Particle Characterization of Pharmaceutical Products by Dynamic Image Analysis

Similar documents
Determination of the particle size and particle shape of pharmaceutical granules with the CAMSIZER digital image processing system

Particle Analyzer CAMSIZER XT

Introducing the Morphologi G3 ID The future of particle characterization

BIO & PHARMA ANALYTICAL TECHNIQUES. Chapter 5 Particle Size Analysis

QbD QUANTITATIVE MEASUREMENTS OF CQAS IN SOLID DOSAGE FORM UNIT OPERATIONS

Particle Measurement Analysis methods in a size range from 0.3 nm to 30 mm

Dynamic Image Analysis and pharmaceutical multi-particulates

Technical Paper. Spray Granulation Gives Solid Materials Customized. The properties of spray-granulated products can be as varied as their appearance

Particle size analysis -Chapter 3

Method Development. Creating the Perfect Standard Operating Procedure (SOP) 2007 HORIBA, Ltd. All rights reserved.

429 LIGHT DIFFRACTION MEASUREMENT OF PARTICLE SIZE

Your partner in Science!

PTG-NIR Powder Characterisation System (not yet released for selling to end users)

Setting Attainable and Practical Particle Size Specifications

TABLE OF CONTENTS. vii

ActiPix SDI300 Surface Dissolution Imaging System

Particle Size Analysis in the Pharmaceutical Industry

F r a U n h o F E r I n s T I T U T E F o r C h E M I C a l T E C h n o l o g y I C T EnErgEtic MatErials ParticlE technology

Thermal Analysis Excellence

Inform is a series of white papers designed to provide advice on material characterization issues. Mie theory The rst 100 years

Method validation for laser diffraction measurements

To Hy-Ternity and beyond. Boris Larchevêque, INDATECH

Pharmaceutical Polymers for Tablets and Capsules

University of Pécs Institute of Pharmaceutical Technology and Biopharmacy

Measuring the flow properties of powders. FT4 Powder Rheometer. freemantechnology

Technical brochure DuraLac H TABLETING DIRECT COMPRESSION ANHYDROUS LACTOSE

Powder. Aulton Chapter 8-10,14,24. Powders

Measurement of Particle Size Distributions

Particle Size and Shape Analysis in PHARMACEUTICAL INDUSTRY.

Analysis of Polymers and Plastics. Innovation with Integrity. Quality Control & Failure Analysis FTIR

CAMSIZER. Catalyst Beads, Powders, Extrudates Presentation. Dynamic Image Analysis ISO conform

Interpreting Your PSA Results

White Paper Optimization of Dry Powder Particle Size Analysis

Philip E. Plantz, PhD Application Note. SL-AN-36 Revision A. Provided By: Microtrac, Inc. Particle Size Measuring Instrumentation

MicroBrook Eyetech. Particle Size, Shape, and Concentration Analyzer. Particle Sizing. Particle Concentration. Particle Shape

Analysis of Polymers and Plastics. Innovation with Integrity. Quality Control & Failure Analysis FT-IR

MORPHOLOGI 4 RANGE. Automated imaging for advanced particle characterization

Particle Analysis at the Touch of a Button. Litesizer series

Particle Characterization of Volcanic Ash by Dynamic Digital Image Analysis

The Elzone II Particle Count and Size Analyzer

The Coulter Principle for Outlier Detection in Highly Concentrated Solutions

Implementing Fundamental Pharmaceutical Science and Materials/Engineer Expertise in Scale-up

Morphologically directed chemical identification Coupling particle characterization by morphological imaging with Raman spectroscopy

Pharmaceutical Characterisation. Dr. Lidia Tajber and Dr. Krzysztof Paluch School of Pharmacy and Pharmaceutical Sciences, Trinity College Dublin

INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE

Optimal conditions to prepare fine globular granules. with a multi-functional rotor processor

Introduction to Fourier Transform Infrared Spectroscopy

Pharmaceutics and Pharmaceutical Technology

A basic guide to particle characterization

Introduction to Fourier Transform Infrared Spectroscopy

Particle Size Analysis with Differential Centrifugal Sedimentation. Marc Steinmetz Sales & Support Manager CPS Instruments Europe

Seeing the Nano-scale: Nanoparticle Tracking Analysis HVM MNT Expo 2006 Oxford. Jeremy Warren CEO, NanoSight Ltd

FACULTY OF PHARMACY. M. Pharmacy I Semester (Suppl.) Examination, November 2015 (Common To All) Subject: Pharmaceutical Analytical Techniques

A Basic Guide to INFORM WHITE PAPER. Abstract:

NIR Chemical Imaging as a Process Analytical Tool

Formulation of Low Dose Medicines - Theory and Practice

PARTICLE SHAPE FACTORS AND THEIR USE IN IMAGE ANALYSIS PART II: PRACTICAL APPLICATIONS As Published in GXP, Autumn 2011 (Vol15/No4) Eric Olson

DVS INTRINSIC. The easy-to-use solution to complex water sorption challenges.

Introduction Fine Particles

What Happens When Rheological Properties Change? Looking into Rheological Properties with Simultaneous Collection of Microscopic Images

Sympatec GmbH, System-Partikel-Technik, Burgstätterstraße 6, D Clausthal-Zellerfeld, Germany

Application of Micro-Flow Imaging (MFI TM ) to The Analysis of Particles in Parenteral Fluids. October 2006 Ottawa, Canada

Holographic Characterization of Agglomerates in CMP Slurries

ANALYSIS OF LOW DENSITY PARTICLES USING DIFFERENTIAL CENTRIFUGAL SEDIMENTATION

DRY METHOD DEVELOPMENT FOR LASER DIFFRACTION PARTICLE SIZE MEASUREMENTS

Discriminating between polymorphs of acetaminophen using Morphologically-Directed Raman Spectroscopy (MDRS)

YOUR PARTNER IN PARTICLE CHARACTERIZATION

BASIC CONCEPTS C HAPTER 1

No turning round: abusing µct as an X-ray transmission microscope

Litesizer Sheet 1. [Home]

INTRODUCTION. Russia

Laser diffraction and automated imaging: complementary techniques for nasal spray development

Automated multi-vapor gravimetric sorption analyzer for advanced research applications

Available online at ScienceDirect. Procedia Engineering 102 (2015 )

The Thermal Challenge

Introduction to Differential Sedimentation

Unit I Unit Operations

Chapter 12. Nanometrology. Oxford University Press All rights reserved.

Determination of Solubility Parameters using Inverse Gas Chromatography

Introduction to Dynamic Light Scattering for Particle Size Determination

An introduction to particle size characterisation by DCS:

Particle Technologies for Food & Beverage Applications

PARTICLE SIZE ANALYTICAL RANGES AND APPLICABILITY. m mm (10-6 m) nm (10-9 m)

My Dosage Design Tool Predicts, Process Analytical Technology Confirms Film Thickness of Fully Formulated Aqueous and Organic Ethylcellulose Coating

next generation Welcome to the Dispersion options Smarter particle sizing Your sample, expertly processed and prepared Particle size

Particle Analysis at the Touch of a Button. Litesizer series

Recommendations by Experts on the Required Parameters for Microplastics Monitoring in the Ocean As of 12 June 2018

Chapter 7 Mixing and Granulation

VESSEL PRESSURES ABOVE THE CRITICAL POINT ON THE PARTICLE FORMATION IN A SUPERCRITICAL ANTISOLVENT PROCESS.

CONTROLLED PRECIPITATION OF ACTIVE PHARMACEUTICAL INGREDIENTS EMPLOYING SUPERCRITICAL FLUIDS: SCALE-UP CONSIDERATIONS

S2 PICOFOX. Innovation with Integrity. Spectrometry Solutions TXRF

What Happens When Rheological Properties Change? Looking into Rheological Properties with Simultaneous Collection of Microscopic Images

ESPRIT Feature. Innovation with Integrity. Particle detection and chemical classification EDS

FORCE TENSIOMETER K100 THE HIGH-END SOLUTION FOR MEASURING SURFACE AND INTERFACIAL TENSION

Water Soluble Polymers For Industrial Water Treatment Applications

Is NIST SRM2806b Responsible for the Sudden Increase in Particle Counts you have Seen on your Oil Analysis Reports?

Morwet naphthalene sulfonates

ENVR 116 Aerosol Technology Laboratory Session Fall 2005

The Dangers of Dilution When Measuring Water Based Pharmaceutical Suspensions and How Nanoflex Overcomes these Obstacles

Transcription:

WHITE PAPER Particle Characterization of Pharmaceutical Products by Dynamic Image Analysis Retsch Technology GmbH Retsch-Allee 1-5 42781 Haan, Germany Phone: +49 24/2333-0 E-Mail: technology@retsch.com www.retsch-technology.com Introduction Higher resolution, better sensitivity, and excellent reproducibility: this is how Dynamic Image Analysis (DIA) improves the characterization of powders and granulates in both quality control and R&D. This white paper outlines how particle size distribution and particle shape are determined by DIA, illustrated by application examples from the pharmaceutical industry which prove the superiority of the method over sieve analysis and laser diffraction. Traditional techniques for particle size distribution analysis of pharmaceutical samples are sieve analysis, microscopy or laser diffraction. These methods are established in the pharmacopoeia and are used in pharmaceutical laboratories all over the world. With the introduction of Dynamic Image Analysis (DIA) as an alternative method, it is now possible to measure particle size and shape of powders, granulates, pellets and suspensions of particles > 1 µm. A number of trials clearly demonstrate the advantages of DIA compared to the traditional particle sizing methods. Many pharmaceutical companies have already recognized the potential of this method and added this technology to their research and quality control labs. Typical application examples are: Micronized active ingredients (APIs) and excipients Pharmaceutical granulates (e. g. tableting mixtures) Size and shape of pellets Monitoring of coating processes Characterization of crystalline APIs and excipients www.retsch-technology.com whitepaper 1

The principle of Dynamic Image Analysis (DIA) Just like microscopes, DIA analyzers consist of an illumination unit, objective lenses and a camera system. Particles are usually detected in transmitted light and the images contain shadow projections of the particles. Unlike microscopy however, the objective of DIA is the detection and evaluation a vast number of particles in a very short analysis time. This is achieved by generating a flow of particles which is photographed by high-speed cameras (Fig. 1). The particle flow can be in free fall for pourable solids, in an air flow generated by a Venturi nozzle for agglomerated powders, or even in a liquid suspension. Depending on the instrument and the application, between and 3 images are acquired and evaluated in real time! An automated microscope will typically need - minutes to capture enough particles to calculate a meaningful size distribution. DIA systems evaluate some hundreds of thousands or even millions of particles within a few minutes. The resulting size distribution is therefore based on a much larger basis, eliminating errors arising from sampling and poor statistics. Additionally, the measurement range of a DIA system is more than times larger than that of a microscope a significant advantage as many samples contain particles from the micron to the millimeter range. Fig 1. Two state of the art DIA instruments: CAMSIZER P4 (left) and CAMSIZER X2 (right). The P4 is optimized for the fast analysis of pourable bulk materials in a size range from µm to mm, whereas the X2 model works in a size range from 0.8 µm to 8 mm. The X2 features several options for sample dispersion, e. g. air-jet nozzle with adjustable pressure and a wet dispersion unit. Image Analysis: What you see is what you get! Imaging techniques provide a direct approach to particle size analysis. The basic idea is simple: What you see is what you get. Based on pictures of individual particles, automatic software algorithms determine size and morphology. Particle length and particle width are directly accessible, as shown in Fig. 2. This shows the versatility of DIA, particularly in combination with shape analysis which runs simultaneously to the size measurement. Some shape parameters are explained in Fig. 3. Fig. 2 Selection of basic size parameters used in image analysis. The size distributions are based on width (red), length (blue) or equal area diameter (green). www.retsch-technology.com whitepaper 2

Fig. 3 Selection of basic shape parameters used in image analysis. Example 1: Detection of agglomerates during pellet production The production of pharmaceutical pellets is typically done by granulation, extrusion with subsequent spheronization or coating. The desired result is a narrow and homogeneous particle size distribution of round particles. In the granulation and coating processes, the formation of agglomerates is an unwanted side effect. Agglomerates can have a negative impact on product properties; they can lead, for example, to changes in the solubility or the release rate of the active ingredients. Therefore, the amount of agglomerates is usually strictly controlled for each product batch. The CAMSIZER is able to detect percentages of agglomerates as low as 0.05 %. Neither laser diffraction nor sieve analysis are suitable methods to reliably detect such minor percentages. Due to the measuring principle laser particle analyzers require a minimum concentration of 2 % to detect agglomerates or undersized particles, such as dust fractions. Smaller amounts may be simply ignored by the software. Particle shape is also an important factor in this context. Elongated particles, for example, can neither be detected with laser diffraction nor with sieve analysis. Fig. 4 Reliable detection of oversized agglomerates by the CAMSIZER P4: Approx. 0.21g of agglomerates, which is equal to about particles, have been repeatedly added to a sample of 21 g to achieve mixtures in the range of 1 % to 11 %. The percentage of oversized particles at 0.7 mm exactly matches the calculated mixing percentages. www.retsch-technology.com whitepaper 3

Example 2: Measuring coating thickness The various coating steps when producing pellets require precise analysis of the coat thickness of the applied layers. The total dosage of the drug layer is defined by its thickness; the thickness of other functional layers can control the drug release rate and dissolution process. The drug release is inversely proportional to the thickness of the polymer membrane layers, and proportional to the surface area of the particles. With Dynamic Image Analysis it is possible to reliably determine variations in the coating thickness of less than 1 micron. The method combines both high resolution and excellent statistics as a great number of particles is analyzed in a very short time. Sieve analysis, however, only offers low resolution, as typically only very few sieve sizes are available in the narrow size range of coated granules. Traditional microscopebased technologies such as SEM or static light microscopy offer excellent resolution but only for very few particles. [2] [3] [4] Fig. 5: Particle size distributions after different process steps during coating, ranging from the small nonpareil starting pellets to the final product after polymer coating. The detection of smallest changes in the average particle diameter allows the precise measurement of the coating thickness. The width of the size distributions enables a characterization of the homogeneity of the coating process. Example 3: Tabletting Mixture In tableting, the goal is usually to generate a product that has low friability and high tensile strength [5]. By analyzing the size distribution, it is possible to predict the suitability of a mixture for tabletting. The starting products are often granulates with a wide size distribution and irregular particle shape. Both properties contribute to the compactability and have an influence on the mechanical properties of the mixture. This is why dynamic image analysis provides valuable information besides particle size to characterize a bulk material and relate these to its mechanical properties. Note that in the example below the DIA result largely matches those of sieve analysis. Hence, it is possible (and advisable) to replace the inaccurate, low resolution and time-consuming sieving method with a faster, more accurate and highly automated technique. This mixture has a size distribution from µm to 1.5 mm and the CAMSIZER X2 DIA analyzer is able to determine the entire distribution without any hardware adjustment. Subsequent examination of individual particles gives a thorough understanding of the material at hand. p3 [%] Fig. 6 Size distribution of a tabletting mixture. CAMSIZER X2 analysis (red curve) and sieve analysis (black *) provide comparable results. Some example images on the right show that the particle shape is irregular. 45 35 25 15 5 0 0 0 0 0 00 10 xc_min [µm] 0 www.retsch-technology.com whitepaper 4

Example 4: Size and shape analysis of starch Starch is a common excipient in pharmaceutical applications. It comes from various vegetable sources and in various size ranges. The example shows the results of two different starch samples. The distribution is slightly different but the median (d) value is almost identical. Considering the particle shape, it becomes very clear that sample 1 consists of round, compact grains whereas sample 2 contains a significant amount of fibres! Fig. 7 Size distribution of two starch samples with the CAMSIZER X2. Both distributions have the same median of 41 µm. 0 0 xc_min [µm] elongated compact Fig. 8 Particle shape analysis with the CAMSIZER X2 of the two starch samples showing significant differences. 0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 b/l www.retsch-technology.com whitepaper 5

A common particle analysis technique is laser diffraction. This method is fast, robust and suitable for routine analysis. Its greatest flaw, however, is the fact that it does not consider the real geometry of the particles but rather relates every measurement signal to the diameter of an equivalent sphere. Thus, it totally neglects particle shape. An additional drawback is that laser diffraction is evaluating a scattering pattern that is simultaneously generated by a collective of particles. Small amounts of oversize and undersize are lost due to the low sensitivity of the method. Fig. 9 shows that the result of laser diffraction analysis correlates with the size definition xarea of the Digital Image Analysis. Fig. 9: Particle size distribution of a starch sample. DIA with the CAMSIZER X2, size definition xc min (red), xarea (green), xfe max (blue). Laser diffraction result (black *). 0 x [µm] Example 5: Unsievable APIs Particle size analysis of micronized APIs can be a nightmare for any lab technician since these powders often show very disadvantageous properties: the material can be sticky, cohesive, charged or highly agglomerated. All this makes an effective sieving process virtually impossible. Furthermore, below 0 µm size resolution and accuracy are limited. Regular vibratory sieving is not applicable here; air-jet sieving is a suitable method but only provides sieve cuts. DIA analyzers like the CAMSIZER X2 provide efficient dispersion systems to handle problematic samples. Fig. CAMSIZER X2 dispersion options. Above: X-Jet module for dry dispersion in an air-flow. Dispersion pressure can be adjusted from kpa to 4 kpa. Below: Liquid dispersion with the X-Flow module. The unit can handle water, alcohol, non-polar organic solvents and plant oil. www.retsch-technology.com whitepaper 6

The powder is either suspended in a suitable liquid (water, alcohol, non-polar organic solvent or plant oil) or the particles are separated in an air-flow generated by a Venturi-nozzle. The dispersion pressure can be adjusted to achieve efficient separation of particles without any unwanted milling effect. Fig. 11 Agglomerated API on a 63 µm sieve after 2 minutes of air-jet sieving (photo). Due to its cohesive nature the sample is unsievable. With the air-jet dispersion of the CAMSIZER X2 the separation of particles can be achieved. The higher the pressure, the finer the result. Perfect dispersion is established at 0 kpa. CAMSIZER X2 results at kpa (orange), kpa (red), 1 kpa (blue). 0 kpa (brown) and 0 kpa (green). 0 0 1 1 1 1 xc_min [µm] www.retsch-technology.com whitepaper 7

3. CONCLUSION Modern DIA analyzers like RETSCH Technology s CAMSIZER P4 or CAMSIZER X2 are not only capable of replacing traditional particle sizing techniques like sieve analysis or laser diffraction; they also provide a wealth of additional information which the other methods lack. If the correct size definition is selected, the results are virtually identical and product specifications can remain unchanged. Users of DIA benefit from highest reproducibility, improved accuracy and sensitivity as well as from short analysis time and additional shape information. References: 1. Cornelia Krüger, Markus Thommes, Peter Kleinebudde, Spheronization mechanism of MCC II-based pellets, Powder Technology, 238:176-187, April 13 2. Grant Heinicke and Joseph B. Schwartz, Particle Size Distribution of Inert Spheres and Pelletized Pharmaceutical Products by Image, Pharmaceutical Development and Technology Vol. 9, No. 4, pp. 359-367, 04 3. Grant Heinicke, Frank Matthews, Joseph B. Schwartz, The Effects of Substrate Size, Surface Area, and Density on Coat Thickness of Multi-Particulate Dosage Forms, Pharmaceutical Development and Technology, 1:85-96, 05 4. Grant Heinicke, Joseph B. Schwartz, Assessment of Dynamic Image Analysis as a Surrogate Dissolution Test for a Coated Multiparticulate Product, Pharmaceutical Development and Technology, 11:3-8, 06 5. Carl Moritz Wagner, Miriam Pein, Jörg Breitkreutz, Roll compaction of mannitol: Compactability study of crystalline and spray-dried grades, International Journal of Pharmaceutics, 453(2) June 13 www.retsch-technology.com whitepaper 8