Determination of Hexavalent Chromium in Drinking Water Using Ion Chromatography

Similar documents
Determination of Cr(VI) in Water, Waste Water, and Solid Waste Extracts

Determination of Trace Cations in Power Plant Waters Containing Morpholine

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System

Determination of Inorganic Cations and Ammonium in Environmental Waters by Ion Chromatography Using the IonPac CS16 Column

METHOD 7199 DETERMINATION OF HEXAVALENT CHROMIUM IN DRINKING WATER, GROUNDWATER AND INDUSTRIAL WASTEWATER EFFLUENTS BY ION CHROMATOGRAPHY

columns IonPac SCS 1 Silica Cation Separator

The Determination of Trace Anions in Concentrated Phosphoric Acid

Determination of Trace Anions in Organic Solvents

Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples

anion-exchange column

Determination of Inorganic Anions in Wastewater by Ion Chromatography

Determination of Sulfate and Chloride in Ethanol by Ion Chromatography

Determination of trace anions in concentrated hydrofluoric acid

Determination of silicate in high-purity water using ion chromatography and online sample preparation

columns IonPac AS18 Anion-Exchange Columns

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Determination of Perchlorate in High Ionic Strength Fertilizer Extracts By Ion Chromatography

Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC

Determination of Silicate in High Purity Water Using Ion Chromatography and AutoPrep

Determination of Perchlorate in High Ionic Strength Fertilizer Extracts by Ion Chromatography

Determination of Trace Anions in Concentrated Glycolic Acid

Determination of Inorganic Anions in Drinking Water by Ion Chromatography

METHOD #: Recommended for Approval for NPDES (August, 1991)

columns IonPac AS17-C Anion Exchange Column

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection

Determination of Trace Anions in Concentrated Hydrofluoric Acid

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Application Note 149. ) as shown in the following set of reactions: 9. BrO I + 3H + 3HOI + Br 3HOI + 3I + 3H + 3I 2 O 3I 2 + 3H 2 3I 3 + 3I

Application Note 187. Brian DeBorba and Jeff Rohrer Thermo Scientific, Sunnyvale, CA, USA

Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2

Trace Level Determination of Bromate in Ozonated Drinking Water Using Ion Chromatography

Determination of an anionic fluorochemical surfactant in a semiconductor etch bath

Determination of Bromate in Bottled Mineral Water Using the CRD 300 Carbonate Removal Device

The determination of trace anions in concentrated phosphoric acid

Inline sample preparation for the determination of anions in sodium hydroxide

Determination of urea in ultrapure water by IC-MS/MS

Application Note 136

Determination of Methylamine in Drug Products

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

Determination of trace anions in high-purity waters by high volume/direct injection ion chromatography

Determination of Glyphosate by Cation-Exchange Chromatography with Postcolumn Derivatization

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory

Determination of Organic Acids and Inorganic Anions in Lithium-Containing Boric Acid-Treated Nuclear Power Plant Waters

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Determination of Verapamil Hydrochloride Purity Using the Acclaim PA Column

Determination of trace anions in high-nitrate matrices by ion chromatography

Accurate and Precise Automated Dilution and In-line Conductivity Measurement Using the AS-AP Autosampler Prior to Analysis by Ion Chromatography

Determination of trace anions in concentrated glycolic acid

Diquat 1,1 -ethylene-2,2 -bipyridium dibromide salt Paraquat 1,1 -dimethyl-4,4 -bipyridium dichloride salt Initial Preparation

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Determination of N-Methylcarbamates by Reversed-Phase HPLC

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

Thermo Scientific. Anion-Exchange Column

Dionex IonPac AS28-Fast-4µm column

Thermo Scientific Dionex IonPac AS23 Anion-Exchange Column

Determination of trace concentrations of oxyhalides and bromide in municipal and bottled waters

A Total Solution for Explosives Analysis by Reversed-Phase HPLC with a Parallel HPLC System

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Determination of Trace Cations in Concentrated Acids Using AutoNeutralization Pretreatment and Ion Chromatography

Thermo Scientific Dionex IonPac AS25 Anion-Exchange Column. Minutes

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Keywords Haloacetic acids, Water analysis, 2-D ion chromatography, ICS-3000

Determination of Trace Anions in Concentrated Bases Using AutoNeutralization Pretreatment and Ion Chromatography

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Hexavalent Chromium by Flow Injection Analysis (FIA)

columns IonPac AS22 Anion-Exchange Column

Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ )

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

METHOD: 1403, Issue 3 EVALUATION: FULL Issue 1: 15 August 1990 Issue 3: 15 March 2003

METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC)

Simultaneous Arsenic and Chromium Speciation by HPLC/ICP-MS in Environmental Waters

Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1

Waters. Ion Chromatography

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

IonPac Trace Cation Concentrator

British American Tobacco Group Research & Development. Method - Determination of ammonia in mainstream smoke

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

Fast methods for the determination of ibuprofen in drug products

Ion Chromatography: Green Chemistry for a Green Environment. Ken Kirkbride Somerset, New Jersey October 12, 2011

Chloride by Flow Injection Analysis (FIA)

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Determination of trace anions in basic solutions by single pass AutoNeutralization and ion chromatography

Direct Determination of Small Organic Acids in Sea Water by IC-MS

Tex-620-J, Determining Chloride and Sulfate Contents in Soil

Fast Protein and Peptide Separations Using Monolithic Nanocolumns and Capillary Columns

Oxygen Demand, Chemical

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

Department of Civil and Environmental Engineering, Stanford University, 473 Via Ortega, Stanford, California 94305, United States

Application of Eluent Generation for Trace Anion Analysis of Borated Waters

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

Transcription:

Application Update 144 Determination of Hexavalent Chromium in Drinking Water Using Ion Chromatography INTRODUCTION Hexavalent chromium, Cr(VI), is the most toxic form of the metal chromium, a primary drinking water contaminant in the U.S. Dissolved hexavalent chromium can be determined as chromate ( ) by ion chromatography in drinking water, groundwater, and industrial wastewater effluents as described in U.S. EPA Method 218.6 and Dionex Technical Note 26. 1,2 Dionex Technical Note 26 uses a 25-µL injection onto a high-capacity IonPac AS7 anion exchange column to separate Cr(III) from Cr(VI) in four minutes. The product of the postcolumn reaction between Cr(VI) and diphenylcarbazide is detected by absorbance at 53 nm, yielding a method detection limit of.4 µg/l in reagent water. The California Department of Health Services (DHS) recently issued a new Public Health Goal (PHG) of 2.5 µg/l for total chromium and.2 µg/l for Cr(VI). In January 21, California DHS added Cr(VI) to the list of unregulated chemicals that must be monitored. As a result of this regulation, public water systems are now monitoring for Cr(VI) in drinking water. 3 EPA Method 218.6 does not allow sufficient sensitivity for analysis at the California PHG level of.2 µg/l. This application update describes modifications to Method 218.6 that significantly increase sensitivity over the existing method. The modifications include lower eluent and postcolumn reagent (PCR) flow rates, a larger reaction coil, and a larger injection volume. The resulting MDL for Cr(VI) as of.2 µg/l is more than sufficient for determinations at the California PHG level. EQUIPMENT A Dionex DX-6 chromatography system consisting of: GS5 Gradient Pump with Vacuum Degas Option AS5 Automated Sampler with Chromatography Compartment AD25 UV/Visible Absorbance Detector 1. ml sample syringe for AS5 (Dionex PN 5566) PC1 Postcolumn Pneumatic Delivery Package, 4-mm (Dionex PN 561) 75-µL knitted reaction coil (Dionex PN 42631) PeakNet 6.1 Chromatography Workstation Syringe filters (Gelman IC Acrodisc.µm, PN 4483) Application Update 144 1

REAGENTS AND STANDARDS Prepare all solutions from analytical reagent grade chemicals (when available). Deionized water, 17.8 MΩ-cm or better Ammonium sulfate, (Mallinckrodt Gen AR # 7725) Ammonium hydroxide (Sigma A6899) 1,5-diphenylcarbazide (JT Baker K62-3) Methanol, HPLC grade (Fisher Optima A454-4) Sulfuric acid, 95 98% (JT Baker Instra-Analyzed #9673) Potassium dichromate (JT Baker 4765-1) CONDITIONS Guard Column: IonPac NG1 Guard 4 5 mm (Dionex PN 39567)I Analytical Column: IonPac AS7 Analytical 4 25mm (Dionex PN 35393) Eluent: 25 mm ammonium sulfate/ mm ammonium hydroxide Eluent Flow Rate: 1. ml/min Sample Volume: 1 µl partial loop with 1 µl cut volume from 11-µL loop Postcolumn Reagent: 2 mm diphenylcarbazide, 1% methanol, 1 N sulfuric acid PCR Flow rate:.33 ml/min Detection: UV/Vis absorbance, 53 nm Noise: 25 5 µau peak-to-peak Backpressure: 12 13 psi Run Time: 1 min (Retention time = 6 7 min) PREPARATION OF SOLUTIONS AND REAGENTS Eluent: 25 mm ammonium sulfate mm ammonium hydroxide Dissolve 66. g of ammonium sulfate in about 1 L of reagent water and add 13. ml of 29% ammonium hydroxide. Dilute to 2. L with water. Sample Adjustment Buffer: 25 mm ammonium sulfate mm ammonium hydroxide Dissolve 3.3 g of ammonium sulfate in about 75 ml of reagent water and add 6.5 ml of 29% ammonium hydroxide. Dilute to 1 ml with water. Postcolumn reagent: 2 mm diphenylcarbazide 1% methanol 1 N sulfuric acid Add 28 ml of 98% sulfuric acid to about 5 ml of water in a 1. L volumetric flask. (Caution: may get hot.) Mix and allow to cool. Add.5 g of 1,5-diphenylcarbazide to about 75 ml of HPLC-grade methanol in a 1 ml volumetric flask, and sonicate to dissolve. Bring to volume with methanol, mix, and add to the cooled sulfuric acid solution. Dilute to 1. L with deionized water, mix, and transfer to the pressurized PCR reagent container. The PCR reagent is stable for several days. Prepare fresh as needed. Standard Add.283 g of potassium dichromate (dried at 1 C to a constant weight) to about 5 ml of deionized water in a 1 ml volumetric flask. Dissolve and bring to volume with deionized water. Or, prepare a 1 mg/l stock solution of Cr(VI) as from a commercially available standard (J. T. Baker, Phillipsburg, NJ). Store the stock standard at 4 C. Prepare working standards fresh daily. Adjust the ph to 9. 9.5 by adding 1 ml of sample adjustment buffer per 1 ml of final volume before bringing to final volume. SAMPLE PREPARATION Clean all sample collection equipment and containers with 1:1 HNO 3 and rinse well with deionized water before use. Collect samples in amber glass bottles with plastic lined caps. Do not filter the samples at the time of collection, but immediately add the sample adjustment buffer dropwise until the sample ph falls in the range of 9. 9.5. Be careful not to contaminate the sample while measuring the ph. Most drinking water samples can be adjusted to ph 9. 9.5 by adding 1 ml or less of the adjustment buffer per 1 ml of sample, which introduces an acceptable 1% dilution error. For more difficult samples, start with a known amount of sample and accurately measure the amount of buffer added so that the amount of Cr(VI) as determined by IC can be corrected for dilution. Cool to 4 C and hold at 4 C during transport and storage. Analyze samples within 24 hr of collection to minimize the potential loss of Cr(VI) through chemical reduction. 4 2 Determination of Hexavalent Chromium in Drinking Water Using Ion Chromatography

Filter drinking water samples through.µm Acrodisc IC syringe filters (Gelman, Ann Arbor, MI) just prior to injection. Discard the first 3 µl of filtrate and filter the remainder directly into a clean plastic autosampler vial. Qualify filters by analyzing a reagent water blank and a 1-µg/L Cr(VI) as standard that has been passed through the filter. The blank should be free of peaks within the retention time window of chromate, and the recovery of the 1-µg/L standard should fall between 8% and 12%. SYSTEM PREPARATION AND SETUP Verify that the pump flow rate is within specifications and recalibrate if necessary. A GP5 should deliver water at 1. ±.5 ml/min against a constant backpressure of 2 psi. Verify that the UV/Vis Absorbance Detector wavelength accuracy is within specifications and recalibrate if necessary. It is a good practice to periodically record the visible lamp output (i.e., the reference cell current in na) and elapsed time as an aid in troubleshooting. Consult the pump or detector manuals for procedural details. The precision and accuracy of the AS5 will vary depending on the mode of injection. The most accurate and precise injections are made with a calibrated sample loop in the full-loop injection mode, which aspirates a total of four times the sample volume to flush the sample loop and make the injection. The largest fullloop injection possible with the AS5 is 3 µl. To inject 1 µl, use the partial-loop injection mode with an 11-µL sample loop, and a programmed sample loop volume of 11-µL and cut volume of 1 µl. This injection procedure should provide peak area precision of <1% RSD. Refer to the AutoSelect AS5 Autosampler operator s manual (Document No. 31169) for a complete discussion of the different injection modes. Install a 1-mL sample syringe and set the syringe speed to four or five to make faster large-loop injections. Enter the correct sample loop size and sample syringe volume in the AS5 Plumbing Configuration screen. Configure the IC with the PCR system as depicted in Figure 3 of Dionex Technical Note 26, and as described in the PC1 Postcolumn Delivery System installation instructions. Pump the eluent at 1. ml/min and set the PC1 pneumatic pressure to 7 psi. To measure the PCR flow rate, collect the effluent from the detector (i.e., the total flow from the IC pump and the PCR module) in a 1-mL graduated cylinder for 5 min. The PCR flow rate is the difference between the total flow rate and that of the IC pump. Adjust the air pressure of the postcolumn delivery module (PC1) and remeasure the flow rate until the correct PCR flow rate of.33 ml/min is established. Variations in the PCR flow rate affect the postcolumn reaction time, ph, dilution, mixing rate, and ratio of the reactants. Stable day-to-day results depend on a well-controlled PCR flow rate. Confirm this flow rate on a daily basis or whenever detector response for a calibration check standard deviates beyond quality control acceptance criteria. The storage solution that the AS7 is shipped with is 3 mm nitric acid. After equilibrating the column with eluent for 6 min, analyze a system blank of 1 µl of reagent water. An equilibrated system has a background signal of less than 2 mau and peak-to-peak noise of less than 5 µau. There should be no peaks eluting within the retention time window of the chromate anion. The column is equilibrated when two consecutive injections of a standard produce the same retention time for chromate. RESULTS AND DISCUSSION Effect of Reaction Coil and Injection Volume The 375-µL reaction coil and higher flow rates recommended in Dionex Technical Note 26 are adequate for the rapid determination of Cr(VI) as at the 1-µg/L level, but a standard 375-µL knitted reaction coil does not provide the maximum peak response for chromate. 5 To optimize the sensitivity of this method, Application Update 144 3

Column: IonPac NG1, AS7 Eluent: 25 mm (NH 4 1 mm NH 4 Inj. Volume: 25 µl Reagent: 2 mm diphenylcarbazide 1% CH 3 RXN Coil: 375, 75, and 15 µl Detector: UV/Vis (53 nm) Cr(VI) Response.4.3.2.1 Peak Height (AU) at 1. ml/min 375 75 15 Reaction coil volume Figure 1. Effect of reaction coil volume on chromate peak response. Conditions: guard column, IonPac NG1; analytical column, IonPac AS7; eluent, 25 mm ammonium sulfate/ mm ammonium hydroxide; flow rate, 1. (A and C) or 1.5 (B and D) ml/min; postcolumn reagent, 2 mm diphenylcarbazide / 1% methanol / 1. N sulfuric acid; reaction coil volume, 375 15 µl as indicated; postcolumn flow rate,.33 (A and C) or.5 (B and D) ml/min; detection, UV/Vis at 53 nm; injection volume, 25 µl; peaks, chromate (1 µg/l). the responses obtained with 375-µL, 75-µL, and 15- µl reaction coils were compared. Figure 1 shows the effect of reaction coil volume on chromate peak height and area at two different flow rates. In both cases, the postcolumn reagent flow rate was adjusted to one-third the eluent flow rate by varying the applied pneumatic pressure. At an eluent flow rate of 1. ml/min (A and C) the postcolumn reagent flow rate was.33 ml/min. At an eluent flow rate of 1.5 ml/min (B and D) the postcolumn reagent flow rate was.5 ml/min. Increasing the reaction coil volume from 375 µl to 75 µl significantly increases the peak response, while the change from 75 µl to 15 µl only marginally increases the response. Larger coil sizes require greater pneumatic pressure to deliver the necessary PCR flow rate. Also, higher eluent flow rates require greater pneumatic pressure to deliver the necessary PCR flow (A) (B) Peak Height (AU) at 1.5 ml/min Peak Area (AU*min) at 1. ml/min (C) (D) Peak Area (AU*min) at 1.5 ml/min Column: IonPac NG1, AS7 Eluent: 25 mm (NH 4 1 mm NH 4 Flow: 1. ml/min Inj. Volume: 25 1 µl Reagent: 2 mm diphenylcarbazide 1% CH 3 RXN Coil: 75 µl Detector: UV/Vis (53 nm) 25 5 75 1 Injection Volume (µl) Figure 2. Effect of injection volume on chromate peak response. Conditions: as for Fig. 1, except; flow rate, 1. ml/min; postcolumn flow rate,.33 ml/min; reaction coil volume, 75 µl; injection volume, 25 1 µl as indicated. rate. For example, a pneumatic pressure of over 1 psi was required to deliver the PCR reagent at a flow rate of.5 ml/min against an eluent flow rate of 1.5 ml/min through a 15-µL reaction coil. We recommend using a 75-µL reaction coil with an eluent flow rate of 1. ml/min and a PCR flow rate of.33 ml/min. This combination provides nearly the maximum peak response while requiring a modest pneumatic pressure of about 7 psi to deliver the postcolumn reagent at the necessary flow rate. Greater sensitivity can be gained by increasing the volume of sample injected, but too large a sample can cause retention time shifts or loss of efficiency when column overloading leads to excessive peak distortion. The effect of injection volume on chromate peak response was studied using the conditions described above. Figure 2 shows a linear increase in both chromate peak height and area as the injection volume is increased from 25 µl to 1 µl. We did not test larger injection volumes because a 1 µl injection volume is the largest that can be made with the AS5 Autosampler. 4 Determination of Hexavalent Chromium in Drinking Water Using Ion Chromatography 17325 Cr(VI) Peak Area (AU*min).3.2.1 Area, R 2 =.9997 Height, R 2 =.9943.15.1.5 Cr(VI) Peak Height (AU) 17326

Column: IonPac NG1, AS7 Eluent: 25 mm (NH 4 1 mm NH 4 Flow: 1. ml/min Inj. Volume: 1 µl Reagent: 2 mm diphenylcarbazide 1% CH 3 RXN Coil: 75 µl Detector: UV/Vis (53 nm) Sample: 1. µg/l Cr(VI).2 Cr(VI) Peak height, AU.15.1.5 y =.14x R 2 = 1 y =.2x R 2 =.9999 5 1 Hexavalent chromium, µg/l Figure 4. Calibration curve for Cr(VI) as chromate..3.2.1 Peak area, AU.min AU Cr(VI) as standard obtained by using the optimized conditions described above. Figure 4 summarizes the calibration data for chromate anion obtained by using the partial-loop injection mode to make 1-µL injections of Cr(VI) as standards at,.4,.1,.2,.5, 1, 2, 5, 1, and 2 µg/l. The calibration curve is linear over the calibra- 2 4 Minutes 6 8 17327 Table I. Method Detection Limits for Chromate Based on a 1-µL Injection Figure 3. Determination of chromate using optimized EPA Method 218.6. Conditions: as for Fig. 2, except injection volume, 1 µl; peaks, chromate (1. µg/l). Sample Adjustment Buffer Method 218.6 requires a solution consisting of 33 g/l ammonium sulfate and 65 ml/l ammonium hydroxide to adjust the sample ph. However, the largeloop injection used in this application update increases the possibility of overloading the analytical column if the sample ionic strength is too high. A solution consisting of 33 g/l ammonium sulfate and 65 ml/l ammonium hydroxide is a suitable substitute to adjust the sample ph and provides results comparable to those obtained by using the Method 218.6 buffer (data not shown). Optimized Method Performance By using a larger volume reaction coil, lower eluent flow rate and increased injection volume, this updated method provides greater than a 1-fold increase in the chromate peak area compared to the response obtained by using the standard conditions specified in Method 218.6. Figure 3 shows a chromatogram of a 1.-µg/L Chromate Conc. Std. Dev. RSD MDL* (µg/l) (µg/l) (%) (µg/l).1.6 6.986.18.2.56 3.193.18 * MDL = (Std. Dev.) x (t s, 99 ), where t s, 99 = 3.14 for n = 7. tion range of.1 1 µg/l for Cr(VI) as with a coefficient of determination of.9999. Method Detection Limits (MDLs) for Cr(VI) as are summarized in Table 1. The MDL is a measure of the precision of replicate injections of a low-level solution and is defined as the minimum concentration of an analyte that can be identified, measured, and reported with 99% confidence that the analyte concentration is greater than zero. 6 We determined the MDL for Cr(VI) as by analyzing seven replicates of reagent water fortified with Cr(VI) as at two concentrations levels of.1 and.2 µg/l (i.e., about 3 5 times the estimated instrument detection limit). Both levels produced a calculated MDL value of.18 µg/l. This permits a minimum limit (ML) for quantitation of.6 µg/l for Cr(VI) as, which is Application Update 144 5

Column: IonPac NG1, AS7 Column: IonPac NG1, AS7 Eluent: 25 mm (NH 4 1 mm NH 4 Flow: 1. ml/min Inj. Volume: 1 µl Reagent: 2 mm diphenylcarbazide 1% CH 3 RXN Coil: 75 µl Detector: UV/Vis (53 nm) Sample: Sunnyvale, CA tap water spiked with.2 µg/l Cr(VI) Solutes: A..55 µg/l Cr(VI) B..245 µg/l Cr(VI) Eluent: 25 mm (NH 4 1 mm NH 4 Flow: 1. ml/min Inj. Volume: 1 µl Reagent: 2 mm diphenylcarbazide 1% CH 3 RXN Coil: 75 µl Detector: Sample: 12% UV/Vis (53 nm) Sunnyvale, CA tap water spiked with.2 µg/l Cr(VI) + -2 mg/l of Cl -, Sulfate Chloride.2 AU Chromate Chromate A B Relative Cr(VI) Peak Response 1% 8% 6% 4% 2% 2 4 6 8 Minutes % 25 5 75 1 2 Anion Concentration (mg/l) 17323 17328/9 Figure 5. Determination of chromate in drinking water. Conditions: as for Fig. 3, except sample, buffered Sunnyvale, CA drinking water (A) and drinking water spiked with.2 µg/l chromate (B); peaks, (A) chromate (.55 µg/l) and (B) chromate (.245 µg/l) adequate for routine analysis at the California PHG level of.2 µg/l. Fig. 5 shows chromatograms obtained by using the optimized conditions of a Sunnyvale, CA, tap water blank (A) and tap water sample spiked with Cr(VI) as at the PHG level of.2 µg/l (B). In both cases, the sample was adjusted to ph 9 by using the solution consisting of 33 g/l ammonium sulfate and 65 ml/l ammonium hydroxide. The presence of the ammonium sulfate and ammonium hydroxide in the sample did not adversely affect the chromate peak shape and a recovery of 96% was obtained for the Cr(VI) as spike at this level. The tap water blank contained a background level of.6 µg/l chromate. Although the IonPac AS7 column specified in Method 218.6 has a relatively high capacity of Figure 6. Effect of common anions on chromate peak response. Conditions: as for Fig.3, except sample, buffered Sunnyvale, CA drinking water spiked with.2 µg/l chromate to which 2 mg/l of sulfate or chloride was added, as indicated. 1 µequivalents/column, the large injection volume used in this application update increases the possibility of interference from other anions in the sample. Hence, the effect of chloride and sulfate on Cr(VI) as response was investigated, as some drinking and ground waters can contain elevated levels of these common anions. Increasing concentrations of sulfate or chloride were added to a series of tap water samples that had been adjusted to ph 9 with the recommended buffer and spiked with.2 µg/l Cr(VI) as. Figure 6 shows the effect of common anions on chromate peak response, relative to the peak area for.2 µg/l Cr(VI) as spiked in tap water containing no added sulfate or chloride. This method provides acceptable performance, for example greater than 8% recovery in the presence of up to 1 mg/l chloride or 2 mg/l sulfate. Typical ground and drinking waters are not expected to contain more than 2 mg/l chloride or 5 mg/l sulfate. 6 Determination of Hexavalent Chromium in Drinking Water Using Ion Chromatography

CONCLUSIONS U.S. EPA Method 218.6, as published, does not allow sufficient sensitivity to determine hexavalent chromium (i.e., Cr(VI) as ) at the California PHG level of.2 µg/l. Modifications to the method, including the use of a lower eluent flow rate and larger reaction coil (to increase reaction time) and a larger injection volume, significantly increase the sensitivity of Method 218.6, resulting in an MDL for Cr(VI) as of.2 µg/l. These modifications allow a minimum limit (ML) of quantitation for Cr(VI) as of.6 µg/l, which is more than sufficient for analysis at the California PHG level. Calibration was linear over the range of.1 1 µg/l, and quantitative recoveries were obtained for Cr(VI) as spiked at.2 µg/l in drinking water. The modified method provides acceptable performance, in terms of peak shape and recovery, in the presence of up to 1 mg/l chloride or 2 mg/l sulfate. REFERENCES 1. Determination of Dissolved Hexavalent Chromium in Drinking Water, Groundwater, and Industrial Wastewater Effluents by Ion Chromatography; U.S. Environmental Protection Agency, Method 218.6; Cincinnati, (1991). 2. Dionex Corporation. Determination of Cr(VI) in Water, Waste Water, and Solid Waste Extracts; Technical Note 26; Sunnyvale, CA. 3. Governor Signs Bill Speeding Water Testing, The Los Angeles Times, Valley Ed., September 3, 2, p. 1. 4. Chromium-6 (Hexavalent Chromium) in Drinking Water, California Department of Health Services, Update, August 22, 21. 5. Raimund Roehl, California Department of Health Services, personal communication, May 21. 6. The Determination of Inorganic Anions in Water by Ion Chromatography; U.S. Environmental Protection Agency, Method 3.; Cincinnati, Ohio (1993). SUPPLIERS VWR Scientific Products, 3745 Bayshore Blvd., Brisbane, CA 945 USA, Tel. 8-935, www.vwrsp.com. Fisher Scientific, 711 Forbes Ave., Pittsburgh, PA 15219 USA, Tel. 8-766-7, www.fishersci.com. Gelman Sciences, 6 S. Waygner Road, Ann Arbor, MI 4816-1448 USA, 8-521-152 www.pall.com/gelman. Sigma Chemical Co., P.O. Box 1458, St. Louis, MO 63178 USA, Tel. 8-325-31, www.aldrich.sial.com. Mallinckrodt Baker, 222 Red School Lane, Phillipsburg, NJ 8665 USA, Tel. 8-582537, www.jtbaker.com. Application Update 144 7

AutoSelect is a trademark, and IonPac and PeakNet are registered trademark of Dionex Corporation. Printed on recycled and recyclable paper. Dionex Corporation Dionex Corporation Dionex U.S. Regional Offices Dionex International Subsidiaries 1228 Titan Way Salt Lake City Technical Center Sunnyvale, CA (48) 737-8522 Austria (1) 616 51 25 Belgium (32) 3-353 42 94 Canada (95) 844-965 China (852428 3282 Denmark (45) 36 36 9 9 P.O. Box 363 1515 West 22 South, Suite A Westmont, IL (63) 789-366 France 1 39 3 1 1 Germany 6126-991- Italy (6) 66 51 5 52 Japan (6) 6885-1213 The Netherlands (161) 43 43 3 Sunnyvale, CA Salt Lake City, UT Houston, TX (281) 847-5652 Switzerland (625 99 66 United Kingdom (1276) 691722 9488-363 8 Determination 84119-1484 of Hexavalent Atlanta, GA (77) Chromium 4381 * Designed, in Drinking developed, and manufactured Water under Using an NSAI registered Ion ISO 91 Chromatography Quality System. (48) 737-7 (81) 979292 Marlton, NJ (856) 596-69 LPN 1495 pdf 1/3 23 Dionex Corporation