Direct Analysis of Surfactants using HPLC with Charged Aerosol Detection

Similar documents
Quantitation of Surfactants in Samples by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution

Enhancement of Linearity and Response in Charged Aerosol Detection

Determination of the Composition of Natural Products by HPLC with Charged Aerosol Detection. Introduction. Black Cohosh. Corona Detector Parameters

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

Quantitation and Characterization of Copper and Nickel Plating Bath Additives by Liquid Chromatography

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Analytical Methods to Qualify and Quantify PEG and PEGylated Biopharmaceuticals

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

Charged Aerosol Detection and Evaporative Light Scattering Detection Fundamental Differences Affecting Analytical Performance

Advances in Sample Preparation for Accelerated Solvent Extraction

Determination of Verapamil Hydrochloride Purity Using the Acclaim PA Column

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

A Platform Method for Pharmaceutical Counterion Analysis by HPLC

Use of Charged Aerosol Detection as an Orthogonal Quantification Technique for Drug Metabolites in Safety Testing (MIST)

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

New Developments in Surfactant Analysis Using HPLC

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Simultaneous Determination of Pharmaceutical Peptides and Acetate Counterions by HPLC Using a Mixed-Mode Weak Anion-Exchange Column

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

Discover what you re missing

HILIC Method Development in a Few Simple Steps

Varian 385-LC. ELSD for Low Molecular Weight Compounds

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Hydrophilic Interaction Liquid Chromatography: Method Development Approaches

Monitoring Protein PEGylation with Ion Exchange Chromatography

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection

Metoprolol and Select Impurities Analysis Using a Hydrophilic Interaction Chromatography Method with Combined UV and Charged Aerosol Detection

APPLICATION NOTE Goal. Introduction. Keywords

Determination of Trace Cations in Power Plant Waters Containing Morpholine

No Chromophore? No Problem! Universal HPLC Detection. The world leader in serving science

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD)

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection

New Stationary Phases for Solid-Phase Extraction. Pranathi R. Perati, Rosanne Slingsby, and Carl Fisher Thermo Fisher Scientific, Sunnyvale, CA, USA

Analysis - HPLC A.136. Primesep 5 µm columns B.136

Thermo Scientific Acclaim Surfactant Plus Column

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Agilent 385-ELSD Evaporative Light Scattering Detector

Direct Determination of Small Organic Acids in Sea Water by IC-MS

Semi-Targeted Screening of Pharmaceutically- Related Contaminants in the Thames Tideway using LC-HRMS

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

Agilent 385-ELSD Evaporative Light Scattering Detector

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Keywords Haloacetic acids, Water analysis, 2-D ion chromatography, ICS-3000

Determination of Carbonyl Compounds in Workplace Air

columns Acclaim Mixed-Mode WCX-1 for Separating Basic Molecules

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Comprehensive Neurochemical Profiling of Brain Tissue Samples

Charged Aerosol Detection 101

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

HPLC-ELSD DETERMINATION OF SODIUM LAURYL SULPHATE AND POLYSORBATE IN NEBIVOLOL DRUG PRODUCT AND DIFFERENT FORMULATION PRODUCTS

columns IonPac AS22 Anion-Exchange Column

Expanded capabilities in ammonium and amine detection

Varian 380-LC. ELSD for Routine HPLC and GPC Analysis. Key Benefits

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

[ Care and Use Manual ]

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Paul A. Ullucci, Bruce Bailey, Ian N. Acworth, Christopher Crafts, Marc Plante Thermo Fisher Scientific, Chelmsford, MA, USA

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Agilent 1260 Infinity Evaporative Light Scattering Detector (ELSD)

Determination of an anionic fluorochemical surfactant in a semiconductor etch bath

columns IonPac SCS 1 Silica Cation Separator

Efficient Separations Obtained by Using Smaller Particle Size Analytical Columns with a High-Pressure Ion Chromatography System

columns IonPac AS17-C Anion Exchange Column

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Analysis of pharmaceuticals using HPLC with evaporative light scattering detection

Impact of Instrument Dispersion on Performance of HPLC Capillary Columns

Determination of trace anions in concentrated hydrofluoric acid

A Total Solution for Explosives Analysis by Reversed-Phase HPLC with a Parallel HPLC System

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Corona Veo the Next Generation Charged Aerosol Detector

Mixed-Mode, Weak Anion-Exchange, Solid-Phase Extraction Method for the Extraction of Niflumic Acid from Human Plasma

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Determination of Bromate in Bottled Mineral Water Using the CRD 300 Carbonate Removal Device

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency

Pharmaceutical Analysis of API and Counter Ions in Complex Formulations in a Single Injection

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Easy Method Transfer from HPLC to RSLC with the Dionex Method Speed-Up Calculator

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory

LC III: HPLC. Originally referred to as High-Pressure Liquid Chromatography. Now more commonly called High Performance Liquid Chromatography

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

Fast methods for the determination of ibuprofen in drug products

Utilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography. Tips and Techniques to Optimize ELSD and MS based Purification

Transcription:

Direct Analysis of Surfactants using HPLC with Charged Aerosol Detection Marc Plante, Bruce Bailey, Ian N. Acworth, Christopher Crafts, Thermo Fisher Scientific, Chelmsford, MA, USA

Overview Purpose: Examples of HPLC methods for the deteration of surfactants using the universal Thermo Scientific Dionex Corona charged aerosol detector with the Thermo Scientific Acclaim Surfactant Plus column were evaluated. Methods: HPLC methods, using buffered mobile phases and different elution programs are outlined. Results: The method was used to generate chromatograms of a mixture of anionic, cationic, and non-ionic surfactants, and samples of Span (80, 83, and 85), TWEEN 80 and 85, Pluronic F68, and a laundry detergent. Introduction Surfactants are a diverse group of chemicals whose structures vary widely but typically contain an oil-soluble hydrocarbon chain and a water-soluble ionic group. Surfactants can be categorized based upon their structure and include nonionic, anionic, and cationic classes. They have widespread use as detergents in shampoos and cleaning products, ion pairing agents used in chromatography, and complex dispersants used to treat oil spills. Many of these commercial surfactants are mixtures of members of a homologous series, and such mixtures can be defined using LC. Chromatographic approaches can separate the molecules on the basis of carbon chain length, chain branching or positional isomer distribution. Surfactants typically do not contain a UVchromophore so are usually measured using RP-HPLC with non-suppressed or suppressed mode conductivity or indirectly using photometric detection. Charged aerosol detection can measure any non-volatile, and many semi-volatile compounds, typically to low ng sensitivity. Furthermore, as response is similar for all compounds and independent of chemical structure, charged aerosol detection is ideal for measurement of surfactant species. Generally, the reproducibility for methods using charged aerosol detection is better than 2% RSD. Sensitive methods are described herein for the analysis of various surfactant classes including anionic alkyl sulfonates (lauryl sulfate), cationic quaternary aes (laurylmethylbenzylae), non-ionic block copolymer (Pluronic F-68), and complex mixtures of oil dispersants (Span 80). The Corona ultra RS charged aerosol detector (CAD ) is a sensitive, mass-based detector, especially well-suited for the deteration of non-volatile and many semivolatile analytes. As shown in Figure 1, the detector uses nebulization to create aerosol droplets. The mobile phase evaporates in the drying tube, leaving analyte particles, which become charged in the mixing chamber. This technology has greater sensitivity and precision than evaporative light scattering detectors (ELSD), and it is simpler to operate than a mass spectrometer (MS). Typical characteristics of chromatography with charged aerosol detection include: low-nanogram on-column (o.c.) sensitivity, over four orders of magnitude of dynamic range, and high precision results, typically less than two percent of peak area RSD. Analyte response is also largely independent of chemical structure, providing clear relationships among different analytes in a sample FIGURE 1. Schematic and functioning of charged aerosol detection. 1 3 10. 2 4 5 8 6 7 9 1. Liquid eluent enters from HPLC system 2. Pneumatic nebulization occurs 3. Small droplets enter drying tube 4. Large droplets exit to drain 5. Dried particles enter mixing chamber 6. Gas stream passes over corona needle 7. Charged gas collides with particles and charge is transferred 8. High mobility species are removed 9. Charge is measured by a highly sensitive electrometer 10. Signal transferred to chromatographic software 2 Direct Analysis of Surfactants using HPLC with Charged Aerosol Detection

Methods Sample Preparation Samples were dissolved in a isopropanol, isopropanol/water (1:1), or acetonitrile/water (1:1) to a concentration of 10 or 20 mg/ml. Liquid Chromatography HPLC System: Thermo Scientific Dionex UltiMate 3000 x2 Dual RSLC system HPLC Column: Acclaim Surfactant Plus, 4.6 250 mm Column Temperature: 30 C (Gradient, Inverse Gradient, Detergent) 40 C (Pluronic F68) Mobile Phase A: mm Ammonium acetate, ph 5.4 Mobile Phase A1: 50 mm Ammonium acetate, ph 5 in water/acetonitrile (9:1) Mobile Phase B: n-propyl alcohol Mobile Phase C: Acetonitrile Mobile Phase C1: 50 mm Ammonium acetate, ph 5 in acetonitrile/methanol/water (4:5:1) Detector: Corona ultra RS Nebulizer Temperature: ambient Filter Setting: 0 Sample Temperature: Ambient Injection Volume: 5.0 µl Gradients: Gradient Elution Inverse Gradient Time () Pluronic F68 Time Flow Rate () (ml/) %A %B -5 0.6 95 5 0 0.6 95 5 5 0.6 30 70 10 0.6 5 95 Detergent Time Flow Rate () (ml/) %A1 %C1-5 0.6 0 20 0 0.6 0 20 2 0.6 0 20 12 0.6 10 90 17 0.6 10 90 Data Analysis All HPLC chromatograms were obtained and compiled using Thermo Scientific Dionex Chromeleon 7.1 SR 1. The inverse gradient was calculated using the Inverse Gradient Calculator using the parameters of void volume difference and setting for maximum acetonitrile content, which increases analyte response. Results Sample Analysis Flow Rate (ml/) Using the conditions above, a mixture of eight surfactants, consisting of five anionic, two non-ionic, and one cationic surfactant were analyzed using the single-pump gradient elution program and the inverse gradient program (both pumps, as programmed with the gradient elution and the inverse gradient conditions). As shown in Figure 2, the column clearly separates the different surfactant classes, including separation of components within more complex surfactants. The use of the inverse gradient provides two benefits: it flattens the baseline and it eliates increases in relative response factors that are associated with increases in nebulization efficiency resulting from increased organic content of the mobile phase. This yields more aesthetic chromatograms which are less error-prone towards peak integration and, more importantly, relative response factors across the gradient are more consistent which allows for improved results on mass-percent values in the sample. Two surfactants, TWEEN 80* (polyoxyethylene-sorbitan monooleate) and TWEEN 85* (polyoxyethylene-sorbitan trioleate), were dissolved in isopropanol and analyzed using the gradient elution conditions. The two chromatograms are overlaid, as shown in Figure 3. Note that not only are the subcomponents of each TWEEN distinguished, but also TWEEN 85 elutes later than the TWEEN 80 due to the greater amount of oleate moieties contained within the polymer. *used in COREXIT 9500 1 %A %C -5 1.0 98 2 0 1.0 98 2 15 1.0 5 95 20 1.0 5 95 Time () Flow Rate (ml/) %A %C -4 1.0 0 0 1.0 0 1 1.0 0 16 1.0 93 7 21 1.0 93 7 3 Direct Analysis of Surfactants using HPLC with Charged Aerosol Detection

FIGURE 2. HPLC with charged aerosol detection chromatogram of a surfactant mix in water/acetonitrile (1:1), containing cationic, anionic, and neutral surfactants. Single-pump eluent gradient conditions in black, and dual-pump inverse gradient conditions in blue. 64.7 pa 60.0 55.0 50.0 45.0 40.0 35.0 30.0 25.0 20.0 15.0 10.0 5.0 0.0 Xylenesulfonates Laurylpyridinium Laurylmethylbenzyl ae Triton X- -5.0-9.7 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0 Cetylbetaine Decylsulfate Dodecylsulfate Linear alkylsulfonates FIGURE 3. Overlaid chromatograms of TWEEN 80 (black) and TWEEN 85 (blue), 20 mg/ml in isopropanol using single gradient conditions. 140 130 pa 120 110 90 80 70 TWEEN 85 60 50 TWEEN 80 40 30 20 10 0-10 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0 16.0 17.0 18.0 19.0 20.0 Six Span surfactants were analyzed, using the single-pump gradient elution parameters shown above. Span-80* (sorbitan monooleate), -83 (sorbitan sesquioleate), and -85 (sorbitan trioleate) were dissolved in isopropanol at a concentration of 20 mg/ml. The similarity between the Span 80 and 83 chromatograms reflects the similarity in composition: Span 83 is similar to Span 80, except that it contains 50% more oleate than Span 80. This may be reflected in the slight increase of the later eluting portions of the Span 83 chromatogram. Taking this difference further, the triolein form of Span 80, called Span 85, contains the greatest amount of later-eluting, hydrophobic material than the other two, which is clearly seen in the chromatogram overlays in Figure 4. A common surfactant in pharmaceutical/biotechnology products is Pluronic F68 (polyoxyethylene-polyoxypropylene block copolymer). Like other surfactants, Pluronic lacks a chromophore, and its polymeric nature makes reversed-phase chromatography difficult, usually resulting in peaks with broad tailing. One recent paper uses a restricted access media column with a step gradient and ELSD. 2 The use of step gradients typically causes baseline disruptions which can interfere with analytical results, especially at low levels. 4 Direct Analysis of Surfactants using HPLC with Charged Aerosol Detection

FIGURE 4. HPLC-CAD chromatogram overlays of Span 80, 83, and 85 at 20 mg/ml in isopropanol. 220 200 pa 180 160 140 120 80 60 40 20 Span 85 Span 83 Span 80 0 0.00 1.25 2.50 3.75 5.00 6.25 7.50 8.75 10.00 11.25 12.50 13.75 15.00 16.25 17.50 18.75 20.00 Use of the Acclaim Surfactant Plus column also generated acceptable chromatography for Pluronic F68, using the Pluronic F68 conditions described in Methods. Triplicate injections of Pluronic F68 at concentrations of 20 mg/ml diluted sequentially to 0.31 mg/ml (or 1.6 µg o.c.) in isopropanol/water (1:1), is shown in Figure 5. Precision was good, with peak area percent RSD values of 0.61 (20 mg/ml) to 6.5 (0.31 mg/ml). A calibration plot, fitted to an inverted second-order polynomial for concentrations, between 0.31 and 10 mg/ml, is shown in Figure 6. The correlation coefficient, r 2, was 0.9995. The signal to noise ratio (S/N) at 1.6 µg o.c. was 139, for an limit of quantitation (LOQ) value of 115 ng o.c., based on a S/N ratio of 10. FIGURE 5. Chromatogram overlays of Pluronic F68, in triplicate, from 20 to 0.31 mg/ml in isopropanol/water (1:1) 300 pa 280 20 mg/ml 1 Pluronic F68-4.695 260 240 10 mg/ml 220 200 180 5 mg/ml 160 140 2.5 mg/ml 120 1.25 mg/ml 80 60 0.63 mg/ml 40 0.31 mg/ml 20 0-40 -50 0.00 1.25 2.50 3.75 5.00 6.25 7.50 8.75 10.02 FIGURE 6. Calibration plot and inverted polynomial fit for Pluronic F68 from 0.31 to 10 mg/ml. 12.0 PluronicF68 External CAD_1 mg/ml 11.0 10.0 9.0 8.0 7.0 6.0 5.0 4.0 3.0 2.0 1.0 0.0 pa* 0.0 10.0 20.0 30.0 40.0 50.0 60.0 70.0 80.0 5 Direct Analysis of Surfactants using HPLC with Charged Aerosol Detection

An off-the-shelf laundry detergent product was diluted in water at a concentration of 50 mg/ml; 5 µl was analyzed using the detergent conditions described in Methods. As can be seen in Figure 7, this detergent appears to contain alcohol ethoxylates, two main varieties of sulfonates, and a variety of more hydrophilic materials that elute before 3 utes. FIGURE 7. Chromatogram of a laundry detergent 160 pa Sulfonates 125 Ions, glycerol, alcohols 75 Alcohol ethoxylates 50 25 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 Conclusion Combining the use of the Surfactant Plus column with the universality, reproducibility, and sensitivity of the Corona charged aerosol detector enables a simplified approach to chromatography method development. Methods used gradient elution for fast, quantitative results, while providing resolution for sample characterization. The methods shown are capable of separating and quantifying many of the typical classes of surfactants, from the simple surfactant to the complex, polymeric surfactants and mixtures. The use of the inverse gradient enabled more consistent response throughout the gradient. Analysis times were less than 21 utes. Similar surfactants were differentiated consistent with their composition. Other surfactants have been analyzed, including Aerosol OT (docusate sodium) as a single peak, as well as Span 20 and 60. References 1. Nalco North America Home Page. http://www.nalco.com/news-andevents/4297.htm (accessed February 29, 2012) 2. Chung H.H.; Zhou C.; Khor H.K.; Qiu J. J. Chromatogr. A. 2011, 1218(15), pp. 2106-13. 6 Direct Analysis of Surfactants using HPLC with Charged Aerosol Detection

www.thermofisher.com/dionex 2016 Thermo Fisher Scientific Inc. All rights reserved. Span is a trademark of Croda International PLC. Pluronic is a registered trademark of BASF Corporation. TWEEN is a registered trademark of Uniqema, a business unit of ICI Americas Inc. COREXIT is a registered trademark of the Nalco Company. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. U.S./Canada (847) 295 7500 Brazil (55) 11 3731 5140 Austria (43) 1 616 51 25 Benelux (31) 20 683 9768 (32) 3 353 42 94 Denmark (45) 36 36 90 90 France (33) 1 39 30 01 10 Germany (49) 6126 991 0 Ireland (353) 1 644 0064 Italy (39) 02 51 62 1267 Sweden (46) 8 473 3380 Switzerland (41) 62 205 9966 United Kingdom (44) 1276 691722 Australia (61) 2 9420 5233 China (852) 2428 3282 India (91) 22 2764 2735 Japan (81) 6 6885 1213 Korea (82) 2 2653 2580 Singapore (65) 6289 1190 Taiwan (886) 2 8751 6655 PN70055_E 07/16S