PILOT OF AN EXTERNAL QUALITY ASSURANCE PROGRAMME FOR CRYSTALLINE SILICA

Similar documents
The Performance of Laboratories Analysing α-quartz in the Workplace Analysis Scheme for Proficiency (WASP)

How to interpret sample volume limitations stated in reference methodologies published for air sampling for Occupational Hygiene purposes

SWGDRUG GLOSSARY. Independent science-based organization that has the authority to grant

Direct reading aerosol monitors

Introduction to IH Analytical Chemistry

Art, Science or Voodoo?

A Comparison of the Performance of Samplers for Respirable Dust in Workplaces and Laboratory Analysis for Respirable Quartz

Reference Materials and Proficiency Testing. CropLife International Meeting October 2015 Gina M. Clapper

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

Accreditation of radiochemical analyses, from NAMAS to ISO 17025:2005 and beyond

EQUATIONS FOR CALCULATING EXPOSURE MANAGEMENT OBJECTIVES

STANDARD PROCEDURE: NWSP R2 (15) Polyacrylate Superabsorbent Powders Determination of the Particle Size Distribution by Sieve Fractionation

Exploration with LA-ICP-MS: Know Your Dirt

SAMPLE SIZE FOR ANALYTICAL SURVEYS USING A PRETEST POSTTEST COMPARISON GROUP DESIGN

Laboratory ID. Laboratory Name. Address. Phone. Lab Manager/Supervisor. QA Officer. Analyst(s) Auditor. Date(s) of Audit

Real-time Gravimetric Personal Monitoring of Respirable Dust QMIHS Conference 16 August 2016

MODULE 2B INDUSTRIAL HYGIENE LABORATORY ACCREDITATION PROGRAM (IHLAP) ADDITIONAL REQUIREMENTS

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

University of Tennessee Safety Procedure

Responsibilities: Effective Date: November Revision Date: February 8, VP, Facilities and Construction Management. Issuing Authority:

Respirable Crystalline Silica (RCS) Understanding the Regulations, Science & Results. Marc Hugus Manager, Business Development

HAZARD COMMUNICATION PROGRAM

Memorandum. Air Sampling

The AAFCO Proficiency Testing Program Statistics and Reporting

Globally Harmonized System Update

HAZARD COMMUNICATION PROGRAM

A Field Evaluation of the BGI and Cowl Type Gravimetric Samplers

Chemical Hygiene Plan for Laboratories

Mount Cotton Quarry Dust Investigation

This procedure describes the monitoring activities in a laboratory quality control (QC) program to ensure the quality of test results.

Improvement of traceability in the field of mass concentration measurements of particles in aerodisperse media carried out in the Russian Federation

Draft PS 18 APPENDIX A Dynamic Spiking Procedure ( ) 1.1 This appendix to Performance Specification 18

1 PRODUCT AND COMPANY IDENTIFICATION. Manufacturer. HAZARDS IDENTIFICATION Classification of the Substance or Mixture

Analytical Measurement Uncertainty

Compact Orifice FlowMeter

Introduction to Fourier Transform Infrared Spectroscopy

Draft Model Memorandum of Understanding to promote voluntary cooperation between Government and Chemical Industry to enhance Precursors Chemicals

1 PRODUCT AND COMPANY IDENTIFICATION. Manufacturer. Description: Natural occuring clay mineral SDS. GHS Safety Data Sheet. PROFILE Products, LLC

Hazard Communication Policy

Research and Development of Nanoparticle Characterization Methods

White Paper. Moisture Analyzer Routine Testing

Towards a responsible development of Nanomaterials

Reference Materials for Trade and Development: Quality and Comparability

Considerations for an Automated SEM-EDX Routine for Characterizing Respirable Coal Mine Dust

A Comparison of Fibre Counting Across Three European National Proficiency Testing Schemes

CURRICULUM VITAE. Department of Physics (D.P), College of Science (C.S), Sudan University of Science and

Quality Assurance Plan. March 30, Chemical Crystallography Laboratory

Hazard Communication Program Hazardous Chemical Program

DOE S RADIOLOGICAL TRACEABILITY PROGRAM FOR RADIOANALYTICAL MEASUREMENTS: WORK IN PROGRESS

ENVH 555. Industrial Hygiene Methods: Laboratory

SAMPLE PAGES. Hazard Communication Program. [Company name]

Industrial Hygiene Report

ENVH 555. Industrial Hygiene Methods: Laboratory

HAZARD COMMUNICATION PROGRAM

centro tecnológico Development of reference materials: a case study on packaging material as a source of taints in foods

APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY

Test Method Development and Validation as Pertaining to Microtrac Particle Size Measuring Instruments

The Globally Harmonized System (GHS) for Hazard Classification and Labelling. Development of a Worldwide System for Hazard Communication

Biol/Chem 4900/4912. Forensic Internship Lecture 5

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction

Investigation of the occurrence of microplastics from the waste water at five different textile production facilities in Sweden

LABORATORY ROUND ROBIN TEST PROJECT: ASSESSING PERFORMANCE IN MEASURING TOXICS IN PACKAGING

Manual stack emissions monitoring Performance Standard for laboratories carrying out analysis of samples

Schedule. Draft Section of Lab Report Monday 6pm (Jan 27) Summary of Paper 2 Monday 2pm (Feb 3)

Pesticide Standards. Sadat Nawaz. Pesticides Residues Analyst CSL, York, UK

International Atomic Energy Agency Learning programme: Radon gas. Module 4: Developing and Implementing a Representative Indoor Radon Survey

Hobart and William Smith Colleges. Hazard Communication Program

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

Working with Rumaila's Supply Chain

Importance of uncertainties in dose assessment to prove compliance with radiation protection standards

Proposed Testing and Research Approach for Pyrrhotite-Induced Concrete Deterioration

Nanomaterials, REACH and CLP - what is going on at the regulatory front

By Lillian Ntshwarisang Department of Meteorological Services Phone:

Introduction to Fourier Transform Infrared Spectroscopy

APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT

On the use of Deming and quadratic calibrations in analytical chemistry

MEASURING NANOPARTICLE EXPOSURE

Monitoring of Ionizing Radiations facilities Experience and challenge

Introduction to the evaluation of uncertainty

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Report of the Inter-laboratory Comparison Project 2004 on Inland Aquatic Environment

C L I E N T A LERT. Ha r m o n i z i n g Ha z a rd Co m m u n i c a t i o n : OSHA Pr o p o s e s t o. I. Ba c k g r o u n d

County of Cortland HAZARD COMUNICATION POLICY

METHOD: 1403, Issue 3 EVALUATION: FULL Issue 1: 15 August 1990 Issue 3: 15 March 2003

Oregon Department of Transportation. Geographic Information Systems Strategic Plan

ALLOWAY METHOD OUTLINE

NANOTECHNOLOGY: AN INDUSTRIAL HYGIENE PERSPECTIVE. Loren Anderson, CIH, ROH

HAZARD COMMUNICATION SAFETY PROGRAM

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012

GUIDELINES OF PROCUREMENT - CHEMICALS FOR RESEARCH CHEMICAL MANAGEMENT CENTRE

Pooja Diagnostic Laboratory Mulund, Mumbai, India

Data Collection & Sample Storage

Agilent Oil Analyzer: customizing analysis methods

WMO Global Data-Processing and Forecasting System Operational weather forecast product delivery relevant to SDSWS

AIR. Ambient, Indoor, Workplace Air and Stack Emissions Proficiency Testing Scheme. Sample Preparation Instructions Round 1

ENGINEERING AND CONSTRUCTION BULLETIN

Protecting Workers A NIOSH Nanomaterials Update

Health, Safety, Security and Environment

METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD)

Transcription:

PILOT OF AN EXTERNAL QUALITY ASSURANCE PROGRAMME FOR CRYSTALLINE SILICA MJ Shai, National Institute for Occupational Health ABSTRACT The mining industry has committed itself to reducing work-related silicosis. As part of this commitment thousands of personal air samples are being collected and analysed every year to establish the exposure of miners to respirable crystalline silica (RCS) dust. There is however no independently established scheme to assess the accuracy of the results obtained and the costly silicosis reduction programme may be being compromised by doubts about the accuracy of silica exposure measurements. Although there are a few expensive international schemes available, there is no national external quality assurance programme for the analysis of respirable crystalline silica dust currently in South Africa. A number of laboratories analysing silica are certified by the South African National Accreditation System (SANAS) but this only confirms that they have adequate management systems in place. An external quality assurance programme, or proficiency testing scheme, is a system for objectively checking laboratory results by means of an external agency. Such a programme for respirable crystalline silica dust in occupational hygiene air samples would provide the mining industry with a considerable degree of assurance that the results being reported by laboratories can be relied on to provide an accurate estimate of exposure. The National Institute for Occupational Health (NIOH) has taken the initiative and piloted an external quality assurance programme for the analysis of respirable crystalline silica dust samples using modern X-Ray diffraction or infra-red methods. The external quality assurance programme is based on the internationally recognised UK Health and Safety Laboratory Workplace Analysis Scheme for Proficiency (WASP) programme. Six laboratories agreed to participate in a six-month pilot scheme. These laboratories initially produced results with poor accuracy, but this improved in later rounds when support and advice to resolve problems was provided by the NIOH. The initial results show that, with this challenging analysis, there is a high potential for error and the production of inaccurate results. With technical support and assistance however, the results can rapidly be improved to acceptable levels. This paper will present the results of the pilot scheme. Page 1 of 8

BACKGROUND An external quality assurance (EQA) programme, or proficiency testing scheme, is a system for objectively checking laboratory results by means of an external agency 1. In the South African mining industry, thousands of personal air samples are being collected and analysed every year to establish the exposure of miners to respirable crystalline silica (RCS) dust. An EQA programme provides an external and independent quality assurance service for laboratories carrying out the analysis of RCS dust in occupational hygiene air samples taken in the workplace environment. There is currently no local external quality assurance (EQA) programme for the analysis of respirable crystalline silica dust in South Africa. Although international quality assurance schemes are available, these are expensive, turnaround times are normally long and inconvenient. The lack of participation in EQA and dependence on internal quality assurance programmes means that there is no simple way of confirming the quality of the results being produced by laboratories carrying out this analysis. This has a deleterious effect on the Elimination of Silicosis programmes put in place by the Department of Minerals and Energy and the Department of Labour. The Occupational Hygiene Section of the National Institute for Occupational Health (NIOH) undertook to develop and pilot an EQA programme for respirable crystalline silica analysis. This is based on similar international models. Participation in this programme will instil confidence in clients and the regulatory authorities that laboratories are reporting RCS results that are correct and reliable. Other benefits of a national EQA system include low cost relative to international schemes, shorter turnaround times, and easy access to advice and support. It should be noted that the EQA programme does not replace the routine internal quality control system to ensure that they are routinely producing accurate and reliable results. Nor does it replace the South African National Accreditation System (SANAS) for evaluating the management and administrative procedures of the laboratory. Page 2 of 8

METHODS Preparation of standard RCS filters Filters with a known mass of respirable crystalline silica deposited on them were prepared 2. This was carried out using a glass flask into which a small amount of respirable crystalline silica dust was placed. The dust was aerosolised with a puff of air and some of that air is then drawn through a pre-weighed filter using a personal dust sampling pump and a dust cyclone. The mass of RCS dust collected on the filter was then accurately determined from the increase in weight of the filter. Confirmation of amount of RCS deposited on standard filters There are several techniques used for determining the amount of RCS on filters. These methods include Infra Red (IR), Fourier Transform Infrared (FTIR) and X-ray diffraction (XRD) spectroscopy. The filter sample is traditionally dry ashed then mixed with KBr and pressed to make a pellet which is then analysed by infrared spectroscopy (IR KBr 3 ), however more modern methods allow a direct-on-filter analysis followed by FTIR or XRD (XRD DOF or XRD DOF 2 ). The direct-on filter method is non-destructive. There is no sample preparation which reduces the likelihood of sample losses and, since the filters are analysed directly, they can be retained for quality control purposes. The mass of RCS dust collected on the standard filters was confirmed by analysing them using both (XRD DOF) & Fourier Transform Infrared (FTIR DOF) spectroscopy. Participation in the EQA pilot scheme Seven laboratories were invited to participate in the NIOH RCS EQA pilot scheme but only four sent in results. The four laboratories, identified as A through D, participated in the pilot study. One of these laboratories used the traditional ashing and KBr pellet method followed by infrared spectroscopy to analyse the RCS dust. One other laboratory analysed the filters using direct-on-filter with both XRD and FTIR spectroscopy. Table 1: Laboratory code and method used for analysis Laboratory code Method used A XRD DOF 2 B XRD DOF 2 C IR KBr 3 D 1 XRD DOF 2 D 2 FTIR DOF 2 Page 3 of 8

Most of the participating laboratories initially struggled to achieve accurate results and to return the results within the agreed timeframe. However, as the pilot programme progressed both these aspects improved considerably. Analysis of results Two indicators were used to look at the results statistically: Firstly, a percentage difference from the target value is calculated for each sample analysed at each participating laboratory. A result within the range of ±15% from the target value was regarded as an indicator of acceptable performance during the pilot stage of the NIOH EQA programme. This is the same as is used in the UK EQA scheme 4. Secondly, an indicator known as the Z score was also used 5. This is a value that standardises the results of the EQA programme and allows comparisons within and between laboratories in different rounds. The Z-score for each sample analyzed corresponds to a point in a normal distribution. It is sometimes called a normal deviate since it describes how much a point deviates from a mean or specified point. In this case, the true value of the crystalline silica result is the specified point and the Z score describes how far from this the participating laboratory s value lies. The Z-score is measured in standard deviations. The Z-score is calculated as indicated below: Z-score= (Laboratory result target value) (SD) The SD is the standard deviation of all the gravimetric and analysis results from the NIOH and the submitting laboratory for each standard filter. Z-scores of up to ±2 are acceptable they contain 95% of values expected in a normal distribution. A Z-score of between ± 2 to 3 provides a warning and outside ±3 is regarded as unacceptable as the value is outside 99.7% of expected values in a normal distribution. Figure 1: Chart of Z-scores indication Page 4 of 8

The Z-score is the Performance Index and is used to allocate the laboratory to a Performance Category of 1 acceptable, 2 - warning or 3 - unacceptable. The single worst result is ignored to make some limited allowance for one-off problems i.e. with analytical equipment. Support As an integral part of the pilot programme, ongoing support was provided to the analysts in the participating laboratories. This mostly involved providing copies of analytical methods, sets of calibration standards, and advice on techniques and the interpretation of data. Page 5 of 8

RESULTS Table 2: Differences of laboratory results from the target values % Difference from known mass of RCS Laboratory Round 1 Round 2 Round 3 Round 4 Round 5 A +7 +7 - - -2 +19 +12 +27 3-5 -2 +6 0-1 -13-10 -8-13 -10-16 B +84 +69 - - 0 +1 +6 +1 2 +5 +9 +2-5 +1-2 -4-3 +3-4 -7 C NR NR - - NR NR NR NR NR NR NR NR NR NR NR NR +77 +91-8 -79 D 1 NR NR - - +56 +55 +33 +59 110-36 +12-10 +4 +10 +11-7 +6 +22 +6 +13 D 2 146-44 +22-8 Table 3: Z-scores Z-Scores* Laboratory Round 1 Round 2 Round 3 Round 4 Round 5 A +1 +1 - - 0 +2 +2 +4 0-1 0 +1 0 0-2 -1-1 -2-1 -2 B +11 +9 - - 0 0 +1 0 0 +1 +1 0-1 0 0-1 0 0-1 -1 C NR NR - - NR NR NR NR NR NR NR NR NR NR NR NR +10 +12-1 -11 D 1 NR NR - - +7 +7 +4 +8 +15-5 +2-1 +1 +1 +1-1 +1 +3 +1 +2 D 2 +19-6 +3-1 NR = No return Page 6 of 8

DISCUSSION AND CONCLUSION In general, most laboratories initially produced results with poor accuracy but this rapidly improved in later rounds. Support and advice to resolve problems was provided by the NIOH. The most common sources of error were the unavailability of high quality reference material and the inability to accurately prepare suitable calibration filters. In some cases, where the participating laboratory obtained very poor results, the samples were returned to the NIOH for re-analysis. Similar results to the target value were again obtained. This suggests that the source of the inaccuracy is at the participating laboratory. These initial results show that, with this challenging analysis, there is a high potential for error and the production of inaccurate results. With technical support and assistance however, the results can rapidly be improved to acceptable levels. Page 7 of 8

REFERENCES 1 UK Health and Safety Laboratory, Methods for the Determination of Hazardous Substances MDHS 71, Analytical quality in workplace air monitoring. March 1991. 2 UK Health and Safety Laboratory, Methods for the Determination of Hazardous Substances, MDHS 101, Crystalline silica in respirable airborne dusts. Direct-on-filter analyses by infrared spectroscopy and X-ray diffraction. February 2005. 3 NIOSH Manual of analytical methods 4 th edition. Silica, Crystalline by IR: Method 7602, Issue3. March 2003. 4 HSE Committee on Analytical Requirements. WASP, Information to participants. 10 th Edition. December 2004. 5 Lentner C Editor. Geigy Scientific Tables. Vol 2. 8 th Edition. 1982. Page 8 of 8