Analysis of trace elements in polymer samples to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES

Similar documents
Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

INVESTIGATION OF ICP-OES ANALYSIS FOR DETERMINATION OF TRACE LEAD IN LEAD-FREE ALLOY

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

Environmental Series - Trace contaminant analysis in brine using a Thermo Scientific icap 6000 Series Duo ICP

Overcoming Interferences with the Thermo Scientific icap 7000 Plus Series ICP-OES

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES

Alphasense Ltd. Compliance Statements and Certificates. Contents. The Restriction of the Use of Certain Hazardous Substances (RoHS)...

Test Report. LEDELS LIGHTING CO., LTD Address. Applicant

Microwave Extraction Provides More Reliable Analysis of High Mass Unit Additives

Analysis of domestic sludge using the Agilent 4200 MP-AES

Accurate Quantification of Cadmium, Chromium, Mercury and Lead in Plastics for RoHS Compliance using the Agilent 7500ce ICP-MS

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

TEST REPORT. NO.: A001R Date: Dec 27, 2013 Page 1 of 13

Verification Report. No. CANEC Date: 12 Apr Page 1 of 9.

Meeting the Challenges of Soil Analysis with the Avio 200 ICP-OES

Sample Preparation of Electronic Device Components for Hexavalent Chromium Analysis by IEC Method :2017

RoHS/WEEE Measurement of Hazardous Substances in Plastics

3M Scotchkote Liquid Epoxy Coating 323+ Brush Grade

ICP-OES Application Note Number 35

OPTIMIZING METALS SAMPLE PREP

3M Scotchkote Liquid Epoxy Coating 323

PROCEDURES. Pharmacopeial Forum 2 Vol. 36(1) [Jan. Feb. 2009]

TEST REPORT. * Note 2 : This report shall not be reproduced except in full without the written approval of the testing laboratory.

FUNDAMENTAL PARAMETERS ANALYSIS OF ROHS ELEMENTS IN PLASTICS

Accurate measurement of elemental impurities in metals and metal alloys using the Thermo Scientific icap TQ ICP-MS

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors

February 20, Joe Cerniglia The International Group for Historic Aircraft Recovery (TIGHAR) Job Number: S0CHG688. Dear Joe:

Applicant: Cactus Technologies Ltd Date : Apr 07, 2010 Suite C 15/F Capital Trade Center 62 Tsun Yip Street Kwun Tong Kowloon Hong Kong

APPLICANT: CACTUS TECHNOLOGIES LTD DATE : JUL 25, 2007 SUITE /F BILLION TRADE CENTER 31 HUNG TO ROAD KWUN TONG KOWLOON HONG KONG

á233ñ ELEMENTAL IMPURITIES PROCEDURES

RoHS COMPLIANCE STATEMENT

Test Method: CPSC-CH-E

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Using cold vapor generation atomic absorption to determine mercury impurities in pharmaceutical products

Test Report No. CRSSA/03608/18 Date : 28/03/2018 Page: 1 of 7 CRS Ref. CRSSA/18/0736/Toray

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Preparation of Food and Feed Samples for Metals Analysis

Pharmacopeial Forum Vol. 36(1) [Jan. Feb. 2010] 1

Test Report REPORT NO. : CRSSA/ /16 CRS REF. : CRSSA/16/3035/Hitachi DATE REPORTED : 06 th Jan., 2017 PAGE : 1 of 5

Certificate of Compliance

TEST REPORT. ADDRESS : 77, Seokam-ro 13-gil, Iksan, Jeonllabuk-do, Korea PAGE: 1 of 8 REPORT NO. RT17R-U2181-E DATE: Dec. 29, 2017

TEST REPORT. The following sample(s) was(were) identified by the customer as : PP-NON WOVEN

Verification Result :.

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

Test Report REPORT NO. : CRSSA/ /16 CRS REF. : CRSSA/16/0030/Hitachi DATE REPORTED : 13 th Jan., 2016 PAGE : 1 of 5

ICP Test Report Certification Packet

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

Verification Result :.

Test Report No. CRSSA/ /17 Date : 04/01/2018 Page: 1 of 5 CRS Ref. CRSSA/17/3272/Toray

LABORATORY ROUND ROBIN TEST PROJECT: ASSESSING PERFORMANCE IN MEASURING TOXICS IN PACKAGING

Simplified workflows for the Thermo Scientific icap TQ ICP-MS

3M MetPak CP2 Press-Fit Header 2 mm Type C, 55 Signal Contacts, Left Guidance CP2 Series

DISCLAIMER: This method:

3M CF Card Header CompactFlash Type I, Low Profile, Long Guides, SMT, Inverse

Test Report NO.: ASZ L-2 Date: Jun 19, 2008 Page 1 of 5

Using FIMS to Determine Mercury Content in Sewage Sludge, Sediment and Soil Samples

TEST REPORT. : The following submitted sample(s) said to be:- SAMPLE RECEIVED : Jan. 07, 2019 TESTING DATE : Jan. 07, 2019 ~ Jan.

This product/component/material was tested to determine the levels of the six hazardous substances in accordance with the standard

Verification Report. No. CANEC Date: 10 Jul Page 1 of 13.

Verification Report. No. CANEC Date: 21 Jun Page 1 of 8.

Test Report. Report No. RLSHD Page 1 of 6

Determination of Total Bromine and Iodine Emission Spectrometric Method (ICP-OES) EuSalt/AS

Elemental Analysis: NCS characterization of paper

: DAM1-125,DAM1-160,DAM1-400,DAM1-630,DAM1-800, DAM1-1600

LUXEON UV U1. Highest power density, superior efficiency, powered by leading Chip Scale Package (CSP) technology ILLUMINATION

ICP Test Report Certification Packet

Test Report No _rev01 Date: 06/DEC/2017 Page 1 of 5

Determination of trace anions in concentrated hydrofluoric acid

Asbestos. Beryllium. Mica. Graphite (CAS No ) Ozone Depleting Substances. Phosphorus. Phthalates 4. California Proposition 65

Test Report. As per applicant s request

Test Report NANTONG HUADA MICROELECTRONICS GROUP CO., LTD NO.99 ZILANG ROAD,NANTONG CITY,JIANGSU PROVINCE

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

TEST REPORT. : The following submitted sample(s) said to be:-

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

3M Serial Attached SCSI (SAS) Connector Boardmount Through-Hole, Vertical Receptacle

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Mass Spectral Studies of Polypropylene Chromatographic Well Plates

Test Report REPORT NO. : CRSSA/ /17 CRS REF. : CRSSA/17/1680/Hitachi DATE REPORTED : 06 th July, 2017 PAGE : 1 of 5

Analysis of noble metals at low levels in geological reference materials and ores

TEST REPORT. Page : 1 of 9 Report No. RT14R-U E Date: Oct. 13, : The following submitted sample(s) said to be:-

Elemental analysis of river sediment using the Agilent 4200 MP-AES

Materials Declaration

TEST REPORT. Address : , Doksan-dong, Geumcheon-gu, Seoul, Korea Page: 1 of 12 Report No. RT16R-S E Date: Jul.

Laboratory 101: A Guide to Understanding your Testing Laboratory

Test Report No Date : 27/NOV/2017 Page 1 of 5

Test Report. Report No. SCL03H Page 1 of 7

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

TEST REPORT. * Note 2 : This report shall not be reproduced except in full without the written approval of the testing laboratory.

Analysis of Cadmium (Cd) in Plastic Using X-ray Fluorescence Spectroscopy

1. Intended use. 2. Introduction. 3. Assay description. 4. Limitations. Instruction For Use Gyrolab Generic PK Kit

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

TEST REPORT. : Lead Frame / Ag is plated on Ni-Fe Alloy-A42 material

Transcription:

APPLICATION NOTE 43145 Analysis of trace elements in polymer s to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES Authors Introduction Sanja Asendorf, Application Specialist, Thermo Fisher Scientific, Bremen, Germany The Waste Electrical and Electronic Equipment (WEEE) and the Restriction of the use of certain Hazardous Substances (RoHS) Directives of the European Parliament and Council give general guidelines to the member governments on the disposal of WEEE and RoHS in electronic equipment sold in, into and out of the European Union. The directives and legislation affect electronics manufacturers as well as commercial, household equipment producers and consumers. The countries and companies affected are those within the European Union (EU) and in regions that trade with the EU. Keywords IEC 62321, Microwave digestion, Polymer, RoHS, Trace analysis, Waste materials, WEEE Goal To demonstrate the ability of the Thermo Scientific icap 7000 Plus Series ICP-OES to monitor analytes in electrical and electronic goods for compliance with WEEE/RoHS regulations. The WEEE1 and RoHS Directives2 are guidelines set out by the European Parliament detailing the goals for the disposal of waste electronic equipment and the restrictions applied to the sale of new equipment in the EU. Individual member governments must determine how best to implement these guideline directives to achieve the EU goals in their respective countries.

The WEEE Directive specifically targets the responsible recycling of separately collected waste electronic equipment and does not govern the hazardous content of that equipment. The RoHS Directive restricts sales in the EU market of any new electrical or electronic equipment containing more than the agreed levels of hazardous substances and is thus used as a pre-emptive control of waste. The guidelines and laws enacted by the individual governments also pertain to the importation of goods into the EU and thus affect a great number of countries around the world 3. The RoHS regulations require analysis of materials used in electronic equipment to restrict and verify the concentrations of cadmium, hexavalent chromium, lead, mercury, polybrominated biphenyls, diphenyl ethers, and phthalates. The maximum limits of hazardous substances are calculated on homogenous materials which means a material that cannot be mechanically separated into its component materials. The maximum allowable quantities in electrical equipment of the hazardous substances listed above are: 0.1% by weight for lead, mercury, hexavalent chromium, polybrominated biphenyls, diphenyl ethers and phthalates 0.01% by weight for cadmium International Electrotechnical Commission analysis protocol IEC 62321 The International Electrotechnical Commission (IEC) is the leading global organization that publishes consensusbased International Standards and manages conformity assessment systems for electric and electronic products, systems and services, collectively known as electrotechnology 4. The IEC 62321 protocol for the analysis of restricted substances provides guidelines for the sampling and analysis of all the restricted substances in electronics and polymers found in electronic parts or equipment. With regards to the ICP-OES technique, only cadmium, lead and mercury are considered for analysis since chromium requires a separation technique to be used to allow the hexavalent form to be analyzed. However, using ICP-OES instrumentation with simultaneous detection capability allows the analysis of total chromium with no extra cost or time penalty and can provide a valuable screening tool to determine if further chromium speciation analysis is required. Instrumentation The Thermo Scientific icap 7200 ICP-OES Duo was used for the analysis of polymer materials which are a common component of materials governed by the RoHS legislation 5. The icap 7200 ICP-OES Duo provides powerful analyte detection capabilities within a small footprint to provide a highly cost effective solution for the routine analysis of smaller batches of liquid s. A standard introduction kit, containing a glass concentric nebulizer and a glass cyclonic spray chamber was used. Sample and standard preparation The Reference Materials (s) chosen to demonstrate analytical performance capabilities were obtained from LGC Limited (UK) and were specifically developed to provide certified standards for the analysis of polymers subject to WEEE-RoHS legislation. The s chosen for this analysis were supplied in small chips and are listed as follows: EC680 polyethylene (low level) EC681 polyethylene (high level) PVC2 polyvinyl chloride PVC1 polyvinyl chloride with low concentrations of analytes, used to determine the method detection limits (MDL) The range of element concentrations present in the polymers are representative of typical ranges that may be found in manufactured polymers. The s were digested and processed as supplied, without further grinding or freeze milling. The IEC 62321 protocol contains sampling recommendations and dissolution guidelines for materials and various s which require analysis for RoHS hazardous substances including metals, polymers and electronic components. The certified reference polymer chips used in this analysis are expected to be homogenous in terms of their elemental composition. However, this may not always be the case and it is therefore recommended to analyze multiple chips from any given in order to establish homogeneity.

The preparation detailed below is based on one of several methods suggested by the IEC 62321 protocol. In this case, a microwave digestion method was utilized due to the potential volatile nature of the analytes as well as the simplicity of the digestion matrix and speed of dissolution. The coarsely ground s (which were used as supplied) were also used by the IEC to determine the efficacy of the IEC 62321 method protocol. The masses were intentionally kept to a minimum so that any matrix differences between s and standards are also minimized. This approach was taken to remove the necessity for internal standards, thereby keeping the method simple and easy to implement. The required quantification of 0.01% (0.1 g L -1 ) for cadmium and 0.1% (1 g L -1 ) for lead, mercury, and chromium at this mass and volume is comfortably within the capabilities of the icap 7200 ICP-OES Duo. A mass of approximately 0.1 g of each was weighed into individual PTFE microwave digestion vessels and 10 ml of analytical grade nitric acid (37% mass by volume) was added to digest the. A trace or catalytic quantity (0.1 ml or less) of hydrogen peroxide may also be added to aid complete digestion and the vessels were then sealed and the s digested in the microwave as detailed in Figure 1. Temperature ( C) 180 160 140 120 100 80 60 40 20 Microwave temperature program 0 0 20 40 60 80 100 120 Time (min) Figure 1. Microwave digestion program. Method development A LabBook was set up using the Thermo Scientific Qtegra Intelligent Scientific Data Solution (ISDS) Software for the analysis of the elements covered by WEEE-RoHS regulation. Method parameters and introduction components are shown in Table 1. Table 1. Method parameters. Parameter Pump Tubing (Standard Pump) Pump Speed Nebulizer Nebulizer Gas Flow Spray Chamber Setting Sample Tygon orange/white Drain Tygon white/white 45 rpm Glass concentric 0.15 MPa Glass cyclonic Auxiliary Gas Flow 1.0 L min -1 Coolant Gas Flow 12 L min -1 Center Tube RF Power Plasma View Exposure Time 2 mm 1150 W Axial UV 10 s, Vis 10 s The s and their spikes were analyzed in a single run after calibration with a suitable set of standards and the standard concentrations are shown in Table 2 for reference. The standards were prepared by diluting traceable 1000 aqueous single element standards and the same high-purity nitric acid that was used for the dissolution was added to the standards to produce a final set of standards containing 5% nitric acid. Table 2. Standard concentrations ( ). Element Blank Standard 1 Standard 2 Standard 3 Cd 0 0.1 0.5 1 Cr 0 0.1 0.5 1 Hg 0 0.1 0.5 1 Pb 0 0.1 0.5 1 The digested s were quantitatively transferred into 100 ml flasks with deionized water and made up to 100 ml in preparation for the analysis by ICP-OES. An aliquot was also decanted from each of the EC680, EC681 and PVC2 s and spiked with the equivalent of 0.1 mg L -1 of all the analytes. (Note: the PVC1 was used solely to provide a matrix blank for the determination of the method detection limit.)

Results The analytical results of the polymer analysis are shown below in Table 3 and the recoveries are further highlighted in Figure 2. The s and the spikes were analyzed in the same run. The majority of the spike recoveries read back within a range of ±10% relative to the accepted values and the spiked concentrations. Two of the analyte results for EC681 (Cr and Hg) were within 15% relative to the accepted values and may be attributed to minor inhomogeneities within the digested chips. It should also be noted that although the low-level analyte results for EC681 show variation from the accepted value, the spike results are well within expectations. Method detection limits (MDL) shown in Table 3 were established for the analytes by analyzing the PVC1 polymer using ten replicates measurements and multiplying the standard deviation of the results by 3 to give a 3σ detection limit. PVC1 has minimal or no detectable concentrations of three out of four restricted elements in a typical polymer matrix and is therefore considered to be appropriate for the purpose of MDL calculation. The MDL figures show the low detection limits which typify the powerful performance of the icap 7200 ICP-OES Duo. Table 3. Results of the analysis of polymer s. Cd 214.438 nm solid EC680 144.6 140.8 102.7 102.6 EC681 22.0 21.7 101.4 103.0 PVC2 37.0 35.0 105.6 97.6 PVC1 MDL 0.09 - - - Cr 267.716 nm solid Hg 194.227 nm solid EC680 110.0 114.6 96.0 101.8 EC681 15.3 17.7 86.3 99.4 PVC2 1.7 NA NA 97.1 PVC1 MDL 0.65 - - - EC680 23.3 25.3 92.1 93.5 EC681 4.8 4.5 105.8 114.9 PVC2 <DL NA NA 90.6 PVC1 MDL 1.19 - - - The results and spike recoveries shown in Table 3 demonstrate the sensitivity and accuracy of the icap 7200 ICP-OES Duo for WEEE-RoHS applications. Pb 220.353 nm solid EC680 106.1 107.6 98.6 100.8 EC681 12.8 13.8 92.5 100.4 PVC2 93.5 89.0 105.0 95.5 PVC1 MDL 1.21 - - - Polymer % and Recovery 120 % Recovery 115 110 105 100 95 90 85 Cd 214.438nm Cr 267.716nm Hg 194.227nm Pb 220.353nm Cd 214.438nm Cr 267.716nm Hg 194.227nm Pb 220.353nm 80 EC680 EC681 PVC2 Sample name Figure 2. Graphical representation of spikes.

Conclusion The analysis of polymer s to verify compliance with WEEE-RoHS legislation is highly efficient and cost effective with the Thermo Scientific icap 7200 ICP-OES Duo. The Thermo Scientific Qtegra ISDS Software allows even for unexperienced users to set up a method in conformity with specific regulations, easy analysis and post-acquisition data manipulation. References 1. EU WEEE Directive http://eur-lex.europa.eu/legal-content/en/txt/ PDF/?uri=CELEX:32012L0019&from=EN 2. EU RoHS Directive http://eur-lex.europa.eu/legal-content/en/txt/ PDF/?uri=CELEX:32011L0065&from=EN and http://eur-lex.europa.eu/legal-content/ EN/TXT/PDF/?uri=CELEX:32015L0863&from=EN 3. EU webpage for links to all WEEE-RoHS legislature http://ec.europa.eu/environment/ waste/weee/legis_en.htm 4. Definition and IEC Standards from the IEC website http://www.iec.ch/ 5. EU webpage on changes to WEEE-RoHS initiatives http://ec.europa.eu/environment/ waste/weee/index_en.htm Find out more at thermofisher.com/icp-oes For Research Use Only. Not for use in diagnostic procedures. 2017 Thermo Fisher Scientific Inc. All rights reserved. Tygon is a trademark of Saint-Gobain Corporation. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. AN43145-EN 0717