Thermocouple Dynamic Errors Correction for Instantaneous Temperature Measurements in Induction Heating. Krzysztof Konopka 1

Similar documents
SOME PRACTICAL ASPECTS OF EXCITATION COIL DESIGN FOR ELECTROMAGNETIC FLOW METER

Application example. Measuring Force Sensors Rigid. Six series Nano, Mini, Gamma, Delta, Theta, Omega. Range of measurement, force ± 36 N..

Introduction to Blackbody Sources

ELECTRONIC SENSORS PREAMBLE. This note gives a brief introduction to sensors. The focus is. on sensor mechanisms. It describes in general terms how

Chapter 3. Experimentation and Data Acquisition

To: Mitch Cottrell, Instructor From: Brian Garber, Crawford Binns, Jeff Hollis (Group 3, Sec. J) Subject: Project Deliverable #3 Date: 4 May 2010

Using a Mercury itc with thermocouples

Veerapong Kanchanawongkul*

Modelling the Temperature Changes of a Hot Plate and Water in a Proportional, Integral, Derivative Control System

Thermocouple Calibrations and Heat Transfer Coefficients

A Nonlinear Dynamic S/H-ADC Device Model Based on a Modified Volterra Series: Identification Procedure and Commercial CAD Tool Implementation

INSTRUMENTATION OF ULTRASONIC HIGH-FREQUENCY MACHINE

PTB S 16.5 MN HYDRAULIC AMPLIFICATION MACHINE AFTER MODERNIZATION

ELG4112. Electromechanical Systems and Mechatronics

Magnetic elds, the Hall eect and electromagnetic induction (Electricity and Magnetism)

Determination of freezing points of pure substances with Cobra4 TEC

1 THE CONCEPT OF TEMPERATURE

INSTRUMENTAL ENGINEERING

Calculation of the heat power consumption in the heat exchanger using artificial neural network

Temperature Sensors & Measurement

Development of device CAP-10 to perform inprocess testing of the single-core wire capacitance per unit length

Theory and Design for Mechanical Measurements

Determine the Inside Wall Temperature of DSTs using an Infrared Temperature Sensor

Transient measurements using thermographic phosphors

Sapphire Fiber-Optic Temperature Sensor Based on Black-Body Radiation Law

Carbonized Electrospun Nanofiber Sheets for Thermophones

Modeling of Power System Components During Electromagnetic Transients

PROPERTY STUDY ON EMATS WITH VISUALIZATION OF ULTRASONIC PROPAGATION

DATA SHEET. Thermocouple module with digital I²C-Interface - THMOD-I²C. Characteristic features. Areas of application. Features.

Temperature Measurement and First-Order Dynamic Response *

MEASUREMENT SYSTEM FOR FRACTIONAL GLOW TECHNIQUE

by M. Felczak, T.Wajman and B. Więcek Technical University of Łódź, Wólczańska 211/215, Łódź, Poland

ESTIMATION OF THE PARAMETERS OF ELECTROMAGNETIC FIELD AT INDUCTION DEVICE BY THE AID OF COMPUTER SIMULATION

Department of Mechanical and Aerospace Engineering. MAE334 - Introduction to Instrumentation and Computers. Final Examination.

Heat transfer coefficient distribution over the inconel plate cooled from high temperature by the array of water jets

VMS-GeoMil. Background

Automatic Control Systems. -Lecture Note 15-

What happens when things change. Transient current and voltage relationships in a simple resistive circuit.

Sensibility Analysis of Inductance Involving an E-core Magnetic Circuit for Non Homogeneous Material

THERMODYNAMICS METHODS OF HEAT TRANSFER RADIATION

Experimentation. Third Edition. Wheeler. Ahmad R. Ganji. School of Engineering. With Third Edition contributions by. V. V. Krishnan. Brian S.

Temporal analysis for implicit compensation of local variations of emission coefficient applied for laser induced crack checking

Electrical and Magnetic Properties of High Temperature Superconductors Using Varying forms of Data Acquisition

8.1 Circuit Parameters

Reflow Soldering Processes Development Using Infrared Thermography

Automatic Adjustment of Integration Time in the NIR Camera for Calibrated Temperature Measurement during Inductive Heating Processes

AN INHERENTLY LINEAR TRANSDUCER USING THERMAL SIGMA-DELTA MODULATOR

Experiment 5: Thermocouples (tbc 1/14/2007, revised 3/16/2007, 3/22,2007, 2/23/2009, 3/13/2011)

INSTITUTE OF AERONAUTICAL ENGINEERING (Autonomous) Dundigal, Hyderabad

What does such a voltage signal look like? Signals are commonly observed and graphed as functions of time:

DESIGN AND ANALYSIS OF AXIAL-FLUX CORELESS PERMANENT MAGNET DISK GENERATOR

Dynamic Modeling of Surface Mounted Permanent Synchronous Motor for Servo motor application

INSTITUTE OF AERONAUTICAL ENGINEERING Dundigal, Hyderabad DEPARTMENT OF ECE QUESTION BANK. : G.Lakshminarayana, Asst.

IEEE TRANSACTIONS ON POWER DELIVERY, VOL. 18, NO. 4, OCTOBER

This is an electronic reprint of the original article. This reprint may differ from the original in pagination and typographic detail.

(A) (B) (D) (C) 1.5. Amplitude (volts) 1.5. Amplitude (volts) Time (seconds) Time (seconds)

Industrial Heating System Creating Given Temperature Distribution

Measurement of a Clock-Tuned Digital Non-Foster Circuit for Positive or Negative Digital Capacitance

Magnetism of materials

Lab 1: Simple Pendulum 1. The Pendulum. Laboratory 1, Physics 15c Due Friday, February 16, in front of Sci Cen 301


COURSE OF Prepared By: MUHAMMAD MOEEN SULTAN Department of Mechanical Engineering UET Lahore, KSK Campus

HT TRANSIENT THERMAL MEASUREMENTS USING THERMOGRAPHIC PHOSPHORS FOR TEMPERATURE RATE ESTIMATES

PRACTICE NO. PD-AP-1309 PREFERRED PAGE 1 OF 5 RELIABILITY PRACTICES ANALYSIS OF RADIATED EMI FROM ESD EVENTS CAUSED BY SPACE CHARGING

MEASUREMENT OF THE MECHANICAL STATE OF SUBSURFACE LAYERS BY EDDY-CURRENT METHOD

New analysis for determination of hemispherical total emissivity by feedback-controlled pulse-heating technique

IMPROVEMENTS IN ACTIVE NOISE CONTROL OF HELICOPTER NOISE IN A MOCK CABIN ABSTRACT

ME 515 Mechatronics. Overview of Computer based Control System

The LM741C Integrated Circuit As A Differential Amplifier Building The Electronic Thermometer

Optimization of Thermal Radiation Source for High Temperature Infrared Thermometer Calibration

Continuous Mass Measurement in Checkweighers and Conveyor Belt Scales

Integral-based Algorithm for Parameter Identification of the Heat Exchanger

Unconventional Double-Current Circuit for Deflection and Temperature Simultaneous Measurement

Szczecin University of Technology Poland ANALOG DOT RECORDING AS AN EXEMPLARY PROBLEM IN THE AREA OF DYNAMIC MEASUREMENTS

MODEL FOR UNCERTAINTY ESTIMATION IN COMPARISON CALIBRATION OF THERMOCOUPLES

Lecture 6: Time-Dependent Behaviour of Digital Circuits

QUESTION BANK DEPARTMENT OF ELECTRICAL AND ELECTRONICS ENGINEERING UNIT I - INTRODUCTION SYLLABUS

Experimental Competition May 7, :30-13:30 hours Marking Scheme

Computer-Based Automated Test Measurement System for Determining Magnetization Characteristics of Switched Reluctance Motors

ROUGHNESS ASPECTS IN THERMAL ANALYSIS OF MACHINE TOOLS

INSTITUTE OF AERONAUTICAL ENGINEERING (Autonomous) Dundigal, Hyderabad

Old Dominion University Physics 112N/227N/232N Lab Manual, 13 th Edition

Exercise 4 Modeling transient currents and voltages

Patrick F. Dunn 107 Hessert Laboratory Department of Aerospace and Mechanical Engineering University of Notre Dame Notre Dame, IN 46556

School of Engineering Faculty of Built Environment, Engineering, Technology & Design

Thermocouples calibration and analysis of the influence of the length of the sensor coating

Induction Heating Spiral Inductor Comparison between Practical Construction and Numerical Modeling

1. Mark the correct statement(s)

Feasibility of Die Attach testing

ENGI 3424 First Order ODEs Page 1-01

A/D Converters Nonlinearity Measurement and Correction by Frequency Analysis and Dither

Temperature Project. Lab: Section K. Masahiro Inano Christopher Casey Adam Tigue

MATLAB SIMULINK Based DQ Modeling and Dynamic Characteristics of Three Phase Self Excited Induction Generator

Light, Electricity and Liquids. By Vitaliy Zamsha and Vladimir Shevtsov

DESIGN, FABRICATION AND ELECTROMECHANICAL CHARACTERISTICS OF A MEMS BASED MICROMIRROR

Rozwiązanie zagadnienia odwrotnego wyznaczania sił obciąŝających konstrukcje w czasie eksploatacji

Managing Emergency Generators

Stray Capacitance Compensation for Non Stray immune ECT systems

Temperature Waves in SRF Research

APPLICATION OF GRAPHENE IN DEVELOPMENT OF HIGH-SENSITIVITY HALL-EFFECT SENSORS. Andrzeja Boboli 8, Warsaw, Poland

Transcription:

Thermocouple Dynamic Errors Correction for Instantaneous Temperature Measurements in Induction Heating Krzysztof Konopka 1 1 Institute of Measurement Science, Electronics and Control, Silesian University of Technology, Gliwice, Akademicka 10, Poland, krzysztof.konopka@polsl.pl Abstract- The most commonly used thermometer to measure high temperatures in metallurgy is the thermocouple, but because of its long response time it can be used mainly for measuring time-averaged temperatures. Research on steel wire patenting using inductive heating required instantaneous temperature values. Pyrometers or thermographic cameras could be used, however they are generally expensive, and temperature measurements are not always reliable due to differing emissivities. In described case expensive infrared thermometers could be replaced by cheap thermocouple provided that dynamic errors were corrected. Correction algorithm would not increase the total cost of the measuring system as no additional hardware is required. Correction algorithm was implemented in LabVIEW. The time constant was determined experimentally. Experiments were carried out to examine improvement in response time of thermocouple with correction algorithm. Keywords: thermocouple dynamic errors, correction algorithm, steel wire patenting, inductive heating. I. Introduction Problem presented is a part of research on steel wire patenting using inductive heating. Patenting is a process of isothermal hardening which is performed in order to obtain specific mechanical properties of wire such as tensile strength, torsion and the ability to be drawn. At a glance, the process involves austenitizing at high temperature between 870 o C and 920 o C and quenching at 450-550 o C [1]. Wires are usually heated in continuous furnaces heated with gas or mazout, and then cooled in lead bath. There is however little information on the properties of wire patented using electrical inductive heating. Induction heating is energy efficient, productive, precise, clean and easily controllable. Constructions of induction heating devices are safe for operators and environment. Therefore the use of induction heating in the process of wire patenting would be beneficial both in terms of wire properties as it allows precise controlling of heating parameters and because of the ecological purity of the process. The test stand was built to study the influence of inductor current frequency and type of inductor on the heating efficiency, the speed of temperature rising and the temperature field distribution along the heated wire. The production efficiency requires high heating rate. There are heated wires of small diameter (2.2-3.5 mm), which requires the research on the shaping of the magnetic field, which would allow heating of load with such a small diameter within required short time. The production effective heating time is about 3 s. With such a short heating time, an important factor in the research are the dynamic properties of the temperature sensor. Most often in the metallurgy for measurements of temperatures of specified order there are used thermocouples for contact measurements and pyrometers for non-contact ones. Pyrometer gives fast response, but as non-contact measurement it is related with the possibility of large errors. The main error source is improperly defined emissivity of measured object. In practice emissivity is determined experimentally by measuring the temperature accurately by thermocouple. However, even then the element surface may be slightly inhomogeneous, and hence the emissivity can vary greatly. Another source of the pyrometer measurement errors in the case described, is a small diameter of the wire, which is connected with difficulties in obtaining the same measurement conditions - slight displacement of the measuring spot is associated with a large change in the angle between pyrometer and measured surface. Thermocouple measurement under the conditions described is associated with too long time constant, which can be larger than the total heating time. In the first approach it completely disqualifies this method of measurement in the presented case. On the other hand, thermocouple measurement is very cheap and accurate in static conditions. It was therefore decided to check whether good enough results can be achieved by implementing correction algorithm into measuring chain. Software dynamic correction algorithm was used and its efficacy examined. ISBN-10: 84-616-5438-2 ISBN-13: 978-84-616-5438-3 508

II. Dynamic error correction algorithm For the experiments in the test stand unshielded K-type thermocouple was used. Such temperature sensor is a first-order system and is governed by the first-order ordinary differential equation dty t Tyt Tx t, (1) dt where T y (t) is the response of the system (the output, temperature indicated by thermocouple) to forcing function T x (t) (the input, temperature of patented wire), τ is the time constant for the system. The time constant τ depends in fact not only on the properties of thermocouple itself (for example dimensions and material constants of sheathed thermocouple), but also on properties of conditioning circuit, especially commonly used low-pass filter. T x (t) Analog conditioning T y (t) DAQ T y t k Correction algorithm T cort k Figure 1. Thermocouple measuring chain Figure 1 presents the considered thermocouple measuring chain. It contains three kinds of processing: analog described by (1), analog-to-digital performed in DAQ board that consists of sampling and quantization and finally digital correction algorithm. It is assumed that measured quantity T x (t) varies in time which causes dynamic errors arising in analog part of the chain. These errors can be decreased using a correction algorithm. The rests of not corrected dynamic errors influence the accuracy of results at the output of the measuring chain, therefore they should be taken into account while final uncertainty is estimated [2][7-11]. Results T ( t ) at the output are estimates of instantaneous values of the input quantity T x (t) at the moments tk kt d cor k, where Td is the sampling period, k is the number of the sampling moment, k 0,1,. The results are T ~ of the analog part output value (t). These obtained on the basis of the quantized samples sequence samples are processed by a dynamic correction algorithm which is built on the basis of the dynamic properties model of the analog part. For the first-order system dynamic correction algorithm is [2][11] 1 ~ ~ Tcor( tk ) T y tk 111T y t k, (2) where 1 y t k Td e Td 11, 1e III. Description of the test stand 1. (3) In order to check the properties of measuring chain with described correction algorithm the test stand was prepared as presented in figure 2. Temperature was measured by unshielded K-type thermocouple, connected to IPAQ-L converter [12] which converts thermocouple voltage to 4-20mA current. The current was measured by data acquisition card NI USB 6009 and then measured signal was processed by program implemented in LabVIEW [13]. A step change in temperature to examine the dynamic response of a thermocouple was obtained by moving it from a room temperature to melted tin. Second channel of DAQ board was used to determine exact moment of thermocouple dipping. Algorithm described by (2) and (3) was implemented in LabVIEW. T y ISBN-10: 84-616-5438-2 ISBN-13: 978-84-616-5438-3 509

K type thermocouple 19 th IMEKO TC 4 Symposium and 17 th IWADC Workshop IPAQ-L 4-20mA AI1 AI2 DAQ NI USB 6009 Computer with LabVIEW Melted tin Figure 2. The test stand to check the properties of measuring chain IV. Identification of the dynamic properties of the object Dynamic parameters of the measuring chain were not analyzed on the basis of analytical equations. It was about the maximum simplification of the procedure for practical applications. The whole part of the measuring chain up to the correction algorithm was interpreted as a black box and its dynamic parameters were determined experimentally. Step response was easiest to implement in practice. The experiment was carried out in such way that thermocouple was immersed in the molten tin. The ambient temperature changed rapidly from room temperature to the molten tin temperature. The moment of immersion, which is the moment of temperature step change, was determined using the second measurement channel of DAQ board. Exemplary temperature curve is shown in figure 3. Figure 3. Thermocouple measuring chain step response Basing on the waveforms measured it was assumed that the measuring chain can be modeled by two transducers: converter generating delay T 0, and the dynamic converter. For unshielded thermocouple it is assumed the first order dynamic model. Other measuring chain elements have also dynamic properties, but it was assumed, that first order model is sufficient for the whole chain. Basing on experimental results time delay is T 0 =0,88s. The time constant of first order dynamic converter was determined using approximate method with assumption, that the converter is the ideal one. Time constant was determined using dependence 0,5, (4) ln 2 where τ is the time constant, 0, 5 is the moment in time when thermocouple indicates the half of real temperature value. Experiments allowed to measure 0, 5 = 0,12s, therefore τ = 0,17s. ISBN-10: 84-616-5438-2 ISBN-13: 978-84-616-5438-3 510

V. Random errors Dynamic error correction algorithms amplify random errors. This amplification is dependent on sampling period T d. The smaller T d, the bigger random errors amplification. On the other hand T d cannot be too long as it would decrease measurement resolution in time domain. As it was described in introduction the aim of measurements is the heating time to achieve the desired temperature. Tables 1, 2 and 3 show the influence of measuring chain parameters on random errors amplification. Noise in directly obtained results were too large and correction algorithm amplified it to an unacceptable level (row 1 and 5 in table 1). So noise filtering is necessary. But most filters makes dynamics worse. In first step capabilities of DAQ device were used. Its maximum sampling rate for two channels is 24 ks/s. It is far enough for performed temperature measurements, so the temperature was measured with frequency f DAQ which was N times assumed frequency for correction algorithm f p 1 fdaq fp. (5) T N p From subsequent portions of N samples arithmetic mean was calculated. In this way, results were obtained with the period of T p and partially eliminated random errors (filter 2 in figure 5). Such filter does not change dynamic properties of measuring chain, but it requires faster sampling. Results in table 1 and 2 show that efficacy of such filtering depends of course on N, but also on T p. With growing f p more and more random errors pass through the filter as they have lower frequencies then f p. For higher f p typical smoothing filter is necessary. There was used moving average with rectangular window (filter 1 in figure 4). Such filter changes of course dynamic properties of measuring chain, therefore new identification of dynamic properties of measuring chain was performed, as described in chapter IV. Time constant of measuring chain with digital smoothing filter was determined as τ = 0,36s. K type thermocouple IPAQ-L 4-20mA AI1 AI2 DAQ NI USB 6009 {Ty(t0),Ty(t1),,Ty(tN)} dt=td/n SMOOTHING FILTER 1 (optional) moving-average rectangular window {Ty(t0),Ty(t1),,Ty(tN)} dt=td/n SMOOTHING FILTER 2 mean value of N points CORRECTION ALGORITHM Tcor(t) dt=td Ty(t) dt=td Figure 4. Thermocouple measuring chain with digital smoothing filters Table 1 Noise amplification by correction algorithm. Influence of N on noise standard deviation. Only filter 2 used, filter 1 inactive. N=1 means that filter 1 is also inactive. Ty -noise standard deviation on input of correction algorithm Tcor -noise standard deviation on output of correction algorithm No f DAQ =N/T p N T p Ty Tcor Hz s K K 1 10 1 0,1 0,97 2,36 2 100 10 0,1 0,32 0,76 3 1000 100 0,1 0,10 0,24 4 10000 1000 0,1 0,02 0,04 5 100 1 0,01 0,97 23,98 6 1000 10 0,01 0,36 8,43 7 10000 100 0,01 0,07 2,18 ISBN-10: 84-616-5438-2 ISBN-13: 978-84-616-5438-3 511

Table 2 Noise amplification by correction algorithm. Influence of N and T p on noise standard deviation. Only filter 2 used, filter 1 inactive. Ty -noise standard deviation on input of correction algorithm Tcor -noise standard deviation on output of correction algorithm No f DAQ =N/T p N T p Ty Tcor Hz s K K 1 10000 100 0,01 0,17 2,51 2 10000 200 0,02 0,11 0,43 3 10000 300 0,03 0,08 0,36 4 10000 500 0,05 0,14 0,20 5 10000 750 0,075 0,16 0,18 6 10000 1000 0,1 0,02 0,04 Table 3 Noise amplification by correction algorithm. Influence of N on noise standard deviation. Both filter 1 and filter 2 used. Ty -noise standard deviation on input of correction algorithm Tcor -noise standard deviation on output of correction algorithm No f DAQ =N/T p N T p Ty Tcor Hz s K K 1 100 1 0,01 0,08 0,10 2 1000 10 0,01 0,12 0,21 3 10000 100 0,01 0,11 0,25 4 24000 240 0,01 0,01 0,36 5 10 1 0,1 0,04 0,04 6 100 10 0,1 0,07 0,07 7 1000 100 0,1 0,16 0,18 8 10000 1000 0,1 0,07 0,09 9 24000 2400 0,1 0,04 0,05 VI. Performance of measuring chain with correction algorithm. Dynamic parameters were identified and noise reduced, so performance of the whole measuring chain was tested. Figures 5 and 6 present the system response to unitary jump. T cor is a temperature calculated by dynamic error correction algorithm. T cor-del is temperature T cor shifted in time to correct delay T 0. Performance of system was tested for different combinations of T p and filtration exemplary graph is in figure 7. 700 600 500 Tout 400 Tin Tcor 300 Tcor-del 200 100 0-0,50 0,00 0,50 1,00 1,50 2,00 2,50-100 Figure 5. Graph of signals before and after correction; T d =0,01s, no noise filter ISBN-10: 84-616-5438-2 ISBN-13: 978-84-616-5438-3 512

600 500 400 300 Tout Tin Tcor Tcor-del 200 100 0-0,50 0,00 0,50 1,00 1,50 2,00 2,50 Figure 6. Graph of signals before and after correction; T d =0,06s, no noise filter 350 300 Ty Tx 250 Tcor Tcor-del 200 150 100 50 0 0 20 40 60 80 100 120 Figure 7. Graph of signals before and after correction; T d =0,16s, smoothing filter I and II Finally system was used to determine the time of induction heating of thin rods to the desired temperature, as mentioned in the introduction. Exemplary heating curve is shown in Figure 8. In this particular case, a small diameter rod resulted in a small magnetic coupling and heating was relatively slow. Figure 8. Heating curve of thin rods with induction heater, rod diameter 2,2mm, power P=7,3 kw, f=61,0 khz; T d =0,1s, no noise filter. VII. Conclusions The most commonly used thermometer to measure high temperatures in metallurgy is the thermocouple because of its low cost and wide experience in its application. But because of its long response time it can be used mainly for measuring time-averaged temperatures. When instantaneous temperature is needed because transient phenomenon is to be observed, other thermometers can be applied such as pyrometers or thermographic camera. However, they are generally expensive, temperature measurements are hindered by differing emissivities and reflections from other surfaces so thermography measurements are not as accurate as contact methods. In many ISBN-10: 84-616-5438-2 ISBN-13: 978-84-616-5438-3 513

cases expensive infrared thermometers could be replaced by cheap thermocouple if its dynamic properties were corrected. Correction algorithm would not increase much the total cost of the measuring system as no additional hardware is required. Results presented in the paper show that because of the delay T 0 measuring chain with correction can be used for research, but harder for controlling. However, despite imperfections, it makes possible such measurements that are impossible to achieve another way. Presented research of induction heating of steel wires in the patenting process can be used as an example. As may be seen in Figure 9 it is not possible to measure the temperature inside the inductor with the pyrometer. Dynamic error correction algorithms amplify random errors, there may therefore be necessary the filtering and the optimum between noise and dynamic properties must be found. This is especially important for measurements during induction heating, which itself adds noise to thermocouple measurements [3][4]. Influence of the algorithm to errors and measurement uncertainty can be also tested through simulations [5][6], making it much easier to carry out experiments. Figure 9. Exemplary inductor used in research of induction heating of steel wires in the patenting process. It is possible to measure the temperature inside the inductor with thermocouple, not with pyrometer References [1]. V. Ya. Zubov, Patenting of steel wire, Metal Science and Heat Treatment, vol. 14, Issue 9, pp 793-800, 1972. [2]. J. Jakubiec, Application of Reductive Interval Arithmetic to Uncertainty Evaluation of Measurement Data Processing Algorithms, Wydawnictwo Politechniki Slaskiej, Gliwice, 2002 [3]. A. Smalcerz, R. Przylucki, Impact of electromagnetic Field upon Temperature Measurement of Induction Heated Charges. International Journal of Thermophysics, pp. 1-13, 2013 [4]. A. Smalcerz, R. Przylucki, K. Konopka, A. Fornalczyk, M. Ślezok.: Multi-variant calculations of induction heating process, Archives of Materials Science and Engineering 58 (2), pp. 177-18, 2012 [5]. K. Konopka, T. Topór-Kamiński, Uncertainty evaluation of measurement data processing algorithm based on its matrix form. Acta Phys. Pol. A, 2011 vol. 120 no. 4, pp. 666-670 [6]. K. Konopka, T. Topór-Kamiński, Identification of measurement data processing algorithm coefficients presented on selected form of FFT algorithm. Fundamental and applied metrology. IMEKO XIX World Congress, Lisbon, Portugal, 2009, pp. 2400-2404, [7]. M. Bojarska, J. Jakubiec, A Method of Modelling Sampling Converter Dynamic Errors. Metrology and Measurement Systems. Vol. VIII - 4/2001, Warsaw 2001, pp. 337-356. [8]. J. Jakubiec, K, Konopka, Identification method of error sources of A/D measuring chain. Proceedings 20th IEEE Instrumentation and Measurement Technology Conference IMTC103, Vail, CO, USA, pp. 1659-1664. [9]. J. Jakubiec, P. Makowski, J. Roj, Error Model Application in Neural Reconstruction of Nonlinear Sensor Input Signal. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 58, NO. 3, MARCH 2009, pp. 649-656 [10]. J. Jakubiec, P. Makowski, J. Roj, Neural Reconstruction of Nonlinear Sensor Input Signal. Instrumentation and Measurement Technology Conference - IMTC 2007 [11]. J. Jakubiec, K, Konopka, Reducing interval arithmetic in dynamic error evaluation. XVI IMEKO World Congress. IMEKO 2000, Wien, 2000, pp. 100-105, [12]. http://www.inor.com/ [13]. http://ni.com/ ISBN-10: 84-616-5438-2 ISBN-13: 978-84-616-5438-3 514