PC235: 2008 Lecture 5: Quantitation. Arnold Falick

Similar documents
Chapter 4. strategies for protein quantitation Ⅱ

NPTEL VIDEO COURSE PROTEOMICS PROF. SANJEEVA SRIVASTAVA

Mass Spectrometry and Proteomics - Lecture 5 - Matthias Trost Newcastle University

Quantitative Proteomics

Quantitative Proteomics

Comprehensive support for quantitation

Tutorial 1: Setting up your Skyline document

Atomic masses. Atomic masses of elements. Atomic masses of isotopes. Nominal and exact atomic masses. Example: CO, N 2 ja C 2 H 4

Quantitation of a target protein in crude samples using targeted peptide quantification by Mass Spectrometry

Isotopic-Labeling and Mass Spectrometry-Based Quantitative Proteomics

Protein Quantitation II: Multiple Reaction Monitoring. Kelly Ruggles New York University

Computational Methods for Mass Spectrometry Proteomics

Workshop: SILAC and Alternative Labeling Strategies in Quantitative Proteomics

SRM assay generation and data analysis in Skyline

Protein Quantitation II: Multiple Reaction Monitoring. Kelly Ruggles New York University

Protein analysis using mass spectrometry

Aplicació de la proteòmica a la cerca de Biomarcadors proteics Barcelona, 08 de Juny 2010

Reagents. Affinity Tag (Biotin) Acid Cleavage Site. Figure 1. Cleavable ICAT Reagent Structure.

6 x 5 Ways to Ensure Your LC-MS/MS is Healthy

SILAC and TMT. IDeA National Resource for Proteomics Workshop for Graduate Students and Post-docs Renny Lan 5/18/2017

We are IntechOpen, the first native scientific publisher of Open Access books. International authors and editors. Our authors are among the TOP 1%

Methods for proteome analysis of obesity (Adipose tissue)

Protein Identification Using Tandem Mass Spectrometry. Nathan Edwards Informatics Research Applied Biosystems

Key questions of proteomics. Bioinformatics 2. Proteomics. Foundation of proteomics. What proteins are there? Protein digestion

Quantitative analysis of the proteome

Figure S1. Interaction of PcTS with αsyn. (a) 1 H- 15 N HSQC NMR spectra of 100 µm αsyn in the absence (0:1, black) and increasing equivalent

Chem 250 Unit 1 Proteomics by Mass Spectrometry

Improved 6- Plex TMT Quantification Throughput Using a Linear Ion Trap HCD MS 3 Scan Jane M. Liu, 1,2 * Michael J. Sweredoski, 2 Sonja Hess 2 *

Amine specific Labeling Reagents for Multiplexed Relative and Absolute Protein Quantitation

Quantitative analysis of the proteome. Proteomics Data Standards

UCD Conway Institute of Biomolecular & Biomedical Research Graduate Education 2009/2010

Modeling Mass Spectrometry-Based Protein Analysis

MS-based proteomics to investigate proteins and their modifications

Chapter 2 What are the Common Mass Spectrometry-Based Analyses Used in Biology?

Proteomics. November 13, 2007

Nature Methods: doi: /nmeth Supplementary Figure 1. Fragment indexing allows efficient spectra similarity comparisons.

High-Throughput Protein Quantitation Using Multiple Reaction Monitoring

Proteome-wide label-free quantification with MaxQuant. Jürgen Cox Max Planck Institute of Biochemistry July 2011

PROTEOMICS IN VASCULAR BIOLOGY

A label-free quantification method by MS/MS TIC compared to SILAC and spectral counting in a proteomics screen

TOMAHAQ Method Construction

Comparative and Quantitative Global Proteomics Approaches: An Overview

DIA-Umpire: comprehensive computational framework for data independent acquisition proteomics

Rapid Distinction of Leucine and Isoleucine in Monoclonal Antibodies Using Nanoflow. LCMS n. Discovery Attribute Sciences

Proteolytic 18O-Labeling Strategies for Quantitative Proteomics

Protocol. Product Use & Liability. Contact us: InfoLine: Order per fax: www:

TUTORIAL EXERCISES WITH ANSWERS

Protocol. Product Use & Liability. Contact us: InfoLine: Order per fax: www:

Tandem mass spectra were extracted from the Xcalibur data system format. (.RAW) and charge state assignment was performed using in house software

CHEMICAL COVALENT LABELING AND TANDEM MASS SPECTROMETRY FOR TARGETED PROTEIN CHARACTERIZATION, QUANTIFICATION AND STRUCTURAL ANALYSIS YALI LU

Introduction to Proteomics & Bottom-up Proteomics

Mass spectrometry has been used a lot in biology since the late 1950 s. However it really came into play in the late 1980 s once methods were

Peter A. DiMaggio, Jr., Nicolas L. Young, Richard C. Baliban, Benjamin A. Garcia, and Christodoulos A. Floudas. Research

Chemical Labeling Strategy for Generation of Internal Standards for Targeted Quantitative Proteomics

PeptideProphet: Validation of Peptide Assignments to MS/MS Spectra. Andrew Keller

See for options on how to legitimately share published articles.

Targeted protein quantification

Isobaric Labeling-Based Relative Quantification in Shotgun Proteomics

Guide to Peptide Quantitation. Agilent clinical research

Effective Strategies for Improving Peptide Identification with Tandem Mass Spectrometry

SERVA ICPL Kit (Cat.-No )

In-gel digestion of immunoprecipitated proteins separated by SDS-PAGE

Effective desalting and concentration of in-gel digest samples with Vivapure C18 Micro spin columns prior to MALDI-TOF analysis.

Rapid and Sensitive Fluorescent Peptide Quantification Using LavaPep

Purdue-UAB Botanicals Center for Age- Related Disease

Workflow concept. Data goes through the workflow. A Node contains an operation An edge represents data flow The results are brought together in tables

LECTURE-13. Peptide Mass Fingerprinting HANDOUT. Mass spectrometry is an indispensable tool for qualitative and quantitative analysis of

SeqAn and OpenMS Integration Workshop. Temesgen Dadi, Julianus Pfeuffer, Alexander Fillbrunn The Center for Integrative Bioinformatics (CIBI)

Translational Biomarker Core

Advances in quantitative proteomics using stable isotope tags

Towards the Prediction of Protein Abundance from Tandem Mass Spectrometry Data

Developing Algorithms for the Determination of Relative Abundances of Peptides from LC/MS Data

Genome wide analysis of protein and mrna half lives reveals dynamic properties of mammalian gene expression

Increasing the Multiplexing of Protein Quantitation from 6- to 10-Plex with Reporter Ion Isotopologues

Lecture 15: Realities of Genome Assembly Protein Sequencing

Electron Transfer Dissociation of N-linked Glycopeptides from a Recombinant mab Using SYNAPT G2-S HDMS

Proteomics: the first decade and beyond. (2003) Patterson and Aebersold Nat Genet 33 Suppl: from

Peptide Targeted Quantification By High Resolution Mass Spectrometry A Paradigm Shift? Zhiqi Hao Thermo Fisher Scientific San Jose, CA

The Agilent 6495 Triple Quadrupole LC/MS: Peptide Quantitation Performance

WADA Technical Document TD2003IDCR

Qualitative Proteomics (how to obtain high-confidence high-throughput protein identification!)

Quantitative Analysis of Opioids Using a Triple-Quadrupole GC/MS/MS

Designed for Accuracy. Innovation with Integrity. High resolution quantitative proteomics LC-MS

Introduction to Proteomics: Fragmentation of protonated peptides and manual sequencing

Exam I Answer Key: Summer 2006, Semester C

Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass Spectrometry

Highly Sensitive and Rugged GC/MS/MS Tool

Optimization and Use of Peptide Mass Measurement Accuracy in Shotgun Proteomics* S

Yifei Bao. Beatrix. Manor Askenazi

WADA Technical Document TD2015IDCR

Quantitative analysis of small molecules in biological samples. Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB.

Protein separation and characterization

Overview - MS Proteomics in One Slide. MS masses of peptides. MS/MS fragments of a peptide. Results! Match to sequence database

Chapter 5. Complexation of Tholins by 18-crown-6:

Biochemistry 3100 Sample Problems Chemical and Physical Methods

Mass Spectrometry Based De Novo Peptide Sequencing Error Correction

HOWTO, example workflow and data files. (Version )

Biological Mass Spectrometry

Mass spectrometry based proteomics

Supporting Information. Experimental details

Transcription:

PC235: 2008 Lecture 5: Quantitation Arnold Falick falickam@berkeley.edu

Summary What you will learn from this lecture: There are many methods to perform quantitation using mass spectrometry (any method worth mentioning has a trendy acronym) Quantitation can be used for studying protein changes, modification changes or as a method for identifying specific results. Quantitation is normally relative (between samples), but can also be absolute. Software for analysis of this type of data is still being developed.

Mass Spectrometry is not Inherently Quantitative There are a number of factors that contribute to peak intensity in a mass spectrum Absolute amount of component Amount of component in relation to other components in the sample Ease of ionization of component Components that are of the same concentration may give drastically different signal intensities E.g., if a protein is digested into peptides, different peptides give very different signal intensities (and some peptides may not be detected at all!)

Mass Spectrometry is not Inherently Quantitative The same sample can look very different based on the analysis technique tryptic digest in CHCA Same digest in DHB

How Can Mass Spectrometry be used for Quantitation? Different types of quantitation can be performed using mass spectrometry Relative quantitation can be performed for chemically similar components present in the same sample Relative quantitation can be performed by comparing peak intensities of the same component in similar samples analyzed sequentially. Comparison to a similar component at a known concentration can be used to estimate absolute quantities. Absolute protein quantities can be crudely estimated without comparison.

MS Methods used to Produce Quantitative Data Stable Isotope Labeling: Incorporation of stable isotopes, samples are mixed and then compared to each other. [Isotopes can be incorporated metabolically or chemically] Label-free Quantitation: Samples are run sequentially and either peak intensities or numbers of peptides identified from a protein are compared.

Enzymatic Incorporation of Stable Isotopes Serine proteases; e.g., trypsin, will exchange carboxyl terminal oxygens with solution. Incubating a sample in H 2 18 O will cause incorporation of two oxygens, producing a 4Da mass shift. Mix unlabeled and labeled samples together and analyze. Advantages Labels all peptides. Does not affect peptide fragmentation. Can be useful for peptide identification as all C-terminal ions (e.g., y-ions) will give pairs of peaks 4 Da apart.

Problems Somewhat difficult to get complete incorporation of both oxygens. 16 O/ 18 O Labeling Label will slowly exchange back once returned to normal water solution. 4 Da is a relatively small mass shift; can get overlap of light and heavy isotope clusters for large peptides. In Summary A quick, easy and cheap quantitation method, but not the most accurate. Wang, J. et al. J Proteom Res (2007) 12 4601-4607

ICAT (Isotope-Coded Affinity Tag) Advantage: Affinity Tag (Biotin) Linker Region Light: 8 1 H Heavy: 8 2 H Cysteine Reactive Group (Iodoacetamide) Can selectively enrich labeled peptides. Problems: Light and heavy reagents do not co-elute in reverse phase, making quantitation of LC-MS data more complicated. Protein must contain Cys The biotin tag causes problems: 1. Difficult to elute quantitatively from streptavidin 2. Tag yields intense fragment ions Fragment biotin rather than peptide Often difficult to identify peptide Gygi, S.P. et al. Nat. Biotechnol. (1999) 17 994-999

cicat (cleavable ICAT) 1. Affinity enrich modified peptides. 2. Cleave tag using TFA, removing the biotin. The smaller isotopically labeled tag fragments less than the whole protein, making peptide identification simpler. Light and heavy reagents co-elute, making quantitation easier. Issue Most proteins are identified by a single peptide, requiring high accuracy and reproducibility in measurements to be able to draw conclusions. Hansen, K. C. et al. Mol Cell Proteomics (2003) 2 299-314

SILAC (Stable Isotope Labeling of Amino Acids in Culture) In cell culture, replace one (or two) amino acids in the culture medium with 13 C and/or 15 N equivalent. Which amino acid/s should you choose? You want to label as many peptides as possible It needs to be an amino acid that is big enough to give a mass shift of at least 4Da. For tryptic digests, lysine (C6, N2) and/or arginine (C6, N4) are sensible choices. Need to be aware that some amino acids can be synthesized by the cell, or can be converted from one amino acid to another. Safest bet is to keep cells well fed, so there is never a shortage of any amino acid. Ong, S-E. et al. Mol Cell Proteom. (2002) 1 5 376-386

SILAC for Modification Quantitation Also possible to label modifications through addition of a labeled version of the donor. S-adenosyl methionine (AdoMet) is the sole donor of methyl groups for protein (and DNA, RNA) methylation. It is synthesized from methionine By adding 13 C 2 H 3 methionine in culture one can label lysine and arginine methylations e.g. in histones. Ong, S-E, Nat Methods (2004) 1 (2) 119-126

Multiplexing SILAC Three samples can be compared in the same experiment. SILAC Summary Strengths: Can get essentially complete labeling (need about 5 population doublings for complete incorporation). Labeling does not affect fragmentation. Weakness: Requires cell culture. Ong, S-E, Nat Chem Biol (2005) 1 (5) 252-262

Heavy Methyl Esterification Can methyl esterify side-chain carboxylic acids Used to reduce non-specific binding in IMAC for phosphopeptide enrichment. Can use normal and deuterated methanol, mix samples and compare. Problem Reaction is sometimes difficult to get to completion. 3 Da mass shift is smaller than ideal (get overlap of peptide isotope patterns).

Amine-Reactive Labels Deuterated versions of many reagents have been used for labeling peptide N-termini and lysine side-chains. E.g.: Acetic anhydride Propionic anhydride Succinic anhydride Phenyl isocyanate

Isotopic Tag for Relative and Absolute Quantitation (itraq) The cleverest and most elegant amine-reactive tagging reagents. Amino-groups are modified (N-terminus, Lys) A set of four reagents, allowing simultaneous comparison of four samples. All reagents attach nominally same mass tag, giving one peak in the MS, but produce different reporter ions in MSMS, allowing relative quantitation. Ross et al. Mol Cell Proteomics (2004) 3, 1154.

itraq Reagents Ross et al. Mol Cell Proteomics (2004) 3, 1154.

Example itraq Data Quantitation MS: Single Peak

When four is not enough 8-plex itraq 8-plex version produces reporter ions at m/z 113, 114, 115, 116, 117, 118, 119 and 121 Have to pick m/z that are not normally observed in peptide CID spectra. Choe, L., et al. Proteomics (2007) 7 (issue#20) 3651-60

β-elimination followed by Michael addition of DTT (BEMAD) Primarily aimed at Ser/Thr post-translational modifications Phosphorylation O-Glycosylation (e.g. O-GlcNAcylation) O-Sulfation Alkylated cysteines also undergo reaction. Using 2 H 6 -DTT, can get quantitative information. Isotopic labeling can also be used to identify type of O-linked modification.

Absolute Quantitation (AQUA) All mass spectrometry quantitation so far has been relative. What if it is relative to a known amount? Absolute quantitation! For highest accuracy, a calibration curve of known amounts spiked into a sample should be obtained, then from this a peak intensity can be converted into an amount. Peak intensity for a given component is roughly linear with respect to intensity Add isotopic labeled version of peptide of known amount to sample, then compare intensity to unlabeled version in the sample 1. 1 Gerber et al. PNAS (2003) 100 (12) 6940-6945

Method of standard addition Sample + 2x std addition signal Sample + std addition Sample conc. Added concentration

itraq for Absolute Quantitation

Mass Spectrometry-based Protein Quantification Methods At what point are the samples combined? Tissue/ Cells SILAC cicat O 16 /O 18 itraq Internal Standards Protein Peptides MS Analysis

Isotopic Differentiation of Interactions as Random or Targeted (I-DIRT) Identifying non-specific interactors isolated as part of an IP/tag pulldown. Tackett et al, J. Prot. Res. (2005) 4, 1752.

Localization of Organelle Proteins by Isotope Tagging (LOPIT) Density centrifugation is used for organelle separation. A given organelle will span several fractions. Most fractions contain proteins from more than one organelle. Distribution of proteins between fractions should be the same for those in the same organelle. If separate fractions are labeled with itraq, the reporter peak intensity pattern can be used to identify organellar location. Sadowski, P.G. et al. Nat Protoc (2006) 1 4 1778-1789

Label-Free Quantitation All approaches so far require purchase of isotopically labeled reagents (can be expensive). What if you want to compare large numbers of samples (10+) What if you can t afford lots of reagents? Peak/Spectral counting Peak area comparison (XIC)

Spectral Counting Count the number of peptides identified from a protein in each sample. Do you count repeat identifications of the same peptide? Not accurate at quantifying magnitude of change, but can be used to determine if there is a difference. In general, need a spectral count difference of about 4 peptides in order to be confident of a difference being real. 1 Most proteins in complex mixtures are identified by less than 4 peptides. 1 Old et al. Mol Cell Proteomics (2005) 4 10 1487-1502

XIC (Extracted Ion Chromatogram) Measure intensity of peak during its elution off HPLC column and into the mass spectrometer. Measure area of peak in XIC. More accurate than selecting peak intensity for one given scan.

XIC with Internal Standards When comparing samples for label-free quantitation, how do you adjust for different overall amounts of protein loaded? Add an internal standard to each sample and normalize against this. Adding a protein digest will give multiple internal standards throughout the run. Can normalize to standard eluting at similar time Minimizes effects due to different levels of ion suppression in different samples. Cutillas et al. Mol Cell Proteomics (2007) 6 9 1560-1573

empai (Exponentially Modified Protein Abundance Index) empai = 10 PAI 1 Where PAI = N observed / N observable What is an observable peptide? Peptides with a precursor mass between 800-2400Da. There is a roughly linear relationship between log protein concentration and the ratio of observable peptides observed in range of 3-500 fmoles. If you know how much total protein you analyzed you can derive absolute abundancies. You can observe more peptides than are observable! Ishihama et al. Mol Cell Proteomics (2005) 4 9 1265-1272

MRM (Multiple Reaction Monitoring) Look for a component of a specific mass that when fragmented forms a fragment of another specific mass. Transition: precursor m/z 612.32 fragment m/z 832.45 Very sensitive and specific.

MRM Best performed on a triple quadrupole instrument. Scans are very fast, so can perform multiple transition scans on a chromatographic time-scale. If you know retention times, you can schedule different scans at different times during the LC-MS run. It is possible to analyze a few hundred transitions in a one hour run. If you perform MRM analysis of isotopically labeled (light and heavy) peptides can perform quantitation analysis at very high sensitivity. With a calibration curve of amount vs intensity, can approximate absolute concentration. This is an approach that is likely to be powerful for biomarker analysis. Lange et al. Mol Cell Proteomics (2008) [Epub Apr 13]

Software for MS Quantitation Analysis Different software is required for different quantitation methods. For most quantitation methods, software needs to be able to read instrument manufacturers raw format. Each manufacturer uses own proprietary format Most quantitation software only works for one type of quantitation on one type of instrument, using one search engine. ProQuant (Applied Biosystems): ICAT, cicat and itraq data acquired on their instruments, searched with their own search engine (Protein Pilot) MSQuant: SILAC data acquired from one of three instrument manufacturers, searched using Mascot Protein Prospector: Any type of isotopic labeling acquired on instruments from two manufacturers.

Issues with MS Quantitation Analysis Should you use all data for quantitation? Minimum peak intensity? Peaks near to signal to noise will have much higher variability in quantitation accuracy. Are proteins identified by a single peptide accurately quantified? What if a peptide contains a modification; e.g. oxidized methionine? What if a peptide is observed with and without a missed cleavage? Results are normally reported with a mean and standard deviation. Trinidad et al. Mol Cell Proteomics (2008) 7 4 684-696

Software for Analysis of Label-free Data Quantitation of label-free data is much more difficult from a bioinformatic perspective. When comparing LC-MS runs, elution times are not identical from run-to-run. Mass measurements may vary slightly. Software needs to be able to distort (stretch / squeeze) chromatogram to be able to compare runs.

Comparison of Quantitation Accuracy What level of quantitation accuracy can you expect with these methods? Dynamic Range: Most MS methods have a dynamic range accuracy of only about 1 order of magnitude: e.g., cannot distinguish reliably between 20 fold and 100 fold increase. Problem is usually the lower value in the comparison is too close to the signal to noise level. What is the minimum change that can be reliably measured? For label-free methods: 2 fold change For isotopic labeling methods: 30% change

Quantitate then MS: Gels Coomassie and Sypro stain densities are relatively linear with respect to amount. Silver stain is less quantitative. Complication: How do you know there is only one major protein component in the band. Greater separation with 2D gels reduces chances of multiple components per spot Multiple components may still be present if starting mixture was complex. 2D gels allow quantitation of a specific-modified version of protein (providing you know what modification state the spot corresponds to!) Will separate different splice variants. Quantitative Westerns: specific, but not very accurate quantitation.

Difference Gel Electrophoresis (DIGE) Fluorescent dyes which react with lysine residues. Fluoresce at different wavelengths, so can be differentially visualized. Labeling based on minimal labeling : only 5% of protein is singly labeled. Viswanathan, S. et al. Nat Protocols (2006) 1 1351-1358

Difference Gel Electrophoresis (DIGE) Label alters migration, so majority of protein (unlabeled) does not coelute Cutting out spot for ID is more complicated. Technology sold by GE Healthcare. Can only detect spots using their scanner. Very powerful analysis software available Decyder. Scanner and software cost a small fortune.

Conclusions Although not intrinsically quantitative, mass spectrometry can be used for many types of quantitation. Isotopic labeling is most accurate. Label-free approaches are simpler (and cheaper) and more appropriate for comparison of large numbers of samples, but are less accurate. Most quantitation is relative, but if one component is at a known concentration then absolute quantitation can be performed. Software for analyzing this type of data is still being developed.