Direct Analysis in Real Time (DART) A New Ionization Technique

Similar documents
M M e M M H M M H. Ion Sources

Chemistry Instrumental Analysis Lecture 37. Chem 4631

Mass Spectrometry. Electron Ionization and Chemical Ionization

1) In what pressure range are mass spectrometers normally operated?

Chemistry Instrumental Analysis Lecture 34. Chem 4631

20.2 Ion Sources. ions electrospray uses evaporation of a charged liquid stream to transfer high molecular mass compounds into the gas phase as MH n

The AccuTOF -DART 4G: The Ambient Ionization Toolbox

Selecting an LC/MS Interface Becky Wittrig, Ph.D.

Lecture 15: Introduction to mass spectrometry-i

TANDEM MASS SPECTROSCOPY

Chapter 1. Introduction

Lecture 8: Mass Spectrometry

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics

Lecture 8: Mass Spectrometry

Mass Spectrometry (MS)

Ionization Techniques Part IV

Chemistry 311: Topic 3 - Mass Spectrometry

Molecular Mass Spectrometry

MS Goals and Applications. MS Goals and Applications

LC-MS Based Metabolomics

Mass Spectrometry. General Principles

Chemical Aspects of Mass Spectrometry

Other Methods for Generating Ions 1. MALDI matrix assisted laser desorption ionization MS 2. Spray ionization techniques 3. Fast atom bombardment 4.

Mass spectrometry and elemental analysis

Mass Spectrometry for Chemists and Biochemists

Harris: Quantitative Chemical Analysis, Eight Edition

Mass Spectrometry in MCAL

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai

LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment

Analysis of Polar Metabolites using Mass Spectrometry

Molecular Mass Spectrometry

CHROMATOGRAPHY AND MASS SPECTROMETER

MS Goals and Applications. MS Goals and Applications

1. The range of frequencies that a measurement is sensitive to is called the frequency

CEE 772: Instrumental Methods in Environmental Analysis

for the Novice Mass Spectrometry (^>, John Greaves and John Roboz yc**' CRC Press J Taylor & Francis Group Boca Raton London New York

Ionization Methods in Mass Spectrometry at the SCS Mass Spectrometry Laboratory

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee

IDENTIFICATION OF ORGANOMETALLIC COMPOUNDS USING FIELD DESORPTION IONIZATION ON THE GCT

Desorption sonic spray ionization for (high) voltage-free ambient mass spectrometry

Auxiliary Techniques Soft ionization methods

MASS SPECTRA measure a compound s Mol. Wt. This ionization type is called: electron impact MS

Presentation Basic Introduction to Instrumentation Matrix Effects Challenges

CHROMATOGRAPHY. The term "chromatography" is derived from the original use of this method for separating yellow and green plant pigments.

Properties of Compounds

5072 CHEMISTRY (NEW PAPERS WITH SPA) TOPIC 1: EXPERIMENTAL CHEMISTRY 5067 CHEMISTRY (NEW PAPERS WITH PRACTICAL EXAM) TOPIC 1: EXPERIMENTAL CHEMISTRY

Mass spectrometry.

Ion sources. Ionization and desorption methods

A process whereby an electron is either removed from or added to the atom or molecule producing an ion in its ground state.

Agilent G3212 GC-APCI Source

Chapter 6. Neutral droplets in high electric fields as a. source of ions. Introduction to field-induced droplet

4. How can fragmentation be useful in identifying compounds? Permits identification of branching not observed in soft ionization.

Dry Deposition of Biogenic Terpenes via Cationic. Oligomerization on Environmental Aqueous Surfaces

MASS SPECTROMETRY. Topics

Introduction. Chapter 1. Learning Objectives

Ch 13. Basics of Mass Spectrometry (I) : Principles & Ionization Sources

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS

Mass Spectroscopy. Base peak. Molecular Ion peak. The positively charged fragments produced are separated, based on their mass/charge (m/z) ratio. M+.

Introduction to LC-MS

KJ3022 MS compendium gives a deeper explanation of what is mentioned in the slides

Electrospray Ion Trap Mass Spectrometry. Introduction

Mass Spectrometry Course

Introduction to Pharmaceutical Chemical Analysis

Chapter 14. Molar Mass Distribution.

10/16/17 ACIDS AND BASES, DEFINED WATER IS AMPHOTERIC OUTLINE. 9.1 Properties of Acids and Bases. 9.2 ph. 9.3 Buffers

Introduction of Hitachi Chromaster 5610 MS Detector for High Performance Liquid Chromatograph

(Refer Slide Time 00:09) (Refer Slide Time 00:13)

MS/MS .LQGVRI0606([SHULPHQWV

Chapter 8. Online field-induced droplet ionization. mass spectrometry probes heterogeneous reaction

Mass Spectrometry. A truly interdisciplinary and versatile analytical method

Selecting Detectors for Compounds with No Optical Absorbance

Mass Spectrometry. Introduction EI-MS and CI-MS Molecular mass & formulas Principles of fragmentation Fragmentation patterns Isotopic effects

Ultra-fast screening of pesticides in foods and agricultural products with desorption corona beam ionization (DCBI) tandem mass spectrometry

Mass spectrometry gas phase transfer and instrumentation

Introduction to the Q Trap LC/MS/MS System

Supporting Information. Use of Mass Spectrometric Vapor Analysis to Improve Canine Explosive Detection Efficiency

Basics of Mass Spectrometry

Planar chromatography meets direct ambient mass spectrometry: current trends

MS Interpretation I. Identification of the Molecular Ion

Saba Al Fayoumi. Tamer Barakat. Dr. Mamoun Ahram + Dr. Diala Abu-Hassan

SEPARATION TECHNIQUES

National Institute of Standards and Technology, Materials Measurement Science Division, Gaithersburg, MD,USA. *

Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass Spectrometry

FAIMS Technology at a Glance

José J. Pérez 1, Glenn A. Harris 1, Joseph E. Chipuk 2, Jennifer S. Brodbelt 2, Michael D. Green 3, Christina Y. Hampton 1, Facundo M.

sample was a solution that was evaporated in the spectrometer (such as with ESI-MS) ions such as H +, Na +, K +, or NH 4

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA

Chapter 5 Basic Mass Spectrometry

Extrel Application Note

Quadrupole Mass Spectrometry Concepts. Mass spectrometers for residual gas analysis: Intermediate Level Users Guide

Chapter 5. Complexation of Tholins by 18-crown-6:

Chapter 3 Engineering Science for Microsystems Design and Fabrication

Ion-mobility mass spectrometry of polyoxometalate Keplerate clusters and their supramolecular assemblies

INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY

Field-Free Atmospheric Ionization: Atmospheric Pressure Chemical Ionization & Electrospray

API Optimization, Source Design, and Other Considerations of LC/MS

Welcome!! Chemistry 328N Organic Chemistry for Chemical Engineers. Professor: Grant Willson

CHAPTER A2 LASER DESORPTION IONIZATION AND MALDI

Olumide Adebolu. Chromatographic Fidelity and Matrix /Analyte Solubility in Complex Polymer Systems using HPLC-MALD/I TOF MS

Transcription:

Direct Analysis in Real Time (DART) A New Ionization Technique Introduction K. P. Madhusudanan Sophisticated Analytical Instrument Facility, Central Drug Research Institute, Lucknow, India - 226001 Direct analysis without sample preparation is an ultimate goal in analytical chemistry. Sample introduction has been one of the limitations of organic mass spectrometry right from the beginning. The sample had to be vaporized and introduced into the ion source maintained at high vacuum for recording electron ionization (EI) mass spectrum. In order to overcome the harsh conditions of electron ionization, another technique called chemical ionization (CI) was developed so as to get molecular weight information from labile molecules. Thermally unstable and polar molecules required field desorption (FD) technique to get ionized intact. The development of fast atom bombardment (FAB) in the early eighties paved the way for the analysis of polar and high molecular weight biomolecles. Electrospray ionization (ESI) and matrix assisted laser desorption ionization (MALDI) revolutionized mass spectrometry by enabling it to analyse very high molecular weight biomolecules. The atmospheric pressure ionization techniques, atmospheric pressure chemical ionization (APCI) and ESI covered the entire spectrum of medium polar and low molecular weight compounds to highly polar and high molecular weight biomolecules. One of the limitations in achieving high sample throughput with ESI or MALDI is the need to dissolve, extract and/or filter the sample prior to analysis. In other words, there was still the requirement of some amount of sample preparation prior to analysis. This included adequate solubility in the desired solvents and freedom from matrices such as salts and buffers. The mass spectrometrist was still looking for a technique which enabled him to present the sample in front of the mass spectrometer for analysis. Two such techniques were developed and announced in 2004 and 2005. These were the desorption electrospray ionization (DESI) and direct analysis in real time (DART). Desorption Electrospray Ionization (DESI) DESI was developed by Cooks and coworkers 1 and uses an aqueous or methanol spray directed at an insulating sample or an analyte deposited on PTFE (Fig. 1). A mixture of liquid similar to that used in liquid chromatograph mobile phases (water-methanol) and controlled by the operating parameters similar to those for ESI conditions is used for the spray. DESI appears to be gentle enough for desorbing ions directly from animal tissue, plant tissue and biological materials. The desorbed ions can be analyzed by a commercial atmospheric pressure ionization mass spectrometer. The working species in DESI appears to be the charged droplets or the desolvated solvent ions generated in the electropray process. DESI generates secondary ions directly from the surface of the material of interest. However, the spray of solvents makes the technique a bit messy.

Solvent Nitrogen MS inlet Spray capillary Gas jet Desorbed ions Spray? Surface Direct Analysis in Real Time (DART) Sample Moving sample stage in air Fig.1 Schematic of DESI source Direct analysis in real time source was developed by Cody and Laramee. 2 DART is a non-contact analysis technique for MS at atmospheric pressure. DART can analyse gases, liquids, solids and materials on surfaces. DART is based on the atmospheric pressure interactions of long-lived electronic excited state atoms or vibronic excited state molecules with atmospheric gases and the sample. Helium or nitrogen is used for producing the ionizing plasma. The gas flows through a chamber where an electrical discharge (corona discharge) produces ions, electrons and excited state (metastable) atoms and molcules. These metastable species, in turn, generate protonated gaseous water clusters in ambient air. The water clusters on proton exchange with the sample give rise to [M+H] + ions of samples. In the negative ion mode oxygen water clusters are reported to be responsible for ionizing the sample. 2 A schematic of the DART source is given in Fig. 2. Nitrogen or helium gas is introduced into a discharge chamber in which a needle electrode is held at a potential of 2-4 kilovolts. The resulting discharge generates a plasma of ions, electrons, and excited-state species. The adjacent chamber has two perforated electrodes, which are biased to remove ions. The neutral gas molecules, including metastable species, pass through a heated chamber. The heated gas exits the source through a grid electrode, which prevents ion-ion and ion-electron recombination and acts as a source of electrons by surface Penning ionization. In addition, it acts as an electrode to promote ion drift towards the mass spectrometer atmospheric pressure inlet interface. The DART gas flow is towards the mass spectrometer inlet orifice across a gap of 5-25 mm. This gap is open to the laboratory room and the material to be analysed is held in this gap in the stream of warm gas.

Needle Electrode 2-4 kv Perforated disk electrodes Grid electrode Sample Gas in MS Inlet Gas Gas out Heater (Ambient to 350 C) Fig. 2 Schematic of DART source and the sampling gap When helium is used the mechanism of ionization involves predominantly the formation of ionized water clusters followed by proton transfer reactions. The helium 2 3 S state has an energy of 19.8 ev and reacts with water efficiently with an estimated reaction cross section of 100 Å 2. He(2 3 S) + H 2 O H 2 O + He (1 1 S) + e - H 2 O + H 2 O H 3 O + OH H 3 O + nh 2 O [(H 2 O)nH] + [(H 2 O) n H] + + M MH + + nh 2 O Electrons produced by Penning ionization or surface Penning ionization are rapidly thermalised by collisions with atmospheric pressure gases and undergo electron capture by atmospheric oxygen producing O 2 - which reacts with the analyte to generate sample anions. M * + surface M + surface + e - (Penning / surface Penning ionization) e - fast + gas e - slow (Thermalised by collision with gas) e - slow + O 2 O 2 - (Electron capture by atmospheric oxygen) Ion formation under DART may also involve other reactions. For example, dopants such as ammonium (from ammonium hydroxide headspace vapour) or chloride (from methylene chloride) ions can modify the chemistry of ion formation. The mechanism involved in desorption of materials from the surfaces by DART is not well understood. It is

postulated that transfer of energy to the surface by metastable atoms and molecules leads to sample desorption and ionization. 2 The mass spectra produced by DART are characterized by M + and/or [M+H] + in positive ion mode and M - or [M-H] - in negative ion mode. Fragment ions are also observed depending on the sample and the source conditions. The degree of fragmentation can be adjusted by the gas temperature or the MS orifice potentials. Unlike ESI, alkali metal ion attachment or doubly charged ions are not observed under DART. Applications Rapid high throughput analysis can be carried out for a variety of sample types using DART. Analysis of drug tablets is one application where direct analysis would be of extreme importance. The DART ion source has been used to analyze an extremely wide range of analytes including drugs (tablets, formulations etc), metabolites in body fluids, explosives, forensics, chemical weapon agents, synthetic organic and organometallic compounds, pesticides, toxic industrial materials, inks, dyes, foods, spices beverages, fatty acids in bacteria and also materials on surfaces such as glass, concrete, paper or currency directly. 3-7 Some examples of such applications we have examined are given below. Spices contain a number of specific components and their characterization can be achieved by directly analyzing them using DART MS. The rhizomes of ginger have long been valued as spice and as a herbal medicine. Dry ginger when placed in front of a DART source gives rise to the spectrum given in Fig. 3 which shows the presence of the reported sesquiterpenes, gingerols, gingerones and shogaols. MS[1];0.739..0.761; / ESI+ / 7JAN050 Sample Comments: DART / Acq. Data Name(FullPath): E:\CDRI\DART_JAN07/JAN07:7JAN050 / Ion Guide RF Volt: 300V Internal Sample Id: DRY GINGER, 250 C Experiment Date/Time: 2/3/2007 2:50:14 PM Intensity (5401) 205.19563 277.1738 CH 3 CH 3 5000 59 H 3 C M. Wt. 204 CH 3 Zingiberene 4000 O CH 3 O (CH 2 ) 4 CH 3 3000 89 HO M. Wt. 276 Shogaol 2000 1000 333.2355 0 100 200 300 400 500 m/z Fig. 3 DART mass spectrum of dry ginger

The spectrum shown in Fig. 4 (upper panel) was obtained when household cooking oil, sunflower oil, on a melting point capillary was placed in front of the DART source in presence of ammonium hydroxide. Ammonium hydroxide was placed in the gap below the sample. The peaks at m/z 872, 896, 998 and 900 are due to ammonium cationized triglycerides. The lower panel in Fig. 4 shows the negative ion DART spectrum of sunflower oil showing the [M-H] - ions of the fatty acids present in the oil. The ions at m/z 255, 279 and 281 are due to palmitic acid, linoleic acid and oleic acid respectively. More such examples will be discussed. M S [ 1 ] ; 0. 2 5 0.. 0. 3 5 9 ; / E S I+ / 6 I2 2 D E C 0 3 8 A S a m p le C o m m e n t s : D A R T / A c q. D a t a N a m e ( F u llp a t h ) : E : \ C D R I\ D A R T _ D E C 0 6 / D E C 2 0 0 6 : 6 I2 2 D E C 0 3 8 A / Io n G u id e R F V o lt : 3 0 0 V In t e r n a l S a m p le Id : S U N F L O W E R O IL, 3 0 0 C, N H 3 E x p e r im e n t D a t e / T im e : 1 / 1 5 / 2 0 0 7 1 : 4 4 : 0 4 P M In t e n s it y ( 8 1 1 0 ) Triglycerides 8 0 0 0 7 0 0 0 6 0 0 0 5 0 0 0 898.781 MNH 4 + 896.770 900.793 4 0 0 0 3 0 0 0 601.5201 872.773 2 0 0 0 1 0 0 0 0 2 0 0 4 0 0 6 0 0 8 0 0 1 0 0 0 m / z M S [ 1 ] ; 0. 5 2 6.. 0. 5 5 9 ; / E S I- / 7 J A N 0 2 6 A S a m p le C o m m e n t s : D A R T / A c q. D a t a N a m e ( F u llp a t h ) : E : \ C D R I\ D A R T _ J A N 0 7 / J A N 0 7 : 7 J A N 0 2 6 A / Io n G u id e R F V o lt : 3 0 0 V In t e r n a l S a m p le Id : S U N F L O W E R O IL, 3 5 0 C N E G E x p e r im e n t D a t e / T im e : 1 / 2 7 / 2 0 0 7 1 0 : 4 2 : 4 6 A M x 1 0 3 In t e n s it y ( 1 8 8 8 5 ) 1 8 1 6 1 4 279.2317 Linoleic acid 1 2 1 0 8 6 4 2 Palmitic acid 255.2324 Oleic acid 281.2467 295.2276 0 2 0 0 4 0 0 6 0 0 8 0 0 1 0 0 0 m / z Fig. 4 Positive (upper panel) and negative (lower panel) ion DART mass spectra of sunflower oil. Conclusion DART is a no-prep sampling technique for MS and works like a solid probe at atmospheric pressure. It operates in open air with no exposed high voltages. DART samples any material held in its ionization stream. It produces instantaneous response and does not need any solvents. Thus DART provides a means for rapid analysis of samples with no solvents or sample preparation. References 1. Z. Takats, J. M. Wiseman, B Gologan and RG Cooks, Science, 306 (2004) 471. 2. R. B. Cody, J. A. Laramee and H. D. Durst, Anal. Chem., 77 (2005) 2297.

3. F. M. Fernández, R. B. Cody, M. D. Green, C. Y. Hampton, R. McGready, S. Sengaloundeth, N. J. White, P. N. Newton, Chem. Med. Chem., 7 (2006)702. 4. R. W. Jones, R. B. Cody, J. F. McClelland, Journal of Forensic Science, 51(2006) 915. 5. C. Y. Pierce, J. R. Barr, R. B. Cody, R. F. Massung, A.R. Woolfitt, H. Moura, H. A. Thompson and F. M. Fernandez, Chem. Commun., 2007 (in press). 6. G. Morlock and W Schwack, Anal. Bioanal. Chem. 385 (2006) 586. 7. J. P. Williams, V. J. Patel, R. Holland and J. H. Scrivens, Rapid Commun. Mass Spectrom., 20 (2006) 1447. Dr. K. P. Madhusudanan, born in 1947, is presently Head of Sophisticated Analytical Instrument Facility, Central Drug Research Institute, Lucknow. After obtaining his M. Sc. Degree from Kerala University in 1970 he did his doctoral research in National Chemical Laboratory, Pune and received his Ph. D. Degree in 1975. Since 1970 he has been working in the area of organic mass spectrometry. He has more than 150 research publications in national and international journals. He has been a member of the Editorial Board of Journal of Mass spectrometry since 1995. He was elected fellow of the National Academy of Sciences, India in 1997. He is a recipient of the ISMAS Eminent Mass Spectroscopist Award (1999). His present interests include effect of metal cationization of organic molecules, applications of mass spectrometry for the characterization of biomolecules and natural products and in drug metabolism and pharmacokinetics and also profiling of natural products and drugs by DART MS.