Analysis of Terpenes in Cannabis Using the Agilent 7697A/7890B/5977B Headspace GC-MSD System

Similar documents
Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Application Note. Abstract. Authors. Introduction

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Applying the Agilent 5977A MSD to the Analysis of USP<467> Residual Solvents with the 7697A Headspace Sampler and 7890B GC

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

USP<467> residual solvents

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Pesticides Analysis Using the Agilent 5977A Series GC/MSD

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Rapid Analysis of Food and Fragrances Using High-Efficiency Capillary GC Columns. Application. Authors. Abstract. Introduction

Author. Abstract. Introduction

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Residual solvents analysis using an Agilent Intuvo 9000 GC system

Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Analysis of Natural Oils and Extracts Using the Low Thermal Mass LTM Series II System

Highly Sensitive and Rugged GC/MS/MS Tool

U.S. EPA Method 8270 for multicomponent analyte determination

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Application. Gas Chromatography February Introduction

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

HIGH SPEED GC-TOFMS FOR ESSENTIAL OILS PROFILING IN ROUTINE QUALITY ASSESSMENT

AppNote 7/2003. Coupling Retention Time Locked Methods and Libraries to Automated SPME or SBSE for Analysis of Flavors and Fragrances

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

Enhanced Sensitivity for Biomarker Characterization in Petroleum Using Triple Quadrupole GC/MS and Backflushing

AppNote 8/2008. A New Purge Tool for Use with Automated Headspace Analysis KEYWORDS ABSTRACT

AppNote 7/2003. Coupling Retention Time Locked Methods and Libraries to Automated SPME or SBSE for Analysis of Flavors and Fragrances KEYWORDS

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Application Note. Abstract. Authors. Introduction. Food Safety

Jack Cochran, Lindsey Shear-Laude, Alex Hodgson VUV Analytics Emerald Conference. #emeraldconference

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Rapid, Robust, and Sensitive Detection of 11-nor-D 9 - Tetrahydrocannabinol-9-Carboxylic Acid in Hair

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Screening of Pesticide Residues in Water by Sequential Stir Bar Sorptive Extraction-Thermal Desorption with GC/MS

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Hydrogen as a Carrier Gas in the Analysis of Polar/Non-Polar Compounds Using the Polyarc System. Application Note. Author. Introduction.

Sulfotepp impurities in Chlorpyrifos EC formulations

A Generic Method for the Analysis of Residual Solvents in Pharmaceuticals Using Static Headspace-GC-FID/MS

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

The 5975 inert MSD Benefits of Enhancements in Chemical Ionization Operation Technical Note

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Determination of Chlorinated Acid Herbicides in Soil by LC/MS/MS Application Note

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

Introduction to Capillary GC

Discovering the Scent of Celery: HS-SPME, GC-TOFMS, and Retention Indices for the Characterization of Volatiles

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

Agilent G2350A Atomic Emission Detector (AED)

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Benzene Contamination in BabyFood and Beverages by New Generation of Static Headspace Autosampler coupled to Fast GC-TOFMS

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Determination of Cannabinoids (THC) in Biological Samples

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

Introduction to Capillary GC

Advantages of Reversed Sandwich Injection for Pesticide Residue Analysis

Fast, Quantitative Analysis of Residual Solvents in Cannabis Concentrates

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Detection of Chemical Warfare Agents by Transportable GC/MS

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Determination of 17 Organotin Compounds in Beverages Using Triple Quadrupole GC-MS/MS System

GAS CHROMATOGRAPHY (GC)

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID)

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

VGA-100 Instrument Detection Limits and Linearity

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

Study of Residual Solvents in Various Matrices by Static Headspace

Transcription:

Analysis of Terpenes in Cannabis Using the Agilent 7697A/789B/977B Headspace GC-MSD System Faster Analysis Time = Greater Productivity Application Note Cannabis, Food Authors Ronald Honnold, Robert Kubas, and Anthony Macherone Agilent Technologies, Inc. 8 Centerville Rd Wilmington, DE 988 Abstract Terpenes and terpenoids are naturally occurring volatile isoprenes constructed from a (C H 8 ) n building block where n is the number of isoprene units linked to form linear or ring molecules. In marijuana (Cannabis sativa), terpenes are biosynthesized in the flower along with other compounds such as psychoactive 9 -tetrahydrocannabinol (THC) and other nonpsychoactive cannabinoids []. Each strain of marijuana is characterized by its unique terpene profile, that creates the distinctive aroma and flavor []. This application note describes the analysis of terpenes common to marijuana in under 6 minutes using the Agilent 7697A/789B/977B headspace-gas chromatography-mass selective detection (HSS-GC-MSD) system with concomitant flame ionization detector (FID) detection using integrated Capillary Flow Technology to split the column effluent in a controllable and precise manner to the two detectors.

Introduction The concentration of individual terpenes in the C. sativa plant varies by strain, harvest time, and drying and storage conditions. Depending on these conditions, the relative terpene concentrations decrease over time for example, after three months of storage, levels can decrease by more than half [,]. Analysis of terpenes in C. sativa strains is typically performed using HSS-GC/MSD, using the Agilent Residual Solvent Analyzer hardware configuration to provide an optimum linear range of detection and mass spectral speciation. Experimental Hardware and consumables Agilent 7697A Headspace Sampler, Agilent 789B GC with FID, and Agilent 977B MSD (Residual Solvent Analyzer) Split/Splitless Inlet Purged Two-Way Splitter: split : FID:MSD Agilent Headspace liner, Ultra Inert, splitless, straight mm (p/n 9-7) Agilent ml Headspace vials (p/n 9-8) Agilent Headspace vial caps (p/n 8-6) Agilent VF- column, m. mm,. µm (p/n CP8877) MSD Restrictor - Deactivated Fused Silica.7 m µm µm FID Restrictor - Deactivated Fused Silica.7 m µm µm Certified terpene reference standards 9-DGEQ and 96 DGEQ were obtained from Restek (Bellefonte, PA). Agilent 7697 Headspace parameters Parameter Instrument settings Vial pressurization gas Loop size Keyboard lock Transfer line type Transfer line diameter System configuration Carrier control Value Helium ml OFF DB-ProSteel. mm Oven temperature C Loop temperature C Transfer line temperature C Vial equilibration Injection duration GC cycle time Vial size GC Instrument. minutes. minutes minutes ml Vial shaking Level Fill mode Fill pressure Loop fill mode Loop ramp rate Loop final pressure Loop equilibration time Carrier control mode Extraction mode Vent after extraction Post injection purge Acceptable leak check Agilent 789B GC conditions Default psi Custom psi/min psi. minutes GC controls carrier Single extraction ON GC oven temperature 6 C Hold time Oven program Equilibration time ml/min for minutes Default,. ml/min. minutes C/min to C, hold minutes then C/min to C, hold. minutes minute Max temperature 6 C Column flow Front S/SL inlet He mode. ml/min Split Heater C Pressure Total flow Septum purge flow Gas saver 9. psi ml/min Off Split ratio : MSD Transfer line 6 C ml/min at minutes

Agilent 977B MSD conditions Acquisition mode Solvent delay Tune file Scan ( m/z m/z)).8 minutes etune.u EM Setting mode Gain = Scan speed Scan parameters Threshold Normal MSD Source temperature C MSD Quadrupole temperature C FID Conditions Detector temperature C H Flow Air flow Makeup flow Flame and electrometer ml/min ml/min ml/min On Aux EPC supplies Column with. ml/min constant flow to Restrictor going to the MSD, FID/MSD split ratio approximately :. Aux EPC FID F.7 psi Standards, sample preparation, and quantitation Standard calibrator preparation For quantitation, a seven-point calibration curve from ppm to, ppm was created from the terpene reference standards by adding µl of each pre-prepared calibration level to a ml headspace vial, and capping the vial. The calibrator levels were:,,,,,, and, ppm. Sample preparation For cannabis plant material, it is recommended that samples be frozen prior to grinding, or that grinding occur under liquid nitrogen. This keeps the samples cold during the grinding process, reducing loss of the more volatile terpenes. The headspace method uses full evaporation technique (FET) standard preparation. Since cannabis product matrices are extremely varied, and plant material will not dissolve in solvent, sampling involves the use of a very small sample amount ( mg) of plant or wax material weighed into the headspace vial, which is then capped and analyzed. For this study, a challenge sample of lemon grass tea was prepared to demonstrate the power of mass spectrometry to differentiate targeted terpenes from interfering chemical compounds often found in real samples. From headspace sampler S/S Inlet m. mm, µm Purged splitter. ml/min F Constant flow mode Column =. ml/min MSD flow =. ml/min FID/MSD split ratio = : MSD psi Quantitation The terpene concentrations in the unknown samples were determined in Agilent MassHunter quantitative software through linear regression analysis of the linear calibration curve constructed from the known calibrator levels. Figure. Capillary flow diagram of the Agilent 789B configured with a two-way splitter for simultaneous collection of FID and MSD data.

Results and Discussion The analytical method developed in this study is ultra fast, robust, and reliable. It leverages the large dynamic range of FID detection and mass spectral confirmation to identify and quantitate targeted terpenes found in cannabis and cannabis wax samples. Figure illustrates the chromatography for both the FID and the MSD at,, and, ppm. All terpenes elute in less than 6 minutes, and the total cycle-time (injection to injection) is minutes. Figure shows typical calibration curves for select terpenes, where the average linear regression coefficient (R ) was.997 for the FID data, and.998 for the MSD. Figure shows the extracted ion chromatogram (EIC) for alpha pinene with the FID and MSD calibration curves over the, ppm range. Table outlines the average retention time (RT), average concentration, the standard deviation, and the percent relative standard deviation (%RSD) for the target concentration, the limit of quantitation (LOQ), the limit of detection (LOD), the average response factor, and the %RSD the average response factor of eight replicate injections of a ppm standard..... 6.. 6 7... 6.. ppm scan ppm FID ppm scan ppm FID, ppm scan, ppm FID.....6.7.8.9.......6.7.8.9.......6.7.8.9.... Acquisition time (min) Figure. FID and MSD chromatograms at ppm, ppm, and, ppm of the -compound terpene mix.

Camphene FID 6 7 8 9,,,, Concentration (ng/ml) beta-myrcene FID 6 7 8 9,,,, Concentration (ng/ml) 6...8.6.. Camphene MS 6 7 8 9,,,, Concentration (ppm) 9 beta-myrcene MS 8 7 6 6 7 8 9,,,, Concentration (ppm) beta-pinene FID 6 8,, Concentration (ng/ml).8.6....8.6.. beta-pinene MS 6 8,, Concentration (ppm) Figure. Typical MSD and FID terpene calibration curves over the ppm to, ppm calibration levels. Counts...9.8.7.6.......9.8.7.6..... A. minutes B MSD Calibration y = 7.798*x + 9.9 R =.999 C FID Calibration y = 7.7*x 7.8997 R =.99767..... Acquisition time (min) Figure. Ten-ppm extracted ion chromatogram (EIC) for alpha-pinene. Calibration levels range from ppm to, ppm.

Table. Analysis of Eight Replicates of ppm Standard Name RT Avg. conc. Std. dev. Conc. %RSD LOQ LOD Avg. resp. Resp. %RSD alpha-pinene FID. 9..6.9.6.9 77 alpha-pinene..7.8.8.8. 87.7 Camphene FID.7 8.88...7.6 79. Camphene....6..9.6 beta-myrcene FID.7 9...9..9.9 beta-myrcene.8 9..8.7 8.6 8. 67.9 beta-pinene FID.6 9.7..9.8.7 6699 beta-pinene.76..8..8 6.8 797. delta--carene FID.9 9.7.8.87.6 878 delta--carene.69.76.9.8 9.6.77 77.8 alpha-terpinene FID.98 9.6.6.7.68. 88.8 alpha-terpinene.6 9..6. 6.7.8 6. D-Limonene FID.6 9.9..9..6 7.9 D-Limonene.68.9.6. 6..8 9. beta-ocimene FID.687 9.....7 9. beta-ocimene.699 9..87.8 8.7.6 8 p-cymene FID.7 8.99....66 868. p-cymene.7..69. 6.98.9 79. Eucalyptol FID.7.98.6. 6..8. Eucalyptol.7 9....6 66 gamma-terpinene FID.787.7. 7...7 869 6.7 gamma-terpinene.797.8... 78 Terpinolene FID.9 9..7.79. 67 Terpinolene.9.8.77. 7.7. 668. Linalool FID.96 9...9..9 6.9 Linalool.97 8..7..7. 98.7 Isopulegol FID.8 8.99..9..7 7.9 Isopulegol..8... 6.69 79 6. Geraniol FID.9 9.68..7.9..7 Geraniol.6 7.9....6 699.6 beta-caryophyllene FID.6.6..6..9 6.6 beta-caryophyllene.7 6.9.... 9. alpha-humulene FID...9.6.9.88 6.6 alpha-humulene. 7.8.... 686. Nerolidol FID.8..7 7. 8.7 9 Nerolidol.9.9.9.8.99.79 9. Nerolidol FID.9.6.9.9.9.7 878.9 Nerolidol.67 9.69..9.7.67 66. Guaiol FID.8.6.79. 7.9.8 7. Guaiol.86..66.7 6.67. 9. Caryophyllene oxide FID.96 7.9.99.7 9.9.97 7.7 Caryophyllene oxide.99.7.8..8.8 9868.6 alpha-bisabolol FID.9.7.7. 7.9. 9. alpha-bisabolol. 9.8.8. 8..9.7 6

7 A 6 Counts Counts 7 B 6.....6.7.8.9.......6.7.8.9.......6.7.8.9... Acquisition time (min) Figure. Total ion chromatogram (TIC) comparison of a ppm terpene reference standard (B) versus a mg challenge sample of lemon grass tea (A) known to contain similar terpenes and other compound that may interfere with proper identification when only using FID detection. Counts Counts. A.......... B....... -......6.7.8.9.......6.7.8.9.......6.7.8.9... Acquisition time (min) Figure 6. FID chromatogram comparing a ppm terpene reference standard (B) with a mg sample of lemon grass tea (A). When using FID detection only, the peak at. minutes may be misidentified as alpha-pinene. 7

Figure 7. The nearly coeluting peak in the challenge sample (lemon grass tea) shown in the top chromatogram and spectrum is clearly different from the alpha-pinene spectrum of the terpene reference standard. This illustrates the advantage of MSD selectivity to rule out interfering species that may otherwise be misidentified as a target analyte when using FID detection alone. The use of both FID and MSD allows for more comprehensive data analysis, as terpenes are often in high percent values, which would saturate the MS, whereas the FID will not saturate. Optimized conditions were obtained using the 7697A method development tools and a parameter increment function built into the software. These tools sequentially optimize the headspace oven temperature, vial equilibration time, and vial shaking level by a set amount over a series of analytical runs. By using the Agilent 7697A tools, the final method can rapidly be optimized versus using manual optimization of other headspace systems. For this analysis, a constant incubation temperature and extraction time was used to ensure volatilization of all terpenes and terpenoids in the sample for reproducible, quantitative results. The Prep Ahead feature of the 7697A headspace system allows an optimum workflow for the short sample analysis time. 8

Conclusion Using the Agilent Residual Solvent Analyzer with an Agilent VF GC column and appropriate restrictors enables full chromatographic separation of targeted terpenes that naturally occur in C. sativa plant material and wax samples and give the plant its distinctive aroma and character. The analysis completed in less than 6 minutes, and uses both FID detection for quantification and extended linear range, and mass selective detection (MSD) for terpene speciation. This ultra-fast methodology almost quadruples laboratory productivity compared to traditional terpene analysis, which takes approximately minutes per sample. References. M. W. Giese, et al. Development and Validation of a Reliable and Robust Method for the Analysis of Cannabinoids and Terpenes in Cannabis Napro Research, California ().. J. M. Parland, E. B. Russo. Cannabis and Cannabis Extracts: Greater than the Sum of Their Parts? The Haworth Press, Pennsylvania ().. E. Russo. Taming THC: Potential Cannabis Synergy and Phytocannabinoid-Terpenoid Entourage Effects British Journal of Pharmacology 6, ().. Chemistry and Analysis of Phytocannabinoids and Other Cannabis Constituents. Humana Press, New Jersey. For More Information These data represent typical results. For more information on our products and services, visit our Web site at www.agilent.com/chem. 9

www.agilent.com/chem Agilent products and solutions are intended to be used for cannabis quality control and safety testing in laboratories where such use is permitted under state/country law. Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 7 Printed in the USA September 6, 7 99-899EN