A fast, easy and green thermal desorption-gc/ms method for the analysis of phthalate esters in PVC

Similar documents
Pyrolysis Resource CD. Application Booklets. Processed Materials. Application Products. Basic Materials. Functional Materials.

Technical Report Abstract: Keywords: RoHS, phthalate esters, Py-GC/MS, Py-Screener 1. Introduction

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

Rapid Catalyst Screening Reactors

Qualitative Profiling of Co-polymer Polyethylene Terephthalate through Multifunctional Pyrolyzer-GC/MS by various Thermal Treatment Techniques

AN INTEGRATED SYSTEM USING TEMPERATURE BASED SAMPLING FOR POLYMER CHARACTERIZATION

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Analysis of Pressure Sensitive Adhesives by GC/MS and GC/AED with Temperature Programmable Pyrolyzer

Determination of 15 phthalate esters by ultra performance convergence chromatography

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Application Note. Authors. Abstract. Xin Sun Shandong Agriculture University, Taian, , China

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Headspace Technology for GC and GC/MS: Features, benefits & applications

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.

Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Understanding Gas Chromatography

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

STANDARD OPERATING PROCEDURES

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

U.S. EPA Method 8270 for multicomponent analyte determination

Solid Phase Micro Extraction of Flavor Compounds in Beer

Chemistry Gas Chromatography: Separation of Volatile Organics

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS

NCFS Sample Preparation and Instrumental Parameters

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

ANNE JUREK. Abstract: soils and are found. in can also with GC/MS. poster. in series. option for. There are problems. analytes. in methanol.

VINYL CHLORIDE. Recommended by NIOSH, partially evaluated by OSHA Laboratory.

DETERMINATION OF PHTHALATES FROM BOTTLED WATER BY GC-MS

Sulfotepp impurities in Chlorpyrifos EC formulations

Test Method: CPSC-CH-E

Gas Chromatography. Introduction

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

H 3 CO H 3 CO S CH 3

BP1 PONA BPX1. GC Capillary Columns BP1 PONA and BPX1

637. Thiamethoxam. HPLC method

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

BUTYL ALCOHOL in urine by GC-FID Code GC05510

Chapter 27: Gas Chromatography

Appendix II- Bioanalytical Method Development and Validation

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Pittcon P

Maximizing Sample Throughput In Purge And Trap Analysis

Chromatography. Gas Chromatography

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

METHADONE and EDDP in urine by GC-MS Code GC Method of Confirmation by GC-MS

GAS CHROMATOGRAPHY (GC)

Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s

High-Resolution Accurate-Mass (HRAM) Phthalate Screening using Direct Analysis in Real Time (DART) Ambient Ionization

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

TEST REPORT. * Note 2 : This report shall not be reproduced except in full without the written approval of the testing laboratory.

Use of High Speed/High Resolution Size-Based Chromatographic Separation of Polymeric Mixtures with Offline Infrared Detection

Methanol Extraction of high level soil samples by USEPA Method 8260C

Material Characterization in the Automotive Industry Using Multi-Mode Pyrolysis GC/MS:

Gas Chromatography (GC)

The end of. mass-speculation. MS Certified Vials Pre-cleaned and certified vials for mass spectrometry

METHYLETHYLKETONE (M.E.K.) IN URINE BY GC/MS in head space Code GC10010

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

PASS TEST REPORT AT-1500

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Increasing Selectivity and Confidence in Detection when Analyzing Phthalates by LC-MS/MS

Analysis of Plasticizer Contaminants in Beverages and Milk using an Automated System Based on Turbulent-flow Chromatography Coupled to LC-MS/MS

ACETONE IN URINE BY GC/MS Headspace Code GC10110

Application. Gas Chromatography March 1998

METHOD: 1403, Issue 3 EVALUATION: FULL Issue 1: 15 August 1990 Issue 3: 15 March 2003

METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Fast Gas Chromatography using Heated Headspace Gas Auto Sampling Techniques: Ethylene Oxide and 1,4-Dioxane in Fatty Acids

Selection of a Capillary

2,5-DIMETHOXY-4-IODOPHENETHYLAMINE Latest Revision: August 22, 2005

Application Note. Abstract. Authors. Introduction

Analytical Trap Comparison for USEPA Method 8260C

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Volatile Organic Compounds in Every Day Food

GC-CI-MS analysis of TMS derivatives

Determination of Pesticides in Aqueous Samples by On-Line Coupling Solid-Phase Extraction to Gas Chromatography with At-Column Concentrating Interface

5-METHOXY-N,N-DIISOPROPYLTRYPTAMINE Latest Revision: June 20, 2005

Chapter 27: Gas Chromatography. Principles Instrumentation Detectors Columns and Stationary Phases Applications

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD)

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

DIQUAT DIBROMIDE. The Determination of Ethylene Dibromide in Diquat Dibromide and Diquat Dibromide / Paraquat Dichloride SL Formulations

Optimizing. Abstract: is standardd. procedures. altered to

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

Analysis of BTEX in Natural Water with SPME

Final Report. Characterisation of Sample Report. Job No 2016/11/12-34 AS No. 1234A. Client Example Contact Sample. Signed Date 2017.

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

Selection of a Capillary GC Column

TDTS. Keywords. Introduction. TD Operation; Calibration; Impedance; Discrimination

ANALYTICAL REPORT. 25B-NBOH (C17H20BrNO3) 2-({[2-(4-bromo-2,5-dimethoxyphenyl)ethyl]amino}methyl)phenol RSUNJYKZRKIBNB-UHFFFAOYSA-N

Transcription:

A fast, easy and green thermal desorption-gc/ms method for the analysis of phthalate esters in PVC A discussion of the central factors that influence data quality when using ASTM D7823 (1) for the determination of phthalates in polymeric substrates D. Randle, R. Freeman, A. Hosaka and C. Watanabe Frontier Laboratories Ltd. 1. Introduction Phthalates, which are esters of phthalic acid, have been used in the manufacture of a wide range of consumer products. They are added to plastics to make them more flexible and harder to break. Although the wide-spread use of phthalates in the manufacture of polymeric products has been largely discontinued, they continue to pose a risk to human health. Phthalates are easily released into the environment because they are not covalently bonded to the polymer. Consequently, exposure can be through direct contact, food, water and the atmosphere. Several phthalates are tightly regulated on a global scale. The United States (2), Canada, the European Union, Japan and many other nations have banned the use of and passed regulations designed to monitor for the concentration of phthalates in consumer products. In the US, Congress has permanently banned three phthalates (, DBP, BBP) in any amount greater than.1 percent (per plasticized component part of a children s toy or child care article). There is also an interim ban on DINP, and DNOP, which only applies to children's toys that can be placed in a child's mouth (3,4). Although a number of different analytical techniques can be used to determine the presence of phthalates, most laboratories use wet chemical techniques to extract the phthalates from the polymer matrix. The extract is then analyzed using GC or HPLC chromatographic methods to separate and quantitate the individual phthalates. "Traditional" sample preparation is, at best, cumbersome, time consuming, costly and does not always completely isolate the phthalates from the polymer matrix. Thermal desorption (TD) is a viable alternative to traditional solvent-based sample preparation. Thermal desorption is based on the fact that there is free volume within the polymer structure through which small molecules (e.g., phthalates) are free to move. The motion is due to intermolecular collisions (i.e., Brownian motion). Random motion from regions of high concentration to regions of lower concentration is called diffusion. Factors affecting the rate of diffusion are the solubility and diffusivity of the small molecules in the polymer. Increasing the temperature of a polymer causes the small molecules on the surface to vaporize and the concentration gradient across the polymer to increase. In addition, the permeability of the polymer increases with temperature. These two factors result an even greater flux of small molecules to the surface. The phthalate vapors are analyzed using GC/MS. Phthalates are identified using both retention data and compound specific ions. The entire process can be automated. 1

2. Overview of ASTM D7823 ASTM D7823 describes the quantitative analysis of six regulated phthalates in polyvinylchloride (PVC). This analytical method uses thermal desorption (TD) in which phthalates in PVC are thermally desorbed and introduced into GC/MS for separation and analysis. In this method, the sample preparation is simple and easy and the amount of organic solvent required is as small as 1 ml for a single sample. Also, the standard addition used in this method provides quantitative values with the highest accuracy and the results are virtually not influenced even by high concentrations of non-regulated plasticizers contained up to several tens of percents. Although ASTM D7823 is written specifically for the six regulated phthalates in PVC; it can easily be extended to include a variety of other phthalates in a number of different polymeric matrices. 3. Sample preparation ASTM D7823 is based on the thermal desorption (thermal extraction ) of the phthalates from a polymeric substrate. The method specifically calls for the dissolution of the polymer in a suitable solvent such as THF; however, subsequent studies have shown that meaningful data can also be obtained when the polymer is analyzed directly. Analysis of the solid is often referred to as a direct method. Analysis of a microliter aliquot of a quantitative solution of the sample is referred to as a thin film method. The entire sample preparation protocol is summarized in Figure 1. Each step is described in the ASTM method. Thin film method Solvent (ml) Sample (µl) Evaporate Sample Crush & Powder (cryo-mill) mg Direct method (2 μg) mg/ml Sample cup TD-GC/MS analysis Sample cup Figure 1. Quick, simple and green sample preparation. The thin film method requires that a quantitative solution of the sample be prepared. An aliquot of the THF solution is placed in the sample cup. Comment 1: The accuracy of the measurement is highly dependent upon three factors: (1) the homogeneity of the portion of the sample selected for analysis, (2) the accuracy of the analytical balance and (3) the accuracy and/or precision of the device (e.g., syringe) used to transfer the dissolved sample into the sample cups. Sample homogeneity is accomplished using a variety of methods including cryo-milling, grinding and sieving, averaging the results from the analysis of multiple samples and/or dissolving the entire sample in a known volume of solvent. Regardless of how the sample is homogenized, microgram quantities are actually analyzed. The ability of the balance to accurately weigh microgram quantities will directly affect the accuracy of the phthalate determination. The analytical precision obtained using the thin film and direct sampling methods is compared in Figure 2. The thin film method is an aliquot of a solution that is, by definition, homogeneous, whereas the direct method is a random sampling of particles, which will inherently lack sample-to-sample reproducibility. 2

Thin film method Direct method : %RSD = 1.15 (n=4) DINCH : %RSD = 4.79 (n=4) DINCH 1 15 2 min 1 15 2 min Figure 2. Influence of the sample preparation method on analytical precision. Sample:.5 mg plastic toy (PVC & DINCH); Thin film: 2 µl THF solution. Direct sampling: powder < 45 mesh. 4. Determination of the EGA thermal desorption zone Evolved gas analysis (EGA) MS (5,6) is used to define the thermal desorption zone for the phthalates. This is a simple method that utilizes the temperature programmability of the Frontier EGA/PY-33D (or 22iD) pyrolyzer. A deactivated stainless steel tube (2.5 m x.15 mm i.d.) connects the injection port to the MS. The sample is heated from a low temperature to a high temperature. As the temperature increases, constituents in the sample vaporize, pass through the tube and are detected by the MS. A plot of sample temperature vs. detector signal is referred to as an EGA thermogram. The EGA thermogram of a polyvinyl chloride sample is presented in Figure 3. Extracted ion chromatograms are used to delineate the thermal zone over which the phthalates evolve. Ion 149, which is common to all of the phthalates of interest, is used to identify the phthalate envelope: 1 to 32ºC. In this example, DOA (Di(2-ethylhexyl)adipate) which is a common plasticizer used in the manufacturing of flexible plastic toys, medical devices and food packaging desorbs along with the phthalates. In order to identify and quantitate the phthalates in the sample, it is necessary to either chromatographically separate the DOA isomers from the phthalates of interest or use ions that are present in the phthalate mass spectra and not found in the DOA spectra. DOA / Phthalates PVC 129 57 147 83 111 185 241 259 341 1 2 3 m/z 5 36 DOA / HCL 55 78 129149 1 2 m/z 91 115 165 141 67 Aromatics 1 2 3 m/z 1 2 4 6 8 1 12 14 16 18 2 22 24 26 28 3 min 2 3 4 5 6 7ºC DOA (Ion 129) 2 4 6 8 1 12 14 16 18 2 22 24 26 28 HCL (Ion 36) 2 4 6 8 1 12 14 16 18 2 22 24 26 28 Phthalates (Ion 149) 2 4 6 8 1 12 14 16 18 2 22 24 26 28 Thermal desorption zone Figure 3. Using extracted ion chromatograms (EIC) to determine the phthalate thermal desorption zone. Sample: PVC + DOA. 3

5. Advantages of thermal desorption over conventional method (solvent extraction) Virtually all analytical methods consist of four steps: sample preparation, method calibration, sample analysis and data processing. Most of the actual sample manipulation occurs during sample preparation and it is during this process that there is the highest probability of introducing error into the determination. Potential sources of errors include: (1) processing a non-representative portion of the sample. (2) sample contamination (e.g., glassware, reagents or solvents) and (3) deviations from the procedure prescribed in the method. Also, it is important to note that sample preparation also represents most of the costs associated with a given analytical procedure. Ideally, the sample should be analyzed as is. Direct analysis of the sample increases laboratory productivity and improves the overall data quality. One of the most important advantages that thermal desorption has over samples prepared using conventional solvent-based techniques is that only that portion of the sample that is desorbed is transported to the column. At 32ºC, the highest thermal desorption temperature prescribed in ASTM D7823, any of the polymers is virtually not decomposed and the polymer remains in the sample cup and is therefore not introduced into the GC inlet, column, or MS. Because only the volatile fraction of the sample is injected, there is no need for a post-run bake out, more samples can be analyzed per unit time and sample-to-sample carry over is virtually non-existent. This has a major positive impact on both data quality and laboratory productivity. 4

6. Analysis In practice, the samples are prepared as previously discussed. If the samples are analyzed directly, approximately 1 µg are placed in a sample cup. If the samples will be analyzed using the thin film method, 1-2 µl of the quantitative solution are added into a sample cup. The solvent is evaporated leaving a thin film of the sample on the cup surface. TD-GC/MS combines the sample preparation (i.e., TD) with GC/MS analysis. This work was done on using a Frontier Lab EGA/PY-33D or 22iD multi-functional pyrolyzer with a 48 position autosampler (Frontier Auto-Shot AS-12E) interfaced via a split/splitless injection port to a GC/MS. The cup is dropped into the furnace, which is at 1ºC. The temperature increases at 2ºC/min until it reaches 32ºC. The initial column temperature is 8ºC, which is low enough to facilitate the condensation of the desorbed vapors into a narrow band at the head of the column. A typical total ion chromatogram (TIC) and the extracted ion chromatograms (EIC) for the six regulated phthalates are shown in Figure 4. The analytical precision (n=6) obtained for each phthalate is shown in Table I. 5 DINCH 25 1 4 2 4 2 3 2 DBP BBP DNOP DINP 7 8 9 1 11 12 13 14 15 min DINCH (Ion 155) 7 8 9 1 11 12 13 14 Phthalates (Ion 149) 7 8 9 1 11 12 13 14 DBP (Ion 223) 7 8 9 1 11 12 13 14 BBP (Ion 26) 7 8 9 1 11 12 13 14, DNOP (Ion 279) 7 8 9 1 11 12 13 14 DINP (Ion 293) 7 8 9 1 11 12 13 14 (Ion 37) 7 8 9 1 11 12 13 14 Figure 4. TD-GC/MS analysis of a PVC-DINCH polymer spiked with the six regulated phthalates. The phthalates are identified based upon their relative retention time and the presence of the phthalatespecific quant ion. The thin film sample preparation as described in D7823 was used for this analysis. Phthalate Quant ion DBP m/z=223 BBP m/z=26 m/z=279 DNOP m/z=279 DINP m/z=293 m/z=37 Replicates (n=6) (n=6) (n=6) (n=6) (n=6) (n=6) % RSD 1.35 1.46 2.54 1.6 1.33 2.52 Table I. Analytical precision based upon the area of the phthalate-specific quant ion for six replicates (n=6) of a PVC/DINCH sample. 5

7. Instrument and analytical conditions Frontier Laboratories EGA/PY-33D or 22iD Pyrolyzer; thermal desorption zone: 1 to 32ºC at 2ºC/min. PY interface: 32ºC (Auto mode); GC injector: 3ºC, GC oven: 8ºC (1 min) 2ºC (5ºC/min) 35ºC (15ºC/min, 2 min hold); Solvent delay: 6 min.; Column: UA-5 (5% Diphenyl- 95%dimethyl polysiloxane) 3m,.25mm i.d,.25 µm film); Column He flow: 1.2 ml/min; Split ratio: 1/2; Mass range: 29-6 m/z; Scan speed: 2.57 scans/sec; Threshold: 5; MS transfer line Temp.: 3ºC; Ion Source (EI) Temp.: 23ºC; Sample wt.: ca. 2 μg. 8. Calibration Of the four calibration methods most often used for quantitation, standard addition is the method of choice. Polymeric products are often very complex blends of additives, co-polymers and oligomers. Many different chemical species can be desorbed along with the phthalates. For example, Figure 4 shows that DOA desorbs with the phthalates. Other common plasticizers, such as Mesamoll II (Alkylsulphonic acid phenyl esters (ASE)) and DINCH (1,2-Cyclohexane dicarboxylic acid di-isononyl ester), also desorb with the phthalates. The potential for interference is significant. Another issue that must be considered is that it is not unusual for a GC/MS to struggle with long term contamination, most of which is inconsequential to the analysis being performed. For example, m/z 27 is a common polysiloxane associated with modern day cross-linked stationary phases. m/z 27 is not an ion that is associated with any of the phthalates on interest; however, very often the system contamination is, in fact a phthalate; standard addition will help mitigate the error associated with the phthalate background. Standard addition is the best calibration protocol when interference is an issue. The calibration is done using the sample matrix in the same time frame as the sample analysis. Interferences and background contamination will be constant for all analyses and its effect on the calculated concentration is negated. Peak area DBP BBP DNOP DINP 1,8, 1,6, 1,4, 1,2, 1,, 8, 6, 4, y = 3169.3x + 61158 R 2 =.9995 y = 2765.2x + 5289 R 2 =.9999 y = 2694.9x + 465186 R 2 =.9999 y = 24.9x + 43726 R 2 =.9996 y = 1941.9x + 358811 R 2 =.9997 y = 1293.9x + 235288 R 2 =.9979 2, -3-2 -1 1 2 3 1,5 1, 5 5 1, 1,5 (ppm) Figure 5. A typical set of single point standard addition calibration curves for the six regulated phthalates in PVC - DOA. The number of calibration points can be increased as specified in the laboratory s data quality objectives (DQO) documents or a project specific Work Plan. 6

PVC DOA Quant ion DBP m/z=223 BBP m/z=26 m/z=279 DNOP m/z=279 DINP m/z=293 m/z=37 ng 181.8 181.8 184.8 172.6 182.1 189.7 ppm 99 99 924 863 911 948 %RSD (n=6) 2.8.76 1.42.85 2.3 2.7 Table II. Calculated amounts of the six regulated phthalates determined using the standard addition calibration method. Example calculation: Equation of the DBP calibration curve is: y = mx + b y = 1293.9x + 235288 (R 2 =.9979). When y= then x = -235288/1293.9 = -181.8 181.8 ng. Concentration in ppm is given by : Value of DBP Assumed content of PVC-DOA = 181.8 (ng)/2 (μg) =.99 (ng/μg) = 99 (ng/mg) = 99 (ppm). Comment 2: A list of the most commonly encountered phthalates is shown in Table III The regulated and monitored phthalates are in red. Note that both and DINP are technical mixtures because of the isomeric nature of the alcohols used to prepare the phthalates. The peaks in the mixtures need to be integrated and summed together. Comment 3: Suitable in-matrix calibration standards are difficult to find from a commercial provider. They must be prepared by the laboratory. Making a quantitative mixture of primary phthalate commercial standards in a suitable solvent is straightforward. However, if microliters of the standard are injected into a sample cup, which is then analyzed using TD, the precision for the most volatile phthalates will be very poor. Because they are dispersed on the cup surface in a thin film, losses due to evaporation are significant. In addition, the phthalate standards should be in-matrix in order to more accurately represent what truly happens during the desorption process. This is another benefit of using standard addition where the standards are spiked into the actual sample prior to analysis. 7

9. Extending ASTM D7823 Although ASTM D7823 is written specifically for the six regulated phthalates in PVC; it can easily be extended to include a larger number of phthalates in a number of different polymeric matrices. Figure 6 shows EGA thermograms of common polymers containing phthalates. It is shown that for any of the polymers, phthalates are desorbed in similar temperature zone. ASTM D7823 prescribes only for PVC, however, the results shown in Figure 6 suggest that this method can used for a number of other common polymeric matrices. Various polymers Phthalates Polycarbonate; (PC) Polystyrene; (PS) CH 2 CH(C 6 H 5 ) n Figure 6. Phthalate thermal desorption zone for four different polymers. Polyethylene (PE) CH 2 CH 2 n Poly (vinyl chloride) ; (PVC) CH 2 CHCl n 2 3 4 5 6 7 8ºC Thermal desorption zone Figure 6 shows the EGA thermograms of four different polymers. The phthalate thermal desorption zone is the same for all four polymers: 1 32ºC. Clearly, it is not necessary to perform an EGA on each batch of samples. Samples arrive in the laboratory essentially ready for analysis. The instrument configuration and GC method will be the same for a wide variety of sample types. 8

There is near universal interest in monitoring phthalates in a number of consumer products. Many private sector companies, in anticipation of regulations directed at a number of common phthalates, have implemented in-house phthalate monitoring programs that are able to report as many as 15 phthalates see Table III. RT Name Acronym CAS No. 7.85 Dimethyl phthalate DMP 131-11-3 8.7 Diethyl Phthalate DEP 84-66-2 9.915 Di-n-propyl phthalate DRPR 131-16-8 1.6 Diisobutyl phthalate DIBP 84-69-5 11.34 Dibutyl phthalate DBP 84-74-2 12.957 Di-n-pentyl phthalate DPP 131-18- 14.686 Di-n-hexyl phthalate DHP 84-75-3 14.778 Benzyl butyl phthalate BBP 85-68-7 15+ Diisoheptyl phthalate DIHP 71888-89-6* 16.328 Dicyclohexyl phthalate DCHP 84-61-7 16.59 Bis(2-ethylhexyl) phthalate 117-81-7 17+ Diisooctyl phthalate DIOP 27554-26-3 * 18.56 Di-n-octyl phthalate DNOP 117-84- 18 + Diisononyl phthalate DINP 68515-48- * 19 + Diisodecyl phthalate 26761-49-1* Table III. These 15 phthalates have been determined using ASTM D7823. The six regulated phthalates are highlighted in red. (* DIHP, DIOP, and DINP are technical mixtures. Care should be taken to ensure that the correct standards are used). Also note that these mixtures will have multiple peaks that need to be integrated and summed together. The column temperature program specified in D7823 is modified to ensure resolution of the extended list. GC oven: 4 2ºC (at 4ºC/min), 2 3ºC (at 5ºC/min, 1 min hold), 3 32ºC (at 2ºC/min, 2.5 min hold). Retention times are provided to give the reader a sense of the relative retention of each phthalate using a 5% phenyl/methyl stationary phase column. 9

1. Phthalates identified in a power cord sheath using ASTM D7823 (modified) The outer PVC material of a power cord was analyzed for an extended list of phthalates using ASTM D7823. Other than the change in the initial column temperature, the samples were analyzed as described in D7823. Commercial standard mixtures of 15 phthalates in methylene chloride (1-3 µg/ml) were used to prepare the standard addition solution. The standard addition method per ASTM D7823 was used. The extracted ion chromatogram (m/z 149) of the sample is shown in Figure 7. Obviously, there are a number of phthalates present in the sample. However, it is also clear that the level of exceeds the linear range of the detector and the sample will have to be diluted and reanalyzed. Concentrations as high as this always raise concerns about carryover and strongly suggest that a series of blank runs be made, in order to ensure that there is no carryover, before proceeding with sample analysis. 15 1 5 BHT (origin inate from THF solvent) DEP DIBP DB BP Unknow wn phthalate Unknown phth halate Unknown phth halate Isomer of 2 4 6 8 1 12 14 16 18 2 22 24 26 Figure 7. Extracted ion chromatogram (m/z=149) of the power cord sample (2 µg in sample cup). 28(min) Analytical method: Thermal desorption temperature: 1-32ºC (2ºC/min) GC Injection port: 3ºC GC Oven: 4 2ºC (4ºC/min) 3ºC (5ºC/min, 1 min hold) 32ºC (2ºC/min, 2.5 min hold) Column: Ultra ALLOY-5 (MS/HT) (5% diphenyl 95% dimethylpolysiloxane, L=3m, i.d.=.25mm, df.=.25µm) Column Flow: 1.2 ml/min Split ratio: 1/2 Scan speed: 4 scans/sec Sample weight: 2.1 & 2 mg dissolved in 1 ml THF ; 2.1-2 µg in sample cup 1

The concentration of is determined by a re-analysis of the sample in which the amount of sample is reduced from 2 to 2.1 µg, or a hundred-fold decrease in the amount of sample analyzed. Figure 8 shows the total ion chromatogram and extracted ion chromatogram (m/z=279) used to quantitate. Note this actual sample contained a whopping 17.13%! 5 3 TIC 1 9.2 9.4 9.6 9.8 1. 1.2 1.4 1.6 1.8 11. 11.2 11.4 11.6 11.8 2 1 EIC m/z 279 amount in sample: 17.13% 9.2 9.4 9.6 9.8 1. 1.2 1.4 1.6 1.8 11. 11.2 11.4 11.6 11.8 Standard addition Sample Name Sample amount (µg) Standard addition (ng) Area PVC Sample 2.15 35993 PVC Sample + spike 2.15 9 1535796 Quant value (%) 17.13 % Figure 8. Total ion chromatogram and extracted ion chromatogram (m/z=279) of 2.15 µg of the power cord. GC oven profile: 4 18ºC (at 4ºC/min) - 28ºC (at 2ºC/min) 32ºC (at 1ºC/min, 2.5 min hold) to 32ºC. was calculated to be 17.13% using the two point standard addition calibration method. Comment 4: Samples, such as this power cord sheath, can have an immediate and potentially disastrous effect on the instrumentation. High level system contamination is common and will requires time and effort to eliminate. If possible, unknown samples should be screened using Infrared (IR) Spectroscopy (7). At a minimum, this will flag the sample and it can be initially analyzed at a very high dilution like was done with the power cord sheath. 11

Comment 5: Although MDL studies yield values for several phthalates in the low ppm range, it may be wise to use the practical quantitation limit (PQL). Other laboratories have selected a reporting limit that be achieved for all targeted phthalates. A value of 5-1 ppm is common. This takes into account changes in the instrument performance and varying levels of contamination that occur as the number of samples analyzed increases. The impact of these changes is somewhat mitigated by using standard addition. Phthalates Identified DEP DIBP DBP DINP Phthalate amount 22 ppm 24 ppm 76 ppm 17.13 % 14 ppm 39 ppm Sample amount analyzed (µg in sample cup) 2 µg 2 µg 2 µg 2.1 ug 2 µg 2 µg Table IV. Quantitative results for each phthalate found in the power cord sheath. 11. Comparing the TD-GC/MS method with the traditional liquid extraction-gc/ms method Figure 9 summarizes the differences in data quality (accuracy and precision) between the thermal desorption/standard addition and solvent extraction/external standard methods. Wet chemical methods often result in low values for the concentration of the volatile phthalates; These can be attributed to losses during sample preparation. High concentration values for DNOP, DINP and are due to interference from the co-eluting DINCH. The TD/standard addition method gives an almost constant response for all six phthalates with an average precision of 1.8 (%RSD). (%).2 Solvent-extraction & ESTD (8.3) (1.7) 1.3e+7 DBP BBP DINCH 6, m/z 279, DNOP.1 (1.4) (13.2) (1.5) (9.3) (2.5) (1.4) DBP BBP DNOP DINP (11.8) (1.6) (1.3) (2.5) TD & Standard Addition 6, m/z 293 DINP 4, m/z 37 7 8 9 1 11 12 13 14 (min) Figure 9. Quantitation and precision using solvent extraction and external standard calibration and thermal desorption and standard addition calibration. %RSD (n=4) values are shown in parenthesis. Phthalate concentrations = 1,25 ppm 12

12. SUMMARY ASTM D7823 is based on thermal desorption (TD) GC/MS. The method is suitable for an extended list of phthalates in a wide range of polymeric materials. The entire analytical sequence is performed in a single step: a known weight of sample (i.e., a known weight of solid sample or a known volume of a quantitative solution is placed in a sample cup which in turn is placed in the auto sampler carousel. Thermal desorption eliminates conventional sample prep regimes; the sample is heated to a point that the phthalates desorb from the polymer. It is fast, uses a minimal volume of solvent, and eliminates the need for expensive glassware. Calibration is based on the areas of specific quant ions for each phthalate which virtually eliminates the interference that may occur from co-eluting non-target chemicals. Quantitation is done using standard addition which takes into account any impact the sample matrix may have on the process. It also ensures that the calibration and sample analysis are performed under identical instrumental conditions. Standard addition does require that each sample be analyzed at least twice, but the benefits far out weigh the investment in the second analysis, especially given the fact that the entire process is automated. 13. References 1. ASTM D7823 (213) Standard Test Method for the Determination of Low Level, Regulated Phthalates in Poly (Vinyl Chloride) Plastics by Thermal Desorption - Gas Chromatography/Mass Chromatography. 2. Section 18: Products Containing Certain Phthalates. http://www.cpsc.gov/about/cpsia/faq/18faq.html 3. Di-(2-ethylhexyl) phthalate (), dibutyl phthalate (DBP), and benzyl butyl phthalate (BBP). Diisononyl phthalate (DINP), diisodecyl phthalate (), and di-n-octyl phthalate (DNOP). 4. Consumer product safety commission, Test Method: CPSC-CH-C11-9.3 - Standard Operating Procedure for Determination of Phthalates, April 1, 21. 5. Applications of a multifunctional pyrolyzer for evolved gas analysis and pyrolysis-gc of various synthetic and natural materials: part 2, S. Tsuge, H. Ohtani, Chu Watanabe and Y. Kawahara, American Laboratory, March, 23. 6. Rapid and Simple Determination in Plastic toys by a Thermal Desorption-GC/MS Method, T. Yuzawa, Chu Watanabe, R. Freeman and S. Tsuge, Anal. Sci., September, 29, vol. 25. 7. ASTM D2124-99 (24) 5, Standard Test Method for Analysis of Components in Poly(Vinyl Chloride) Compounds Using an Infrared Spectrophotometric Technique. 13