Impact of Instrument Dispersion on Performance of HPLC Capillary Columns

Similar documents
Improving HPLC Column Selection and System Performance

Luna 2.5 µm C18(2)-HST. Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Easy Method Transfer from HPLC to RSLC with the Dionex Method Speed-Up Calculator

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis

High Resolution Fast LC

Method Transfer between HPLC and UHPLC Instruments Equipment-related challenges and solutions

LC Technical Information

Determination of Verapamil Hydrochloride Purity Using the Acclaim PA Column

New 5-micron HALO-5 columns based on Fused- Core particle technology boost the performance of HPLC. Compared to other HPLC columns, HALO-5 columns

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

Introduction to Chromatographic Separations (Chapter 1) Many determinations involve separation followed by analysis chromatography electrophoresis

Welcome to our E-Seminar: Choosing HPLC Columns for Faster Analysis Smaller and Faster

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Prep 150 LC System: Considerations for Analytical to Preparative Scaling

Separation Methods Based on Distributions in Discrete Stages (02/04/15)

Fast methods for the determination of ibuprofen in drug products

High Speed 2D-HPLC Through the Use of Ultra-Fast High Temperature HPLC as the Second Dimension

Important Guidelines for Optimizing Speed and Sensitivity in Small-Molecule LC UV and LC MS

HPLC Praktikum Skript

Thermo Scientific Accucore XL HPLC Columns. Technical Manual

C18 Column. Care & Use Sheet

Routine MS Detection for USP Chromatographic Methods

Impact of Pore Exclusion on Reversed-Phase HPLC Column Performance

Chromatography. Gas Chromatography

Transferring HPLC methods from the HP 1090 Series to the Agilent 1100 Series HPLC system

Dionex IonPac AS28-Fast-4µm column

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Chemistry Gas Chromatography: Separation of Volatile Organics

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter

MODERN HPLC FOR PRACTICING SCIENTISTS

Trends in Method Development for HPLC and UHPLC... the movement toward using solid-core particles

Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

Agilent 1200 Series Rapid Resolution LC and Rapid Resolution LC/MS Optimization Guide

for Acclaim Mixed-Mode HILIC-1 Column

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation

Changing the way you think about HPLC

Understanding Gas Chromatography

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Figure 1. Structures for Vitamin B2 and Vitamin B1.

High Performance Liquid Chromatography

High Performance Liquid Chromatography

GPC/SEC Practical Tips and Tricks. Thomas Dent Applications Scientist Agilent Technologies. October, 2011 Gulf Coast Conference

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

HPLC COLUMNS WILEY-VCH. Theory, Technology, and Practice. Uwe D. Neue with a contribution from M. Zoubair El Fallah

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

LC800. Smart HPLC. Until now UHPLC From now Smart HPLC

HPLC & LC-MS. Techniques to Improve Speed, Resolution and Sensitivity without Investing in Capital Equipment

Gemini-NX Performance paired with Axia Technology

Practical Applications of Method Translation Using the Agilent Method Translation Tool Rita Steed Inside Application Engineer Agilent Technologies

Introduction to Capillary GC

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Using the Agilent 1290 Infinity II Multicolumn Thermostat with Extreme Temperatures and Flow Rates

The Secrets of Rapid HPLC Method Development. Choosing Columns for Rapid Method Development and Short Analysis Times

penta-hilic UHPLC COLUMNS

Chromatography- Separation of mixtures CHEM 212. What is solvent extraction and what is it commonly used for?

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Analytical Methods to Qualify and Quantify PEG and PEGylated Biopharmaceuticals

Introduction to Capillary GC

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

HPLC Background Chem 250 F 2008 Page 1 of 24

Successfully Scaling and Transferring HPLC and UPLC Methods

High Performance Liquid Chromatography

ProPac WCX-10 Columns

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

Liquid Chromatography

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Fall 2012 Due In Class Friday, Oct. 19. Complete the following on separate paper. Show your work and clearly identify your answers.

"Theory and Practice of High Speed Chromatography for Bioanalysis" Stuart Coleman March 20, 2007

A Quality by Design (QbD) Based Method Development for the Determination of Impurities in a Peroxide Degraded Sample of Ziprasidone

Chemistry Instrumental Analysis Lecture 31. Chem 4631

Advanced HPLC Method Development Using Quality by Design

Direct Analysis of Surfactants using HPLC with Charged Aerosol Detection

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Pharmaceutical Analysis of API and Counter Ions in Complex Formulations in a Single Injection

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

What is Chromatography?

Simultaneous Determination of Pharmaceutical Peptides and Acetate Counterions by HPLC Using a Mixed-Mode Weak Anion-Exchange Column

Thermo Scientific Syncronis HPLC Columns. Remarkable separations guaranteed time after time

Expanded capabilities in ammonium and amine detection

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

How to Troubleshoot a Failed System Suitability Test

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Corona Veo the Next Generation Charged Aerosol Detector

Shodex TM ODP2 HP series columns

Comparing the Microlab 600 to Volumetric Glassware and Air Displacement Pipettes. January 2013

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application

ultimate efficiency unleashed [ Cortecs 1.6 µm columns ]

Modernizing the USP Monograph for Acetaminophen

Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water

Transcription:

Impact of Instrument Dispersion on Performance of HPLC Capillary Columns Wendy Roe, Richard A. Henry, and Hillel Brandes Supelco, Div. of Sigma-Aldrich Bellefonte, PA 16823 USA T413131 sigma-aldrich.com/analytical

Introduction The HPLC system flow path comprises column plus instrument components that include valves, tubing, and detector(s). Band spreading occurs as the sample traverses through each component. The band spreading occurring within the column is minimal and related to the chromatographic process while extra-column band spreading occurs outside the column and is more rapid and undesirable. Mathematically, each of the components within the flow path contributes equally to the total observed peak variance as demonstrated by equation 1. Variances associated with instrument components and tubing are considered instrument bandwidth (IBW) and require minimization to optimize column performance. As column diameters decrease, so does the column s impact in equation 1 rendering the extra-column contributions more significant to the total system s performance. 2

Capillary HPLC Instrument Thermo Scientific UltiMate 3000 RSLCnano System (6) (as supplied): Pump equiped with capillary flow meter Heated column compartment UV detector equiped with a 45 nl flow cell Pulled loop autosampler equiped with a 1 µl sample loop, 2.4 µl needle, and electronic Peltier elements for sample cooling 50 µm I.D. tubing connected all components 3

Experimental Measuring Instrument Bandwidth Several methods exist for the determination of extra-column contributions or IBW (1-5) in HPLC and UHPLC systems. An adaption is described below. 1. Replace the column with a low volume connector 2. Prepare mobile phase; 40:60 water:acetonitrile 3. Prepare test probe in mobile phase; naphthalene, 55 µg/ml in mobile phase 4. Set detector sampling rate to adequately capture the peak; 25 Hz 5. Inject low sample volumes; 1 µl, 500 nl, 70 nl 6. Record peak efficiency, N, and retention time, t r 7. Calculate σ = (t r x flow) / N; IBW = 4σ 4

Figure 1. Instrument Bandwidth Performance (No column Installed) Full Loop (1,000 nl) 500 nl, Partial Loop (1,000 nl) 70 nl, Partial Loop (1,000 nl) Flow Rates: 0.5 µl/min 1.0 µl/min 10.0 µl/min Ghost Peak 5

Figure 2. Chromatographic Separation Showing Ghost Peaks Isocratic Method 1 μl Sample Loop, Partial Loop Mode Step Gradient Method 1 μl Sample Loop, Partial Loop Mode 300 nl Injection 70 nl Injection 70 nl Injection Gradient Step 0 2 4 6 Time (min) Conditions: Column: Acclaim PepMap RSLC C18, 150 mm x 300 µm, 2 µm, Mobile phase: 40:60 water:acetonitrile, Flow rate: 10 µl/min, Column temperature: 35 ºC, Sample: uracil (7 µg/ml), acetophenone (7 µg/ml), benzene (750 µg/ml), toluene (775 µg/ml), naphthalene (55 µg/ml) in 60:40 water:acetonitrile, Detector: UV, 254 nm, Collection rate: 25 Hz, Detector cell volume: 45 nl 0 2 4 6 8 Time (min) Conditions: Column: Acclaim PepMap RSLC C18, 150 mm x 300 µm, 2 µm, Mobile phase: (A) water, (B) acetonitrile, gradient: stepwise 5% B for 1 min, 60% B for 9 min; 0.1 min between steps, Flow rate: 10 µl/min, Column temperature: 35 ºC, Sample: uracil (7 µg/ml), acetophenone (7 µg/ml), benzene (750 µg/ml), toluene (775 µg/ml), naphthalene (55 µg/ml) in 60:40 water:acetonitrile, Detector: UV, 254 nm, Collection rate: 6 25 Hz, Detector cell volume: 45 nl

Figure 3. Origin of Ghost Peak from Injection Valve During Partial Loop Injections Step 1 Step 2 1 µl Loop Port to port volume: 348 nl 1 µl Loop Buffer line To Syringe Buffer line To Syringe Buffer line Needle (2.4 µl) is filled with sample. 1 µl Loop Step 3 Buffer line To Syringe Valve switches to load position and partially fills the sample loop. Emptying of sample loop does not completely clear the 1-2 port of sample and allows it to diffuse into sample loop. Valve switches to injection position and the 70 nl sample is flushed onto the column. Diffuse sample from port follows the sample band. 7

Figure 4. Effect of Full Loop Injection Volume on System Volume at 0.5 µl/min 1,000 nl Injection 70 nl Injection mau 0 100 200 300 Instrument Volume: 6.1 min = 3.1 μl Sample Volume: 2.8 min = 1.4 μl mau 0 100 200 300 Instrument Volume: 5.9 min = 3.0 μl Sample Volume: 0.5 min = 0.3 μl 0 2 4 6 8 10 12 Time (min) Total System Volume: 3.1 μl + 1.4 μl = 4.5 0 2 4 6 8 10 12 Time (min) Total System Volume: 3.0 μl + 0.3 μl = 3.3 Peak width taken at baseline 8

Figure 5. IBW for Different Injection Volumes vs. Flow Rate No column installed; Peak width at half height; IBW = 4σ 1,000 nl 500 nl 70 nl IBW for different injection volumes vs flow. Mobile phase: 40:60 water:acetonitrile, Column temperature: 35 ºC, Sample: naphthalene, 55 µg/ml in mobile phase, Detector: UV, 254 nm, Collection rate: 25 Hz, Detector cell volume: 45 nl, Chromelon 6.8 used to calculate efficiency. 9

Figure 6. Minimizing Dispersion Through Changing Sample Solvent Strength 1 3 4 5 2 N naphthalene = 17,200 N naphthalene = 19,300 N naphthalene = 19,200 1.0 2.0 3.0 4.0 5.0 6.0 7.0 Time (min) Elution Order: 1. Uracil 2. Acetophenone 3. Benzene 4. Toluene 5. Naphthalene Sample Solvent (3 Shown): 90:10 Water:Acetonitrile 70:30 Water:Acetonitrile 40:60 Water:Acetonitrile Mobile Phase: 40:60 Water:Acetonitrile Conditions: Column: Acclaim PepMap RSLC C18, 150 mm x 300 µm, 2 µm, Mobile phase: 40:60 water:acetonitrile, Flow rate: 10 µl/min, Column temperature: 35 ºC, Injection: Full loop 70 nl, Detector: UV, 254 nm, Collection rate: 25 Hz, Detector cell volume: 45 nl 10

Figure 7. Column Performance with Optimized Injection Loop and Sample Solvent Focusing Optimum Flow Rate: 8 µl/min 0 2 4 6 8 Time (min) Elution Order Component USP Tailing Efficiency 1 Uracil 1.38 3,356 2 Acetophenone 1.26 11,847 3 Benzene 1.17 17,040 4 Toluene 1.14 18,393 5 Naphthalene 1.1 19,230 Van Deemter Plot for optimized injection loop and sample solvent focusing. Column: Acclaim PepMap RSLC C18, 150 mm x 300 µm, 2 µm, Mobile phase: 40:60 water:acetonitrile, Column temperature: 35 ºC, Injection: full loop (70 nl), Detector: UV, 254 nm, Collection rate: 25 Hz, Detector cell volume: 45 nl 11

Discussion Figure 1 demonstrates that the largest injection volume exhibited a square peak shape showing dispersion only at the peak s leading and tailing edges. Smaller injection volumes exhibited a normal Gaussian peak shape. The smallest injection volume also showed a ghost peak in Figure 1. To identify the source of this ghost peak, a chromatographic test using a five component sample was introduced in 300 nl and 70 nl injection volumes (Figure 2). The 70 nl sample volume showed ghost peaks for each analyte. A step gradient performed with the same sample and 70 nl injection volume exhibited Gaussian peaks for the retained analytes and the same ghost peak behavior for the void marker (Figure 2). This indicated that the ghost peaks derived from the injection process and could be eliminated for retained analytes by focusing the sample on the top of the column. 12

Discussion (contd.) Consultation with the vendor revealed that the ghost peaks originated from a diffusion of the sample from the autosampler valve port. Figure 3 demonstrates the contamination of the sample loop. First, the needle is filled with sample and the waste is drawn into the buffer line. When the valve switches to load the sample loop, the slot connecting ports 6-1 moves to ports 1-2. If not enough sample is drawn into the sample loop to clear the slot, the sample diffuses into the backside of the sample loop. It is this small portion of sample that generates the observed ghost peak(s). The vendor suggested matching the sample loop volume to the injection volume. A 30 µm x 100 mm tube (70 nl volume) was substituted for the 1 µl supplied sample loop and corrected the ghost peaks at this injection volume. Other design solutions are currently being investigated to improve all small volume injections. The 70 nl sample loop was evaluated for dispersion. It exhibited much lower IBW than the 1,000 nl injection volume and reduced the total system volume (Figure 4). 13

Discussion (contd.) Figure 5 summarizes the impact of injection volume on dispersion. As injection size decreases so does the IBW. The 1 µl sample loop does not impact the observed IBW, even for 70 nl injections. It is possible to replace the 1,000 nl sample loop with a 70 nl sample loop and not experience any additional extra-column effects. Samples prepared in eluotropically weaker solvents than the mobile phase focused the sample at the top of the column (Figure 6). This reduced the sample bandwidth upon entering the column and minimized the extracolumn dispersion ahead of the column. This is evident from the increased efficiency and improved plate height for naphthalene. Figure 7 demonstrates the reference column s performance via a Van Deemter Plot. The instrument was optimized for small volume injections using a 70 nl sample loop, and dispersion effects ahead of the column were minimized by a weak sample solvent. The similar plate heights achieved for the three most retained analytes indicated the band spreading was effectively minimized to optimize the column performance. 14

Conclusions Choice of sample volumes required careful consideration when using the Thermo Scientific UltiMate 3000 RSLCnano System. Large sample volumes, >500 nl, yielded large IBW values. Small sample volumes, <100 nl, experienced diffusion from the buffer line and required a smaller volume sample loop to perform well. IBW values for the smaller sample loop were comparable to the larger sample loop. It is critical to optimize the sample loop size to the desired sample volume. Focusing the sample at the entrance to the column minimized the extracolumn dispersion. This technique is recommended for optimal column performance when using this system for isocratic capillary HPLC applications. Other systems could benefit from this technique to improve column performance. Future experiments to further optimize the system include additional design changes to improve small volume injections. As hardware changes are made, further IBW experiments are necessary to ensure extra column effects are minimized. 15

References 1. R.E. Majors LCGC NORTH AMERICA, Vol. 21 No. 12, 1124-1133 (December 2003). 2. M. W. Dong, Modern HPLC for Practicing Scientists (Wiley-Interscience, New York 2006). 3. F. Gritti, C.A. Sanchez, T. Farkas and G. Guiochon, J. of Chromatography A, 1217, 3000 3012 (2010). 4. R.A. Henry and D. S. Bell, LCGC NORTH AMERICA, Vol. 23 No. 5, 2-7 (May 2005). 5. R. A. Henry, H. K. Brandes, D. T. Nowlan and J. W. Best, Practical Tips for Operating UHPLC Instruments and Columns, LCGC North America April 2013 (article in press). 6. Dionex website. http://www.dionex.com/en-us/webdocs/57741-man-lc- U3000-WPS-PL(RS)+FC-Operation-DOC4828-2050-1-5.pdf (accessed March 1, 2013). 16

Trademarks Acclaim and UltiMate are registered trademarks of Thermo Fisher Scientific Inc. PepMap is a trademark of Thermo Fisher Scientific Inc. 17