Study of absorption and re-emission processes in a ternary liquid scintillation system *

Similar documents
Survey of the performance of scintillation materials at low temperatures

arxiv: v1 [physics.ins-det] 27 Feb 2014

A Guide to Recording Fluorescence Quantum Yields

Measurements of Liquid Scintillator Light Yield for Future Neutrino Experiments

DETERMINATION OF RELATIVE FLUORESCENCE QUANTUM YIELD USING THE AGILENT CARY ECLIPSE

Scintillation detectors

Platinum resistance. also wirewound versions. eg

arxiv: v1 [physics.ins-det] 3 Feb 2011

Synthesis of plastic scintillator. Ildefonso León Monzón Universidad Autónoma de Sinaloa

Chapter 4 Scintillation Detectors

R O Y G B V. Spin States. Outer Shell Electrons. Molecular Rotations. Inner Shell Electrons. Molecular Vibrations. Nuclear Transitions

SCINTILLATION DETECTORS AND PM TUBES

The fast light of CsI(Na) crystals

The near-infrared spectra and distribution of excited states of electrodeless discharge rubidium vapour lamps

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

Isotopic effect of Cl + 2 rovibronic spectra in the A X system

D ue to considerable losses by absorption and reflections, short-wavelength light transmission over long

Liquid Scintillator Timing Property

LENA Scintillator Development

05 - Scintillation detectors

Determining Riboflavin Content in a Multivitamin: Things to do before the lab

Reflection = EM strikes a boundary between two media differing in η and bounces back

Skoog Chapter 6 Introduction to Spectrometric Methods

Supplementary information for. Observation of photovoltaic action from photoacid-modified Nafion due to light-driven ion transport

AS 101: Day Lab #2 Summer Spectroscopy

Scintillation Detectors

CHEM6416 Theory of Molecular Spectroscopy 2013Jan Spectroscopy frequency dependence of the interaction of light with matter

Spectroscopy. Page 1 of 8 L.Pillay (2012)

arxiv: v1 [physics.ins-det] 17 Aug 2016

ZICOS - New project for neutrinoless double beta decay experiment using zirconium complex in liquid scintillator -

Supporting Information. Hexagonal Cobalt Oxyhydroxide-Carbon Dots Hybridized Surface: High Sensitive Fluorescence Turn-on Probe for Monitoring of

LENA. Investigation of Optical Scintillation Properties and the Detection of Supernovae Relic Neutrinos. M. Wurm. January 18, 2006 LENA. M.

A Study on Radiation Damage in PWO-II Crystals

A Fluorometric Analysis of Quinine in Tonic Water

Characterisation of Luminescent Down-Shifting Materials for the Enhancement of Solar Cell Efficiency

Chem 321 Lecture 18 - Spectrophotometry 10/31/13

arxiv: v1 [physics.ins-det] 27 Jun 2017

Neutrino Physics. Neutron Detector in the Aberdeen Tunnel Underground Laboratory. The Daya Bay Experiment. Significance of θ 13

Production and Properties of the Liquid Scintillators used in the Stereo Reactor Neutrino Experiment

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

two slits and 5 slits

1 WHAT IS SPECTROSCOPY?

Production of Gadolinium-loaded Liquid Scintillator for the Daya Bay Reactor Neutrino Experiment

Chem 310 rd. 3 Homework Set Answers

PHYS 3446 Lecture #12

arxiv:physics/ v1 3 Aug 2006

FLUORESCENCE QUANTUM YIELDS OF SOME RHODAMINE DYES

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 55, NO. 4, AUGUST /$ IEEE

Lecture 0. NC State University

Ultraviolet-Visible and Infrared Spectrophotometry

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Unit 2. Instrumentation. Experts Teaching from Practical Experience

Lecture 3: Light absorbance

X-ray induced singlet oxygen generation by. nanoparticle-photosensitizer conjugates for. photodynamic therapy: determination of singlet

Wavelength Shifters as (new) light sensors

GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY

DAY LABORATORY EXERCISE: SPECTROSCOPY

1901 Application of Spectrophotometry

Next Generation Scintillation Detectors: Development of Quantum Dot Doped Scintillator. Lindley Winslow University of California Los Angeles

3342 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 43, NO. 6, DECEMBER 1996

Overview of Scintillation Systems

Scintillation and Light Sensitive Detectors - I. Mark Chen Queen s University

Energy. Position, x 0 L. Spectroscopy and the Particle-in-a-Box. Introduction

Overview about the Work with Liquid Scintillators

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters )

FLUORESCENCE APPLICATIONS

A Demonstration of Light Guides for Light Detection in Liquid Argon TPCs

UV-VIS Spectroscopy and Fluorescence Spectroscopy (Part 2 of 2)

Atomic Spectroscopy. Objectives

Double-distance propagation of Gaussian beams passing through a tilted cat-eye optical lens in a turbulent atmosphere

A method to measure transit time spread of photomultiplier tubes with Cherenkov light

Test bench for measurements of NOvA scintillator properties at JINR D.S. Velikanova, 1 A.I. Antoshkin, 1 N.V. Anfimov, 1 O.B.

Vacuum ultraviolet 5d-4f luminescence of Gd 3+ and Lu 3+ ions in fluoride matrices

No. 9 Experimental study on the chirped structure of the construct the early time spectra. [14;15] The prevailing account of the chirped struct

Investigation of M1 transitions of the ground-state configuration of In-like Tungsten

Basics of UV-Visible Spectroscopy *

The TDCR method in LSC. P. Cassette Laboratoire National Henri Becquerel CEA/LNE, France

Wavelength λ Velocity v. Electric Field Strength Amplitude A. Time t or Distance x time for 1 λ to pass fixed point. # of λ passing per s ν= 1 p

Interaction mechanism for energy transfer from Ce to Tb ions in silica

Supporting Information

Supplementary Information

Gamma Spectroscopy. References: Objectives:

Near-infrared dual luminescence from an extended zinc porphyrin

2001 Spectrometers. Instrument Machinery. Movies from this presentation can be access at

Chemistry 524--Final Exam--Keiderling May 4, :30 -?? pm SES

Supporting Information. Synthesis and Upconversion Luminescence of BaY 2

arxiv: v2 [physics.ins-det] 12 Feb 2014

Chapter 17: Fundamentals of Spectrophotometry

What is spectroscopy?

Energy transfer and optical gain studies of FDS: Rh B dye mixture investigated under CW laser excitation

Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications

MOF-76: From Luminescent Probe to Highly Efficient U VI Sorption Material

Geant4 Monte Carlo code application in photon interaction parameter of composite materials and comparison with XCOM and experimental data

Fast detectors for Mössbauer spectroscopy )

PHYS 3650L - Modern Physics Laboratory

C101-E145 TALK LETTER. Vol. 17

John Ellison University of California, Riverside. Quarknet 2008 at UCR

single-molecule fluorescence resonance energy transfer

Solution reduction synthesis of amine terminated carbon quantum dots

Single Emitter Detection with Fluorescence and Extinction Spectroscopy

Transcription:

CPC(HEP & NP), 2010, 34(11): 1724 1728 Chinese Physics C Vol. 34, No. 11, Nov., 2010 Study of absorption and re-emission processes in a ternary liquid scintillation system * XIAO Hua-Lin( ) 1;1) LI Xiao-Bo( ) 2 ZHENG Dong( ) 3 CAO Jun( ) 2 WEN Liang-Jian( ) 2 WANG Nai-Yan( ) 1 1 College of Nuclear Science and Technology, Beijing Normal University, Beijing 100875, China 2 Institute of High Energy Physics, Chinese Academy of Sciences, Beijing 100049, China 3 Analytical and Testing Center, Beijing Normal University, Beijing 100875, China Abstract Liquid scintillators are widely used as the neutrino target in neutrino experiments. The absorption and emission of different components of a ternary liquid scintillator (Linear Alkyl Benzene (LAB) as the solvent, 2,5-diphenyloxazole (PPO) as the fluor and p-bis-(o-methylstyryl)-benzene (bis-msb) as wavelength shifter) are studied. It is shown that the absorption of this liquid scintillator is dominant by LAB and PPO at wavelengths less than 349 nm, and the absorption by bis-msb becomes prevalent at the wavelength larger than 349 nm. The fluorescence quantum yields, which are the key parameters to model the absorption and re-emission processes in large liquid scintillation detectors, are measured. Key words PACS 29.40.Mc liquid scintillator, absorption and re-emission, fluorescence quantum yield 1 Introduction Liquid scintillators have been used extensively in neutrino experiments. Recently, a ternary liquid scintillator has been proposed for the Daya Bay Reactor Neutrino Experiment [1 3]. LAB is chosen as the solvent, and 3 g/l PPO and 15 mg/l bis-msb are chosen as the primary solute and the secondary solute, respectively. The liquid scintillator doped with 0.1% Gadolinium will be used as the neutrino target and the undoped one will be used as the Gamma-Catcher. Some properties of this ternary liquid scintillator have been studied in Refs. [2, 4 6]. In this paper, we will study the absorption and emission properties. When a liquid scintillation solution is exposed to ionizing radiation, the primary radiation energy is primarily absorbed mostly by the solvent which makes up the bulk of the solution, then partially transferred to the solute via two processes. One is a non-radiative process, which involves the dipoledipole interaction and the short-distance collision between the donor and acceptor molecules when they come sufficiently close to each other. The other is a radiative process, which involves the absorption and re-emission of the photons [7, 8]. In a ternary liquid scintillator, the primary solute is always utilized as a fluor in order to obtain a higher scintillation efficiency and the secondary solute is always utilized as the wavelength shifter in order to minimize the selfabsorption of the scintillation and provide a better match between the photomultiplier spectral response and the fluorescence emission spectrum. Generally, the fluorescence efficiency of the solvent itself is small [7]. The radiative energy transfer process between the solvent and the fluor is insignificant and most of the excitation energy is migrated to the fluor by a nonradiative process. Hence, most of the scintillation is primarily generated by the fluor. Both of scintillation generated by the fluor and the Cerenkov photons produced by charge particles could be absorbed and the photons can be re-emitted by the liquid scintillator, which makes the scintillation spectrum red-shift [9] and modifies the time profile of the scintillation [10]. It is important to model the optical process to study the behaviors of the scintillation detector. In this work, the absorptions of different Received 25 January 2010 * Supported by Nation Natural Science Foundation of China (211202037) 1) E-mail: xiaohl@mail.bnu.edu.cn 2010 Chinese Physical Society and the Institute of High Energy Physics of the Chinese Academy of Sciences and the Institute of Modern Physics of the Chinese Academy of Sciences and IOP Publishing Ltd

No. 11 XIAO Hua-Lin et al: Study of absorption and re-emission processes in a ternary liquid scintillation system 1725 components of this ternary liquid scintillator and their emission spectra are studied. The fluorescence quantum yields, which are defined as the re-emission probability after an absorption by the PPO and the bis-msb, are measured. 2 Absorption and emission The attenuation lengths of the four samples listed in Table 1 are shown in Fig. 1. They were measured with a cuvette of 2 mm optical path. The attenuation lengths of the sample B and sample D at λ < 345 nm were not obtained because the transmission was close to 0. As shown in Fig. 1, the attenuation of the LAB is negligible at λ > 340 nm. Moreover, the attenuation length curve of sample C overlaps with that of sample D at λ > 360 nm and crosses with the curve of sample B at 349 nm. This denotes that bis- MSB has a dominant contribution to the attenuation at λ > 349 nm. At λ < 349 nm, the attenuations by the LAB and PPO become prevalent. Figure 2 shows the attenuation length of the ternary liquid scintillator sample D (345 nm < λ < 417 nm). Four cuvettes (2 mm, 4 mm, 1 cm and 10 cm optical path) were used. The cuvettes of different optical paths provided a cross check with each other. The attenuation curves measured by the cuvettes of different optical paths largely overlap. The attenuation lengths at longer wavelengths cannot be obtained because the transmissions even for the 10 cm optical path are close to 100%, but this measurement can be performed by a long tube, which has been described in Ref. [4]. The absorption cross section σ can be expressed as 1/Λ, where Λ is the optical attenuation length, which is defined as the distance at which the incident light intensity I 0 reduces to its 1/e. The attenuation length Λ can be calculated by I(d)/I 0 = e d/λ, (1) where I 0 is the original light intensity and I(d) is the light intensity after a path d. The attenuation length Λ can be obtained by measuring the transmission, i.e. I(d)/I 0, by an ultraviolet spectroscopy method. In order to study the attenuations of different components in the ternary liquid scintillator, we prepared four samples. Their recipes are listed in Table 1. The attenuation length measurements were performed by a double beam UV-Vis spectrophotometer. Fig. 1. The attenuation lengths of the pure LAB (1), sample B (2), sample C (3) and sample D (4). Table 1. Samples in the attenuation length measurement. sample solvent PPO bis-msb A LAB 0 0 B LAB 3 g/l 0 C LAB 0 15 mg/l D LAB 3 g/l 15 mg/l Fig. 2. The attenuation length of the ternary liquid scintillator sample D. The transmission measurements were performed with the cuvettes of 2 mm, 4 mm, 1 cm and 10 cm optical path. Photons could be re-emitted after the absorption. The emission spectrum can be measured with a fluorescence spectrometer. In order to measure the intrinsic emission spectra of the LAB, the PPO and the bis- MSB, we diluted them with cyclohexane, which can be regarded as transparent in our wavelength region of interest. Meanwhile, we prepared a 3 g/l PPO in cyclohexane to study the emission spectra at high concentration. These samples were placed in 2 mm cuvettes for the emission spectra measurements. To study the energy transfer between the PPO and the bis-msb, a small amount of the ternary liquid scintillator sample D was painted on a glass surface and

1726 Chinese Physics C (HEP & NP) Vol. 34 its emission spectrum was measured. The thickness of this sample is so small that the radiative energy transfer process (emission after absorption process) can be ignored and only the non-radiative process can occur in this thickness. The LAB sample and the bis-msb sample were excited by ultraviolet light at 280 nm and 360 nm, respectively. The dilute PPO sample, the 3 g/l PPO sample and the ternary liquid scintillator sample were excited at 310 nm. Most of the excitation energy was absorbed by the PPO in the ternary sample when they were excited at 310 nm. Figure 3 shows their emission spectra measured by a Fluorolog-3 fluorescence spectrometer. The emission spectra were measured in a front-face geometry to minimize the self-absorption of the fluorescence. We see from Fig. 3 that the emission spectra of the 3 g/l PPO and the dilute PPO show no significant difference in shape and peaks. This indicates that the formation of the transient dimers (see Ref. [11]) is not obvious, which changes the shape and peaks of the emission spectrum and lowers the fluorescence yield. Hence, the concentration quenching of the PPO is not important at the concentration of 3 g/l. It is also observed that the fluorescence of the bis-msb was not present in the emission spectrum of the thin ternary liquid scintillator sample D. This shows that the nonradiative energy transfer process between the PPO and the bis-msb is negligible. In the ternary liquid scintillator, the possible energy transfer process between the PPO and the bis-msb is the absorption and re-emission process. In the next section, we will report the measurement of the re-emission probability of the PPO and the bis-msb. Fig. 3. The emission spectra of the dilute LAB (1), the dilute PPO (2), the 3 g/l PPO (3), the ternary liquid scintillator sample D painted on the glass surface(4) and the dilute bis-msb (5). The spectra were collected at a front-face geometry and all spectra shown in the figure are in arbitrary units. 3 Fluorescence quantum yield measurement 3.1 Measure fluorescence quantum yield with relative method The fluorescence quantum yield is defined as the re-emission probability after an absorption. One of the practical methods to measure the fluorescence quantum yield is the relative method, which employs the fluorescence spectrometer to measure the fluorescence intensities of the unknown fluorescence quantum yield sample and the known fluorescence quantum yield sample (the reference sample). The expression for the unknown fluorescence quantum yield is [12] s B x (λ)dλ r n Φ x = Φ r 2 x, (2) B x n s r (λ)dλ 2 r where the subscripts r and x refer to the reference sample and the unknown fluorescence quantum yield sample, Φ is the quantum yield, s(λ) is the emission spectrum, the item (n x /n r ) 2 is the correction of the refractive index effect and B is the fraction of the excitation light absorbed by the sample, which is given by B = 1 T, (3) where T is the transmission of the sample. The transmission can be measured by an UV-Vis spectrophotometer. 3.2 Samples The reliability of the fluorescence standard is quite important for the relative method. Quinine sulfate in 0.1 mol/l sulfuric acid, whose fluorescence quantum yield is 0.55 [13], is an ideal fluorescence standard. It has been widely used as the fluorescence quantum yield standard. Its fluorescence quantum yield can be regarded to be independent of the excitation wavelength over 200 390 nm [12]. In order to have a cross check of our method and instrument, fluorescence quantum yield standard harmane in 0.1 mol/l sulfuric acid was prepared as well. The fluorescence quantum yield of harmane is 0.83 [14]. As described in Section 2, the concentration quenching effect is not obvious in the 3 g/l PPO sample. The fluorescence quantum yield of the PPO when the PPO concentration is 3 g/l should be equal to that in the lower concentration. It is reported that

No. 11 XIAO Hua-Lin et al: Study of absorption and re-emission processes in a ternary liquid scintillation system 1727 the fluorescence collection efficiency will change because of the drift in scintillation gravity when the absorption of the sample is high in the fluorescence intensity measurement (see Ref. [12]). The change in fluorescence collection efficiency will change the fluorescence intensity and bring great errors to the fluorescence quantum yields. We carefully calibrated this effect with the standard samples of different absorptions which were excited at the same wavelength, and we observed that the fluorescence intensities were proportional to the absorptions when they were lower than 0.5. We used the diluted samples with absorptions lower than 0.5 for both the samples and the references, thus the effect of the drift of scintillation gravity can be ignored. We diluted the PPO in cyclohexane to measure the fluorescence quantum yield of the PPO (the PPO concentration was 125 mg/l). Moreover, we prepared several LAB/bis-MSB samples (the bis-msb concentrations were 1 mg/l, 5 mg/l, 7.5 mg/l and 15 mg/l, respectively) to measure the fluorescence quantum yield of the bis-msb. As described in Section 1, the absorption of the LAB is negligible when the wavelength of excitation light is larger than 340 nm compared with the bis-msb. 3.3 Corrections and calibration In our measurements, the fluorescence spectra were corrected by the wavelength sensitivity curve of the photomultiplier and the variation of the intensity of the excitation light by the reference detector, which was mounted on the fluorescence spectrometer. Figure 4 shows the integrated fluorescence intensities of unit absorption, i.e. s r (λ)dλ/b r in Eq. (2), as a function of the excitation wavelength of the two standard samples. These samples were placed in 2 mm cuvettes during the absorption and emission measurements and the fluorescence was collected in a frontface geometry during the fluorescence spectra measurements. The intensity of unit absorption of the quinine sulfate, which is proportional to the fluorescence quantum yield, is flat within 2%. The average intensities of unit absorption of the quinine sulfate sample and the harmane sample are 2.44 10 10 and 3.83 10 10, respectively. Since the fluorescence spectra of the two standard samples are similar and both of the samples are diluted, the refractive indices of the two standard samples can be regarded as equal. The ratio of the fluorescence quantum yields of the two standard samples can be obtained from Eq. (2). This value is 1.56. The ratio of the fluorescence quantum yields of the two standard samples which were reported in Refs. [13, 14] is 1.51 (0.83/0.55). 3.4 Results In the absorption and fluorescence intensity measurements, the PPO and the bis-msb samples were placed in 2 mm cuvettes. The absorptions at different wavelengths were measured with the spectrophotometer. The fluorescence intensities were measured with the Fluorolog-3 spectrometer and the fluorescence was collected in the front-face geometry in order to minimize the self-absorption of the fluorescence. The excitation wavelength of the PPO sample varied from 300 nm to 340 nm and the bis-msb from 360 nm to 380 nm. Fig. 5 shows the fluorescence intensities at different absorptions and corresponding linear fits. Since the fluorescence quantum yields are Fig. 4. The integration fluorescence intensities of unit absorption (i.e. s(λ)dλ/b) in the Eq. (2) of the quinine sulfate sample and the harmine sample as a function of excitation wavelength. Fig. 5. The fluorescence intensities of the PPO and the bis-msb samples at different absorptions. The excitation wavelength of the PPO sample was from 300 to 340 nm and bis-msb from 360 nm to 380 nm. The absorption is the B described in Eq. (2).

1728 Chinese Physics C (HEP & NP) Vol. 34 independent of excitation wavelength, the slopes of linear fits which denote the fluorescence intensities of unit absorption are proportional to their fluorescence quantum yields. Table 2. The fluorescence quantum yield of the PPO and the bis-msb. material fluorescence quantum yield PPO 0.77 bis-msb 0.96 The average refractive indices of the samples were calculated by the expression s(λ)n(λ)dλ/s(λ)dλ, where s(λ) and n(λ) are the corresponding emission spectrum and the refractive index, respectively. The refractive index of the LAB was measured by Abbe prism and V prism methods. The average refractive index of the LAB for the bis-msb fluorescence is 1.50. The refractive index of the 0.1 mol/l sulfate acid and the cyclohexane were obtained from Ref. [15] and Ref. [16]. Their average refractive indices are 1.34 and 1.44, respectively. The fluorescence quantum yields of the PPO and the bis-msb can be obtained from Eq. (2) by comparing their fluorescence intensities of unit absorption with that of the quinine sulfate sample. Table 2 shows the results. Both of the fitting errors of the PPO and the bis-msb fluorescence quantum yields are 0.03. 4 Conclusion The absorption and emission properties of a ternary liquid scintillator were studied. The absorption was dominant by LAB and PPO at wavelengths less than 349 nm and the absorption of bis-msb become privilege at wavelength larger than 349 nm. The fluorescence quantum yield of the PPO and the bis- MSB were measured. This makes it possible to model the absorption and re-emission process in the studied ternary liquid scintillator. References 1 Daya Bay Collaboration. arxiv:hep-ex/0701029v1 2 DING Y et al. Nucl. Instrum. Methods A, 2008, 584: 238 243 3 Yeh M et al. Nucl. Instrum. Methods A, 2007, 578: 329 339 4 LIU Jin-Chang et al. HEP & NP, 2007, 30(1): 76 79 (in Chinese) 5 ZHONG W L et al. Nucl. Instrum. Methods A, 2008, 587: 300 303 6 XIAO H L, DENG J S. arxiv:nucl-ex/0904.1329v3 7 Birks J B. The Theory and Practice of Scintillation Counting Oxford, EnglandPergamon Press, 1964. 278 312 8 Cohen S G et al. Proc. Phys. Soc. B, 1956, 69: 593 605 9 D Ambrosio C et al. Appl. Spectrosc., 1991, 45(3): 484 490 10 Martinho J M G, Maçanita A L, Berberan-Santos M N. J. Chem. Phys., 1989, 90(1): 53 59 11 Berlman I B. J. Chem. Phys., 1961, 34: 1083 1084 12 Demas J N et al. J. Chem. Phys., 1971, 75(8): 991 1023 13 Melhuish W H. J. Phys. Chem., 1961, 65: 229 14 Pardo A et al. Journal of Luminescence., 1992, 51: 269 274 15 Daimon M, Masumura A. Appl. Opt., 2007, 46: 3811 3820 16 Rheims J et al. Meas. Sci. Technol., 1997, 8: 601 605