WM2015 Conference, March 15 19, 2015, Phoenix, Arizona, USA

Similar documents
DESIGN AND PERFORMANCE OF THE INTEGRATED WASTE ASSAY SYSTEM (IWAS)

A New MCNPX PTRAC Coincidence Capture File Capability: A Tool for Neutron Detector Design

WM 05 Conference, February 27 - March 3, 2005, Tucson, AZ INTEGRATED WASTE ASSAY SYSTEM (IWAS) AND ANALYSIS ENHANCEMENTS

Pulsed Neutron Interrogation Test Assembly - PUNITA

Active Mode Calibration of the Combined Thermal Epithermal Neutron (CTEN) System

High Precision Nondestructive Assay to Complement DA. H.O. Menlove, M.T. Swinhoe, and J.B. Marlow Los Alamos National Laboratory

Simulation of Rate-Related (Dead-Time) Losses In Passive Neutron Multiplicity Counting Systems

Research in NDA Techniques for Waste Characterization at the JRC

CALIBRATION OF SCINTILLATION DETECTORS USING A DT GENERATOR Jarrod D. Edwards, Sara A. Pozzi, and John T. Mihalczo

Effect of Cosmic-ray Shielding in Passive Neutron Coincidence Counting

Estimation of Performance of an Active Well Coincidence Counter Equipped with Boron-Coated Straw Neutron Detectors 13401

Characterization of a Portable Neutron Coincidence Counter Angela Thornton and W. Charlton Texas A&M University College Station, TX

Integrated Waste Assay System (IWAS)

AN ALTERNATIVE PERSPECTIVE ON THE WEIGHTED POINT MODEL FOR PASSIVE NEUTRON MULTIPLICITY COUNTING

SNM detection by means of thermal neutron interrogation and a liquid scintillation detector

A VERY HIGH EFFICIENCY NEUTRON COUNTER FOR THE MEASUREMENT OF PLUTONIUM IN DECOMMISSIONING WASTES. J.A. Mason/A.N. Technology Ltd.

Verification measurements of alpha active waste

Digital simulation of neutron and gamma measurement devices

FISSILE MATERIALS DETECTION VIA NEUTRON DIFFERENTIAL DIE-AWAY TECHNIQUE

Verification measurements of alpha active waste

BWXT Y-12 Y-12. A BWXT/Bechtel Enterprise SMALL, PORTABLE, LIGHTWEIGHT DT NEUTRON GENERATOR FOR USE WITH NMIS

The Absolute Calibration of Active Neutron Assay Instruments

Neutron Multiplicity Counting for Future Verification Missions: Bias When the Sample Configuration Remains Unknown

Performance Review of the High Efficiency Neutron Counter

Photoneutron Interrogation of Uranium Samples by a 4 MeV LINAC. A Feasibility Study

Fysikgå rden 4, SE Gö teborg, Sweden b Swedish Radiation Safety Authority, SE Stockholm, Sweden

Neutron Multiplicity Measurements for the AFCI Program

WM2010 Conference, March 7-11, 2010, Phoenix, AZ

M. S. Krick D. G. Langner J. E. Stewart.

Henzl, Vladimir Menlove, Howard O. Thron, Jonathan L.

The Neutron Diagnostic Experiment for Alcator C-Mod

Evaluation of Nondestructive Assay Characterization Methods for Pipe- Over-Pack Containers

IMAGING BOXED WASTE ASSAY SYSTEM

Rice University Physics 332 LIFETIME OF THE MUON I. INTRODUCTION...2! II. MEASUREMENT PROCEDURES...3! III. ANALYSIS PROCEDURES...7!

Quartz-Crystal Spectrometer for the Analysis of Plutonium K X-Rays

Neutron and Gamma Ray Imaging for Nuclear Materials Identification

He-3 Neutron Detectors


Nuclear Reactions A Z. Radioactivity, Spontaneous Decay: Nuclear Reaction, Induced Process: x + X Y + y + Q Q > 0. Exothermic Endothermic

Fast Neutron Multiplicity Counter: Development of an active-mode counter UM-INL Collaboration

Evaluation of radiation dose rates due to calibration and interrogation neutron sources while using the AWCC in horizontal configuration

Figure 1. Decay Scheme for 60Co

PRECISION OF THE ACCIDENTALS RATE IN NEUTRON COINCIDENCE COUNTING

Detection of Highly Enriched Uranium Using a Pulsed IEC Fusion Device

ACTIVATION ANALYSIS OF DECOMISSIONING OPERATIONS FOR RESEARCH REACTORS

Computational study of Passive Neutron Albedo Reactivity (PNAR) measurement with fission chambers

Chapter 11: Neutrons detectors

A Monte Carlo Simulation for Estimating of the Flux in a Novel Neutron Activation System using 252 Cf Source

in Cross-Section Data

2 Korea Atomic Energy Research Institute, Yuseong gu, Daejeon, South Korea, Abstract

CHARACTERIZING THE DETECTOR RESPONSE AND TESTING THE PERFORMANCE OF A NEW WELL COUNTER FOR NEUTRON COINCIDENCE MEASUREMENTS OF PLUTONIUM IN RESIDUES

WM2014 Conference, March 2 6, 2014, Phoenix, Arizona, USA

DESIGN AND OPERATION OF THE COMBINED TECHNOLOGY AUTOMATED WASTE CHARACTERISATION SYSTEM

Nuclear Cross-Section Measurements at the Manuel Lujan Jr. Neutron Scattering Center

Nuclear Physics and Astrophysics

Reactivity monitoring of a subcritical assembly using beam-trips and current-mode fission chambers: The Yalina-Booster program

ENEN/Reactor Theory/ Laboratory Session 1 DETERMINATION OF BASIC STATIC REACTOR PARAMETERS IN THE GRAPHITE PILE AT THE VENUS FACILITY

The Thick Target (α, n) Production Yield of Sulphur

Detection of explosives and fissile material based on neutron generators, survey of techniques and methods. M. Bruggeman

YALINA-Booster Conversion Project

Numerical Tools for the Evaluation of Super-Compacted Radioactive Waste Residues

Detekce a spektrometrie neutronů. neutron detection and spectroscopy

A Proposal of Nuclear Materials Detection and Inspection Systems in Heavily Shielded Suspicious Objects by Non-destructive Manner.

Neutron Sources Fall, 2017 Kyoung-Jae Chung Department of Nuclear Engineering Seoul National University

Emerging Capabilities for Advanced Nuclear Safeguards Measurement Solutions

Reactivity Monitoring with Imposed Beam Trips and Pulsed Mode Detectors in the Subcritical Experiment YALINA-Booster

DESIGN OF NEUTRON DOSE RATE METER FOR RADIATION PROTECTION IN THE EQUIVALENT DOSE

Determining Plutonium Mass in Spent Fuel with Non-destructive Assay Techniques NGSI Research Overview and Update on NDA Techniques, Part I

Low dimensional neutron moderators for enhanced source brightness

NEUTRONIC ANALYSIS STUDIES OF THE SPALLATION TARGET WINDOW FOR A GAS COOLED ADS CONCEPT.

WM2013 Conference, February 24 28, 2013, Phoenix, Arizona, USA

DEVELOPMENT OF A PORTABLE NEUTRON COINCIDENCE COUNTER FOR FIELD MEASUREMENTS OF NUCLEAR MATERIALS USING THE

Fissile material experiments at the Device Assembly Facility

The Attenuation of Neutrons in Barite Concrete

Observation of High Multiplicity Bursts of Neutrons During Electrolysis of Heavy Water with Palladium Cathode Using the Dead-Time Filtering Technique

PARAMETERISATION OF FISSION NEUTRON SPECTRA (TRIGA REACTOR) FOR NEUTRON ACTIVATION WITHOUT THE USED OF STANDARD

Beam diagnostics: Alignment of the beam to prevent for activation. Accelerator physics: using these sensitive particle detectors.

Modelling moderated proportional neutron counters using the Geant4 toolkit and the application to detection of fast neutron burst

Sh. Zeynalov, O. Zeynalova, V. Smirnov

FAILURE MODE AND EFFECTS ANALYSIS (FMEA) FOR THE ADVANCED MIXED WASTE TREATMENT PROJECT (AMWTP) RETRIEVAL BOX ASSAY SYSTEM (RBAS)

The Use of Self-Induced XRF to Quantify the Pu Content in PWR Spent Nuclear Fuel

Recent Activities on Neutron Calibration Fields at FRS of JAERI

Researchers at the University of Missouri-Columbia have designed a triple crystal

Shielded Scintillator for Neutron Characterization

MC simulation of a PGNAA system for on-line cement analysis

Comments on the possible observation of d-d fusion in sonoluminescence (Reference-31 in Taleyarkhan et al. [2002] 1 )

The investigations in the field of advanced nuclear power systems for energy production and transmutation of RAW in NAS of Belarus

Distillation purification and radon assay of liquid xenon

Neutron Interactions with Matter

SIMULATION OF THE THERMAL NEUTRON SEMICONDUCTOR DETECTOR RESPONSE USING MCNPX CODE

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

Introduction to neutron rem meters

MUSE-4 BENCHMARK CALCULATIONS USING MCNP-4C AND DIFFERENT NUCLEAR DATA LIBRARIES

Nuclear Science and Engineering Institute, University of Missouri, USA and Sidney Kimmel Institute for the Nuclear Renaissance

Nuclear Data Measurements at the RPI LINAC

An introduction to Neutron Resonance Densitometry (Short Summary)

WM2013 Conference, February 24 28, 2013, Phoenix, Arizona USA. Screening and Spectral Summing of LANL Empty Waste Drums 13226

Neutron induced reaction and neutron sources

Precision neutron flux measurement with a neutron beam monitor

Performance Characterization of A New Cam System M.J. Koskelo 1, J.C. Rodgers 2, D.C. Nelson 2, A.R. McFarland 3 and C.A. Ortiz 3

Transcription:

A Study of Differential Die Away Performance with Neutron Interrogation Pulse Width 15407 Ludovic Bourva *, Christophe Bedouet **, Rosemary Lester *, Ken Lambert *, Bruno Vernet ** * Canberra UK Ltd - AREVA Group, Harwell Oxford, Building 528.10 unit 1, OX11 0DF United Kingdom ** SODERN 20 Avenue Descartes, Limeil-Brévannes 94114 France ABSTRACT The Differential Die-Away (DDA) technique is an active non-destructive assay methodology that relies on the production of short pulsed bursts of neutrons to interrogate and quantify the fissile nuclear materials present in waste. Pulsed neutron sources like 14 MeV Deuterium-Tritium (D-T) fusion generators are commonly used for such applications and devices producing in the range of 10 8 to 2 x 10 9 neutrons per second at pulse frequencies of 100 Hz are readily available from commercial vendors. The DDA technique is well established and well suited to detect very low masses of fissile materials with detection limits for short assays (<5 min) down to a few milligrams of 239 Pu eq in 208-l drums. Canberra UK Ltd has extensive experience in the commissioning and calibration of such systems. The modelling work reported in this study shows how the time width of the neutron burst used for each interrogation pulse may impact on the thermal neutron interrogation time profile, the fast neutron detection profiles of Fast Neutron Detector Packages (FNDP) and their interrogation neutron background. Discussion on the optimal time setting of the FNDP s gate structure is provided for several pulse width scenarios covering a 10 to 200 µs range. As a result, we have evaluate the impact of using larger pulse widths and higher frequencies compared to the traditional 10 µs/100 Hz settings of early DDA systems. Although potentially penalizing in terms of detection limits these new pulse settings may significantly extend the lifetime of the generator system. This could therefore result in a significant cost benefit to operators of DDA systems while preserving the overall excellent detection capabilities of the differential die-away measurements. INTRODUCTION Differential Die Away Method The differential die-away (DDA) technique is a non-destructive assay technique based on the pulsed neutron interrogation of fissile materials [1]. The detection and potential quantification of the fissile content in an item is performed by measuring the fast (in practice epi-cadmium) neutrons produced by induced fission events in the fissile nuclear material ( 235 U, 239 Pu, 241 Pu, 233 U ). Segregation of the neutron signals attributed to the fissile material against the interrogation neutrons is performed by exploiting the intrinsic difference in die-away time constants observed between the primary neutron interrogation flux and the secondary fast neutron flux originating from the thermal induced fission events. DDA chambers traditionally rely to detect neutrons on Fast Neutron Detector Packages (FNDP), made of one or many thermal neutron detectors, like 3 He gas proportional counters, embedded in a good neutron moderator, like High Density Polyethylene (HDPE). These detector packages are made insensitive to thermal neutrons by tightly wrapping them with a high thermal neutron absorber, like cadmium. This is illustrated in Figure 1. 1

Figure 1. Basic schematic of a Fast Neutron Detector Package used for DDA measurements As a result, the die-away time constant of the detected pulsed interrogation neutrons is representative on the time taken for these neutrons to fully drop below the cadmium neutron capture threshold (about 0.4 ev). This time constant is closely dependent on the geometrical design and materials used in the construction of a DDA assay system. Early DDA systems [2, 3, 4] introduced graphite as a reflector material surrounded by HDPE. DDA systems have also been built purely with HDPE [5]. HDPE being a more efficient moderator than graphite, these assay systems showed a shorter primary die-away time constant than their graphite counterparts, as shown in Figure 2. Figure 2. Comparison of interrogation flux time profile for a HDPE DDA chamber and a Graphite/HDPE chamber. On the other hand, graphite chambers present the advantage of having longer die-away time constant for thermal neutrons than HDPE based systems. They hence provide opportunities over longer period for the thermal neutron flux built-up in the assay cavity to interact with the fissile material. Although design specific, typical detectable fissile signals do not tend to extent well over 4 milliseconds for pure HDPE chambers whereas values up to 10 milliseconds can be achieved in graphite chambers [6]. Practically, the DDA assay electronics allows the repeated accumulation of the neutron detection events between interrogation pulses. The resulting spectrum is then analyzed using a time gated structure (early/late gate) to derive the net neutron count rate associated to fissile material. This is illustrated in Figure 3, where the time evolution of the intensities of the primary and secondary neutron fluxes in the assay cavity is overlaid. 2

Figure 3. Schematic of the FNDP response for primary interrogation neutrons and secondary fissile neutron [7] To reach the best detection capability the position of the Early Gate (EG) relative to the start of the pulse emission if usually adjusted to include a small portion of the remaining interrogation background, as waiting for the complete decay of the interrogation background is significantly penalizing in terms of performances. Also, for waste assay, the interrogation background is dependent on the moderating/absorbing neutron properties waste materials. The introduced background correction is therefore treated as a matrix dependent factor which may introduce additional systematic errors in the assay. Neutron Generator Systems Production of the interrogation neutrons for DDA assays has been mainly developed around pulsed D-T fusion generator tubes. These devices generate fast neutron through Deuterium-Tritium fusion reactions as shown in Equation 1. (Eq 1) Commercial product like the Zetatron N-250 from Activation Technology Corporation or the D-211 system from Thermo Fisher, shown in Figure 4, have been used in many DDA system since the mid 1980 s. These produce 4-π neutron flux up to 2 x 10 8 n.s -1, and although adjustable, use trigger frequencies and pulse width of 100 Hz and 10 µs respectively. Figure 4. Photograph of the ATN N-250 system components (LHS) and the Thermo Fisher D-211 system (RHS) Both these units generate the very high voltage current used to accelerate the Deuterium ions at the level of the neutron emitting module. They only require high voltage (up to 600-V) to operate. SODERN is also commercializing neutron generator for nuclear measurement applications. The latest Genie 16 NG 3

series of generators, shown in Figure 5, matches the ATC and Fisher Thermo neutron emission capability and provide a modern computer based interface to control the system. Figure 5. Photographs of the SODERN Genie 16 NG system showing the NEM (LHS), Control Chassis and VHV power supply (Centre) and the System s Software Interface (RHS) Pulsed emission characteristics of the Genie 16 NG are directly configurable by the user to operate at the selected frequency and interrogation pulse width. In comparison with the other two systems, the SODERN system include a single VHV power supply unit to directly deliver up to 100 kv, 85µA current to the Neutron Emitting Module (NEM). The life time of neutron generator tubes for pulsed applications usually cover only a few 100 hours of neutron emissions. Such equipment represents a significant investment to operators of NDA systems and contributes to the overall operating cost of the systems. The possibility to investigate the dependence of DDA performances with emission characteristics of the neutron generator system that may improve his longevity is therefore of interest to improve their cost effectiveness. DDA Assay Chamber Design As mentioned earlier in this introduction, the design of a DDA assay chamber play an important role in the expected detection time profile of the system s FNDP s. To study the impact of the neutron emission pulse width on the performances of a DDA assay, we have based this study on the specifics of a HDPE DDA chamber. Note that the chamber is not purely design to meet DDA requirements but also include passive neutron detection capabilities and requirements as it common for NDA system to provide both passive and active assay capabilities. Consequently the design represents a compromise. Figure 6 shows a horizontal 2-D cross section plot of the assay system as well as a 3-D illustration of the chamber main body geometry. The system is designed to assay 208-l waste drum raised from the floor with a hoist powered drum grab assembly. Figure 6. Graphical illustration of the NDA chamber. 4

The design incorporate 6 FNDP fitted with a single 2-inch, 48-inch active length 3 He tube pressurized at 2 atmosphere. Neutron generation is based on the Genie 16 NG system. Data acquisition is performed using CANBERRA Multiport-II multichannel analyzers operated in MCS mode synchronized on the generator pulse trigger. The system is also fitted with a flux and barrel 3 He monitors and a 252 Cf Add-a-Source matrix interrogation system to provide both passive and active matrix correction capabilities. NUMERICAL MODELLING This section describes a series of MCNPX simulations aimed at first investigating the DDA s FNDP response with variable neutron interrogation pulse width. It has provided an initial evaluation of the chamber response to this generator operational characteristic when modelling inert waste and 235 U bearing waste. This work has been further supported by tallying for several neutron energy groups the time evolution of the interrogation neutron flux intensity within the waste volume for these two cases. These calculations provided good insight in the inner workings of the DDA chamber and are reported below. Using the FNDP data a preliminary evaluation of impact of the pulse width on the performance and its impact on the positioning of the DDA early gate is presented. Fast Neutron Detector Package Response The response in the FNDP 3 He neutron detector was calculated in the MCNPX simulations by using a neutron flux tally (F4:N) and tally multiplier cards set filter only (n,p) neutron detection reactions in the 3 He gas volume. A time bin structure set from the time T 0 =0 to 5000 µs was also used to generate time profiles. The time T 0 corresponds to the start of the neutron emission from the neutron generator target point. The source definition of the models was set to simulate the isotropic production of 14.1 MeV neutrons with a time distribution probability uniformly distributed between 0 and the desired length of the neutron pulse. Five settings of 10, 20, 50, 100 and 200 µs were consistently used in the models. Figure 7 shows the obtained FNDP detection time profiles up to 1600 µs obtained for the cases of an inert material (simply air in the drum) and the theoretical case of a quasi-infinitely dilute mass of 235 U uniformly distributed in the same volume. 5

Figure 7. FNDP neutron detection time profiles the case of inert (empty) and fissile ( 235 U) materials with D-T interrogation pulse widths between 10 and 200 µs. All FNDP profiles show, independently of the pulse width setting, the expected behavior of neutron detections of a DDA system. The inert cases show a relatively rapidly decaying detection profile over the first 1000 µs. This is consistent with a large HDPE base system. In the presence of fissile material the detection profile is a two component profile following closely up to about 300 to400 µs the empty chamber profile and then decaying to longer time with a much longer time constant. Neutron Flux in Waste Drum The FNDP calculations have been complement with two sets of results which aimed at further understanding for the assumed chamber design the slowing down process of the interrogation neutrons and the production of secondary fast fission neutron in the nuclear material, when present. The timing of these phenomena and their potential dependence to the interrogation neutron pulse width was of particular interest. In Figure 8 the time evolution of four neutron energy groups, G1 with E< 0.5 ev (Thermal Neutrons), G2 with 0.5 ev < E < 100 ev (Epi-Cadmium), G3 with 100 ev < E < 1 MeV (Epithermal) and G4 with 1MeV < E < 14 MeV (fast), is shown. The figure present the results obtained when modelling the 5 aforementioned pulse widths and no fissile material in the simulated waste. G4 and G3 show very short die away time in all cases while G2 progressively increases during the pulse width. It then dies-away with a time constant comparable to the one observed for the FNDP neutron detection profiles. This is consistent with the assumption that this population of neutron is sufficiently energetic to cross the cadmium liner of the FNDPs and hence contribute to the measured neutron signals. G4, independently of the pulse width builds-up from 50 µs onwards to dominate the total neutron flux as time extent above 300 µs. These thermal neutrons, on the other hand do not appear in the FNDP counts as they are effectively 6

suppressed by the Cadmium absorber. The 100 and 200 µs gate width seems to show some relatively small difference in the timing of the cross over point between the G1 and G2 curves; that is roughly 225 and 300 µs rather than less than 200 µs for the other three cases. Also note that the time values given in this example are directly related to the modelled chamber design and would need to be evaluated for any specific chamber design. Figure 8. FNDP neutron detection time profiles the case of inert (empty) and fissile ( 235 U) materials with D-T interrogation pulse widths between 10 and 200 µs. A similar study was repeated for models were 235 U was modelled in the drum volume, as shown in Figure 9. This study showed that when fissile material is present in the waste the previously depressed G4 and G3 groups are now, after an initial fast die-away of the interrogation neutrons, tallying the G1 time 7

profile. This is consistent with secondary fast neutron originating for thermal induced fission event in the waste. Figure 9. Time dependence in the waste volume of the D-T interrogation neutron flux for four neutron energy groups during the assay 235 U with interrogation pulse widths between 10 and 200 µs As for the no fissile material case, the 100 and 200 µs pulse width models shows some relative degree of difference with the other 3 profiles. Interrogation Background The extension of the interrogation pulse width is anticipated to impact on the die-away profile of the interrogation neutrons and hence, for a given DDA counting gate structure, result into an increase of the 8

interrogation background noise in the DDA early gate. To evaluate this impact, the MCNPX FNDP profiles obtained for the fissile and non-fissile cases were analyzed. For a fixed early gate end time of 1600 µs, which significantly exceed the interrogation background, and a variable DDA early gate start time, the relative weight of the interrogation background signal to the fissile signal was computed. The result of such an analysis is presented in Figure 10 for the 5 interrogation pulse width cases. Figure 10. Relative importance of the D-T interrogation background relative to the fissile signal with neutron interrogation pulse widths between 10 and 200 µs Figure 10 shows that, as expected, increasing the neutron interrogation pulse width tends, for a given early gate start time, to increase the relative weight of the interrogation background, and as a result the detection limit of the DDA assay. Nevertheless, an increase from 10 to 50 µs interrogation pulse width tends to bring only a small variation. The increase in the importance of the interrogation background for the interrogation longer pulse widths can be compensated by shifting the early gate to a later time. However this would also translate into reducing the fraction of fission signal collected in the DDA early gate, impacting performances. To illustrate this and define the optimum position of the early gate, Figure 11 present a plot of the interrogation background corrected fissile FNDP signal as a function of the early gate settings. This suggests that a small increase to the interrogation pulse width has very small impact on the DDA performance. However note that the plot shown only considers a statistical argument in expressing a gate structure for which the best signal to noise ratio is established. It doesn t consider the aforementioned dependence of the interrogation background on the waste matrix neutron moderation/absorption characteristics. As a consequence, this simple graph implies that the most efficient gate settings may tolerate a level of interrogation noise which is quite high (>10%), but ignore that the variability and accuracy by which it can be corrected in real measurement cases may affect the overall accuracy of the measurements. Despite this comment, numerical simulation tend to indicate that the use of interrogation pulse width between 10 to 50 µs have little impact on the DDA performances. 9

Figure 11. Impact of the DDA early gate start time on the net fissile signal for different interrogation pulse widths DISCUSSION The numerical simulations results reported in this publication show that increasing the generator pulse width for 10 up to 50 µs does not significantly modifying the characteristics of the thermal neutron flux interrogating the nuclear material in the waste. As a result, DDA performances with 50 µs bursts should remain almost unchanged despite using an operating regime that is less demanding for the neutron generator components. Extension of the interrogation pulse width to 100 µs is shown to roughly double the interrogation background in the early DDA gate. This translates into shifting the early DDA gate start position by about 50 µs to maintain the interrogation background to similar level to that obtained with a 10 µs pulse width. Consequently, the achievable Lower Limit of Detection of the DDA system will be affected, although probably still providing adequate performances for many measurement applications. Based on the numerical results, extension of the pulse width in this HDPE chamber to 200 µs is not advised as time profiles were shown to be affected by the longer spread of the neutron production in the assay system. The neutron emission rate of the SODERN neutron generator increases with the voltage and intensity of the current applied to accelerate the ion beam. The current intensity is a function of the established D-T pressure in the neutron generator D-T tube. Consequently, for a given pulse frequency and a given neutron output the increase of the pulse width, which defines the workload the neutron tube, corresponds to an operating point with lower D-T pressure or lower ion beam acceleration voltage. These two parameters, D-T pressure and ion beam acceleration voltage have a direct incidence on the longevity of the neutron tube. For example, these are directly correlated to the probability of phenomena like field effect electronic emissions that trigger high voltage breakdown. A reduction of these two parameters therefore contributes to reducing the chances and the intensity of such degrading events and other events that may lead to the premature failure of the neutron tube. However, note that operating the system with 100 µs pulse width with a 100 Hz frequency or 50 µs pulses and a 200 Hz frequency are equivalent in terms of D-T pressure as they correspond to equivalent workloads. CONCLUSION Simulation of the FNDP response with neutron generator pulse width from 10 up to 50 µs indicated that the DDA assay of nuclear waste could be performed with no significant loss of performance. This increase of the neutron pulse width should, although difficult to truly quantify, result in the extension of 10

the neutron generator lifetime. This work will be supplemented with a set of empirical measurements with various waste matrices to further study the impact of neutron generation characteristics on the performance of DDA measurements and confirm the advantage of operating DDA system with longer pulse width while preserving the nuclear material detection/quantification of this assay technic. ACKNOWLEDGMENTS The author would like to thanks Dr. Stephen Croft from Oak Ridge Laboratory to have kindly accepted to review the content of this publication. REFERENCES [1] J.T Caldwell, R.D. Hastings, G.C. Herrera, W.E. Kunz, E.R. Shunk; The Los Alamos second generation system for passive and active neutron assay of drum size containers ; Los Alamos National Laboratory, LA-10774-MS, UC-15, September 1986. [2] C. Passard, A. Mariani, F. Jallu, J. Romeyer-Dherber, H. Recroix, M. Rodriguez, J. Loridon, C. Denis, PROMETHEE: an alpha low level waste assay system using passive and active neutron measurement methods. ; Nucl. Technol. 140, Dec 2002, 303 314 [3] J.T Caldwell et al; Apparatus and Method for Quantitative Assay of Generic Transuranic Wastes from Nuclear Reactors ; US Patent Number 4,483,816, 31 st March 1982 [4] J.T Caldwell et al; High Sensitivity Imaging Passive and Active Neutron (IPAN) Transuranic Assay System ; Proceedings of the 14 th Annual ESARDA Symposium on Safeguards and Nuclear Material Management, Salamanca, Spain, 5 th -8 th May 1992. [5] R.D. McElroy, Jr., S. Croft, B. Young, and L. C-A Bourva; Design and Performance of the Integrated Waste Assay System (IWAS) ; Proceedings of the Institute of Nuclear Materials Management, 44 th Annual Meeting, Phoenix (2003) [6] B.H.Armitage, P.M.J. Chard, S. Croft, B.C. Syme and C.G. Wilkins; Limitations to Accuracy in Measurements of Wastes in Safeguards ; Proceedings of the 14 th Annual ESARDA Symposium on Safeguards and Nuclear Material Management, Salamanca, Spain, 5 th -8 th May 1992. [7] NDA 2000 Technical Manual Canberra Industries Document 9231595F V5.1 (2011) 11