A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas

Similar documents
Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

Methanol Extraction of high level soil samples by USEPA Method 8260C

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

Optimizing. Abstract: is standardd. procedures. altered to

Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

Solid Phase Micro Extraction of Flavor Compounds in Beer

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler

Optimizing Standard Preparation

Maximizing Production While Minimizing Costs

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

Solid Phase Micro Extraction of Flavor Compounds in Beer

Analytical Trap Comparison for USEPA Method 8260C

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Maximizing Sample Throughput In Purge And Trap Analysis

Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

EPA TO-17 Volatile Organic Compounds

Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

2017 EPA Method Update Rule and EPA Method 624.1

Ed George and Anaïs Viven Varian, Inc.

Determination of Volatile Organic Compounds in Air

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

Performance of a Next Generation Vial Autosampler for the Analysis of VOCs in Water Matrices

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: CTRA Project: 6454 T Lab Order: A 6454 Aqueous 3/1/2013.

Application News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: Cascade Thornapple River Assoc. Project: Water Analysis Lab Order:

Volatile Organic Compounds in Every Day Food

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

A Single Calibration for Waters and Soil Samples Performing EPA Method Anne Jurek Applications Chemist

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Target Compound Results Summary

AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL

Reduced VOC Sample Analysis Times Using a New Dual Purge-and-Trap System

National Pollutant Discharge Elimination System (NPDES); BADCT Limits for Volatile Organic Compounds

ANALYTICAL REPORT. Job Number: Job Description: Transform Complete

CALA Directory of Laboratories

Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS

ANALYTICAL REPORT. Results relate only to the items tested and the sample(s) as received by the laboratory. Page 1 of 5

Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS

UCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring

Rapid Determination of TO-15 Volatile Organic Compounds (VOCs) in Air

The following report includes the data for the above referenced project for sample(s) received on 5/15/2017 at Air Toxics Ltd.

Meeting NJ Low Level TO-15 Air Testing Method Requirements

STANDARD OPERATING PROCEDURES

McCAMPBELL ANALYTICAL INC Willow Pass Road Pittsburg CA

OREGON Environmental Laboratory Accreditation Program ORELAP Fields of Accreditation ALS Environmental - Simi Valley

METHOD 5021 VOLATILE ORGANIC COMPOUNDS IN SOILS AND OTHER SOLID MATRICES USING EQUILIBRIUM HEADSPACE ANALYSIS

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Techniques for Optimizing the Analysis of Volatile Organic Compounds in Water Using Purge-and-Trap/GC/MS Application

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Thermal Desorption Technical Support

Building 280A and Building 450 Soil Sampling Results Richmond Field Station Site Berkeley Global Campus at Richmond Bay, Richmond, California

Associates. Project No.: October 25, 2017

Copies: Dave Favero, RACER File

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.

Volatile Organic Compounds in Water PBM

Standard Operating Procedure for the Analysis of Purgeable Organic Compounds (VOC s) in Water by Capillary Column Gas Chromatography/Mass Spectrometry

ANALYTICAL REPORT. Job Number: SDG Number: Job Description: Joint Base Cape Cod

Enhanced Preconcentrator for the Analysis of Vapor Phase Volatile Organic Compounds

Date Issued: April 01, 2013 Expiration Date: June 30, 2013

Method 1624 Revision C Volatile Organic Compounds by Isotope Dilution GCMS

ANNE JUREK. Abstract: soils and are found. in can also with GC/MS. poster. in series. option for. There are problems. analytes. in methanol.

SUMMARY REPORT OF AIR MONITORING FOR LEED CERTIFICATION PHASE 2 SWANSFIELD ELEMENTARY SCHOOL 5610 CEDAR LANE COLUMBIA, MD PREPARED FOR:

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS

ANALYTICAL RESULTS. Project: Mayflower, AR Pipeline Incident

Study of Residual Solvents in Various Matrices by Static Headspace

R.E.A.C.T. Roxbury Environmental Action CoaliTion P.O. Box 244 Ledgewood, N.J Website:

Headspace Technology for GC and GC/MS: Features, benefits & applications

317 Elm Street Milford, NH (603) Fax (603)

Table 1 Soil and Groundwater Sampling Summary Table

ORGANIC COMPOUNDS IN WATER BY GAS CHROMATOGRAPHY/MASS SPECTROMETRY USING NITROGEN PURGE GAS

Laboratory Report Number(s) Lone Tree Road (Deep) 10/5/15, 10/19/15, 6/13/16. Address Sample Date(s) Analyses Requested

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

METHOD 5030C PURGE-AND-TRAP FOR AQUEOUS SAMPLES

STONY HOLLOW LANDFILL, INC S. Gettysburg Ave. Dayton, OH (937) (937) Fax

Detection of Environmental Contaminants Caused by the Oil Spill in the Gulf of Mexico by GC and Hplc

/J~ GOLDER. Golder Associates Inc. golder.com. Project No April 25, 2018

2. September 2015 Sub-Concrete Slab Soil Investigation

Transcription:

A Comparison of Volatile Organic Compound When Using Nitrogen as a Gas Application Note By: Anne Jurek Abstract For many years Helium has been the gas of choice for purging Volatile Organic Compounds (VOCs). However, in the past few years, the price and demand for helium gas has increased substantially, thus making the use of Helium for the purge gas in and Trap (P&T) very expensive. The expense of Helium has provoked interest in finding a viable alternative purge gas. This application note investigates VOC compound responses when purged using Helium and Nitrogen in order to validate Nitrogen as a possible alternative to Helium for P&T applications. Introduction Teledyne Tekmar developed a combination P&T Concentrator/ Vial Autosampler, the Atomx. The Atomx was developed to fully automate water, soil, and methanol extraction in accordance with the USEPA methods for volatile analyses. One of the beneficial features of the Atomx highlighted in this study is the use of an electronic mass flow controller that is calibrated for either Nitrogen or Helium. The controller is employed for both fritted glass sparging used for aqueous samples and in vial sparging used in soil applications. Since the mass flow controller is electronic, flow rates can be programmed via the software interface for various modes of operation. This patent pending ability allows for the end user to simplify the potential switch by simply changing the configuration rather than manually adjusting pressures and flows as seen in traditional regulator/needle flow controller systems. In this study, data was collected to evaluate compound response when using Nitrogen as a purge gas as opposed to the traditional Helium purge gas. Furthermore, as water samples are purged in the sparge vessel and soils are purged in the sample vial, an additional comparison was done to see if the analytes responded differently when purged in the vial with Nitrogen in contrast to a Helium purge. Experimental-Instrument Conditions The Atomx, an Agilent 7890A GC and a 5975C inert XL MSD were used for this analysis. The Atomx was equipped with a #9 adsorbent trap. Tables 1 and 2 display the GC, MSD conditions while Tables 3 and 4 display the P&T Concentrator/Autosampler conditions for water and soil matrices respectively.

GC Parameters MSD Parameters GC: Agilent 7890A MSD: 5975C Inert XL Column: Oven Program: J&W Scientific DB-VRX 30m x 0.250mm x1.4um 35 C for 4 min; 16 C/min to 85 C for 0 min; 30 C /min to 210 C for 3 min, 14.29 min runtime Source: 230 C Quad: 150 C Inlet: 220 C Solvent Delay: 0.5 min Column Flow 1.2mL/min Scan Range: m/z 35-300 Gas: Helium Scans: 4.51 scans/sec Split: 80:1 Threshold: 400 Pressure: 9.3 psi MS Transfer Line Temp. Inlet Split/Splitless Tables 1 & 2: GC and MSD Parameters 230 C Atomx Water Parameters Variable Value Variable Value Valve Oven Temp 140 C Dry Flow 100mL/ min Transfer Line Temp 140 C Dry Temp 20 C Sample Mount Temp 90 C Methanol Needle Rinse Off Water Heater Temp 90 C Methanol Needle Rinse Volume 3.0mL Sample Vial Temp 20 C Water Needle Rinse Volume 7.0mL Sample Equilibrate Time 0.00 min Sweep Needle Time 0.50 min Soil Valve Temp 125 C Desorb Preheat Temp 245 C Standby Flow 10mL/ min GC Start Signal Start of Desorb Ready Temp 40 C Desorb Time 2.00 min Condensate Trap Standby 45 C Drain Flow 300mL/min Presweep Time 0.25 min Desorb Temp 250 C Prime Sample Fill Volume 3.0mL Methanol Glass Rinse On Sample Volume 5.0mL Number of Methanol Glass Rinses 1 Sweep Sample Time 0.25 min Methanol Glass Rinse Volume 3.0mL Sweep Sample Flow 100mL/min Number Of Bake Rinses 1 Sparge Vessel Heater Off Water Bake Rinse Volume 7.0mL Sparge Vessel Temp 20 C Bake Rinse Sweep Time 0.25 min Prepurge Time 0.00 min Bake Rinse Sweep Flow 100mL/min Prepurge Flow 0mL/min Bake Rinse Drain Time 0.40 min Time 11.00 min Bake Time 4.00 min Flow 40mL/min Bake Flow 250mL/min Temp 20 C Bake Temp 280 C Condensate Temp 20 C Condensate Bake Temp 200 C Dry Time 0.50 min Table 3: Atomx Water Parameters (Parameters highlighted in yellow were not used.)

Calibration Atomx Soil Parameters Variable Value Variable Value Valve Oven Temp 140 C Time 11.0 min Transfer Line Temp 140 C Flow 40mL/min Sample Mount Temp 90 C Temp 20 C Water Heater Temp 90 C Condensate Temp 20 C Sample Vial Temp 40 C Dry Time 1.00 min Prepurge Time 0.00 min Dry Flow 100mL/ min Prepurge Flow 0mL/min Dry Temp 20 C Preheat Mix Speed Off Methanol Needle Rinse On Sample Preheat Time 0.00 min Methanol Needle Rinse Volume 3.0mL Soil Valve Temp 125 C Water Needle Rinse Volume 7.0mL Standby Flow 10mL/min Sweep Needle Time 0.25 min Ready Temp 40 C Desorb Preheat Temp 245 C Condensate Temp 45 C GC Start Signal Start of Desorb Standby Presweep Time 0.25 min Desorb Time 2.00 min Water Volume 10mL Drain Flow 300mL/min Sweep Water Time 0.25 min Desorb Temp 250 C Sweep Water Flow 100mL/min Bake Time 4.00 min Sparge Vessel Heater Off Bake Flow 250mL/min Sparge Vessel Temp 20 C Bake Temp 280 C Mix Speed Medium Condensate Bake Temp 200 C Table 4: Atomx Soil Parameters (Parameters highlighted in yellow were not used.) A 50ppb working calibration stock standard was prepared in methanol. Calibration standards were prepared in a 50mL volumetric flask and filled to volume with de-ionized water. In this study, a linear calibration was performed for both the water and the soil matrices for 95 analytes. The range for the water study was 0.5-200ppb and the soil range was 1.0-200ppb. The water standards were transferred to headspace free 40mL vials for analysis while the soil standards were transferred to 40mL vials in 5mL aliquots. A 5.0 milliliter (ml) purge volume was used for the water curve. For the soil curve, a 5g sample with 10mLs of reagent water was simulated. Conditions and specifications outlined in USEPA Method 8260 were utilized for both matrices. The calibration data was analyzed using Agilent Chemstation software. The average compound response for the water and soil matrices with the Helium and Nitrogen purge gases is outlined in Table 5. The relative response factors of all of the analytes of interest were evaluated for linearity and response and the average %RSD of the respective curves are summarized in Table 6. Method Detection Limit (MDL) A statistical determination of the MDL s was determined for all of the compounds by analyzing seven replicate standards of a low calibration standard. The average detection limits are provided in Table 6.

Compound Water Soil Pentafluorobenzene (IS) N/A N/A N/A N/A Dichlorodifluoromethane 0.337 0.517 0.807 0.507 Chloromethane 0.551 0.655 0.993 0.692 Vinyl Chloride 0.675 0.620 1.194 0.763 Bromomethane 0.379 0.393 0.717 0.660 Chloroethane (Ethyl Chloride) 0.461 0.415 0.693 0.501 Trichlorofluoromethane 0.962 0.735 1.526 1.035 Diethyl Ether 0.604 0.484 0.783 0.540 1,1-Dichloroethene 0.839 0.686 1.416 0.528 Carbon Disulfide 1.069 1.417 2.423 1.111 1,1,2-Trichlorofluoroethane (Freon ) 0.283 0.349 0.805 0.366 Iodomethane 0.469 0.475 0.801 0.462 Allyl Chloride 0.713 0.615 0.859 0.441 Methylene Chloride 0.694 0.614 1.067 0.377 Acetone 0.263 0.203 0.252 0.194 trans-1,2-dichloroethene 0.675 0.712 0.874 0.768 Methyl Acetate 0.493 0.494 0.336 0.187 MTBE 1.507 1.827 2.721 1.780 TBA 0.088 0.078 0.098 0.046 Diisopropyl Ether 1.394 1.668 1.711 1.565 Chloroprene 0.671 0.851 0.945 0.878 1,1-Dichloroethane 0.874 0.953 1.996 1.119 Acrylonitrile 0.308 0.263 0.299 0.104 Vinyl acetate 0.855 0.636 1.146 1.025 ETBE 1.462 1.776 1.520 1.596 cis-1,2-dichloroethene 0.617 0.704 1.390 0.855 2,2-Dichloropropane 0.682 0.485 0.928 0.915 Bromochloromethane 0.397 0.392 0.476 0.422 Chloroform 0.915 1.005 1.094 0.974 Carbon Tetrachloride 0.571 0.727 0.710 0.715 1,1,1-Trichloroethane 0.726 0.892 0.880 0.844 THF 0.143 0.169 0.105 0.125 Dibromofluoromethane (Surrogate) 0.464 0.502 0.502 0.510 Methyl Acrylate 0.511 0.576 0.438 0.466 1,1-Dichloropropene 0.614 0.706 0.740 0.807 2-Butanone (MEK) 0.209 0.217 0.208 0.158 Benzene 1.977 2.214 2.316 2.243 Propionitrile 0.591 0.637 0.612 0.567 tert Amyl Methyl Ether (TAME) 1.449 1.787 1.216 1.524 1,2-Dichloroethane 0.729 0.815 0.777 0.728 Isobutyl Alcohol 0.451 0.193 0.458 0.148 Isopropyl Acetate 0.931 1.134 0.813 0.873 Trichloroethene 0.482 0.617 0.588 0.619 1,4-Difluorobenzene (IS) N/A N/A N/A N/A Dibromomethane 0.160 0.197 0.148 0.172 1,2-Dichloropropane 0.253 0.300 0.275 0.270 Bromodichloromethane 0.330 0.466 0.391 0.393 Methyl Methacrylate 0.227 0.317 0.180 0.238 n-propyl Acetate 0.347 0.411 0.270 0.323 2-Cleve 0.162 0.192 0.110 0.149 Table 5: Compound Summary

Water Soil Compound cis-1,3-dichloropropene 0.377 0.460 0.370 0.444 Toluene-d8 (surr) 0.830 1.092 0.790 1.089 Toluene 0.999 1.249 1.034 1.276 2-Nitropropane 0.339 0.425 0.270 0.323 Tetrachloroethene 0.266 0.493 0.235 0.300 4-methyl2-pentanone 0.036 0.050 0.068 0.034 1,1,2-Trichloroethane 0.230 0.301 0.205 0.235 Ethyl Methacrylate 0.192 0.255 0.124 0.177 Dibromochloromethane 0.207 0.329 0.213 0.248 1,3-Dichloropropane 0.420 0.506 0.354 0.410 1,2-Dibromoethane 0.230 0.301 0.187 0.233 n-butyl Acetate 0.361 0.463 0.272 0.339 2-Hexanone 0.174 0.213 0.128 0.149 Chlorobenzene-d5 (IS) N/A N/A N/A N/A Chlorobenzene 0.739 0.908 0.803 0.949 Ethylbenzene 1.207 1.550 1.321 1.573 1,1,1,2-Tetrachloroethane 0.242 0.336 0.270 0.287 M&P Xylene 0.961 1.264 1.042 1.300 Ortho Xylene 1.014 1.314 1.115 1.290 Styrene 0.675 0.920 0.747 0.913 Bromoform 0.131 0.240 0.143 0.164 Isopropylbenzene 1.200 1.502 1.218 1.541 n-amyl Acetate 0.471 0.533 0.399 0.401 BFB (surr) 0.364 0.501 0.390 0.501 n-propylbenzene 1.335 1.720 1.452 1.924 trans-1,4-dichloro-2-butene 0.104 0.116 0.085 0.091 Nitrobenzene 0.016 0.041 0.017 0.020 Bromobenzene 0.536 0.675 0.555 0.661 1,1,2,2-Tetrachloroethane 0.413 0.429 0.381 0.366 1,3,5-Trimethylbenzene 0.917 1.260 0.972 1.262 2-Chlorotoluene 0.869 1.113 0.914 1.127 cis-1,4-dichloro-2-butene 0.143 0.152 0.109 0.125 4-Chlorotoluene 0.886 1.113 0.933 1.223 Tertbutylbenzene 0.756 1.035 0.813 1.050 1,2,4-Trimethylbenzene 0.930 1.259 0.961 1.260 sec-butylbenzene 1.216 1.578 0.388 1.717 p-isopropyltoluene 0.992 1.281 1.081 1.367 1,3-Dichlorobenzene 0.506 0.629 0.546 0.671 1,4-Dichlorobenzene-d4 (IS) N/A N/A N/A N/A 1,4-Dichlorobenzene 0.552 0.678 0.580 0.691 n-butylbenzene 0.936 1.159 1.127 1.360 1,2-Dichlorobenzen 0.507 0.642 0.481 0.600 1,2-Dibromo-3-Chloropropane 0.070 0.097 0.049 0.061 Hexachlorobutadiene 0.125 0.169 0.157 0.199 1,2,4-Trichlorobenzene 0.313 0.439 0.293 0.413 Naphthalene 1.098 1.461 0.685 1.026 1,2,3-Trichlorobenzene 0.318 0.442 0.264 0.364 Table 5: Compound Summary (cont.)

Matrix %RSD N2 %RSD He MDL N2 MDL He Compound Compound Water 10.08 8.30 0.32 0.33 0.602 0.708 Soil 10.30 8.10 0.48 0.53 0.718 0.689 Table 6: Experimental Results Summary Figure 1: Overlay of 50ppb water standard purged in Helium and in Nitrogen Figure 2: Overlay of 50ppb soil standard purged in Helium and in Nitrogen

Conclusions The Atomx and Trap Concentrator Multi-Matrix Autosampler in conjunction with an Agilent GC/MS system performed very well for both the water and the soil calibration range. These findings support the option of moving to Nitrogen as an alternative to Helium. Considering Helium can cost as much as three times the price of Nitrogen, this switch can save companies performing typical USEPA methodologies considerable amounts of costs over the long term. In addition the use of Nitrogen generators capable of producing 99.999 or greater purity offer yet another solution to the cost associated with the analysis by removing the need for cylinders.