PHS-25CW/3BW Benchtop ph Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD

Similar documents
PHS-25CW/3BW Benchtop ph Meter Instruction Manual

Bante210 Benchtop ph Meter Instruction Manual

User s Guide. Waterproof Palm ph Meter Model PH220. Introduction

Bante900P Portable Multiparameter Water Quality Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD

Appendices towards the Operation Manual of Ultrasonic milk analyzer Lactoscan

STANDARD OPERATING PROCEDURE

INE-PHS-3E ph Meter Operating Manual

PH700 Benchtop ph Meter Instruction Manual

This sensor is now also available with a sturdier flat electrode which enables testing even smaller samples than the conventional electrode.

Rapid RH Portable ph Meter. User's Manual

ph 5 Meter Manual The LaMotte 5 Series

PC100 Handheld ph/cond./tds/sal. Meter. Instruction Manual

Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter

GENERAL INSTRUCTIONS GENERAL PREPARATION

symphony Reference Electrodes

symphony Multisensor Probes

Chemistry MultiMeasure Sensor PS-2170

ph Probe Silver / silver chloride Reads Range / Resolution 95% in 1s Response time 100 PSI Max pressure 60m (197 ft) Max depth 1 99 C

Sodium Sealed Electrode Sensor Bundle Product Number: ENSOD051

OPERATING INSTRUCTIONS

INE-DDB-303A Portable Conductivity Meter Instruction Manual

Fluoride Sealed Electrode Sensor Bundle Product Number: ENFLU049

Guide to ph Analysis.

GENERAL INSTRUCTIONS GENERAL PREPARATION

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

Operating Instructions for Ammonium ISE Specifications

Instruction Manual Portamess 911 ph

DOC /2013, Edition 1. User Manual

Ammonium Sealed Electrode Sensor Bundle Product Number: ENAMN020A

All about ph! Is it as simple as it seems? The ph Workshop. Rick Noone Thermo Scientific Water Analysis Instruments, Orion products May 22, 2014

10 samples per second

ph Probe Spear Tip Silver / silver chloride Reads Range / Resolution 95% in 1s Response time 100 PSI Max pressure 60m (197 ft) Max depth

The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium ions in aqueous solutions.

Standard Operating Procedure for GLNPO Specific Conductance: Conductivity Bridge

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com INSTRUCTION MANUAL. Milwaukee Refractometer

COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

1) A ph/mv meter or an ion meter, either line operated or portable.

ph electrode Instruction Manual For use with the RAH-210 & RPH-250

Understanding ph Troubleshooting and Diagnostic information

COLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL

POTASSIUM ION SELECTIVE ELECTRODE

Use a pipet bulb when pipeting cyanide solutions, as these solutions are highly toxic.

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

INSTRUCTION MANUAL Milwaukee Refractometer MA881 Refractometer for Invert Sugar Measurements

ION 5 and ION 6 Acorn Series Meters

COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

Electrodes for ph measurements. - when you need to be sure...

[Agilent Technologies] [Electrochemistry Products] Quick Reference

S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) Fax: +44 (0)

User Manual. Portamess 911 ph and Portamess 911 X ph

Richter Optica. Instructions for Model: S850 Stereo Zoom Microscope

Instruction Manual. Waterproof Portable Colorimeter C401 / 301 / 201 / 101 / 102 / 103 / 104 / 105. Part of Thermo Fisher Scientific

User instructions. Gel filled ORP/Redox electrode: Model MTC10101, MTC10103, MTC10105, MTC10110, MTC10115 or MTC10130

Advanced Sensor Technologies, Inc. TEL: TOLL FREE: WOW-ASTI ( ) FAX:

The Cole-Parmer Nitrate Ion Electrodes are used to quickly, simply, accurately, and economically measure nitrate in aqueous solutions.

ph Probe ReallyEasyData com

C = concentration as ppm sulfide V t = volume of titrant at endpoint V s = volume of standard used (10 ml)

ph Electrode basics ph Electrodes Model 19 B /1 =1 m 24 B /1 =1 m

PRETREATMENT TECHNICAL DATA SHEET CHROMIUM-FREE ORGANIC PASSIVATING RINSE PRODUCT DESCRIPTION

GENERAL INSTRUCTIONS GENERAL PREPARATION

NeoMet. Instruction Manual. istek, Inc. Portable ph/ise/mv/orp, DO/O2/Air, Conductivity/TDS/Salinity/Temp Meter 29D 47C 415CP 815PDC

Buffer Preparation. Learning Objectives:

Differential ph Probe Type 18

METHOD 9040B. ph ELECTROMETRIC MEASUREMENT

INSTRUCTION MANUAL Milwaukee Refractometer MA886 Refractometer for Sodium Chloride Measurements

COLE-PARMER LABORATORY NITRATE ION ELECTRODE INSTRUCTION MANUAL

Laboratory Certification Workshop

COLE-PARMER INDUSTRIAL CALCIUM ION ELECTRODE INSTRUCTION MANUAL

PRETREATMENT TECHNICAL DATA SHEET CHROMIUM-FREE ORGANIC PASSIVATING RINSE PRODUCT DESCRIPTION

Instruction Manual C401/301/201/102/101. Portable Colorimeter. 99 Washington Street Melrose, MA Phone Toll Free

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com INSTRUCTION MANUAL. Milwaukee Refractometer

INSTRUCTION MANUAL Milwaukee Refractometer MA888 Refractometer for Ethylene Glycol Measurements

NO 500 DIN, NO 500 BNC NO 800 DIN, NO 800 BNC

National Optical & Scientific Instrument Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210)

Our Drop Counter sensor now features housing for two electrode sensors, an anti-twist mechanism, an indicator LED and two cable guides.

ORP Probe ORP. Reads. +/-2000mV. Range. 95% in 1s. Response time 100 PSI. Max pressure. 60m (197 ft) Max depth 1 80 C. Temperature range C.

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

NITRATE ION SELECTIVE ELECTRODE

ORP Probe. Mini ORP. Reads. +/-2000mV. Range SMA. Connector. 95% in 1s. Response time 100 PSI. Max pressure. 60m (197 ft) Max depth 1 80 C

COLE-PARMER LABORATORY CYANIDE ION ELECTRODE INSTRUCTION MANUAL

PRETREATMENT TECHNICAL DATA SHEET A CHROME-FREE FINAL RINSE PRODUCT DESCRIPTION

testo 230 ph- and temperature measuring instrument Instruction manual

Ion Selective Electrode Probe

COLE-PARMER LABORATORY CALCIUM ION ELECTRODE INSTRUCTION MANUAL

INSTRUCTION MANUAL. Type: ORP500 C, ORP500 CD, ORP222 CD, ORP223 CD EN Introduction Helpful Operating Tips. 2

To measure ph s in a variety of solutions and mixtures and to account for the results obtained.

Operation Manual. SPECTRO-NANO4 Nucleic Acid Analyzer PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION

The Oakton Cadmium Ion Electrodes are used to quickly, simply, accurately, and economically measure cadmium ions in aqueous solutions.

Instruction Manual HI Refractometer for Invert Sugar Measurements

HI Refractometer for Sodium Chloride Measurements INSTRUCTION MANUAL

Razi Vaccine & Serum Research Institute

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

INSTRUCTION MANUAL Milwaukee Refractometer MA872 Refractometer for Fructose Measurements

HI Propylene Glycol Refractometer INSTRUCTION MANUAL

User Guide. Green ph Electrodes

Compact ph Measuring Instruments With Pro Engineering

COLE-PARMER REPLACEABLE MEMBRANE CALCIUM ION ELECTRODES INSTRUCTION MANUAL

Introduction. Objectives

QAM-I-111 Operation and Calibration of the Conductivity Meter

Transcription:

PHS-25CW/3BW Benchtop ph Meter Instruction Manual BANTE INSTRUMENTS CO., LTD

PHS-W Series Benchtop ph Meter 1 Introduction Thank you for selecting the PHS-W series benchtop ph meter. This manual provides a step-by-step guide to help you operate the meter, please carefully read the following instructions before use. Unpacking Before unpacking, ensure that the current work environment meets following conditions. Relative humidity is less than 80 %. Ambient temperature is greater than 0 o C and less than 60 o C. No potential electromagnetic interference. The following list describes the standard components of the meter. After the unpacking, please check all components are complete. If any are damaged or missing, please contact nearest distributor. ph Meter Electrode Arm ph4.01/7.00/10.01 Buffer Pouches ph Electrode Temperature Probe (Only for PHS-3BW) DC9V Power Adapter

PHS-W Series Benchtop ph Meter 2 Display PHS-W series benchtop ph meter is equipped with an easy-read LCD display that used to show measured values and mode icons. The following table describes the function of each icon. INDEX: Measurement mode icon: Indicates the meter is in the measurement mode. Electrode slope icon: Indicates the average slope of the ph electrode. Calibration mode icon: Indicates the meter is in the calibration mode. ERR Error alarm icon: Indicates the ph buffer or electrode has expired. Temperature setting icon: Indicates the meter is in the temperature setting mode. ATC Automatic Temperature Compensation: Indicates the temperature compensation is enabled. Keypad The meter has a succinct membrane keypad, names and symbols describe the each function key controls. KEY ON CAL ph mv o C ENTER FUNCTION Switches the meter ON/OFF. Starts calibration. Enters the ph measurement mode. Enters the mv measurement mode. Sets the temperature. Increase the setting value. Decrease the setting value. Confirms the calibration or settings.

PHS-W Series Benchtop ph Meter 3 Connectors PHS-W series benchtop ph meter provides 3 connectors are located on the rear panel. Listed in the below table are the details of these connectors. NO. CONNECTOR DESCRIPTION 1 ELECTRODE Used for connecting the ph or ORP electrode. 2 ATC Used for connecting the temperature probe (Only for PHS-3BW meter) 3 POWER Used for connecting the power adapter. Installing the Electrode Holder Take out the electrode arm from the packaging. The base plate of the electrode holder has an irregular round hole, the electrode arm has a connecting rod. Insert the connecting rod into the irregular round hole and swivel the electrode arm 90 o. Electrode holder is now ready to swing into desired position. Adjustment of Electrode Arm After installation, if the electrode arm automatically rises or falls, you need to adjust the screws until arm locate at any position. 1. Remove the plastic cover from the electrode arm. 2. Use the screwdriver to tighten the screw moderately. 3. Insert the plastic cover to previous position. Installation is completed.

PHS-W Series Benchtop ph Meter 4 Connecting the Electorde 1. Take out the ph electrode from the packaging. Follow the steps below to place the electrode into left or right side of the electrode arm. 1. Insert the electrode 2. Hook up the cable 3. Hook up the cable 2. Insert the BNC connector into the connector socket labeled ELECTRODE. Rotate and push the connector clockwise until it locks. After the connection is completed, DO NOT pull on the cable. Always make sure that the connector is clean and dry. Connecting the Power Adapter 1. Before plugging in the power adapter, ensure that its voltage matches the local main voltage. 2. Insert the connector to the power socket. The meter is now ready for use.

PHS-W Series Benchtop ph Meter 5 Preparing the ph Buffer Solutions PHS-W series benchtop ph meter is packaged with three ph buffer packets required for calibration. Open the ph7.00 buffer packet, place the powder into a 250ml volumetric flask. Pour distilled water 250ml to scale line, mix the solution until the reagent is completely dissolved. Preparation of ph4.01 and 10.01 standard buffer solutions are the same as above. Prepared standard buffer solutions should be stored in hermetically sealed glass containers. Prior to Use Remove the protective cap from the bottom of the electrode. If the glass sensitive membrane has dried out, soak the electrode in 3M KCL solution (ph adjusted to 4.0) for at least 15 minutes. DO NOT store the electrode in distilled or deionised water that will render the electrodes useless. Power On/Off Press the ON key to turn on the meter, the display shows the measured value. Press and hold the ON key for 5 seconds, the meter will turn off.

PHS-W Series Benchtop ph Meter 6 Temperature Compensation For better accuracy, we recommend that using a temperature probe for the calibration and measurements. Manual Temperature Compensation 1.1 Use a high-accuracy thermometer to measure the temperature of sample. 1.2 Press the o C key to enter the temperature setting mode. 1.3 Press the or key to set the temperature value. 1.4 Press the Enter key to confirm, the meter returns to the measurement mode. Setting is completed. Automatic Temperature Compensation (Only for PHS-3BW meter) 2.1 The PHS-3BW benchtop ph meter is supplied with a TP-100K temperature probe. Place the temperature probe into the round hole of the electrode arm. 2.2 Insert the phone plug to the connector socket labeled ATC. Ensure the connector is fully seated. When the connection is done, the ATC indicator will show on the display. The meter is now switched to the automatic temperature compensation mode.

PHS-W Series Benchtop ph Meter 7 ph Calibration PHS-W series benchtop ph meter can perform 2 points calibration. The first calibration point must be ph7.00. If your sample is belong to acidic substances, select the ph4.01 buffer solution as second calibration point, if the alkaline substances, select the ph10.01 buffer solution. The meter will automatically recognize and calibrate to following standard buffer values. USA Standard Buffers ph4.01, 7.00, 10.01 NIST Standard Buffers ph4.01, 6.86, 9.18 The meter must be calibrated prior to first use or new electrode replaced. To ensure accuracy, regular calibration is recommended. DO NOT reuse the ph buffer solution after calibration, contaminants in solution will affect the calibration and eventually the accuracy of the measurement. 1. Rinse the ph electrode with distilled water. Press the Cal key, the meter enters the calibration mode, the display shows CAL/pH7.00. 2. Place the ph electrode into the ph7.00 standard buffer solution, the end of the electrode must be completely immersed into the calibration solution. Stir the electrode gently to create a homogeneous solution. 3. Press the Enter key, the meter begins the calibration. Wait for the reading to stabilize, the displayed value will flash 3 times indicating that first calibration point is completed. The display automatically shows --- and waits for calibrating the second point.

PHS-W Series Benchtop ph Meter 8 4. Rinse the ph electrode with distilled water. Place the electrode into the ph4.01 (or 10.01) standard buffer solution. Press the Enter key, the meter automatically recognizes the ph buffer solution and begins the calibration (e.g., ph4.01) 5. Wait for the reading to stabilize, the displayed value will flash 3 times, the meter automatically shows electrode slope and returns to the measurement mode. Calibration is completed. Electrode Slope 99% Exit the Calibration During the calibration process, press the ph key, the meter will exit the calibration and return to the measurement mode. The electrode icon ( ) shows the average slope of the ph electrode when the calibration is completed. If the slope or calibration result does not meet measurement requirements, the icon will disappear on the display. If the meter shows ERR on the display, please check the ph electrode and ensure the ph buffer solutions are fresh and uncontaminated.

PHS-W Series Benchtop ph Meter 9 ph Measurement 1.1 Press the ph key to enter the ph measurement mode. 1.2 Rinse the ph electrode with distilled water to remove any impurities adhering to the probe body. 1.3 Place the electrode into the sample solution, stir the electrode gently. 1.4 Record the measured value when the reading is stable. mv Measurement 2.1 Press the mv key to enter the mv measurement mode. 2.2 Rinse the electrode thoroughly with distilled water. 2.3 Place the electrode into the sample solution, stir the electrode gently. 2.4 Record the measured value when the reading is stable. ph Electrode Care and Maintenance Since ph electrode is susceptible to dirt and contamination, clean as necessary depending on the extent and condition of use. After measuring: rinse the electrode in distilled water, store the electrode into the 3M KCL solution. Salt deposits: soak the electrode in warm tap water to dissolve deposits, then thoroughly rinse with distilled water. Oil or Grease film: wash the glass sensitive membrane of electrode gently in some detergents and water. If necessary, using the alcohol to clean the sensitive membrane, then rinse with distilled water. Place the electrode in the 3M KCL solution for at least 30 minutes. Clogged reference junction: heat a diluted KCl solution to 60 o C to 80 o C. Place the electrode into the heated solution for about 10 minutes. Allow the electrode to cool in some unheated KCl solution. Protein deposits: prepare a 1% pepsin solution in 0.1M of HCL. Place the electrode in the solution for 10 minutes. Rinse the electrod with distilled water. Rectivating the Electrode If stored and cleaned properly, the electrode should be ready for immediate use. However, a dehydrated sensitive membrane may cause sluggish response. To rehydrate the sensitive membrane, immerse the electrode in a ph4.01 buffer solution for 10 to 30 minutes. If this fails, the electrode requires activation. 1. Soak the electrode in 0.1M HCl for 5 minutes. 2. Remove and rinse with deionized water, then place in 0.1M NaOH for 5 minutes. 3. Remove and rinse again, and soak in 3M KCL solution for at least 30 minutes.

PHS-W Series Benchtop ph Meter 10 Optional ph Electrodes PHS-W series benchtop ph meter comes with a general purpose ph electrode that is used to measure the ph of the liquids. If this electrode can not meet your measurement requirements, please refer to the table below to select an applicable probe. SAMPLE TYPE P11 P12 P13 P15 P16 P18 P19 P21 E201 E202 Agar Beer Blood Products Bread, Dough Cement Cosmetics Dairy Products Education Fats/Cream Field Use Fish Products Lab Flasks Low Ionic Meat, Cheese Micro Samples Paint Photographic Soil Surface Test Tubes Tris Buffer Viscose Samples Troubleshooting LCD DISPLAY CAUSE CORRECTIVE ACTION ERR Electrode dried out Measured value is out of range Incorrect ph buffer solutions Electrode has expired Soak the ph electrode in 3M KCL solution at least 20 minutes. Check the electrode whether clogged, dirty or broken. Using the fresh ph buffer solutions for calibration. Replace the ph electrode.

PHS-W Series Benchtop ph Meter 11 Specifications Model PHS-25CW PHS-3BW Range 0.00~14.00pH Accuracy ± 0.05pH ± 0.01pH ph mv Temperature Resolution 0.01pH Calibration Points 2 points Calibration Solutions USA (ph4.01/7.00/10.01) Temperature Compensation 0~100 C, Manual 0~100 C, Automatic or Manual Range -1999~1999mV Accuracy ± 10mV ± 1mV Resolution 1mV Range --- 0~100 C Accuracy --- ± 1 C Resolution --- 1 C Connector BNC Others Power Requirements Dimensions Weight DC9V, using AC adapters, 220VAC/50Hz 210(L) 205(W) 65(H)mm 1.5kg

Hazardous Substance Statement Bante Instruments is committed to the reduction and eventual elimination of all hazardous substances in both the manufacturing process and finished products we supply. We have an active manufacturing and procurement program to minimize and eliminate the use of harmful heavy metals such as cadmium, lead, mercury and the like. New technologies and design parameters are also promoting these efforts and we expect to have little or no such materials in our product in the coming years. We welcome our customer suggestions on how to speed up these efforts. Warranty The warranty period for meter is one year from the date of shipment. Above warranty does not cover sensor and calibration solutions. Out of warranty products will be repaired on a charged basis. The warranty on your meter shall not apply to defects resulting from: Improper or inadequate maintenance by customer. Unauthorized modification or misuse. Operation outside of the environment specifications of the products. For more information, please contact the nearest authorized distributor. BANTE INSTRUMENTS CO., LTD E-mail: banteinstruments@yahoo.com Phone: +8621-6404-1598 Website: www.bante-china.com The information in this document is subject to change without notice. Copyright Bante Instruments, 2013. All rights reserved.