 trinitrobenzene as well as 2,4,6 trinitrobenzene and tetryl. TABLE 4. Automated SPE conditions for 14 explosives in tap water using the Dionex SolEx SPE HRPHS cartridge and Dionex AutoTrace 280 Solid-Phase Extraction instrument. Step No. Step Description SPE Function 1 Condition cartridge with 5.0 ml of MeOH into solvent waste Condition 2 Condition cartridge with 10.0 ml of water into aqueous waste Condition 3 Dry cartridge with gas for 5 min Dry 4 Load 100.0 ml of sample onto cartridge at 10 ml/min Load 5 Rinse cartridge with 5.0 ml of water into aqueous waste Wash 6 Collect 5.0 ml fraction into sample tube using acetonitrile Elute FIGURE 2. Separation of 14 explosives in tap water using the Acclaim Explosives E2 column and Dionex SolEx SPE HRPHS cartridge for sample preparation with the Dionex AutoTrace 280 Solid-Phase Extraction instrument for automated SPE. Absorban nce [mau] 8.0 6.3 5.0 3.8 2.5 1.3 0 1 2 3 4 5 Sample Prep Instrument: Dionex AutoTrace 280 SPE for 6 ml cartridge Cartridge: 6 ml, 500 mg resin HRPHS Sample: 100 ml, 2 ppb 14 explosive Standard Flow Rate: 10 ml/min Elution: Acetonitrile 6 7 8 9 10 11 12 14 13-2 0 5 10 15 20 25 30 35 40 45 50 Retention Time in min Analytical Column: Acclaim Explosives E2, 3 µm 3.0 250 mm Mobile phase: 48/52 v/v MeOH/H 2 O Temperature: 25 C Flow rate: 0.3mL/min Inj. volume: 5µL Detection: UV at 254 nm Peaks: (EPA 8330 mix, 2 ppm each) 1. HMX 2. RDX 3. 135-trinitrobenzene 1,3,5 4. 1,3-Dinitrobenzene 5. Nitrobenenze 6. 2,4,6-trinitrobenzene 7. Tetryl 8. 2,6-Dinitrotoluene 9. 2,4-Dinitrotoluene 10. 2-Nitrotoluene 11. 4-Nitrotoluene 12. 3-Nitrotoluene 13. 4-Amino-2,6-Dinitrotoluene 14. 2-Amino-4,6-Dinitrotoluene 4 New Stationary Phases for Solid-Phase Extraction

TABLE 5. Recoveries of 14 explosives in Sunnyvale tap water by automated SPE. Peaks Recovery% RSD% HMX 108.61% 5.61% RDX 95.19% 4.01% 1,3,5-Trinitrobenzene 100.31% 3.98% 1,3-Dinitrobenzene 106.39% 5.15% Nitrobenzene 108.38% 4.98% 2,4,6-Trinitrotoluene 101.68% 5.20% Tetryl 111.58% 1.76% 26Dinitrotoluene 2,6-Dinitrotoluene 103.01% 01% 6.04% 2,4-Dinitrotoluene 104.98% 3.86% 2-Nitrotoluene 119.98% 4.27% 4-Nitrotoluene 119.63% 2.67% 3-Nitrotoluene 117.96% 4.00% 4-Amino-2,6-Dinitrotoluene 107.60% 3.11% 2-Amino-2,4-Dinitrotoluene 103.00% 5.23% Good recoveries of all 14 explosives were achieved ranging from 95% for RDX to 120% for 2-nitrotoluene, indicating that the method is robust. Method B: Automated solid-phase extraction of triclosan in tap water using hydrophilic reversed-phase cartridges and HPLC with UV detection Triclosan is a powerful antibacterial and antifungal agent that is an ingredient added to many consumer products to reduce or prevent bacterial contamination. Its safety is currently under review by the U.S. FDA based on recent studies that indicate it may cause endocrine disruption in the body and possibly lead to the emergence of drugresistant super bacteria. Triclosan-contaminated samples of tap water were simulated by adding known amounts of triclosan-containing liquid hand soap to Sunnyvale tap water. The Dionex AutoTrace 280 Solid-Phase Extraction instrument along with the Dionex SolEx SPE HRPHS cartridge were used to concentrate triclosan. TABLE 6. Automated SPE conditions for triclosan in soap water using the Dionex SolEx SPE HRPHS cartridge and Dionex AutoTrace 280 Solid-Phase Extraction instrument. Step No. Step Description Step Function 1 Condition cartridge with 5.0 ml of MeOH into solvent waste Condition 2 Condition cartridge with 10.0 ml of water into aqueous waste Condition 3 Dry cartridge with gas for 5 min Dry 4 Load 100.0 ml of sample onto cartridge at 10 ml/min Load 5 Rinse cartridge with 5.0 ml of water into aqueous waste Wash 6 Collect 5.0 ml fraction into sample tube using methanol Elute Thermo Scientific Poster Note PN71398-EN 1114S 5

FIGURE 3. Chromatogram showing the separation of triclosan using the Dionex Acclaim 120 column and the Dionex SolEx HRPHS cartridge for sample preparation with the Dionex AutoTrace 280 Solid-Phase Extraction instrument for automated SPE. Absorban nce [mau] 20 15 10 5 0 1 Sample Prep Instrument: Dionex AutoTrace 280 SPE for 6 ml cartridge Cartridge: 6 ml, 200 mg resin HRPHS Sample: 100mL 0.033% liquid antibacterial hand soap in Sunnyvale, CA tap water Flow Rate: 10 ml/min Elution: Acetonitrile or methanol Analytical Column: Acclaim column 120, 2 50 mm Eluent: 65/35 acetonitrile/water Flow Rate: 0.21 ml/min Detection: UV 254 nm Peak: 1. Triclosan 10 mg/l Final Concn Recovery: 103% for acetonitrile; 96% for methanol 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 Retention Time in Minutes Recovery of triclosan from soap water is 96% using methanol as elution solvent and 103% using acetonitrile with RSDs less than 3% in both cases. Conclusion Five polymer-based stationary yphases are developed for large-volume SPE. 14 explosives regulated in U.S. EPA Method 8330 in tap water was successfully analyzed by automated SPE using the Dionex AutoTrace 280 Solid-Phase Extraction instrument and Dionex SolEx SPE HRPHS cartridge. Triclosan in tap water was recovered by automated SPE using the Dionex SolEx SPE HRPHS cartridge in two different elution solvents without laborious sample preparation. References 1. Thermo Scientific Application Note 1081: Automated Solid Phase Extraction of Triclosan in Tap water Using Hydrophilic-Reversed Phase Cartridges and HPLC with UV Detection, Sunnyvale, CA. [Online] http://www.dionex.com/enus/webdocs/115038-an1081-spe-hplc-triclosan-tap-water-an70884_e.pdf (accessed April 17, 2014). 2. Thermo Scientific Application Note 1086: Automated Solid Phase Extraction of 14 Explosives in Tap Water Based on U.S. EPA Method 8330 Using Hydrophilic Reversed-Phase Cartridge Followed by HPLC with UV Detection, Sunnyvale, CA. [Online] http://www.dionex.com/en-us/webdocs/115039-an1086-spe-hplc- Explosives-Tap-Water-AN70897 AN70897_E.pdf (accessed April 17, 2014). PO71398-EN 1014S 6 New Stationary Phases for Solid-Phase Extraction

www.thermoscientific.com/sampleprepration 2014 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Thermo Fisher Scientific, Sunnyvale, CA USA is ISO 9001:2008 Certified. Africa +43 1 333 50 34 0 Australia +61 3 9757 4300 Austria +43 810 282 206 Belgium +32 53 73 42 41 Brazil +55 11 3731 5140 Canada +1 800 530 8447 China 800 810 5118 (free call domestic) 400 650 5118 PN71398-EN 1114S Denmark +45 70 23 62 60 Europe-Other +43 1 333 50 34 0 Finland +358 9 3291 0200 France +33 1 60 92 48 00 Germany +49 6103 408 1014 India +91 22 6742 9494 Italy +39 02 950 591 Japan +81 6 6885 1213 Korea +82 2 3420 8600 Latin America +1 561 688 8700 Middle East +43 1 333 50 34 0 Netherlands +31 76 579 55 55 New Zealand +64 9 980 6700 Norway +46 8 556 468 00 Russia/CIS +43 1 333 50 34 0 Singapore +65 6289 1190 Sweden +46 8 556 468 00 Switzerland +41 61 716 77 00 Taiwan +886 2 8751 6655 UK/Ireland +44 1442 233555 USA +1 800 532 4752