White Paper. Application of GC-MS/MS in the Analysis of Taints and Off Flavours. Alan Rose Technical Specialist. Donna Byrom Technical Specialist

Similar documents
Odours from Food Packagingan Analytical Challenge?

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401

Determination of volatiles in Food sample preparation strategies and techniques Dr Kathy Ridgway

Automated and accurate component detection using reference mass spectra

Multi-residue analysis of pesticides by GC-HRMS

Cerno Application Note Extending the Limits of Mass Spectrometry

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Characterization of Odor & Taste Components Using Olfactory Detection and Solid Phase Microextraction GC-MS

Benefits of Equipping your Lab with a Time-of-Flight Mass Spectrometer

Quantitative Analysis of Opioids Using a Triple-Quadrupole GC/MS/MS

U.S. EPA Method 8270 for multicomponent analyte determination

GAS CHROMATOGRAPHY MASS SPECTROMETRY. Pre-Lab Questions

Odor Detection and Analysis using GC/SAW znose

A New Way to Extract Off-Flavor Compounds in the Aquatic Environment. Stir Bar Sorptive Extraction (SBSE)

WADA Technical Document TD2003IDCR

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Determination of Volatile Substances Proof of Food Adulteration

SIFT-MS. technology overview

SIFT-MS SELECTED ION FLOW TUBE MASS SPECTROMETRY TECHNOLOGY OVERVIEW \ SYFT TECHNOLOGIES

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Benchtop NMR Combined with GC/MS Confirms Identity of Forensic Case Sample

Meeting Challenging Laboratory Requirements for USEPA Method 8270 Using a Highly Sensitive, Robust, and Easy-to-Use GC/MS

Advances in the Identification of Tastes and Odours in. Drinking Water. Gavin Mills CChem FRSC Principal Scientist

Application. Gas Chromatography February Introduction

Highly Sensitive and Rugged GC/MS/MS Tool

Hans-Joachim Hubschmann. Handbook ofgc-ms. Fundamentals and Applications. Third Edition. Wl LEY-VCH. VerlagGmbH & Co. KG aa

Analysis of Pesticides in Baby Foods Using a Triple-Quadrupole GC/MS/MS

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Mass Spectral Studies of Polypropylene Chromatographic Well Plates

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Analysis of TO-15/TO-17 air toxics in urban air using TD GC/TOF MS and automated compound identification software

Impurities Sources, Detection and Measurement HPLC-MS Analyses. Dr Courtney Milner Chemical Analysis

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Multi-residue GC-MS analysis. Richard Fussell CSL York, UK

ACD/Labs Software Impurity Resolution Management. Presented by Peter Russell

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS

Thermo Scientific Pesticide Explorer Collection. Start-to-finish. workflows for pesticide analysis

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS

Impurity Profiling of Pharmaceutical Starting Materials Using Gas Chromatography Coupled with High- Resolution Accurate Mass Spectrometry

QUALITY CONTROL CRITERIA FOR CHEMISTRY EXCEPT RADIOCHEMISTRY.

AppNote 1/2004. Use of a Mass Spectral Based Chemical Sensor to Discriminate Food and Beverage Samples: Olive Oils and Wine as Examples KEYWORDS

AN INTEGRATED SYSTEM USING TEMPERATURE BASED SAMPLING FOR POLYMER CHARACTERIZATION

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer

Author. Abstract. Introduction

GC Analysis of Polybrominated Flame Retardants Application

high performance OPTIC 3 and applications a total analytical solution

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

SPME-GC-MS/MS for Identification and Quantification of Migration Contaminants in Paperboard Food Packaging

Application of multi-dimensional GC techniques to the analysis of cigarette smoke

GUIDELINES ON THE USE OF MASS SPECTROMETRY (MS) FOR IDENTIFICATION, CONFIRMATION AND QUANTITATIVE DETERMINATION OF RESIDUES CAC/GL

The Use of Tandem Quadrupole GC/MS/MS for the Determination of Polycyclic Aromatics Hydrocarbons (PAHs) in Food Products

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs

Characterisation of allergens in cosmetics by GC GC TOF MS with Select-eV variable-energy electron ionisation

Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

Innovative Thermal Desorption From Agilent Technologies. Vivek Dhyani Application Chemist

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

PesticideScreener. Innovation with Integrity. Comprehensive Pesticide Screening and Quantitation UHR-TOF MS

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Introduction. Chapter 1. Learning Objectives

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

Implementing GC-HRAM MS for More Efficient and Effective Routine Pesticide Residues Analysis

Enhanced Simultaneous Total ion and Selective Multi-Ion GC/MS Using a Combination of New Software and Low-Bleed Columns

Chromatographic Methods of Analysis Section: 5 Gas Chromatography (GC) Prof. Tarek A. Fayed

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems

Chemical Analysis Problem

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

Workflow Using Hyphenated Mass Spectrometry Techniques

HYPHENATED TECHNOLOGY GUIDE

EGASUS. 4D GCxGC TOFMS. Delivering the Right Results

Overview. Introduction. André Schreiber 1 and Yun Yun Zou 1 1 AB SCIEX, Concord, Ontario, Canada

Schedule. Draft Section of Lab Report Monday 6pm (Jan 27) Summary of Paper 2 Monday 2pm (Feb 3)

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.

Analysis of alkylphenols using GC-MS/MS and automated SRM development

Counterfeit and illegal pesticides: Strategies for addressing the issue in the analytical laboratory

MassHunter Software Overview

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

Environmental Mass Spectrometry

Determination of releasable 2,4,6-trichloroanisole in wine by cork stoppers (Resolution OIV-Oeno 296/2009)

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Technical Note. Introduction

HYPHENATED TECHNOLOGY GUIDE

The Theory of HPLC. Quantitative and Qualitative HPLC

all suspects Identify Thermo Scientific Extractables & Leachables Workflow Unknown impurity identification and quantification workflows for:

Determination of Organochlorine Pesticides and Polychlorinated Biphenyls Using GC/MS/MS Operated in the MRM Mode

environmental Rtx -CLPesticides and Rtx -CLPesticides2 Columns: The Ideal Confirmational Pair for Analyzing Polychlorinated Biphenyls (PCBs)

Cerno Application Note Extending the Limits of Mass Spectrometry

Agricultural. Chemistry. Agricultural production: crops and livestock Agrichemicals development: herbicides, pesticides, fungicides, fertilizers, etc.

Enhanced odour profiling of strawberries using TD GC TOF MS

Introduction to E&Ls 1

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences

is given for the isotopic fingerprinting methodology.

Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ

Transcription:

White Paper Application of GC-MS/MS in the Analysis of Taints and Off Flavours Alan Rose Technical Specialist Donna Byrom Technical Specialist

Application of GC-MS/MS in the Analysis of Taints and Off Flavours Introduction Taint and off-flavour compounds can be present in food and drinks causing an atypical smell or taste. Extreme examples can influence the consumer s perception of a product and can lead to product recalls. These compounds are often present at very low levels that traditional gas chromatography-mass spectrometry (GC-MS) based methods can struggle to detect and identify. Fortunately developments in mass spectrometry technology have meant that instrument sensitivity can be significantly increased. This technology has been assessed to determine whether it can play a role in supporting the food and drinks industry in the investigation of taints and off flavours. We assess GC-MS/MS as a tool to identify a number of the common tainting compounds found in the food and drinks and compare this to traditional GC-MS, to determine whether an increase in sensitivity can be seen, and in turn address an age old problem MS/MS is used to fragment specific ions in the compound of interest, generating a product ion which should be free of interferences form the matrix and thereby increasing sensitivity. Taints and Off Flavours Taint compounds are contaminants which are not intended constituents of the product and can originate from a variety of sources, which include chemicals that are transferred from packaging, storage bags, wooden pallets, shipping containers, storage facilities and corks of bottles. Moreover the use of cleaning products near food production can also be a source. Typical taints cause musky, earthy or even medicinal smells. Compounds from a number of chemical classes, most commonly halophenols, and haloanisoles are responsible. In addition, pyrazines, geosmin and 2-methyl-isoborneol are well documented taint compounds in food. On the other hand off flavour compounds are present in products as a result of chemical alteration of the product, which include oxidative, biological and enzymatic degradation of ingredients such as fats and oils. Typical off flavours cause a rancid smell of the product. Halophenols Halophenols such as chlorophenols, bromophenols and iodophenols, are compounds that are used as chemical intermediates or as finished products in industry. These compounds have low sensory thresholds and may be present as taints in foods and drinks at concentrations in the ppb (parts per billion; µg per kg) or even ppt (parts per trillion; ng per kg) range. At this level traditional GC-Ms techniques struggle to pick up these compounds even though they can be obvious during sensory evaluation. Products that have been contaminated with halophenols have an aroma that is described as disinfectant or medicinal. Chlorophenols consist of many different compounds with different degrees of chlorination. They are used as fungicides, pesticides, herbicides and disinfectants. W wellknown chlorophenol is pentachlorophenol, which has been used extensively for wood preservation. Chlorophenols in drinking water can also result from chlorination of natural phenolic compounds that are present in the treated water, whereas iodophenols can be formed during the use of iodine as a drinking water disinfectant. Bromophenols and their derivatives are widely used as fire-retardants, but also for the preservation of wood and wood products, leather, textiles, paint, plastic, paper and pulp. An additional important source of halophenols and especially chlorophenols in drinking water is diffusion of organic additives from high density polyethylene water pipes. These additives consist of impurities or by-products of typical phenolic additives that are used as antioxidants in the production of the pipes. Haloanisoles Chloro-, bromo- and iodoanisoles are formed by methylation of the hydroxide group of halophenols by moulds and fungi under humid and warm conditions. This methylation is a biological mechanism to detoxify the halophenols. Haloanisoles are easily transferred when moulds grow in the vicinity of a product and contaminated products exhibit musty and mouldy odours. These compounds can be detected on aroma when present at extremely low concentrations (as low as 0.02 ppt or ng/kg) and are responsible for the cork taint in wine. Page 2

Taint and Off Flavour Investigation Typical Taint Analysis Workflow analytical instrument is capable of. Therefore, in addition to concentrating the samples, the most sensitive analytical techniques are required to analyse for these compounds. A very powerful and sensitive technique is tandem mass spectrometry. Tandem Mass Spectrometry (MS/MS) The first objective in a taint and off flavour investigation is to identify the compound or compounds that are the cause of the atypical smell or taste. When possible and safe to do so, an informal sensory analysis is performed on the affected product, which will help to decide what analytical strategy will be pursued. The headspace of the samples will be analysed if the odour is very strong or a compound is suspected to be highly volatile (e.g. example, by solidphase microextraction gas chromatographymass spectrometry (SPME-GC_MS)). If this does not provide the required information or sensory suspect a less volatile compound, the same is extracted (e.g. using Liken-Nickerson extraction) and the extract will be concentrated (e.g. using a Kuderna Danish apparatus) to increase the relative amount of the tainting compound in the extract. If a taint or off-odour compound has been identified, the next stage in the investigation is aimed to find the root course of the problem. Therefore a targeted analytical approach is required to analyse affected samples, raw materials, storage materials, etc. Sample amounts are often limited. Therefore, in addition to concentrating extracts of samples, the most sensitive analytical techniques are required to analyse for these compounds. Confirmatory data may also be needed to prove the origin of the contamination, for example, for legal or insurance purposes. Most taint and off flavour compounds are potent odour compounds that can be detected by humans in much lower concentrations than any In order to be able to analyse complex mixtures of volatile compounds, gas chromatography is coupled to mass spectrometry (GC-MS). The chromatography separates the components then a mass spectrometer operated in scanning mode produces mass spectral data for each component of the chromatographic analysis. This means that detected compounds can be identified from their mass spectra. These spectra can be compared to recorded spectra from commercially available libraries. Analysis time has to be dedicated to scan and record a signal for each mass. Whilst this provides a large amount of data, the sensitivity is lower and this can also be time consuming to generate and analyse. Compounds can be monitored and quantified by selecting a single mass and ignoring all the other masses, constructing an extracted-ion chromatogram. Sensitivity can be greatly improved by only scanning for selected masses, since analysis time is spend on recording signals for a reduced number of masses. Although this technique of selected-ion monitoring (SIM) is very sensitive, there is still a potential that compounds with similar masses cause interference with the masses of the compounds of interest. Moreover this relies on the expertise of the analyst to select the correct ions to search for. In each of these instances there is the chance that the tainting compound will be missed. Selectively and therefore sensitivity can be improved by using single-reaction monitoring (SRM) using a triple-quadrupole mass spectrometer. Rather than monitoring a specific ion (as in SIM), selected precursor ions are fragmented and a product ions are monitored. This is highly selective, because for many compounds, it is unlikely that a compound elutes very closely to the compound of interest that can be characterised by the same precursor/product ion (SRM transition). In many gas chromatography-tandem mass spectrometry (GC-MS/MS) applications matrix interference is virtually eliminated. In many instances, this will remove or reduce the need for sample clean-up, which can simplify the analytical procedure and increase efficiency. Page 3

In concentrated extracts, the detection limits of GC-MS/MS are lower than the odour thresholds form the majority of taint and off flavour compounds without the need to extract and concentrate excessive quantities of sample. This is especially important, since the amount of tainted sample is often limited, especially in the case of customer complaints. Furthermore, GC-MS/MS is a well-established and widely accepted analytical technique. This means that GC-MS/MS data can be used for legal cases or to aid insurance claims. Gas Chromatography-Tandem Mass Spectrometry (GC-MS/MS) Analysis Samples are analysed for a large suite of taint and off-flavour compounds using a Thermo Scientific GC that is coupled to a TSQ 8000 triple quadrupole MS. The gas chromatograph is equipped with a TG-WASMX column and samples are injected into a split-splitless injector operated in splitless mode. An oven temperature programme is used to separate the compounds that are injected onto the column. The mass spectrometer is operated is operate in positive electron impact MS/MS mode at 70eV. SRM transitions of the taint and off-flavour compounds were determined with the aid of the automatic algorithms of the Thermo Xcalibur 2.2 software sing commercial standards. Example of the Analysis of a Taint Compound To illustrate the analysis of a taint compound by GC-MS/MS, a food sample was analysed by both GC-MS and GC-MS/MS looking for 2, 4- dichlorophenol (Figure 1). At a concentration of 0.05 ppb (ng/ml) in the sample, the compound could not be detected from the total-ion current chromatogram (TIC). However, when the molecular ion (m/z 162) was selected in selective ion monitoring (SIM), a signal could be observed, but was difficult to distinguish from the baseline noise. When the same sample was analysed by GC-MS/MS, a significant increase in the signal to noise ratio of the SRM chromatogram (m/z 162 to 98) was observed. The taste threshold in water for 2,4-dichlorophenol is reported to be 0.3 ppb so this clearly demonstrates the benefits of the increase in selectivity (and therefore sensitivity) at a level significantly below the taste threshold. Figure 1 Top Partial chromatograms (from top to bottom: TIC, SIM and SRM) Bottom Spectrum of 2, 4-dichlorophenol Conclusions The use of GC-MS/MS for the analysis of taint and off flavour compounds in food and potable water samples provides a very sensitive and effective analysis technique. Experience demonstrates that food and drinks manufacturers face challenges in retrieving affected samples from the consumer for analysis. Therefore for the majority of occasions, the amount of tainted product that can be analysed is limited to relatively small sample amounts (<100 g). Sensitivity is therefore essential. GC-MS/MS has demonstrated that as a tool for taint and off flavour analysis it provides very sensitive, reliable quantitative and confirmatory data, which is needed to identify the root cause of the contamination of a product. References Dietz, F. & Traud, J. (1973 Vom Water, 41, 137 Page 4

Alan Rose Technical Specialist Alan is a Technical Specialist in the Investigative Analysis Department of Reading Scientific Services Ltd. A Chartered Chemist and Member of the Royal Society of Chemistry, he has over 30 years laboratory experience working on taints and off flavours in food. Donna Byrom Technical Specialist Donna has recently joined RSSL with 30 years of experience in a variety of analytical techniques on varied matrices. Her experience includes method development, validation and special analysis on tobacco and related products; pesticide analysis on teas and various crops, and on pesticides for regulatory purposes. She was also a Forensic Scientist for 16 years in the field of drugs and toxicology. For further information, please contact Customer Services on enquiries@rssl.com Page 5