Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

Similar documents
Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Hydrogen as a Carrier Gas in the Analysis of Polar/Non-Polar Compounds Using the Polyarc System. Application Note. Author. Introduction.

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Author. Abstract. Introduction

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

Introduction to Capillary GC

1.0 Reactive Gas Chromatography Design and Setup Procedure for Catalyst Pelletization & Inlet Packing Software and Method.

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application

Analysis of Terpenes in Cannabis Using the Agilent 7697A/7890B/5977B Headspace GC-MSD System

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

U.S. EPA Method 8270 for multicomponent analyte determination

Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Enhanced Sensitivity for Biomarker Characterization in Petroleum Using Triple Quadrupole GC/MS and Backflushing

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Introduction to Capillary GC

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

Application Note. Abstract. Authors. Introduction

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

Agilent UltiMetal Plus Stainless Steel Deactivation for Tubing, Connectors, and Fittings

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Application Note. Abstract. Authors. Introduction. Food Safety

Chapter 27: Gas Chromatography

Agilent G2350A Atomic Emission Detector (AED)

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

2401 Gas (liquid) Chromatography

Volatile organic compounds (VOCs):

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Activity in the FID Detection Port: A Big Problem if Underestimated

A New PEG GC Column with Improved Inertness Reliability and Column Lifetime Agilent J&W DB-WAX Ultra Inert Polyethylene Glycol Column

CH 2252 Instrumental Methods of Analysis Unit V Gas Chromatography. M. Subramanian

Characterisation of allergens in cosmetics by GC GC TOF MS with Select-eV variable-energy electron ionisation

Highly Sensitive and Rugged GC/MS/MS Tool

Fast Determination of Impurities in Propane- Propylene Streams Using a Pulsed Flame Photometric Detector (PFPD) and a New Capillary.

Use of the SFC-MS Splitter Kit G

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

Understanding Gas Chromatography

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Rapid Analysis of Food and Fragrances Using High-Efficiency Capillary GC Columns. Application. Authors. Abstract. Introduction

Eliminating the Need for Standards and Calibration in GC/FID Analysis of Hydrocarbons and Oxygenates

Determination of Volatile Substances Proof of Food Adulteration

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

The Polyarc System Product Brochure. A Breakthrough Innovation Revolutionizing GC/FID Technology

Analysis of Sulfur-Containing Flavor Compounds By GC/MS With A PFPD

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Methods of Analysis of Polychlorinated Biphenyl Congeners Using an Application-Specific Capillary GC Column

Agilent J&W PoraBOND Q PT Analyzes Oxygenates in Mixed C4 Hydrocarbon Streams by GC/FID and GC/MSD

Spectra Library Electron Capture Negative Ion (ECNI) Mass Spectra of Selected Polybrominated Diphenyl Ethers (PBDEs)

Selection of a Capillary GC Column

Selective Formation of Benzo[c]cinnoline by Photocatalytic Reduction of 2,2 Dinitrobiphenyl with TiO 2 and UV light irradiation

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.

New ZB-5HT Inferno The World s Highest Temperature Non-Metal GC Column

Development and validation of a pneumatic calculator for the reversed-flow differential flow modulator for comprehensive GC GC

Blood Alcohol Determination using Static Headspace Analysis with Optimized Sample Throughput

Application. Gas Chromatography February Introduction

Detection of Chemical Warfare Agents by Transportable GC/MS

Medical Cannabis Pesticide Screening using the SRI 8610C GC

Analysis of Natural Oils and Extracts Using the Low Thermal Mass LTM Series II System

Virtually Particle-Free Rt -Silica BOND Columns

Study of Residual Solvents in Various Matrices by Static Headspace

Agilent G3212 GC-APCI Source

Solid Phase Micro Extraction of Flavor Compounds in Beer

Chapter 31 Gas Chromatography. Carrier Gas System

Determination of 17 Organotin Compounds in Beverages Using Triple Quadrupole GC-MS/MS System

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

Characterization and Classification of Heroin from Illicit Drug Seizures Using the Agilent 7200 GC/Q-TOF

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

Gas Chromatography (GC)

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform

MS n Analysis With Fast Polarity Switching in the Agilent 1100 Series LC/MSD Trap SL. Application Note. Christine Miller Agilent Technologies

Residual solvents analysis using an Agilent Intuvo 9000 GC system

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs

Gas Chromatography. Introduction

Secrets of GC Column Dimensions

Ultra-Inert chemistry for Trace Level Analysis

GC Instruments. GC Instruments - Sample Introduction

VUV ANALYTICS VGA-100 GC DETECTOR Gas Chromatograph Agilent 6890 equipped with a 7683 model autosampler Restek 30m x 0.25mm x 0.

Gas Chromatography. Chromatography Laboratory Course. Dr. Christian Jungnickel Chromatography Course GC September 2005

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

HIGH SPEED GC-TOFMS FOR ESSENTIAL OILS PROFILING IN ROUTINE QUALITY ASSESSMENT

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE

GAS CHROMATOGRAPHY (GC)

Expectations for GC-MS Lab

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Transcription:

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS Application Note Multi-detector Splitter Authors Charlie Spanjers and Andrew Jones Activated Research Company 7561 Corporate Way Eden Prairie, MN 553 Charlie.Spanjers@activatedresearch.com Erick White National Renewable Energy Laboratory 15013 Denver West Parkway Golden, CO 8001 Erick.White@nrel.gov Abstract Quantification of unknowns with gas chromatography traditionally requires time-consuming and costly calibration steps including purchasing, preparing, and analyzing calibration standards, and applying the calibration results to determine analyte concentration. In this application note, we describe a method for identifying and quantifying unknowns in a single injection using a parallel Polyarc /FID and mass spectrometer (MS. converting all organic molecules to methane, water, and non-carbonaceous by-products for a uniform FID response. The response of the FID per carbon atom becomes identical for all carbon-containing compounds and calibrations are no longer necessitated. In this application note, we describe a method for configuring a Polyarc /FID and MS in parallel leading to a single system that can identify and quantify unknowns in a single injection. The simultaneous analysis of compounds with two or more detectors and a single inlet and column requires the use of a post-column flow splitter prior to detection. A constant split ratio as a function of temperature is necessary to obtain accurate results when using calibration-free methods. Keeping the split ratio constant requires the use of a pressurized splitter, such as the Agilent Two-Way CFT splitter. The configuration of a system utilizing a splitter is shown in Figure 1 below. Introduction The mass spectrometer (MS has long been the go-to instrument for qualitative identification of analytes in gas chromatography. Quantification, on the other hand, is most often performed with a flame ionization detector (FID because of its linearity (over seven orders of magnitude and its sensitivity to low-level organic analytes. The FID, however, does not provide a uniform response for different molecules, and therefore, requires detailed calibration of the detector response for accurate quantification. FID calibration is time consuming, and in many cases, impossible when standards do not exist or are prohibitively expensive. The Polyarc system solves these problems by Figure 1. Configuration of a Polyarc /FID and MS for simultaneous compound identification and quantification. In this figure, it can be seen that the column connects directly to the port adjacent to the He inlet. The next port adjacent to the column inlet is fed to the Polyarc via the metal capillary restrictor built-in to the Polyarc (R. The last port is connected to the MS via a fused silica capillary restrictor (. The restrictors provide backpressure for the added He flow at the CFT splitter and, control the split ratio between the two detectors, 1

and limit the flow to the MS. Note: If using a threeway splitter for configuring this method, the port adjacent the He inlet (the column port in Figure 1 should be plugged with a no-hole ferrule and the column should be attached at the port adjacent the plugged port. Flow in capillary tubes is governed by the theory of ideal flow of compressible gases, as shown below: F 60πr 16η(TL (T ref T i P o (1 where F is flow rate (sccm, r is the radius of the capillary tube (cm, η(t is the temperaturedependent viscosity of the gas a s, L is the length of the capillary tube (cm, P i and P o are the inlet and outlet pressures a, respectively, T is the temperature (K, T ref is the reference temperature (98 K, and is the reference pressure (101,300 Pa. The split ratio (S between MS and Polyarc detectors is defined as the ratio of the gas flows to each detector, as shown below (complete derivation is in the Appendix: S F Polyarc F MS r R L split P Polyarc R r L split ( where F MS is the flow rate to the MS, F Polyarc is the flow rate to the Polyarc, P split is the pressure of the splitter, P MS is the pressure of the MS split P MS, so that P split - P MS P split, P Polyarc is the pressure of the Polyarc, r and r R are the inner radii of restrictor and R, respectively, and L and L R are the lengths of the restrictor capillaries and R, respectively. P split is controlled by the Aux EPC and should be programmed to remain constant throughout all runs. P Polyarc is the pressure inside the Polyarc and will vary slightly with temperature. However, the variation is small. Experimental A-MFC-A09 and to the FID. The effluent of the GC column was connected to an Agilent 3-way CFT splitter (G3183-60500 according to Figure 1. Because a 3- way splitter was used to perform a -way split, the port adjacent to the He inlet was plugged with an Agilent no-hole ferrule to minimize dead volume. The MS was connected to the splitter via a retention gap column (Agilent, 160-635-5, 0.61 m, 0.1 mm ID, labeled in Figure 1. The inlet capillary to the Polyarc was connected directly to the splitter according to Figure 1. The splitter was controlled by an EPC (with restrictor frit removed. It is very important that the restrictor frit is removed from the EPC to maintain a constant pressure at the splitter. GC conditions Front inlet Split Split ratio 10:1 Inlet temperature 50 C Column flow 3 sccm Septum purge flow 3 sccm Inlet liner Agilent 1870-80190 Oven 0 C (5 min, 15 C/min to 15 C (0 min, 5 C/min to 50 C ( min Column DB-5MS UI (30 m 0.5 mm 1 µm Syringe 10 µl Injection volume 0.5 µl MS conditions Energy 70 ev Scanning range 9-500 m/z Source temp. 30 C Quadrupole temp. 150 C Transfer line temp 50 C Transfer line length 0.6 m Transfer line diam 0.1 mm ID FID conditions Temperature 300 C H 1.5 sccm Air 350 sccm Makeup 5 sccm (He Sampling rate 100 Hz Polyarc reactor conditions Setpoint 93 C H 35 sccm Air.5 sccm An Agilent 7890A GC equipped with a split/splitless inlet, capillary-optimized FID, mass spectrometer (Agilent 5973, and Polyarc reactor (ARC PA-RRC- A0 were used for the analysis. Helium (99.999%, Praxair was used for carrier and FID makeup. Air (zero grade, Praxair and H (99.999%, Praxair were supplied to the ARC electronic flow control module

Results and Discussion The theory above was verified by analyzing an 8- component mixture of alcohols and alkanes (50 ppm each in dichloroethane: Restek Polar ISO to ensure the split ratio remains constant with changing oven temperature. The average error in concentration was 1.5%, consistent with the typical accepatance criteria of this mixture of 3%. An application of the parallel Polyarc /FID and MS to the analysis of E85 gasoline is shown in Figure. Quantification of ethanol with a single injection without calibration yields a measured concentration of 88 wt. % ethanol. The actual concentration of the sample measured by ASTM D5501 methods is 86 wt. %. The small error between the two methods of.3% is within the accuracy of calibrations and the coelution of some minority species. Figure. Parallel Polyarc /FID (A and MS (B for the quantification and identification of the components of E85 gasoline Basic steps for system configuration (contact ARC for more information: Parts needed for an Agilent 7890 configuration (contact ARC for other systems: Polyarc System with Electronic Flow Control A-SYC-01 Polyarc MS Split Kit A-SPL-KT7 Agilent Aux 3-Channel EPC A-EPC-3C7 Configuration instructions on an Agilent 7890 (contact ARC for other systems: 1. Install an Agilent 3-Channel EPC. Important: Remove the restrictor frit from the EPC module.. Install an Agilent -way CFT splitter and connect the input to the EPC. 3. Connect the column to the CFT splitter (see Figure 1.. Install the Polyarc system with electronic flow control, following the installation manual. 5. Connect the inlet of the Polyarc directly to the CFT splitter according to Figure 1 (do not change the length of the Polyarc inlet transfer capillary. 6. Install a 0.6 m, 0.1 mm ID transfer line (Agilent, 160-635-5 to the MS. 7. Configure the Aux EPC, and set the EPC pressure to psi in all runs. 8. Configure the GC method with the configuration above to ensure column flow rates are set correctly. Notes: 1. Ensure that the splitter is properly installed and configured in the method.. The flow control frit in the EPC flow module must be removed or replaced with a zero resistance tube (removal of the frit ensures the pressure at the splitter is equal to the pressure set by the EPC. 3

Conclusions The combined Polyarc /FID and MS system proves to be a highly useful tool for identification and quantification of analytes in a single injection. A constant split ratio with oven temperature ramps can be obtained with a constant splitter pressure controlled by an EPC. This allows for the unparalleled simultaneous quantification and identification without the need for calibration using equimolar carbon response of the Polyarc/FID and fragmentation databases for the MS. Contact Us For more information or to purchase a Polyarc system, please contact us at 61-787-71 or contact@activatedresearch.com. Please visit our website for details and additional technical literature, www.activatedresearch.com. Activated Research Company shall not be liable for errors contained herein, or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. 017 Activated Research Company, LLC Printed in the USA October 13, 017 PA-APP-1713

Appendix Derivation of Eq. (: S F Polyarc F MS 60πr R ( T ref 16η(TL R T split P Polyarc 60πr ( T ref 16η(TL T split P MS r R L split P Polyarc R r L split P MS r R L split P Polyarc R r L split ( 5