High resolution neutron spectrometry with liquid scintillation detectors for fusion applications

Similar documents
Unfolding of neutron spectra with an experimentally determined diamond detector response function

Data analysis for neutron spectrometry with liquid scintillators: applications to fusion diagnostics

Diagnostic Capabilities of Line-Integrated Neutron Pulse Height Spectra Measurements

Characterization of the 3 MeV Neutron Field for the Monoenergetic Fast Neutron Fluence Standard at the National Metrology Institute of Japan

Neutron Emission Spectroscopy Measurements with a Single Crystal Diamond Detector at JET

MONTE-CARLO SIMULATIONS OF TIME- RESOLVED, OPTICAL READOUT DETECTOR for PULSED, FAST-NEUTRON TRANSMISSION SPECTROSCOPY (PFNTS)

STUDY ON THE ENERGY RESPONSE OF PLASTIC SCINTILLATION DETECTOR TO MEV NEUTRONS ABSTRACT

The Deuterium-Deuterium Neutron Time-of- Flight Spectrometer TOFED at EAST

Recent Activities on Neutron Calibration Fields at FRS of JAERI

Update on Calibration Studies of the Canadian High-Energy Neutron Spectrometry System (CHENSS)

Fast-Neutron Production via Break-Up of Deuterons and Fast-Neutron Dosimetry

Improved Plasma Confinement by Ion Bernstein Waves (IBWs) Interacting with Ions in JET (Joint European Torus)

Neutron and gamma ray measurements. for fusion experiments and spallation sources

Improved Plasma Confinement by Ion Bernstein Waves (IBWs) Interacting with Ions in JET

Energy Resolution of LaBr 3 (Ce) Gamma-Ray spectrometer for Fusion Plasma Studies on JET

A Measurement of Monoenergetic Neutrons from 9 Be(p,n) 9 B

Recent Activities on Neutron Standardization at the Electrotechnical Laboratory

Diamond Neutral Particle Spectrometer at JET and proposal for ITER

arxiv: v2 [physics.ins-det] 1 Sep 2017

Scintillation efficiency measurement of Na recoils in NaI(Tl) below the DAMA/LIBRA energy threshold

A Project for High Fluence 14 MeV Neutron Source

(Tandem Collimators for the Tangential GammaRay Spectrometer - KM6T-TC)

New developments in JET gamma emission tomography

Development of a neutron time-of-flight source at the ELBE accelerator

V. 3. Development of an Accelerator Beam Loss Monitor Using an Optical Fiber

Measurement of nuclear recoil responses of NaI(Tl) crystal for dark matter search

Tandem Collimators System

A gamma-ray spectrometer system for fusion applications

Neutron diagnostics for reactor scale fusion experiments: a review of JET systems

arxiv: v1 [physics.ins-det] 16 May 2017

GA A22407 THRESHOLD BUBBLE CHAMBER FOR MEASUREMENT OF KNOCK-ON DT NEUTRON TAILS FROM MAGNETIC AND INERTIAL CONFINEMENT EXPERIMENTS

Measurement of the energy spectrum from the neutron source planned for IGISOL

PMT Charge Response and Reconstruction of Position and Energy for a Reactor θ 13 Experiment

High Energy Neutron Scattering Benchmark of Monte Carlo Computations

Monte Carlo Methods for Uncertainly Analysis Using the Bayesian R-Matrix Code SAMMY

Measurements of liquid xenon s response to low-energy particle interactions

A new neutron monitor for pulsed fields at high-energy accelerators

PFC/JA MEASUREMENT OF THE D-D FUSION NEUTRON ENERGY SPECTRUM AND VARIATION OF THE PEAK WIDTH WITH PLASMA ION TEMPERATURE

Neutronics experiments for validation of activation and neutron transport data for fusion application at the DT neutron generator of TU Dresden

Activation measurement of (n,t) reactions

Shielded Scintillator for Neutron Characterization

Calibration of the Modular Neutron Array (MoNA)

Response function measurements of an NE102A organic scintillator using an 241 Am-Be source

The Versatile Array of Neutron Detectors at Low Energy

Status of ISS-RAD. Cary Zeitlin, Southwest Research Institute On behalf of the ISS-RAD Team

Neutrons for Science (NFS) at SPIRAL-2

Upgrade of the Magnetic Proton Recoil (MPRu) spectrometer for MeV neutrons for JET and the next step

This paper should be understood as an extended version of a talk given at the

Neutron emission spectroscopy of fusion plasmas with a NE213 liquid scintillator at JET

Digital imaging of charged particle track structures with a low-pressure optical time projection chamber

arxiv: v1 [physics.plasm-ph] 20 Feb 2019

Dipartimento di Fisica, Università degli Studi di Milano-Bicocca, Milano 2. Istituto di Fisica del Plasma, CNR, Milano

Neutronics experiments for validation of activation and neutron transport data for fusion application at the DT neutron generator of TU Dresden

Photoneutron reactions studies at ELI-NP using a direct neutron multiplicity sorting method Dan Filipescu

Measurement of the photodissociation of the deuteron at energies relevant to Big Bang nucleosynthesis

Diamond detectors in Bonner Spheres A Novel Approach for Real-time Neutron Spectroscopy

Preparatory experiments for cold-neutron induced fission studies at IKI

Measurements of the Proton Light Output Function of the Organic Liquid Scintillator NE213 in Several Detectors

Principles of neutron TOF cross section measurements

Recent developments in neutron metrology at the Institute for Radiological Protection and Nuclear Safety (IRSN)

Fast Neutron and Gamma-Ray Detectors for the CSIRO Air Cargo Scanner

Feasibility Studies for the EXL Project at FAIR *

Unfolding techniques for neutron spectrometry

Measurement of 40 MeV Deuteron Induced Reaction on Fe and Ta for Neutron Emission Spectrum and Activation Cross Section

A comprehensive study of a wide energy range proportional counter neutron spectrometer

Activities of the neutron standardization. at the Korea Research Institute of Standards and Science (KRISS)

Scintillation Detector

Study of Collimated Neutron Flux Monitors for MAST and MAST Upgrade

On the Core Deuterium-Tritium Fuel Ratio and Temperature Measurements in DEMO

PoS(FNDA2006)033. Radiography Using Fission Neutrons. Thomas Bücherl. Christoph Lierse von Gostomski

Characterizations and Diagnostics of Compton Light Source

II. 5. Study for NaI(Tl) and Scintillation Fiber with 80 MeV Proton Beam Toward ESPRI Experiment at NIRS-HIMAC, RIKEN-RIBF

Fast ion physics in the C-2U advanced, beam-driven FRC

Hands on Project: Large Photocathode PMT Characterization

New Neutron-Induced Cross-Section Measurements for Weak s-process Studies

Measurement of the inelastic neutron scattering cross section of 56 Fe

Bonner Sphere Spectrometer. Cruzate, J.A.; Carelli, J.L. and Gregori, B.N.

Evaluation of the Nonlinear Response Function and Efficiency of a Scintillation Detector Using Monte Carlo and Analytical Methods

PoS(TIPP2014)088. Diamonds for beam instrumentation. Erich Griesmayer 1. Pavel Kavrigin

D- 3 He Protons as a Diagnostic for Target ρr

INTENSE PULSED NEUTRON GENERATION TO DETECT ILLICIT MATERIALS.

Quality Assurance. Purity control. Polycrystalline Ingots

Characterization and Monte Carlo simulations for a CLYC detector

SCI-O11. Design of a Compton Spectrometer Experiment for Studying Electron Response of a Scintillator

High Resolution Optical Spectroscopy

arxiv: v1 [nucl-ex] 8 Aug 2012

Diagnosing OMEGA and NIF Implosions Using the D 3 He Spectrum Line Width

High-energy calibration data for neutron activation analysis

Response of LaBr 3 (Ce) scintillators to 2.5 MeV fusion neutrons

RESPONSE FUNCTION STUDY FOR ENERGY TO LIGHT CONVERSION IN ORGANIC LIQUID SCINTILLATORS

Beam diagnostics: Alignment of the beam to prevent for activation. Accelerator physics: using these sensitive particle detectors.

PoS(ICHEP2016)474. SoLid: Search for Oscillations with a Lithium-6 Detector at the SCK CEN BR2 reactor

EEE4106Z Radiation Interactions & Detection

Measurement of the 30 Si mole fraction in the new Avogadro silicon material by Neutron Activation and high resolution spectrometry

Neutral beam plasma heating

Investigation of fast neutron spectroscopy capability of 7 Li and 6. Li enriched CLYC scintillator for nuclear physics experiments

PoS(PD07)020. Timing Properties of MCP-PMT. Kenji Inami. Nagoya university, Nagoya, Japan

Neutron based instruments for application to fusion and spallation sources

Method of active correlations in the experiment 249 Cf+ 48 Ca n

Plasma Accelerator for Detection of Hidden Objects by Using Nanosecond Impulse Neutron Inspection System (NINIS)

Transcription:

High resolution neutron spectrometry with liquid scintillation detectors for fusion applications Andreas Zimbal *, Horst Klein, Marcel Reginatto, Helmut Schuhmacher Physikalisch-Technische Bundesanstalt Bundesallee 1, D-38116 Braunschweig, Germany E-mail: Andreas.Zimbal@PTB.DE Luciano Bertalot Associazione Euratom-EEA sulla Fusione C.P. 65, I-44 Frascati, Roma, Italy Andrea Murari Consorzio RFX-Ass. Euratom-EEA sulla Fusione Corso Stati Uniti 4, I-35127 Padua, Italy and JET-EFDA Contributors See annex of J. Pamela et al., 23 Overview of recent JET results, ucl. Fusion 43 International Workshop on Fast eutron Detectors University of Cape Town, South Africa April 3 6, 26 * Speaker Copyright owned by the author(s) under the terms of the Creative Commons Attribution-onCommercial-ShareAlike Licence. http://pos.sissa.it

Organic liquid scintillation detectors, e.g. E213, EJ-31 or BC51A, are routinely used as high resolution neutron spectrometers in many nuclear experiments. The neutron energy spectrum can be reliably determined from the pulse height spectrum provided the response function of the detector is well determined, counting statistics is sufficient, a suitable unfolding method is applied and care is taken to ensure the gain stability of the system during the measurement. The Physikalisch-Technische Bundesanstalt (PTB) has many years of experience with the development, detailed characterization and use of liquid scintillation detectors. Recent work done at PTB in collaboration with EEA and EFDA-JET concerning the use of liquid scintillation detectors as neutron diagnostic instruments in fusion experiments is discussed. An E213 liquid scintillator system has been carefully characterized and successfully used in a series of measurements at the Joint European Torus (JET). We describe the improvements that were made to the system to meet the challenging measurement conditions at JET, as well as the processes used to calibrate and characterize the system. For the determination of the neutron light output function and for a comparison of experimentally determined response functions with those calculated with Monte Carlo techniques, measurements have been performed at the PTB facility in mono-energetic neutron reference fields and in fields with a broad energy distribution applying time of flight spectrometry. Selected examples of measured neutron spectra are discussed in terms of energy resolution and compared to the expected neutron spectra, both for measurements at the PTB accelerator and at JET. 2

1. Introduction Organic liquid scintillation detectors such as E213, EJ-31 or BC51A are routinely used as spectrometers for neutron and photon radiation in many nuclear experiments. The intrinsic properties of these scintillators permit the separation of neutron and photon induced events by pulse shape analysis and allow for spectrometry in mixed neutron/photon radiation fields [1]. The possibility of using these detectors as high resolution neutron spectrometers, e.g. at fusion experiments like the Joint European Torus (JET), has been demonstrated in recent experimental campaigns [2]. For these applications it is necessary to ensure gain stability of the system and to know precisely the detector response function needed for the application of dedicated methods of data analysis [3]. The envisaged installation of compact neutron detection systems at the International Thermonuclear Experimental Reactor (ITER) require further developments in the field of data acquisition [4, 5] as well as the investigation of alternative methods for the characterization of the detectors. 2. The E213 scintillation detector For the experiments at JET [2], an E213 scintillation detector (5.8 cm in diameter and in length) was used. The scintillator cell is coupled through a partially coated light guide to an XP22 Photomultiplier. In this way, the light collection system is designed for an optimized pulse height resolution [6]. An LED signal fed into the light guide was used as a monitor for count rate dependent gain variations of the photomultiplier. eutron spectrometry in the energy range from 1.5 MeV < < 2 MeV and photon spectrometry in the energy range from.2 MeV < E γ < 1 MeV can be performed with this detector system. 3. Determination of the detector response functions for neutrons The approach used up to now at PTB for the characterization of scintillation detectors [1, 7] is based on a combination of measurements in mono-energetic neutron and photon reference fields and Monte Carlo calculations using specially developed programs [8]. The calculations are partly based on parameters which need to be determined from the experimental results. Although this method has proven to be very powerful, there are certain limitations to it. Small deficiencies in the response functions used for the unfolding can result in oscillations in the resulting neutron spectra in the case of broad neutron energy distributions or in spurious peaks in the case of nearly mono-energetic neutron spectra. We report on another calibration technique [9, 1] that improves upon the standard approach. A neutron beam with a broad energy distribution was produced using the 9 Be(p,n)-reaction with the ns-pulsed proton beam (E p = 19 MeV) from the PTB cyclotron and a thick (3 mm) Be-target. The spectral neutron distribution of this reaction determined with the time-of-flight technique (TOF) is shown in Figure 1. 3

5x1 6 4x1 6 Φ E -1 cm -2 3x1 6 2x1 6 1x1 6 2 4 6 8 1 12 14 16 18 Figure 1: Spectral fluence of neutrons Φ Ε from of the 9 Be(p,n)-reaction using the ns-pulsed proton beam (E p = 19 MeV) from the PTB cyclotron and a thick (3 mm) Be-target analyzed from a time-of-flight measurement. The E213 detector was placed at a distance of 27.35 m from the target behind a collimator. During a 4 h run, the amplitude of the integrated signal from the 9 th dynode of the photomultiplier tube (PMT), the shape of the anode pulse and the TOF of each event were analyzed with conventional IM standard electronic modules and stored in list mode using a MPA-3 * multi parameter data acquisition system. The TOF information was recorded with 8192 channels using a time calibration of.5 ns/channel, the pulse height converted with a gain factor of 5 kev ee /channel obtained by a calibration with a 27 Bi γ-source in terms of electron equivalent energies and the pulse shape information were recorded with 496 channels each. From these data, a two-dimensional histogram was extracted representing the pulse height response of the detector as a function of the time-of-flight. This two-dimensional histogram is shown in Figure 2, where all neutron and photon events are reported. The uncertainty in the determination of the neutron energy can be estimated to be 58 kev for a neutron energy of 14 MeV taking into account the width of the prompt γ-peak in the TOF spectrum, the time calibration of the Time to Amplitude Converter used in this experiment (including differential nonlinearity and calibration against a DCF77 based frequency standard) and the time response of the detector chiefly determined by its length. A total of 161 time slices corresponding to neutron energies between 1.5 MeV and 17. MeV were extracted from the pulse height response of the detector shown in Figure 2, after gating to neutron induced events only and after correction of the satellite peaks from the * www.fastcomtec.com 4

cyclotron. The energy width of the channels in the neutron time of flight spectrum increases with neutron energy (from 8 kev at = 2.5 MeV to 27 kev at = 14 MeV) and is a factor of 15 smaller than the resolution of the corresponding pulse height spectrum even for the highest neutron energies, which is important for achieving an energy resolution 4-5 times smaller than the corresponding pulse height resolution with unfolding methods. After fluence normalization, these 161 individual functions for mono-energetic neutrons form the response matrix needed for the unfolding process. The RESP7 code [8] was used for the fluence normalization of these pulse height spectra, considering only the pulse height range resulting from (n,p)-scattering (see Figure 3). The threshold for this normalization was varied with the neutron energy. The proton light output function of the detector, an important input for the RESP7 calculations, was also extracted from the 161 pulse height spectra. The result of this analysis is shown in Figure 3. The measured response matrix was then used for the unfolding of data collected in the PTB reference neutron fields (section 4) and at JET (section 5). Figure 2: Pulse height response of the detector vs. TOF measured with the 9 Be(p,n)-reaction. The prompt γ-peak and the neutron TOF-distribution can clearly be seen, but also contributions from satellite -pulses of the cyclotron which can be identified in the TOF-region between the prompt γ-peak and the neutron events, or at high pulse amplitudes for long TOF, respectively. The detector response matrix was extracted from this data set selecting only neutron events by pulse shape analysis. In total 9. 1 7 neutron events were recorded during a 4 h run. 5

.1.1.4.8.6 = 2.5 MeV.8.6 = 3. MeV.3 = 4. MeV.2.4.4.2.2.1...2.4.6.8 1..2.15.1.5.8.6.4.2. L...5 1. 1.5 2. 2.5 3..1 L = 12. MeV 1 2 3 4 5 6 7 8 L = 6. MeV..15.5..2.4.6.8 1. 1.2 1.4.1 = 8. MeV En...5 1. 1.5 2. 2.5 3. 3.5 4. 4.5 5..14.12.1.8.6.4.2. L L = 14. MeV 1 2 3 4 5 6 7 8 9 1 L...2.4.6.8 1. 1.2 1.4 1.6 1.8 2..1.8.6.4.2...5 1. 1.5 2. 2.5 3. 3.5 4. 4.5 5. 5.5 6..14.12.1.8.6.4.2. L = 1. MeV L = 16. MeV normalization with RESP7 1 2 3 4 5 6 7 8 9 1 Figure 3: Response functions of the E213 scintillation detector for 9 neutron energies selected between 2.5 MeV and 16 MeV by time-of-flight slices. The experimental spectra (black histogram) are compared with and normalized to responses calculated with the RESP7 code (red lines). 4. Test in PTB reference fields Measurements in mono-energetic neutron reference fields were carried out at the PTB accelerator facility to test the new calibration procedure of the detector. Among the different neutron fields available at PTB, two specific reactions provide the opportunity of producing mono-energetic neutron beams with energies of 2.5 MeV and 14 MeV, i.e. nearly the same energy as the neutrons of nuclear fusion discharges and in particular with line widths comparable to line widths expected for a thermal fusion plasma with a Maxwellian temperature around 1 kev. Due to these conditions, the measurements performed at PTB are an excellent test of the diagnostics capabilities of this neutron spectrometer for fusion application. The energy of neutrons produced in the reactions D(d,n) and T(d,n) in the laboratory frame is shown in Figure 4 for different incoming deuteron energies. The targets used at PTB to produce neutrons from these reactions are 1-2 mg/cm 2 thick deuterium and tritium loaded Ti-targets on a Ag backing in which the deuterons are fully stopped. The neutrons produced are therefore not mono-energetic except for one emission angle in the laboratory frame, which can be identified in Figure 4 as the crossing point of the kinematics for different deuteron beam energies, i.e. 16 in case of the D(d,n)-reaction and 98 in case of the T(d,n)-reaction. The very L 6

3.8 3.6 3.4 15.4 15.2 15. 3.2 3. 2.8 2.6 2.4 2.2 2. = 5 kev = 45 kev = 4 kev = 35 kev = 3 kev = 25 kev 1.8 2 4 6 8 1 12 14 16 18 Angle / degree 13. 2 4 6 8 1 12 14 16 18 Figure 4: Kinematics of the d + D 3 He + n (left) and the d + T 4 He + n (right) reaction in the laboratory frame. The neutron energies considered here are those relevant for the solid Ti-targets used at PTB with an energy of the incoming deuteron beam of = 5 kev for the D(d,n) and of = 242 kev for the T(d,n) case. 14.8 14.6 14.4 14.2 14. 13.8 13.6 13.4 13.2 = 242 kev = 2 kev = 15 kev = 1 kev = 8 kev Angle / degree small energy distribution resulting from the kinematics of the reaction is broadened due to angle scattering of the incoming deuterons in the Ti-target. Calculations of the neutron energy distribution for the targets used have been performed with the TARGET code [11] and the results are shown in Figures 5 and 6 (thin lines) for the D(d,n) and the T(d,n) case. Pulse height spectra (PHS) were been acquired in these neutron reference fields. To ensure the stability of the detector system in terms of energy calibration and pulse height resolution, measurements with a 27 Bi γ-source were made regularly. The measured γ-phs were analyzed with the same procedures that were used during the determination of the response matrix. The neutron PHS were unfolded with the maximum entropy deconvolution code MAXED [12, 13] using the measured response matrix (section 3) and "flat" default spectra (i.e. default spectra which did not contain any information of the peak structure expected for the solution spectra). The χ 2 -value of the solution was determined with a method ( L-curve method ) described briefly in [2]. The results of the unfolding for the different angles are shown by the thick lines in Figures 5 and 6. In the D(d,n) case (Figure 5), the TARGET calculation for 16 results in a Gaussian shaped distribution with a full width at half maximum (FWHM) of 84 kev, whereas the unfolded measurement results in a value of 79 kev. In the case of the measurement at, the shape of the calculation is confirmed in the unfolding, except for a small shift of the high energy edge of the distribution. The reason for this in not clear up to now and might be due to some unknown properties of the neutron producing target, namely the density distribution of the deuterium-atoms in the Ti-layer, which will influence the energy distribution of the neutron beam much more in the case than under 16. The authors would like to thank Dr. D. Schlegel for performing the calculations of the expected line shapes for the different experimental settings with the TARGET calculation tool. 7

E213, MAXED, Measured Response Matrix 3 / TARGET (Schlegel, PTB) 25 16 Φ E -1 cm -2 2 15 1 5 2. 2.2 2.4 2.6 2.8 3. 3.2 3.4 3.6 Figure 5: Calculated (thin line) and measured (thick line) energy distribution for neutrons using the d + D 3 He + n reaction at different positions of the detector (neutron emission angles of 16 and relative to the incoming deuteron beam). The unfolding was performed using the experimentally determined response matrix (see section 3). The same comparison was performed for the T(d,n) case (Figure 6). In this case, the TARGET calculation gives a value of 2 kev for the FWHM at 98, whereas the unfolding gives a significant larger value of 425 kev. The measurements and calculations for agree very well. The reason for the discrepancy between TARGET calculation and measurement under 98 is believed to be the result of the low statistics of the measured response functions, which is not as good for neutrons with energies around 14 MeV as for neutrons with energies around 2.5 MeV (see Figure 3). This view is supported by the fact that much better agreement between the TARGET calculation and measurement under 98 is obtained if a response function purely based on the RESP7 Monte Carlo calculations (red lines in Figure 3) is used for the unfolding. The result of this unfolding is shown in Figure 7. In this case, the FWHM of the TARGET calculation and the unfolded measurement for the 98 case agree very well (2 kev respectively 22 kev). The calculated response matrix has the advantage of resolving smaller structures in the results of the unfolding. But, as can be seen in Figure 7, spurious peaks appear at the low energy side of the 14 MeV neutron peak, which are due to the imperfect calculated response matrix [1]. The problem with the low statistics in the measured response functions can be solved by either making measurements for the determination of the detector response functions with longer counting times or by using methods which remove the noise due to the statistics, e.g. applying smoothing procedures or introducing smooth parameterized functions that can be fitted to the high energy part of the response. 8

2 18 16 14 E213, MAXED, Measured Resonse Matrix / TARGET (Schlegel, PTB) 98 Φ E -1 cm -2 12 1 8 6 4 2 11. 11.5 12. 12.5 13. 13.5 14. 14.5 15. 15.5 16. Figure 6: Calculated (thin lines) and measured (thick line) energy distribution for neutrons using the d + T 4 He + n reaction at different positions of the detector (neutron emission angles of 98 and relative to the incoming deuteron beam). The unfolding was performed using the experimentally determined response matrix (see section 3). Φ E -1 cm -2 2 18 16 14 12 1 8 E213, MAXED, Calculated Response Matrix / TARGET (Schlegel, PTB) 98 6 4 2 11. 11.5 12. 12.5 13. 13.5 14. 14.5 15. 15.5 16. Figure 7: Calculated (thin lines) and measured (thick lines) energy distribution for neutrons using the d + T 4 He + n reaction at different positions of the detector. The unfolding was performed using the calculated response matrix (red lines in Figure 3). 9

5. Measurements at the Joint European Torus (JET) During two campaigns in 22 and 23 at the Joint European Torus (JET), the E213 detector was installed in the roof lab with a vertical line of sight, at a distance of 19 m from the centre of the plasma. Pulse height spectra (PHS) were measured for various heated plasma scenarios [2]. The analysis of these data with the newly determined response matrix illustrates the capabilities of this spectrometer for high resolution neutron spectrometry in fusion experiments. The PHS have been unfolded with the MAXED code using "flat" default spectra (i.e. with default spectra which do not contain any information of the peak structure expected for the solution). Examples of such data analysis are shown in Figures 8 and 9. For pure ohmic heating of the plasma, the expected shape of the neutron energy distribution corresponds to a Gaussian distribution with a FWHM related to the ion temperature (T i ) of the plasma as follows: FWHM = 88.2 Ti kev 1/2 (T i in kev) [14]. The result of the unfolding of the measured PHS (shown as black histogram in the insert of Figure 8) is a Gaussian shaped line with a FWHM of 126 kev. The ion temperature calculated from this value results in T i = 2.3 kev. From other plasma diagnostics ( charge exchange measurements) it is known that a typical value for this kind of plasma scenario is T i = 2. kev, which leads to good agreement between these two independent methods. Φ E -1 cm -2 25 2 15 1 / s -1 22 2 18 16 14 12 1 8 6 4 2.3.4.5.6.7.8.9 1. E e 5 1.5 2. 2.5 3. Figure 8: Pure ohmic heating. The insert shows the measured PHS (black) as well as the PHS that results from folding the analyzed neutron spectrum with the response functions (red) [JET pulse o. 57112 57117 + 57151 57153 (52.7 s 59.7 s), 56 s total @ 17 counts per second (cps)]. 1

For other heating scenarios, like neutral beam injection (BI) or radio frequency (RF) heated plasmas, the expected line shape is not Gaussian but shows different structures in the neutron energy distribution which are produced by the non thermal, i.e. high energy, population of ions in the plasma. As an example, the measurement of an ohmic and BI heated plasma is shown in Figure 9. The shape of the fusion peak shows a broad distribution with a FWHM of around 575 kev, which is in agreement with the result of calculations (53 kev [15]). Φ E -1 cm -2 3 25 2 15 1 5 / s -1 2 18 16 14 12 1 8 6 4 2.3.4.5.6.7.8.9 1. 1.5 2. 2.5 3. 3.5 4. 4.5 E e Figure 9: Ohmic + BI heating. The insert shows the measured PHS (blue) as well as the PHS that results from folding the analyzed neutron spectrum with the response functions (red) [JET pulse o. 6139 (53. s 64. s), 11 s total @ 16 cps]. 6. Conclusions The application at JET of liquid scintillation detectors, well characterized at the PTB facilities, has shown that these detectors can be used as high resolution compact neutron spectrometers in fusion research. The new calibration technique using the broad energy distribution of the 9 Be(p,n)-reaction allows one to determine experimentally the detector response function which plays a key role in the unfolding methods used for the determination of the neutron energy spectra. This is a clear advantage as compared to the methods used at PTB up to now. Future activity will be performed at JET and PTB on the use of liquid scintillators with fast digital acquisition to optimize the system for higher count rates and further investigations will be done to study the influence of the statistics in both the PHS and in the response matrix on the results of the unfolding. 11

References [1] H. Klein, S. eumann, eutron and photon spectrometry with liquid scintillation detectors in mixed fields, uclear Instruments and Methods in Physics Research A 476 (22) 132 142 [2] et. al., Compact E213 neutron spectrometer with high energy resolution for fusion applications, Review of Scientific Instruments, Vol. 75 o. 1 (24) 3553 [3] M. Reginatto,, Bayesian methods for the analysis of pulse height spectra from liquid scintillation detectors, to be published [4] B. Esposito et. al., eutron measurements on Joint European Torus using an E213 scintillator with digital pulse shape discrimination, Review of Scientific Instruments, Vol. 75 o. 1 (24) 355 [5] D. Marocco et. al., A digital data acquisition system optimized for spectrometry with liquid scintillation detectors, these proceedings [6] H. Klein, H. Schölermann, Improvement of the light collection in scintillation detectors, IEEE transactions on uclear Science, Vol. S-26, o. 1 (1979) 373 [7] D. Schmidt, B. Asselineau, R. Böttger, H. Klein, L. Lebreton, S. eumann, R. olte, G. Pichenot, Characterization of liquid scintillation detectors, uclear Instruments and Methods in Physics Research A 476 (22) 186 189 [8] G. Dietze and H. Klein, RESP4 and EFF4: Monte Carlo Codes for the Calculation of eutron Response Functions and Detection Efficiencies for E213 Scintillation Detectors, PTB Report PTB-D-22, 1982, ISS 572-717 [9] E. Dekempeneer et al., A spectrometer for double-differential neutron-emission cross section measurements in the energy range 1.6 to 16 MeV, uclear Instruments and Methods in Physics Research A 256 (1987) 489 498 [1] A. Hildebrand, H. Park, S. Khurana, R. olte, D. Schmidt, Experimental Determination of the Response Matrix of a BC51A Scintillation Detector using a Wide eutron Spectrum: A Status Report, Laboratory Report PTB-6.42-5-1, 25 [11] D. Schlegel, Target User's Manual, Laboratory Report PTB-6.42-5-2, 25 [12] M. Reginatto, P. Goldhagen, MAXED, A Computer Code for Maximum Entropy Deconvolution of Multisphere eutron Spectrometer Data, Health Phys. 77 (1999) 579-583 [13] M. Reginatto, P. Goldhagen, S. eumann, Spectrum unfolding, sensitivity analysis and propagation of uncertainties with the maximum entropy deconvolution code MAXED, uclear Instruments and Methods in Physics Research A 476 (22) 242 246 [14] L. Ballabio, G. Gorini, J. Kallne, Energy spectrum of thermonuclear neutrons, Review of Scientific Instruments, Vol. 68 o. 1 (1997) 585 [15] L. Bertalot et. al., eutron energy measurements of trace tritium plasmas with E213 compact spectrometer at JET, Europhysics Conference Abstracts Vol. 29C (25) http://eps25.ciemat.es/papers/pdf/p1_78.pdf 12