Nitrate plus Nitrite and Nitrite in Seawater by Segmented Flow Analysis (SFA)

Similar documents
Chloride by Flow Injection Analysis (FIA)

Standard Operating Procedure for the Analysis of Nitrate/Nitrite and Nitrite in Fresh Waters CCAL 31B.1

Hexavalent Chromium by Flow Injection Analysis (FIA)

Alkalinity, USEPA by Flow Injection Analysis (FIA)

Bromide by Flow Injection Analysis (FIA)

Total Kjeldahl Nitrogen by Gas Diffusion and Flow Injection Analysis (FIA)

DETERMINATION OF NITRATE AND NITRITE BY ZONE FLUIDICS

Method for Nitrite determination on Low-Range Samples

O Sodium Bicarbonate, NaHCO 3 Ammonium Molybdate Tetrahydrate, (NH 4. O 4 Sodium Chloride, NaCl Ascorbic Acid, C 6 H 8 O 6

1) Alkaline Persulfate Digestion for TN, TDN and TDPTotal Dissolved Nitrogen. a) Scope and Application

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Sulfate by Flow Injection Analysis (FIA)

Nitrogen, ammonia, colorimetry, salicylate-hypochlorite, automated-segmented flow

Total Phosphorus, USEPA by Flow Injection Analysis (FIA)

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Standard Operating Procedure for the Analysis of Silica in Fresh Waters CCAL 32A.2

METHOD 9200 NITRATE. 1.2 The applicable range of concentration is 0.1 to 2 mg NO -N per liter. 3 of sample.

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Hach Company TNTplus TM Phosphorus Spectrophotometric Measurement of Phosphorus in Water and Wastewater

Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1

SECTION D.2 AMMONIA NITROGEN

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

Total Kjeldahl Nitrogen (TKN), USEPA by Flow Injection Analysis (FIA)

METHOD DETERMINATION OF SULFATE BY AUTOMATED COLORIMETRY. Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

Ohio EPA Total (Extracellular and Intracellular) Microcystins - ADDA by ELISA Analytical Methodology Ohio EPA DES Version 2.

EnAlgae SOP sheet: Silicate - SFA. 1. Purpose This procedure is to analyse seawater and freshwater for silicate, SiO2.

Standard Operating Procedure VUSP M

Glossary of Common Laboratory Terms

Standard Methods for the Examination of Water and Wastewater

Flow Solution FS 3700 Automated Chemistry Analyzer

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

-fr-[*-] ^'A: Approval: /": f. z /zs- /, e. Determination of Nitrate/Nitrite as Nitrogen. EffectiveDate: 3 it

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

METHOD 9035 SULFATE (COLORIMETRIC, AUTOMATED, CHLORANILATE)

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows

Standard Operating Procedure VUSP N

ALLOWAY METHOD OUTLINE

METHOD DETERMINATION OF TOTAL CYANIDE BY SEMI-AUTOMATED COLORIMETRY

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

DR/4000 PROCEDURE NITRATE. 2. The display will show:

METHOD 9012 TOTAL AND AMENABLE CYANIDE (COLORIMETRIC, AUTOMATED UV)

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

Spectrophotometric Determination of Ferrocyanide in Effluents

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right.

DETERMINATION OF NITRATE IN 2M KCl SOIL EXTRACTS BY FLOW INJECTION ANALYSIS NITRATE REDUCTASE

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV

Phenolphthalein and Total Alkalinity Method to 4000 mg/l as CaCO 3 Digital Titrator

Nitrogen, Total Inorganic

METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD)

Method to 0.50 mg/l NH 3 N Powder Pillows

Scope and application: For wastewater, seawater, drinking water, surface water and process water.

Method 630. The Determination of Dithiocarbamate Pesticides in Municipal and Industrial Wastewater

DETERMINATION OF NITRATE IN BRACKISH WATERS BY FLOW INJECTION ANALYSIS NITRATE REDUCTASE

Method to 0.50 mg/l NH 3 N Powder Pillows

Flow Solution FS 3700 Automated Chemistry Analyzer

Product Safety Reference Manual. Book 5 - Laboratory Policies and Procedures C

METHOD DETERMINATION OF PHOSPHORUS BY SEMI-AUTOMATED COLORIMETRY

L Mott SALT WATER AQUACULTURE. Model AQ-4/AQ-5 Code / test kit instruction manual

NITROGEN, AMMONIA, High Range, Test N Tube

Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY

METHOD DETERMINATION OF AMMONIA NITROGEN BY SEMI-AUTOMATED COLORIMETRY

Sodium Chloride - Analytical Standard

United States EPA Method 415.3

3. Chemical industry. Because of their modular design, the instruments in the TOC-L series can be equipped for any possible measurement

Tex-620-J, Determining Chloride and Sulfate Contents in Soil

METHOD 9252A CHLORIDE (TITRIMETRIC, MERCURIC NITRATE)

O'DONEL HIGH SCHOOL CHEMISTRY 2202

Unit 3 Chemistry - Volumetric Analysis

DR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water.

The sample is not preserved and analysed as soon as possible, at least within 24 hours after collection. Store the sample at 4 C till analysis.

Chlorine, Total. USEPA DPD Method 1 Method to mg/l as Cl 2 Chemkey Reagents. Test preparation. Before starting.

Chlorphosphonazo Colorimetric Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow. Instrument Adapter Sample cell

Nutrient Analysis, Nitrate, Nitrite, Silicate, Phosphate and Ammonium

National Research Council Institute for Ecosystem Study Verbania Pallanza - Italy

QAM-I-116 Preparation of Labware

Cadion Method Method to 0.30 mg/l Cd TNTplus 852

Chemical Analysis Instruction Book

Timberline Ammonia-001. Determination of Inorganic Ammonia by Continuous Flow Gas Diffusion and Conductivity Cell Analysis

Relative Solubility of Transition Elements

METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC)

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

METHOD 9012B TOTAL AND AMENABLE CYANIDE (AUTOMATED COLORIMETRIC, WITH OFF-LINE DISTILLATION)

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM

ASTM Designation: D Standard Test Method for Determination of Iodine Number of Activated Carbon

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

Method for estimation of iodine in urine

Standard Operating Procedure for the Analysis of Fresh Water Samples for Orthophosphorus CCAL 34B.2

Parameter Method Range # of Tests Code Page. Acidity (as % Oleic acid) titration % acidity 6 HI

DOUBLE DISPLACEMENT REACTIONS. Double your pleasure, double your fun

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials

Standard Operating Procedure for GLNPO Specific Conductance: Conductivity Bridge

LOW-RANGE TOTAL CHLORINE

Procedures for Preparing Reagents and Media used in Firearm and Tool Mark Examinations

Exercise 6: Determination of Hardness of Water

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure

Transcription:

Methodology Nitrate plus Nitrite and Nitrite in Seawater by Segmented Flow Analysis (SFA) (Cartridge Part #A002603) 1.0 Scope and Application 1.1 This method is used to determine the concentration of nitrate (NO 3 ) plus nitrite (NO 2 ) or nitrite singly in seawater. 1.2 The Method Detection Limit (MDL) of this method is 0.007 µmoles/l. The applicable range of the method is 0.02 40 µmoles/l nitrate plus nitrite and nitrite. The range may be extended to analyze higher concentrations by sample dilution. 2.0 Summary of Method 2.1 Nitrate is reduced quantitatively to nitrite by cadmium metal. Nydahl (Reference 15.11) provides a good discussion of nitrate reduction by cadmium metal. The nitrite formed, in addition to any nitrite originally present in the sample, is diazotized with sulfanilamide (SAN) and subsequently coupled with Ν-(1-naphthyl)ethylenediamine dihydrochloride (NED). The resulting highly colored azo dye is colorimetrically detected at 540 nm (Reference 15.7). A calibration curve allows for accurate quantitation of the detected nitrite. 2.2 Nitrite singly may be measured by performing the same analysis as in Section 2.1 but without the cadmium reduction. Without the cadmium, nitrate is not reduced to nitrite and is not detected since only nitrite forms the azo dye. 2.3 Both nitrate and nitrite may be measured simultaneously by using a two channel flow analyzer. One channel is used to measure nitrate plus nitrite (Section 2.1), while the second channel is used to measure nitrite only (Section 2.2). Using WinFLOW software, the results of the nitrite analysis may be subtracted from the results of the nitrate plus nitrite analysis, thus providing quantitative nitrate results. 2.4 For additional references, Patton (Reference 15.8) and Fox (Reference 15.4) provide discussions of the mechanisms and kinetics of the color forming reactions used in this method. 2.5 The quality of the analysis is assured through reproducible calibration and testing of the Segmented Flow Analysis (SFA) system. 1

2.6 A general flow diagram of the SFA system is shown below (see Section 17.0 for a detailed flow diagram). NED Reagent SAN Reagent Nitrogen To Waste 540 nm 60 C 60 C Buffer Open Tube Cadmium Reactor (OTCR) Sample 3.0 Definitions Definitions for terms used in this method are provided in Section 16.0, Glossary of Definitions and Purposes. 4.0 Interferences 4.1 Turbid samples may interfere with the photometric detector s ability to measure the true absorbance of the sample. Filter turbid samples prior to analysis. 4.2 Iron, copper, and other metals may interfere with the analysis by binding with the nitrate and/or nitrite in the sample, thus blocking the color formation reaction. The imidazole buffer eliminates this interference. 4.3 Samples that are outside the functional ph range of the ammonium chloride buffer may affect the results obtained from this method. Adjust the ph of these samples to within a range of 5 9 using either concentrated hydrochloric acid (HCl) or sodium hydroxide (NaOH). 4.4 Oil and grease will coat the cadmium surface, thus reducing its reduction efficiency. Extract samples containing large concentrations of oil and grease with an appropriate organic solvent (Reference 15.1). 4.5 Sulfide in the presence of cadmium will form cadmium sulfide (CdS), which will precipitate from solution. Samples containing sulfide cannot be determined by this method without first removing the sulfide by precipitation with cadmium salts (Reference 15.9). 4.6 Chlorine may reduce the reduction efficiency of the cadmium reactor. Samples that may contain residual chlorine should be tested for reduction efficiency through the analysis of Matrix Spike/ Matrix Spike Duplicate (MS/MSD) samples (Section 9.3). When necessary, dechlorinate samples with sodium thiosulfate (Na 2 S 2 O 3 ). 4.7 Method interferences may be caused by contaminants in the reagents, reagent water, glassware, etc., which may bias the results. Care should be taken to keep all such items free of contaminants. 2

4.8 Norwitz and Keliher (References 15.1 and 15.2) have compiled a comprehensive study of interferences in the spectrophotometric analysis of nitrite. 5.0 Safety 5.1 The toxicity or carcinogenicity of each compound or reagent used in this method has not been fully established. Each chemical should be treated as a potential health hazard. Exposure to these chemicals should be reduced to the lowest possible level. 5.2 For reference purposes, a file of Material Safety Data Sheets (MSDS) for each chemical used in this method should be available to all personnel involved in this chemical analysis. The preparation of a formal safety plan is also advisable. 5.3 The following chemicals used in this method may be highly toxic or hazardous and should be handled with extreme caution at all times. Consult the appropriate MSDS before handling. 5.3.1 Cadmium, Cd (FW 112.40) 5.3.2 Cupric Sulfate Pentahydrate, CuSO 4 5H 2 O (FW 249.61) 5.3.3 Hydrochloric Acid, concentrated, HCl (FW 36.46) 5.3.4 Imidazole, C 3 H 4 (FW 68.08) 5.3.5 Magnesium Sulfate Heptahydrate, MgSO 4 7H 2 O (FW 246.48) 5.3.6 N-(1-naphthyl)ethylenediamine Dihydrochloride, C 12 H 14 2HCl (FW 259.18) 5.3.7 Potassium Nitrate, KNO 3 (FW 101.11) 5.3.8 Potassium Nitrite, KNO 2 (FW 85.11) 5.3.9 Sodium Chloride, NaCl (FW 58.44) 5.3.10 Sodium Hydroxide, NaOH (FW 40.00) 5.3.11 Sulfanilamide, C 6 H 8 O 2 S (FW 172.21) 5.4 Unknown samples may be potentially hazardous and should be handled with extreme caution at all times. 5.5 Proper personal protective equipment (PPE) should be used when handling or working in the presence of chemicals. 5.6 This method does not address all safety issues associated with its use. The laboratory is responsible for maintaining a safe work environment and a current awareness file of OSHA regulations regarding the safe handling of the chemicals specified in this method. 3

6.0 Apparatus, Equipment, and Supplies 6.1 Segmented Flow Analysis (SFA) System (OI Analytical Flow Solution IV) consisting of the following: 6.1.1 Model 502 Multichannel Peristaltic Pump 6.1.2 Random Access (RA) Autosampler 6.1.3 Expanded Range (ER) Photometric Detector with 10-mm path length flowcell and 540-nm optical filter 6.1.4 Data Acquisition System (PC or Notebook PC) with WinFLOW software 6.1.5 Nitrate+Nitrite and Nitrite in Seawater Cartridge (Part #A002603) 6.2 Sampling equipment Sample bottle, amber glass, with polytetrafluoroethylene (PTFE)-lined cap. Clean by washing with detergent and water, rinsing with two aliquots of reagent water, and drying by baking at 110 150 C for a minimum of one hour. 6.3 Standard laboratory equipment including volumetric flasks, pipettes, syringes, etc. should all be cleaned, rinsed, and dried per bottle cleaning procedure in Section 6.2. 7.0 Reagents and Calibrants 7.1 Raw Materials 7.1.1 Brij -35, 30% w/v (Part #A21-0110-33) 7.1.2 Cupric Sulfate Pentahydrate, CuSO 4 5H 2 O (FW 249.61) 7.1.3 Deionized Water (ASTM Type I or II) 7.1.4 Hydrochloric Acid, concentrated, HCl (FW 36.46) 7.1.5 Imidazole, C 3 H 4 (FW 68.08) 7.1.6 Magnesium Sulfate Heptahydrate, MgSO 4 7H 2 O (FW 246.48) 7.1.7 N-(1-naphthyl)ethylenediamine Dihydrochloride, C 12 H 14 2HCl (FW 259.18) 7.1.8 Potassium Nitrate, KNO 3 (FW 101.11) 7.1.9 Potassium Nitrite, KNO 2 (FW 85.11) 7.1.10 Sodium Chloride, NaCl (FW 58.44) 4

7.1.11 Sodium Hydroxide, NaOH (FW 40.00) 7.1.12 Sulfanilamide, C 6 H 8 O 2 S (FW 172.21) 5