Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution

Similar documents
Enhancement of Linearity and Response in Charged Aerosol Detection

Direct Analysis of Surfactants using HPLC with Charged Aerosol Detection

Determination of the Composition of Natural Products by HPLC with Charged Aerosol Detection. Introduction. Black Cohosh. Corona Detector Parameters

Quantitation of Surfactants in Samples by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Quantitation and Characterization of Copper and Nickel Plating Bath Additives by Liquid Chromatography

Charged Aerosol Detection and Evaporative Light Scattering Detection Fundamental Differences Affecting Analytical Performance

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

Determination of Verapamil Hydrochloride Purity Using the Acclaim PA Column

Analytical Methods to Qualify and Quantify PEG and PEGylated Biopharmaceuticals

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

Use of Charged Aerosol Detection as an Orthogonal Quantification Technique for Drug Metabolites in Safety Testing (MIST)

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

Advances in Sample Preparation for Accelerated Solvent Extraction

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Monitoring Protein PEGylation with Ion Exchange Chromatography

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Fast methods for the determination of ibuprofen in drug products

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Metoprolol and Select Impurities Analysis Using a Hydrophilic Interaction Chromatography Method with Combined UV and Charged Aerosol Detection

Clinical Toxicology. Biomass Component Extraction: The uneaten cooked plant specimen was prepared for

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Determination of Carbonyl Compounds in Workplace Air

Simultaneous Determination of Pharmaceutical Peptides and Acetate Counterions by HPLC Using a Mixed-Mode Weak Anion-Exchange Column

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD)

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Semi-Targeted Screening of Pharmaceutically- Related Contaminants in the Thames Tideway using LC-HRMS

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Paul A. Ullucci, Bruce Bailey, Ian N. Acworth, Christopher Crafts, Marc Plante Thermo Fisher Scientific, Chelmsford, MA, USA

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

Comprehensive Neurochemical Profiling of Brain Tissue Samples

Keywords Haloacetic acids, Water analysis, 2-D ion chromatography, ICS-3000

Discover what you re missing

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

Introduction to Fourier Transform Infrared Spectroscopy

Hydrophilic Interaction Liquid Chromatography: Method Development Approaches

Direct Determination of Small Organic Acids in Sea Water by IC-MS

Efficient Separations Obtained by Using Smaller Particle Size Analytical Columns with a High-Pressure Ion Chromatography System

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Varian 385-LC. ELSD for Low Molecular Weight Compounds

Varian 380-LC. ELSD for Routine HPLC and GPC Analysis. Key Benefits

Simultaneous Determination of Aromatic Amines and Pyridines in Soil

A Platform Method for Pharmaceutical Counterion Analysis by HPLC

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter

Introduction to Fourier Transform Infrared Spectroscopy

APPLICATION NOTE Goal. Introduction. Keywords

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

New Stationary Phases for Solid-Phase Extraction. Pranathi R. Perati, Rosanne Slingsby, and Carl Fisher Thermo Fisher Scientific, Sunnyvale, CA, USA

Easy Method Transfer from HPLC to RSLC with the Dionex Method Speed-Up Calculator

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

Analysis of Polyphenols in Saffron Petal Extracts with UHPLC/UV/MS

Automated and accurate component detection using reference mass spectra

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

Detection of Mycotoxins in Corn Meal Extract Using Automated Online Sample Preparation with Liquid Chromatography-Tandem Mass Spectrometry

Thermo Scientific Accucore XL HPLC Columns. Technical Manual

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

No Chromophore? No Problem! Universal HPLC Detection. The world leader in serving science

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

A Total Solution for Explosives Analysis by Reversed-Phase HPLC with a Parallel HPLC System

IDENTIFICATION OF STEROIDS IN COSMETIC PRODUCTS BY TLC AND HPLC 1 02/12/2005 ACM 007 A. THIN LAYER CHROMATOGRAPHY (TLC)

Thermo Scientific HyperSep Dispersive SPE Products. Efficient sample preparation and clean-up using the QuEChERS Method

[ Care and Use Manual ]

Selecting Detectors for Compounds with No Optical Absorbance

Introduction. Chapter 1. Learning Objectives

HPLC-ELSD DETERMINATION OF SODIUM LAURYL SULPHATE AND POLYSORBATE IN NEBIVOLOL DRUG PRODUCT AND DIFFERENT FORMULATION PRODUCTS

HILIC Method Development in a Few Simple Steps

Supported Liquid Extraction (SLE) Guide and FAQ s. User Guide

HPLC analysis of VFAs, Furfural and HMF

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

Transcription:

Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution Marc Plante, Bruce Bailey, Christopher Crafts, Ian N. Acworth Thermo Fisher Scientific, Chelmsford, MA, USA

Overview Purpose: To provide a sensitive analytical method for resolving and quantifying triterpenoids Methods: A Thermo Scientific Acclaim C30 column provided high resolution of the different triterpenoids, and the Thermo Scientific Dionex Corona charged aerosol detector (CAD ) enabled the sensitive and direct quantitation of these analytes. Results: Resolution between ursolic and oleanolic acid was greater than any reports in literature. No analyte derivatization was required, and limits of quantitation were < 2 ng on column. A sample of a dried spice and quantitation of ursolic acid in apple peel are provided. Introduction Triterpenoids are a class of naturally occurring compounds found in plants, and their beneficial effects on the immune system and in the treatment of cancer are currently being investigated. Some triterpenoids of interest include glycyrrhetinic, betulinic, oleanolic, and ursolic acids, where ursolic has been recently reported to exist at high concentration in apple skin. 1 Separation techniques using gas chromatography (GC), and micellar electrophoretic capillary chromatography, are either laborious (derivatization for GC) or have issues with reproducibility. With high performance liquid chromatography (HPLC) separation, a problem often encountered is poor resolution between oleanolic and ursolic acids, which tend to co-elute in many liquid chromatographic (LC) methods. One difficulty in making accurate determinations by UV is that these compounds do not have a strong chromophore moiety and sensitivity is compromised. A recent LC paper using a C18 column and UV absorption at 210 nm reported limits of detection of 50 and 135 ng on column for oleanolic and ursolic acids, respectively, with a resolution of 1.53 between these acids. 2 As reported in this work, a novel method using a C30 HPLC column and charged aerosol detection, with a resolution of 2.73 between oleanolic and ursolic acids is described. The limit of quantitation was also improved with an LOQ of 2 ng on column (o.c.) achieved for all four analytes. An example of triterpenoids in apple skin and basil is also demonstrated. The charged aerosol detector is a sensitive, mass-based detector, especially wellsuited for the determination of non-volatile and many semi-volatile analytes. As shown in Figure 1, the detector uses nebulization to create aerosol droplets. The mobile phase evaporates in the drying tube, leaving analyte particles, which become charged in the mixing chamber. The charge is then measured by a highly sensitive electrometer, providing reproducible, nanogram-level sensitivity. This technology has greater sensitivity and precision than evaporative light scattering (ELS), and it is simpler and less expensive to operate than a mass spectrometer (MS). Typical characteristics of chromatography with charged aerosol detection include: low-nanogram on column (o.c.) sensitivity, over four orders of magnitude of dynamic range, and high precision results, typically less than two percent of peak area RSD. Analyte response is also largely independent of chemical structure, providing clear relationships among different analytes in a sample FIGURE 1: Schematic and functioning of charged aerosol detection. 1 3 10 2 4 8 6 9 5 7 1. Liquid eluent enters from HPLC system 2. Pneumatic nebulization occurs 3. Small droplets enter drying tube 4. Large droplets exit to drain 5. Dried particles enter mixing chamber 6. Gas stream passes over corona needle 7. Charged gas collides with particles and charge is transferred 8. High mobility species are removed 9. Charge is measured by a highly sensitive electrometer 10. Signal transferred to chromatographic software 2 Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution

Methods Sample Preparation Standards were dissolved in ethanol, except for ursolic acid, which was dissolved in acetone. Samples were sonicated in methanol/chloroform (1:1) for one hour, centrifuged at 13,000 g for 5 minutes, and diluted with one equivalent of methanol to assist in maintaining retention and analyte peak shape. Liquid Chromatography HPLC System: Thermo Scientific Dionex UltiMate 3000 SD HPLC Column: Acclaim C30, 5 µm, 250 x 4.6 mm Column Temperature: 30 C Mobile Phase A: 1 w/v% Ammonium acetate in water Mobile Phase B: Acetonitrile/methanol (750 : 250) Flow Rate: 1.0 ml/min Detector: Corona ultra RS Nebulizer Temperature: 15 C Filter Setting: 4 Sample Temperature: Ambient Injection Volume: 5.0 µl Gradient: Time (min) %A %B Curve 0.0 92 8 -- 20.0 20 80 1 20.0 92 8 5 25.0 92 8 5 Data Analysis All HPLC chromatograms were obtained and compiled using Thermo Scientific Dionex Chromeleon 6.8 SR11 Chromatography Data Station. Results. Method Development The structures of the four triterpenoids are shown in Figure 2. These analytes have a very similar structure, often differing only by the position of a methyl group, making chromatographic separation difficult. One publication focused on the separation of oleanolic and ursolic acids, investigated nine different conditions, and at best achieved a resolution of 1.65 for these two analytes, respectively, using a C18 column. 2 FIGURE 2: Chemical structures of the four triterpenoids included in this study. Ursolic Acid, R1 = CH 3, R2 = H, R3 =COOH Oleanolic Acid, R1 = H, R2 = CH 3, R3 =COOH Glycyrrhetinic Acid, R1=H, R2=COOH, R3=H, ketone at (*) Betulinic Acid A better separation for these compounds was achieved using a C30 column. The C30 solid phase is known to provide alternate selectivity over other column materials. 3 Using the analytical conditions given above, this column gave a resolution of 2.73 for oleanolic and ursolic acids. 3 Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution

Standards were dissolved in ethanol (ursolic acid in acetone) to provide 4 to 2000 ng (o.c.) amounts. A chromatogram at 500 ng o.c. is shown in Figure 3, and calibration curves for the four triterpenoids (analyzed in triplicate) are presented in Figure 4. The peak area relative standard deviations for amounts greater than 30 ng o.c. was 5%. The data were fit to inverted second-order polynomials, yielding correlation coefficients, R 2, greater than 0.998 for the four analytes. FIGURE 3: Analysis of the four triterpenoids at 500 ng on-column. 80.0 70.0 pa 60.0 50.0 40.0 30.0 20.0 10.0 0.0 min -10.0 0.0 2.0 4.0 6.0 8.0 10.0 12.0 14.0 16.0 18.0 20.0 22.0 24.0 25.0 FIGURE 4: Calibration curves for glycyrrhetinic, betulinic, oleanolic, and ursolic acids, from 4 to 2000 ng on-column. 2500 Amount on Column (ng) 2000 1500 1000 500 0 0 5 10 15 20 25 30 Corona Peak Area (pa * min) Glycyrrhetinic Betulinic Oleanolic Ursolic Sensitivity values were determined using a signal-to-noise (S/N) of 3.3 for limit of detection (LOD) and 10.0 for limit of quantitation (LOQ). Using results from the 4 ng o.c. analysis, the LOQ was calculated to be < 2 ng o.c., and the LOD < 1 ng o.c, as shown in Table 2. Calibration precision (RSD) was < 7% for all analytes. Table 2. Calibration precision and sensitivity values for triterpenoids by HPLC and charged aerosol detection. Analyte %RSD LOD (ng o.c.) LOQ (ng o.c.) Glycyrrhetinic acid 2.62 0.6 1.7 Betulinic acid 2.40 0.6 1.7 Oleanolic acid 4.82 0.5 1.5 Ursolic acid 6.82 0.5 1.5 4 Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution

Sample Analysis Two samples were analyzed for triterpenoids: dried basil and apple peel. A qualitative sample preparation was performed on dried basil, using 19 mg/ml of basil in methanol/chloroform (1:1), diluted with one equivalent of methanol. A chromatogram, overlaid with a 32 ng o.c. standard chromatogram, is presented in Figure 5, and shows the presence of all four analytes, with ursolic and oleanolic acids being the two major triterpenoids, as reported by GC-FID. 4 FIGURE 5. HPLC chromatogram of basil extract (blue) and 32 ng on-column triterpenoid standard solution (black). 12.0 pa 10.0 8.0 6.0 4.0 2.0 min 0.0 8.00 9.00 10.00 11.00 12.00 13.00 14.00 15.00 16.00 17.00 The peel (3.037 g) from a red delicious apple was finely chopped and placed in a 30 ml centrifuge tube to which 12 ml of methanol/chloroform (1:1) was added. The mixture was sonicated for 1 hour, and then centrifuged at 13,000 g for five minutes. To a 500 µl aliquot of supernatant, 500 µl of methanol was added, mixed, and the resulting solution was analyzed. A chromatogram of apple peel extract, overlaid with 500 ng o.c. standard, is shown in Figure 6. The four triterpenoids in the standard solution were found in the sample, along with a number of other peaks, which are unidentified. The amount of ursolic acid in the peel was found to be 0.14 (mass-percent) which is in good agreement to literature values (0.15%). 5 Figure 6. HPLC chromatogram of apple peel extract (blue) and 500 ng oncolumn triterpenoid standard solution (black). 80.0 pa 70.0 60.0 50.0 40.0 30.0 20.0 10.0 0.0 min -10.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0 16.0 17.0 18.0 5 Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution

Method selectivity is the highest in the reported literature, with baseline separation of oleanolic and ursolic acids. Sensitivity is high, with an LOQ of <2 ng o.c., giving relative LOQ values of 0.0003 mass-percent. Method selectivity was not evaluated for other analytes within the ursane, oleanane, and lupane classes of triterpenes. Clearly, other compounds are present in the chromatography, and these can be easily identified with mass spectrometry, with which the chromatography is completely compatible. Corona charged aerosol can be used as the primary method of quantitation for the determination of these compounds, once identification has been established by an orthoganol approach, namely mass spectrometry. Conclusion A reversed-phase HPLC method was developed for the analysis of triterpenoids using the selectivity of a 3 µm C30 column and the sensitivity and precision of the Corona ultra RS detector. The combination of acetonitrile and methanol with small-particle size C30 solid phase provides excellent separation of analytes. The method has unmatched resolution and detection limits for triterpenoids. Analysis is direct and samples do not require derivatization as is required for GC. References 1. Liang, Z.; Jiang, Z.; Fong, D.; HAO, Z. Journal of Food and Drug Analysis, 2009, 17(2), 69-77. 2. Yamaguchi, H.; Noshita, T.; Kidachi, Y.; Umetsu, H.; Hayashi, M.; Komiyama, K.; Funayama, S.; Ryoyama, K. Journal of Health Science, 2008, 54(6) 654 660. 3. Snyder, L.; Kirkland, J. Introduction to modern liquid chromatography, 3 rd edition; John Dolan, Ed.; Wiley & Sons: New Jersey, 2010; p 249. 4. Jäger, S.; Trojan, H.; Kopp, T.; Laszczyk, M.N.; Scheffler, A. Molecules, 2009, 14, 2016-2031. 5. He, X. and Liu, R.H. J. Agric. Food Chem., 2007, 55, 4366 4370. 6 Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution

www.thermofisher.com/dionex 2016 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. U.S./Canada (847) 295 7500 Brazil (55) 11 3731 5140 Austria (43) 1 616 51 25 Benelux (31) 20 683 9768 (32) 3 353 42 94 Denmark (45) 36 36 90 90 France (33) 1 39 30 01 10 Germany (49) 6126 991 0 Ireland (353) 1 644 0064 Italy (39) 02 51 62 1267 Sweden (46) 8 473 3380 Switzerland (41) 62 205 9966 United Kingdom (44) 1276 691722 Australia (61) 2 9420 5233 China (852) 2428 3282 India (91) 22 2764 2735 Japan (81) 6 6885 1213 Korea (82) 2 2653 2580 Singapore (65) 6289 1190 Taiwan (886) 2 8751 6655 PN70037_E 07/16S