Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Similar documents
Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS

Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES

Trace Analyses in Metal Matrices Using the ELAN DRC II

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Overcoming Interferences with the Thermo Scientific icap 7000 Plus Series ICP-OES

Fergus Keenan Shona McSheehy Julian Wills

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS

Using cold vapor generation atomic absorption to determine mercury impurities in pharmaceutical products

Accurate measurement of elemental impurities in metals and metal alloys using the Thermo Scientific icap TQ ICP-MS

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

Analysis of trace elements in polymer samples to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Simplified workflows for the Thermo Scientific icap TQ ICP-MS

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

Sample Analysis Design PART II

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors

February 20, Joe Cerniglia The International Group for Historic Aircraft Recovery (TIGHAR) Job Number: S0CHG688. Dear Joe:

Determination of Carbon d13c isotopic composition from DIC samples

Elemental analysis of river sediment using the Agilent 4200 MP-AES

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

Samples compliant with the WATER

OES - Optical Emission Spectrometer 2000

Advanced ICP-MS techniques for overcoming interferences in LA-ICP-MS bioimaging

Analysis of domestic sludge using the Agilent 4200 MP-AES

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

Implementing EPA Method 6020 with the Agilent ICP-MS Portfolio

Sample Analysis Design PART III

Determination of Elements at Sub-ppb Concentrations in Naphtha Mixtures Using the NexION 300 ICP-MS

Analysis of noble metals at low levels in geological reference materials and ores

Rapid, simple, interference-free analysis of environmental samples using the XSERIES 2 ICP-MS with 3 rd generation CCT ED

SmartNotes. What is meant by the term interference?

Environmental Series - Trace contaminant analysis in brine using a Thermo Scientific icap 6000 Series Duo ICP

Basic and Application Inductively coupled plasma mass spectrometry (ICPMS)

Application. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid. Authors. Introduction. Abstract.

ICP-OES Application Note Number 35

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES

Hands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment

Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3

Determination of Phosphorus and Sulfur with Dynamic Reaction Cell ICP-MS

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Performance of the Agilent 7900 ICP-MS with UHMI for high salt matrix analysis

Following documents shall be used for reference on quantities, units, prefixes and other technical vocabulary in this document:

Elemental Scientific. Spike Recovery. Elemental Scientific

Defining quality standards for the analysis of solid samples

Agilent ICP-MS. Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application

Elemental Scientific. Spike Recovery. Elemental Scientific

Thermo Scientific ICP-MS solutions for the semiconductor industry. Maximize wafer yields with ultralow elemental detection in chemicals and materials

Sample Analysis Design Polyatomic Interferences

Elemental Scientific. Stability. Elemental Scientific

Key Words Nanoparticles, spicp-ms, ICP-MS, Qtegra ISDS Software, AF4-ICP-MS

icap TQ ICP-MS Applications Compendium

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry

China Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

Low level Os isotopic measurements using multiple ion counting

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

Rapid and precise calcium isotope ratio determinations using the Apex-ACM desolvating inlet system with sector-field ICP-MS in low resolution

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Elemental Analysis: NCS characterization of paper

Robust ICP-MS with ease of use and high productivity for routine laboratories

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS

Determination of trace anions in concentrated hydrofluoric acid

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Handling Interferences in the Modern Laboratory with Advanced Triple Quadrupole ICP-MS Technology

ELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW

Analysis Repeatability of Trace and Major Elements in a Water Sample

Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS

Redefining triple quadrupole technology with unique ease of use

Key Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company

Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS

Automated and accurate component detection using reference mass spectra

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Outsmart Interferences with Triple Quadrupole ICP-MS. The world leader in serving science

Meeting the Challenges of Soil Analysis with the Avio 200 ICP-OES

Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS)

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

- A spark is passed through the Argon in the presence of the RF field of the coil to initiate the plasma

On-line isotope dilution analysis with the 7700 Series ICP-MS: Analysis of trace elements in high matrix samples

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Automated Determination of PPQ Levels of Thorium in High Purity Copper

Ar Ar + e - INDUCTIVELY+COUPLED+ PLASMA+SPECTROMETRY+ What+is+Plasma?+ FuncDon+of+Plasma+

Transcription:

TECHNICAL NOTE 4322 Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS Keywords Argon gas dilution, AGD, High matrix samples, Seawater Goal To critically assess the use of Argon Gas Dilution (AGD) on the Thermo Scientific icap RQ ICP-MS for the direct analysis of high matrix samples such as seawater. Introduction High matrix samples remain a challenge in ICP based applications. For example, high levels of dissolved solids can cause ionization suppression that reduces sensitivities during the analysis of real samples by both ICP-OES and ICP-MS. For ICP-MS the situation is further complicated as high matrix levels may lead to blockage of the narrow aperture cones used in the interface between the ICP ion source and the mass spectrometer. While continual advances in interface design have improved ruggedness, manual or automated sample dilution using a suitable solvent (usually dilute acid) is still commonly used to reduce sample matrix levels to ~% total dissolved solids (TDS) before ICP-MS analysis. Depending on both the application and required sample throughput level, however, this may not be ideal as it increases both the risk of contamination and the cost per analysis (due to the additional time and consumables).

Argon Gas Dilution (AGD) is a useful technique to address these fundamental limitations. In AGD-ICP-MS analyses, the argon (Ar) gas flow through the nebulizer is reduced while the total Ar gas flow to the plasma is maintained by the addition of a make-up Ar gas flow to the aerosol leaving the spray chamber. The sample aerosol is therefore diluted with Ar gas inside the ICP-MS sample introduction system. Since the absolute amount of sample entering the plasma is limited, the dissociation of heavy matrix samples in the ICP is improved, reducing matrix deposition on the ICP-MS interface. Overall plasma robustness improves in AGD mode as demonstrated by a reduction in metal oxide formation. Additionally, dilution with clean Ar gas minimizes contamination by reducing sample handling without increasing analysis times or requiring extra (potentially costly) consumables. The use of AGD is not, however, without disadvantages. In contrast to automated, sample specific dilution, all samples are diluted in AGD mode, reducing absolute instrumental sensitivity. Furthermore, since the proportion of sample to Ar delivered to the ICP is reduced, the plasma runs slightly drier lowering the ionization efficiency for elements with higher st ionization potentials (Zn, Cd for example) leading to reduced recoveries. Approaches to overcome this inherent characteristic of Ar gas based sample dilution, including re-humidifying the plasma and the addition of excess carbon, are investigated in this note. Instrument configuration A Thermo Scientific icap RQ ICP-MS fitted with the Argon Gas Dilution Kit was used for all analyses (see Figure ). As part of this kit, a pergo Argon Nebulizer Gas Humidifier (Elemental Scientific, Omaha, US) is used to humidify an additional argon gas supply added to the sample aerosol. Instrument parameters are shown in Table. Table. Instrument parameters. Parameter Value Figure. Argon Gas Dilution, AGD (PN: 3783) on the icap RQ ICP-MS. Method development In order to assess the AGD technique, the icap RQ ICP-MS was used to determine trace elements in undiluted seawater (NRCC and reference materials) using a series of different AGD conditions. Seawater, with approximately 3.5% (35, μg ml - ) total dissolved solids (TDS), over a factor of higher than the normal TDS limit for ICP-MS analysis, was chosen for these tests as it is one of the most challenging sample matrices analyzed by ICP-MS. The two seawater samples were spiked with a multielement standard solution at concentration levels between 2 and 2 μg L -. A sample of 2% HNO 3 spiked at the same level was also prepared. Forward Power 55 W Nebulizer Glass, concentric (pumped, 4 μl.min - ) Nebulizer Gas Flow 5 L min - Add. Gas Flow (Ar).95 L min - Interface Sample Introduction QCell Conditions Ni sampler and skimmer cones High matrix insert Orange/Yellow tubing, 2 rpm 5 ml min - He, 3V KED

An internal standard solution, containing Sc, Ge, Rh and Ir in 2% HNO 3 (and % isopropanol, where appropriate), was added to all samples via a T-piece (a : online dilution) before entering the spray chamber. Higher dilution factors (of up to :9) can be achieved by combining peristaltic pump tubing with different internal diameters so that samples with higher amounts of total dissolved solids can be analyzed using the same configuration. Dry Ar The following AGD configurations were tested: 9 Be 27 Al 5 V 52 Cr 55 Mn 57 Fe 59 Co 6 Ni 63 Cu 66 Zn 7 Ga 75 As 78 Se 85 Rb 95 Mo 7 Ag Cd 25 Te 37 Ba 25 Tl 28 Pb 29 Bi 238 U Argon gas dilution using dry Ar. Humidified Ar Argon gas dilution using humidified Ar. Argon gas dilution using humidified Ar + % CH 4 (in Ar) added after the humidifier at a flow rate of ml min -. Argon gas dilution using humidified Ar + % (v/v) isopropanol in the internal standard solution. The spiked seawater and 2% HNO 3 samples were then analyzed using the four AGD settings on the icap RQ ICP-MS. 9 27 5 52 55 Be Al V Cr Mn 57 59 6 63 Fe Co Ni Cu 66 Zn 7 75 78 Ga As Se 85 Rb 95 Mo 7 Ag Cd 25 Te 37 Ba 25 28 Tl Pb 29 238 Bi U Humidified Ar with CH 4 Results Relative recovery values, calculated from blank subtracted intensities in the seawater and 2% HNO 3 matrices, were used to assess the extent of signal suppression for each AGD setting tested. No internal standard correction was made. As an acceptance criteria, signal recovery in the seawater matrix should be between 8 2% for all elements, as commonly accepted for environmental analysis. 9 Be 27 Al 5 V 52 Cr 55 Mn 57 Fe 59 Co 6 Ni 63 Cu 66 Zn 7 Ga 75 As 78 Se 85 Rb 95 Mo 7 Ag Cd 25 Te 37 Ba 25 Tl 28 Pb 29 Bi 238 U Relative recovery values for each AGD configuration tested are shown in Figure 2. Humidified Ar with IPA 9 Be 27 Al 5 V 52 Cr 55 Mn 57 Fe 59 Co 6 Ni 63 Cu 66 Zn 7 Ga 75 As 78 Se 85 Rb 95 Mo 7 Ag Cd 25 Te 37 Ba 25 Tl 28 Pb 29 Bi 238 U Figure 2. Trace metal recoveries for each AGD configuration in undiluted seawater relative to 2% HNO 3. Red-shaded areas indicate recoveries of between 8 2%.

Results for the four AGD settings tested are summarized below: Dry Ar: Dilution with dry Ar shows acceptable (8 2%) recoveries for most trace metals in the ICP-MS analysis of the two seawater samples analyzed. Significant signal suppression is observed, however, for several critical analytes with relatively high st ionization potentials (Zn, Se, Cd and Te), where recoveries as low as 5% (Te in ) are observed. These low recoveries demonstrate the unsuitability of dry Ar for multielemental AGD-ICP-MS analysis of high matrix samples such as seawater. Humidified Ar: With an argon humidifier to re-wet the plasma for AGD-ICP-MS analysis, signal recoveries for the higher ionization potential (IP) elements are improved, but still do not meet the 8% lower recovery limit commonly used in environmental analyses. Humidified Ar with CH 4 : The introduction of carbon to the plasma in the form of methane addition (% CH 4 in Ar) does not significantly improve recoveries for the higher IP elements and a slight reduction in recovery is observed for others. Humidified Ar with IPA: When carbon is introduced to the plasma via the addition of isopropanol (% v/v) to the internal standard solution, AGD-ICP-MS recoveries for all trace elements, including those with higher st ionization potentials, are between 8 2%. To verify the robustness of the humidified Ar with IPA approach for AGD-ICP-MS analysis, the two spiked seawater reference materials ( and ) were quantified against an external calibration curve (prepared in dilute HNO 3 ) and analyzed over a 7.5-hour period regularly interspersed with quality control samples. The AGD-ICP-MS method developed minimizes signal suppression so that undiluted seawater can be quantified against simple HNO 3 calibration standards. During the entire analysis, minimal signal drift was observed, as shown in Figure 3. Figure 3. Internal standard recoveries during a 7,5-hour analysis of undiluted seawater. Average recoveries and precisions over the 7.5-hour analysis sequence for the spiked and seawaters are shown in Table 2. Table 2. Accuracy and precision data from a 7.5-hour AGD-ICP-MS analysis of the spiked seawater reference materials and using the icap RQ ICP-MS. Average trace metal accuracies are expressed as the relative percentage of the spiked concentrations. Isotope (%) Precision (%RSD) (%) Precision (%RSD) 5 V 95 2.2 96 2.2 52 Cr 95 3.5 97 2.9 55 Mn 86 4.2 89 4.7 56 Fe 85 2.9 85 4.2 59 Co 9 2.8 9 2.5 6 Ni 9 3.4 9 3. 63 Cu 94 2.7 97 2.3 68 Zn 85 2.7 86.9 7 Ga 95 3. 96 2.7 75 As 2.3 3 2.2 78 Se 9 2.5 9 2.4 85 Rb 99 2.7 99 2.9 95 Mo.5.6 7 Ag 89 8.3 Cd 87 2.2 88.7 37 Ba 99 3. 99 3.2 25 Tl 94.6 93 3.5 28 Pb 93.9 93 2. 29 Bi 85 2. 86 2. 238 U 95 2. 96.9

As can be seen from the percentage recovery data in Table 2, the AGD-ICP-MS method developed for the icap RQ ICP-MS provides accurate data for all trace metals determined in the spiked seawater reference materials. Concentration precisions of below 5% RSD demonstrate the robustness of the AGD-ICP-MS technique for the long term direct analysis of undiluted seawater. Software implementation AGD based analysis on the icap RQ ICP-MS is fully supported by the Thermo Scientific Qtegra Intelligent Scientific Data Solution (ISDS) Software. Both nebulizer and additional gas supplies are controlled by Qtegra ISDS Software and an AGD specific autotune is provided, making method development straightforward. Instrument performance under AGD conditions can be assessed using a Qtegra ISDS Software supplied Performance Report. AGD can be integrated with any Thermo Scientific QCell based solution for interference removal, i.e. collision based He KED or reaction based CCT. Multi-mode AGD analysis methods can be easily defined to provide accurate, interference-free analyses in even the most challenging sample matrices. Performance Lowering the nebulizer gas flow in AGD based analyses reduces sample transfer to the plasma decreasing absolute instrumental sensitivity when compared to non-agd based conditions. At the same time however, metal oxide formation is reduced. Typical performance in STD, STD-AGD and KED-AGD analysis modes is shown in Table 3. While oxide and doubly charged formation are shown as percentages. Table 3. Typical performance of an icap RQ ICP-MS in STD, STD-AGD and KED-AGD analysis modes. Sensitivities are expressed as cps ppb -. STD STD-AGD KED-AGD 7 Li 788 434 23 59 Co 3645 7627 335 5 In 27964 2739 8953 238 U 383232 3582 4238 CeO/Ce.8%.9%.5% Ba ++ /Ba +.8%.%.8% Conclusion Argon gas dilution significantly improves plasma robustness allowing the direct analysis of high matrix samples such as undiluted seawater using the icap RQ ICP-MS. Optimum performance for high IP analytes in such samples is only possible, however, after online addition of water vapor and carbon. Find out more at thermofisher.com/sq-icp-ms 27 Thermo Fisher Scientific Inc. All rights reserved. pergo is used in trade by Elemental Scientific Inc. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. TN4322-EN 27