Paper # # Verification of Coriolis Flow

Similar documents
Fine Tuning Coriolis Flow Meter Calibrations Utilizing Piece- Wise Linearization

KNOWLEDGE. Allow Smart Meter Verification to Reduce your Proving and Proof-Test Costs. Abstract. Introduction

KNOWLEDGE

Custody Transfer Measurement and Calibration Round Robin Testing for Natural Gas with Coriolis Meters

CORIOLIS MASS FLOW METERS FOR GAS AND LIQUID MEASUREMENT. Michael Keilty (Presented by Hubbard Donald) Endress + Hauser Flowtec AG

Rocky Mountain Measurement Seminar Denver August 20, 2015 Coriolis Meters

Evaluating New Generation Vibrating Tube Sensor for Density Measurement under Process Conditions

ISM Evolution. Elscolab. Nederland BV

Guidelines for the Selection and Operation of Provers with Micro Motion ELITE Coriolis Flow Meters

Coriolis Mass Flowmeter. (MCM Series) 100% Customer Satisfaction MAXIFLO TM. Coriolis Mass Flowmeter Series MCM

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

ALASTAIR MCLACHLAN Applications Cameron

Describe Coriolis Mass Flowmeters

Coriolis in the Marine Industry TRICOR Coriolis Technology. All rights reserved.

Applied Fluid Mechanics

Process Gas Flow Measurement and Performance Evaluation API TECHNICAL REPORT 2571 SECOND EDITION DRAFT, FEBRUARY 2018

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

Applied Fluid Mechanics

FINAL DRAFT: October 25, 2010 API TECHNICAL REPORT FUEL GAS MEASUREMENT. FINAL DRAFT: October 25, 2010 Page 0

Manual of Petroleum Measurement Standards Chapter 5 Metering

Commissioning and Testing of AVR and PSS

Pro-V Multivariable Flowmeter Model M22 In-line Vortex

Armin Rasch * Abstract

DS 300. Electronic Pressure Switch. with IO-Link interface. Stainless Steel Sensor. accuracy according to IEC 60770: 0.35 % FSO.

Control Loop Investigations

Linearity of Coriolis Flowmeters at Full Scale Value. Dr. Jesús Aguilera - Endress + Hauser Paul Ceglia - Endress + Hauser

SITRANS F flowmeters. SITRANS F US System information SITRANS F US Ultrasonic flowmeters 4/181

Improving the Accuracy of Dynamic Vibration Fatigue Simulation

TS02 - Pre-Instructional Survey

Innovative Solutions from the Process Control Professionals

ACCEPTANCE PROCEDURE FOR ROMET ROTARY METER BODIES AND/OR MECHANICAL NON-CONVERTING AND CONVERTING MODULES

Pro-V Multivariable Flowmeter Model M23 Insertion Vortex

NORTH SEA FLOW MEASUREMENT WORKSHOP 2004 In. St Andrews, Scotland

E v a l u a t i o n C e r t i f i c a t e

Reference 1 Thermal Energy Metering. Abstract: Some aspects of the selection and installation requirements for thermal energy metering are canvassed.

Calibration Verification Linearity in the Clinical Lab. Did I Pass or Fail? 2017 ASCLS New Jersey

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

Safety analysis and standards Analyse de sécurité et normes Sicherheitsanalyse und Normen

Fundamentals of Particle Counting

Best selling products overview of company AURA a.s.

Ultrasonic Gas Flow Meter

APPLICATION OF DENSITOMETERS TO LIQUID MEASUREMENT Class #

DEPARTMENT OF MECHANICAL ENIGINEERING, UNIVERSITY OF ENGINEERING & TECHNOLOGY LAHORE (KSK CAMPUS).

Instrumentation & Data Acquisition Systems

An Essential Requirement in CV Based Industrial Appliances.

Minute Impulse Clock Controller I01DN

P R O G N O S T I C S

ISOCON-6. 24V AC or DC POWERED ISOLATING SIGNAL CONVERTER

MODEL: CR002, CR003, CR004. ADC12 Display shield. Liquid (Density range: 0.3 to 2.0g/mL) Operating temperature range

Advisory Circular. U.S. Department of Transportation Federal Aviation Administration

Monitoring Radar Mechanical Drive Systems FAA's choice of monitoring solutions

Review of Anemometer Calibration Standards

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012

1.4 System Accuracy* B. G. LIPTÁK (1982, 1995, 2003) DEFINITIONS OF TERMS

POWER UNDERSTANDING MEASUREMENT UNCERTAINTY IN DP FLOW DEVICES

The Fundamentals of Moisture Calibration

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone

Densitometers Calibration Sorted? But what about in Service? Dr Norman F Glen NEL

Lab 11 Simple Harmonic Motion A study of the kind of motion that results from the force applied to an object by a spring

BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple

Perseverance. Experimentation. Knowledge.

Public Outreach Meeting. Proposed Revision of Onshore Orders 3, 4, and 5

Flow Monitoring Technologies in Oil and Gas. Don Ford, Technical Specialist Spartan Controls Ltd.

A Case for Periodic Calibration or Verification of RTDs

Large Rocket Engine Environments

24 volts (0.25 amps current-limited)

REPORT ON API 21.1 EGM STANDARD Brent E. Berry ABB-Totalflow

PRACTICAL CONTROL OF DYNAMIC METERING SYSTEM AND DATA ACQUISITION

REDUCING PROCESS VARIABLITY BY USING FASTER RESPONDING FLOWMETERS IN FLOW CONTROL

Standard Practices for Air Speed Calibration Testing

INDUSTRIAL APPLICATION EXPERIENCES OF NEW TYPE FLOW- METERING SYSTEM BASED ON ULTRASONIC-DOPPLER FLOW VELOCITY-PROFILE MEASUREMENT

The new 11-R (one part of the Mini WRAS) and GRIMM MINI-WRAS Andreas Jaksch. Symposium Stockholm Mai

792A AC/DC Transfer Standard

BOXA-II ON-STREAM X-RAY FLUORESCENCE ANALYZER. Technical Description Version 2 BGRIMM AUTOMATION

How to avoid pressure oscillations in district heating systems

Understanding the Uncertainty when Using a Type Coefficient for Temperature Corrections of a Coriolis meter. Dean M. Standiford Emerson-Micromotion

Mechanical Principles

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator

Experiment # 5 5. Coupled Water Tanks

IDEA Campus Energy 2016

Using the Coriolis Acceleration to Design a Digital Mass Flow Meter

Series 7500 Formed Metal Bellows Catalog 974C

CE 321 Sample Laboratory Report Packet

8.2 AG Die Sachverständigen für

Anyone who observes. By Jesse Yoder

Full Scale Structural Durability Test Spectrum Reduction by Truncation Coupon Testing

2008 International ANSYS Conference

SELECCIÓN DE CAUDALIMETROS. Tablas de Fabricantes

E C t y p e - e x a m i n a t i o n C e r t i f i c a t e

Flow-meter calibrations for the ID of the ATLAS Experiment

Meteorological Instruments for Solar Energy

Turbine Meter TRZ 03 PRODUCT INFORMATION. Reliable Measurement of Gas

FLOW MEASUREMENT. INC 102 Fundamental of Instrumentation and Process Control 2/2560

FOCUS 30/FOCUS 35 Field Calibration with Survey Pro Field Software

TransCanada / Daniel Using Advances Diagnostics to Identify Measurement Uncertainties

04/01/1998 Developments in DP Flowmeters By Jesse Yoder

DMP 331Pi. Precision Pressure Transmitter. pressure ports and process connections with flush welded stainless steel diaphragm

o0, and since the period of vibration of the scale is related to the mass,

Flow Measurement. SITRANS F M System information SITRANS F M electromagnetic flowmeters. 4/22 Siemens FI

Transcription:

1

Paper # #2015-0164 Verification of Coriolis Flow Meter Calibration Timothy J. Cunningham Principal Engineer Micro Motion, Inc. Boulder, Colorado 80301 2

Agenda Coriolis Mass Flowmeter Theory of Operation Proving, Verification, and Coriolis Robustness Coriolis Meter Verification Review Statistical Hypothesis Testing Direct Stiffness Verification Results Conclusion 3

Agenda Coriolis Mass Flowmeter Theory of Operation Proving, Verification, and Coriolis Robustness Coriolis Meter Verification Review Statistical Hypothesis Testing Direct Stiffness Verification Results Conclusion 4

Coriolis Meters Directly Measure Massflow Process fluid enters the sensor and flow is split with half the flow through h each tube Tubes are oscillated at the first out-of-phase bending mode by closed-loop loop control system 5

How Does It Measure Mass Flow? No flow no Coriolis effect Pickoff signals in phase with each other With flow, Coriolis forces are induced Coriolis forces cause flow tubes to twist Twist causes a time delay, t, between pickoffs t mass flow rate 6

Flow Calibration Factor and Zero Coriolis meters are inherently linear Flow Calibration Factor (FCF) slope 100 80 Zero offset ( t z ) intercept 40 m FCF ( t t Z ) 20 Zero verification not discussed here mass flow rate - e.g. gm/sec Mass flow vs t showing FCF & zero 140 mass flow vs t 120 FCF 60 1 dt z =0.5 0 0 1 2 3 4 5 t - e.g. sec Many techniques, much information on zeroing Modern Coriolis meters have stable zeros, use factory zero Flow Calibration Factor (FCF) units of mass flow/time e.g. (gm/sec)/ sec FCF m t FCF for mass flow is correlated to stiffness 7

Coriolis Density & Volumetric Flow Measurement Density is measured independently of mass flow Both measurements are used to calculate volumetric flow rate Density of instantaneous mixture in tubes Coriolis meters also output standard volume Instantaneous, standard, sampled, calculated density Coriolis transmitters totalize flow Total mass Total volume 8

Agenda Coriolis Mass Flowmeter Theory of Operation Proving, Verification, and Coriolis Robustness Coriolis Meter Verification Review Statistical Hypothesis Testing Direct Stiffness Verification Results Conclusion 9

Terminology Calibration: Establishing the relationship between flow and signal produced by sensor. Relates the flow to the time delay to give FCF Proving, Validation: Confirming flow performance by comparing a primary flow standard to meter under test Provings generates a meter factor Verification: Establishing confidence in performance by analysis of secondary variables correlated with primary flow measurement Verifications gives a Go/No Go message 10

Coriolis Robustness No change in FCF expected over the life of the meter Meter proving gives consistent meter factors Measurement not dependent on flow profile Expect no repair/rebuild over the life of the meter No moving parts means no change in calibration Coriolis MTBF Exida, SIS life expectancy Coriolis flowmeter robustness is key to its value Very low operating costs offset higher initial price 11

Coriolis Meters Show Stable Provings Coriolis meters can be proven Must be configured correctly Coriolis provings are very stable 13 meters, 375 points over 13 years No change in meter factor Provings are expensive $175,000 Value add? 1.01 1.008 Long-term Proving Data for 6 Coriolis flowmeters 1.006 Mete er Factor 1.004 1.002 1 0998 0.998 0.996 04/97 09/98 01/00 05/01 10/02 02/04 07/05 11/06 04/08 08/09 12/10 Date - month/year 12

Directive 17 Coriolis Exceptions Directive 17 recognizes Coriolis meter stability Provides for proving exceptions 13

Directive 17 Coriolis Diagnostics Directive 17 recognizes that Coriolis vendors have developed d new diagnostics and verification techniques in response to user requests 2.6.1 Exceptions 1. If a meter used to measure fluids at flow-line conditions is a type that uses no internal moving parts (e.g., orifice meter, vortex meter, cone meter, Coriolis meter, ultrasonic meter), it does not require proving, provided that all the following conditions are met: The flow through the meter must be continuous (not intermittent) or the meter must qualify for bench proving or be a Coriolis-type meter with sufficient structural integrity internal diagnostics (see below) The internal components of the primary meter device must be removed from service at the same frequency as indicated in Table 2.1, inspected, replaced or repaired if found to be damaged, and then placed back in service, in accordance with procedures specified by the API, the AGA, other relevant standards organizations, other applicable industry-accepted procedures, or the device manufacturer s recommended procedures, whichever are most applicable and appropriate. Internal metering diagnostics may be used to determine if the structural integrity of the primary measurement element is within acceptable operating parameters and checked at the same required intervals as an internal inspection. Then internal inspection is not required until an alarm or error is generated by the device or as recommended by the manufacturer. An initial baseline diagnostic profile must be performed and documented during the commissioning process. The operator must maintain documentation on the diagnostic capability of the measurement system and make that available to the ERCB on request. 14

Coriolis Structural Integrity Verification Verification results mirror stable provings Around 70 verifications over 6 months Some bias and variation apparent Data is stationary, statistics unchanging over time Stiffness results are in-spec, 4 mass FCF is unchanged Directive 17 says no need to prove (in many applications) Extend proving interval Save money Stiffness Un ncertainity % 3 2 1 0-1 -2-3 Structural Integrity Normalized Stiffness K Outlet K Inlet -4 0 10 20 30 40 50 60 70 Meter Verification Counter 15

Agenda Coriolis Mass Flowmeter Theory of Operation Proving, Verification, and Coriolis Robustness Coriolis Meter Verification Review Statistical Hypothesis Testing Direct Stiffness Verification Results Conclusion 16

Key Verification Points Verification will not replace proving Proving regulated by legal & contractual arrangements Proving checks entire flow measurement system Verifications verify calibration (span) Configuration, I/O, zero not necessarily verified Built-in to verification technique/procedure/software? If not, need verify these separately? Coriolis robustness underlies value of verification Documentation and commissioning are critical Multiple verifications at startup Set up routine verifications and documentation at the beginning 2.6.1 Exceptions 1. If a meter used to measure fluids at flow-line conditions is a type that uses no internal moving parts (e.g., orifice meter, vortex meter, cone meter, Coriolis meter, ultrasonic meter), it does not require proving, provided that all the following conditions are met: An initial baseline diagnostic profile must be performed and documented during the commissioning process. The operator must maintain documentation on the diagnostic capability of the measurement system and make that available to the ERCB on request. 17

Verification Overview Why are you going to verify? Leverage Directive 17 to extend proving intervals Flowmeter or process troubleshooting How are you going to verify? Flowmeter device/local display, standalone laptop computers, integrated into DCS Analog I/O or digital comm, wired or wireless What communications protocol; HART, Modbus, Dedicated Asset Management software How often are you going to verify? Verifications are much easier, cheaper, and quicker than proving Done more frequently, generate statistical confidence Who is going to do it? Operations Maintenance, Instrument techs Meter Vendor Third party, e.g. proving service provider, cloud-based system 18

Verification Methods Focus on stiffness based methods Recall Mass Flow Calibration Factor (FCF) has units of mass flow/time, e.g. (gm/sec)/ sec Dimensional analysis shows FCF Stiffness FCF FCF Mass m FCF Time t Time 2 Force / acceleration Force /( Length / Time ) Time Time Force Time Time Length FCF fundamental units of stiffness (force/length) No change in flow tube stiffness no change in FCF 19

Coriolis Stiffness Verification Relationship between FCF & stiffness led to development of stiffness-based verification techniques No absolute stiffness number generated Measure change in stiffness from a baseline Several different stiffness verification methods Known density Drive/Coriolis mode frequency ratio Wall thickness verification Direct stiffness measurement All approach verification from structural dynamics theory Different requirements and assumptions in each method 20

We can measure stiffness statically k d k F d k = Stiffness of spring or flow tube F = Force or drive force on flow tube d = Amplitude of movement or flow tube deflection F One equation, one unknown Determine stiffness exactly by measuring F and d once Mass is not involved in measurement

Dynamics and Verification Verification measures stiffness dynamically Dynamic stiffness measurement must solve for 3 unknowns: stiffness, mass, & damping Frequency Response Function Requires at least 3 10 3 Peak ~1/Damping equations, or, some assumptions 10 2 (constraints) FRF j 2 M jc K se Magnitude Respon 10 1 10 0 Stiffness Mass 10-1 10 2 Frequency (Hz) 22

Stiffness Verification Techniques Common assumptions and constraints Damping is small Use resonant frequency, ratio of stiffness/mass Techniques vary in ease of use Stop flow? Stop measurement? Need additional hardware? Need remote host (PC, DCS, asset management system)? Review stiffness verification methods Known density Drive/Coriolis mode frequency ratio Wall thickness verification Direct stiffness measurement 23

Known Density Verification 1 equation Uses 1 resonant frequency Assumes fluid density is accurately known Typically air or water Vendor independent, d compatible with all Coriolis meters that measure density Known Density Verification 10 Can be challenging or not 3 depending on installation 10 2 Confirms drive frequency 10 is unchanged from 1 baseline Response Magn nitude 10 0 Stiffness Peak ~1/Damping Mass 10-1 Frequency (Hz) 10 2 24

Coriolis Stiffness Verification Known density Drive/Coriolis mode frequency ratio Wall thickness verification Direct stiffness measurement 25

Drive/Coriolis mode frequency ratio 2 equations Uses 2 resonant frequencies Measure Coriolis mode frequency by putting transmitter in verification mode Ratioing drive and Coriolis frequencies eliminates effect of process fluid, verifies tube stiffness Unchanged ratio unchanged stiffness good FCF Response Magn nitude Drive/Coriolis Mode Frequency Verification 10 3 Peak ~1/Damping 10 2 10 1 10 0 Drive Mode Frequency Coriolis Mode Frequency Stiffness Mass 10-1 Frequency (Hz) 10 2 26

Coriolis Stiffness Verification Known density Drive/Coriolis mode frequency ratio Wall thickness verification Direct stiffness measurement 27

Wall thickness verification 2 equations Drive frequency plus 1 tone at 120% of drive frequency Transforms FRF equation from stiffness to wall thickness Verifies unchanging wall thickness Wall Thickness Verification 10 3 Peak ~1/Damping 10 2 Response Magn nitude 10 1 10 0 Stiffness Mass 10-1 Frequency (Hz) 10 2 28

Coriolis Stiffness Verification Known density Drive/Coriolis mode frequency ratio Wall thickness verification Direct stiffness measurement 29

Direct Stiffness Verification 10 equations Drive frequency plus 4 tones at both pickoffs Directly solves for stiffness, mass, and damping independently Overdetermined solution reduces variation Measured stiffness is compared to factory value to confirm FCF Response Magn nitude Direct Stiffness Verification 10 3 10 2 10 1 10 0 Stiffness Peak ~1/Damping Mass 10-1 Frequency (Hz) 10 2 30

Agenda Coriolis Mass Flowmeter Theory of Operation Proving, Verification, and Coriolis Robustness Coriolis Meter Verification Review Statistical Hypothesis Testing Direct Stiffness Verification Results Conclusion 31

Hypothesis Testing Overview Used in many fields: medicine, pharmaceutical, radar, target recognition, law Sample Hypothesis Testing Matrix Verdict of Jury Guilty Innocent Did the Deserved Back on Yes Defendant Jail Time the Street Commit Fl Falselyl Proven No the Crime Imprisoned Innocent Great tool to understand and compare Coriolis verification techniques 2x2 matrix can be organized in many different ways, nomenclature can vary 32

Proving Metrology Proving requires (assumes?) that prover is 3x to 10x more repeatable, accurate, reproducible than meter under test Easy to understand: Prover (jury) is always correct Meter is always suspect Hypothesis testing matrix reduced to 1 column Correct by Corrective action is clear Table 1. Proving Metrology Table Prover Correct by Definition Meter matches Pass Meter prover Do nothing Condition Meter doesn't match prover Fail Adjust meter factor 33

Verification Metrology Takes into account that jury might be wrong Accounts for verification possibilities and statistics Accounts for robustness and low failure rates of Coriolis meters Note organization of hypothesis testing matrix, nomenclature for positive/negative Pass Diagonal terms vs off-diagonal terms Table 2. Verification Hypothesis Testing Matrix Verification Result Fail Meter is True False Alarm Meter accurate Positive Condition Meter is Covert True Negative inaccurate Failure 34

Hypothesis Testing Statistics 101 Hypothesis testing statistics around Coriolis flowmeters and verification give confidence in Directive 17 s exceptions for proving Coriolis failures are very rare (Hard to get data for second row of matrix) Verification statistics and results should be treated differently than proving statistics and results Verification variation under lab conditions on a par with flow repeatability Table 2. Verification Hypothesis Testing Matrix Verification Result Pass Fail Meter is True False Alarm Meter accurate Positive Condition Meter is Covert True Negative inaccurate Failure Variation may increase with field effects Verification failure limits flow/proving failure limits accurate Positive 35

Agenda Coriolis Mass Flowmeter Theory of Operation Proving, Verification, and Coriolis Robustness Coriolis Meter Verification Review Statistical Hypothesis Testing Direct Stiffness Verification Results Conclusion 36

True Positives Getting results for any cell in the hypothesis testing ti matrix requires both flow testing ti and verification results Vast majority of tests will be true positives Flow accuracy in spec Verification passes Meter Condition Meter is accurate Meter is inaccurate Verification Result Pass True Positive Covert Failure Fail False Alarm True Negative 37

True Positive (#1) 1.01 Cavern storage application Meter passes proving Meter passes verification 4 Meter Condition Meter is accurate Meter is inaccurate Verification Result Pass Fail True Positive False Alarm Covert Failure True Negative Verification results within 0.5%, well within failure limits Long-term Proving Data for 6 Coriolis flowmeters Structural Integrity Normalized Stiffness Meter Factor 1.008 1.006 1.004 1.002 1 0.998 Stiffness Unc certainity % 3 2 1 0-1 -2-3 K Outlet K Inlet 0.996 04/97 09/98 01/00 05/01 10/02 02/04 07/05 11/06 04/08 08/09 12/10 Date - month/year -4 0 10 20 30 40 50 60 70 Meter Verification Counter 38

True Positive (#2) NMi testing 1 & 2 meters Meter Gasoil 10C, 30C, and 50C (NEL) Condition Natural gas 16 and 50 bar (Pigsar) Multiple rates Verifications run at all process conditions All flow results within specification Verifications all pass within 0.6% Meter is accurate Meter is inaccurate Verification Result Pass Fail True Positive False Alarm Covert Failure True Negative 39

True Negatives Meter failures are very rare Worked with repair and quality departments Identified 2 meters that failed both calibration and direct stiffness verification Meter Condition Meter is accurate Meter is inaccurate Verification Result Pass True Positive Covert Failure Fail False Alarm True Negative 40

True Negative (#1) Meter used in acidic slurry Meter is Meter accurate Damage visible on inlet manifold Condition Meter is inaccurate Inlet pickoff stiffness -10%, Outlet pickoff -8% Factory flow testing showed density reading ~0.25 gm/cm 3 high, h mass flow error of over +10% highh All consistent with a decrease in stiffness Tube thinning detected on one tube Verification Result Pass Fail True Positive False Alarm Covert Failure True Negative 41

True Negative (#2) Meter is Verification Result Pass Fail True False Alarm Meter wouldn t prove after FAT Meter accurate Positive Condition Meter is Covert inaccurate Failure Pickoff stiffness increased over 10% Factory flow testing showed density reading ~0.2 gm/cm3 low, mass flow error of ~4% low All consistent with a increase in stiffness Inspection clearly showed overpressurized tubes True Negative 42

False Alarms & Covert Failures Statistical analyses required to calculate l off-diagonal outcomes Coriolis meter failures are very rare False alarms can be expensive Meter Condition Meter is accurate Meter is inaccurate Verification Result Pass Fail True Positive False Alarm Covert Failure True Negative Rerun verification to see if verification failure repeats If meter failure is expected in a particular application, consult vendor for methods to minimize covert failures Directive 17 does not require these analyses 43

Agenda Coriolis Mass Flowmeter Theory of Operation Proving, Verification, and Coriolis Robustness Coriolis Meter Verification Review Statistical Hypothesis Testing Direct Stiffness Verification Results Conclusion 44

Conclusion Coriolis flowmeters should not change calibration over the life of the meter Several verification techniques available for Coriolis meters to assure valid calibration Run verification frequently Leverage Directive 17, extend proving intervals Work with your vendor! 45